Desco 19430 Installation Manual

TECHNICAL BULLETIN TB-3048
Digital Static Locator Installation, Operation, and Maintenance
The POWER button turns the Digital Static Locator on and off. If the 19430 is left on for 5 minutes, it will automatically turn itself off. To resume operations, press the red POWER button once. To disable the AutoOff feature and allow continuous operation, press the red Power button and hold down for at least 3 seconds when turning the Digital Static Locator on.
The HOLD button holds the display of a measured charge. HOLD mode turns off the ring range lights. No measurements are possible when in the HOLD mode. The hold measurement feature enables the user to use
Figure 1. Digital Static Locator Item #19430
Description
Desco’s Digital Static Locator is a compact
static charges.
The Digital Static Locator measures static voltages within ±20kV (20,000V) at a distance of 1". Results are simultaneously displayed numerically and in bar graph format. POWER on/off, ZERO adjustment, Ion Balance (IB) and HOLD are all push button operation. The HOLD button allows the display to retain the static charge reading. This is especially useful where display is difcult to see during measurement.
the eldmeter in inaccessible areas. When the HOLD button is pressed once during a measurement, the digital numerical value and the bar graph will be held on the display. This function allows the operator to move the eldmeter where it can be more easily read. The HOLD mode is displayed above the battery capacity display. To cancel HOLD mode and resume measurement, press HOLD once more.
The IB button and prepares the eldmeter for Ion Balance (offset voltage) measurement.
The ZERO button sets the display to a zero static charge. When the digital meter is turned on, the zero adjustment is inoperative if it is in HOLD mode.
Made in America
Two LED guide ring range lights help position the eldmeter at the right distance from a charged surface. The conductive case and ground snap facilitate grounding for accurate measurement. The circuitry of the 19430 has been designed to make measurements in areas using air ionization.
Packaging
1 Digital Static Locator
1 Plate Assembly 1 Ground Cord
1 Battery
Features
The 19430 has four push buttons. The buttons operate by pressing once lightly. There is no need to press the button repeatedly. Its effective life may be decreased by repeated misuse.
NOTE: For reliable measurement, the eldmeter must be grounded properly using the ground lead provided. The plastic case of the 19430 Electrostatic Fieldmeter is made of conductive resin. The grounding terminal provides the reference potential for the electrical circuit. This terminal must be grounded properly for accurate measurement.
Operation For Static Charge Measurement
1. Turn the ring range lights and sensor toward a surface to be evaluated.
2. Press the POWER button. A single beep for one second indicates proper operation. The battery capacity indicator and bar graph will be displayed.
3. The meter reading should be zero (0). Use the ZERO button if it is not.
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
TB-3048 March 2011 Page 1 of 5
© 2010 DESCO INDUSTRIES INC.
Employee Owned
4. Hold the eldmeter 1” away from the charged surface. Two beams of light will form concentric circles when the meter is the correct distance, 1” away, from the surface. The measurement accuracy is dependent on a stable ground reference and the 1 inch measuring distance. It is also dependent on the “aspect ratio”, relating the size of the object to be measured to the measurement distance. This aspect ratio should be at least 3 for best accuracy, i.e. the object should be at least a 3 inch square when measuring at a 1 inch distance. Accurate measurements may be made at other measurement distances by scaling the meter range and observing the proper aspect ratio. For example, at a measurement distance of 3 inches, multiply the meter reading by 3 to give a range of 0 to 60 kilovolts. For accuracy, the object being measured at this distance should be at least a 9 inch square.
Operation For Ion Balance (Offset Voltage) Measurement
NOTE: The Digital Static Locator Ion Balance
measurement feature should not be used to measure the ion balance of static bars. It may be used to mea­sure other ionizers for ion balance (offset voltage).
1. Depress the plate assembly release button on either side of the eldmeter to remove the charge plate assembly.
2. Insert the plate assembly into the other end of the eldmeter (the end with the Desco logo) until it
clicks into place.
NOTE: Ion Balance plate assembly must not move during measurement.
5. A bar graph and a digital display on the 19430 displays the voltage of the charge. Press HOLD if you want to hold the display of the measured charge.
6. To turn off meter: Press POWER if you have not used the HOLD mode.
If you have pressed HOLD and
wish to retain the
meter reading for the next use, press the POWER button to turn off the 19430.
If you have pressed HOLD and
do not wish to retain the meter
reading for next use, press
Figure 3.
HOLD then POWER.
If the 19430 is left on for ve minutes, the unit will beep
ve times then turn itself off automatically. To disable this feature, simply press and hold the red power button for at least three seconds when turning the 19430 on. Three beeps and the A-OFF display indicates that the feature has been disabled.
3. The conductive case must be grounded using the supplied ground cord. It is imperative for accurate ion balance and to drain any charge that has developed prior to measurement.
4. To turn on the ion balance measuring circuit, press the blue IB button.
5. Turn sensor toward an uncharged area.
6. The meter reading should be zero (0). Press the Zero button if it is not.
7. Hold the eldmeter centered in front of the ionizer at any convenient, repeatable distance, such as 1 foot.
8. A bar graph and a digital display on the 19430 displays the ion balance information. Press HOLD if you want to hold the display of the measured
charge.
9. To turn off the meter:
Press POWER if you have not used the HOLD
mode.
Results
A display of [+] or [-] next to the digital numeric value represents the polarity of the static charge being measured. In addition, the color of the bars in the bar
graph indicates the polarity of the static charge. Red
and blue bars indicate positive and negative charges respectively.
During a static charge measurement, if the digital display starts ashing along with a continuous beep, stop taking the measurement. The static charge is greater than the meter can measure. Attempts to measure too high a charge may damage the sensor.
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
TB-3048 Page 2 of 5
If you have pressed HOLD and wish to retain the meter reading for the next use, press the POWER button to turn off the 19430.
If you have pressed HOLD and do not wish to retain the meter reading for next use, press HOLD
then POWER.
10. Depress the charge plate assembly release buttons
on either side of the eldmeter to remove the charge plate assembly. Store the plate on the opposite end, pressing it until it clicks into place.
© 2010 DESCO INDUSTRIES INC.
Employee Owned