Thank you for purchasing our Detecto icon® Digital Clinical Scale. It has been manufactured
with quality and reliability at our factory in Webb City, MO USA. Your scale has been tested
before leaving our factory to insure accuracy and dependability for years to come.
This manual is provided to guide you through installation and operation of your scale. Please
read it thoroughly before attempting to install or operate your scale and keep it handy for future
reference.
COPYRIGHT
All rights reserved. Reproduction or use, without expressed written permission, of editorial or
pictorial content, in any manner, is prohibited. No patent liability is assumed with respect to
the use of the information contained herein.
DISCLAIMER
While every precaution has been taken in the preparation of this manual, the Seller assumes
no responsibility for errors or omissions. Neither is any liability assumed for damages resulting
from use of the information contained herein. All instructions and diagrams have been
checked for accuracy and ease of application; however, success and safety in working with
tools depend to a great extent upon the individual accuracy, skill and caution. For this reason
the Seller is not able to guarantee the result of any procedure contained herein. Nor can they
assume responsibility for any damage to property or injury to persons occasioned from the
procedures. Persons engaging the procedures do so entirely at their own risk.
FCC COMPLIANCE STATEMENT
This equipment generates, uses and can radiate radio frequency and if not installed and used
in accordance with the instruction manual, may cause interference to radio communications. It
has been designed within the limits for a Class A computing device pursuant to Subpart J of
Part 15 of FCC rules to provide reasonable protection against such interference when
operated in a commercial environment. Operation of this equipment in a residential area may
cause interference in which case the user will be responsible to take whatever measures
necessary to correct the interference.
You may find the booklet “How to Identify and Resolve Radio TV Interference Problems”
prepared by the Federal Communications Commission helpful. It is available from the U.S.
Government Printing Office, Washington, D.C. 20402. Request stock No. 001-000-00315-4.
Before using this scale, read this manual
and pay special attention to all
"NOTIFICATION" symbols:
PRECAUTIONS
IMPORTANT ELECTRICAL
WARNING
3300-0192-0M Rev D 01/17 3
I
Page 4
PROPER DISPOSAL
When this device reaches the end of its useful life, it must be properly disposed of. It must not
be disposed of as unsorted municipal waste. Within the European Union, this device should
be returned to the distributor from where it was purchased for proper disposal. This is in
accordance with EU Directive 2002/96/EC. Within North America, the device should be
disposed of in accordance with the local laws regarding the disposal of waste electrical and
electronic equipment.
It is everyone’s responsibility to help maintain the environment and to reduce
the effects of hazardous substances contained in electrical and electronic
equipment on human health. Please do your part by making certain that this
device is properly disposed of. The symbol shown to the right indicates that
this device must not be disposed of in unsorted municipal waste programs.
CAUTION
CAUTION: RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN
INCORRECT TYPE. DISPOSE OF USED BATTERIES ACCORDING TO THE
INSTRUCTIONS.
ATTENTION: RISQUE D'EXPLOSION SI LA BATTERIES EST REMPLACE'E PAR
UN TYPE INCORRECT. REJETEZ LES BATTERIES UTILISE'ES SELON LES
INSTRUCTIONS.
300 kg x 0.1 kg / 500 kg x 0.2 kg
Weighing/Height Units: Pounds/Inch (lb, in) or Kilograms/Centimeters (kg, cm)
Platform Size: 17 in W x 17 in D x 1.5 in H
(43 cm W x 43 cm D x 3.8 cm D)
Overall Dimensions: 17 in W x 18.5 in D x 88.5 in H
(43 cm W x 47 cm D x 224.8 cm H)
Sonar Height Rod Measures: 24” (2’) – 86” (7’2”) x 0.2 in
(60 cm – 218 cm) x 0.5 cm
Display Type: 3.2 in (diagonal) full color TFT display
320x240 resolution
Number of Characters: Weight: 5 digit, 0.8 in (20 mm) high
Height: 4 digit, 0.35 in (9 mm) high
BMI: 3 digit, 0.35 in (9 mm) high
Touch Screen: Resistive touch panel covering the display
Keys: Mechanical switch type, Power, Zero, Lock/Release, ID,
Cycle Display Mode, Menu, Print
Power Requirements: Included 100 to 240 VAC 50/60Hz 12 VDC 1A wall plug-in
UL/CSA listed AC power adapter (Cardinal part number
6800-1045) or optional 12 “AA” cell Alkaline, Ni-Cad or NiMH
batteries (not included)
Operating Environment:
Communication Interfaces: RS232, USB
Standard Features:
Dual Range Weight
Auto Weight Lock Feature
StableSENSE
® 1
8-Digit Numeric Patient ID
Optional Features:
Wi-Fi
Bluetooth
1
StableSENSE® is a digital filter utilizing proprietary software algorithms to remove or greatly
reduce changes in the weight display resulting from movement on the scale platform.
