Denar Mark II User Manual

Page 1
The Dénar
®
Mark II System
TECHNIQUE MANUAL
Page 2
Page 3
The Dénar
®
Mark II System
TECHNIQUE MANUAL
Whip Mix Corporation - West
1730 East Prospect Rd., Suite 101
Fort Collins, CO 80525
Toll-Free: 1-800-201-7286
Fax: 1-970-472-1793
www.whipmix.com
Page 4
ACKNOWLEDGEMENTS
The Mark II System was developed to fill a need that existed primarily in dental schools, but also among practicing dentists and laboratory technicians. The Schools of Dentistry expressed a desire for an effective teaching system that was competi­tively priced. Practicing dentists and technicians expressed a need for a simple Arcon articulator that was anatomically accurate which could be used for simple restorative procedures and to mount diagnostic casts when illustrating occluso-condylar rela­tions to patients for case presentations. There was also a need for an instrument sys­tem offering an intermediary step to the incorporation of the principles of occlusion as well as one offering upward potential to more advanced systems.
We set these needs as our objectives and proceeded to accomplish the task with the help of many professionals who provided us with both direction and assistance for which we are most grateful.
Doctors Rex Ingraham, Patrick M. Walker, Donald C. Curntte, Albert Solnit, Howard M. Landesman, Glen D. Richardson, all at the University of Southern California, gave us extremely valuable inputs with respect to the needs of the undergraduate stu­dents. We are indebted to them particularly for their constructive criticism, even though painful at the time.
A special word of appreciation is expressed to Doctors Sumiya Hobo and Frank V. Celenza for their contributions in the early design phases of the instrumentation system.
In planning the preparation of the Mark II Technique Manual, it was our intent that it encompass more than just mechanical instruction in the use of the Denar
®
Mark II System. We wanted to offer more by also providing related instruction in the theory of occlusion as it directly pertains to the use of the instrument. Special credit must go to Dr. Niles F. Guichet for his contributions and the time he spent working with us, particularly in view of the demands of his teaching schedule.
We wish to acknowledge the direction and wisdom that we received from Doctors L. D. Pankey, Loren Miller, Henry Tanner, James Zuccarella, Mel Steinberg, and Mr. Jack Snyder, of the Pankey Institute, with respect to how the system can be used by prac­titioners wishing to render quality dentistry through the incorporation of the principles of occlusion. A great deal of encouragement in this area was also received from Dr. Peter E. Dawson to whom we are equally grateful.
To insure the System’s compliance with the purest theories of gnathology we are most indebted to Dr. Peter K. Thomas. He made what seems like “impossible demands” to arrive at perfection. Fortunately, his toughness was matched with a great deal of patience.
Through the development phases many different opinions were expressed, but there was at all times one common goal: to provide dentistry with a quality occlusal instru­mentation system. We believe we accomplished this goal. This feat took the efforts, contributions, dedication and assistance of many more people not mentioned, and to all of them as well: we are very grateful.
Page 5
CONTENTS
I. Introduction
Who Should Use The Mark II System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Why Should The Mark II System Be Used . . . . . . . . . . . . . . . . . . . . . . . . . 6
II. The Denar®Mark II Articulator (Semi-adjustable)
Articulator Manipulations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Articulator Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Hand Grasps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
III. Relating Condylar Movements to Occlusal Anatomy . . . . . . . . . 12-16
IV. The Immediate and Progressive Side Shift
Adjustments (Bennett Shift) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
V. The Denar®Mark II Facebow/Earbow (Model D31AB)
Locating Three Reference Points on Patient’s Face . . . . . . . . . . . . . . . . . 19
Making the Facebow/Earbow Registration . . . . . . . . . . . . . . . . . . . . . . . . 21
Transferring the Facebow/Earbow to the Articulator. . . . . . . . . . . . . . . . . 24
VI. Mounting the Casts in the Articulator
The Maxillary Cast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
The Mandibular Cast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
VII. Setting the Articulator to Checkbite Records
Simulating the Orbiting Condylar Paths . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Simulating the Protrusive Condylar Paths. . . . . . . . . . . . . . . . . . . . . . . . . 31
VIII. Incisal Table Adjustments
Custom Incisal Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Adjustable Incisal Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
IX. Treatment Procedures
Fixed and Removable Partial Denture Restorations . . . . . . . . . . . . . . . . . 35
Complete Dentures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Page 6
APPENDICES
A. Checkbite Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
B. Denar
®
System Protocol for Dentist-Laboratory Relations . . . . . . . 42
C. Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
D. Occlusal Plane Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
E. Selecting Instruments for Occlusal Treatment. . . . . . . . . . . . . . . . . 55
F. Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
G. Care and Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
H. Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
IMPORTANT
Your Denar®Mark II Articulator and Facebow/Earbow are precision instruments man­ufactured to precise tolerances and designed to give you years of troublefree service. Like all precision instruments they must be handled carefully to avoid damage. A thorough study of the information contained in this instruction manual will insure you of the benefits which these instruments offer.
DO NOT ADJUST the factory adjustment screws illustrated in figure 2A. They are for factory use only. Adjustment of these screws in the field can alter the instrument’s operation and may require factory repair.
Also the microadjustable adjustment lockscrews illustrated in figure 2B are used to calibrate the centric relation position of your Mark II articulator to tolerances of plus or minus one thousandths of an inch from the factory reference position. Adjustment of these screws can alter the precise factory alignment of your articulator. Do not modify the setting of these adjustment screws until after you have read this instruc­tion manual and thoroughly understand the function of these adjustment screws as explained in Appendix D. These microadjustment screws should only be used in con­junction with a Denar
®
Field Inspection Gage to calibrate your articulator.
fig. 2A
fig. 2B
Factory Adjustments
Microadjustment Screws
Page 7
5
I. INTRODUCTION
WHO SHOULD USE THE DÉNAR®MARK II SYSTEM
DOCTORS WHO WANT:
• To mount casts quickly and easily on a semi-adjustable instrument that mechanically and accurately reproduces mandibular movements.
• To produce restorations by means of checkbite records and/or the func- tionally generated path (FGP) techniques.
• A useful tool for diagnosis and for the fabrication of restorations when not utilizing a pantograph and fully adjustable articulator.
• A semi-adjustable articulator with the immediate side shift adjustment capability.
• Casts mounted on a very rigid articulator in the position of maximum inter­cuspation.
• To incorporate the benefits of the Denar
®
Two Instrument System.
TECHNICIANS WHO WANT:
• A practical, rigid and easy to use articulator.
• To efficiently produce restorations that require fewer remakes than restora­tions constructed on articulators of lesser adjustment capability.
• To receive mounted casts.
• To service dentists using the Denar
®
Two-Instrument System.
EDUCATORS WHO WANT:
• An economically priced articulator and facebow for student issue without violating sound anatomical principles.
• Occlusal instruments that fulfill the needs of all restorative departments.
• To avoid the need for the student having a separate articlulator for each restoration under construction.
STUDENTS WHO WANT:
• To study Occlusion and the movement of the temporomandibular joint. A means of progressing to a fully adjustable articulator and pantograph.
Page 8
6
WHY SHOULD THE MARK II SYSTEM BE USED
Simple and Practical to Use
The Denar®Mark II Articulator and Facebow System enables the user to quickly and easily mount casts of a patient’s teeth on an instrument that is both an equivalent of their natural rela­tionship and which also can be mechan­ically programmed to simulate the mandibular movements of the patient. To accomplish this mechanical equiva­lence, the Mark II articulator has adjust­ment capability to duplicate the more clinically significant movements of the mandible. Those condylar paths of movement of lesser clinical significance have not been ignored, but instead, are constructed to average anatomic dimensions.
Accurate Diagnostic and
Treatment System
The Mark II System is a particularly use­ful tool for the diagnosis and for the fab­rication of restorations when not utilizing a pantograph and fully adjustable artic­ulator. The simplicity and accuracy with which the system may be used enables the user to produce precision occlusal restorations that require significantly less in the mouth modifications than restorations constructed on articulators of lesser adjustment capability.
Excellent Learning Tool
The Mark II System is built around sound principles of human anatomy and is consequently ideal for study of TMJ characteristics and theories of occlu­sion. Understanding this system facili­tates progress to a fully adjustable artic­ulator as the movements and adjust­ments are the same.
Constructed with Clinically
Needed Features
The Mark II System is easy to use. The articulator is rigid with a very positive centric lock. It is of the Arcon construc­tion with the back designed for maxi­mum lingual visibility to the casts. The upper and lower bows come apart and lock together in the open and closed positions (no rubber bands needed). The articulator can be placed level in the upside down position for mounting the casts without the need of a plastering stand. The facebow sidearms are inde­pendently adjustable and can be locat­ed to either the hinge axis or the exter­nal auditory meatus (opening) of the ears.
Economical
The Mark II Articulator which is compet­itively priced is also two instruments in one. Not only is it a semiadjustable articulator with the features described above, but also because of the microm­eter adjustments in the condylar areas, the condyles can be adjusted three dimensionally with the Denar
®
Field Inspection Gage to tolerances of plus or minus .001 inches (.025 mm) which allows transfer of the mounted casts to other Denar
®
articulators. This feature reduces the need for a separate articu­lator for every restoration under con­struction.
Page 9
7
In order to be proficient in the use of the articulator to diagnose condylar paths of movement and to fabricate occlusal restorations, the operator must have knowledge of:
• The articulator condylar controls and how to adjust them.
• The proper hand grasps for manip­ulating the articulator through its excursive movements.
In this section of this manual we will first discuss the simple procedure of how to open and close the articulator properly and how to operate the centric latch. Secondly we will discuss the location of the articulator condylar controls and how to adjust them. Lastly we will cover the proper hand grasps for manipulating the articulator through its excursive movements. How to adjust the articula­tor to checkbite records is discussed in another section of this manual.
ARTICULATOR MANIPULATIONS
To assemble the articulator hold the upper bow approximately parallel to the lower bow and simply place it on the lower bow.
Centric Latch Operation
When the articulator is closed it can be locked in the centric position by pushing the centric latch to the down position. The centric latch is opened by placing the index finger on the trigger located underneath the center of the crossbar of the lower bow and by placing the thumb on top of the upper bow (fig. 3). Next apply pressure with the index finger; it must be released by the trigger. When the centric latch is open the upper bow can be removed from the lower bow by lifting it vertically while maintaining the upper bow parallel to the lower bow.
To Open the Articulator
Apply downward pressure to the top of the upper bow over the centric latch and fossa assemblies with the heel of the palm of the left hand and simply open the articulator. The opening movement of the articulator automatically engages the centric latch and locks it (fig. 4).
II. THE DÉNAR®MARK II ARTICULATOR
fig. 3
fig. 4
Page 10
8
To Lock in the Open Position
When the articulator is opened, the upper bow can be moved downward toward the lower bow so that the condyles move forward in their fossae to engage the lock open position (fig. 5).
To Close the Articulator
To close the articulator move the upper bow up and forward to disengage the lock-open position and close the articu­lator (fig. 6).
ARTICULATOR ADJUSTMENTS
The articulator is a mechanical equiva­lent of the lower half of the head
__
a mechanical jaw so to speak. In order to discuss the adjustments of the articula­tor or specifically the fossa controls, it would be helpful to discuss the condylar paths of movement of the human mandible. In a lateral mandibular move­ment the condyle on the side toward which the mandible moves is termed the rotating condyle (fig. 7). The condyle on the side opposite the side towards which the mandible moves is termed the orbiting condyle.
In a lateral mandibular movement the orbiting condyle moves inward, down­ward and forward and orbits about the
r
otating condyle which is simultaneous­ly rotating and moving outward during the mandibular side shift.
orbiting path - the path of move­ment of the orbiting condyle. rotating path - the path of move­ment of the rotating condyle. protrusive path - the path of movement of the condyle in a straight protrusive movement.
fig. 5
fig. 6
fig. 7
Page 11
9
Protrusive Adjustments
The inclination of the protrusive condy­lar path can be adjusted by loosening the protrusive adjustment lockscrew. The protrusive condylar path inclination scale is below the protrusive adjustment lockscrew and is calibrated in incre­ments of 5 degrees (fig. 8).
