DeltaOHM HD32.1 Operating Manual

Operating manual
HD32.1
www.deltaohm.com
English
Keep for future reference.
Companies / Brands of GHM
HD32.1 - 2 - V2.1
TABLE OF CONTENTS
1. GENERAL CHARACTERISTICS...................................................................................................................... 4
2. WORKING PRINCIPLE..................................................................................................................................... 8
2.1 THE OPERATING PROGRAM A: MICROCLIMATE ANALYSIS ......................................................................... 8
2.1.1 WBGT Index............................................................................................................................................ 8
2.1.2 Turbulence Intensity (Tu index)................................................................................................................ 9
2.1.3 WCI Index...............................................................................................................................................10
2.1.4 Average Radiation Temperature tr............................................................................................................11
2.2 THE OPERATING PROGRAM B: DISCOMFORT ANALYSIS .............................................................................13
2.2.1 Unsatisfied with the vertical difference of temperature .............................................................................13
2.2.2 Unsatisfied with the floor temperature......................................................................................................14
2.2.3 Unsatisfied with the radiant asymmetry....................................................................................................14
2.3 THE OPERATING PROGRAM C: PHYSICAL QUANTITIES ...............................................................................15
3. THE USER INTERFACE...................................................................................................................................16
3.1 THE DISPLAY ...................................................................................................................................................16
3.1.1 The Operating Program A, Microclimate Analysis, Display......................................................................16
3.1.2 The Operating Program B, Discomfort Analysis, Display.........................................................................17
3.1.3 The Operating Program C, Physical Quantities, Display...........................................................................17
3.2 THE KEYBOARD ...............................................................................................................................................24
4. OPERATION ......................................................................................................................................................26
4.1 THE OPERATING PROGRAM A: MICROCLIMATE ANALYSIS..................................................................................26
4.1.1 DR Index – Draught Risk.........................................................................................................................27
4.1.2 WBGT Index...........................................................................................................................................28
4.1.3 WCI Index...............................................................................................................................................29
4.1.4 Radiation Temperature Tr........................................................................................................................29
4.1.5 The unit of measurement “Unit”...............................................................................................................29
4.1.6 The maximum, minimum and average values of the captured quantities....................................................29
4.1.7 Instrument Setup......................................................................................................................................30
4.1.8 Start of a new logging session..................................................................................................................30
4.2 THE OPERATING PROGRAM B: DISCOMFORT ANALYSIS.....................................................................................31
4.2.1 The unit of measurement “Unit”...............................................................................................................31
4.2.2 The maximum, minimum and average values of the captured quantities....................................................32
4.3 THE OPERATING PROGRAM C: PHYSICAL QUANTITIES.......................................................................................32
4.3.1 Setting the pipeline section ......................................................................................................................33
4.3.2 The maximum, minimum and average values of the captured quantities....................................................34
5. MAIN MENU ......................................................................................................................................................35
5.1 INFO MENU......................................................................................................................................................35
5.2 LOGGING MENU...............................................................................................................................................36
5.2.1 Log Interval.............................................................................................................................................36
5.2.2 Self Shut-off mode...................................................................................................................................37
5.2.3 Start/stop time – Automatic start..............................................................................................................38
5.2.4 Cancel auto start......................................................................................................................................39
5.2.5 Log File Manager....................................................................................................................................40
5.3 SERIAL MENU (SERIAL COMMUNICATION)........................................................................................................42
5.3.1 Baud Rate................................................................................................................................................43
5.3.2 The Print Interval.....................................................................................................................................43
5.4 RESET..............................................................................................................................................................44
5.5 CONTRAST.......................................................................................................................................................44
5.6 FIRMWARE.......................................................................................................................................................45
5.7 TIME/DATE ......................................................................................................................................................45
5.8 CALIBRATE......................................................................................................................................................46
5.9 KEY LOCK........................................................................................................................................................46
5.10 PASSWO RD ....................................................................................................................................................47
HD32.1 - 3 - V2.1
6. PROBES AND MEASUREMENTS....................................................................................................................49
6.1A AND B OPERATING PROGRAM PROBES : ........................................................................................................49
6.1.1 Warnings, care and maintenance of the probes .........................................................................................65
6.2 PROBES FOR THE OPERATING PROGRAM C: PHYSICAL QUANTITIES ....................................................................67
6.2.1 Temperature measurement using the probe Pt100 complete with SICRAM module...................................67
6.2.2 Technical information on temperature probes Pt100 using SICRAM module ............................................68
6.2.3 Measurement of relative humidity using the combined humidity/temperature probe..................................69
6.2.4 Technical information on relative humidity and temperature probes using SICRAM module...........................70
6.2.5 Wind speed measurement ........................................................................................................................72
6.2.6 AP471S… Hot-wire wind speed measurement probes complete with SICRAM module.............................74
6.2.7 Technical information on Hot-wire wind speed measurement and temperature probes using SICRAM
module....................................................................................................................................................76
6.2.8 AP472S… Vane wind speed measurement probes complete with SICRAM module...................................79
6.2.9 Technical information on Vane wind speed measurement probes using SICRAM module..........................82
6.2.10 Light measurement ................................................................................................................................83
6.2.11 Technical characteristics of photometric and radiometric probes complete with SICRAM module to be
connected with the instruments on line.....................................................................................................84
6.2.12 HD320A2 probe for the measurement of CO Carbon monoxide..............................................................90
6.2.13 HD320B2 probe for the measurement of CO2 carbon dioxide concentration.............................................94
7. SERIAL INTERFACE AND USB.......................................................................................................................96
7.1 THE OPERATING PROGRAM A: MICROCLIMATE ANALYSIS..................................................................................96
7.2 THE OPERATING PROGRAM B: DISCOMFORT ANALYSIS.....................................................................................98
7.3 THE OPERATING PROGRAM C: PHYSICAL QUANTITIES.......................................................................................99
7.4 STORING AND TRANSFERRING DATA TO A PC .................................................................................................101
7.4.1 The Logging Function ...........................................................................................................................101
7.4.2 The Erase Function: clearing the memory...............................................................................................101
7.4.3 The Print Function.................................................................................................................................101
8. INSTRUMENT SIGNALS AND FAULTS .......................................................................................................102
9. BATTERY SYMBOL AND BATTERY REPLACEMENT – MAINS POWER SUPPLY..............................103
9.1 WARNING ABO UT BAT TERY USE ......................................................................................................................104
10. INSTRUMENT STORAGE ............................................................................................................................104
11. MEASUREMENT REPORTS PRINTING ....................................................................................................105
12. TECHNICAL CHARACTERISTICS.............................................................................................................141
13. EXPLANATORY TABLES ON THE MICROCLIMATE PROBES USAGE..............................................143
13.1 DIAGRAM OF THE PROBES FOR HD32.1 OPERATING PROGRAM A: MICROCLIMATE ANALYSIS .........................144
13.2 DIAGRAM OF THE PROBES FOR HD32.1 OPERATING PROGRAM B: DISCOMFORT ANALYSIS .............................147
14. ORDERING CODES.......................................................................................................................................148
14.1 A AND B OPERATING PROGRAMS PROBES ......................................................................................................149
14.2 PROBES FOR THE OPERATING PROGRAM C: PHYSICAL QUANTITIES ................................................................150
14.2.1 Temperature probes complete with SICRAM module...........................................................................150
14.2.2 Relative Humidity and Temperature combined probes complete with SICRAM module........................150
14.2.3 Wind Speed and Temperature combined probes complete with SICRAM module.................................151
14.2.4 Photometric/Radiometric probes for Light measurement complete with SICRAM module ....................151
14.2.5 Probes for CO2 carbon dioxide measurement complete with SICRAM module.....................................152
14.2.6 Probes for the measurement of CO carbon monoxide complete with SICRAM module .........................152
HD32.1 - 4 - V2.1
1. GENERAL CHARACTERISTICS
The HD32.1 Thermal Microclimate has been designed for microclimate analysis in the workplace; the instrument is used to detect the necessary parameters to establish if a certain workspace is suitable to perform certain activities.
The instrument can manage three operating programs that can be loaded by the user, according to the measurements program and the quantities being detected.
The instrument has eight inputs for probes with SICRAM module: The probes are fitted with an electronic circuit that communicates with the instrument. The calibration settings are memorized inside.
All SICRAM probes, except the vane probe, can be plugged into any input: They will be automatically detected when you turn the instrument on.
NOTE: The vane probe, complete with SICRAM module, for wind speed measurement can be exclusively connected to input 8.
The instrument comes with a barometric pressure sensor. The barometric pressure is displayed only by the Microclimate Analysis operating program.
