DeVilbiss SleepCube nCPAP, SleepCube DV54-P, SleepCube DV53, SleepCube DV54, SleepCube DV53 PAP Service Manual

SleepCube™ nCPAP Device
Model DV53/DV54/DV54-P
Service Manual
2
EN
Table of Contents
1. General Information 3
A. Safeguards 3 B. Initial Inspection 3 C. Travel 4 D. DC Power 4 E. Setting Pressures and Features 4 F. Images: DeVilbiss DV54/DV53 PAP 4
2. Description of Normal Operation 5
3. Cleaning and Disinfecting 6
A. Routine Cleaning--Patient 6 B. Disinfecting 6
4. Maintenance 9
A. Initial Inspection—Equipment Provider 9 B. Routine Maintenance—Patient 9 C. Required 3-year Maintenance—
Equipment Provider 9
5. Testing 10
A. Pressure Accuracy Test 10 B. Keypad Test 10 C. Auto-ON / Auto-OFF Test 10 D. Backlight Test for LCD and Keypad 11
6. Alerts and Device Faults 12
A. Alerts Visible to Patients 12 B. Device Faults Visible to Patients 12 C. Reading and Clearing the Last Device Fault Code 12
7. Calibration 13
A. Manual Calibration 13 B. Calibration Errors 14 C. Auto-Calibration 14 D. Details on Tc Serial Command 14
8. Troubleshooting 15
9. Service Instructions 16
A. Cover Removal 16 B. Cover Replacement 16 C. Control PC Board Removal 16 D. Control PC Board Replacement 17 E. Power Supply Board Removal 18 F. Power Supply Board Replacement 18 G. Blower Removal 19
H. Blower Replacement 19 I. KeyPad & LCD Display Removal and Replacement 20
10. Unit Specifications 21
11. General Information- DV5HH 22
A. Safeguards 22 B. Travel 22 C. DC Power 22 D. Product Description 22
12. Description of Normal Operation (DV5HH) 23
13. Maintenance and Testing (DV5HH) 24
A. Cleaning and Disinfection 24 B. Maintenance 24 C. Testing 24
14. Troubleshooting (DV5HH) 25
15. Service Instructions (DV5HH) 26
A. Removing the DV5HH base cover 26 B. Replacing the DV5HH base cover 26 C. Removing and replacing silicone manifold 27 D. Removing and replacing the latch 27
16. Unit Specifications (DV5HH) 28
A. Ordering Non-Warranty Replacement Parts 29 B. Ordering Warranty Replacement Parts 29 C. Returning Warranty Defective Parts 29 D. Placing orders 29
17. Ordering and Returning Parts 29
18. Parts List 30
DV54/DV53 31 DV5HH 31
19. Warranties 31
3
EN
1. General Information
A. Safeguards
When using electrical products, basic safety precautions should always be followed. Read all instructions before using this device. Important information is highlighted by the fol­lowing terms.
DANGER: Urgent safety information for hazards
that will cause serious injury or death.
WARNING: Important safety information for hazards
that might cause serious injury.
CAUTION: Information for preventing damage to the
product.
NOTE: Information to which you should pay
special attention.
PLEASE READ ALL INSTRUCTIONS
BEFORE USING THIS DEVICE.
DANGER!
ELECTRIC SHOCK HAZARD – Do not use while bathing.
ELECTRIC SHOCK HAZARD – Do not immerse this device into water or any other liquid.
ELECTRIC SHOCK HAZARD – Do not attempt to open or re­move the enclosure. There are no user-serviceable internal components. If service is required, return the product to your home care provider. Opening or tampering with the product will void the warranty.
WARNING!
• The DeVilbiss IntelliPAP™ and SleepCube™ devices should be used only with masks recommended by DeVilbiss, your physician or respiratory therapist.
• To avoid rebreathing of exhaled air, do not use a CPAP mask unless the device is turned on and providing a supply of air. Venting in the mask should never be blocked. When the device is turned on and providing a fresh supply of air, exhaled air is fl ushed out of the mask vent. However, when the device is not operating, exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suf­focation. This warning applies to most CPAP devices.
• The DeVilbiss IntelliPAP and SleepCube devices are not a life support devices and may stop operating with certain device faults or with a power failure. It is intended to be used on spontaneously breathing individuals weighing 66 lbs/30 Kg or greater.
• To avoid electric shock, always unplug power cord from wall outlet power source when performing cleaning.
• Use only accessories recommended by DeVilbiss.
CAUTION– Product Damage: The circular data port connector
located on the back of the IntelliPAP and the SleepCube is used to attach accessories to the device. The connector must only be used with accessories approved for use by DeVilbiss. Do not attempt to attach any other device to this connector as it may damage the CPAP or the acces­sory device.
CAUTION– Product Damage: Never rinse or place the device in
water. Never allow liquids to get into or around any of the ports, switches or air fi lter; doing so will result in device damage. If this occurs, discontinue use and remove the power cord from the power source. Allow the device to completely dry before use.
CAUTION– Product Damage: Do not place the IntelliPAP or
SleepCube where it can be bumped onto the fl oor or where the power cord may create a trip hazard.
CAUTION– Product Damage: Only the DeVilbiss DV5 series
Heated Humidifi er system is recommended for use with the IntelliPAP and SleepCube. Other humidifi er systems may prevent the device from detecting snoring and may cause inappropriate pressure levels in the mask.
CAUTION– Product Use: Oxygen is a prescription gas and
should only be administered under the supervision of a physician.
B. Initial Inspection
DeVilbiss recommends equipment inspection upon delivery
Power up the DV54/DV53 PAP using AC power. Test pressure accuracy using an outlet cap (DV54/
DV53D-620), a calibrated pressure gauge and the proce­dures listed under Pressure Accuracy Test.
Test the keys on the keypad using the procedures listed
under Keypad Test.
4
EN
C. Travel
The DeVilbiss DV54/DV53 PAP:
• Automatically adjusts for altitudes between sea level and 9000 ft (2750 m)
• Automatically accepts line voltages of 100-240 V, 50/60 Hz
• Needs power cord appropriate for area o USA DV54/ DV53D-606 o Europe, except UK DV54/DV53D-607 o UK DV54/DV53D-608 o Australia DV54/DV53D-609 o Set of 3 (UK, Europe, Australia) DV54/DV53D-611
D. DC Power
The DeVilbiss DV54/DV53 PAP:
• Automatically accepts 12V DC power
• Operates on DC power only if AC power is not present
• Operates on AC power if both DC and AC power are pres­ent
• Optional 12V, 60 amp hour, deep cycle marine battery
• Optional DC to AC inverter: minimum 200 watts @ 100/120 VAC
minimum 400 watts @ 220 VAC
• Appropriate DC power cord
o DV54/DV53D-619 DC accessory cable for DC plug-in
adapters
o DV54/DV53D-696 DC battery clamp-on adaptor for
stand-alone battery
E. Setting Pressures and Features
Use the following steps to enter Clinical/SetUp mode:
