DeVilbiss Pulmo-Aide 3655D Instruction Manual

DeVilbiss®Compact Compressor/Nebulizer Instruction Guide
Pulmo-Aide®Compact Compressor Model 3655D
CAUTION– Federal (U.S.A.) law restricts this device to sale by, or on the order of a physician.
Made in U.S.A. Latex free. Entela certified to UL1431, CAN/CSA C22.2 No601.1 and IEC60601-1.
Compresor/ Nebulizador Compacto DeVilbiss®Guía de instrucciones
Compresor Compacto Pulmo-Aide Modelos 3655D
PRECAUCION– La ley federal (EE.UU.) restringe la venta de este aparato para ser realizada
por, o bajo prescripción de, un médico.
Hecho en EE.UU. No Contiene Látex. Certificación Entela para UL1431, CAN/CSA C22.2 Nº 601.1, e IEC 60601-1.
Guide dInstructions du Compresseur/Nébuliseur compact DeVilbiss
®
Compresseur compact Pulmo-Aide modèle 3655D
ATTENTION– La loi fédérale (États-Unis) limite la vente de ce dispositif par ou sur ordon-
nance d’un médecin.
Fabriqué aux États-Unis. Garanti sans Latex. Homologation Entela aux normes UL1431, CAN/CSA C22.2 N° 601.1 et IEC 60601-1
Table of Contents
A-655
2
A-655
Important Safeguards
3
IEC SYMBOLS
Attention, consult instruction guide
Alternating current
Type BF applied part
Double Insulated
IMPORTANT SAFEGUARDS
When using electrical products, especially when children are present, basic safety pre­cautions should always be followed. Read all instructions before using. Important information is highlighted by these terms: DANGER– Urgent safety information for hazards that will cause serious injury or
death.
WARNING–Important safety information for hazards that might cause serious injury. CAUTION– Information for preventing damage to the product. NOTE– Information to which you should pay special attention.
READ ALL INSTRUCTIONS BEFORE USING.
DANGER
To reduce the risk of electrocution:
1. Always unplug this product immediately after using.
2. Do not use while bathing.
3. Do not place or store product where it can fall or be pulled into a tub or sink.
4. Do not place in or drop into water or other liquid.
5. Do not reach for a product that has fallen into water. Unplug immediately.
To reduce the risk of burns, electrocution, fire or injury to persons:
1. A product should never be left unattended when plugged in.
2. Close supervision is necessary when this product is used by, on, or near children or invalids.
3. Use this product only for its intended use as described in this guide. Do not use attachments not recommended by the manufacturer.
4. Never operate this product if: a. It has a damaged power cord or plug. b. It is not working properly. c. It has been dropped or damaged. d. It has been dropped into water. Return the product to an authorized Sunrise service center for examination and repair.
ENGLISH
IEC Symmbols ................................................................................ 3
Important Safeguards...................................................................... 3
Introduction .................................................................................. 4
Important Parts of Your DeVilbiss Compressor ..................................... 5
Important Parts of Your Disposable Nebulizer...................................... 5
Important Parts of Your Reusable Nebulizer ........................................ 5
How to Operate Your Compressor....................................................... 6
Nebulizer Cleaning.......................................................................... 8
Compressor Cleaning ....................................................................... 9
Filter Change ................................................................................. 9
Maintenance .................................................................................. 9
Specifications ................................................................................ 10
Warranty........................................................................................ 11
ESPAÑOL
Simbolos IEC.................................................................................. 12
Precauciones Importantes................................................................ 12
Introducción.................................................................................. 13
Partes Importantes de Su Compresor DeVilbiss .................................... 14
Partes Importantes de Su Nebulizador Desechable ............................... 14
Partes Importantes de Su Nebulizador Reutilizable............................... 14
Cómo Operar Su Compresor............................................................... 15
Limpieza del Nebulizador ................................................................. 17
Limpieza del Compresor ................................................................... 18
Cambio de Filtro ............................................................................. 19
Mantenimiento............................................................................... 19