StableSENSE
movement or vibration on the scale. Any application affected by vibration or movement on
the scale platform can benefit using StableSENSE
®
can be used with clinical scales to lessen the effects of the patient’s
Operated Temperature Range: 14 to 104 ºF (-10 to +40 ºC)
Humidity: 0 to 90% non-condensing
1 RS232 Serial Port
1 USB-B Port
Adjustable Filtering Sonar Height Rod
WAConnect, Welch Allyn® Interface
®
.
3300-0192-0M Rev D 01/17 1
Page 8
This page intentionally left blank.
3300-0192-0M Rev D 01/17 2
Page 9
SITE PREPARATION REQUIREMENTS
The icon Digital Clinical Scale is a precision weight indicating instrument. As with any
precision instrument, it requires an acceptable environment to operate at peak performance
and reliability. This section is provided to assist you in obtaining such an environment.
Environmental
For indoor use only.
Suitable for dry environments only RH < 90% and non-condensing environments.
NEVER allow scale to get wet.
The scale meets or exceeds all certification requirements within a temperature range
of 14 to 104 °F (-10 to +40 °C).
The scale should be placed out of direct sunlight and to provide adequate air circulation,
keep the area around the scale clear.
Do not place the scale directly in front of a heating or cooling vent. Such a location will
subject it to sudden temperature changes, which may result in unstable weight readings.
Insure that the scale has good, clean AC power and is properly grounded. In areas subject
to lightning strikes, additional protection to minimize lightning damage, such as surge
suppressors, should be installed.
Electrical Power
The icon Digital Clinical Scale has been designed to operate from a 100 to 240 VAC
50/60Hz 12 VDC 1A wall plug-in UL/CSA listed AC power adapter. Note that a special
order is not required for operation at 230 VAC.
The socket-outlet supplying power to the scale should be near the scale and should be
easily accessible.
On installations requiring 230 VAC power, it is the responsibility of the customer to
have a qualified electrician install the proper power adapter plug that conforms to
national electrical codes and local codes and ordinances.
Electrical Noise Interference
To prevent electrical noise interference, make certain all air conditioning and heating
equipment, lighting or other equipment with heavily inductive loads, such as welders, motors
and solenoids are on circuits separate from the system. Many of these disturbances can
seriously affect the operation of the system. These sources of disturbances must be
identified and steps must be taken to prevent possible adverse effects on the system.
Examples of available alternatives include isolation transformers, power regulators,
uninterruptible power supplies, or simple line filters.
3300-0192-0M Rev D 01/17 3
Page 10
UNPACKING
Carefully remove the icon Digital Clinical Scale from the shipping carton and inspect it for any
damage that may have taken place during shipment. Keep and use the original carton and packing material for return shipment if it should become necessary. The purchaser is
responsible for filing all claims for any damages or loss incurred during transit. Remove all
plastic wrapping, foam fillers and cardboard material from scale platform, display and other
components. You should have the following components:
Column with attached Display and Sonar Bracket
Column Cover
Scale Base
Scale Base Cover
Sonar Height Rod
AC Power Adapter
ASSEMBLY
TOOLS REQUIRED:
3/16” Allen wrench
#2 Phillips Screwdriver
1. Place the scale base on a level floor.
2. Insert the column cover onto the column.
3. Position the column over the opening in the scale base and then plug the modular
connector from the column into the socket in the scale base.
4. Insert the column into the scale base.
5. Using the 3/16” Allen wrench, evenly tighten the (4) Allen head screws on the back of
the scale base to secure column to scale base.
NOTE: The nuts on the front of the column opening in the scale base may turn
as you are tightening the Allen head screws.
This is normal. DO NOT use a wrench or other tool to hold the nut to keep it
from turning!
6. Slide the column cover down onto the base. Note that you may have to pull the cover
out slightly to clear the AC power jack.
7. Remove (6) screws securing sonar bracket to column and set bracket aside.
8. Align wire connector from the end of Sonar tube with the wire connector from the back
of column and plug together.
9. Insert the connector into the hole on back of column.
10. Place the sonar tube on back of the column and hold in place.
11. Place the bracket over the sonar tube and align the holes in the bracket with the holes
in the column.
IMPORTANT! Make sure the bend of the sonar tube is against the back throat of the
column.
12. Insert the screws removed earlier and using the #2 Phillips screwdriver, tighten the
screws to secure the sonar bracket to the column.
13. Place the scale base cover on the scale base and press down on both sides of the
cover. Note that a clicking sound will occur when the cover is snapped in place.
14. Scale is now ready for operation.
NOTE: For instructions on powering the scale using the AC power adapter or for
instructions on how to install batteries, refer to the INTERCONNECTIONS
section of this manual.
3300-0192-0M Rev D 01/17 4
Page 11
Step 1 Step 2 Step 3
Step 4 Step 5 Step 6
Step 7 Step 8 Step 9
Step 10 Step 11 Step 12
3300-0192-0M Rev D 01/17 5
Page 12
PLACING THE SCALE
For indoor use only.
Place scale on a flat, level floor or low-cut carpet away from heating and cooling vents.
Make certain the AC Power Adapter cord is out of the way of normal traffic to avoid a
trip hazard.
