Desco 19219 Installation Manual

TECHNICAL BULLETIN TB-3049
Ground Hub Continuous Monitor Operation, Installation and Maintenance
The Ground Hub Continuous Monitor is available in the following models:
Item Voltage Plug
19219 120VAC N. America
19224 220VAC Asia / UK
Packaging
1 Ground Hub Monitor 6 10-32 x 1/4" Screws 6 #10 Split Washers 1 Certificate of Calibration
Figure 1. Desco 19219 Ground Hub Continuous Monitor
Features and Components
Ground Hub
Made in America
Neutral
Load / Hot
Ground
Figure 2. Desco 19224 Ground Hub Continuous Monitor
Description
When the Ground Hub Continuous Monitor is plugged into an AC outlet, the green LED illuminates when both the outlet’s wiring is correct and the path to equipment ground via the equipment grounding conductor is intact. The monitor provides 12 verified ground points when in PASS condition. The red LED illuminates when either the outlet’s wiring is incorrect or the path to equipment ground is defective.
Use the Ground Hub Continuous Monitor to fulfill the ANSI/ ESD S6.1 section 6.3.1 requirement. “The hot, neutral and equipment grounding conductor shall be verified to be in the proper wiring orientation in accordance with the National Electric Code (ANSIINFPA-70).”
Outlet Status
FRONT VIEW BACK VIEW
Figure 3. Desco 19219 Ground Hub Continuous Monitor features and components
Earth
Load / Hot
Neutral
Outlet Status
Earthing Point
FRONT VIEW BACK VIEW
Figure 4. Desco 19224 Ground Hub Continuous Monitor features and components
TB-3049 Page 1 of 3
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Website: Desco.com
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
February 2011
© 2011 DESCO INDUSTRIES, INC.
Employee Owned
Operation
Plug the Ground Hub Monitor into an outlet to see if the green PASS LED illuminates. This signifies that the outlet is correctly wired and the path to equipment ground is intact. The screw that holds the outlet cover on may be used to secure the Ground Hub Monitor so that it cannot be unplugged. Use the included screws and split washers to connect ring terminals (not included) to the Ground Hub or the banana jacks (not included) to connect banana plugs (see Figure 5).
(Wires not included)
Figure 5. Installing ring terminals and banana plugs to the Ground Hub
Calibration
The Ground Hub Monitor is calibrated to NIST traceable standards. Calibration may be performed to insure that the tester is operating within limits. We recommend annual calibration of our monitors.
TESTING THE GROUND HUB MONITOR (19219 ONLY)
NOTE: The following procedure should only be done by someone familiar with voltage hazards. This procedure does not apply to item 19224 because Asia and the United Kingdom do not reference GROUND to NEUTRAL as the United States does.
Isolate the ground plug from the Ground Hub Monitor by inserting the monitor into a 3 to 2 plug adapter.
A. PASS OUTLET - Connect a 5 kilohm resistor @ 5% tolerance across the Neutral Plug and Ground Plug. The PASS LED should remain illuminated until the resistor is removed. This test confirms the PASS point for the neutral-to-ground resistance.
B. FAIL OUTLET - Connect a 12.5 kilohm resistor @ 5% tolerance across the Neutral Plug and Ground Plug. The FAIL LED should remain illuminated until the resistor is removed. This test confirms the FAIL point for the neutral-to-ground resistance.
Figure 6. Ground Hub Continuous Monitor in-use photo
Specifications
Input Voltage:
120VAC ±15% (Item 19219) 220VAC ±15% (Item 19224)
Current Drain: < 10mA
Equipment Ground Conductor Resistance Limits:
10 kilohms (nominal) (19219 only)
Dimensions:
2.0" x 3.0" x 1.3" (5.1cm x 7.6cm x 3.2cm) (Item 19219)
2.0" x 3.0" x 2.0" (5.1cm x 7.6cm x 5.1cm) (Item 19224)
Operating Temperature:
0° - 40°C
TB-3049 Page 2 of 3
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Website: Desco.com
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
© 2011 DESCO INDUSTRIES, INC.
Employee Owned