Immediate Side Shift Adjustment
The medial fossa wall can be displaced straight medially by means of the imme­diate side shift adjustment. The scale for the immediate side shift adjustment is lateral to the adjustment lock screw on top of the fossa (fig. 9). The scale is a Vernier scale calibrated in .2 millimeter increments. The scale reads medialward from its lateral extremity.The index in fig. 10 indicates an immediate side shift set­ting which is more than 0 but less than 1 full millimeter. By reading the Vernier scale on the lower portion of the scale medialward from the index it can be determined that the immediate side shift is .6 of a millimeter since it is the third graduation that lines up with a millimeter graduation on the upper portion of the scale.
Progressive Side Shift Adjustment
The angle of inclination of the medial fossa wall to the sagittal plane be adjusted by loosening the progressive side shift adjustment lockscrew and moving the insert from 5 to 15 degrees. The scale for the progressive side shift adjustment is anterior to the adjustment lockscrew and is calibrated in 5 degree increments. (fig. 11).
fig. 8
fig. 9
fig. 10
Scale
Lockscrew
Page 12
10
Rear Wall Inclination
The posterior fossa wall of the Mark II Articulator is nonadjustable but is con­structed to average anatomic dimen­sions. It is inclined posteriorly 25 degrees to allow for a backward move­ment of the rotating condyle as it moves outward (fig. 12).
HAND GRASPS
To use the articulator properly the oper­ator must master the proper hand grasps. A right handed operator curls the fingers of the left hand under the mandibular crossbar and places his left thumb on top of the upper bow (fig. 13). To effect both left and right lateral excur­sive movements the left thumb guides the back of the upper bow while the forefinger and thumb of the right hand holds the incisal pin moving it in the desired direction. These hand grasps can be best described as the underhand push__pull grasps and the underhand protrusive grasps. In order to execute lateral movements the centric latch must be open.
Underhand Push Grasps
To effect a right lateral mandibular movement be sure the latch is open and move the upper bow to the left. (The left side of the articulator is the left side of the instrument as it is viewed from the rear.) Pressure is applied with the left thumb to insure the left orbiting condyle maintains contact with its superior and medial fossa walls and the right rotating condyle maintains contact with its supe­rior and rear fossa walls (fig. 13).
fig. 11
fig. 12
fig. 13
Rear
Wall
Page 13
11
Underhand Pull Grasps
To effect a left lateral mandibular move­ment the upper bow is moved to the right and pressure is applied with the left thumb to insure that the right orbit­ing condyle maintains contact with its superior and medial fossa walls and the left rotating condyle maintains contact with its superior and rear fossa walls (fig. 14).
Underhand Protrusive Grasps
(Protrusive Push Grasp)
To effect a straight protrusive movement the upper bow is moved straight poste­riorly with the right hand and the left thumb is used to apply downward pres­sure on the back of the upper bow so that the condyles maintain contact with their superior fossa walls.
The overhand grasps as contrasted from the underhand grasps are also useful in manipulating the articulator and are required to efficiently set a fully adjustable articulator to a pantographic record. Figure 15 illustrates the
Overhand Push Grasp; fig. 16 the Overhand Pull Grasp, and fig. 17 the Overhand Protrusive Grasp. When
employing the overhand grasps to manipulate the articulator be sure to apply pressure to the back of the articu­lator to insure that the condyles main­tain contact with their respective fossa bearing surfaces.
fig. 14
fig. 15
fig. 16
fig. 17
Page 14
12
The Dénar®Mark II Articulator is of the Arcon construction; i.e., the condyles are attached to mandibular bow and the fossa assemblies are fixed to the maxil­lary bow. This construction which is a facsimile of the anatomical structures, enables the articulator to more accu­rately simulate condylar paths of move­ment. In addition this construction makes it easy to understand the relation of condylar paths of movement to occlusal anatomy.
An understanding of the relationships which exist between condylar paths of movement and occlusal anatomy is an invaluable aid in the use of an articulator for diagnosis and treatment. The follow­ing exercises which utilize the articulator as a teaching method are helpful to enable you to quickly understand these relationships.
Set the left immediate side shift adjust­ment to 1 millimeter and left progressive side shift adjustment to 15 degrees. By observing the articulator movements from the back of the articulator it is easy to understand why the immediate and progressive side shifts are so named (fig. 18). Hold the articulator in centric relation. Since the left medial fossa wall is set to permit a one millimeter immedi­ate side shift, centric relation is achieved when the right condyle touches its medial fossa wall.
Move the articulator in a right lateral mandibular movement until the left orbiting condyle contacts its medial fossa wall and note that the rotating condyle and mandible move immediate- ly to the right. As you continue the right lateral mandibular movement the orbit­ing condyle move downward, forward and inward. Note that during this move­ment of the orbiting condyle the rotating
condyle and mandible move progres­sively more to the right as the orbiting condyle advances. Repeat this articula­tor movement and note that the rotating condyle moves immediately to the right and then progressively more to the right as the orbiting condyle advances.
mandibular side shift (Bennett Shift): the bodily side shift of the mandible which occurs during a lateral jaw movement. immediate side shift: a mandibular side shift in which the orbiting condyle moves essentially straight medially as it leaves centric relation. progressive side shift: a mandibular side shift which occurs at a rate or amount which is directly proportional to the forward movement of the orbit­ing condyle.
By observing a right lateral mandibular movement from the front of the articula­tor you can see that the path of move­ment of the orbiting condyle (orbiting path) as it moves inward, downward and forward is guided by the superior, rear and medial fossa walls (fig 19). This condylar path of movement is associat­ed with and has its principal effect on the balancing inclines of cusps on the orbiting side (fig. 20B).
III. RELATING CONDYLAR MOVEMENTS
TO OCCLUSAL ANATOMY
fig. 18
Page 15
13
o
f the marginal ridge, fossa, or central
groove of the tooth (fig.21).
An increase of the progressive side shift movement of the articulator has an effect of flattening the balancing inclines of cusps on the orbiting side mediolaterally (fig. 22).
The closer a cuspal incline is to a condylar path of movement the greater is the influ­ence that condylar control has on occlusal anatomy. Consequently due to the fact that the orbiting condyle is moving down­ward so rapidly as it moves forward, we observe that as we move more distally in the dental arches the lingual cusps of maxillary molars project increasingly downward and the buccal cusps of mandibular molars project increasingly upward to harmonize the occlusion to condylar paths of movements (fig 23).
Again by observing a right lateral condylar movement from the front of the articulator you can see that the path of movement of the rotating condyle (rotating path) as it moves outward is guided by the rear and top fossa walls (fig.24). This path of move­ment is most closely associated with and has its principle effect on the working inclines of cusps on the working side (fig. 20A).
The rotating condylar path may be inclined upward or downward as the rotating condyle moves outward. This upward and downward inclination of the rotating condylar path in the coronal plane has its principle influence on the height of the working inclines of posterior cusps on the rotating side (fig. 25). If the rotating condy­lar path is inclined upward the cusps must be flatter (fig. 25A). If the rotating condylar path is inclined downward the cusps may be steeper (fig. 25C). The Mark II Articu­lator cannot be adjusted to upward or downward movements of the rotating condyle.
fig. 19
fig. 20
fig. 21 fig.22
Three articulator adjustments establish the character of the orbiting path on the artic­ulator: the immediate side shift adjustment, the progressive side shift adjustment and the protrusive inclination of the superior fossa wall.
An increase of the immediate side shift movement of the articulator has an effect of increasing the bucco-lingual dimension
Page 16
14
The rotating condylar path may be inclined forward or backward as the rotating condyle moves outward. This forward and backward inclination of the rotating condylar path in the horizontal plane has its principle effect on the intermeshing of the working inclines of cusps on the work­ing side (ridge and groove direction).
The Dénar
®
Mark II Semi-adjustable Articulator has the rotating condylar path reset to the average anatomic inclination (out and backward 25 degrees).
Figure 26A illustrates a frontal view of molar tooth relations in a right lateral mandibular movement. Although the rotat­ing condyle moves straight outward the functioning tooth inclines on the rotating side have a slight downward inclination due to the fact that the path of movement of the orbiting condyle is inclined remark­able downward.
fig. 24
fig. 25
fig. 23
Page 17
15
The closer the functioning tooth incline is to the condylar path of movement the more the tooth incline simulates that condylar path of movement.The interrelat­ing tooth inclines on the orbiting side in figure 26A have steep inclines to comple­ment the path of movement of the orbiting condyle.
Figure 26B illustrates a left lateral move­ment. Due to the fact that the left rotating condyle is moving straight outward the left maxillary buccal cusps must be kept short to allow the left mandibular buccal cusps to escape. It is this influence of the rotat­ing and orbiting condylar paths on occlusal anatomy that establishes the
Curve of Wilson. The more posteriorly we progress in the dental arches the mandibular teeth take on a greater lingual inclination and the maxillary teeth take on a greater buccal inclination to harmonize occlusal anatomy to condylar paths of movement (fig. 27). The condyle tracks a path in its fossa just as a buccal cusp of a lower molar tracks a path in its fossa on the occlusal surface of an upper molar. For all practical purposes in the use of articu­lators to establish dental articulation, the temporo-mandibular joint can just be thought of as another tooth, the fourth molar__another anatomic control of jaw movement (figs. 26C and 27).
fig. 26 fig. 27
Page 18
16
To facilitate a clear understanding of the relation of the orbiting condylar path to occlusal anatomy study fig. 28. Illustrated are the cuspal inclines of the left bicuspids and molars which are associated with the orbiting condylar path. It is the distal aspects of the max­illary lingual cusps’ buccal inclines (shaded) which interrelate with the mandibular buccal cusps’ lingual inclines, mesial aspects. In your mind’s eye it is helpful to dissect out these cus­pal inclines (fig. 28A) and visualize what influence a change in the character of the orbiting path on the articulator would have on these aspects of the
cusps. Three articulator controls estab­lish the character of the orbiting path on the articulator
__
the immediate side shift adjustment and the inclination of the medial and superior fossa walls. In­creasing the immediate side shift adjust­ment on the articulator increases the clearance between these cuspal inclines (fig. 28B). Increasing the progressive side shift movement of the articulator (increasing the inclination of the medial fossa wall) flattens the cuspal inclines mediolaterally (fig. 28C). A decrease of the inclination of the superior fossa wall flattens the cuspal inclines anterio-pos­teriorly (fig. 28D).
fig. 28
Page 19
17
It should be noted that unlike most semi-adjustable articulators the Denar
®
Mark II Semi-adjustable Articulator has the capability of more accurately simu­lating the mandibular side shift (Bennett Shift) by more accurately simulating the component condylar movements: the immediate side shift and the progressive shift. The immediate side shift is expressed in units of tenths of a millime­ter (fig. 29A). The progressive side shift is expressed in degrees (fig. 29B).
The immediate side shift of the mandible has primary influence on the width of the central groove of posterior teeth.The progressive side shift has its principal influence on the balancing inclines of posterior cusps on the orbiting side and on the direction of the ridges and grooves of posterior teeth, primarily on the orbiting side.
Figure 30 illustrates the protrusive, orbit­ing and rotating path records of the right and left temporomandibular joints of 50 patients (100 TMJ records). 1 The X and Y axes are calibrated in increments of 1 millimeter. You will note that the orbiting path is divided essentially into two com­ponents: immediate side shift and pro­gressive side shift. Furthermore with few exceptions once the immediate side shift has occurred the progressive side shift records are approximately parallel to each other and are inclined approxi­mately five to seven degrees to the sagittal plane. The biggest variable is the immediate side shift component of the orbiting path.