NEVER allow scale to get wet.
Check to make certain the scale is level. The level indicator is
located at the rear of the scale. If the scale is not level (the
bubble will not be centered), position the scale as required to
center the bubble and attain a level scale.
NOT LEVEL LEVEL
NOTE: Any time the scale is moved or re-located, be sure to check the level
bubble to ensure the scale is level before using.
INTERCONNECTIONS
The output and power connections to the scale are made on the rear of the scale base.
AC Power Adapter
To power the scale using the 12VDC wall plug-in UL/CSA listed AC power adapter, connect
the plug from the adapter into the power jack on the back of the scale base and then plug
the power adapter into the proper electrical outlet. On models requiring 230 VAC, it is the
customer’s responsibility to obtain the correct power adapter plug.
USB
The USB port on the scale is a device (or upstream) port and uses readily available cables
with the industry standard “Micro-B” connector.
The USB port may be connected to a computer for transmission of weight and associated
data to a PC-based EMR (electronic medical record) software program. The data can be
transmitted on demand (pressing the
computer.
Height Rod – Detecto Sonar Height Rod
The Detecto Sonar Height Rod connects to the scale by a snap-in modular connector
socket. Insert the modular connector of the Sonar Height Rod into the socket on the back of
the scale column it locks in place (a clicking sound will be heard when it is locked in place).
Batteries
The scale can use 12 "AA” size Alkaline, Ni-Cad or NiMH batteries (not included). You must
first obtain and install batteries before operations can begin. Batteries are contained in two
(2) battery holders inside the scale base. Access is via two (2) removable panels in the
scale base, accessible after removing the scale platform cover.
CAUTION! The scale can be operated from Alkaline, Ni-Cad or NiMH batteries. All
twelve (12) batteries must be of the same type. They must all be Alkaline, all Ni-Cad
or all NiMH. DO NOT mix Alkaline and Ni-Cad or NiMH batteries.
NOTE: The icon scale does not have a battery charging circuit. Should you wish to
use Ni-Cad or NiMH batteries, they must be fully charged before installing. When
discharged, Ni-Cad or NiMH batteries must be removed and placed in an external
charger to recharge.
key) or on receipt of a command from the
3300-0192-0M Rev D 01/17 6
Page 13
Battery Status
The battery status will be displayed when batteries are installed. A number will be displayed
in the upper right of the screen indicating the level of charge remaining in the batteries. For
example 100 = 100% charge, 75 = 75% charge, etc.
When the batteries are the point they need to be replaced (Alkaline) or recharged (NiCad or
NiMH) and the battery voltage drops too low for accurate weighing, the scale will
automatically shut off and you will be unable to turn it back on.
Using Alkaline Batteries
When the batteries are low enough that the scale turns off, remove the old batteries and
replace with new ones.
Using NiCad or NiMH Batteries
When the batteries are low enough that the scale turns off, remove the discharged
batteries and replace with fully charged ones. Place the discharged batteries in an
external charger to recharge.
Battery Installation/Replacement
To install or remove the batteries, the following steps should be followed:
1. Make sure the AC power adapter is unplugged.
2. Remove the platform cover from the scale base.
3. Referring to Figure No. 1, locate the two (2) rectangular panels in the scale base.
4. To install or replace the batteries, first remove each battery holder cover by pushing in
on the tab and lifting it up exposing the battery holders. See Figure No. 2 and 3.
5. If installing new batteries, proceed to step 6. If replacing the batteries, remove all 6
batteries from the battery holder and then proceed to step 6.
6. Referring to Figure No. 4, install six (6) "AA" size batteries in the each holder, noting the
polarity markings located in the battery holder.
7. After placing all six (6) batteries in each holder, replace the battery covers.
8. Install the platform cover on the scale base and press the key.
9. If display turns on, batteries have been installed correctly. If not, remove the platform
cover and then the battery covers and check for one or more improperly positioned
batteries.
10. The scale is now ready for operation.
3300-0192-0M Rev D 01/17 7
Page 14
UNITS SELECTION
When the scale is powered on for the first time, the display will prompt for which weighing and
height units to use, LB/IN or KG/CM.
1. Press
2. The scale display will show the software version for a few seconds, the Detecto logo
briefly and then change to the Select Units screen.
key to turn the scale on.
Select Units
LB/IN KG/CM
3. Press the LB/IN key at the bottom left of the screen to select pounds for weighing units
and inches for height measurement.
4. Press the KG/CM key on the bottom right of the screen to select kilograms for weighing
units and centimeters for height measurement.
5. The scale is now ready for operation.
6. Once the units setting is selected, it will be kept.
NOTE: If it is desired to change the units setting, refer to the SCALE SETUP
section, Setup Menu Page 1, Units: selection.
3300-0192-0M Rev D 01/17 8
Page 15
KEYPAD FUNCTIONS
DO NOT operate the keys or touchscreen with pointed objects (pencils, pens,
etc.). Damage to keys or touchscreen resulting from this practice is NOT
covered under warranty.