1. Apply AC power to the unit
2. Verify that LCD displays OFF
3. Press and hold the Down Arrow key and the Delay key.
4. While holding two keys, press the ON/OFF key.
5. The blower will begin operating and the LCD will display ‘Clinical Menu’.
6. Use the Left and Right Arrow keys to scroll through the options
7. Use the Up and Down Arrow keys to select the option’s value.
8. Press the ON/OFF key at any time to exit Clinical Mode.
F. Images: DeVilbiss DV54/DV53 PAP
Figures A and B: DV54/DV53 Device Back and Bottom
1. Air supply port on back of device
2. Air supply port on bottom of device (for optional humidifi er)
3. Air supply port plug
4. Heater power connector (for optional humidifi er)
5. AC power connector
6. DC power connector
7. Data por t
8. Air inlet fi lter opening
Figure C: Keypad
1. ON/OFF
2. Previous item
3. Next item
4. Delay
5. Decrease value
6. Increase value
7. Heater power LED (for optional humidifi er)
C
5
1
2
4
6
3
7
B
2
4
A
1
3
56 7
8
5
EN
2. Description of Normal Operation
An AC line cord (100-240 VAC, 50/60 Hz) or a DC line cord (12 VDC) supplies electrical power to the DV54/DV53 PAP. The PAP converts the AC input voltage to DC voltage by means of an internal switch mode power supply and uses the DC voltage to power the internal electronics of the unit, such as microcontroller, motor control circuitry, blower, LCD display, etc. NOTE: The heated humidifi er will not operate when the PAP uses external DC power.
The PAP produces positive pressure by spinning a reverse­curved impeller with a brushless DC motor. Room air is drawn through a fi lter into the blower, pressurized in the blower, optionally passed through a heated humidifi er chamber, and then discharged through a 22mm-ID, smooth-bore tube. Pressure regulation is achieved using measured pressure as feedback.
The DV54/DV53 PAP senses patient breathing by monitoring changes in motor current: Breathing produces variations in airfl ow; the airfl ow variations produce changes to the blower load and result in variations to the motor current. A pressure transducer triggers the auto-ON / auto-OFF functionality and mask-OFF alerts.
6
EN
A. Routine Cleaning–Patient
• Unplug the DV54/DV53 PAP and wipe enclosure with a clean, damp cloth every few days to keep dust free. Allow the device to dry completely before returning to power source.
• Check the air-inlet fi lter every 10 days. Wash the dark outer foam fi lter in a solution of warm water and mild detergent. Rinse with water and allow the fi lter to dry completely before returning to the device.
• Check the optional fi ne particle fi lter every 10 days and replace if dirty or damaged.
• Clean the air supply tubing every day. Remove the tubing from the device and mask and wash the inside of the tub­ing in a solution of warm water and mild detergent. Rinse with water and allow to air dry before replacing device and mask.
• Clean the mask and headgear according to the manufac­turer’s instructions.
B. Disinfecting
CAUTION– Always work in an ESD (electro static discharge)
safe environment when assembling or disassembling electronic components.
NOTE– DeVilbiss does not require disinfection as part of unit
maintenance. If disinfection is desired, the following pro­cedure is recommended.
NOTE– Parts needed to perform the disinfection procedure
are NOT covered under the DV54/DV53 PAP warranty as listed in this manual.
NOTE– Follow manufacturer’s instructions for cleaning and
disinfection solutions.
NOTE– Disinfection requires the following parts sold in Dis-
infection Kit, DV54/DV53D-682.
• Blower assembly, balanced
• Blower isolator-silicone
• Grommet-silicone
• 1/4 inch blower wrap foam
• 3 1/2 inch silicone tubing
• Air-inlet fi lter
• 5 1/2 inch silicone tubing— discard, not needed for this model
• Tubing/Humidifi er air supply support plug
• 1/2 inch blower mount foam-round
• 1/2 inch top blower foam
• 1/8 inch inside bottom cover foam
• 5 inch silicone tubing
• Fine particle fi lter (optional)
3. Cleaning and Disinfecting
Disinfection Procedure for the DV Series PAP
1. Make certain that the device is not connected to a power source.
2. Dispose of the mask, headgear, air supply tubing, gray air inlet fi lter and white fi ne particle fi lter, if applicable.
3. Place the unit, top side down, onto a clean work surface. Remove the four (4) screws from the bottom cover and remove the bottom cover from the unit. (Fig. 1)
4. Remove the black 1/8 inch foam from inside the bottom cover assembly and discard. This foam may stick to the chassis of the main unit. (Fig. 2)
5. Remove the black 1/2 inch top blower foam from the chassis assembly and discard.
6. Turn the unit top side up and remove the top cover.
7. Using a T-15 driver, remove the screw holding the PC board to the chassis. (Fig. 3)
8. Disconnect the 8-conductor blower wire harness from the underside of the PC board. (Fig. 4)
9. Disconnect the two (2) pieces of tubing and discard. (Fig. 5)
NOTE– Pull the tubing straight off the connectors to prevent
damaging the chassis.
10. With the ports side of the chassis toward you, tilt the PC board and its respective wire harnesses upward, and remove the wire connections. Lift the PC board slowly so that you can assist any wires that may become hung up on the chassis or power supply. (Fig. 6)
7
EN
41
3
2 5
6
8
EN
7
8
9
10
11
11. Remove the grommet that seals the blower’s wire harness. (Fig. 7)
12. Remove the heater wire harness connector from the bot­tom of the chassis by sliding the connector outward while pulling the top retaining tab upward slightly. (Fig. 8)
13. Remove the 1/4” blower wrap foam, the blower, and the silicone isolator from the underside of the chassis.
14. Carefully spray, wipe or soak the air intake, fl ow path, and blower chassis inside and outside using a 2% glutaralde­hyde solution.
15. Wipe the power cord and the outside of both halves of the PAP case with the same solution.
16. Let all components fully dry
17. Reassemble the unit using parts from the Disinfection Kit—DV54/DV53D-682.
NOTES:
• Route the silicone tubing correctly, as shown, and ensure that there are no kinks. (Fig. 9)
• Ensure that the silicone isolator is fully seated in the chas­sis and the blower.
• Ensure that the grommet is fully seated around the blower wire harness.
• Route the wire harnesses properly, using the provided strain reliefs on the chassis. (Fig. 10 & 11)
• Secure the power supply board within the mount clips on the chassis before replacing the unit covers. (Fig. 10)
9
EN
4. Maintenance
A. Initial Inspection—Equipment Provider
DeVilbiss recommends equipment inspection upon delivery
Power up the DV54/DV53 PAP using AC power. Test pressure accuracy using an outlet cap (DV54/
DV53D-620), a calibrated pressure gauge and the proce­dures listed under Pressure Accuracy Test.
Test the keys on the keypad using the procedures listed
under Keypad Test.
B. Routine Maintenance—Patient
Wipe case with damp cloth every few days Check air-inlet fi lter every 10 days and clean as needed Clean air supply tubing daily Clean mask and headgear per manufacturer’s instructions.
C. Required three years Maintenance—
Equipment Provider
DeVilbiss requires the following maintenance from the equip­ment provider:
Test pressure accuracy using procedures listed under Pres-
sure Accuracy Test. If the device is out of calibration, follow corrective procedures as listed.