The machine can be programmed to perform the logging (capture) of a measurement session. You can specifically set a sampling interval per each session.
Furthermore, the auto-start function can be used to activate the setting of the measurements’ initial date and time, and the automatic start and end of the data logging session.
NOTE: The set capture interval is valid for all the probes connected to the machine.
Other operating programs user selectable/settable common parameters are:
The units of measurement for the displayed temperature quantities: °C, °F, °K.
The system date and time
The display of the maximum, minimum, and average statistic parameters and their
deletion.
The data transfer speed via the serial RS232 port.
The setting and enabling/disabling of the keyboard protection password.
The operating programs are:
prog. A: HD32.1 Microclimate Analysis
prog. B: HD32.1 Discomfort Analysis
prog. C: HD32.1 Physical Quantities
The operating program A: Microclimate Analysis, HD32.1, can simultaneously detect the following quantities:
Globe thermometer temperature
Natural ventilation wet bulb temperature
Environment temperature
Atmospheric pressure
Relative humidity
Wind speed
HD32.1 - 5 - V2.1
The operating program A: Microclimate Analysis displays also:
The local turbulence intensity Tu, for DR (Draught Rating) calculation.
The WBGT index (Wet Bulb Glob Temperature) with or without solar radiation.
The WCI index (Wind Chill Index)
The average radiation temperature tr.
The operating program B: Discomfort Analysis, HD32.1, can simultaneously detect the following quantities:
Air temperature detected at head height (1.7 m for a standing person; 1.1 m for a seated
person).
Air temperature detected at abdomen height (1.1 m for a standing person; 0.6 m for a seated
person).
Air temperature detected at ankle height (0.1 m).
Temperature at floor level.
Temperature of the net radiometer.
Net radiation.
Radiant asymmetry temperature.
The operating program B: Discomfort Analysis is used to calculate the local discomfort indexes due to vertical temperature gradients or radiant asymmetry temperature.
The operating program C: Physical Quantities, HD32.1, can simultaneously detect the following physical quantities:
Temperature
Relative humidity
Illuminance, luminance, PAR, irradiance
Wind speed
carbon monoxide concentration CO.
carbon dioxide concentration CO2
By using the operating program C: Physical Quantities, HD32.1 can simultaneously manage up to six different probes complete with SICRAM module: one or two light probes and a probe for each of the other physical quantities. If two or more probes of the same physical quantity are connected (light probes excluded), the instrument recognizes only one probe. If, for example,
you insert two Pt100 temperature probes complete with SICRAM module to inputs 1 and 2, the probe connected to input 1 is immediately recognized while the probe connected to input 2 is ignored. The probe is detected during turn on, therefore if a probe is changed, it is necessary to turn the machine off and on. During turn on the instrument performs a scan from input 1 to input 8. The probes arrangement in relation to the inputs is arbitrary, except for the vane probes.
The vane probe, complete with SICRAM module, for wind speed measurement can be exclusively connected to input 8.
HD32.1 - 6 - V2.1
Thermal Microclimate
HD32.1
HD32.1 - 7 - V2.1
HD32.1
1. ON/OFF key: Turns the instrument on and off.
2. TIME key: Allows the display of date and time, in the first line for about 8 seconds.
3. SHIFT FNC key: Activates the Shortcut window.
4. Graphic display.
5. Function keys F1, F2, F3: Activate the function in the bottom line of the display.
6. ENTER key: In the menu, confirms the data entered.
7. ESC key: Allows exiting from the menu or, in case of a submenu, exiting from the current level
display.
8. Navigation keys ▲▼◄►: Allows navigation through the menus.
9. PRINT key: Starts and ends the data transfer to the serial/USB communication port.
10. MEM key: Starts and ends the recording of the data.
11. SETUP key: Allows entering and exiting the instrument’s functioning parameter setting menu.
12. SICRAM inputs for the probes.
13. Battery cover.
14. RS232 serial port.
15. USB port.
16. Power supply input.
HD32.1 - 8 - V2.1
2. WORKING PRINCIPLE
2.1 THE OPERATING PROGRAM A: MICROCLIMATE ANALYSIS
By microclimate is meant those environmental parameters that influence the heat exchange between the person and the surrounding spaces, and that determinate the so-called " thermal well-being".
The microenvironment climatic factors, together with the type of work performed, generate a series of biological responses linked to thermal well-being (Comfort) or uneasiness (Discomfort).
The human organism, indeed, tends to maintain a thermal balance so that the body temperature is optimum.
The HD32.1, Thermal Microclimate, through its operating program A: Microclimate Analysis measures the following quantities:
tnw: natural ventilation wet bulb temperature
tg: globe thermometer temperature
ta: environment temperature
pr: at mospheric pressure
RH: relative humidity
va: wind speed
In addition to the direct measurements performed with the probes connected, the instrument can directly calculate and display the following well-being data:
WBGT index
Tu index
WCI index
Average radiation temperature tr
2.1.1 WBGT Index
WBGT (Wet Bulb Globe Temperature) is one of the indexes used to determine the thermal stress of a person in a hot environment. It represents the value, related to the metabolic output linked to a specific work activity, that causes a thermal stress when exceeded. The WBGT index combines the measurements of the natural ventilation wet bulb temperature tnw with the globe thermometer temperature tg and, in some situations, with the air temperature ta. The calculation formula is the following:
Inside and outside buildings without solar radiation:
WBG
enclosed spaces
= 0.7 tnw + 0.3 t
g
Outside buildings with solar radiation:
WBGT
outdoor spaces
= 0.7 tnw + 0.2 tg + 0.1 ta
where:
t
nw
= natural ventilation wet bulb temperature;
t
g
= globe thermometer temperature;
t
a
= air temperature.
The measured data should be confronted with the limit values prescribed by regulations; when exceeded you have to:
HD32.1 - 9 - V2.1
Reduce directly the thermal stress on the workplace being examined;
Proceed to a detailed analysis of the thermal stress.
In the following table are reported the thermal stress index WBGT limit values as provided for by ISO 7243:
METABOLIC RATE, M WBGT LIMIT VALUE
METABOLIC
RATE CLASS
RELATIVE TO A
UNIT AREA OF
THE SKIN
W/m
2
TOTAL (FOR A
SKIN SURFACE
AVERAGE AREA
OF 1.8 m
2
)
W
PERSON ACQUAINTED TO THE
HEAT
°C
PERSON NOT ACQUAINTED TO
THE HEAT
°C
0
(RESTING)
M 65 M 117 33 32
1 65 < M 130 117 < M 234 30 29
2 130 < M 200 234 < M 360 28 26
3 200 < M 260 360 < M 468
DEAD AIR
25
NON DEAD AIR
26
DEAD AIR
22
NON DEAD AIR
23
4 M > 260 M > 468 23 25 18 20
NOTE – THESE VALUES HAVE BEEN DETERMINED BY USING A MAXIMUM REFERENCE RECTAL TEMPERATURE OF 38 °C
FOR THE PEOPLE BEING EXAMINED.
In order to calculate the WBGT index, the following probes should be connected:
The natural ventilation wet bulb temperature probe.
The globe thermometer probe.
The dry bulb temperature probe, if the measurement is performed with solar
radiation.
In order to measure the WBGT index, you should refer to the following regulations:
ISO 7726
ISO 7243
2.1.2 Turbulence Intensity (Tu index)
Turbulence Intensity: Local turbulence intensity percentage, defined as the ratio between the local wind speed standard deviation and the local air average speed (ISO 7726):
100×=
a
v
SD
Tu
where:
v
a
= average local wind speed
SD
= local wind speed standard deviation
()
=
=
n
i
aa
vv
n
SD
i
1
2
1
1
From the turbulence calculation, knowing the average values of the local wind speed and environment temperature, you can get the DR (Draught Rating), according to ISO 7730:
()()(
)
14,337,005,034
62,0
+= TuvvtDR
aaa
HD32.1 - 10 - V2.1
The discomfort from air current is defined as an undesired local cooling of the body due to air mot io n. The
DR indicates the percentage of unsatisfied people due to air current. The DR index is
calculated when the temperature goes from 20 °C to 26 °C and the average wind speed is < 0.5 m/s.
The DR index is calculated using the DeltaLog10 software.
2.1.3 WCI Index
WCI (Wind Chill Index) allows a synthetic evaluation of the effects of cold environments on man.
It shows the cooling index due to the wind. It allows evaluating the discomfort perceived during exposure to low temperatures and wind. The index does not consider the clothing and the work intensity.
The WCI index is calculated by the instrument in presence of air under 10°C.
