DH Instruments MOLSTIC-S SUPPLY ONLY PLATFORM 1-4 IN. AND 1-2 IN, molstic-S, 401876, 401877 Installation Sheet

The 1/4 in. and 1/2 in. supply only molstic-S platforms, P/N 401876 and 401877, provide inlet gas pressure regulation to either the Device Under Test (DUT) or the downstream molstic-S mounting systems. They are used to control the pressure supplied to an upstream DUT, or to adjust the flow rate through the system while calibrating a downstream DUT.
High pressure gases are potentially hazardous.
Energy stored in these gases can be released unexpectedly and with extreme force. High pressure systems should be assembled and operated only by personnel who have been instructed in proper safety practices.
molstic-S™ Supply Only Platform
1/4 in. and 1/2 in.
P/Ns 401876 and 401877
Installation Sheet
Figure 2 1/2 in. Supply Only molstic-S
1. Quick connect adaptor (includes stem with 1/2 in. SWG - not shown)
2. Pressure reducing regulator
3. Flow metering/shut-off valve
4. Outlet adaptor, 1/2 in. tube x 1/2 in. male VCR
OPERATION OF SUPPLY ONLY MOLSTIC-S
Set regulator outlet pressure to minimum.
Rotate the knob of the pressure reducing regulator
in the counterclockwise direction until the spring force of the diaphragm is relieved. Insure that the shut-off valve is in the closed position.
Be sure that the molstic-S supply only
platform shut-off valve is closed prior to connecting the quick connect stem to the
Figure 1 1/4 in. Supply Only molstic-S
1. Quick connect adaptor (includes stem with 1/4 in. SWG - not shown)
2. Pressure reducing regulator
3. Flow metering/shut-off valve
4. Outlet adaptor, 1/4 in. tube x 1/4 in. male VCR
Connect the gas supply.
Connect the gas supply to the quick connect stem
(either 1/4 in. or 1/2 in. tube). The gas supply pressure should not exceed 1.7 MPa (250 psig). Insert the stem into the quick connect body.
Connect the outlet of the supply only platform to
either the DUT or the inlet of the molstic-S.
quick connect body (located on the inlet side of the regulator).
The outlet pressure of the pressure reducing
regulator is controlled by adjusting the position of the control knob. Rotating the knob clockwise raises the outlet pressure. Counterclockwise rotation, coupled with venting of the downstream side of the regulator plumbing, lowers the outlet pressure.
Make final adjustments of control pressure in the
direction of increasing pressure in order to obtain the most accurate and stable set point.
INSTALLATION OF THE PRESSURE REDUCING REGULATOR ONTO MOLSTIC-S PLATFORM
The pressure reducing regulator can be removed from the supply only molstic-S platform and mounted onto the molstic-S platform. Use the following procedure to install the pressure reducing regulator onto the molstic-S platform:
Disconnect the gas supply from the supply only
platform and vent any residual pressure.
Using a pair of properly sized open-end wrenches
loosen the VCR side of the VCR double male union between the regulator and the flow control/shut-off valve. Leave the double male union assembled to the regulator.
Using the proper size hex wrench, remove and
retain the socket head cap screws, washers, and nut plate that mount the regulator and its bracket to the supply only molstic-S platform, along with the VCR double male union. Do not disassemble the bracket from the regulator.
Using a pair of properly sized wrenches remove
the VCR x tube adaptor on the inlet end of the molstic-S, remove the VCR gasket and discard. Retain the adaptor for future use.
Place a new VCR gasket (P/N 102183 for 1/4 in. or
102923 for 1/2 in.) securely against the sealing surface inside the female VCR swivel nut of the molstic-S inlet fitting.
Place the regulator with its bracket on top of the
molstic-S platform.
®
swivel nut on the downstream
Align the regulator’s male VCR fitting with the
mating female swivel nut and thread the nut onto the fitting.
Tighten the female nut finger-tight.
Hold male body stationary with a backup wrench.
Tighten the female nut 1/8 turn past finger-tight.
Over-tightening will damage the sealing
beads and possibly cause system leakage. Do not re-use VCR gaskets.
Place the flat side of the 60 mm nut plate under
the molstic-S platform. Align it with the holes in the platform and the regulator bracket. Hold it in place.
Thread the socket head cap screws through the
regulator bracket and tighten securely into the nut plate.
ACCESSORIES INCLUDED WITH THE SUPPLY ONLY MOLSTIC-S SHIPMENT
The following accessories are included in the supply only molstic-S shipment:
P/N 401941
1/4 in. Supply Only molstic-S Accessory Kit
DESCRIPTION QTY PART NO.
Gasket, VCR, 1/4 in. 2 102183
O-ring, VCR face seal, 1/4 in. 2 102070
Tubing, 1/4 in. PFA, 50 cm 1 101450-Z
Quick connect stem, 1/4 in. 1 102458
P/N 401949
1/2 in. Supply Only molstic-S Accessory Kit
DESCRIPTION QTY PART NO.
Gasket, VCR, 1/2 in. 2 102923
O-ring, VCR face seal, 1/2 in. 2 102912
Tubing, 1/2 in. PFA, 50 cm 1 103227-Z
Quick connect stem, 1/2 in. 1 103235
molbloc and molbloc-S are trademarks, registered and otherwise, of DH Instruments, Inc.
Viton is a registered trademarks of DuPont deNemours Company.
Document 560049a 030108
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
VCR is a registered trademark of the Swagelok Company.
Tel 602.431.9100
Fax 602.431.9559 dhi@dhinstruments.com www.dhinstruments.com
Loading...