Instruct patient on fi lter maintenance described in Instruc-
tion Guide: Standard Filter o Inspect fi lter every ten days o Clean as needed o Change every 6 months
Optional Fine Particle Filter
o Inspect fi lter every ten days
o Replace if dirty
o Change every 30 days
10
EN
The following test procedures verify correct operation of the DV54/DV53 PAP and should be performed on all repaired units.
NOTE– Verify that the DV54/DV53 PAP contains the latest
fi rmware release before conducting maintenance or repairs. Go to http://www.devilbisshealthcare.com/Prod­ucts to download the latest fi rmware update. Download and installation instructions are provided on the web site.
A. Pressure Accuracy Test
Additional equipment required: outlet cap (DV54/ DV53D-620), a calibrated pressure gauge (0-30 cmH2O, ac­curacy ±0.25 cmH2O)
1. Connect the pressure gauge to the unit outlet port.
2. Apply AC power to the unit. If the blower is running, press the ON/OFF key on the keypad so that the blower stops and the LCD displays ‘OFF’.
3. While pressing both the Down Arrow and Delay keys, press the ON/OFF key to start the blower and enter clinical mode.
4. Press the Next Item key until Set Pressure is displayed. Use the Up and Down Arrow keys to adjust the pressure to 12 cmH2O. Wait at least 30 seconds for the pressure to stabilize.
5. Verify that the displayed pressure matches the pressure measured on the gauge (pressure setting ±1.0 cmH2O). Accuracy at other pressure settings can be checked if needed.
6. If the pressure is out of calibration, turn the blower OFF for approximately 1 minute to allow auto-calibration to occur. Turn the blower back ON and allow 30 seconds for the pressure to stabilize. If the pressure is still out of calibra­tion, perform a full calibration using the Manual Calibra­tion Procedure.
7. If manual calibration does not improve pressure accuracy, there may be a problem with the fl ow path or control board. Refer to Troubleshooting.
5. Testing
B. Keypad Test
No additional equipment required.
1. Apply AC power to the unit. If the blower is running, press the ON/OFF key on the keypad so that the blower stops and the LCD displays ‘OFF’.
2. While pressing both the Down Arrow and Delay keys, press the ON/OFF key to start the blower and enter clinical mode.
3. Press the Next Item key until Delay Time is shown on the LCD display.
4. Note the Delay Time; then change it by pressing the Up and Down arrow keys. Use the arrows keys to return the Delay Time back to the original setting.
5. Use the Previous Item key to return to the Set Pressure screen.
6. If any of the keypad keys did not function correctly dur­ing this test, replace the PC control board using Service Instruc tions.
11
EN
C. Auto-ON / Auto-OFF Test
Additional equipment required: 6 ft x 22 mm smooth bore tubing.
1. Apply AC power to the unit. If the blower is running, press the ON/OFF key on the keypad so that the blower stops and the LCD displays ‘OFF’.
2. While pressing both the Down arrow and Delay keys, press the ON/OFF key to start the blower and enter clinical mode.
3. Press the Next Item key until Enable Menu is displayed, then press the Down Arrow key to enter the sub-menu.
4. Press the Next Item key until the Auto-ON setting is dis­played. Use the UP or DOWN arrow keys to set the value to Enabled.
5. Press the Next Item key until the Auto-OFF setting is dis­played. Use the UP or DOWN arrow keys to set the value to Enabled.
6. Press the ON/OFF key on the keypad so that the blower stops and OFF is displayed on the LCD.
7. Connect the 6 ft x 22 mm smooth bore tubing to the CPAP and block the airfl ow with your hand or a plug.
8. Turn the blower ON by pressing the ON/OFF key on the keypad.
9. Keep the airfl ow blocked for two minutes. Verify that the blower continues running and no mask alerts appear.
10. Unblock the smooth bore tubing to open the unit’s air­fl ow.
11. Verify that after approximately 10 seconds the LCD shows ‘Mask OFF’. If the mask alert did not appear, go to step 14.
12. After another 20 seconds the blower should turn OFF via the Auto-OFF feature. Wait approximately two minutes after the blower turns OFF. Verify that the blower does not turn back ON without breathing present.
13. Tap the palm of your hand on and off the end of the tub­ing to simulate breathing and verify that the blower turns ON via the Auto-ON feature.
14. If the Auto-OFF or Auto-ON features do not work correctly, perform a full calibration using the Manual Calibration Procedure.
D. Backlight Test for LCD and Keypad
No additional equipment required.
1. Apply AC power to the unit. If the blower is running, press the ON/OFF key on the keypad so that the blower stops and the LCD displays OFF.
2. Press the ON/OFF key and verify that the LCD and keypad backlights are lit.
3. Wait one minute and verify that the LCD and keypad back­lights reduce intensity.
4. Press any keypad key, except ON/OFF, and verify that the LCD and keypad backlights return to full intensity.
5. If the LCD or keypad backlights did not function correctly during this test, replace the LCD, keypad and/or PC control board.
12
EN
6. Alerts and Device Faults
A. Alerts Visible to Patients
Alert Description
Delay running XX minutes left
Comfort Delay is active
Mask off. Please check mask fi t.
High airfl ow during use (Poor mask fi t or mask removed)
Device Fault E0X. Call provider.
Device error (See next table.)
B. Device Faults Visible to Patients
Only device faults classifi ed as critical are visible to the patient on the LCD display. These faults can be corrected by a service technician.
Device
Fault
Description
E01
Read Setting Error. Stored prescription set­tings are corrupted. Unit is in fail-safe state (blower off) when this error displays. See Troubleshooting and Service Instructions to correct fault, or return unit for repair.
E03
Motor Error. Unit attempts to drive the mo­tor, but the blower is not spinning (zero motor speed). See Troubleshooting and Service Instructions to correct fault, or return unit for repair.
E04
Locked Rotor Error. Average motor speed is low, while the motor current is high, for several seconds. See Troubleshooting and Service Instructions to correct fault, or return unit for repair.
E07
Motor Fault Error. The motor control chip fault output goes active low for a fi xed period of time. See Troubleshooting and Service Instruc­tions to correct fault, or return unit for repair.
C. Reading and Clearing the Last Device
Fault Code
NOTE– The last device fault code is stored in EEPROM to assist
in fi eld service of the unit.
Additional equipment: IBM or Compatible PC with a free serial port, PAP to PC cable (DV54/DV53D-615), a terminal program (Microsoft Windows HyperTerminal or equivalent).
1. Connect the DV54/DV53 PAP to the PC serial port with PAP to PC cable.
2. Apply AC power to the unit. If the blower is running, press the ON/OFF key on the keypad so that the blower stops and the LCD displays ‘OFF’.
3. Use a terminal program (such as Windows HyperTerminal, or equivalent) with COM settings 9600 baud, no parity, 8 data bits, 1 stop bit, and Flow Control set to none.
4. When connection is established with the PAP, the terminal program will show: ‘RI;’ every 0.5 seconds. This display indicates that the PAP is waiting for a handshake.