Points A, B and C on one orbiting path represent three different condylar posi­tions at which lateral checkbite position­al records may be taken on one patient. It should be noted that if an articulator possessing a progressive side shift and not an immediate side shift adjustment were set to each of the three condylar positions A, B and C as shown in fig. 30. It would produce three different progres­sive side shift inclinations correspon­ding to the three dotted lines in figure 30
__
all of which inclinations would be wrong. On the other hand, if an articula­tor possessing a progressive as well as an immediate side shift adjustment
IV. THE IMMEDIATE AND PROGRESSIVE SIDE
SHIFT ADJUSTMENT (Bennett Shift)
fig. 29
1. Lundeen, Harry C. and Wirth, Carl G.: Condylar Movement Patterns Engraved in Plastic Blocks, J. Prothet. Dent. December 1973.
Pages 870-875.
Page 20
18
(Dénar®Mark II) were adjusted so that the progressive side shift was pre-set to the average anatomic dimension of six degrees, one immediate side shift adjustment setting would intersect with all three condylar position checkbite records (A, B, and C) which remarkably
reduce the amount of irritation that oth­erwise might be introduced in the occlu­sion. Therefore when adjusting the Mark
II Articulator to lateral checkbite records, always set the progressive side shift adjustment to the 6˚ average anatomic dimension for this diagnostic procedure.
fig. 30
Page 21
19
The Dénar®Mark II Facebow/Earbow is used to register the correct position for the patient’s maxillary cast to be mount­ed in the articulator. In other words, the facebow/earbow records the relation of the patient’s maxillary dental structures to the horizontal reference plane and transfers this relationship to the articula­tor.
The use of the Denar
®
Facebow/ Earbow
involves three overall procedures:
A. Locating three reference points
on the patient’s face
B. Assembling the face-
bow/earbow on the patient
C. Transferring the face-
bow/earbow to the articulator
The detailed steps of each of these three procedures is as follows.
LOCATING THREE REFERENCE POINTS ON THE PATIENT’S FACE
The components of the facebow kit needed are: the reference plane locator and reference plane marker. These two items are used to locate three anatomi­cal reference points on the patient’s face. Of these three points, two are pos­terior and one is anterior.
There are two means of locating the posterior points. The first is by precise location of the terminal hinge axis with a
hinge axis locator.
1
The second means is by locating the points by average anatomical measurement, which is sim­pler and faster, and is the procedure described in this section of this manual.
V. THE DÉNAR®MARK II
FACEBOW/EARBOW (Model D31AB)
1. Other reference material must be consulted for detailed instructions on how to precisely locate the terminal hinge axis position (i.e., the “Denar
®
Procedures Manual__Procedures for Occlusal
Treatment” or the “Denar
®
Office Tutor” ).
´
´
fig. 31
fig. 32
´
REFERENCE ROD INDEX
MILLIMETER SCALE
43mm
NOTCH
POSTERIOR REFERENCE POINT HOLE
Page 22
20
Average measurement may be used to locate the posterior reference points whenever you do not vary the vertical dimension of the casts on the articula­tor, or, in other words, when the mandibular cast is to be transferred to the articulator by means of an interoc­clusal record taken at the correct verti­cal dimension and the vertical dimen­sion is not going to be changed on the articulator.
Place the “reference plane locator” along the right side of your patient’s face. It should extend from the middle of the upper border of the external audito­ry meatus to the “outer canthus” of the eye. In other words, the reference plane locator should extend from the middle of the upper border of the ear-hole to the outer corner of the eye (fig. 33).
There is a small hole in the upper poste­rior area of the locator. Once the locator is in position on the patient’s face, use your felt-tipped pen to gently mark through the hole onto the face (fig. 34).
Make the mark on both sides of the patient’s face.
The position of the “anterior reference point” is measured up 43 millimeters
from the “incisal edges” of the central or lateral incisor, toward the inner corner of the eye. The notched out area of the “reference plane locator” is used to make this measurement. The notch is 43 millimeters in length.
Simply rest the lower edge of the notch on the incisal edge of the right central or lateral incisor. On an edentulous patient measure up from the low lip line. The “low lip line” is the lower border of the upper lip when it is in repose. In either case, mark the anterior reference point below the inner canthus of the right eye where the top point of the locator touch­es the patient’s face (fig. 35).
fig. 33 fig. 34
fig. 35
Page 23
21
M
easure the distance between the ante­rior reference point and the inner can­thus of the eye (fig. 36). Record this measurement in the patient’s file for future reference. In this way, if the ante­rior teeth are removed or modified the same anterior reference point can be located by measuring downward from the fixed immovable inner canthus of the eye.
The final step is to mark the “horizontal reference plane” on the right side of the patient’s face. Just line the ruler up between the anterior and posterior ref­erence points. Hold the ruler so that it is just out of contact with the patient's skin, so that it will not displace the skin, and then draw a short line on the side of the face. This line represents the “hori­zontal reference plane” (fig. 37).
Y
ou will therefore notice that the hori­zontal reference plane is identified on the face of the patient by two posterior reference points in the area of the termi­nal hinge axis and one anterior reference point located 43 millimeters above the incisal edges of the maxillary anterior teeth or low lip line of the patient.
MAKING THE FACEBOW/ EARBOW REGISTRATION (Assembling the Facebow/Earbow on the patient)
The components of the kit needed are: the bitefork, anterior crossbar, reference rod, reference rod clamp, and the right and left facebow side arms with nylon earplugs at the ends of the posterior ref­erence slides (fig. 31).
Attach the bite fork to the crossbar so the reference rod clamp is to the patient’s right, and the u-shaped part in the bite fork is above the crossbar (fig.
38). Then load the upper surface of the bite fork with two thicknesses of base­plate wax (fig. 39). Soften the wax to a dead soft consistency in warm water or an open flame, and then put the loaded bite fork in the patient’s mouth to get a light indexing impression of the maxil­lary teeth. When the bite fork is first placed in the mouth, be certain to line up the crossbar so that it is parallel to the coronal and horizontal planes of the patient. Also be sure to be careful not to depress or displace any mobile teeth
__
all you really need is a slight impression of the tips of the cusps (fig. 40).
Remove the bite fork from the patient’s mouth, and place the maxillary cast, if available, in the bite fork to confirm accurate seating. If the maxillary cast seats accurately in the bite fork, you can now begin assembly of the facebow record.
Put the bite fork assembly back in the patient’s mouth, indexing it to the maxil­lary teeth. Have the patient hold the bite
fig. 36
fig. 37
Page 24
22
f
ork in place (fig. 41). This can be done most conveniently by having the patient bite on the index and middle fingers of the left hand. Alternatively position cot­ton rolls on the occlusal surface of the lower posterior teeth and instruct the
p
atient to maintain the bite fork in place
with gentle biting pressure.
Adjust the reference rod clamp so it is parallel to the reference plane marked on the patient’s face (fig. 42).
fig. 38 fig. 40
fig. 39 fig. 41
Page 25
23
At this point you will be ready to attach the facebow side arms. Note that they are marked right and left and refer to the patient’s right and left. Make sure that the scales on the posterior reference slides are adjusted to their zero posi­tions.
A. Facebow Application
Remove the nylon earpieces on both the posterior reference slides and begin the attaching of the side arms (fig. 43). First you will need to locate the right side arm on the crossbar so that the lockscrew on the crossbar clamp faces upward and the posteri­or reference pin at the end of the posterior reference slide lightly touches the posterior reference point. Secure the side arm clamp to the anterior crossbar. Then attach the left side arm similarly.
B. Earbow Application
Make sure the nylon earpieces are on both posterior reference slides. Position the right arm on the cross­bar so that the lockscrew on the crossbar clamp faces upward and the nylon earpiece fits snugly in the external auditory meatus. Secure the side arm clamp tightly and attach the left side arm similarly.
At this point the facebow/earbow records the relationship of the maxillary
dental structure to the posterior refer­ence points. The only thing remaining to be done is to relate the maxillary dental structures to the anterior reference point.
Insert the reference rod into its clamp bringing it up from underneath the clamp, with the step in the rod facing toward the patient’s right (fig. 44). Hold the reference plane locator between the thumb and index finger of your left hand so that you can read the instructions on the back. The semilunar notch on the locator’s inferior surface should be placed over the bridge of the nose. Note the small hole in the center of the loca­tor. Turn the locator down flat so that the instruction side faces downward and index the hole on the locator over the small dowel-like projection on the upper extremity of the reference rod. Position your eye approximately six inches in front of the locator. By line of sight adjust the locator by inclining it anterio­posteriorly and medio-laterally until a projection of its broad surfaces pass through both posterior reference points as indicated by the posterior reference slides (fig. 45). At this time the anterior reference point marked on the patient’s face may be above or below the refer­ence plane locator.
Adjust the height of the reference rod so that a projection of the locator’s broad
fig. 42 fig. 43
Clamp
Page 26
24
surfaces passes through the anterior reference point as well as the posterior reference points (fig. 45). Then with your left hand pull the locator to the side so that the reference rod is wedged in its clamp. With the wrench in your right hand tighten the clamp to secure the reference rod in its support.
OPTIONAL The Anterior Reference
Pointer, Part No. D145, shown on page 58, is an optional accessory to the face­bow which may be used in lieu of the ref­erence plane locator (ruler) to adjust the reference rod to the anterior reference point marked on the patient’s face.
Now slightly retract the posterior refer­ence slides so that they will not scratch the patient’s face or hurt the ears as you remove the facebow/earbow assembly from the patient’s face. Once removed repostion the slides to their original zero positions. The facebow can now be used to accurately locate the maxillary cast on the articulator.
TRANSFERRING THE FACEBOW/ EARBOW TO THE ARTICULATOR
To prepare the articulator to accept the facebow, set the immediate side shift adjustments to zero, the progressive side shifts to 5 degrees and protrusive condylar paths to 30 degrees. The verti­cal dimension of the incisal pin should also be set to zero (fig. 46).
Next secure a mounting plate to the upper bow of the articulator and a max­illary cast support to the lower bow (fig.
47). The maxillary cast support fits onto the lower bow of the articulator in lieu of a mounting plate. It will help support the weight of the cast and prevent deflec­tion of the facebow.
When attaching a mounting plate to the bow of an articulator always turn the mounting plate in the same direction the lockscrew is turned as the mounting plate is secured to the articulator bow. (fig. 48).
fig. 44 fig. 45
fig. 47
fig. 46
Progressive Side Shift 5˚ Protrusive Adjustment 30˚ Incisal Pin & Table 0˚
All Other Settings 0˚
Page 27
25
A. Facebow Transfer
If the facebow/earbow is used as a face­bow (the posterior reference points are oriented to the posterior reference points marked on the side of the patient’s face in the area of the terminal hinge axis) the facebow will be attached to the articulator by indexing the poste­rior reference pins into the facebow indexes on the lateral aspects of the condyles (fig. 50).
A. Earbow Transfer
If the facebow/earbow is used as an earbow the nylon earplugs are removed for attachment of the bow to the articu­lator and the posterior reference pins are indexed into the earbow index holes on the lateral aspects of the fossae (fig.
50).
If the facebow was oriented to the patient to posterior reference points located precisely with a hinge axis loca­tor, marked readjustment of the posteri­or reference slides of the facebow to accommodate it to the articulator could introduce a slight mounting error. In order to minimize this error, mounting studs are inserted into the lateral aspects of the condylar elements to minimize the amount of adjustment of the posterior reference slides (fig. 51).
A
ttach the facebow (or earbow) assem­bly to the articulator by equally adjusting both posterior reference slides so that both scales give the same reading when the posterior reference pins are firmly
fig. 48
fig. 50
fig. 51
Facebow Index
Earbow Index
Page 28
26
s
eated in their respective condylar or earbow index holes. This can be done most conveniently by indexing both posterior reference pins in their respec­tive holes. Observe the setting to which the slides are adjusted. The sum of these settings divided by 2 is the setting to which the posterior reference slides are adjusted for proper transfer of the facebow to the articulator. (For example, the sum of +30 and -10+=20 divided by 2=+10.)
