DHI PK-PPC-BG-DVU User Manual

Page 1
INTERCONNECTIONS KIT, P/N 401699, INCLUDES
PK-PPC-BG-DVU Dual Volume Unit
and Interconnections Kit
P/N 401699
Installation Sheet
QTY PART NO
1 401697 N/A Dual Volume Unit (DVU)
1 101450-Z 1 Tubing, PFA, 1/4 in. x 160 cm
2 103279 2
4 101477 3
2 101985 4 Union, 4P
2 101487 5 Tee, 1/4 in. tube
2 101482 6
FIGURE
CALLOUT DESCRIPTION
Adaptor, 1/4 in. tube x O-seal 1/8 in. NPT M
Adaptor, 1/4 in. tube x 1/4 in. NPT M
Adaptor, 1/4 in. tube x 1/4 in. NPT F
TO INSTALL THE DUAL VOLUME UNIT (DVU)
Identify a location for the DVU™ that minimizes
the length of tubing between the DVU connections and the PPC BG0002™ TEST(+) and TEST(-) ports. The PPC BG0002 and the DVU are designed to stack one on top of the other. The DVU may be rack mounted using the rack mount kit (P/N 400440). Recommended configurations are for the PPC and DVU to be stacked directly one on top of the other or with their rear panels face to face.
TO CONNECT THE DVU AND PPC BG0002
You will be making and connecting two identical
assemblies, one for the TEST(+) port and one for the TEST(-) port.
Install a union (4) into a 1/4 in. tube x 1/4 in. NPT
M adaptor (3) (do this twice). Screw the 1/4 in. NPT M end of the adaptors into the two ports of the DVU using Teflon
®
tape as a sealer.
Figure 1 Interconnections Kit
Install a 1/4 in. tube tee (5) onto each of the
union/adaptor assemblies that were just installed into the DVU ports in Step
. Orient the tee considering the connection to the PPC BG0002 (schematic above is for PPC stacked on top of the DVU).
Install the appropriate adaptors onto the PPC BG0002
TEST(+) and TEST(-) ports:
For PPC3™ Install the 1/4 in. tube x O-seal
1/8 in. NPT M (2).
For PPC2+™ Install the 1/4 in. tube x 1/4 in.
NPT M adaptor (3) using Teflon tape as a sealer.
Cut lengths of PFA tubing as needed to connect
the 1/4 in. tube tee to the adaptors on the PPC TEST ports. Take care that the cut is at a right angle to the tube and only cut the length needed.
Optional for a connection “For DUT” of 1/4 in. NPT M
rather than 1/4 in. tube:
• Cut a small length of PFA tubing to use as
a union between the 1/4 in. tube tee (5) and the 1/4 in. tube x 1/4 in. NPT F adaptor (6) (see schematic above).
DVU, PPC, PPC2+, PPC3, and PPC BG0002 are trademarks, registered and otherwise, of DH Instruments, Inc.
Document 560041a 030211
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
®
Teflon
is a registered trademark of E.I. du Pont de Nemours and Company.
Tel 602.431.9100
Fax 602.431.9559 dhi@dhinstruments.com www.dhinstruments.com
Loading...