Especificaciones ............................................................................. 20
Garantía........................................................................................ 21
FRANÇAIS
Symboles CEI ................................................................................. 22
Mises en Garde Importantes ............................................................. 22
Introduction .................................................................................. 23
Parties importantes de votre compresseur DeVilbiss ............................. 24
Pièces importantes de votre nébuliseur jetable ................................... 24
Pièces importantes de votre nébuliseur réutilisable ............................. 24
Comment faire fonctionner le nébuliseur ............................................ 25
Nettoyage du nébuliseur.................................................................. 27
Nettoyage du compresseur ............................................................... 28
Changement du filtre....................................................................... 28
Entretien....................................................................................... 28
Spécifications ................................................................................ 29
Garantie........................................................................................ 30
A-655
Important Parts
5
1
2
4
5
3
6
7
Figure C
IMPORTANT PARTS OF YOUR DEVILBISS COMPRESSOR (FIGURES A & B)
1. Power switch
2. Filter
3. Air-outlet connector
4. Carrying handle
5. Power cord storage recess
IMPORTANT PARTS OF YOUR DISPOSABLE NEBULIZER (FIGURE C)
1. Mouthpiece
2. T-piece
3. Cap
4. Medication Cup
5. Nebulizer Air-Inlet Connector
6. Baffle
7. Tubing
Nebulizer parts are not sold separately. Additional neb­ulizers (part #4650D-621) are available through your Sunrise provider.
Figure A
Important Safeguards
A-655
4
5. Keep the power cord away from heated surfaces.
6. Never block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air open­ings free of lint, hair, and the like.
7. Never use while drowsy or asleep.
8. Never drop or insert any object into any opening or hose.
9. Do not use outdoors. This product is for household use only.
10. Do not use in an oxygen enriched environment.
11. Do not use this product with a DC to AC converter, or with any AC voltage and/or frequency other than specified.
NOTE– This unit employs both a temperature and current limiting device that will shut
off the entire product should a malfunction occur. Contact your qualified Sunrise provider for a replacement unit.
If replacement of the cord or plug is necessary, contact your qualified Sunrise provider.
This appliance has a polarized plug (one blade is wider than the other). As a safety feature, this plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electri­cian. Do not attempt to defeat this safety feature.
SAVE THESE INSTRUCTIONS
INTRODUCTION
Your doctor has prescribed a liquid medication to treat your respiratory condition. To best utilize this liquid medication, he has prescribed a DeVilbiss brand compressor/nebulizer. Your DeVilbiss compressor/nebulizer works to convert the med­ication to a high-quality mist of fine particles that penetrates deep into the lungs.
Be sure that you read and understand the information in this instruction guide. By fol­lowing these simple instructions and the advice of your physician, your compressor will become an effective addition to your therapeutic routine.
Statement of Intended Use
The DeVilbiss compressor/nebulizer Model 3655 includes an AC powered air compressor that provides a source of compressed air for home health care use. The compressor is used in conjunction with a jet (pneumatic) nebulizer to convert certain inhalable drugs into an aerosol form for inhalation by a patient. The device can be used with adult or pediatric patients.
NOTE– If using a nebulizer other than one distributed by Sunrise Medical, verify that
the pressure and/or flow characteristics of the nebulizer (while driven by the Model 3655) meet the nebulizer manufacturer’s recommendation.
1
2
3
4
5
Figure B
5
Figure 2
A-655
How to Operate Your Compressor
7
How to Operate Your Compressor
A-655
6
9. Attach tubing to nebulizer air-inlet connector (
Fig. 5).
10. Press the power switch “On” to start the com­pressor.
11. Begin treatment by placing the mouthpiece between teeth. With mouth closed, inhale deeply and slowly through mouth as aerosol begins to flow, then exhale slowly through the mouthpiece (
Fig. 6). If treatment needs to be
interrupted, simply press power switch “Off”.
NOTE– Some physicians recommend a “clearing breath”
after every five to seven treatment breaths. Remove mouthpiece from mouth and hold breath for at least five seconds (ten is better), then exhale slowly.