This is the Power key. With the scale off, pressing this key will apply power to
it and turn on the display. If the scale is already on, pressing the key will turn it
off.
This is the Zero key. Press and release this key to reset the display to zero, up
to 100% of the scale capacity.
This is the Lock/Release key. Press and release this key, or touch the weight
readout on the display to cause the weight, height, and BMI to lock on the
current values until the key is pressed again or the weight readout on the
display is touched again. While the weight is locked, the weight status will
change to LOCKED in order to denote that the weight is being held.
This is the ID key. Press and release this key to open the ID entry screen
where you can use the touch screen to enter in a numeric ID. This ID is NOT
used for tracking patient measurements. The ID that is entered is only
transmitted out of the communication port when the
change or edit the ID, simply press the ID key again to return to the ID entry
screen.
This is the Display Mode key. Press and release this key to cause the display
to cycle between display modes: Weight/Height/BMI or Weight only.
key is pressed. To
3300-0192-0M Rev D 01/17 9
Page 16
This is the Menu key. Press and release this key to launch the Menu screen.
The first screen that is displayed is the Setup Review screen where you can
see all of the settings at a glance. To continue into setup, touch the Setup key
at the bottom right of the touch screen.
This is the Enter key. Press and release this key to signal completion of the
entry of data and cause the scale to output the current weight, height, and
BMI for data capture for EMR/EHR purposes. The options include Cardinal
SMA, Welch Allyn, Tanita, or IEEE 11073-10415 formats.
3300-0192-0M Rev D 01/17 10
Page 17
ANNUNCIATORS
The annunciators are displayed on the Weight screen to show that the scale is in the mode
corresponding to the annunciator label or that the status indicated by the label is active.
WEIGHT: POUNDS/GROSS AC
Status
Messages
Weight
Display
Weight Mode
CENTER ZERO
Power
Status
(AC or ##)
0.0
HEIGHT: IN BMI
Height
Display
AC
This is shown on the Weight screen to indicate that the scale is powered by the AC adapter.
##
This is shown on the Weight screen to indicate that the scale is powered by batteries and
the percentage of their remaining charge.
POUNDS/GROSS
This is shown on the Weight screen to indicate that the weight displayed is gross weight in
pounds.
KILOGRAMS/GROSS
This is shown on the Weight screen to indicate that the weight displayed is gross weight in
kilograms.
CENTER ZERO
This is shown on the Weight screen to indicate that the weight is within +/- 1/4 division of the
center of zero.
LOCK
This is shown on the Weight screen to indicate that the scale is locked onto the displayed
weight. In operation after obtaining a stable weight value, pressing the
the scale to lock onto the weight and turn on the annunciator. Pressing the
second time will unlock the display and turn off the annunciator.
IN
This is shown on the Weight screen to indicate that the displayed height measurement is in
inches.
CM
This is shown on the Weight screen to indicate that the displayed height measurement is in
centimeters.
BMI
This is shown on the Weight screen to indicate the calculated body fat (Body Mass Index).
0.00.0
BMI
Display
key will cause
key a
3300-0192-0M Rev D 01/17 11
Page 18
OPERATION
ALWAYS assist the patient in stepping on and off the scale platform to
ensure they do not fall. NEVER leave a patient unattended while they are on
the scale platform. Failure to maintain control of the patient at all times can
result in serious injury to you and/or the patient.
Zero Weight Display
1. In Gross Weight mode (POUNDS/GROSS or KILOGRAMS/GROSS shown on weight
screen), if scale is not showing zero weight on weight screen press 0 key.
2. Weight screen will return to zero (0.0) and CENTER ZERO will be displayed to show
that the scale is ready for use.
Basic Operation
To Weigh
1. Press
2. If required, press 0 key to zero weight screen.
3. Assist patient on scale
4. When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup. NOTE: If more time is needed,
press the
5. Read weight display.
6. If desired, press the
7. Assist patient off scale.
Basic Operation with ID – (No BMI)
To Weigh
1. Press
2. If required, press 0 key to zero weight screen.
key to turn scale on.
key to hold the locked weight reading.
key to output EMR/EHR data.
key to turn scale on.
3. Press
4. Display will change to show PATIENT ID screen.
5. Using numeric keys, enter up to an 8 digit numeric identification number.
6. Press Save key.
7. Assist patient on scale
8. When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup. NOTE: If more time is needed,
press the
9. Read weight display.
10. If desired, press the
11. Assist patient off scale.
key.
key to hold the locked weight reading.
key to output EMR/EHR data.
3300-0192-0M Rev D 01/17 12
Page 19
Body Mass Index (BMI) Operation – (No ID)
To Weight, Measure Height and Calculate BMI
1. Press
key to turn scale on.
2. If required, press 0 key to zero weight screen.
3. Assist patient onto scale.
NOTE: In order to obtain an accurate height measurement, instruct patient to
look straight ahead and not down at scale display.
4. When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup. NOTE: If more time is needed,
press the
key to hold the locked weight reading.
