DeVilbiss PD1000 Service Manual

DeVilbiss® PulseDose® ComPaCt ConserVing DeViCe
erViCe manual
s
CAUTION-Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
Model PD1000
GENERAL INFORMATION / THEORY OF OPERATION
TABLE OF CONTENTS
I. GENERAL INFORMATION
A. Initial Inspection ................................................................... 2
B. Maintaining the PD1000............................................................. 2
C. General Description ................................................................ 2
II. THEORY OF OPERATION ............................................................. 2
III. INSTALLATION AND OPERATION
A. Installation ....................................................................... 3
B. Operation ........................................................................ 3
IV. MAINTENANCE PROCEDURES
A. Testing .......................................................................... 4
B. Cleaning......................................................................... 4
V. TROUBLESHOOTING................................................................. 5
VI. SERVICE INSTRUCTIONS
A. Regulator Seal Replacement......................................................... 6
B. Battery Replacement ............................................................... 6
C. Cover / Regulator and Gauge Removal ................................................ 6
D. Cannula Fitting Removal/Replacement ................................................. 7
E. Solenoid/Vacuum Switch Testing and Removal........................................... 7
VII. PARTS AND IllUSTRATIONS ........................................................... 8
VIII. PNEUMATIC DIAGRAM ............................................................... 9
IX. UNIT SPECIFICATIONS ............................................................... 9
X. PROVIDER'S NOTES ................................................................. 9
XI. DEVILBISS GUIDANCE AND MANUFACTURER’S DECLARATION ............................ 10
XII. PARTS RETURN AND ORDERING POLICY ............................................... 11
XIII. WARRANTY ........................................................................ 11
I. GENERAL INFORMATION
A. Initial Inspection
An initial inspection should be performed on the PD1000 as soon as possible after receipt. When removed from carton, an inspection should be made for any damage due to shipping. If shipping damage has occurred, call DeVilbiss Healthcare at 800-338-1988 (814-443-4881) for replacement instructions.
B. Maintaining the PD1000
The PD1000 should be periodically maintained according to the guidelines set forth in Section IV. Maintenance, and
testing should only be done by qualied service personnel. Failure to follow the procedures set forth in this manual
may void the warranty.
C. General Description
The PD1000 PulseDose system delivers a pulse or "bolus" of oxygen at the leading edge of inspiration. This bolus is
delivered at both the proper ow and volume so that it is delivered deep into the lungs where gas exchange takes
place. The PD1000 PulseDose is rate responsive from 6 BPM to 40 BPM.
II. THEORY OF OPERATION
The PD1000 utilizes a vacuum switch to detect the negative pressure at the beginning of each inspiration (approximately .1 inches water column due to inhalation). That, in turn, opens the solenoid for a time interval that
corresponds to the ow rate selected on the rotary selector. At higher ow rates, the valve is open longer resulting in
increased pulse volumes. Atmospheric pressure compensation occurs automatically because the pressure side of the vacuum switch is open
to atmosphere.
All PD1000 units are set on 2 liters per minute continuous ow from the factory. Changing to a different continuous ow rate involves changing the cannula tting as outlined in Section VI, D.
The PD1000 delivers 16.5 cc 02 per setting number (i.e. setting 2 = 33cc 02).
LT-1859
2
INSTALLATION AND OPERATION
III. INSTALLATION AND OPERATION
A. Installation
The PD1000 utilizes a rotary selector that has three modes for use by the patient. Those modes are: OFF, "PulseDose", and Continuous Flow. The PD1000 is battery operated and is turned off by turning the Rotary Selector to the "OFF" position. The unit requires (2) "AA" batteries to operate in PulseDose mode (Figure
1). The batteries will not become discharged as a result of not turning the Rotary Selector to the "OFF" position.
1
Press Latch to Open Battery Door
The PD1000 mounts on a standard CGA870 type post
using guide pins and the knob (Figure 2). It can be used on C, D, E, ML-6, M4 and M-6 size tanks at
pressures between 500-2250 PSIG. Also verify that the regulator seal (Part #9286-RD) is in place and in good condition (Section VI, A). Position the guide pins into
the tank post holes, and tighten the knob until the
PD1000 is securely in position and there are no seal
leaks.
The tank valve can now be slowly opened and the rotary selector can be set to the prescribed ow rate.
Verify that a pulse is being delivered at the leading
edge of each inhalation. As the ow rate is increased,
so is the duration of the pulse (Figure 3).
To use the PD1000 in the "Continuous Flow" mode, turn the Rotary Selector to the "CF" position (Figure 3). See Note in the Important Parts section of the instruction guide A-1000.
3
"CF"
Rotary Selector
The "CF" ow rate is set at 2 liters per minute. This ow rate can be changed by changing cannula ttings
(Figure 4). See Service Instructions Section VI, D.
4
2
Cannula Fitting
B. Operation
When using the PD1000 in the PulseDose mode, the patient must breathe through the nose only. A standard nasal cannula must be used. Do not use a pediatric or
low-ow cannula. The cannula with tubing can be up to
35 feet in length, but a 10 foot maximum is recommended to lessen the chances of the oxygen cylinder tipping over while in use.
Do not use on patients who can only mouth breathe. The PD1000 should only be used on patients capable of nose breathing.
3
LT-1859
INSTALLATION & OPERATION / MAINTENANCE PROCEDURES
Remaining battery life can be observed as the Pulse Indicator Lights illuminate with each breath (Figure 5).
5
Battery Indicator Lights
If the Green Pulse Indicator Light illuminates with each
breath, the batteries have sufcient power (8 or more
hours).
If the Red Pulse Indicator Light illuminates with each breath, the batteries have between 4-8 hours of battery life remaining.
If the Red Indicator Light illuminates continuously, the batteries must be changed. The unit can be used on
the continuous ow "CF" setting if no batteries are
immediately available.
NOTE-Turn the Rotary Selector to the OFF setting prior to changing batteries.
NOTE-The oxygen cylinder will not last as long in
continuous ow mode as it would in PulseDose mode.
The contents gauge indicates the approximate amount
of oxygen left in the tank. It reads 1/4, 1/2, 3/4, and full. It also has a red area to emphasize when the tank
needs to be changed (Figure 6).
IV. MAINTENANCE PROCEDURES
A. Testing
NOTE-The following should be performed after repair
or between patients
1. Install known good batteries.
2. Connect the unit to a pressurized oxygen cylinder as described in Section III-A and connect a nasal cannula.
3. Open the tank valve and verify that the PD1000 contents gauge indicates that there is a full tank.
4. Verify that there are no leaks around the tank seal between the PD1000 and the oxygen tank.
5. Select any ow rate and simulate an inhalation
through the nasal cannula while verifying that a "pulse" or bolus of oxygen is delivered with each simulated breath.
6. Position the Rotary Selector to the "CF" position and verify that the Continuous Flow Mode of operation is functioning.
7. Verify that the top and bottom covers are not
cracked or broken.
8. Verify that the label can be read and is not damaged.
9. Turn the Rotary Selector to the "OFF" position,
dose the tank valve, and remove PD1000 from
cylinder.
B. Cleaning
Wipe with a damp cloth having a maximum 5.25% Sodium Hypochlorite (Bleach) or 3% Hydrogen
peroxide solution. Avoid getting uids or debris such as
sand or dirt inside the oxygen connections. Do not immerse in water.
6
LT-1859
Contents Gauge
4
Loading...
+ 8 hidden pages