5. Type ‘ri’ (all commands are case sensitive) and press Enter. The PAP should respond with the fi rmware version in the form “V0.XX mm/dd/yyyy.” Repeat this step if necessary to complete the handshake.
6. Request the last alert code and date by typing ‘El’ and pressing Enter. The return value will be in the form ‘Exx dd/mm/yyyy’. Exx is the alert code and dd/mm/yyyy is the date the error occurred.
7. Clear the last alert code by typing ‘Ec’ and pressing Enter. The PAP will respond with ‘OK’.
13
EN
A. Manual Calibration
DeVilbiss recommends calibrating the DV54/DV53 PAP if replacing the PC control board or when testing or trouble­shooting indicate calibration is required. The manual calibration procedure accesses a terminal program via the unit’s communication port to calibrate the internal pressure sensor and the motor current circuitry. Progress and confi r­mation messages display on the terminal screen throughout the procedure. If a failure occurs during calibration, the error displays as well. See Calibration Errors below.
Additional equipment: IBM or Compatible PC with a free serial port, PAP to PC cable (DV54/DV53D-615), a terminal program (Microsoft Windows HyperTerminal or equivalent) and a calibrated pressure gauge (0 to 30 cmH
2
O).
Please read all steps before beginning this process.
NOTE– Details on the Tc serial commands used in this proce-
dure are listed in Section D.
1. Use the PAP to PC cable to connect the DV54/DV53 PAP to a PC serial port and apply AC power to the PAP.
2. Use a terminal program (such as Windows HyperTerminal, or equivalent) with COM settings 9600 baud, no parity, 8 data bits, 1 stop bit, Flow Control set to none.
3. When connection is established with the PAP, the terminal program will show: ‘RI;’ every 0.5 seconds. This display indicates that the PAP is waiting for a handshake.
4. Type ‘ri’ (all DV54/DV53 commands are case sensitive) and press ENTER. The PAP will respond with the fi rmware version in the form ‘V0.XX dd/mm/yyyy’. Repeat this step if necessary to complete the handshake.
5. Start the calibration process by typing ‘Tc= s’ and pressing ENTER. The PAP will respond with ‘Calibration Started’. If the unit responds with ‘WF’, resend the ‘Tc=s’ command.
6. The PAP will begin sending messages and prompting for responses to complete the calibration. NOTE: If a calibra­tion failure is encountered at any stage during the pro­cedure, the PAP will respond with ‘Cal failed: EXX’, where XX is a two digit error code.
a. PAP sends ‘Waiting for blower to stop spin-
ning’.
b. PAP sends ‘Cal press offset’ when spinning is
stopped.
During this stage of the procedure the PAP stops the
blower and saves the pressure sensor and motor current offsets. After a short delay, the PAP increases the pressure to approximately 20 cmH
2
O.
c. PAP sends ‘Set pressure to 20 cmH2O using
Tw+, Tw-, or Tw=’ ‘Send Tc=y when ready
d. Connect the pressure gauge to the PAP outlet (the
7. Calibration
outlet should be in a deadhead or zero fl ow condition) and, using the computer keyboard (‘Tw+’ followed by ENTER increases the pressure and ‘Tw-’ followed by ENTER decreases the pressure), adjust the PAP pressure to 20 cmH2O (± 0.1 cmH2O), then type ‘Tc=y’ followed by ENTER.
e. PAP sends ‘Reading and saving pressure hi
cal’ after confi rming that the pressure is correct. During this stage of the procedure, the PAP calibrates the pressure sensor by reading and saving the high pressure calibration value.
NOTE– As an alternative to steps d and e, the current PAP
pressure gauge reading can be sent to the PAP from the terminal. Perform the following steps for this option: Use the ‘Tw+’, ‘ Tw-’, or ‘Tw=X’ (where X is a number between 0 and 1023 to adjust the pressure to a value between 15 and 30 cmH2O. If the pressure is already within this range no adjustment is needed. NOTE: A more accurate calibration will be obtained, if the value is close to 20 cmH2O.
7. Send the pressure to the PAP in the format ‘Tc=yxxxxx’, where xxxxx is the gauge pressure in cmH2O multiplied by 1000 (Example: 20 cmH2O is sent as ‘Tc=y20000’. )
8. After a short delay the PAP will respond with ‘cal suc- cessful’. Calibration is now complete. Please check the PAP pressure settings for accuracy.
14
EN
B. Calibration Errors
Code Description
E95
Calibrate Motor Stop Error. The blower did
not stop during device calibration. This error only occurs during the calibration procedure.
C. Auto-Calibration
The DV54/DV53 PAP automatically adjusts calibration to maintain accuracy over time and under varying operating conditions. Each time the blower is turned OFF, the PC control board reads and saves the offset voltages from the pressure transducer. The PAP uses these voltages to maintain calibra­tion accuracy as the unit is used over time.
D. Details on Tc Serial Command
Tc serial commands are used in the manual calibration pro­cedure listed in Section A. The ‘Tc’ command on the terminal program reads the unit’s current calibration values, starts the calibration procedure, and handles user feedback dur­ing the calibration process. Refer to the manual calibration procedure for details on the calibration process.
• Sending ‘Tc=s’ to the PAP starts the calibration process.
• Sending ‘Tc’ without a modifi er requests the current
calibration values. A comma-separated list of all cal values is returned.
Comma-separated response to ‘Tc’ with no modifi er:
llll,hhhh,mmmm<cr>
o llll is the pressure low cal (sensor offset). Convert to
voltage by dividing by 1024 and multiplying by 5.
o hhhh is the pressure high cal (sensor output at 20
cmH2O). Convert to VDC by dividing by 1024 and mul­tiplying by 5.
o mmmm is the motor current calibration (adc offset)
• Feedback is provided to the calibration task by sending
Tc= y’ when prompted.
15
EN
Symptom
Action (for symptoms listed at left, follow the
steps listed below)
Blower does not start when powered up
1. Remove the cover. See Service Instruc­tions.
2. Search for disconnected wire harnesses and reconnect, if needed.
3. Determine if fuses are open.
Without AC power to the unit, measure
the resistance across the F1 fuse with an ohmmeter. If 0.0 to 0.5 ohms of resistance are not present across the fuse, the fuse is open and the power supply board should be replaced. See Service Instructions.
Without AC power to the unit, measure
the resistances across the F2 and F3 fuses with an ohmmeter. If 0.0 to 0.5 ohms of resistance are not present across either fuse, the fuse is open and the PC control board should be replaced. See Service Instruc tions.
4. If all fuses are closed, install a blower known to operate properly, see Service Instructions, and power up the unit.
If the new blower operates properly,
replace the old blower.
If the new blower does not operate prop-
erly, replace the control board. See Service Instruc tions.
5. Replace the cover. See Service Instruc­tions.
E03, E04, or E07 errors display OR display is blank
E01 error code displays
Replace the control board
Blower is loud
1. Remove the cover and control board. See Service Instructions.
2. Inspect the assembly of the foam, blower, and silicone isolator. See Blower Replace­ment instructions for correct placement.
