Desco TB-2082 Installation Manual

TECHNICAL BULLETIN TB-2082
Monitor Calibration Unit Operation Instructions
Figure 1. Desco 98220 Monitor Calibration Unit
The Desco 98220 Monitor Calibration Unit is designed to
simplify the process of calibrating Workstation Continuous Monitors. The unit allows the user to quickly and easily verify whether a tester is operating within specifications. The Monitor Calibration Unit is a passive device and requires no power source.
Made in America
Calibrating the Full Time Continuous Monitor
The Full Time Continuous Monitor is available as following item number:
Item Voltage
19225 120 VAC
TESTING THE OPERATOR CIRCUIT Step 1: Connect the Calibration Unit’s banana plug wire
labeled “GROUND” to a known ground. Do not plug the ground lead from the Calibration Unit to the monitor’s common point ground banana jack. There is a 47 kilohm resistance to ground.
Step 2: Insert the Calibration Unit’s banana plug labeled
“OPERATOR” into the jack labeled “OPERATOR” on the Full Time Continuous Monitor (see Figure 2). Power the monitor.
“A Compliance Verification Plan shall be established to ensure the Organization’s fulfillment of the technical requirements of the ESD Control Program Plan (ANSI/ ESD S20.20-2007 section 7.3). The product qualification for the Continuous Monitors per ANSI/ESD S20.20-2007 Table 3 is “User defined” test method and required limits, and compliance verification is per ESD TR53 with the required limit being “Manufacturer defined.” Per ESD TR53 Compiance Verification “Compliance verification of the constant monitor device should follow the manufacturer’s instructions.”
Packaging
1 Monitor Calibration Unit 1 Alligator Clip 1 Stacking Snap Banana Jack Adapter 1 Banana Jack to Wire Adapter 1 Certicate of Calibration
Calibration Test Procedures
The Monitor Calibration Unit was specifically designed for use in calibration of Desco brand equipment. The following step by step procedures will cover calibration for specific test units. The procedures will not cover adjustment of the test equipment. For detailed information regarding adjustment of specific Desco brand testers contact our Customer Service Department at 909-627-8178.
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Website: Desco.com
Figure 2. Connecting the OPERATOR test lead from the Monitor Calibration Unit to the Full Time Continuous Monitor
Step 3: Start the Calibration Unit’s selector knob at the full
counter-clockwise position. When testing the OPERATOR circuit, test only the first four limits on the Calibration Unit. Rotate the selector knob to the “LOW FAIL” position; the red “OPERATOR GROUND” LED on the monitor should illuminate and alarm.
TB-2082 Page 1 of 6
Rev. December 2009
© 2009 DESCO INDUSTRIES, INC.
Employee Owned
Step 4: Rotate the selector knob clockwise to the “LOW
LIMIT” position of the green pass sector. The monitor’s green “OPERATOR GROUND” LED should illuminate.
Step 5: Rotate the selector knob clockwise to the “HIGH
PASS” position at the end of the green pass sector. The monitor’s green “OPERATOR GROUND” LED should remain illuminated.
Step 6: Rotate the selector knob clockwise to the “HIGH
FAIL” position. The green “OPERATOR GROUND” LED should turn off and the red LED should illuminate and the alarm should sound.
Step 7: Disconnect the Monitor Calibration Unit from the
monitor.
TESTING THE MAT CIRCUIT Step 1: Connect the Calibration Unit’s banana plug wire
labeled “GROUND” to a known ground.
Step 2: Connect the included banana jack to wire adapter to
the Calibration Unit’s banana plug labeled “MAT.” Insert the wire end of the adapter to the monitor’s green terminal block labeled “MAT” (see Figure 3).
Calibrating the Dual Operator Continuous Monitor with Two Satellites
The Dual Operator Continuous Monitor with Two Satellites is available as following item numbers:
Item Voltage
19232 120 VAC 19233 220 VAC
Note: If calibrating discontinued items 19208 or 19209,
disconnect the remote jacks (if used) from the unit. Remote jacks may cause intermittencies in calibration. Leave the matt connection to the 19208 or 19209 in place.
TESTING THE OPERATOR CIRCUIT Step 1:
labeled “GROUND” to a known ground.
Step 2: Insert the Calibration Unit’s banana plug labeled
“OPERATOR” into the jack labeled “OPERATOR” on the satellite remote (see Figure 4).
Connect the Calibration Unit’s banana plug wire
Figure 3. Connecting the MAT test lead from the Monitor Calibration Unit to the Full Time Continuous Monitor
Step 3: Rotate the Calibration Unit’s knob switch to the MAT
FAIL 10M position. The monitor’s red “MAT GROUND” LED should illuminate and the alarm should sound.
Step 4: Rotate the selector knob counter-clockwise to the
MAT PASS 10M position. The monitor’s “MAT GROUND” green LED should illuminate.
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Website: Desco.com
TB-2082 Page 2 of 6
Figure 4. Connecting the OPERATOR test lead from the Monitor Calibration Unit to the satellite remote
Step 3: Start the Calibration Unit’s selector knob at the full
counter-clockwise position and power the monitor. When testing the OPERATOR circuit, test only the first four limits on the Calibration Unit. Rotate the selector knob to the “LOW FAIL” position; the corresponding operator red LED on the monitor should illuminate and alarm.
Step 4: Rotate the selector knob clockwise to the “LOW
LIMIT” position of the green pass sector. The monitor’s corresponding green “OPR1” or “OPR2” LED should illuminate.
© 2009 DESCO INDUSTRIES, INC.