ASSEMBLY INSTRUCTIONS FOR PL-ICKIT
UNPACKING
Open package containing PL-ICKIT.
You should have the following components:
ITEM PART NO. DESCRIPTION QTY
TOOLS REQUIRED
3/16 Allen wrench 1/2" open end, box end, or combination wrench 1/4 Allen wrench 9/16" open end, box end, or combination wrench
ASSEMBLY
Referring to the figure below:
1. Attach the Connecting Links (1) to the PLXXX scale (Load Cell) using one of the Socket-Head Shoulder Bolt (2), 2 Spacers (4) (1 on each side of the Load Cell) and 5/16-18 Elastic Nut (3). Securely tighten using the 3/16 Allen wrench and 1/2 wrench.
2. Place the Yoke Pin into the Connecting Links (1) and then secure the Yoke Pin to the Connecting Links (1) with a Socket-Head Shoulder Bolt (2) and 5/16-18 Elastic Nut (3). Securely tighten using the 3/16 Allen wrench and 1/2 wrench.
3. Attach the Adapter (7) to the PLXXX scale using the .375-16 x .625 Socket-Head Shoulder screw and .375-16 Elastic Nut (6). Securely tighten using the 1/4 Allen wrench and 9/16 wrench.
4. Attach the entire assembly to the patient lift arm using the remaining Socket-Head Shoulder Bolt (2) and 5/16-18 Elastic Nut (3). Securely tighten using the 3/16 Allen wrench and 1/2 wrench.
6
2
4
7
5
3
3
1
2
0046-M509-O1 Rev A 10/04
Patient Lift Arm
(Not Included)
PL-400 or PL-
600
(Not Included)
Yoke Pin
(Not Included)
DO NOT OVER TIGHTEN MOUNTING HARDWARE.
This will damage mounting brackets. Brackets must pivot. The bolt threads must protrude past the elastic stop nut to ensure that locking feature functions.
Page 2
2
Page 3
2
Page 4
ASSEMBLY INSTRUCTIONS FOR PL-ICKIT
UNPACKING
Open package containing PL-ICKIT.
You should have the following components:
ITEM PART NO. DESCRIPTION QTY
TOOLS REQUIRED
3/16 Allen wrench 1/2" open end, box end, or combination wrench 1/4 Allen wrench 9/16" open end, box end, or combination wrench
ASSEMBLY
Referring to the figure below:
1. Attach the Connecting Links (1) to the PLXXX scale (Load Cell) using one of the Socket-Head Shoulder Bolt (2), 2 Spacers (4) (1 on each side of the Load Cell) and 5/16-18 Elastic Nut (3). Securely tighten using the 3/16 Allen wrench and 1/2 wrench.
2. Place the Yoke Pin into the Connecting Links (1) and then secure the Yoke Pin to the Connecting Links (1) with a Socket-Head Shoulder Bolt (2) and 5/16-18 Elastic Nut (3). Securely tighten using the 3/16 Allen wrench and 1/2 wrench.
3. Attach the Adapter (7) to the PLXXX scale using the .375-16 x .625 Socket-Head Shoulder screw and .375-16 Elastic Nut (6). Securely tighten using the 1/4 Allen wrench and 9/16 wrench.
4. Attach the entire assembly to the patient lift arm using the remaining Socket-Head Shoulder Bolt (2) and 5/16-18 Elastic Nut (3). Securely tighten using the 3/16 Allen wrench and 1/2 wrench.
6
2
4
7
5
3
3
1
2
0046-M509-O1 Rev A 10/04
Patient Lift Arm
(Not Included)
PL-400 or PL-
600
(Not Included)
Yoke Pin
(Not Included)
DO NOT OVER TIGHTEN MOUNTING HARDWARE.
This will damage mounting brackets. Brackets must pivot. The bolt threads must protrude past the elastic stop nut to ensure that locking feature functions.
Loading...