The WCI index calculation formula is:
WCI = 13.12 + 0.6215 t
a
– 11.37 v
a
0.16
+ 0.4275 ta v
a
0.16
where:
t
a
: air temperature (in °C);
v
a
: Wind speed (in km/h) calculated at 10 m from the ground.
As the instrument measures the wind speed at 1.5 m from the ground, the formula is corrected as follows:
WCI = 13.12 + 0.6215 t
a
– 11.37(1.5 v
1.5
)
0.16
+ 0.4275 ta (1.5 v
1.5
)
0.16
where v
1.5
is the wind speed measured by the instrument at 1.5 m from the ground.
The following tables report some WCI values and the relevant frostbite risks (source: NOAA – National Weather Service).
Air temperature °C
10 5 0 -5 -10 -15 -20 -25 -30 -35 -40 -45 -50
10 8.6 2.7 -3.3 -9.3 -15.3 -21.1 -27.2 -33.2 -39.2 -45.1 -51.1 -57.1 -63.0
15 7.9 1.7 -4.4 -10.6 -16.7 -22.9 -29.1 -35.2 -41.4 -47.6 -51.1 -59.9 -66.1
20 7.4 1.1 -5.2 -11.6 -17.9 -24.2 -30.5 -36.8 -43.1 -49.4 -55.7 -62.0 -68.3
25 6.9 0.5 -5.9 -12.3 -18.8 -25.2 -31.6 -38.0 -44.5 -50.9 -57.3 -63.7 -70.2
30 6.6 0.1 -6.5 -13.0 -19.5 -26.0 -32.6 -39.1 -45.6 -52.1 -58.7 -65.2 -71.7
35 6.3 -0.4 -7.0 -13. 6 -20.2 -26.8 -33.4 -40.0 -46.6 -53.2 -59.8 -66.4 -73.1
40 6.0 -0.7 -7.4 -14. 1 -20.8 -27.4 -34.1 -40.8 -47.5 -54.2 -60.3 -67.6 -74.2
45 5.7 -1.0 -7.8 -14. 5 -21.3 -28.0 -34.8 -41.5 -48.3 -55.1 -61.8 -68.6 -75.3
50 5.5 -1.3 -8.1 -15. 0 -21.8 -28.6 -35.4 -42.2 -49.0 -55.8 -62.7 -69.5 -76.3
55 5.3 -1.6 -8.5 -15. 3 -22.2 -29.1 -36.0 -42.8 -49.7 -56.6 -63.4 -70.3 -77.2
Wind speed v
a
Km/h
60 5.1 -1.8 -8.8 -15. 7 -22.6 -29.5 -36.5 -43.4 -50.3 -57.2 -64.2 -71.1 -78.0
The values that can cause frostbite within 30 minutes are reported in bold.
HD32.1 - 11 - V2.1
Wind Chill (°C) Frostbite risk
> -28 Low
-28 to -39 Medium: The exposed body parts can freeze within 10 to 30 minutes
-40 to -44 High: The exposed body parts can freeze within 5 to 10 minutes (*)
Alarm level
-44 to -47
High: The exposed body parts can freeze within 2 to 5 minutes (*)
-48 and colder High: The exposed body parts can freeze in less than 2 minutes (*)
(*): With va higher than 50 km/h the frostbite process can be faster.
In order to calculate the WCI index, the following probes should be connected:
The dry bulb temperature probe for air temperature measurement t
a
.
The hot-wire probe for wind speed measurement.
In order to measure the WCI index, refer to the following:
ISO 7726
NOAA Specifications, National Weather Service.
2.1.4 Average Radiation Temperature tr
The average radiation temperature is defined as the temperature of a thermally uniform simulated environment that would exchange with a man the same thermal radiation power exchanged in the real environment.
In order to evaluate the average radiation temperature you have to measure: The globe thermometer temperature, the air temperature and the wind speed measured near the globe thermometer. The average radiation temperature calculation formula is the following:
In case of natural convection:
() ()
273
1025.0
273
4/1
4/1
8
4
×
⎟ ⎠
⎜ ⎝
×
++=
ag
ag
g
gr
tt
D
tt
tt
ε
In case of forced convection:
() ()
273
101.1
273
4/1
4.0
6.08
4
⎥ ⎦
⎢ ⎣
×
××
++=
ag
g
a
gr
tt
D
v
tt
ε
where:
D = globe thermometer diameter
ε
g
= 0.95 assumed emissivity of the globe thermometer
t
g
= globe thermometer temperature
t
a
= air temperature.
v
a
= wind speed
The average radiation temperature does not correspond to air temperature: If within a room very high temperature surfaces are present (for example, a fireplace), these hot areas remarkably influence the average radiation temperature.
The average radiation temperature is detected with the globe thermometer. A temperature probe formed by a 150 mm copper ball, painted black matt, with emissivity equal to ε
g
= 0.95 (according
HD32.1 - 12 - V2.1
to ISO 7726), and a Pt100 sensor inside. The globe thermometer temperature could be remarkably higher than the air temperature. For example in a mountain cottage, in which air is at 0°C but the presence of a fireplace produces an average radiation temperature of 40°C, ensuring a comfortable condition. In normal conditions, maintaining a certain difference between average radiation temperature and air temperature (where
T
MR
is remarkably higher than TA) is preferable in order to get a better environment quality. In houses, where fireplaces and ranges are absent, generally the average radiation temperature is equal to the air temperature, or even lower. These conditions (mainly in buildings with large window surfaces) are not particularly healthy as the warm humid air facilitates the development of pathogenic organisms. From this point of view, using lamps or radiating panels is more healthy. It is much more hygienic to use a higher average radiation temperature than the air temperature, in order to guarantee comfort conditions. The regulations use air temperature and not average radiation temperature to assess heating system, but this is wrong.
In order to calculate the average radiation temperature you have to connect the following probes:
Globe thermometer probe
Air temperature measurement probe
Hot-wire probe for wind speed measurement
In order to measure the average radiation temperature, you should refer to the following regulations:
ISO 7726
HD32.1 - 13 - V2.1
2.2 THE OPERATING PROGRAM B: DISCOMFORT ANALYSIS
The HD32.1, Thermal Microclimate, through its operating program B: Discomfort Analysis, measures the following quantities.
t
h
head temperature: air temperature detected at head height
tb: body temperature: air temperature detected at abdomen height
tk: ankle temperature: air temperature detected at ankle height
tf: floor temperature: air temperature at floor level
P: net radiation: net radiation, measured in Wm
-2
.
By knowing the temperatures at head t
h
, abdomen tb, ankles tk and floor t
f
height, you can
determine, according to
ISO 7730, the following local thermal discomfort indexes:
Unsatisfied with the vertical difference of temperature;
Unsatisfied with the floor temperature;
Unsatisfied with the radiant asymmetry.
2.2.1 Unsatisfied with the vertical difference of temperature
By detecting the temperatures at various heights you can check the presence of a air temperature vertical gradient. This gradient could cause a local discomfort feeling. In the following figure you can see the percentage value of the
unsatisfied with the vertical difference of temperature PD
v
, according to the temperature gradient between head (1.10 m) and ankles (0.10 m) for a seated person.
This index is calculated using the DeltaLog10 software.
In order to calculate the PDv index (unsatisfied with the vertical difference of temperature) you have to connect the following probes:
Probe for the air temperature detected at head height
Probe for the air temperature detected at ankle height
The PDv index is calculated by the DeltaLog10 software.
In order to calculate the PDv index (unsatisfied with the vertical difference of temperature) you should refer to the following regulations:
ISO 7730
HD32.1 - 14 - V2.1
2.2.2 Unsatisfied with the floor temperature
By measuring the floor temperature you can calculate the
percentage index of the unsatisfied with
the floor temperature. The following diagram shows the PD
f
index trend according to the floor
temperature.
In order to calculate the PDf index (unsatisfied with the floor temperature) you have to connect the following probe:
Floor temperature probe
The PDf index is calculated by the DeltaLog10 software.
In order to calculate the PDf index (unsatisfied with the floor temperature) you should refer to the following regulations:
ISO 7730
2.2.3 Unsatisfied with the radiant asymmetry
The radiant temperature asymmetry Δt
pr
is the difference between the radiant temperatures measured with the net radiometer. The index of the unsatisfied with the radiant asymmetry is calculated according to a vertical (ceiling-floor) or horizontal (wall-wall) asymmetry. Form the radiant temperature Δt
pr
measured by the instrument, you can determine the percentage of the
unsatisfied with the radiant asymmetry
using the DeltaLog10 software. This is reported in the following figure.