5. Scale will have completed measuring patient’s height and calculating their BMI.
6. Read weight, height and BMI displayed.
7. If desired, press the
key to output EMR/EHR data.
8. Assist patient off scale.
Body Mass Index (BMI) Operation with ID
To Weight, Measure Height and Calculate BMI
1. Press
2. If required, press 0 key to zero weight screen.
key to turn scale on.
3. Press
key.
4. Display will change to show PATIENT ID screen.
5. Using numeric keys, enter up to an 8 digit numeric identification number.
6. Press Save key.
7. Assist patient onto scale.
NOTE: In order to obtain an accurate height measurement, instruct patient to
look straight ahead and not down at scale display.
8. When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup. NOTE: If more time is needed,
press the
key to hold the locked weight reading.
9. Scale will have completed measuring patient’s height and calculating their BMI.
10. Read weight, height and BMI displayed.
11. If desired, press the
key to output EMR/EHR data.
12. Assist patient off scale.
3300-0192-0M Rev D 01/17 13
Page 20
Body Mass Index (BMI) Operation
Manually Entering Known Height
1. Press
2. If required, press 0 key to zero weight screen.
3. Press HEIGHT on display screen.
4. If POUNDS/INCHES was selected for Units: in setup, display will change to show
HEIGHT IN INCHES on screen.
a. Use numeric keys to enter up to 3 digits and decimal point (# #.#) for height in
5. If KILOGRAMS/CENTIMETERS was selected for Units: in setup, display will change to
show HEIGHT IN CENTIMETERS on screen.
a. Use numeric keys to enter up to 2 digits and decimal point (#.#) for height in
6. Assist patient onto scale.
7. When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup. NOTE: If more time is needed,
press the
8. Scale will have completed calculating patient’s BMI.
9. Read weight and BMI displayed.
10. If desired, press the
11. Assist patient off scale.
key to turn scale on.
inches and then press Save key.
centimeters and then press Save key.
key to hold the locked weight reading.
key to output EMR/EHR data.
Body Mass Index (BMI) Operation with ID
Manually Entering Known Height
1. Press
2. If required, press 0 key to zero weight screen.
key to turn scale on.
3. Press
4. Display will change to show PATIENT ID screen.
5. Using numeric keys, enter up to an 8 digit numeric identification number.
6. Press Save key.
7. Press HEIGHT on display screen.
8. If POUNDS/INCHES was selected for Units: in setup, display will change to show
HEIGHT IN INCHES on screen.
a. Use numeric keys to enter up to 3 digits and decimal point (# #.#) for height in
9. If KILOGRAMS/CENTIMETERS was selected for Units: in setup, display will change to
show HEIGHT IN CENTIMETERS on screen.
a. Use numeric keys to enter up to 2 digits and the decimal point (#.#) for height in
10. Assist patient onto scale.
11. When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup. NOTE: If more time is needed,
press the
12. Scale will have completed calculating patient’s BMI.
13. Read weight and BMI displayed.
14. If desired, press the
15. Assist patient off scale.
key.
inches and then press Save key.
centimeters and then press Save key.
key to hold the locked weight reading.
key to output EMR/EHR data.
3300-0192-0M Rev D 01/17 14
Page 21
SCALE SETUP
Your icon Digital Clinical Scale has been pre-configured at the factory and should not require
changes for use in most applications. However, if the factory settings do not meet the
requirements of your operation, the following describes the setup process for your scale.
NOTE: The keypad and touchscreen are not to be operated with pointed
objects (pencils, pens, fingernails, etc.). Damage to the keypad or touchscreen
resulting from this practice will NOT be covered under warranty.
To Enter Setup
1. Press key to turn the scale on.
2. The scale will show the Detecto logo briefly and then change to the Weight screen.
3. With the Weight screen displayed, press the Menu key.
4. The display will change to show the Setup Review screen.
MENU PAGE 0
Setup Review
Audit Trail – Cal=16, Cfg=30
Range 1: 600.0 x 0.2
Range 2: 1000.0 x 0.5
Weight Units=lb Height Units=IN
Auto Lock=7 Sec Sonar Ht=223.3 cm
Motion=3 div. Filter=Off
ZeroTrack=10/2 d Power up zero =Yes
Gravity=1.000000 OIML=No
Exit Setup
5. Press Setup on the touchscreen.
6. The display will change to show the ENTER PASSWORD screen.
7. Using the 10-key on the touchscreen, enter 64870 and then press the Save key.
8. The scale is now ready for setup and calibration
Menu Functions
After pressing the Menu key to enter the Setup menu, the first screen of the menu is a
Setup Review screen where all of the settings can be reviewed at a glance. The Audit
Trail Counters for the metrological settings and the non-metrological settings are available
for review on this screen as well. Each prompt in the menu screens will show the current
value of the setting next to the prompt to easily identify what the current setting for that
parameter is.
Setup Menu Keys
Press the Setup key on the Setup Review screen to enter into the scale configuration.
Press the >>> key on the bottom right of the Setup Menu screen to advance to the next
menu page.