3. Inspect the blower for visible space be­tween the blower halves and press halves back together, if needed.
4. If the assembly is not correct, reassemble the components.
5. If assembly is correct, replace the blower. See Service Instructions.
6. Replace the control board and cover. See Service Instructions.
8. Troubleshooting
Symptom
Action (for symptoms listed at left, follow the
steps listed below)
Pressure out of tolerance
(Pressure ±
1.0 cmH
2
O
of setting)
1. Remove the cover. See Service Instruc­tions.
2. Search for disconnected, occluded, or kinked tubing on the pressure sensor or chassis. Correct any issues.
3. Search for damage to any foam pieces and/or tubing and replace the compo­nent, if damaged.
4. Remove the control board. See Service Instructions.
5. Search for signs of water damage to the blower assembly, pressure sensor and replace the component, if damaged.
6. Inspect the blower. If a space is visible between the top and bottom sections, press the halves together to eliminate the space.
7. Repeat the pressure test.
8. If the pressure remains out of tolerance, re-calibrate the unit. See Service Instruc­tions.
9. If the pressure remains out of tolerance, replace the control board. See Service Instruc tions.
10. Replace cover. See Service Instructions.
Calibration failure– Error code E95
Display on terminal program during calibration.
1. Remove the cover. See Service Instruc­tions.
2. Search for damage, and/or loose con­nections, on the air intake fi lter, intake foam components, and all internal tubing and replace or reconnect components, if needed. Search for disconnected, occlud­ed, or kinked tubing on the pressure sen­sor and chassis and reconnect or replace tubing, if needed.
3. Remove the control board. See Service Instruc tions.
4. Search for signs of water damage to the blower assembly, pressure sensor and replace, if damaged.
5. Inspect the blower. If a space is visible between the top and bottom sections, press the halves to eliminate the space.
6. Repeat Manual Calibration Procedure.
7. If the error codes are not resolved, replace the control board. See Service Instruc­tions.
8. Replace the cover. See Service Instruc­tions.
16
EN
C. Control PC Board Removal
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. Remove the cover. See instructions above.
2. Wearing an anti-static device, lift the cover off the chassis and place it, with the control board facing up and the blower facing down, onto a clean fl at surface.
3. Disconnect the two pieces of tubing from the outlet port side of the chassis.
4. Disconnect accessible wire harnesses: one behind and one to the left of the tubing in step 3, one at the right corner under the keypad, and one on the outside of the power supply board.
5. Remove one T-15 screw, located near the middle of the board beside LCD screen. Lift the board off the chassis slowly and disconnect any remaining wire harnesses from the control board.
6. Place control board keypad facing down onto a clean, fl at, static-free surface, if returning to the unit.
NOTE– After performing any service on the DV54/DV53 PAP,
please test the unit to ensure proper operation and cali­brate, if necessary.
A. Cover Removal
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. If using a DV5HH, remove the heated humidifi er and un­plug the tubing/humidifi er air supply support plug from the back of the cover and re-plug it in the outlet on the bottom cover.
2. Position the PAP on a clean, fl at surface with the keypad facing down.
3. Remove four T-20 screws from the bottom cover.
4. Lift the bottom cover straight up off the top cover.
NOTE– Attempting to reorient the PAP in order to lift the
top cover, instead of the bottom cover, may damage this device.
5. Wearing an anti-static device, turn the unit over and lift the top cover off the chassis.
B. Cover Replacement
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. Wearing an anti-static device, search for and connect all loose wire harnesses.
2. Place the top cover onto the chassis, with the keypad facing the cover and fi tting the control board around the screw posts in the cover.
3. Connect all tubing, carefully creating the correct loops and eliminating any kinking. (Fig. 1)
4. With the 1/8 inch inside bottom cover foam oriented correctly and lying fl at, place the bottom cover over the chassis and onto the top cover, fi tting the screw posts into proper alignment with the chassis and moving any pro­truding wire harnesses and tubing away from the edges of the cover.
5. Replace four T-20 screws.
9. Service Instructions
1
17
EN
D. Control PC Board Replacement
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. See instructions above to remove the cover and the con­trol PC board.
2. Wearing an anti-static device, verify that the silicone rat­tails holding the keypad and LCD display onto the control board are pulled through the appropriate holes in the board. See Keypad & LCD Display instructions.
3. Place the chassis on a clean fl at surface with the blower facing down. Orient the replacement board so that the keypad is over the rounded end of the chassis.
4. Tilt the board and connect the two wire harnesses near the power supply board and insert the brown and blue wires into the notch on the chassis. Tilt the board and connect the two wire harnesses near the outlet port. Connect the wire harness near the right corner of the keypad. (Fig. 2)
5. Position the board (Fig. 3) so that: the hole on the board (A), located below the ON/OFF key, is over the peg on the chassis; the small hole (B), located near the LCD screen, is over the screw hole on the chassis; and the cut-out (C) on the left rear corner fi ts around the protrusion above the air outlet port on the chassis. Move any stray tubing from between the board and the chassis, verify that all wire har­nesses are properly routed, and replace the T-15 screw on the board.
6. Replace the cover. See instructions above.
7. Connect the CPAP to a PC serial port with cable DV54/ DV53D-615 and turn the CPAP on.
8. Use a terminal program (such as Windows HyperTerminal, or equivalent) with COM settings 9600 baud, no parity, 8 data bits, 1 stop bit, Flow Control set to none.
9. When connection is established with the CPAP, the terminal program will show: “RI;” every 1/2 seconds. This display indicates that the CPAP is waiting for a handshake.
10. Type ‘ri’ and press ENTER. The CPAP should respond with the fi rmware version in the form “V0.XX dd/mm/yyyy.” Please note that all commands are case sensitive. Repeat this step if necessary to complete the handshake.
11. Type ‘Mn<cr>’. The CPAP will report the model number saved in memory, it should be “000000” if the PC board is new.
12. Read the model number from the CPAP label on the bot­tom of the unit (EX: DV54SE).
13. Type ‘Mn=nnnnnn<cr>’ where nnnnnn is the 6 digit model number read in the previous step (use a space for the last digit if necessary to make the string 6 digits). The CPAP should return the new six-digit model number.
14. Type ‘Sn<cr>’. The CPAP will report the serial number
2
3
A
B
C
saved in memory, it should be “00000000” if the PC board is new.
15. Read the serial number from the CPAP label on the bottom of the unit.
16. Type ‘Sn=nnnnnnnn<cr>’ where nnnnnnnn is the 8 digit serial number read in the previous step. The CPAP should return the new eight-digit serial number.
18
EN
17. Read the Hour Meter (“Uh<cr>”)– a new PC board should have 00000 hours. It is recommended that the original hour meter value be restored if the original blower is used. Set the hour meter by sending ‘Uh=nnnn<cr>’ where nnnn is the hour meter reading x 10 (Ex: 100.5 hours would be set by typing ‘Uh=1005<cr>’).
18. Read the Compliance Meter (“Up<cr>”)– a new PC board should have 00000 hours. Clear the Compliance meter if it is not zero by typing ‘Up=c<cr>’. The CPAP will return
000000.0 hours.
