DHI PK-7000-MPC1-D User Manual

Page 1
Calibration Solutions for Pressure and Flow™
Interconnections Kit
INTERCONNECTIONS KIT, P/N 401645, INCLUDES
PK-7000-MPC1-D
Installation Instructions
100319 Adaptor, 1/8 in. tube x 1/8 in. NPT M 100317 Adaptor, 1/8 in. tube x 1/4 in. NPT F 401459 Quick Connect and Stand Assy
102184 Reducer, 4P x 2P 101988-Z Tubin g, SS, . 125 in. OD x 300 cm 102116 Quick Connect, 1/8 in. NPT F
PARTS ON TEST PORT OR IN ACCESSORIES OF PG7000
101200 Collar
100272 Gland
102116
100319
101988-Z
TO P 7000G
101200
TO RP M3
100272
400518
TO MP C1-D REF HI or REF LO
100319
400518 Adaptor , 1/8 in. tube x DH 200 M
102184
401459
101988-Z
TO MPC1-D TEST HI or TEST LO
INCLUDED WITH P 7000G
100317
100319
TO PNEUMATICALLY CONNECT THE PG7000S OR RPMS TO THE MPC1-D
MPC1-D is intended for controlling pressures in a high line differential pressure system where there is a “high” side and
a “low”. The instructions below should be repeated, once for the high side and once for the low side. See the MPC1-D Operation and Maintenance manual for a schematic of a typical system layout.
If you are using RPMs (not PG7000s) go directly to Step pb.
n Remove gland, collar and plastic plug from the PG7000 TEST port. o Install gland and collar onto 1/8 in. tube x DH200 male adaptor supplied in PG7000 accessories.
Note that the collar threads left handed. Assure it is properly attached. The orange plastic plug is not used.
pa PG7000 - Screw the adaptor (as assembled) into the PG7000 TEST port. Torque to 15 Nm (Newton
meter).
pb RPM3 - Screw 1/8 in. tube x 1/8 in. NPT male adaptor into the RPM3 TEST port using Teflon tape as
a seal.
Page 2
PK-7000-MPC1-D Interconnections Kit
q Screw 1/8 in. tube x 1/8 in. NPT male adaptor into MPC1 TEST port using Teflon tape as a seal. ra To use quick connect stand for TEST connection: Install reducing union into quick connecting head.
Mount quick connect stand on bench using mounting holes provided.
rb To use 1/4 in. NPT female adaptor for TEST connection: Install 1/8 in. tube x 1/4 in. NPT female
adaptor onto ends of tube from MPC1-D TEST port (Step u below).
s Screw 1/8 in. tube x 1/8 in. NPT male adaptor into MPC1 REF port using Teflon tape as a seal. t Connect PG7000 or RPM3 to MPC1-D REF port by cutting a piece of 1/8 in. SS tubing to desired
length and swage together.
u Connect quick connect stand or 1/8 in. tube x 1/4 in. NPT female adaptor to MPC1-D TEST port by
cutting a piece of 1/8 in. SS tubing to desired length and swage together.
1/8 in. SS tubing can be bent by hand to desired shape but use caution not to kink the tube.
PG7000, RPM, PPC and MPC are trademarks, registered and otherwise, of DH Instruments, Inc. Document 560039 010509
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
Tel 602.431.9100
Fax 602.431.9559 dhi@dhinstruments.com www.dhinstruments.com