1 Hot ceiling 2 Cold wall 3 Cold ceiling 4 Hot wall
HD32.1 - 15 - V2.1
In order to calculate the percentage of the unsatisfied with the floor temperature you have to connect the following probe:
Net radiometer probe for radiant temperature measurement
The percentage of the unsatisfied with the radiant asymmetry is calculated using the DeltaLog10 software.
In order to calculate the percentage of the unsatisfied with the radiant asymmetry, you should refer to the following regulations:
ISO 7730
2.3 THE OPERATING PROGRAM C: PHYSICAL QUANTITIES
The HD32.1, Thermal Microclimate, through its operating program C: Physical Quantities, measures the following quantities:
Temperature.
Relative humidity and resulting measurements.
Illuminance, luminance, PAR, irradiance.
Wind speed and resulting measurements.
carbon monoxide concentration CO.
carbon dioxide concentration CO
2
HD32.1 - 16 - V2.1
3. THE USER INTERFACE
The user interface consist of an LCD display and of the power, function, and setting keys. Turn the instrument on and off with the
ON/OFF key. When you turn the instrument on, the logo
and model will be displayed for a few seconds, and then the main display.
3.1 THE DISPLAY
The display changes according to the loaded operating program.
3.1.1 The Operating Program A, Microclimate Analysis, Display
The Thermal Microclimate HD32.1 main display, is divided in three areas:
The first area displays the battery’s charge status and the current time on the first line and the
measured quantities arranged on two columns:
Tnw: natural ventilation wet bulb temperature Tg: globe thermometer temperature Ta: environment temperature Pr: atmospheric pressure RH: relative humidity Va: wind speed
The second area displays the resulting measurements, that is, the indoor or outdoor WBGT indexes (WI or WO), the WCI and the average radiant temperature Tr ( see the previous chapter for further details).
The third area displays the
F1, F2 and F3 options. Please see the SHIFT FNC key in the following
paragraph.
HD32.1 - 17 - V2.1
3.1.2 The Operating Program B, Discomfort Analysis, Display
The Thermal Microclimate HD32.1 main display, is divided in three areas:
The first area displays the battery’s charge status and the current time on the first line and the
measured quantities arranged on two columns:
Th: head temperature: air temperature detected at head height Tb: body temperature: air temperature detected at abdomen height Tn: net temperature: temperature of the net radiometer Tk: ankle temperature: air temperature detected at ankle height Tf: floor temperature: air temperature at floor level P: net radiation: net radiation, measured in Wm
-2
The second area displays the radiant asymmetry temperature DT. By knowing this parameter you can obtain the
percentage of the unsatisfied with the radiant asymmetry using DeltaLog10,
according to ISO 7730.
The third area displays the
F1, F2 and F3 options. Please see the SHIFT FNC key in the following
paragraph.
3.1.3 The Operating Program C, Physical Quantities, Display
The
Thermal Microclimate, HD32.1, display of the Operating Program C: Physical Quantities,
shows the quantities of the probes connected to the instrument inputs. The display is divided into three areas: The first area (first line) displays the
battery’s charge status, the current date and time and the
type of probe:
Displayed message Displayed SICRAM probe TEMPE Pt100 probe RH-TEMPE RH-temperature combined probe AIR-TEMP Vane probe or hot-wire probe and temperature probe LUX Light probe CO CO probe CO2 CO2 probe
The second area of the display shows the measurements detected by the probes. Each display is associated to the relevant probe. In order to display the measurements, press F3 when the message
next appears on the last line.
The third area displays the
F1, F2 and F3 options. Please see the SHIFT FNC key in the following
paragraph.
HD32.1 - 18 - V2.1
If a Pt100 temperature probe with SICRAM module is present, the following is displayed:
The first display line shows “
TEMPE.” indicating that the temperature of a Pt100 probe complete
with SICRAM module is displayed.
By repeatedly pressing
F1 with unit on the last line, you can change the unit of measurement: The
available units are °C and °F; by pressing
F1 again you can return to °C.
By pressing
F3 with next on the last line, you go to the combined humidity/temperature probe
complete with SICRAM module:
The first display line shows “
RH-TEMPE” indicating that the measurements of a combined
humidity/temperature probe complete with SICRAM module are displayed.
The second line displays the relative humidity. The third line displays the temperature.
In order to change the unit of measurement, press
F2 with sel displayed on the last line.
HD32.1 - 19 - V2.1
By pressing F1 when the second line is selected, you can change the relative humidity unit of measurement:
The available units of measurement are:
RH: % of relative humidity (%RH - Relative Humidity)
SH: Grams of vapor in a kilogram of dry air (g/Kg - Specific Humidity, calculated)
AH: Grams of vapor in a cubic meter of dry air (g/m3 - Absolute Humidity, calculated)
Pa: Partial vapor pressure (hPa, calculated)
H: Enthalpy (J/g, calculated)
Td: Dew point (°C or °F, calculated)
Tw: Wet bulb temperature (°C or °F)
By pressing
F1 when the third line is selected, you can change the temperature unit of
measurement: The available units are °C and °F; by pressing
F1 again you can return to °C.
HD32.1 - 20 - V2.1
By pressing F3 with next on the last line, you go to the combined speed/temperature probe complete with SICRAM module: The vane probes for wind speed measurement can be
exclusively connected to input 8. The display is as follows:
The first display line shows “
AIR-TEMP” indicating that the measurements of a combined
speed/temperature probe complete with SICRAM module are displayed.
The second line displays the wind speed. The third line displays the flow rate measurement. In order to get this measurement, you have to set the duct section (see next paragraph). The fourth line displays the temperature, if required.
In order to change the unit of measurement, press
F2 with sel displayed on the last line.
By pressing
F1 when the second line is selected, you can change the wind speed unit of
measurement:
HD32.1 - 21 - V2.1
The units of measurement for the wind speed are:
m/s
km/h
ft/min
mph (mile/hour)
knot
By pressing
F1 when the third line is selected, you can change the flow rate unit of measurement:
The units of measurement for the flow rate are:
l/s (liter/s)
m
3
/s
m
3
/min
m
3
/h
ft
3
/s
ft
3
/min
By pressing
F1 when the fourth line is selected, you can change the temperature unit of
measurement: The available units are °C and °F; by pressing
F1 again you can return to °C.
HD32.1 - 22 - V2.1
By pressing F3 with next on the last line, you go to the light probes complete with SICRAM module:
The first display line shows “
LUX” indicating that the measurement of a light probes complete
with SICRAM module are displayed.
To change the unit of measurement, select one of the rows with F2 key and then press repeatedly
F1:
The available units of measurement depend on the type of probe:
Type of measurement Unit of Measurement
Illuminance (Phot) lux - fcd
Irradiance (RAD - UVA - UVB - UVC)
W/m
2
- μW/cm2
PAR
μmol/( m
2
s)
Luminance (LUM 2) cd/m2
Pushing the
function key F3 when there is the writing next on the last line you go on visualizing
the measurements detected by
CO
2
probe complete with SICRAM module. The visualization is the
following one:
HD32.1 - 23 - V2.1
The first line of the display indicates the acronym “CO2”, indicating that the measurement of
carbon dioxide concentration detected by the CO2 probe
with SICRAM module is visualized on
the display.
The measurement unit in
ppm (parts per million) can’t vary.
When you start the instrument up, it realizes a warm-up (the heating) of at least 30 seconds of the probe before visualizing CO
2
measurement. The writing “warm-up” stays on for other 30 seconds together with the measurement to indicate that the indicated values could not already be within the declared accuracy limits. When the writing “warm-up” switches off, the instrument is working.
The probe is calibrated by the company and usually doesn’t request any other intervention by the user.
However, there is the possibility to calibrate again: see the chapter that deals with the Probe "
Probe
HD320B2 for the measurement of CO2 carbon dioxide concentration".
Pushing the
function key F3 when there is the writing next on the last line you go on visualizing
the measurements detected by
CO
2
probe complete with SICRAM module. The visualization is the
following one:
The first line of the display indicates the symbol
“CO”, indicating that the measurement of carbon
monoxide concentration detected by the CO probe
with SICRAM module is visualized on the
display.
The measurement unit in
ppm (parts per million) can’t vary.
When you start the instrument up, it realizes a warm-up (the heating) of at least 30 seconds of the probe before visualizing CO measurement. The writing “warm-up” stays on for other 30 seconds together with the measurement to indicate that the indicated values could not already be within the declared accuracy limits. When the writing “warm-up” switches off, the instrument is working.
The probe is calibrated by the company and usually doesn’t request any other intervention by the user.
However, there is the possibility to calibrate again: see the chapter that deals with the Probe "
Probe
HD320A2 for the measurement of CO carbon monoxide concentration".