12. If an aerosol mask is used, place the mask over mouth and nose (
Fig. 7). As aerosol begins to
flow, inhale deeply and slowly through mouth, then exhale slowly.
13. When treatment is complete, turn unit off by pressing the power switch to the “Off” position. Unplug the unit from the wall power outlet.
Figure 5
Figure 6
Figure 7
HOW TO OPERATE YOUR COMPRESSOR
NOTE– Before initial operation, your nebulizer should be
cleaned following cleaning instructions, or as rec­ommended by your physician or Sunrise provider.
1. Make sure the power switch is in the “Off” posi­tion (
Fig. 1).
2. Unwrap power cord.
3. Place compressor so that the rubber feet are resting on a level, sturdy surface (
Fig. 2) and
upon which the controls can be easily reached when seated. Never block the air openings of the product or place it on a soft surface, such as a patient’s lap or a bed or couch where the air openings may be blocked.
4. Plug power cord into an appropriate wall outlet.
DANGER
DeVilbiss compressor/nebulizer must be operat­ed on the specified power source to avoid risk of electric shock and damage to the compressor.
5. Wash hands.
6. Connect one end of the nebulizer tubing to the compressor air-outlet connector (
Fig. 2).
NOTE– During high-humidity weather conditions, con-
densation (water build up) can occur in the nebulizer tubing.
7. Assemble clean nebulizer parts by placing baffle down in medication cup. Add prescribed med­ication into the medication cup (
Fig. 3).
Holding cup stationary, screw on nebulizer cap (
Fig 4).
8. Assemble mouthpiece and t-piece (if applicable) and insert into the top of the nebulizer cap. If using an aerosol mask, insert the bottom part of the mask directly into the top of the nebuliz­er cap.
Figure 1
Figure 3
Figure 4
A-655
Cleaning
9
Cleaning
A-655
8
To prevent possible risk of infection from contaminated cleaning solutions, always prepare fresh solution for each cleaning cycle and discard solution after each use.
3. Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing does not have to be washed because only filtered air passes through it.
NOTE– The DeVilbiss Disposable Nebulizer will last at least 15 days and possibly longer,
depending on usage. Proper cleaning will also help extend the life of the nebu­lizer. Because it is disposable, we recommend that an extra nebulizer be kept on hand at all times. Sunrise also manufactures the DeVilbiss Reusable Nebulizer that is dishwasher safe and may be cleaned and reused for up to one year.
COMPRESSOR CLEANING
1. With power switch in the "Off" position, unplug power cord from wall outlet.
2. Wipe outside of the compressor cabinet with a clean damp cloth every few days to keep dust-free.
DANGER
Do not submerge in water; doing so will result in compressor damage.
FILTER CHANGE
1. Filter should be changed every 6 months or sooner if filter discolors.
2. Remove filter
(Fig. A) by grasping it firmly and pulling out of the unit.
Discard old filter.
3. Replace with a new DeVilbiss filter (part #3655D-601). Additional filters should be purchased from your Sunrise provider.
4. Push new filter into position.
CAUTION– Do not reuse or wash filter or substitute any other material such as cotton
for a DeVilbiss air-inlet filter. It will result in compressor damage.
MAINTENANCE
All maintenance must be performed by a qualified Sunrise provider or authorized ser­vice center.
DANGER
Electric shock hazard. Do not remove compressor cabinet. All disassembly and maintenance must be done by a qualified Sunrise provider.
NEBULIZER CLEANING
All parts of the nebulizer, except tubing, should be cleaned according to the following instructions. Your physician and/or Sunrise provider may specify a certain cleaning procedure. If so, follow their recommendations.
To prevent possible risk of infection from con­taminated medication, cleaning of the nebulizer is recommended after each aerosol treatment. Disinfecting is recommended once a day.
Clean After Every Use:
1. With power switch in the “Off” position, unplug power cord from wall outlet.
2. Disconnect tubing from the air-inlet connector and set aside.
3. Disassemble mouthpiece or mask from cap. Open nebulizer by turning cap counterclockwise and removing baffle.
4. Wash all items, except tubing, in a hot water/dishwashing detergent solution. Rinse under hot tap water for 30 seconds to remove detergent residue. Allow to air dry.