19. Check the fi rmware and hardware version information. Read and record the following:
• BIOS rev “VB<cr>”
• Firmware rev “VF<cr>”
Check the present version of fi rmware to make sure the
board is up to date. Install new fi rmware if the board has an old version.
20. Check for last saved error, type ‘El<cr>’. The CPAP will respond with “Exx dd/mm/yyyy” where xx is the last saved error code and dd/mm/yyyy is the date that the error was detected. Clear the last error by typing ‘Ec<cr>’, the CPAP will return “OK.” Verify that the error code is “00” by typing ‘El<cr>’.
21. Set all device settings to their default values by typing ‘TD<cr>’. The CPAP should return “OK.”
22. Recalibrate the unit.
E. Power Supply Board Removal
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. See instructions above to remove the cover and the con­trol PC board.
2. Wearing an anti-static device and with the power supply board facing you, apply thumb pressure to top left corner of the board while pulling the top right corner up and out of the mount clip at its bottom.
3. Pulling the top left corner out of its mount clip, lift the board off the chassis.
4. If returning the power supply board to the unit, place the board onto a clean, fl at, static-free surface with its large components facing up OR if discarding the power supply board, remove the two wire harnesses for use with the replacement board or purchase a new wire harness kit.
F. Power Supply Board Replacement
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. See instructions above to remove the cover, the control PC board and the power supply board.
2. Wearing an anti-static device, connect the wire harnesses from the old power supply board to the new power supply board or use a new wire harness kit.
3. Position the chassis so that the blower faces down and the inlet and outlet ports face you. Position the power supply board so that the components are facing away from you with the large brown component in the bottom left cor­ner and the small blue component in the top right corner.
4. Place the board into the mount clips at the bottom of the power supply board cavity and press down fi rmly on the top corners until the board is fully mounted.
5. See instructions above to replace the control PC board and the cover.
G. Blower Removal
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. See instructions above to remove the cover and control PC board
2. Carefully remove the silicone grommet holding the blower wire harness. (Fig. 4)
3. Holding the wire harness in your hand, turn the chassis over so that the blower is facing up.
4. Remove the foam covering the blower and the foam around the blower.
5. Reach into the blower cavity and push the silicone isolator off the blower outlet. (Fig. 5)
6. Slowly lift blower out of the chassis and carefully pull the wire harness through the hole in the chassis. If not return­ing blower to unit, remove foam from bottom of blower and save for new blower.
19
EN
6. Insert the top blower foam so that it conforms to the edges of the blower cavity portion of the chassis.
7. Turn the chassis over so that the blower is facing down. Gently pull any slack from the wire harness. Insert the silicone grommet into the hole in the chassis so that the slit in the grommet is toward the power supply board location, all the wires are in the center hole of the grom­met and the grommet fl ange is fully sealed. Route the wire harness along the top of the chassis and down the front of the blower cavity portion of the chassis. (Fig. 8)
8. See instructions above to replace the control PC board and the cover.
H. Blower Replacement
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. See instructions above to remove the cover, control PC board and blower.
2. Insert the blower mount foam on the blower base so that the holes in the foam align with the protrusions on the blower and the blower wire harness is through the center hole in the foam. (Fig. 6)
3. Turn the blower over so that the motor is facing down. Guide the wire harness through the hole in the bottom of the blower cavity portion of the chassis.
4. Lower the blower with its outlet port pointing at the silicone isolator. Position the blower’s outlet port into the top edge of the silicone isolator then tilt the outlet port into the isolator opening without folding the isolator. Push the outlet port into the isolator until the port is evenly inserted into the isolator. See isolator image in Blower Removal instructions.
5. Insert the blower wrap foam so that the single-notched end is beside the blower outlet port. Wrap the remain­ing foam strip clockwise around the blower so that the notched foam corresponds to the latches on the blower body. Wrap the double-notched end around the blower and over the outlet port. The top edge of the foam should be almost fl ush with the blower. (Fig. 7)
4
6
7
Slit direction
harness direction
8
20
EN
I. KeyPad & LCD Display Removal and
Replacement
CAUTION– Failure to wear anti-static equipment during ser-
vice may damage this device.
1. See instructions above to remove cover and control PC board.
2. Wearing an anti-static device, disconnect the two LCD display wire harnesses. (Fig. 9)
NOTE– Do not touch the exposed contact area on the end of
the ribbon cable. Dirt and oils may damage it.
3. Gently roll the silicone frame away from the edges of the LCD display screen (Fig. 10), lift the display off the keypad and place it on a clean, fl at, anti-static surface, if returning to the unit.
4. Lift the keypad off the board, gently pulling three silicone rattails through the board.
NOTE– Do not touch the keypad contacts on the PC board.
Dirt and oils may damage them.
5. Replace the keypad by inserting the silicone rattails on the replacement keypad through appropriate holes on the control board and gently pulling the rattails through the holes.
6. Insert the LCD display into the silicone frame on the keypad and connect the wire harnesses to the appropriate connectors on the board.
7. See instructions above to replace the control PC board and the cover.
9
10
21
EN
Unit Specifi cations
Size 4.2” H x 6.5” W x 6.9” D (10.7 cm x 16.5 cm x 17.5 cm)
Weight 2.7 lbs. (1.22 kg)
Electrical Requirements 100-240V~, 50/60 Hz
D.C. Operation 12 volts at 5 amps
Maximum Power Consumption 65 watts max from AC power source (PAP device only)
CPAP Pressure Range 3-20 cmH
2
O
Operating Temperature Range 41°F to 104°F (5°C to 40°C)
Operating Humidity Range 0% to 95% RH non-condensing
Operating Atmospheric Conditions Sea level to 9,000 feet
Storage & Transportation Temperature Range -40°F to 158°F (-40°C to +70°C)
Storage & Transportation Humidity Range 0% to 95% RH non-condensing
Maximum Limited Pressure 20 cmH
2
O under normal use
Sound Level 26 dBA
Filter Specifi cations
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Max Flow Rates
83 L/m @ 6.5 cmH2O (1/3 max pressure)
134 L/m @ 13 cmH
2
O (2/3 max pressure)
163 L/m @ 20 cmH
2
O (max pressure)
Pressure Accuracy ± 1.0 cmH
2
O
Pressure Swings (cmH
2
O peak to peak at 500mL tidal volume, sine wave profi le)
Breaths per minute
10 15 20
@ 6.5 cmH
2
O 0.5 0.5 0.5
@ 13 cmH
2
O 1.0 1.0 1.0
@ 20 cmH
2
O 1.0 1.0 1.0
Other Specifi cations
Equipment classifi cation with respect to protection from electric shock
Class II
Degree of protection from electric shock Type BF Applied Part
Degree of protection against ingress of liquids IPX1
Drip Proof Equipment not suitable for use in the presence of a fl ammable anesthetic mixture with air or with oxygen or nitrous oxide.
Mode of operation Continuous
Clinical Specifi cations for Digital Outputs
Pressure 0 to 25.5 cmH2O ± 1.0
22
EN
11. General Information DV5HH
A. Safeguards
When servicing electrical products, basic safety precautions should always be followed. Important safety information in this manual is highlighted by the following terms.
