Disabling auto power off
The auto power off function is disabled during switching on of the instrument, by pressing simultaneously, for at least 1 second, the On/Off and
°C/°F buttons.
When the auto power off function is disabled the battery symbol flashes.
Instrument calibration code
The instrument can operate with 3 calibration or work codes (calibration
codes). These three codes are:
• C01 calibration code set in the factory, cannot be changed
• C02 calibration code for the instrument, may be changed
• C03 calibration code for the instrument plus probe, may be
changed
When the instrument is switched on, for 2 seconds it shows the previously set work code on the display with the messages C01, C02 or C03.
Changing the work code (instrument calibration code) to be used
1. Switch on the instrument, holding down the On/Off and CAL buttons
simultaneously for at least 1 second.
During this phase the auto power off function is automatically disabled (the battery symbol flashes).
After performing test display, the instrument shows the CAL message and, after a few moments, C01, C02 or C03 (depending on the
work code used previously) with the message C flashing.
2. Using the sand tbuttons, select the work code that you intend to
use, C01, C02 or C03.
3. Confirm the work code by pressing the CAL button. The instrument
switches off, after having displayed the CAL message.
If you have selected a work code different from those available, the
instrument switches off but the previously set work code is not
changed.
4. Switch the instrument on again and check that the desired work code
is used.
Displaying measurements in pH or in mV
By pressing the pH/mV button it is possible to display the measurement
taken at the instrument input in pH or mV units.
The unit in which the value is measured is indicated on the display.
When switched on, the instrument is set for pH measurement.