Press the <<< key at the bottom left of the Setup Menu screen to return to the previous
menu page.
Press the Exit key to exit Setup and return to the Weight screen.
3300-0192-0M Rev D 01/17 15
Page 22
Setup Menu Page 1
MENU PAGE 1
Cal=16, Cfg=30
Units: lb, in
Auto Locking: 7 sec.
Sonar Ht: 223.3 cm
<<< Exit >>>
Cal=XX, Cfg=YY – Audit Trail Counters
These are the Audit Trail Counters for the number of times that the metrological and nonmetrological settings have been changed. This is for information purposes only.
Units:
Touching this key will toggle the weighing and height units between (lb, in) or (kg, cm).
Note that the units cannot be toggled if OIML is enabled.
NOTE: When you calibrate the scale, ensure that the proper base units are selected
here before attempting to calibrate. It is safe to change the units without calibrating, just
be sure not to change any other metrological settings when you are using the converted
units as this may place your calibration to an invalid state.
Allowable settings are: lb, in (pounds, inches) or kg, cm (kilograms, centimeters)
Auto Locking:
This setting is used by the scale to hold a stable patient weight for a desired amount of time.
For example, if a value of 5 seconds is used, then when the scale locks onto a stable
patient weight, it will remain locked for 5 seconds before automatically releasing the weight.
Touching this key will open the entry screen for the auto-locking feature of the scale. The
value entered is a time in seconds. Using the numeric keys, enter a new value and then
press the Save key to save it.
Allowable values for auto locking are: 0 to 60.
Sonar Ht:
This is the height in centimeters from the scale platform cover to the bottom of the sonar
sensor enclosure. This value is used to determine the patient’s height and should be
measured and entered as accurately as possible to guarantee accurate height
measurements using the sonar height rod.
Touching this key will open the entry screen for the sonar sensor height. The value entered
is in centimeters. Using the numeric keys, enter a new value and then press the Save key
to save it.
Press the >>> key to advance to the Setup Menu Page 2.
Press the <<< key to return to the Setup Review screen.
Press the Exit key to exit Setup and return to the Weight screen.
Touching this key will open the capacity screen. This will allow you to set the capacity of
the first range. Using the numeric keys, enter a new value and then press the Save key to
save it.
The first range of the scale should be set to 600.0.
Capacity2:
Touching this key will open the capacity entry screen for the second weight range. This
value MUST be greater than Capacity1 if used. Using the numeric keys, enter a new value
and then press the Save key to save it.
The second range of the scale should be set to 1000.0.
NOTE: To disable the second weight range, set Capacity2 to a value of (0) zero.
Interval1:
Touching this key will open the interval entry screen. This will allow you to set the scale
interval for the first range. Using the numeric keys, enter a new value and then press the
Save key to save it.
Allowable values for interval 1 are” 1, 2, or 5.
Interval2:
If a second range capacity has been set, then Interval 2 will be used as the interval of the
second weight range if enabled (see Capacity2). Using the numeric keys, enter a new
value and then press the Save key to save it.
Allowable values for interval 1 are: 1, 2, or 5.
Press the >>> key to advance to the Setup Menu Page 3.
Press the <<< key to return to the Setup Menu Page 2.
Press the Exit key to exit Setup and return to the Weight screen.
Touching this key will open the decimal entry screen. This will allow you to set the decimal
point precision for the first weight range. Using the numeric keys, enter a new value and
then press the Save key to save it.
Allowable values for decimal point precision are: 0 – 3.
Decimal2:
Touching this key will open the decimal entry screen. This will allow you to set the decimal
point precision for the second weight range if enabled (see Capacity2). Using the numeric
keys, enter a new value and then press the Save key to save it.
Allowable values for decimal point precision are: 0 – 3.
Motion Range:
Touching this key will open the motion range entry screen. This will allow you to set the
number of scale divisions of movement that will be allowed for stable. Using the numeric
keys, enter a new value and then press the Save key to save it.
Allowable values for motion are: 0 – 20.
Filter Mode:
Pressing this key will open the weight filtering screen. This will allow you to set the amount
of digital filtering being applied to the scale. Using the numeric keys, enter a new value and
then press the Save key to save it.
Allowable settings are: 0 = Off, 1 = Minimal, 2 = Moderate, and 3 = Maximum
Press the >>> key to advance to the Setup Menu Page 4.
Press the <<< key to return to the Setup Menu Page 3.
Press the Exit key to exit Setup and return to the Weight screen.
3300-0192-0M Rev D 01/17 18
Page 25
Setup Menu Page 4
MENU PAGE 4
Zero Tracking: 10
Power up Zero: Yes
Gravity: 1.000000
OIML: No
<<< Exit >>>
Zero Tracking:
Touching this key will open the zero tracking screen. This will allow you to set the number
of half (1/2) divisions that the scale will attempt to maintain zero. Using the numeric keys,
enter a new value and then press the Save key to save it.
Allowable settings for zero tracking are: 0 – 100.