DANGER: Urgent safety information for hazards
that will cause serious injury or death.
WARNING: Important safety information for hazards
that might cause serious injury.
CAUTION: Information for preventing damage to the
product.
NOTE: Information to which you should pay
special attention.
PLEASE READ ALL INSTRUCTIONS BEFORE SERVICING THIS DEVICE.
DANGER!
Electric Shock Hazard – Do not use while bathing.
Electric Shock Hazard – Do not immerse this device into water or any other liquid.
Electric Shock Hazard – Do not attempt to open or re­move the enclosure. There are no user-serviceable internal components. If service is required, return the product to your home care provider. Opening or tampering with the product will void the warranty.
WARNING!
• Use the Heated Humidifi er system only for its intended use as described in this manual.
• Be sure to read and understand all safety instructions supplied with your IntelliPAP or SleepCube fl ow genera­tor device.
• Use only accessories recommended by DeVilbiss.
• Always remove the water chamber from the humidifi er cradle for fi lling.
• Use only tubing recommended or supplied with the fl ow generator.
• If water has been spilled onto the cradle or it has been submerged into water, unplug the power cord from the power source immediately. Allow the device to completely dry before use.
• If the device has been dropped, refer to the troubleshoot­ing guide for instructions. Contact your homecare provider for evaluation of the device.
• Never block the air openings of the cradle or chamber. Do not insert any objects into any openings or tubes.
• For proper operation of the humidifi er, the device must be placed on a fl at, level surface.
• Never place the system on a soft surface such as a bed or couch during operation.
• This heated humidifi er is intended for single-patient use only.
B. Travel
The DeVilbiss DV5HH humidifi er system
• Automatically accepts line voltages of 100-240 V, 50/60 Hz
• Needs power cord appropriate for area o USA DV54/DV53D-606 o Europe, except UK DV54/ DV53D-607 o UK DV54/DV53D-608 o Australia DV54/ DV53D-609 o Set of 3 (UK, Europe, Australia) DV54/ DV53D-611
C. DC Power
The DeVilbiss DV5HH heated humidifi er system
• Will NOT operate, if the DC power is coming directly from a 12V DC power source.
• Will operate, if the DC power coming to the PAP passes through an inverter so that AC power is delivered to the system.
• See DC Power information under the DV54/DV53 section of this manual
D. Product Description
The DV5HH Standard Heated Humidifi er System is intended to warm and add moisture to the pressurized air supplied to the patient during the treatment of Obstructive Sleep Apnea (OSA). The addition of heated humidifi cation to the air relieves dry­ness and irritation to the patient’s airway during OSA therapy. The heated humidifi er is used in conjunction with the DV5X series PAP devices and consists of a humidifi er chamber and humidifi er heater.
The humidifi er chamber holds enough water for a mini­mum of 8 hours of operation, with the heating unit at the maximum temperature setting at ambient conditions of 23°C and 25% R.H. with a system fl ow rate of 60 L/Min. The chamber slides out of the unit for cleaning and fi lling, without adjusting tubing connections. The chamber’s halves separate for easy cleaning but, when assembled, maintain a leak free seal to a maximum operating pressure of 30 cmH2O.
Heat transferred from the humidifi er heater to the humidi­fi er chamber raises the effi ciency of the DV5HH system. The patient controls the heater plate temperature using the up/ down arrow keypad keys on the attached PAP device to pro­vide the appropriate level of humidifi cation depending upon environmental conditions and individual preference.
23
EN
DV5HH
The heated humidifi er connects mechanically and electrically to a DV5X series PAP between the air stream going from the PAP to the patient. The keypad on the PAP controls the heated humidifi er’s output and ON/OFF functions. Before operation, the humidifi er chamber is fi lled with water to a predefi ned and visible water-fi ll level with the patient tubing and patient interface attached to the outlet port on the heated humidi­fi er system.
During operation, the DV5X series PAP device routes AC electrical power (100 – 240 VAC 50/60 Hz) to the heated humidifi er’s drive circuitry, which is contained within the PAP device. The drive circuitry directs the AC power to the heating element on the heated humidifi er’s heater plate. A thermistor on the heater plate provides continuous feedback back to the control circuitry so that the temperature of the heater plate is consistent with the setting selected via the PAP’s keypad. The PAP’s internal switch-mode power supply routes DC electrical power to the heater’s low-voltage circuitry.
The heated humidifi er’s heater plate transmits heat to the chamber, and then to the water, through a heat transfer plate located on the bottom of the humidifi er chamber. As the wa­ter is heated, its evaporation effi ciency increases. Air fl owing into the humidifi er chamber from the PAP device is humidi­fi ed by the evaporation and is then transmitted to the patient through the patient tubing and interface.
24
EN
A. Cleaning and Disinfection
See the DV5HH Instruction Guide (A-DV5HH) for information on cleaning the humidifi er chamber and heater base.
The DV5HH Standard Heated Humidifi er is single patient use only and therefore does not require disinfection.
B. Maintenance
The DV5HH Standard Heated Humidifi er System does not require regular maintenance.
C. Testing
Since the DV5HH device is powered and controlled by the PAP, it is important to determine which unit is at fault when there is an issue with humidity/heat. Use the following procedures to determine whether the DV5HH device or the PAP device is faulty.
1. Connect the DV5HH device to a known good PAP device.
2. After turning the system on,
• if the LCD display does NOT show “Heat:XXX” on the bottom line, proceed to the troubleshooting section.
• if the DV5HH device does NOT produce heat, proceed to the troubleshooting section.
3. Connect the PAP to a known good DV5HH device, if the
LCD display does NOT show “Heat:XXX” on the bottom line, proceed to the troubleshooting section.
If you detect excessive air or water leaks in the DV5HH sys­tem, proceed to the troubleshooting section.
13. Maintenance and Testing DV5HH
25
EN
14. Troubleshooting
HEATER
Symptom Action
The PAP does not recognize the heater. (With the PAP blower ON, the PAP display does not indicate Heat:xxx on the bot­tom line.)
Connect the heater to a known good PAP device, if one is available. If PAP display still does not recognize the heater, replace the heater.
If a known good PAP device is not available, remove the PAP from the heater and check the thermistor for resistance. The thermistor pins are located in the connector and are the ones closest to each other. See image below. The resistance should be 10 K ohms +/- 1% at 25° C (77°F). If open or shorted, replace the heater. See Service instructions. Alternative­ly: Replace the heater plate assembly, DV5H-607. If the resistance is in the correct range, the issue is in the PAP device. Troubleshoot as indicated below.
Heater plate does not respond to the PAP heat setting.
(With the PAP blower ON and the heat setting at 1 or above, the heater plate does not get warm.)
Connect the heater to a known good PAP device, if one is available. If the heater still does not respond to the heat setting, replace the heater.