The maxillary cast support is adjusted so that the support crossbar firmly con­tacts the undersurface of the bitefork without lifting the reference rod from its bearing surface (fig. 52).
fig. 52
Page 29
27
The procedures for constructing casts are not discussed in this manual. Once you have obtained impressions of the upper and lower arches for purposes of constructing the casts, you will in addi­tion require four interocclusal checkbite records in order to mount the mandibu­lar cast in the articulator and to adjust the fossa controls of the articulator: a centric relation record, one right and one left lateral checkbite record and a protrusive record. The selection of the method and material used for obtaining these records is left to the preference of the operator. One recommended check­bite procedure is presented in Appendix A of this manual.
THE MAXILLARY CAST
First secure the maxillary cast to the bite fork with sticky wax or with a light elas­tic band (fig. 53).
With the mounting plate secured in the upper bow of the articulator, fill the mounting plate with stone and also apply stone to the top of the cast. Be sure the mounting stone completely fills the recesses of the mounting plate. The cast is secured to the mounting plate with a “one mix” or “two mix” proce­dure.
“One Mix” Procedure Use one mix
of fast set mounting stone to secure the maxillary cast to the mounting plate. With a little experience a neat mounting can be achieved with this technique.
“Two Mix” Procedure Use one mix
of fast set stone to completely fill the recesses of the mounting plate and to tack the cast to the mounting plate. After the stone is set, remove the mounted cast from the articula­tor and use a second mix of fast set
stone to complete the mounting. The “two mix” procedure is recommend­ed to easily obtain a neat mounting.
Bring the upper bow down on top of the cast, so the stone bonds the two together. Now engage the centric latch. Throughout this procedure be sure that the condyles maintain centric relation and the incisal pin touches the incisal platform (fig. 54). Once the stone has set, remove the maxillary cast support from the articulator.
VI. MOUNTING THE CASTS IN THE ARTICULATOR
fig. 53
fig. 54
Page 30
28
W
ith the facebow application (not the earbow) it is usually easier to fill the mounting plate with stone and stock a pile of mounting stone on the maxillary cast if the maxillary bow is removed and inverted on the working surface beside the facebow-mandibular bow assembly (fig. 55).
THE MANDIBULAR CAST
With the maxillary cast in the articulator, separate the upper bow from the lower bow and turn the upper bow upside down. Orient the mandibular cast to the maxillary cast by accurately seating and luting the centric relation record between the two casts. Secure the casts assembly with sticky wax or a light rubber band (fig. 56).
If the centric relation record was taken at an increased vertical dimension, esti­mate the distance the vertical dimension was increased by the centric relation record and adjust the vertical dimension of the incisal pin to this dimension.
Fill the lower bow mounting plate with fast set stone and put an appropriate amount of stone on the base of the mandibular cast (fig. 57). Then invert the lower bow and seat it on top of the upper bow. Make sure the condyles are seated in their fossae (fig. 58). Make sure that the incisal table is on the incisal pin. Lock the centric latch to its most closed portion. Complete the
m
ounting of the mandibular cast with the “one mix” or “two mix” procedure previously described.
An alternate method of mounting the mandibular cast is to first load the mounting plate with fast set stone and then put an appropriate amount of stone on the base of the mandibular cast. Next grasp the maxillary bow and sup­port the centrically related casts with the thumb, index and middle fingers as illus­trated in fig. 59. Invert the maxillary bow-casts assembly over the mandibu­lar bow and firmly seat the condyles in their fossae. Close the articulator until the incisal pin contacts the incisal table. While maintaining this hand grasp have an assistant move the centric latch to its most closed position. After the initial set of the mounting stone has occurred to
fig. 55
fig. 56
fig. 57
Page 31
29
support the mandibular cast, the fingers are removed from the assembly. Sub­sequently the mounted mandibular cast is removed and the mounting is com­pleted with the “two mix” method.
A recommended procedure is to initially obtain three centric relation check-bite records to verify the accuracy of the centric relation mounting as detailed in Appendix B of this manual.
fig. 58 fig. 59
Page 32
30
To set the articulator to checkbite records you will need the articulator to which have been mounted both the maxillary and mandibular casts. You will also need right and left lateral checkbite records and a protrusive record. (Alternate techniques require fewer checkbite records and use average anatomical dimensions for those condy­lar path dimensions not recorded and measured.)
In this section we will first diagnose and record the characteristics of the patient’s orbiting paths of movement from the two lateral checkbite records, and then diagnose and record the incli­nations of the protrusive condylar paths with protrusive checkbite record.
SIMULATING THE ORBITING CONDYLAR PATH
First remove the maxillary bow from the articulator and confirm that both pro­gressive side shift adjustments are set to the 6˚ settings. The reason for this setting was explained in the previous section on the Immediate and Progressive Side Shift Adjustments.
Next loosen the protrusive and immedi­ate side shift adjustment lockscrews on
both sides of the articulator (a total of four screws). Set the protrusive condylar paths to 0˚ and move the medial fossa walls medially to the limit of their range of movement. Do not tighten the lockscrews.
Seat the right lateral checkbite record on the mandibular casts. Firmly seat the maxillary cast in the checkbite record by grasping the maxillary cast as illustrated in fig. 60 or by applying pressure to the top of the upper bow to immobilize the maxillary cast (due to the fact that the Mark II Articulator has the rotating condylar paths built to average anatom­ic dimensions, impingement of the rotat­ing condyle against its rear and superior fossa walls may sometimes prevent complete seating of the maxillary cast in the checkbite record). At this time the left condyle is positioned inward, down­ward, and forward from its centric relat­ed position (fig. 61). Increase the inclina­tion of the left protrusive condylar path until the superior wall of the fossa con­tacts the top of the condyle (fig. 62). Secure the protrusive condylar path in this position by tightening the lockscrew.
Move the left medial fossa wall laterally until it contacts the condylar element.
VII. SETTING THE ARTICULATOR
TO CHECKBITE RECORDS
fig. 60 fig. 61
Page 33
31
Lock the immediate side shift adjust­ment lockscrew.
Note: These three articulator adjust­ments, the immediate side shift, the pro­gressive side shift and the protrusive inclination of the superior fossa wall establish the character of the orbiting path on the left side of the articulator.
Use the left lateral checkbite record and follow the same procedure to adjust the settings of the right articulator fossa and diagnosis the character of the orbiting path of the right condyle.
Record the articulator settings on the patient’s record.
Note: It is the adjustment of the right
medial fossa wall medialward that allows for a mandibular side shift to the left as the right condyle moves medially to bear and move against its medial fossa wall. Therefore, when the operator writes on the patient’s record “right immediate side shift .6mm” the reference is to the articulator adjustment on the right side of the articulator and not to the side to which the mandible moves. The right articulator adjust­ment will allow for a mandibular side shift to the left. The articulator’s right side is the right side of the
a
rticulator. A medialward adjust­ment of the right medial fossa wall (right immediate side shift adjust­ment of the articulator) allows for a mandibular side shift to the left. Repeat. It is important to note that the right immediate side shift adjustment refers to the articulator setting on the right side of the artic­ulator which allows for a mandibular side shift to the left and not to the right.
SIMULATING THE PROTRUSIVE CONDYLAR PATHS
The inclinations of the protrusive condy­lar paths are diagnosed in the following manner.
Again loosen the lockscrews of the pro­trusive adjustment on both sides of the articulator. Set the protrusive condylar path inclinations to zero degrees. Do not tighten the lockscrews. Seat the protru­sive checkbite record on the mandibular cast and seat the maxillary cast in the checkbite record. Apply downward pressure to the maxillary cast or upper bow to stabilize the maxillary cast in the record. Note that the condyles do not contact their superior fossa walls. Increase the inclination of the protrusive condylar path on both fossae until the superior fossa wall contact their respec­tive condyles. Lock the protrusive adjustments lockscrews. The inclina­tions of the patient’s protrusive condylar path have now been diagnosed. Record the protrusive condylar path settings on the patient’s record.
A recommended manner for utilizing the diagnostic data obtained by adjusting the articulator to protrusive and lateral checkbite records for fixed and remov­able prosthodontics is presented in the “Treatment Procedures” section of this manual on page 35.
fig. 62
Page 34
32
CUSTOM INCISAL TABLE
There are two different custom incisal tables that fit Dénar®Articulators. They are shown in fig 63. Part No. D41 is used with articulators with the long centric adjustment on the foot of the incisal pin. Part No. D4 1 AB is used with articula­tors having the rounded foot on the incisal pin.
To use either of the custom incisal tables first attach a small mount of cold cure acrylic to the posterior portion of the incisal table. Part No. D41 has a machined precision stop on its anterior portion to maintain the correct vertical dimension and care should be taken to ensure that no acrylic is positioned on the top of the stop. Also the long centric
adjustment at the foot of that incisal pin should be turned so that the rounded end will more efficiently mold the acrylic. This is done by adjusting the foot so that its anterior extremity is flush with the anterior surface of its dovetail support.
When the acrylic has reached a rather firm consistency, the articulator is moved in right lateral, left lateral, and protrusive excursions allowing the ante­rior teeth which are kept in contact to guide these excursive movements thereby functionally generating a cus­tom incisal guide. This recording is transferred to the incisal table and small fleur-de-lis is generated in the cold cure acrylic (fig. 64). This is later perfected with a vulcanite burr.
The custom incisal table can be used to best advantage in adjusting to the verti­cal and horizontal overlap relation of the anterior teeth.This is particularly true in adjusting the incisal guidance of the articulator to natural teeth which exhibit varying amounts of horizontal overlap of the teeth which bear the horizontal load in the protrusive, right lateral, and left lateral excursive movements.
ADJUSTABLE INCISAL TABLE
Long Centric Adjustment.
The Long Centric Adjustment is located on the foot of the incisal pin. When there is a horizontal overlap of the anterior teeth, or a “long centric”, the foot of the incisal pin is adjusted to complement this tooth relationship in the following manner. Set the protrusive inclination of the incisal table to its maximum angle. Loosen the foot of the incisal pin and slide it forward (fig. 65). Move the upper bow posteriorly until the lower anterior teeth contact the lingual surface of the upper anterior teeth (fig. 66).
VIII. INCISAL TABLE ADJUSTMENTS
fig. 64
fig. 63
Page 35
33
With the right index finger, push the foot of the incisal pin back until it just con­tacts the inclined platform of the incisal table, and tighten the incisal pin foot lockscrew in that position (fig. 67). By allowing the incisal pin to come forward into the centric related position again, a space will be noticeable between the foot of the pin and the incisal table
__
this space is equal to the horizontal overlap of the anterior teeth. (fig. 68).
Protrusive Adjustment
To adjust the incisal table for the inclina­tion of the lingual bearing surfaces of the anterior teeth, first loosen the incisal platform and set it back to zero. Do not tighten the lockscrew. Then, move the upper bow posteriorly until the incisal edges of the lower anterior teeth con-
tact the lingual surface of the upper anterior teeth just lingual to their incisal edges. Note that the incisal pin does not touch the incisal table now (fig. 69). Increase the angle of the table until it touches the foot and lock the table in that position (fig. 70).