Pushing the
function key F3 at the end of the cycle, when on the last line of the display there is the
indication
next, you go back to the visualization of the measurements detected by the Pt100 probe.
If one of the probe described is not connected to the instrument when switching on, the corresponding visualization does not appear.
HD32.1 - 24 - V2.1
3.2 THE KEYBOARD
The keys on the instrument perform the following functions:
ON-OFF key
Turns the instrument on and off. When turning on the instrument using this key, the first screen will be displayed. After few seconds the measured quantities will be displayed.
NOTE: If no probes were connected on turning on, only the barometric pressure will be displayed.
The other quantities will be indicated by dashes, instead of the value.
TIME key
It allows the display of
year/month/day and hour/minutes/seconds, in the first line for about 8
seconds. Normally the display shows, on the left, the
icon for the battery’s charge status, on
the right, hour/minutes/seconds. The battery symbol becomes
[~] when the external power supply
is connected.
SHIFT FUNCTION key
Activates the Shortcut window. The figure shows the Shortcut menu for the
Microclimate Analysis
operating program.
Function keys F1, F2, F3
These are “function keys”: They activate the function in the last line of the display (indicated by the arrow in the figure); the function, enabled by
SHIFT FNC, is selected and displayed in “reverse”
(e.g. in the figure the
Microclimate Analysis operating program “WBGT indoor” function is
enabled).
HD32.1 - 25 - V2.1
SETUP key
Allows entering and exiting the instrument’s functioning parameter setting menu.
ENTER key
In the menu, confirms the entered data.
ESC key
Allows exiting from the menu or, in case of a submenu, exiting from the current level display.
MEM key
Allows starting and ending a “logging” session; the data sending interval must be set in the menu.
PRINT key
Allows direct printing of the data via serial port; the data sending interval must be set in the menu.
Arrow keys
Allow navigation through the menus.
HD32.1 - 26 - V2.1
4. OPERATION
Before turning on the instrument, connect the SICRAM probes to the inputs: 8-pole male DIN45326 connectors, located in the lower part of the instrument (see figure on page 2), according
to the measurement being performed.
NOTE: Connect the probes when the instrument is off. If a probe is connected and the instrument is on, it will be ignored. In this case, it is necessary to turn it off and on.
If a probe is connected when the instrument is on, you will get an acoustic signal (one beep per second) and an indication on the display relevant to the physical quantity being disconnected. The “LOST” message will be displayed.
If you connect multiple probes of the same type, only the first recognized probe is accepted (the first two if light probes): The probes scanning starts fro input 1 to input 8.
The barometric pressure sensor is internal: Upon turning on the instrument, should no probes be connected, only the atmospheric value is displayed.
During turning on, the following message is displayed for about 10 seconds:
In addition to the
Delta Ohm logo the tab name and the operating program code are displayed:
prog. A: HD32.1 Microclimate Analysis
prog. B: HD32.1 Discomfort Analysis
prog. C: HD32.1 Physical Quantities
4.1 THE OPERATING PROGRAM A: MICROCLIMATE ANALYSIS
Connect the probes. Turn on the instrument: After 10 seconds, the measurement display mode will appear:
The battery charge symbol and current time are up on the left (for further details, see the par. 10).
HD32.1 - 27 - V2.1
The following quantities are reported:
Tnw: Humid temperature, measured by a natural ventilation wet bulb probe Tg: Globe thermometer temperature, measured by a globe thermometer probe Ta: Environment temperature, measured by a Pt100 probe Pr: Barometric pressure, measured by an internal sensor RH: Relative humidity, measured by a combined humidity/temperature probe Va: Wind speed, measured by a hot-wire probe
A resulting quantity is displayed in the central part of the display: In this example, the WI index, that is, the WBGT index measured indoor or outdoor without solar radiation.
In order to select the displayed index, press
SHIFT FNC: A drop-down menu is shown with the
following information:
draught: DR index: draught risk
wbgt: WBGT index: wet bulb globe temperature
wci: WCI index: wind chill index
trad: average radiation temperature Tr
unit: temperature measurement unit
data: maximum, medium, average values
4.1.1 DR Index – Draught Risk
In order to calculate the DR index you need to know the
turbulence intensity Tu obtained from the
wind speed. For the turbulence intensity Tu calculation, the instrument starts an automatic procedure to capture the wind speed within a preset interval; at the end, the instrument displays the turbulence intensity percentage value. The DeltaLog10 software is then used to obtain the DR index. To start the Tu index calculation, proceed as follows after opening the drop-down menu with
SHIFT FNC:
1.
Use the arrow keys ▲▼ to select “draught”;
2.
Press ENTER to confirm: the Tu message is displayed in the central line of the display;
3.
Press ESC to exit the drop-down menu without making any change.
The following screen will appear:
HD32.1 - 28 - V2.1
Press
F1 to start the capture procedure:
On the first line, the blinking
TU symbol and the start message replaced by running, indicate
that the procedure was started.
After few seconds the
TU symbol will disappear, the running message will be replaced by start
and the turbulence intensity value will be displayed.
4.1.2 WBGT Index
To display the WBGT index, proceed as follows after opening the drop-down menu with
SHIFT
FNC
:
4.
Use the arrow keys ▲▼ to select WBGT;
5.
Press ENTER to confirm: The selected quantity is displayed in the central line of the display;
6.
Press ESC to exit the drop-down menu without making any change.
Now you can display the Indoor (WI) or Outdoor (WO) values, by selecting them using the
F1 or
F3 keys (see figure).
HD32.1 - 29 - V2.1
4.1.3 WCI Index
To display the WCI index, proceed as follows after opening the drop-down menu with
SHIFT
FNC:
1.
Use the arrow keys ▲▼ to select WCI;
2.
Press ENTER to confirm: the selected quantity is displayed in the central line of the display;
3.
Press ESC to exit the drop-down menu without making any change.
4.1.4 Radiation Temperature Tr.
To display the radiation temperature Tr, proceed as follows after opening the drop-down menu with
SHIFT FNC:
1.
Use the arrow keys ▲▼ to select TRAD;
2.
Press ENTER to confirm: the selected quantity is displayed in the central line of the display;
3.
Press ESC to exit the drop-down menu without making any change.
4.1.5 The unit of measurement “Unit”
By using the “Functions” menu, you can display the temperature in °C (Celsius), °F (Fahrenheit) or °K (Kelvin) degrees, as follows:
1.
Use SHIFT FNC to open the drop-down menu;
2.
Use the arrow keys ▲▼ to select unit;
3.
Press ENTER to confirm: the selected quantity is displayed in the central line of the display;
4.
The three different temperature units of measurement are shown in the bottom line of the
display, using
F1, F2 or F3: The unit is selected and displayed near the value shown in the
central line;
5.
Press ESC to exit the drop-down menu without making any change.
4.1.6 The maximum, minimum and average values of the captured quantities
In order to display the maximum, minimum and average values of the measured quantities, proceed as follows:
1.
Use SHIFT FNC to open the drop-down menu;
2.
Use the arrow keys ▲▼ to select data;
3.
Press ENTER to confirm: the selected quantity is displayed in the central line of the display;
4.
The three quantities max (maximum), min (minimum) and avg (average) are shown in the
bottom line of the display, using
F1 or F2.
HD32.1 - 30 - V2.1
NOTE: Once selected, for example max, all displayed quantities represent the maximum value. The average is calculated on the first five minutes of samples, and then on the current average.
The
F3 key allows choosing to clear (Clr) the maximum, minimum and average data of the
captured measurements:
1.
In order to clear the data, select Clr with F3;
2.
Another drop-down menu will open;
3.
Use the arrow keys ▲▼ to select yes;
4.
Press ENTER to confirm.
5.
Press ESC or select no, to exit without clearing the data.
4.1.7 Instrument Setup
In order to set the instrument, you have to open the main menu by pressing
SETUP. See the next
chapter for further details.
4.1.8 Start of a new logging session
Press
MEM to start a Logging session: This key starts and stops the logging of a data block to be
saved in the instrument's internal memory. The data logging frequency is set in the
"Log interval"
menu parameter. The data logged between a start and subsequent stop represent a measurement block.
When the logging function is on, the LOG indication is displayed, the battery symbol blinks and a beep is issued each time a logging occurs. To end the logging, press
MEM again.
The instrument can turn off during logging between one capture and the next: The function is controlled by the
Auto_shut_off_Mode parameter. When the logging interval is less than one
minute, the logging instrument remains on; with an interval of at least one minute, it turns off between one capture and the next.
