DHI PPCMPC User Manual

Calibration Solutions
PPC™/MPC™ Vacuum Kit
for Pressure and Flow™
THE 110V VACUUM KIT INCLUDES
1 102381 [102034] Fast Vac Vacuum Pump
1 101322 [102382] 3-Way Solenoid Valve
1 102123 Adaptor AN4F Swivel x 1/8 in. NPTM
2 102033 Adaptor 1/4 in. Swage x 1/8 in. NPTM
1 100109 Adaptor 1/4 in. Swage x 1/4 in. NPTM
1 101450-Z 1/4 in. O.D. Tubing (200cm)
The [ ] above denotes the part number required for a 220V Vacuum Kit
100109 or
P/N 400922 (110V)
P/N 401160 (220V)
The PPC/MPC Vacuum Kit should be installed as illustrated above. Fill the vacuum pump with oil before using. Read Fast Vac instruction sheet for additional information.
A three-way solenoid valve has been installed to protect the vacuum pump when turned off. When decreasing pressure in a PPC the test gas is exhausted through the reference port. This ensures that the gas will not exhaust into the pump and possibly cause damage. The valve is electrically connected to the vacuum pump power switch and will be actuated when the pump is turned on.
When the vacuum pump is turned off, the reference or vacuum port of the PPC (or MPC) will be connected to atmosphere bypassing the vacuum pump.
When the vacuum pump is turned on, the reference port of the PPC (or the vacuum port of the MPC) will be connected to the vacuum pump enabling pressure below atmosphere to be generated.
PPC, RPM and MPC are trademarks, registered and otherwise, of DH Instruments, Inc.
Document 560004b 040317
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
Tel 602.431.9100
Fax 602.431.9559 dhi@dhinstruments.com www.dhinstruments.com