Lateral Wing Adjustment
The lateral wings of the incisal table are adjusted to complement the lateral rela­tionships of the anterior teeth
__
most frequently the vertical and horizontal overlap relationships of the cuspid teeth. To make this adjustment, first remove the incisal table and loosen the lateral wing lockscrews (fig. 71). Then replace the incisal table on the articula­tor to observe visually the angle to which the lateral wings must be adjusted to
fig. 67fig. 65
fig. 68fig. 66
Page 36
34
complement the anterior tooth relation­ships in the following manner. Hold the articulator in a right lateral mandibular position so that the right cuspids are in function. (The upper bow is pushed to the left. ) Note that the incisal pin does not contact the incisal table. With the right index finger, reach behind the tilt­ing platform, and elevate the left wing of
the table until it touches the foot of the incisal pin (fig. 72). Visually take a read­ing on the front scale to note its angular setting. Do this same procedure on the other side wing. Now remove the incisal table from the articulator and tighten the lockscrew wings at those settings. Then attach the table to the articulator again.
fig. 69 fig. 71
fig. 70
fig. 72
Scale
Page 37
35
The rationale for utilizing the diagnostic data obtained from protrusive and later­al checkbite records is as follows.
When the protrusive inclination of the superior fossa wall is adjusted to the lat­eral checkbite record, a characteristic of the orbiting condylar path is diagnosed. This characteristic is associated with the balancing inclines of posterior teeth on the orbiting side__the mandibular buc­cal cusps’ lingual inclines’ mesial aspects and the maxillary lingual cusps’ buccal inclines distal aspects.
When the protrusive inclination of the superior fossa wall is adjusted to the protrusive checkbite record, the inclina­tion of the patient’s protrusive condylar path is diagnosed. This inclination of the superior fossa wall is associated with the protrusive contacts of posterior teeth__the mesial aspects of mandibular cusps and the distal aspects of maxil­lary cusps.
The orbiting path inclination of the supe­rior fossa wall adjusted to lateral check­bite records is always equal to or greater than the protrusive path inclination of the superior fossa wall adjusted to the protrusive checkbite record.
FIXED RESTORATION AND REMOVABLE PARTIAL DENTURE RESTORATIONS
Adjusting the protrusive inclination of the superior fossa to an angle which is slightly less than the patient’s protrusive condylar path (5 to 10 degrees less) when the restoration is fabricated will prevent the fabrication of protrusive contacts, or balancing contacts on the orbiting side of posterior teeth in the laboratory. This is due to the fact that when the restoration is seated in the patient’s mouth and the patient’s
condyle tracks a steeper protrusive and orbiting condylar path, the posterior teeth will separate in both the protrusive excursion and in the lateral excursion on the orbiting side.
COMPLETE DENTURES
Method 1. Adjusting the articulator to
the patient’s protrusive condylar path inclinations for both the pro­trusive and lateral excursive move­ments allows the fabrication of pro­trusive balance in the laboratory, When the restoration is seated in the patient’s mouth and the patient executes a lateral mandibular movement with the teeth in contact, the patient will feel the primary occluding pressures on the working side. If the patient’s orbiting path is slightly steeper than the articulator setting when the restoration was fabricated, the patient would per­ceive minimal occluding pressures on the balancing side. However orbiting side occlusal contact would prevent loss of peripheral seal.
Method 2. Adjusting the inclinations of
the superior fossa wall to the patient’s orbiting path inclinations allow the fabrication of bilateral bal­anced occlusion in lateral excursive movements in the laboratory. Subsequent adjustment of the articulator to the patient’s protru­sive condylar path inclinations per­mit adjustment of the occlusion to protrusive balance. This method permits development of full arch balanced occlusion. (Note: Ob­taining accurate eccentric check­bite records on extensive tissue borne restorations, although theo­retically attainable, is a difficult accomplishment).
IX. TREATMENT PROCEDURES
Page 38
36
CHECKBITE PROCEDURE
The mandibular cast is transferred to the articulator by:
1. occluding it with the maxillary cast
2. checkbite procedure
3. a combination of the above
In the transfer of the mandibular cast to the articulator, if the mandibular cast can be occluded accurately in the desired most intercusped position, the most accurate way to transfer the mandibular cast to the articulator is by occluding it with the maxillary cast. However, if deflective contacts exist in the patient’s mouth, they must either be removed by grinding prior to the time impressions are taken so the cast can be accurately oriented in the desired most intercusped position or the centric relation record must be taken at an increased vertical dimension by check­bite procedure.
There are many satisfactory recording agents such as waxes, bite registration pastes, etc., and many techniques with which these agents can be employed to advantage in checkbite registration pro­cedures. However, there is no universal method which proves optimum for all situations. Therefore, the dentist must be knowledgeable of the anatomy of the related structures, neuromuscle physiol­ogy, mandible manipulation techniques and in the physical properties of the dental materials employed which deter­mine checkbite accuracy. This will enable him to effect a method of choice as varying situations present them­selves.
It is not within the scope of this text to present an exhaustive discourse on checkbite procedure. However, one rec­ommended procedure which effects basic principles of checkbite techniques is presented. Variations of this technique
can be employed as circumstances dic­tate.
The following basic principles should always be employed in any checkbite technique.
1. The checkbite record should be obtained as close to the vertical dimension at which the restoration is to be constructed as possible.
2. The dentist should always manipu­late the patient’s mandible in obtaining the checkbite. The patient must not be instructed or allowed to bite or close down in such a manner as to cause flexion of the mandible or movement of teeth in obtaining the record.
3. The recording media interposed between the teeth should be of a dead soft consistency as the mandible is brought to the desired position. This prevents deflection of the mandible and/or mobile teeth, or the programming of mus­cle function.
TECHNIQUE FOR WORKING CASTS
The following technique can be used to register the mandiblular position at the correct vertical dimension or at an increased vertical dimension. It can be used to register centric or eccentric position. Described will be the registra­tion of the centric relation position of the mandible at the minimal increase of ver­tical dimension which will result in lack of posterior tooth contact.
Fabricate an occlusal programmer by softening a small piece of base plate wax in a water bath to a soft consisten­cy and adapt it over several teeth in the anterior segment of the mouth as illus­trated in fig. 73. Manipulate the mandible
APPENDIX A
Page 39
37
i
nto anterior centric relation and gently guide the patient repeatingly in terminal hinge closure to lightly tap the soft wax to establish simultaneous contact of several mandibular teeth on the wax. The wax is removed from the mouth and chilled in tap water. The number of thicknesses of wax used should be the minimum number to effect posterior dis­clusion.
A small amount of denture adhesive cream is applied to the wax occlusal programmer in the area that adapts to the lingual surfaces of the maxillary anterior teeth (fig. 74) The occlusal pro­grammer is repositioned in the mouth by reindexing it to the maxillary teeth. The occlusal programmer thus establishes a predetermined stop to vertical closure with the condyles in centric relation. There is an absence of contacting deflective inclines posteriorly as a result of group functioning of teeth in the ante­rior segment of the mouth where the occluding pressures are the least. The occlusal programmer reprograms mus­cle function so as to eliminate resistance to terminal hinge closure previously pro­grammed by deflective contacts.The dentist can now more easily manipulate the mandible in terminal hinge closure to obtain the centric relation checkbite record.
Similar occlusal programmers can be fabricated in eccentric positions to facil­itate obtaining of eccentric checkbite records.
Suitable checkbite recording media such as Bosworth’s Superbite ( a rapid and hard setting zinc oxide and eugenol base material) is placed on the mandibular posterior teeth and the mandible is guided in terminal hinge clo­sure until the anterior teeth make light contact with the occlusal programmer. Since it is undesirable to register deep fissures, grooves and undercuts on teeth and cavity preparations with the
checkbite record media, these areas are blocked out by wiping Vaseline or soft wax over the teeth with the index finger before the zinc oxide and eugenol paste material is used for checkbite registra­tion. The checkbite record media is allowed to harden without the patient executing a biting pressure. Figure 75 illustrates an occlusal programmer and checkbite records accomplished with a hard fast setting zinc oxide and eugenol (ZOE) base material.
If the patient brings the mandible mov­ing muscles into function during the set­ting of the record media, as may occur during swallowing or bruxing, flexion of the mandible or displacement of teeth may occur which would result in an inaccurate record. To insure that the patient does not bring the muscles into
fig. 73
fig. 74
Page 40
38
function during the securing of the cen­tric relation checkbite record, the dentist should periodically gently arc the mandible in small terminal hinge rotary movements preceding the set of the record media and immediately following the set. Resistance to this movement immediately prior to or after the set indi­cates functions of the patient’s mandible moving muscles and the checkbite record should be discarded and a new one taken. If resistance to this move­ment is not encountered, the dentist can be reasonable sure the patient did not effect muscle function. This technique is termed “dynamic technique” and the record termed “dynamic checkbite” because of the small movements of the patients mandible executed by the den­tist during the securing of the record.
The record media selected for this tech­nique should allow an adequate, consis-
t
ent and not extended working time. It should have a flash set to a hard con­sistence that will fracture without warp­ing so the parts can be accurately reassembled if fractured. The material should be of a dead soft consistency when mixed but it must also have suffi­cient body to resist flow so that it will stay in place when positioned on the teeth.
To facilitate the placement of the record media on the occlusal surfaces of the mandibular teeth construct small paper cones from sheets of paper approxi­mately 4 x 6 inches in dimension. Twist the papter around the index finger as illustrated in fig. 76 to fashion a cone that has an opening at its large end approximately 1-1/4 inches in diameter (fig. 77A). The margin of the paper is sealed with Scotch Tape so the paper will not unwrap. The small end of the
fig. 75
fig. 76
fig. 77
Page 41
39
c
one is cut with a scissors to produce an opening at the apex approximately 1/8 inch in diameter (fig. 77B).
Mix record media and gather it into one mass on the end of the cement spatula. Load the media into the bottom of a paper cone (fig. 77C). Squeeze the paper cone behind the record media to capture the media near the apex of the cone and to clean the spatula as it is withdrawn (fig. 77D). After the spatula is removed the cone is squeezed with the index finger and thumb of both hands to push all the media to the apex of the cone and then the cone is folded back on itself to prevent movement of the media toward the large end of the cone when it is compressed (fig. 77E).
The cone is squeezed to dispense the media into the central groove of the mandibular teeth (fig. 78). As soon as a cone is loaded the operator can pro­ceed to apply the media to the teeth as the assistant immediately begins to mix another batch of material for securing additional centric relation records.
An inexpensive file jacket is used as a mixing pad and provides for convenient disposal of the record media after use. It is good procedure to obtain three cen­tric relation records. One is used to mount the mandibular cast and the remaining two are used to double check the mounting for accuracy with the Denar
®
Vericheck technique presented
in Appendix B of this manual.
The setting times of zinc oxide eugenol record medias vary with temperature, humidity, spatulation time and batch number. Under certain conditions the material will set very rapidly and the operator must work quickly to get the material to place before the set occurs. Alternately the material may set slowly. Since it is difficult for the operator to maintain the patient’s mandible in cen­tric relation for extended periods of
t
ime,
time, the operator using a slow setting record media should wait for a period of time after the media is placed on the mandibular teeth until the set is about to occur before manipulating the patient’s mandible to centric relation. Therefore, the operator should dispense record media for four batches of material on the mixing pad (fig. 79). The first batch is use to time the set of the media and is discarded. The remaining three batches are used to obtain the centric relation records.
To time the set, mix the first batch of material according to the manufacturers recommendations. Load the cone and apply the first batch of record media to the mandibular teeth. Do not bring the mandible to centric relation. The opera­tor should then start to count slowly until the set occurs. Once the set has
fig. 78
fig. 79
Page 42
40
b
een timed the operator knows how long to pause after the media has been placed on the lower teeth before bring­ing the mandible to centric relation. It is desirable to bring the mandible to cen­tric relation just before the set occurs. For best results it is important that the operator and assistant work with consis­tent technique throughout this procedure.
TECHNIQUE FOR DIAGNOSTIC CASTS
Checkbite records made of hard base plate wax can be used to advantage when mounting diagnostic casts. The wax used should be of a hard, brittle consistency and fracture without bend­ing at room temperature. For obtaining centric relation records develop wax wafers composed of two thicknesses of wax to the dimension illustrated in fig.
80. It is best not to include the anterior teeth in the centric relation record if suf­ficient posterior teeth are present to sta­bilize the casts in the record.