HD32.1 - 31 - V2.1
4.2 THE OPERATING PROGRAM B: DISCOMFORT ANALYSIS
Connect the probes. Turn on the instrument: After few seconds, the measurement display mode will appear:
The battery charge symbol and current time are up on the left (for further details, see the par. 10). The following quantities are reported:
Th: Air temperature detected at head height (1.7 m for a standing person; 1.1 m for a seated
person)
Tb: Air temperature detected at abdomen height (1.1 m for a standing person; 0.6 m for a seated
person)
Tn: Temperature of the net radiometer Tk: Air temperature detected at ankle height (0.1 m) Tf: Temperature at floor level P: Net radiation DT: Radiant asymmetry temperature
By pressing
SHIFT FNC, a drop-down menu is shown with the following information:
unit: allows selection of the unit of measurement
data: allows display of the maximum, minimum, and average values
4.2.1 The unit of measurement “Unit”
Proceed as follows to access unit funct ion:
Use SHIFT FNC to open the drop-down menu;
Use the arrow keys ▲▼ to select unit;
Press ENTER to confirm: the selected quantity is displayed in the central line of the
display;
The three different temperature units of measurement are displayed in the bottom line of the
display, using
F1, F2 or F3: The unit is selected and displayed near the value shown in the
central line;
Press ESC to exit the drop-down menu without making any change.
HD32.1 - 32 - V2.1
4.2.2 The maximum, minimum and average values of the captured quantities
Proceed as follows to access data function:
Use SHIFT FNC to open the drop-down menu;
Use the arrow keys ▲▼ to select data;
Press ENTER to confirm: the selected quantity is displayed in the central line of the
display;
The three quantities max (maximum), min (minimum) and avg (average) are shown in the
bottom line of the display, using
F1 or F2.
The
F3 key allows choosing to clear (Clr) the maximum, minimum and average data of the
captured measurements:
In order to clear the data, select Clr with F3;
Another drop-down menu will open;
Use the arrow keys ▲▼ to select yes;
Press ENTER to confirm.
Press ESC or select no, to exit without clearing the data.
4.3 THE OPERATING PROGRAM C: PHYSICAL QUANTITIES
Connect the probes. Turn on the instrument: After few seconds, the measurement display mode will appear (according to the selected page): To shift from one display to the other, press
F3.
fig. 1-a fig. 1-b fig. 1-c
fig. 1-d fig. 1-e fig. 1-f
HD32.1 - 33 - V2.1
fig. 1-a: Display of the measurement using the Pt100 SICRAM probe fig. 1-b: Display of the measurement using the combined humidity/temperature SICRAM probe fig. 1-c: Display of the measurement using the combined speed/temperature SICRAM probe fig. 1-d: Display of the measurement using the photometric/radiometric SICRAM probe fig. 1-e: Display of the measurement using the CO2 SICRAM probe fig. 1-f: Display of the measurement using the CO SICRAM probe
By pressing
SHIFT FNC, a drop-down menu is shown with the following information:
unit: allows selection of measuring unit for the actual variable.
data: Allows display of the maximum, minimum, and average values.
section: Allows setting the pipeline section for flow rate calculation
Cal CO2: allows starting of calibration procedure for the CO2 probe.
Cal CO: allows starting of calibration procedure for the CO probe and sensor change.
4.3.1 Setting the pipeline section
Proceed as follows to access section function:
Use SHIFT FNC to open the drop-down menu;
Use the arrow keys ▲▼ to select section;
Press ENTER to confirm;
The Shortcut menu will appear:
Use the ▲▼ navigation keys to select section and press enter. The following screen will
appear:
Use the ◄► navigation keys to highlight the digit in the section. Use the ▲▼ navigation
keys to modify the highlighted digit.
The area comprised must be between 0.0001 m2 (1 cm2) and 1.9999 m2.
Press ◄► to select the digit
Press ▲▼ to modify the digit
HD32.1 - 34 - V2.1
Pressing F1 toggles the display between m2 and inch2;
Press enter to confirm the information and exit from the setting section.
Press ESC to exit the drop-down menu without making any change.
4.3.2 The maximum, minimum and average values of the captured quantities
Proceed as follows to access
data function:
Use SHIFT FNC to open the drop-down menu;
Use the arrow keys ▲▼ to select data;
Press ENTER to confirm: the selected quantity is displayed in the central line of the
display;
The three quantities max (maximum), min (minimum) and avg (average) are shown in the
bottom line of the display, using
F1 or F2.
The
F3 key allows choosing to clear (Clr) the maximum, minimum and average data of the
captured measurements:
In order to clear the data, select Clr with F3;
Another drop-down menu will open;
Use the arrow keys ▲▼ to select yes;
Press ENTER to confirm.
Press ESC or select no, to exit without clearing the data.
HD32.1 - 35 - V2.1
5. MAIN MENU
To access the programming menu press SETUP: The setting menu will be displayed with the following items:
0) Info 5) Firmware
1) Logging 6) Time/date
2) Serial 7) Calibrate
3) Reset 8) Key lock
4) Contr. 9) Password
If you do not press any key within 2 minutes, the instrument goes back to the main display. Use the arrow keys ▲▼ ◄► and press
ENTER to select an item.
To exit the selected item and return to the previous menu, press
ESC.
To exit immediately from the main menu, press
SETUP again.
5.1 INFO MENU
Once you enter the main menu by pressing SETUP, press and ENTER to access the Info menu, The following information on the instrument will be displayed:
Instrument code and operating program, firmware date and version, serial number, instrument calibration date, user identification code.
Model HD32.1 Prog. A: Microclimate Analysis Operating Program
Model HD32.1 Prog. B: Discomfort Analysis Operating Program
Model HD32.1 Prog. C: Physical Quantities Operating Program
To change the
USER ID, press ENTER. Using the arrows , select the digit you want to
change and modify it with arrows ▲ ▼. Proceed for all other digits and, at the end, confirm with the
ENTER key. Note: The USER ID can also be changed by software.
Press
ESC to return to the main menu. Press SETUP to exit the menu.
HD32.1 - 36 - V2.1
5.2 LOGGING MENU
Once you enter the main menu by pressing SETUP, to access the Logging menu proceed as follows:
1.
Use the arrow keys ▲▼ to select Logging;
2.
Press ENTER:
The parameter setting submenu for the logging sessions (measured data capture) will be displayed.
You can set the data capture frequency (Log interval) and the automatic l
ogging start (Start/stop
time). The capture interval is the same for all probes.
5.2.1 Log Interval
Use this item to set the LOG interval (interval between two subsequent sample captures): To enter this setting, proceed as follows: Once you have accessed the LOGGING submenu (previous par.) use the arrow keys to select Log Interval:
1.
Use the arrow keys to select the interval duration from 15 seconds to one hour;
2.
Press ENTER to confirm and return to the Logging menu;
3.
Press ESC to return to the Logging menu without making any change;
4.
Press ESC again to return to the main menu;
5.
Press SETUP to exit immediately from the menu.
These are the available values: 15 seconds - 30 seconds - 1 minute - 2 minutes - 5 minutes - 10 minutes - 15 minutes - 20 minutes - 30 minutes - 1 hour
Storage interval Storage Capacity Storage interval Storage Capacity
15 seconds About 11 days and 17 hours 10 minutes About 1 year and 104 days 30 seconds About 23 days and 11 hours 15 minutes About 1 year and 339 days
1 minute About 46 days and 22 hours 20 minutes About 2 years and 208 days 2 minutes About 93 days and 21 hours 30 minutes About 3 years and 313 days 5 minutes About 234 days and 17hours 1 hour About 7 years and 261 days
HD32.1 - 37 - V2.1
5.2.2 Self Shut-off mode
The Self shut-off mode item controls the instrument's automatic turning off during logging, occurring between the capture of a sample and the next one.
When the interval is lower than 60
seconds, the instrument will always remain on. With intervals greater than or equal to 60
seconds, it is possible to turn off the instrument between loggings: it will turn on at the moment of sampling and will turn off immediately afterwards, thus increasing the battery life.
Once you have accessed the LOGGING submenu (previous par.) use the arrow keys to select Self shut_off mode:
If the set Log Interval (see previous par.) is lower than 60 seconds, the following will be
displayed
If the set Log Interval (see previous par.) is greater or equal to 60 seconds, the following will
be displayed
1.
By using the arrow keys you can select:
STAY ON (the instrument stays on) SHUT OFF (the instrument stays off)
2.
Press ESC to return to the Logging menu ;
3.
Press ESC again to return to the main menu;
4.
Press SETUP to exit immediately from the menu.
