LT-2026
15
EN
Symptom
Action (for symptoms listed at left, follow the
steps listed below)
Blower
does not
start when
powered
up
1. Remove the cover. See Service Instructions.
2. Search for disconnected wire harnesses
and reconnect, if needed.
3. Determine if fuses are open.
Without AC power to the unit, measure
the resistance across the F1 fuse with an
ohmmeter. If 0.0 to 0.5 ohms of resistance
are not present across the fuse, the fuse is
open and the power supply board should
be replaced. See Service Instructions.
Without AC power to the unit, measure
the resistances across the F2 and F3 fuses
with an ohmmeter. If 0.0 to 0.5 ohms of
resistance are not present across either
fuse, the fuse is open and the PC control
board should be replaced. See Service
Instruc tions.
4. If all fuses are closed, install a blower
known to operate properly, see Service
Instructions, and power up the unit.
If the new blower operates properly,
replace the old blower.
If the new blower does not operate prop-
erly, replace the control board. See Service
Instruc tions.
5. Replace the cover. See Service Instructions.
E03, E04,
or E07
errors
display
OR display
is blank
E01
error code
displays
Replace the control board
Blower is
loud
1. Remove the cover and control board. See
Service Instructions.
2. Inspect the assembly of the foam, blower,
and silicone isolator. See Blower Replacement instructions for correct placement.
3. Inspect the blower for visible space between the blower halves and press halves
back together, if needed.
4. If the assembly is not correct, reassemble
the components.
5. If assembly is correct, replace the blower.
See Service Instructions.
6. Replace the control board and cover. See
Service Instructions.
8. Troubleshooting
Symptom
Action (for symptoms listed at left, follow the
steps listed below)
Pressure
out of
tolerance
(Pressure ±
1.0 cmH
2
O
of setting)
1. Remove the cover. See Service Instructions.
2. Search for disconnected, occluded, or
kinked tubing on the pressure sensor or
chassis. Correct any issues.
3. Search for damage to any foam pieces
and/or tubing and replace the component, if damaged.
4. Remove the control board. See Service
Instructions.
5. Search for signs of water damage to the
blower assembly, pressure sensor and
replace the component, if damaged.
6. Inspect the blower. If a space is visible
between the top and bottom sections,
press the halves together to eliminate the
space.
7. Repeat the pressure test.
8. If the pressure remains out of tolerance,
re-calibrate the unit. See Service Instructions.
9. If the pressure remains out of tolerance,
replace the control board. See Service
Instruc tions.
10. Replace cover. See Service Instructions.
Calibration
failure–
Error code
E95
Display on
terminal
program
during
calibration.
1. Remove the cover. See Service Instructions.
2. Search for damage, and/or loose connections, on the air intake fi lter, intake
foam components, and all internal tubing
and replace or reconnect components, if
needed. Search for disconnected, occluded, or kinked tubing on the pressure sensor and chassis and reconnect or replace
tubing, if needed.
3. Remove the control board. See Service
Instruc tions.
4. Search for signs of water damage to the
blower assembly, pressure sensor and
replace, if damaged.
5. Inspect the blower. If a space is visible
between the top and bottom sections,
press the halves to eliminate the space.
6. Repeat Manual Calibration Procedure.
7. If the error codes are not resolved, replace
the control board. See Service Instructions.
8. Replace the cover. See Service Instructions.