To develop the wax wafers soften a sheet of base plate wax in a water bath or open flame. Flame one surface of the wax and fold the wax over on itself on the flamed surface to bond the wax into a wafer consisting of two thicknesses of wax. With a warm spatula trim the wax slightly oversize of the desired dimen­sion. With the wax slightly softened so the wafer is pliable, position the wafer on the maxillary teeth and using the index fingers adapt the wax to the teeth to obtain very light indexing impressions of the cusps tips and to contour the wafer to the plane of occlusion. Do not impress the mandibular teeth into the wax at this time. Chill the wax with air so that it will maintain the shape of the plane of occlusion and remove the wafer from the mouth. With a warm scalpel trim the wafer to the desired dimensions. Check the wafer in the mouth or on a maxillary cast to confirm that the wafer is contoured to the plane of occlusion and is stable on the maxillary teeth.
Soften the lateral edges of the wax in an open flame until it is dead soft. Quickly temper the wax in a water bath and position the wafer on the dry maxillary teeth to obtain light impressions of the tips of the cusps as previously described. Guide the mandible in cen­tric relation closure to obtain light impressions of the mandibular cusp tips in the record. It is important that the operator guides the mandible to the desired position.The patient should not be instructed to bite down or close. As soon as the impressions are obtained, chill the wax with air and remove it from the mouth. The record is chilled in cool water and returned to the mouth and checked for accuracy. The wafer is placed on the maxillary teeth and the mandible is arced in centric relation clo­sure to occlusal contact. The mandibu­lar cusp tips should come into simulta­neous even contact with their impres­sions in the record. When the mandibu­lar teeth are brought into extemely light occlusal contact with the record the patient should preceive no premature contact or uneven pressure on ques­tioning.
Lateral Checkbite Records
A technique for fabricating a right lateral checkbite record is described. The left lateral record is obtained in a similar manner.
Train the patient to allow you to arc the mandible in hinge retation with the patient’s rotating condyle in its most
fig. 80
F
igure 80 courtesy
Dr. P. Dawson and C.V. Mosby Co.
Page 43
41
retruded position and the orbiting condyle advanced approximately seven millimeter. This can be done with the “thumb on chin” method with the thumb on the anterio-inferior border of the mandible in the area of the left cuspid.
Construct a wax wafer composed of two thicknesses of hard base plate wax to the dimensions illustrated in fig. 81. This is done by fabricating an oversized wafer as previously described and adapting it to the maxillary teeth with the index fingers to conform the wafer to the plane of occlusion and to obtain light imprints of the cusp tips. Chill the wafer with air and remove it from the mouth. Use a warm scalpel to trim the wafer to the desired dimensions. Return the wafer to the mouth, and position it on the maxillary teeth to confirm that it is stable on the maxillary teeth and con­forms to the plane of occlusion. Guide the mandible in right lateral closure to contact the wax and determine the
number of additional strips of wax which must be added to the lateral extremities of the mandibular side of the wafer to establish even contact with the mandibular teeth. Lute additional thick­nesses of wax to the mandibular side and then heat the lateral extremities of the wafer in an open flame to soften the wax. Quickly temper the wax in a wafer bath and position the wafer on the dry maxillary teeth. Guide the mandible in right lateral closure to obtain light imprints of the cusp tips in the wax. Chill the wax with air and remove the wafer from the mouth and place it in cool water. The wafer can be returned to the mouth and checked for accuracy.
fig. 81
Page 44
42
The following protocol was developed by a panel of experts in occlusal treat­ment on the philosophy that a dentist seeking good laboratory support will give the technician complete specifica­tions for fabrication of the occlusal aspects of the restoration.
Recognizing that the position of maxi­mum intercuspation is of paramount importance, it is the responsibility of the dentist who takes the centric relation or centric occlusion interocclusal record to mount the mandibular cast. Complete specifications furnished by the dentist to his laboratory for fabrication of the occlusal aspects of restorations include:
1. Full arch master casts mounted in an instrument with trimmed dies or dies with margins that are easily identifiable.
2. Instructions for adjusting articula­tor condylar controls (if adjustable) when the restoration is fabricated plus records for adjustment of the incisal table.
3. Specifications as to which teeth are to bear the load in eccentric bruxing movements.
4. Specifications for the class of occlusal anatomy desired and the character of the position of maxi­mum intercuspation as described below.
DISCUSSION
Mounted Casts with Trimmed Dies
To specify that it is the responsibility of the dentist to mount the mandibular cast does not mean that he cannot del­egate this procedure to a dental auxil­iary. For example, in limited restorative procedures in which there are sufficient unprepared teeth to accurately index full
arch casts in the position of maximum intercuspation, the mounting procedure could certainly be delegated to a dental auxiliary. However, as the restoration becomes more extensive, if the dentist delegates the mounting of the casts to a dental auxiliary and an error is intro­duced, the dentist should still assume full responsibility for the mounting error.
Instructions for Adjustment of Condylar Controls
To specify that the dentist should pro­vide instructions for adjusting condylar controls does not mean that the restora­tion is to be constructed on a fully adjustable articulator. An example of instructions for adjustment of the condylar controls could be: “Lock the instrument in centric relation throughout the laboratory procedure”. It is the responsibility of the dentist to specify to his technician if he wants consideration given to eccentric factors of occlusion in the laboratory, and if so, to what extent.
Dentist__Laboratory Relations
Currently, most dentists send unmount­ed casts to the laboratory for fabrication of prosthetic restorations. Restorations are returned to the dentist on unmount­ed casts. If the occlusion is found to be in error when the restoration is inserted in the patient’s mouth, who made the error__dentist or technician? This condi­tion can result in strained dentist-labo­ratory relations.
It is the responsibility of the dentist who takes the centric relation record to mount the mandibular cast. When the restoration is returned to the dentist on full arch mounted casts the dentist has a basis for accurate communication with his laboratory. If the laboratory services
APPENDIX B
DÉNAR SYSTEM PROTOCOL FOR DENTIST-LABORATORY RELATIONS
Page 45
43
a
re found to be satisfactory and occlusal discrepancies are identified in the mouth, the dentist is confident that the error was not introduced in the labo­ratory. He can then re-evaluate his pro­cedure to determine the source of error. In other words, if the restoration fits the mounted casts but not the mouth, the technician cannot be expected to par­ticipate in the additional cost of remakes not attributed to laboratory procedures.
CLASSIFICATIONS OF OCCLUSAL CARVINGS
The following classification of occlusal carvings as illustrated in figure 82 enables the dentist to specify to the lab­oratory the relative amount of time and effort he desires the laboratory to expend (and consequently the laborato­ry fee) to fabricate the prescribed occlu­sion.
Alloy Restorations
Class A__Refined Occlusal Carving
The restoration is fabricated in the man­ner the technician might employ when using a simple crown and bridge hinge mechanism. When the occlusion is semi-developed, the technician simply bruxes the instrument to burnish away eccentric interferences and refines the anatomy of the remaining wax. Eccentric irritations are removed from the occlusion in the laboratory. This is the application of principles of occlusion in its simplest form and consequently is a minimum fee laboratory service.
Class AA__Modified Drop Wax
Basically, this is a Class A carving with more refinement. The technician carves in supplemental grooves and then accentuates triangular, cuspal and mar­ginal ridges with the drop wax tech­nique. The ultimate appearance will be similar to a Class AAA carving. A Class AA carving requires more laboratory time than a Class A carving and so the
laboratory fee should reflect 10% to 30% increase in the labor factor of the Class A carving.
Class AAA__Complete Drop Wax
The drop wax technique is employed to restore the total occlusal surface and all restored axial surfaces to obtain the maximum esthetic cusp height, opti­mum cusp distribution, and to harmo­nize the occlusal anatomy with condylar paths of movement. This type of occlusal fabrication requires more time and skill than a Class A or Class AA carving and consequently the laboratory fee reflects a 50% to 100% increase in the labor factor of a Class A carving. However, this can only be done on a restoration involving extensive tooth reduction. For example, if the restora­tions are MOD inlays or onlays with min­imum tooth reduction, the cusp distribu­tion has already been accomplished and the laboratory only needs to do the fill­in-procedures of the drop wax tech­nique. Consequently, a Class AA or AA+ carving would suffice.
Centric Contacts
The character of the centric contacts must produce axial loading of the teeth. This can be accomplished by:
1. Point Contact (•). There is an absence of contacting inclines.
fig. 82
Page 46
44
C
entric stops are on high spots or
low spots of cusps.
2. Tripod Contacts (...). Centric cusps do not hit on their tips but are sup­ported by three or sometimes only two centric contacts on the perimeter of cusp tips.
Developing tripodized centric contacts (...) requires more laboratory time than point contacts (•). Consequently, an additional laboratory fee is required. When the dentist’s prescription speci­fies a Class AA (...) occlusion, it means he desires a Class AA carving with tripodized centric stops and he antici­pates a laboratory fee which will allow the technician adequate time to perform this service.
PORCELAIN RESTORATIONS
An extension of the above classification for alloy restorations is used for specifi­cations of porcelain restorations.
Class A porcelain restorations are carved to look like Class A alloy restora­tions. Class AA porcelain restorations are carved with more detail to accentu­ate supplemental grooves and cusp height to look like a Class AA or AAA alloy restoration. A plus symbol (+) denotes additional staining and charac­terization.Thus a Class AA (•)+ porcelain restoration is carved to be similar to a Class AA alloy, has point contact centric stops with an absence of contacting inclines and has special staining charac­terization.
DISCUSSION
Laboratory Fees and Instruments for Occlusal Treatment
The laboratory fee is for the time and skill which the technician employs in harmonizing the occlusion to condylar paths of movement and is not deter­mined by the instrument used. A fully adjustable or semi-adjustable articulator can be bruxed as easily as a simple crown and bridge hinge mechanism that has springs for condylar posts.
The above classification of occlusal carvings is not an attempt to establish laboratory fees. It is offered as a means of establishing more accurate communi­cations between dentist and laboratory.
SUMMARY
The protocol which requires the dentist to mount the mandibular cast gives the dentist “laboratory-control”. This control of services usually results in better labo­ratory work. The laboratory on the other hand receives “dentist control” and less reduced fee remake discussions.
The dentist who mounts his mandibular cast and provides the technician with complete specifications for fabrication of the occlusal aspects of the restora­tion enjoys the benefits of improved communications with his laboratory, laboratory control, better laboratory work, and improved dentist__laborato­ry relations.
Page 47
45
The Dénar®Field Inspection Gage is an optical inspection gage used to cali­brate Dénar
®
Articulators and laboratory Relators to tolerances which allow the transfer of mounted casts between cali­brated articulators and Lab Relators.
NOTE: This manual details the use of the Dénar
®
Field Inspection Gage to cal-
ibrate the Dénar
®
Mark II Semi­adjustable Articulator. It does not detail the calibration procedure for the Dénar
®
D5A Fully Adjustable Articulator and D6 Centric Lab Relator; nor does it describe the recommended applications of the Dénar
®
Two-Instrument System. For complete information on these subjects the reader is referred to the Dénar
®
Office Tutor (Cat. #D1269) and the Dénar
®
Field Inspection Gage
Instruction Manual (Cat. #D166).
The gage consists of an upper member called the “scope” and a lower member called the “stage.” The gage is supplied with two inter-bow gage pins. Inter-Bow Gage Pin #1 and Inter-Bow Gage Pin #2 (fig. 83).
CARE OF DÉNAR
®
FIELD
INSPECTION GAGE
The Dénar
®
Field Inspection Gage is a precision instrument capable of making measurements in three directions to an accuracy of one ten thousandths of an inch (.0001”). To insure its continued accuracy, the components should be handled carefully and not be allowed to come to rest on a hard surface with a force which would cause a nick or deformation of the bearing surface. When a component is not in use, it should be kept in its protective case. It is advisable to have a piece of cork, rub­ber, cardboard or other soft, lint-free surface to rest the gage components on when they are not mounted in their case or on a Dénar
®
instrument.