HD32.1 - 38 - V2.1
5.2.3 Start/stop time – Automatic start
The logging start and end can be programmed by entering the date and time. When called, the function suggests the current time plus 5 minutes as the start time: Press <ENTER> to confirm or set the date and time using the arrow keys. Then you are asked to set the data to end the recording: By default the instrument suggests the start time plus 10 minutes. The default suggested values are such to allow the user to setup an instrument ready for measurement.
NOTE: By default the set time is 5 minutes after the current time.
To enter this setting, proceed as follows.
Once you have accessed the LOGGING submenu (previous par.) use the arrow keys to select Start/Stop time: The following message “Enter start time” will be displayed:
1.
Use the arrow keys to select the data to be changed (year/month/day and
hour:minutes:seconds);
2.
Once selected, the data will start blinking;
3.
Use the arrow keys ▼▲ to change its value;
4.
Confirm by pressing ENTER;
6.
Press ESC to return to the Logging menu without making any change;
7.
Press ESC again to return to the main menu;
8.
Press SETUP to exit immediately from the menu.
After setting the logging start time, the logging end time (enter stop time) window will be displayed:
1.
Use the arrow keys to select the data to be changed (year/month/day and
hour:minutes:seconds);
2.
Once selected, the data will start blinking;
3.
Use the arrow keys ▼▲ to change its value;
4.
Confirm by pressing ENTER;
5.
Press ESC to return to the Logging menu without making any change;
6.
Press ESC again to return to the main menu;
7.
Press SETUP to exit immediately from the menu.
HD32.1 - 39 - V2.1
NOTE: By default the acquisition end time is 10 minutes after the logging session start time.
8.
Once both values have been set, a summary will be displayed: Start and end time of the LOG
session.
9.
Press ENTER to confirm or ESC to exit without enabling the automatic start: In both cases, you
will return to the LOGGING menu.
10.
Press SETUP to exit immediately from the main menu.
When the instrument starts automatically a LOG session, a beep is issued on each capture and the blinking
LOG message is shown at the top of the display.
Press
MEM to stop the session before the set time.
To cancel the automatic start setting, use the
Cancel auto start function as illustrated in the
following paragraph.
NOTE: The automatic logging session is started even when the instrument is off. If it is off
when the automatic logging session is started, the instrument is turned on few seconds earlier and remains on at the end of logging. If it is powered by the battery, it is turned off when idle for some minutes at the end of the logging session. See paragraph 4.2.2 to set the automatic shut off.
5.2.4 Cancel auto start
Once the LOG session start and end times are set, you can prevent the session automatic start by using Cancel auto start. Once you have accessed the LOGGING submenu:
1.
Use the arrow keys to select Cancel auto start
2.
The LOG session start and end times will be displayed:
HD32.1 - 40 - V2.1
3. By pressing the following message will be displayed: ”Self timer not active”;
4. Press ENTER to cancel the automatic start;
5.
Press ESC to exit without cancelling the automatic start;
6.
Press ESC again to exit from the submenus;
7.
Or press SETUP to exit immediately from the main menu.
See the previous paragraph to set a new automatic start time after cancelling the previous one.
5.2.5 Log File Manager
This item allows managing the captured logs: the instrument allows printing the files of the captured data (Print selected log)and deleting all memory (Erase ALL logs) . The instrument can store up to 64 LOG sessions numbered progressively from 00 to 63, in a 4-line and 4-column layout. If there are more than 16 sessions, press F1 (
Page-) to go back to the previous screen and
F3 (
Page+) to go to the next one. The current page (0,1,2 or 3) and the total pages with stored data
are displayed in the upper right corner: in the example below, “0/3” refers to page 0 of 3 with stored data.
Once you have accessed the LOGGING submenu:
1.
Use the arrow keys to select Log File manager: the following submenu will display
HD32.1 - 41 - V2.1
0) Print selected log
1)
Erase ALL logs
2)
Log time interval
2.
Use the arrow keys ▲▼ to select a menu item;
3.
Press ENTER to confirm;
4.
Press ESC to go back to menu;
5.
Press SETUP to exit the main menu directly.
NOTE: you can connect a PC or a serial port printer to the instrument RS232 serial port. If you
connect a parallel port printer, you will need a parallel-serial converter between the instrument and the printer (not supplied with the instrument).Before starting the printing via the RS232C port, set the baud rate. To do so, select Baud Rate
in the Serial menu (see par. 5.3.1 Baud Rate
) and select the maximum value equal to
38400 baud. If you connect a printer, set its maximum
value allowed.
The instrument to Pc or printer communication is possible provided that the instrument baud rate is the same as that of computer or printer.
0) Print selected log:
By selecting this item, the page of the log to be printed will be displayed:
1.
Use the arrow keys ▲▼◄►to select the log to be printed ;press F1 and F3 to go to another
page;
2.
once you select a file, the acquisition start date and time and the number of samples in the file
(Rec)are displayed in the lower corner of the display.
Files are stored in ascending order.
Each file is identified by date and time only, both shown on the display. In the example above, the 00 file is selected: logging began on 27Th January 2006 at 18.50. The file contains 6 samples.
3.
Press ENTER to print the selected log (or press ESC to return to the previous menu, without
printing);
NOTE: You can print a file only by using the same operating program that generated the data.
4.
The data transfer message will be displayed a few seconds, then the instrument will go back to
the
Print selected log page to select another log to be printed;
5.
Repeat the procedure to print the required sessions or press ESC to exit this menu;
6.
Press SETUP to exit immediately from the main menu.
1) Erase all memory
If you select this item, “ERASE ALL FILES” will display:
HD32.1 - 42 - V2.1
1. press ENTER to erase all files;
2. press ESC to undo and back to the previous menu;
3. press SET to exit the main menu directly.
2) Log time interval
It refers to logging time: when the set time interval expires, logging stops. Press the MEM key to stop logging before the set time interval expires. To disable this function, set time at 0. In this case, if you press the MEM key or the memory is full, logging will stop.
Use the arrow keys to change the set time, the allowed maximum value is 1 hour. Press
ENTER to confirm.
Press
ESC to exit this submenu without saving changes.
Press
SETUP to exit the main menu directly.
5.3 SERIAL MENU (SERIAL COMMUNICATION)
The Serial submenu allows setting the data transfer speed via serial port (Baud rate) and the record printing interval (Print Interval).
The LOG sessions can be downloaded on a PC, through serial
RS232 or USB connection.
In case of serial connection, the transfer speed can be set by the user (see next par.) but it can not be higher than 38400 bps. In case of USB connection, the transfer speed is fixed at 460800 bps.
After downloading the data on the PC, using the dedicated software, they will be processed by this software for graphic display and the calculation of the comfort/stress indexes.
The instrument can be connected directly to an 80 column serial printer.
To access the Serial submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Serial;
3.
Press ENTER;
HD32.1 - 43 - V2.1
4. You will get the Serial submenu.
5.3.1 Baud Rate
The Baud Rate indicates the speed used for the serial communication with the PC. To set the Baud rate, proceed as follows:
1.
Use the arrow keys ▲▼ to select the item;
2.
Press ENTER: You will get the following message:
3.
Use the arrow keys ▼▲ to set the value;
4.
Press ENTER to confirm and return to the previous page, or press ESC to cancel the change and
exit the menu item;
5.
Press ESC over and over to exit from the submenus;
6.
Press SETUP to exit immediately from the main menu.
WARNING: The communication between instrument and PC (or serial port printer) only works if the instrument and PC baud rates are the same. If the USB connection is used this
parameter value is automatically set.
NOTE: When setting the baud-rate, check the printer speed.
5.3.2 The Print Interval
To set the Print Interval, proceed as follows:
1.
Use the arrow keys ▲▼ to select the item;
2.
Press ENTER: You will get the following message:
HD32.1 - 44 - V2.1
3.
Use the arrow keys ▼▲ to set the value;
4.
Press ENTER to confirm and return to the previous page, or press ESC to cancel the change and
exit the menu item;
5.
Press ESC over and over to exit from the submenus;
6.
Press SETUP to exit immediately from the main menu.
The print interval can be set from 0 seconds to one hour: 0 s - 15 s - 30 s - 1 min. - 2 min. - 5 min. - 10 min. - 15 min. - 20 min. - 30 min. - 1 hour.
5.4 RESET
To enter the Reset submenu in order to carry out a complete reset of the instrument, proceed as follows:
1.
Press SETUP
2.
Use the arrow keys ▲▼ to select Reset
3.
Press ENTER: You will get the following message
4.
Use the arrow keys ▲▼ to select Reset
5.
Press ENTER to confirm, or press ESC over and over to exit from the submenus
6.
Press SETUP to exit immediately from the main menu.
5.5 CONTRAST
This menu item allows increasing or decreasing the contrast on the display: To access the Contrast submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Contr.
3.