Power up Zero:
Touching this key will toggle the power up zeroing of the scale to Yes or No. If enabled, this
will cause the scale to attempt to zero the scale on power up.
Allowable values for Power up Zero are: Yes or No.
Gravity:
Gravity compensation accounts for latitudes and elevations that are different from where the
scale was calibrated at. In order to calculate the value for this parameter, use the
gravitational constant of the location where the scale was calibrated divided by the
gravitational constant of where the scale will be installed:
Gravitational Constant (Calibration location)
---------------------------------------------------------- = value
Gravitational Constant (Operation location)
This should give you a value close to 1 that you can enter in to compensate for variation in
gravity due to elevation/latitude.
Touching this key will open the gravity compensation entry screen. Using the numeric keys,
enter a new value and then press the Save key to save it.
NOTE: If you do not wish to use the compensation feature, it must be set to 0 (zero) in order
to disable it.
OIML:
Touching this key will toggle the OIML setting to Yes or No. The prompt window will show
the current state of the OIML setting. This setting should be enabled for all EU models to
comply with regulations.
Allowable values for OIML are: Yes or No.
Press the >>> key to advance to the Setup Menu Page 5.
Press the <<< key to return to the Setup Menu Page 4.
Press the Exit key to exit Setup and return to the Weight screen.
3300-0192-0M Rev D 01/17 19
Page 26
Setup Menu Page 5
MENU PAGE 5
Model: Icon
Calibrate Scale
<<< Exit
Model:
This will change the default settings and operation of the scale. For instance, select “Icon”
as the model in order to use the stand-up type scale with the sonar height measurement.
For wheelchair type scales with tare functionality, select “Wheelchair” as the model.
Touch this key to open the model selection screen. Using the numeric keys, enter a new
value and then press the Save key to save it.
Allowable settings are: 0 = icon, 1 = Wheelchair and 2 = Other
Calibrate Scale
Touching this key will allow for calibration of the scale. The scale display will first prompt for
the loaded calibration weight. Enter in the weight that will be used to calibrate the scale and
touch the Save key. The scale will perform the loaded calibration and then prompt for you
to unload the scale. Make sure all weight has been removed from the scale before
selecting Save to perform the unloaded calibration.
You will be returned to the menu when calibration is complete.
Press the <<< key to return to the Setup Menu Page 4.
Press the Exit key to exit Setup and return to the Weight screen.
3300-0192-0M Rev D 01/17 20
Page 27
SCALE CALIBRATION
To Enter Calibration Setup
1. Enter the Setup menu of the scale by pressing the Menu key.
2. Enter in the desired setup parameters for the scale.
3. Select Calibrate Scale from the Setup Menu Page 5.
4. Make certain the scale platform is empty and free of debris.
5. Place the desired amount of calibrated test weights on the scale platform. A minimum
of 50% of scale's capacity is required. However 70% to 100% is recommended.
6. Enter the amount of the test weight into the Calibration menu screen and touch the
Save key.
7. Wait while the scale performs the loaded calibration.
8. When the loaded calibration is complete, the scale display will prompt to unload the
scale.
9. Remove any weight from the scale platform and touch the Save key.
10. Wait while the scale performs the zero calibration.
11. The scale display will return to the Setup Menu screen when the calibration is
complete.
Press the Exit key to exit Setup and return to the Weight screen.
3300-0192-0M Rev D 01/17 21
Page 28
This page intentionally left blank.
3300-0192-0M Rev D 01/17 22
Page 29
ERROR AND OPERATION MESSAGES
The icon scale is equipped with diagnostic software that tests various portions of the scale's
circuitry and verifies proper operation. Should a problem be detected, an error or status
message will be displayed. The following lists these messages and their meaning.
AUTO LOCKED
This message appears if the weight/height has been automatically locked.
BELOW ZERO
The scale weight reading is below zero.
CAL REQUIRED
The scale requires calibration, weight will show as dashes. Consult your scale service
representative.
CENTER ZERO
The scale weight reading is at center of zero.
ERROR CANNOT ZERO
Scale cannot zero due to motion on scale.
LOCKED
This message appears if the weight/height has been manually locked.
MOTION
The scale is in motion based on motion range setting.
OVER CAPACITY
The scale weight exceeds scale capacity.
OVERFLOW ERROR
This message appears if there are too many characters to display on the screen. This
message appears if there are too many characters to display on the screen.
3300-0192-0M Rev D 01/17 23
Page 30
BEFORE YOU CALL FOR SERVICE
Problem Possible Solutions
Display does not turn on AC Operation:
Is AC power supply fully inserted into wall receptacle?
Check wall receptacle for proper AC power. Try another
electrical appliance in same receptacle, does it work?
Check circuit breaker.
Has there been power failure?
Battery operation:
Check if batteries are installed and correctly.
If Alkaline, remove old batteries and replace with new ones.
If NI-CAD or NiMH, remove discharged batteries and
replace with fully charged ones. Place discharged batteries
in an external charger to recharge.
Incorrect weight is
displayed
Weight is not displayed Refer to Error and Operation Messages.