IMPORTANT There are several socket set screws on the scope member of the Field Inspection Gage. Do not attempt to turn these screws with a wrench. These screws are for use of factory authorized personnel only. Manipulation of these screws can affect the accuracy of the instrument.
For accuracy, it is important that the bearing surfaces of the Field Inspection Gage components, Articulator, Lab Relator, as well as the mounting plates, be wiped clean of foreign material prior to their use.
VERTICAL ALIGNMENT
Prior to using the Field Inspection Gage to align a Dénar
®
Articulator or Lab Relator the distance between the maxil­lary and mandibular bows (inter-bow distance) of the articulator or Lab Relator must be precisely established by means of the inter-bow gage pin #1, as illustrated in fig. 84.
APPENDIX C
CALIBRATION PROCEDURE
fig. 83
Page 48
46
To use this pin the adjustable incisal pin and the incisal table are removed and the inter-bow gage pin #1 is secured in position as shown.
Carefully and without forcing, mount the gage components flush to the articula­tor bows. Make sure to turn the gage component relative to the articulator bow in the same direction to which the lockscrew is turned as the gage is secured to the bow of the articulator (fig.
85).
It is important that the operator use a consistent amount of torque when tight­ening lockscrews to secure mounting plates or a Field Inspection Gage to the bows of an articulator or Lab Relator.
Tighten the lockscrew without forcing.
W
hen placing a gage component onto the bow of an articulator the index holes on the surface of the gage that faces the articulator bow should be carefully and visually located over their respective index pin on the bow of the articulator to prevent damage of the bearing surface of the gage. DO NOT SLIDE THE GAGE laterally over the index pins in an attempt to locate the index pins in the gage by feel. This can cause damage if done with too much force.
Figure 86 illustrates the Field Inspection Gage and inter-bow gage pin 1 mount­ed in the articulator. At this time you will note that a space of approximately 1/4 to 1/2 millimeter exists between the scope and the stage.
Optional: If you wish to confirm that your gage is accurate proceed in the follow­ing manner. On the back of the stage you will find a calibration record as illus­trated in fig. 87. The two large circles represent the faces of the dial indicators on the right and left sides of the gage as viewed from the rear. On each circle is a scribed calibration mark which is usual­ly found between the +10 and +20 grad­uations. Each graduation represents one half-thousandth of an inch (.0005”). These readings are the values which the dial indicators on the respective sides of the instruments should read when the scope is allowed to rest on the stage
fig. 84
fig. 85
fig. 86
Space
Page 49
47
by loosening the maxillary lockscrew while the Field Inspection Gage is mounted in an articulator or Lab Relator.
To confirm that your field gage is accu­rate, simply loosen the maxillary lockscrew which secures the scope to the maxillary bow and firmly press on the lockscrew as illustrated in fig. 88 to insure that the scope rests on the stage.
If the dial indicators do not register the calibration values scribed on the cali­bration record on back of the stage, simply loosen the lockscrew on the face of the dial indicator (fig. 88) and rotate the face of the indicator until the indica­tor reads the desired value and then lock the face in position. The field gage is now calibrated. Resecure the scope to the face of the maxillary bow, remem­bering to make sure that you rotate the face of the indicator until the indicator reads the desired value and then lock the face in position. The field gage is now calibrated. Resecure the scope to
the face of the maxillary bow, remem­bering to make sure that you rotate the scope in the same direction that you turn the lockscrew as you secure the scope to the face of the articulator. At this time the dial indicators will register “O” if both condylar elements of the articulator are at the correct height.
If the height of the condylar elements are out of specification proceed in the following manner. On the posterior aspects of the horizontal crossbar are vertical adjustments (A) and vertical adjustment lockscrews (B) as illustrated in fig. 89. Loosen the vertical adjustment lockscrews on both sides of the articu­lator. While applying downward pres­sure on the maxillary bow above the condyles, manipulate both vertical adjustments until both dial indicators read +5 to +10. A plus reading indicates that the condyles are below the desired height. Note that when you change the height of one condyle it affects the read­ing on both dial indicators. Snug the vertical adjustment lockscrews to firmly seat the pins supporting the condylar elements in their housings but do not tighten all the way. At this time both condylar elements are just below the desired height. Alternately tighten the vertical adjustment screws in small increments on the right and left sides of the instrument until both dial indicators read “O”. Tighten the vertical
fig. 87
fig. 88
Page 50
48
adjustment lockscrews to secure the condylar elements in this position.
HORIZONTAL ALIGNMENT
The articulator must be in specification in the vertical dimension before horizon­tal alignment is made.
On the scope member of the Field Inspection Gage are two forty power (40x) monocular scopes. You are to look through these scopes one at a time. In each scope is a reticle graduate in one thousandths of an inch (.001”) incre­ments in four directions off of a centric dot as illustrated in figure 90A. On the lateral wings of the stage are crosshair targets as illustrated in figure 90B. If an articulator is in perfect specification anterior-posteriorly and medio-laterally, the intersection of the target crosshairs will touch the centric dot of the reticle as in figure 90C when viewed through the scope. If the articulator is out of specifi­cation as illustrated in figure 91, the amount it is out of specification can be measured with the graduated reticle. For example, figure 91 illustrates the view through the scope of an articulator out of specification ten thousandths of an inch (.010”) medio-laterally and eight thousandths of an inch (.008”) anterio­posteriorly.
To align an articulator in the horizontal plane, proceed in the following manner.
O
n the back of the stage is a calibration record as illustrated in figure 87. The two small circles on the lower portion of the record are the calibration records for the scope’s on the respective sides of the articulator. Each graduation off of the centric dot represents one thou­sandths of an inch (.001”). the total diameter of each small circle represents a little over four thousandths of an inch (.004”), or about the thickness of normal writing paper. In each of the small circles you find a scribed cross mark. This mark represents the true centric position. In other words, when you look through the monocular scopes, the true centric position in some gages is that position indicated by the scribed cross marked on the calibration record.
fig. 89
fig. 90
fig. 91
A
B
Page 51
49
T
his calibration system is used because in manufacture, slight creepage of the centric dot of the reticle can occur. This calibration system allows for corrections of this error and maximum accuracy.
In the following instructions to adjust the horizontal alignment of the articulator, it is assumed that the scribed crosses marked on the calibration record coin­cides with the centric dots of the reticle.
Loosen the horizontal adjustment lockscrews illustrated in figure 92 on both sides of the articulator the mini­mum amount necessary to allow hori­zontal crossbar. Engage the centric latch of the articulator. While maintain­ing slight downward pressure on the horizontal crossbar of the lower member of the articulator to keep the horizontal crossbar seated flush on the crossbar supports, slide the horizontal crossbar and maxillary bow assembly in the hori­zontal plane until the centric dots are on the junction of the crossbar targets as viewed through both scopes. Then while carefully maintaining this crossbar posi­tion incrementally tighten in a crisscross sequence the four horizontal adjustment lockscrews.
The articulator in now calibrated in the centric position.
RUN-OUT INSPECTION PROCEDURE
The articulator was aligned by means of the Field Inspection Gage with the pro­trusive condylar path inclinations set to 30 degrees. Readjust the right protru­sive condylar path to 10 degrees and tighten the lockscrew. Record the read­ing on the right dial indicator. The two readings indicate the amount of run-out of the articulator adjustment. Factory specifications require that all articulators have a run-out of less than plus or minus five thousandths of an inch (+.005”).
REINSERTION OF THE ADJUSTABLE INCISAL PIN
After an articulator is properly aligned by means of a Field Inspection Gage, the most convenient way to set the adjustable incisal pin to an accurate zero position is to remove the incisal pin #1 and reinsert the incisal table and adjustable incisal pin set to zero. If the dial indicators do not read zero the height of the adjustable incisal pin can be minutely adjusted until the dial indi­cators read zero.
INTER-BOW GAGE PIN #2
Inter-bow gage pin #2 is used to pre­cisely adjust incisal pins to the zero dimension in articulators and Lab Relators known to be in specification. The incisal pin is raised and the inter­bow distance is established by means of the inter-bow gage pin #2, as illustrat­ed in figure 93.
The adjustable incisal pin is then adjust­ed and secured to maintain the inter­bow distance established by the inter­bow gage pin #2. It is important to apply downward pressure on the top of the adjustable incisal pin as the lockscrew is tightened to prevent upward creepage of the incisal pin when tightening the lockscrews as illustrated in figure 94.
fig. 92
Horizontal Adjustment Screws
Page 52
50
Use of the inter-bow gage pin #2 requires a little care and skill. Therefore, after this adjustment is made be sure to check carefully to confirm that the inter-bow dimension of the gage pin was in fact transferred to the adjustable incisal pin.
fig. 93 fig. 94
Page 53
51
The Broadrick Occlusal Plane Analyzer is used for analyzing the Curve of Spee and developing an acceptable curve of occlusion, and is recommended for lab­oratory procedures employing the Functionally Generated Path Technique.
Note: This text describes the mechanical function of the Broadrick Occlusal Plane Analyzer. The reader is urged to refer to other textual material as well as classes devoted to the clinical application of this pro­cedure in occlusal correction.
A Broadrick Occlusal Plane Analyzer, fig. 95, consists of (1) Card Index #300061, (1) Bow Compass #300087 with graphite leads of varying hardness,
1
(1) extra center points #300088, (`1) Scribing Knife #3017, (1) Needle Point #300089, and (2) Plastic Record Cards #300066.
To use the Occlusal Plane Analyzer the casts are mounted in the articulator and the articulator condyles and incisal con­trols are adjusted in the conventional manner. The Occlusal Plane Analyzer is attached to the upper bow of the articu­lator as illustrated in fig. 96 by securing it beneath the screws which hold the maxillary mounting plate and incisal pin assembly in place.
It is pointed out that during the ensuing procedure, the maxillary cast must be removed from the articulator during the survey.
Secure a plastic record card on the right side of the flag. The plastic record cards are matte finished on both sides and readily accept pencil or ink markings.
The relatively small divergence between arcs of 3 3/4”, 4”, and 5” radii over the functional occlusal surfaces on the lower posterior teeth is shown in fig. 97. An average of 4” radius may be used in the majority of surveyed cases.Variation is only necessary when a pronounced Curve of Spee may require a selection of up to a 5” radius.
APPENDIX D
OCCLUSAL PLANE ANALYZER
1. Graphite leads of preferred hardness may be obtained locally from drafting supply houses.
fig. 95
fig. 96
Page 54
52
A view through both lower second molars, fig. 98, illustrates the small divergence between arcs of the same three radii at the functional occlusal sur­faces on the Curve of Wilson.
Insert a piece of graphite lead into the bow compass, tighten thumbscrew, and sharpen to a suitable point. Adjust the bow compass to the radius selected (in this instance 4”) (fig. 99).
Remove the upper cast, fig. 100, and position the center point of the bow compass, set at the 4” radius, on the anterior survey point (A.S.P.) which is usually the disto-incisal angle of the cuspid. If the cuspid is worn flat, the A.S.P. may be at the incisal edge. In any event, this point must be selected as the most desirable to “beam” the line and plane of occlusion posteriorally. Once selected, it is marked on the cuspid and NOT CHANGED. With the center point of the bow compass positioned on the A.S.P., apply a long arc with the graphite lead (about 3”) on the plastic record card. The occlusal plane survey center
(
O.P.S.C.) will ultimately be located on
some point on this arc.
Select the posterior survey point (P.S.P.) at the disto-buccal cusp tip of the last lower molar (fig. 101). Should non molars exist, replace the upper cast and place soft modeling compound over the lower ridge and close the articulator until the incisal in contacts the incisal guide in centric relation. Chill the com­pound and carve away any excess, leaving only the compound contacting the upper fossae (simulating the lower buccal cusp). Remove the upper cast and select a P.S.P. on the modeling compound in the same manner as the P.S.P. was selected on the last molar as described above.
