Desco TB-5512 Installation Manual

© 2009 DESCO INDUSTRIES INC.
Employee Owned
TB-5512 January 2009 Page 1 of 5
Bench Top Ionizer Installation, Operation and Maintenance
Figure 1. 94020 Bench Top Ionizer
Description
The SPI Bench Top Ionizer (SP4511) is a compact and lightweight steady state DC auto-balancing bench top ionizer. The unit is normally placed at one end of the workbench or area to be neutralized. It may also be mounted to a wall or shelf. The Ionizer's neutralization discharge time will be best approximately 12" to 48" directly in front of the unit and will increase as the distance from the unit increases.
Ionizers are useful in preventing electrostatic charge generation, ElectroStatic Discharge, ElectroStatic Attraction, as well as preventing equipment latch-up and safety related shock. Per ANSI/ESD S20.20 section 6.2.3.1. Protected Areas Requirement states: "Ionization or other charge mitigating techniques shall be used at the workstation to neutralize electrostatic fields on all process essential insulators if the electrostatic field is considered a threat." Air ionization can neutralize the static charge on insulated and isolated objects by producing separate charges in the molecules of the gases of the surrounding air. When an electrostatic charge is present on objects in the work environment, it will be neutralized by attracting opposite polarity charges from the ionized air. Note that ionization systems should not be used as a primary means of charge control on conductors or people. (Reference: EN 61340-5-2:1paragraph 5.2.9)
"The primary method of static charge control is direct connection to ground for conductors, static dissipative materials, and personnel. Acomplete static control program must also deal with isolated conductors that cannot be grounded, insulating materials (e.g., most
common plastics), and moving personnel who cannot use wrist or heel straps or ESD control flooring and footwear.
Air ionization is not a replacement for grounding methods. It is one component of a complete static control program. Ionizers are used when it is not possible to properly ground everything and as backup to other static control methods. In clean rooms, air ionization may be one of the few methods of static control available." (ESD Handbook TR20.20 Ionization, section 5.3.6.1 Introduction and Purpose / General Information)
The SPI Bench Top Ionizer operates on DC, steady state. Steady DC systems consist of separate negative and positive ion emitters connected by a pair of high­voltage cables to their respective high-voltage power supplies. The spacing between emitters varies depending on the design, and DC power is constantly applied to the emitter points.
Ionizer Selection
ANSI/ESD S20.20 paragraph 6.1.1.2. ESD Control Program Plan Guidance states: "The Plan should include a listing of the specific type of ESD protective materials and equipment used in the Program." When selecting an ionizer, life cycle costs should be considered, including:
equipment cost
installation cost
operation and maintenance cost
SPI Ionizers meet the ANSI/ESD S20.20 minimum recommended technical requirement range of less than +/- 50 volts voltage offset tested in accordance with ANSI-EOS/ESD S3.1. All SPI Bench Top Ionizers greatly exceed the requirement providing ±5 to ±25 volt auto-balancing.
Packaging
1 Bench Top Ionizer 1 Power Cord (94020 only) 1 Emitter Point Cleaner Pack 1 Certificate of Calibration
Made in America
TECHNICAL BULLETIN TB-5512
ESD Systems.com • 432 Northboro Road Central • Marlboro, MA 01752 • (508) 485-7390 • Fax (508) 480-0257 • Website: ESDSystems.com
ESD Systems.com • 432 Northboro Road Central • Marlboro, MA 01752 • (508) 485-7390 • Fax (508) 480-0257 • Website: ESDSystems.com
© 2009 DESCO INDUSTRIES INC.
Employee Owned
TB-5512 January 2009 Page 2 of 5
Installation
Place the unit at a desired location where that the airflow will not be restricted. Be sure that the ON/OFF switch located on the rear of the unit is in the OFF position. Plug the power cord into the unit and then into the appropriate AC power source.
Operation Air Injection Test
I. Set the fan speed switch on the rear of the unit to the LOW, MED, or HI position (See Figure 1). Higher airflow will result in faster neutralization rates. II. Position the ionizer so that maximum airflow is directed towards the items or area to be neutralized.
III. Turn the unit ON. When the unit is first turned on, it conducts a self-test. The audible alarm will sound and the LED will cycle through the colors red, yellow, and green. The LED will remain green during normal operation.
Figure 2. Rear View of the Bench Top Ionizer
Maintenance
"All ionization devices will require periodic maintenance for proper operation." Maintenance intervals for ionizers vary widely depending on the type of ionization equipment and use environment. Critical clean room uses will generally require more frequent attention. It is important to set-up a routine schedule for ionizer service. Routine service is typically required to meet quality audit requirements." (ESD Handbook TR 20.20 paragraph 5.3.6.7 Maintenance / Cleaning)
EIA-625, recommends checking ionizers every 6 months, but this may not be suitable for many programs particularly since an out-of-balance may exist for months before it is checked again. The SPI Bench Top Ionizer has a very desirable feature as it will provide visual and audible alarms when an out of balance exists. ANSI/ESD S20.20 paragraph 6.1.3.1 Compliance Verification Plan Requirement states: "Test equipment shall be selected to make measurements of appropriate properties of the technical requirements that are incorporated into the ESD program plan."
Calibration AND Adjustments
Balance Adjustment
The SPI Bench Top Ionizer is an auto-balancing unit. However, tuning or manual adjustment can be accomplished by inserting a small screwdriver or trimmer adjustment tool into the balance adjustment hole located at the rear of the unit (See Figure 1). To increase the output in a positive direction, turn the potentiometer clockwise. To increase the output in a negative direction, turn the potentiometer counter­clockwise.
Maintenance / Alarms
WARNING - RISK OF ELECTRIC SHOCK
THESE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY. DO NOT PERFORM ANY SERVICING OF INTERNAL PARTS UNLESS YOU ARE QUALIFIED TO DO SO. NOTE: The AC power cord MUST always be disconnected before the unit is disassembled.
The input voltage may be verified or reset by removing the 3 screws located on the back of the unit then removing the back case.
The input voltage can be selected using the two internal jumpers shown in Figures 2 and 3.
If the supply voltage drops from 110 Volts to below 85
Figure 3. 110 Volt Jumper Setting
Figure 4. 220 Volt Jumper Setting
Loading...
+ 3 hidden pages