Insure that scale platform isn't touching an adjacent object.
Have proper operation procedures been followed?
3300-0192-0M Rev D 01/17 24
Page 31
CARE AND CLEANING
NOTE: The scale contains no user-serviceable parts and maintenance should be
limited to an occasional cleaning and battery replacement as required.
Do not submerge the scale in water, pour or spray water directly on it to clean. The
scale is not waterproof and covering it with water will damage it and void the
warranty.
Always remove power before cleaning.
Do not use wire brushes, abrasives, or cleaning tools such as steel pads and
scrapers, which will scratch the painted surface. Instead, use soft cloths or plastic
scouring pads for cleaning.
When possible use treated water. Hard water can leave behind deposits. Soft water
is much gentler on the painted steel’s surface.
Avoid the use of acetone, thinner or other volatile solvents and abrasive type
cleaners for cleaning. If required, a mild solvent such as mineral spirits can be used
to remove oil, grease, tars, wax, and similar substances. Use a cloth dampened with
mineral spirits and apply only to areas that are contaminated. Follow up the use of
this mild solvent with detergent cleaning and rinsing.
#4-24 X 1/4, PHIL. DR. Z-PLATE
6 1 6021-1509 SCW PAN-HEAD, SHEET METAL #8X.75
7 1 6650-1114 O-RING 2 ID X 2-3/16 OD X 3/32 THK
8 A/R 6710-1021 TAPE 1.88” X 60YDS DUCT ROLL
23 A/R 6560-1125 ADHESIVE LOCTITE 290 THRD LCKR, WICKING
24 1 6690-0001 LEVEL, "BULLSEYE" TYPE
25 A/R 6710-1021 TAPE 1.88” X 60YDS DUCT ROLL
26 1 593GR986 SERIAL TAG ASSY
27 4 FP-250 LOAD CELL, FLAT PLATE, 250 LBS
NUT #10-32 HEX Z/P
VIBRATION MOUNT, 1/4” – 20 MALE-FEMALE, 1” OD,
1/2” BASE HEIGHT
3300-0192-0M Rev D 01/17 30
Page 37
PARTS IDENTIFICATION
Base Sub-Assembly
[SH1][SH2]
3300-0192-0M Rev D 01/17 31
Page 38
STATEMENT OF LIMITED WARRANTY
Detecto Scale warrants its equipment to be free from defects in
material and workmanship as follows: Detecto warrants to the
original purchaser only that it will repair or replace any part of
equipment which is defective in material or workmanship for a
period of two (2) years from date of shipment. Detecto shall be
the sole judge of what constitutes a def ect.
During the first ninety (90) days Detecto may choose to replace
the product at no charge to the buyer upon inspection of the
returned item.
After the first ninety (90) days, upon inspection of the returned
item, Detecto will repair or replace it with a remanufactured
product. The customer is responsible for paying for t he freight
both ways.
This warranty does not apply to peripheral equipment not
manufactured by Detecto; this equipment will be covered by
certain manufacturer’s warranty only.
This warranty does not include replacement of expendable or
consumable parts. This does not apply to any item which has
deteriorated or damaged due to wear, accident, misuse, abuse,
improper line voltage, overloading, theft, lightning, fire, water or
acts of God, or due to extended storage or exposure while in
purchaser’s possession. This warranty does not apply to
maintenance service. Purchased parts will have a ninety (90) day
repair or replacement warranty only.
Detecto may require the suspect product to be returned to the
factory; item(s) must be properly packed and shipping charges
prepaid. A return authorization number must be obtained for all
returns and marked on the outside of all returned packages.
Detecto accepts no responsibility for loss or damage in transit.
3300-0192-0M Rev D 01/17 32
Page 39
STATEMENT OF LIMITED WARRANTY
Conditions Which Void Limited Warranty
This warranty shall not apply to equipment which:
A.) Has been tampered with, defaced, mishandled or has had
repairs and modifications not authorized by Detecto.
B.) Has had serial number altered, defaced, or removed.
C.) Has not been grounded according to Detecto’s recommended
procedure.
Freight Carrier Damage
Claims for equipment damaged in transit must be referred to the
freight carrier in accordance with freight carrier regulations.
This warranty sets forth the extent of our liability for breach of any
warranty or deficiency in connection with the sale or use of the
product. Detecto will not be liable for consequential damages of
any nature, including but not limited to, loss of profit, delays or
expenses, whether based on tort or contract. Detecto reserves
the right to incorporate improvements in material and design
without notice and is not obligated to incorporate improvements in
equipment previously manufactured.
The foregoing is in lieu of all other warranties, express or implied
including any warranty that extends beyond the descripti on of the
product including any warranty of merchantability or fitness for a
particular purpose. This warranty covers only those Detecto
products installed in the forty-eight (48) contiguous continental
United States.
08/27/2014
Printed in USA
D268-WARRANTY-DET-A
3300-0192-0M Rev D 01/17 33
Page 40
Printed in USA
3300-0192-0M Rev D 01/17
3300-0192-0M Rev D 01/1734