Position the center point of the bow compass on the P.S.P. and apply an arc with the graphite lead to intersect the arc from the A.S.P.
Alternate to the molar P.S.P. is a position on the condylar element of the articulat-
fig. 97
fig. 98
fig. 99
Page 55
53
or at the middle of the anterior margin of the hole on the lateral aspect of the condyle (fig. 102). Position the center point of the bow compass on the condylar posterior survey point (C.P.S.P.) and apply an arc with the graphite lead to intersect the arc formed from the A.S.P.
Continue with fig. 101 or 102 and sub­stitute the needle point for the graphite lead. Place the center point of the bow compass adjusted to the 4” radius at the intersection of arcs on the plastic record card (initial occlusal plane survey cen­ter). Sweep the needle point over the occlusal surfaces of the lower posterior teeth to see how the arc conforms to the existing occlusal plane. Shift this occlusal plane survey center (O.P.S.C.) on the long arc on plastic record card (A.S.P. line) until the most acceptable line and plane of occlusion is found. To raise the line and plane of occlusion at the distal end, move the point anterior to the arc intersection.
To lower the line and plane of occlusion, move the point posterior of the intersec­tion. By trial and retrial, the ideal survey
center forming the most acceptable line and plane of occlusion will be located.
Upon thorough and considered study, locate the best possible line and plane of occlusion for the lower posterior teeth to harmonize with all other factors. The center point of the bow compass is now pierced into this ideal O.P.S.C. on the plastic record card and circled with pen­cil or ink for subsequent relocation. It may be advantageous to mark “R” (right) in the upper corner of the plastic index card for identification (fig. 103).
fig. 100
fig.101
fig.102
Page 56
54
A
plastic record card is now secured to the left side of the flag and marked “L.” Repeat the procedure commencing with figure 99 for the left survey.
The scribing knife, as furnished, is for placement into the bow compass for scribing or cutting plaster, compound, or wax during the occlusal plane correc­tion. The edge of the scribing knife may be sharpened to individual requirements as the edge supplied may not meet your preference.
fig.103
Page 57
55
Articulators and other occlusal instru­mentation exists to facilitate both the diagnostic and treatment procedures for the fabrication of restorations from the single tooth to the complete dentition. The question often arises regarding the selection of the instrument of choice for treating a particular prosthondontic problem. To answer this question it is of value to first briefly review the progress that has taken place with the develop­ment of instruments for occlusal treat­ment and its impact on prosthodontic techniques. This will also shed light on why there are so many different instru­ments being used.
The earliest articulators, which were of the hinge type, were developed primari­ly for full denture construction; i.e., to fill the need for a mechanical device to relate casts in an anatomically correct position for the arrangement of artificial teeth.
The next breakthrough resulted from the desire to duplicate nature’s scheme and provide patients with improved appear­ance and speech. This led to improved denture teeth of anatomical form. Also, understanding the anatomy and func­tions of the temporomandibular joints and their relation to the occlusal scheme brought about the development of artic­ulators of anatomical dimensions with some adjustment capability to simulate the more pronounced condylar move­ments of a given patient. These articula­tors were adjusted primarily to interoc­clusal records.
Since the articulators were still not com­pletely adequate, the mouth was found to be the best articulator to fabricate wax patterns and casting that would be attached to the natural teeth and be compatible with the excursive move-
ments of the jaw. These early articula­tors and occlusal schemes, that were used to construct dentures resting on mobile tissues, had to be supplemented with more “in the mouth” procedures to produce satisfactory results for fixed restorations.
A great advance came in the late 1940s and early fifties with the popularization of the “indirect technique” i.e., the use of impressions, casts and dies to dupli­cate the dentition so that most of the work could be done in the laboratory at a bench rather than “in the mouth.” The introduction of improved dental materi­als to conveniently develop accurate impressions and full arch working casts coupled with the advent of high speed cutting tools in dentistry popularized restorative procedures for the general public which heralded in a new era in the development of instruments for occlusal diagnosis and treatment.
Dentists sought more practical means to accuratelly record and simulate the movements of the jaw in order to deliver better restorations in less time through both improved laboratory work and operating efficiency. The goal was to produce better dentistry faster. It was this effort and series of events that resulted in the development of the Dénar
®
Pantograph and Fully Adjustable Articulator and subsequently the Dénar®Mark II System and the Dénar
®
Two Instrument System.
Questions often asked are: “Should a pantograph and fully adjustable articula­tor be used to fabricate a single gold crown?” and “How extensive must the restoration be before the use of a panto­graph is indicated?” The simplest response to these questions is as fol­lows. If after tooth preparation, there
APPENDIX E
SELECTING INSTRUMENTS FOR OCCLUSAL TREATMENT
Page 58
56
a
re sufficient unprepared teeth to pro­vide positive stops and guidelines to which to construct the restoration, the restoration can be constructed in har­mony with the existing occlusion and the use of a pantograph is not neces­sarily indicated. On the other hand, if after tooth preparation, there are insuffi­cient guidelines in the remaining occlu­sion to which to construct the restora­tion, the restoration should be con­structed in harmony with the patient’s temporomandibular joint characteris­tics. These characteristics can be diag­nosed most efficiently with a panto­graph and fully adjustable articulator although other treatment methods are available.
In order to better understand the criteria for selecting the instrument of choice for occlusal treatment a discussion of the examples on the chart below is helpful. In this discussion reference is made pri­marily to the pantograph and fully adjustable articulator; however the Mark II System utilizing checkbites and sup­plemented with occlusal equilibration and the functional generated path tech-
n
ique, when indicated, can be used to
produce excellent results.
To keep the discussion pertinent, it is assumed that none of the patients in the following examples have temporo­mandibular joint disease.
PATIENT A is a finished orthodontic patient who has a full complement of teeth in good alignment. The patient has a slight occlusal prematurity which has triggered facial pain. The patient has a pathologic occlusion and occlusal treat­ment is indicated. The treatment is pri­marily occlusal equilibration. An articu­lator can be used in diagnostic proce­dures although it is not always needed in treatment.
PATIENT Z on the other hand exhibits many missing teeth. The remaining teeth are badly broken down. There is exten­sive drifting of teeth and the occlusion is totally disorganized. This patient requires a complete mouth reconstruc­tion. This restoration could be fabricated most efficiently by employing a panto­graph and fully adjustable articulator.
Page 59
57
A
t what point between the two extremes represented by Patient’s A and Z is the use of a pantograph and fully adjustable articulator indicated? The answer to this question is presented in the following discussion of Patient’s B through E who exhibit increasingly complex occlusal problems.
PATIENT B has a missing right second bicuspid and a minor centric prematuri­ty. There are no symptoms which can be related to the occlusal condition. The patient has a physiologic occlusion. The recommended treatment is a three unit bridge. To perform this laboratory pro­cedure the cast can be accurately relat­ed at the correct vertical dimension in a nonrigid cast relating device such as a Johnson Olgesby Articulator (which has springs for condylar posts so that the casts can be gnashed together) and the bridge fabricated in harmony within the anatomical guides provided by the pre­vailing physiologic occlusion. This pro­cedure could be used successfully if the restoration were one, two or three crowns on one, two or three of the teeth involved in the bridge. Alternately a method preferred by many operators is the use of the functional generated path technique (FGP) to fabricate this pros­thesis.
PATIENT C has the same mouth condi­tions and occlusal scheme as Patient B with the exception that the occlusal scheme has triggered bruxism and the sequelae of bruxism (occlusal disease). This patient first requires occlusal treat­ment and in addition, a three unit bridge. The recommended procedure is to first equilibrate the natural occlusion to estab-
l
ish a physiologic occlusion and subse­quently fabricate the three unit bridge (or inlays) as outlined for Patient B.
PATIENT D has the right second bicus­pid missing and in addition the right second molar has indications for full coverage. Slight occlusal discrepancies exist. Whether this patient has related symptoms or not, the procedure many dentists would recommend when this many teeth are involved in the prosthe­sis is to fabricate the restoration in cen­tric relation. Since after tooth prepara­tions there are no more anatomical con­trols in the occlusion on the right side of the mouth to use as anatomical guides to which to fabricate the restoration, then the next most posterior anatomical control, the temporomandibular joint, is used as a control for the fabrication of the prosthesis (see figures 26C and 27 and related discussion). This could be accomplished most efficiently by utiliz­ing a pantograph and fully adjustable articulator.
PATIENT E. If Patient B had in addition to the missing right second bicuspid the right second molar missing, the recom­mended treatment would be the same as for Patient D.
Restorations more involved than those previously listed could all be operated more efficiently by employing a panto­graph and fully adjustable articulator or alternately by means of the Mark II System supplemented with the func­tional generated path technique and/or occlusal equilibration as described above.
Page 60
58
APPENDIX F
Dénar®Semiadjustable Articulator
Mark II Articulator (includes #300197 pin and D41AB platform)
Mark II-P2T2 Articulator (includes #110093 pin and D41 platform)
Mark II-P2T3 Articulator (includes #110093 pin and D46 adjustable incisal table)
P1 (#300197)-round incisal pin functions with the D41AB incisal Platform which has a detent in its superior surface.
P4 (#300270) round short Incisal Pin functions with the D49 Incisal Platform which has a flat surface
P2 (#11093) Adjustable Incisal Pin functions with the D46 Adjustable Incisal Table and the D41 custom Incisal Platform
Earbow/Facebow D31AB Bite Fork - Standard
Dentulous
Bite Fork - Edentulous
Anterior Reference Pointer
Page 61
59
Maxillary Cast Support Remount Record Jig
Mounting Plates Magnetic Mounting System
Dawson Fossa-Guide Pin Pantograph Mounting Fixture
Page 62
60
Your Whip Mix®Articulator is a precision instrument and requires care and main­tenance. Periodic cleaning and lubricat­ing as described below will assure pro­longed life and dependable service from the instrument. Failure to follow these instructions will void your warranty.
CLEANING
Use a mild soap and water solution with the aid of a brush to dissolve accumula­tions of wax and to wash away car­borundum grit. Then air dry and lubri­cate. DO NOT use strong detergents, alkalines, gasoline or naptha as cleaning agents.
LUBRICATION
Lubricate the working and bearing com­ponents with thin film of sewing machine or high speed handpiece type oil. Wipe off excess oil to prevent accu­mulations of dust or grit.
A thin coating of petroleum jelly must be applied to all articulator surfaces that will be contacted by the gypsum mount­ing material
STORAGE
Store the articulator in a clean, dry atmosphere free of plaster and car­borundum dust; away from acids, alka­lies or corrosive medicaments. Wait a
full day after mounting casts before storing the articulator in a carrying case or corrugated carton. Moisture
dissipation from the stone in an enclosed area causes alkalinity of the stone mixture which can damage the articulator surface.
WARRANTY
Whip Mix Corporation warrants the articulator system to be free from defects in material and/or workmanship for a period of one year. In the event of a defect, please notify the factory in writing of the defect prior to returning the instrument. Whip Mix will, at its option, either repair, replace or issue credit for such defects.
Because Whip Mix Corporation is con­tinually advancing the design of its products and manufacturing methods, it reserves the right to improve, modify or discontinue products at any time, or to change specifications or prices without notice and without incurring obligations.
APPENDIX G
ARTICULATOR
CARE AND MAINTENANCE
Page 63
Page 64
Page 65
Page 66
Page 67
Page 68
FN 80471-F AE R1008
Whip Mix Corporation - West
1730 East Prospect Rd., Suite 101
Fort Collins, CO 80525
Toll-Free: 1-800-201-7286
Fax: 1-970-472-1793
www.whipmix.com
Denar®and logo are registered
trademarks of Whip Mix Corporation.
©2008 Whip Mix Corporation
Dénar
®
The Dénar
®
Mark II System TECHNIQUE MANUAL
Loading...