Disinfect Daily:
1. Using a clean container or bowl, soak items in three parts hot water to one part white vinegar for 30 minutes (
Fig. 8) OR use a medical bacte-
rial-germicidal disinfectant available through your provider. For reusable nebulizer only, clean daily in dishwasher using top shelf. Be sure to follow manufacturer's instructions carefully.
2. With clean hands, remove items from disinfectant solution, rinse under hot tap water, and air dry on a clean paper towel. Store in a zip-lock bag.
NOTE– Do not towel dry nebulizer parts; this could
cause contamination.
CAUTION– The DeVilbiss Reusable Nebulizer is dish-
washer safe, but do not place any dispos­able nebulizer parts in an automatic dish­washer; doing so could cause damage.
Figure 8
A-655
Warranty
11
Specifications
A-655
10
FIVE-YEAR LIMITED WARRANTY ON COMPRESSOR PARTS
All DeVilbiss Compact Compressor models are warranted to be free from defective work­manship and material for a period of five years from the date of purchase. Any defec­tive part(s) will be repaired or replaced at Sunrise’s option if the unit has not been tampered with or used improperly during that period. Make certain that any malfunc­tion is not due to inadequate cleaning or failure to follow the instructions. If repair is necessary, contact your authorized Sunrise provider or Sunrise Service Department at 800-333-4000 (814-443-4881) (Canada 800-263-3390 or 905-660-2459) for instruc­tions. NOTE— This warranty does not cover providing a loaner compressor, compensat­ing for costs incurred in compressor rental while said unit is under repair, or costs for labor incurred in repairing or replacing defective part(s).
THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS, OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.
This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
MANUFACTURER'S NOTE
Thank you for choosing a DeVilbiss compressor/nebulizer. We want you to be a satis­fied customer. If you have any questions or comments, please send them to our address on the back of this manual.
SPECIFICATIONS
Size................................................................................7.5" W x 4.0" H x 7.2" D
(19.1 cm x 10.2 cm x 18.3 cm)
Weight ....................................................................................4.8 lbs. (2.18 kg)
Electrical Requirements
.....................................................2.5 amps, 115 VAC (±10%) 60 Hz
Sound Level.............................................................................................58 dBA
Power Consumption ...............................................................140 watts maximum
Max. Compressor Pressure..........................................35 psig or greater* (241 kPa)
Compressor Free Air Flow ...........................................................8 lpm or greater*
Operating Temperature Range .......................................................+40° to +104°F
(+5 to +40°C)
Operating Humidity ......................................................up to 95% non-condensing
Storage/Transport Temperature Range ............................................-40° to +158°F
(-40° to +70°C)
Storage/Transport Humidity ..........................................up to 95% non-condensing
Supplied Nebulizer Operating Pressure .................................9 psig (min) (62 kPa)
Supplied Nebulizer Liter Flow ........................................................5.5 lpm (min)
Supplied Nebulizer Capacity .................................................................6 ml (cc)
Supplied Nebulizer Nebulization Rate .....................................0.15 ml/min or greater*
Supplied Nebulizer MMAD .......................................................5 microns or less**
*Conditions may vary based on altitude above sea level, barometric pressure and temperature. **Value determined with minimum flow rate of 5.5 lpm through the nebulizer (20° C, sea level).
This product complies with the following electromagnetic compatibility standard: IEC 60601-1-2:2nd Edition
Equipment Classifications:
Protection against electric shock: ................................................................Type II
Degree of protection against electric shock: .................................................Type BF
Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide
Mode of operation:.................................................................Continuous operation
FOR SERVICE CALL YOUR AUTHORIZED SUNRISE PROVIDER:
____________________________________________________ ____________________________________________________
____________________________________________________ Phone ___________________________________________________ Purchase Date _____________________________________________ Serial # __________________________________________________