If a known good PAP device is not available, connect the PAP, using the serial interface cable DV54/DV53D-615, to a PC using a terminal program (such as Windows HyperTer­minal, or equivalent) with Com settings 9600 baud, no parity, 8 data stop bits, 1 stop bit, Flow Control set to none. Turn the PAP blower ON and increase the heat setting to 10 using the PAP keypad (UP key). In the terminal program, connect to the PAP device by typing ‘ri’ followed by ENTER; then type ‘Ht’ followed by ENTER. Repeat the ‘Ht’ – ENTER sequence every 2-3 minutes and verify that the returned value increases. If the value in­creases, the system is operating normally. If the value does not increase, remove the PAP device from heater and measure the heater resistance. The heater terminals are the ones diagonal to each other on the connector.
heater terminals
t
hermistor terminals
The resistance should be between 180 to 220 ohms; if not replace the heater. See Service instructions. Alternatively: Replace the heater plate assembly, DV5H-607. If the resistance is in the correct range, the issue is in the PAP device. Troubleshoot as indicated below.
PAP
Symptom Action
The PAP does not recognize a known good DV5HH heater. (With the PAP blower ON, the PAP display does not indicate Heat:xxx on the bottom line.)
Check the wire harness on the PAP / heater connector; repair or replace, if needed. If the wire harness and connections are correct, replace the control PC board. See Service Instructions.
Heater plate on a known good DV5HH heater does not respond to the PAP heat setting.
(With the PAP blower ON and the heat setting at 1 or above, the heater plate does not get warm.)
Check the wire harness on the PAP / heater connector; repair or replace, if needed. If the wire harness and connections are correct, replace the control PC board. See Service Instructions.
MANIFOLD and O-RING
Symptom Action
Excessive air leak Remove the humidifi er chamber. Inspect the manifold in the heater base and replace, if
torn or worn. Inspect the O-Ring in the humidifi er chamber and replace, if worn or missing.
Water leak from the humidifi er chamber.
Remove the humidifi er chamber. Inspect the chamber for cracks in the plastic or water leak­ing from the heat transfer plate. If cracks or transfer plate leaks are found, replace the water chamber. Inspect the O-Ring in the humidifi er chamber and replace, if worn or missing.
DV5HH
26
EN
A. Removing the DV5HH base cover
1. Remove the humidifi er chamber from the DV5HH base.
2. Push the release button (Fig. 1) on the DV5HH base and
lift the PAP device off the heater base and set aside in a safe place.
3. Remove two T-10 screws from the top cover of the heater
base.
4. Turn the base over so that the bottom of the unit faces up
and remove four T-10 screws.
5. Lift the bottom cover off the unit
6. Remove the fi nal T-10 screw beside the wire harness clip.
(Fig. 2)
7. Hold the heater plate onto the base and turn the unit over.
8. Gently pull the top of the unit off the chassis and carefully
tilt the top without pulling the wire harness out of the connector.
9. Carefully release the wire harness from the harness clip.
Reposition the heater plate so that the orange side is fac­ing the harness clip and carefully push the heater plate through the opening. (Fig. 3)
B. Replacing the DV5HH base cover
1. Position the heater plate so that the orange side is fac-
ing the harness clip and carefully push the heater plate through the opening in the base. See image above.
2. Hold the heater plate onto the base and carefully insert
the harness wires, two at a time, into the harness clip. See image above.
3. Install the top of the base onto the heater chassis with the
wire harness and manifold clear of all edges and the latch working smooth and free. Fasten top to chassis with two T-10 screws.
4. Holding the heater plate onto the base, turn the unit over
and position heater plate into the shallow opening with the metal side facing down and the orange plastic side facing up. Move the harness wires between wire guides. (Fig. 4)
5. Replace the T-10 screw beside the wire harness clip.
6. Verify that the silicone spacer is in place on the inside of
the bottom cover. (Fig. 5)
7. Install the cover onto the bottom of the unit matching the
edges. Replace the 4 remaining T-10 screws in the bottom cover.
8. With the air supply support plug attached to the back of
the PAP device, place the front of the PAP onto the front of the heater base against the brackets and push the back of the PAP down so that the latch closes and the electrical connector is tight.
9. Slide the humidifi er chamber into the heater base until it
clicks into place.
DV5HH
1
harness clip
T-10 screw
2
3
4
27
EN
DV5HH
spring post
latch control
screw post
latch retainer
short leg
8
6
rattails
7
5
C. Removing and replacing silicone
manifold
1. Remove the cover as described above.
2. Squeeze the inlet port on the silicone manifold and push the port down into the hole in the top cover, gently releas­ing the manifold from all other connections. Discard the manifold or set aside in a safe place, if returning to the unit. (Fig. 6)
3. To replace, pinch the inlet port coupler on the replacement silicone manifold and insert it, from the inside of the cover, through the cover hole so that the coupler collar is fully and evenly positioned on the outside of the cover.
4. Press the manifold body into place and insert a fi nger into each of the rectangular openings, pressing the manifold’s rattails through the slots in the top cover until they are fully seated. (Fig. 7)
5. Replace the cover as described above.
D. Removing and replacing the latch
1. Remove the cover and silicone manifold as described above.
2. Raise the latch control away from the top cover so that it lifts the attached spring. Remove the spring from the latch and discard or set aside in a safe place, if returning it to the unit. (Fig. 8)
3. Still holding the latch control in a raised position, push it into the latch retainer and rotate the latch so that the spring post faces up. Gently push the short leg of the latch up and through the slot in the cover. Discard faulty latch.
4. Orient the replacement latch over the bottom side of the top cover so that short leg of the latch is over the slot near the screw post and the spring post is facing up.
5. Insert the latch into the slot by rotating the latch stem back toward the screw post and gently applying pressure to the short leg until it fi ts through the slot.
6. Place one end of the spring over the spring post and insert the other end into the channel on the cover by slightly compressing the spring. Push the latch control several times to ensure that the system is confi gured properly.
7. Replace the silicone manifold and cover as described above.
28
EN
Humidity Output (in the operating fl ow range) ≥10 mgH2O/L air
Chamber/Cradle Dimensions
Size 2.6” H x 6.3” W x 8.4” D (6.6 cm x16.0 cm x 21.3cm)
Weight 1.75 lbs. (0.794 Kg)
Elec trical Rating:
Electrical Supply Frequency 50/60 Hz
Power Consumption 85 Watts
Voltage and Current
DV5HH 100-240VAC, 0.95 A
Heater Plate Cutout 152°C (305°F)
Power/Temperature Control Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature approx. 29°C to approx. 65°C (84°F to 149°F)
Operating, Transport & Storage
Operating Temperature Range 41°F to 104°F (5°C to 40°C)
Operating Humidity Range 0 to 95% R.H. non-condensing
Operating Atmospheric Pressure Range 70.0 – 106.0 kPA
Transport & Storage Temperature Range -40°F to 158°F (-40°C to +70°C)
Transport & Storage Humidity Range 0 to 95% R.H. non-condensing
Transport & Storage Atmospheric Pressure Range 50.0 – 106.0 kPA
Class II Equipment; Type BF Applied Parts; Continuous Operation
IPX1, Drip-proof vertical
16. Unit Specifications DV5HH
Loading...
+ 64 hidden pages