Press ENTER:
4.
You will get the following message
HD32.1 - 45 - V2.1
5.
Use the arrow keys ◄► to decrease or increase the contrast;
6.
Press ENTER or ESC to return to the main menu;
7.
Press SETUP to exit immediately from the main menu.
5.6 FIRMWARE
This menu item allows changing the instrument’s operating program.
To access the Firmware submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Firmware;
3.
Press ENTER;
4.
You will get the following display:
5.
Use the arrow keys ▲▼ to select the operating program that you want to install;
6.
Press ENTER to confirm and wait for the chosen program self-installation;
7.
At the end the instrument will reset and get ready for the chosen program.
Note: The operating program must be present in the instrument.
5.7 TIME/DATE
This menu item allows setting the date and time that will be shown at the top of the display. To access the Time/date submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Time/date
3.
Press ENTER:
4.
You will get the following message
HD32.1 - 46 - V2.1
5.
Use the arrow keys to select the data to be set (year/month/day and hour:minutes);
6.
Once selected, the data will start blinking;
7.
Use the arrow keys ▼▲ to enter the correct value;
8.
Press ENTER to confirm and return to the main menu;
9.
Or press ESC to return to the menu without making any change;
10.
Press SETUP to exit immediately from the main menu.
NOTE: In regard to the time, you can set hours and minutes. The seconds are always set to 00 (set
00 seconds!!).
5.8 CALIBRATE
The Calibrate menu is reserved to Technical Support. It reports the calibrations and the last calibration performed: To access the Calibrate submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Calibrate
3.
Press ENTER:
4.
You will get the following message:
5.
Press ENTER or ESC to go back to the main menu: you cannot change anything: only
Technical Support can calibrate the instrument.
6.
Press SETUP to exit the main menu directly.
NOTE: You cannot change the calibration date.
5.9 KEY LOCK
This menu item allows LOCKING/UNLOCKING the instrument, when the password has been input: See the next chapter for further details.
HD32.1 - 47 - V2.1
To access the Key lock submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Key lock
3.
Press ENTER:
4.
You will get the following message: “Enter password”
5.
Use the arrow keys ▼▲ to enter the correct password;
6.
Press ENTER to confirm (or ESC to cancel);
By pressing
ENTER you return to the main menu and the instrument is locked: A “key” is
displayed at the top left of the display;
WARNING! When the instrument is locked by a password, all keys are locked, except MEM, used
to start the LOG session and
SETUP, ENTER and ESC that allow entering the main menu to
unlock the instrument. Therefore the user has to set all required parameters, protect the instrument using the KEY LOCK function and start the LOG session, in order to prevent any undesired access by unauthorized personnel.
To
unlock the instrument, repeat the steps above: Enter the main menu and unlock the instrument
using the Key lock and entering the password.
If the password is wrong, you will get the message “Wrong password”.
5.10 PASSWORD
This menu item allows setting a password to protect the instrument from unauthorized access. There are two types of passwords available,
both consisting of four characters:
The default password consists of four zeros: 0000.
The user password: can be set by the user to protect the instrument from unauthorized access;
The factory password is reserved to Technical Support.
To access the Password submenu, proceed as follows:
1.
Press SETUP;
2.
Use the arrow keys ▲▼ to select Password
3.
Press ENTER:
4.
The following message will appear:
HD32.1 - 48 - V2.1
5.
Use the arrow keys ▲▼ to select the current password
6.
Press ENTER to confirm (or ESC to cancel);
7.
The following message will appear:
8.
Use the arrow keys ▼▲ to enter the new password;
9.
Press ENTER to confirm (or ESC to cancel) and go back to the main menu;
10.
Press SETUP to exit the main menu directly.
WARNING! The User password allows you to lock/unlock the instrument (see paragraph 5.9 Key
lock).
HD32.1 - 49 - V2.1
6. PROBES AND MEASUREMENTS
6.1 A AND B OPERATING PROGRAM PROBES : A: Microclimate Analysis B: Discomfort Analysis
TP3207
Temperature probe Sensor type: Thin film Pt100
(*)
Measurement uncertainty: Class 1/3 DIN Measurement range: -40°C … +100°C Connection: 4 wires plus SICRAM
module Connector: 8-pole female DIN45326 Cable length: 2 meters Dimensions: Ø=14 mm L=140 mm
(**)
Response Time T95 15 minutes
HD32.1 - 50 - V2.1
TP3275
Globe thermometer probe Ø=150 mm according to ISO 7243 - ISO 7726 Sensor type: Pt100
(*)
Measurement uncertainty: Class 1/3 DIN Measurement range: -10°C … +100°C Connection: 4 wires plus
SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
(**)
T
95
Response Time 15 minutes
HD32.1 - 51 - V2.1
TP3276
Globe thermometer probe Ø=50 mm Sensor type: Pt100
(*)
Measurement uncertainty: Class 1/3 DIN Measurement range: -10°C … +100°C Connection: 4 wires plus SICRAM
module
Connector: 8-pole female
DIN45326
Cable length: 2 meters
(**)
Response Time T95 15 minutes
HD32.1 - 52 - V2.1
TP3227K
Probe composed of 2 standalone temperature probes, Pt100 sensor. Used for local discomfort measurement Due to vertical temperature gradient in order to study standing or seated persons. Adjustable height, complete with extension code
TP3227.2 (L=450
mm, Ø=14) Sensor type: Thin film Pt100
(*)
Measurement uncertainty:
Class 1/3 DIN
Measurement range: -10°C … +100°C Connection: 4 wires plus 2-input
SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
(**)
Upper probe T95 15 minutes Lower probe T95 4 minutes
The
TP3227K probe can be used for simultaneous
measurement of temperature at 1.10 m and 0.10 m.
In order to perform simultaneous measurements at different heights:
In case of a standing person: 1.70 m, 1.10 m and 0.10 m from the floor
In case of a seated person: 1.10 m, 0.60 m and 0.10 m from the floor
You can use the following probes:
TP3227K
Dual probe capable of measuring:
Temperature at 1.70 m and 1.10 m from the floor in case of standing person;
Temperature at 1.10 m and 0.60 m from the floor in case of seated person;
TP3227PC
Dual probe for temperature measurement at floor level and at ankle height (0.10 m).
HD32.1 - 53 - V2.1
TP3227PC
Probe composed of 2 standalone temperature probes. Used for local discomfort measurement due to vertical temperature gradient. Suitable for temperature measurement at floor level and at ankle height (0.10 m). The TP3227PC has priority on the TP3227.1, if both are connected.
Sensor type: Thin film Pt100
(*)
Measurement uncertainty:
Class 1/3 DIN
Measurement range: -10°C … +100°C Connection: 4 wires plus 2-input
SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
(**)
Ankles T
95
4 minutes
Floor T
95
20 minutes
HD32.1 - 54 - V2.1
TP3207P
Temperature measurement probe at floor level, used for local discomfort measurement due to vertical temperature gradient. Sensor type: Thin film Pt100
(*)
Measurement uncertainty:
Class 1/3 DIN
Measurement range: -10°C … +100°C Connection: 4 wires plus SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
(**)
Response Time T95 20 minutes
HD32.1 - 55 - V2.1
TP3207TR
Combined probe for radiant temperature measurement. Used to assess the unsatisfied with the radiant asymmetry percentage. Sensor type: Pyranometer / NTC
(*)
Measurement uncertainty:
NTC ± 0.15 Typical spectral sensitivity 10µV/(W/m
2
) Measurement range: 0°C … +60°C Connection: 4 wires plus SICRAM
module Connector: 8-pole female DIN45326 Cable length: 2 meters
(**)
Net radiometer T
95
90 seconds
NTC T95 20 minutes
HD32.1 - 56 - V2.1
AP3203 / AP3203-F
Omni directional hot-wire probe. Sensor type: NTC 10Kohm
(*)
Measurement uncertainty: ± 0.2 m/s (0.1 … 1 m/s)
± 0.3 m/s (1 … 5 m/s)
Measurement range: 0.1 … 5 m/s
0°C … +80°C (AP3203) / -30°C … +30°C (AP3203-F)
Connection: 7 wires plus SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
HD32.1 - 57 - V2.1
HP3201
Natural ventilation wet bulb probe for WBGT index measurement Sensor type: Pt100
(*)
Measurement uncertainty:
Class A
Measurement range: +4°C … +80°C Connection: 4 wires plus SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters Cotton wick length: 16 cm Tank capacity: 15 cc Tank autonomy: 96 hours with RH=50%, t=23°C
(**)
Response time T95 15 minutes
HD32.1 - 58 - V2.1
TP3204S
Natural ventilation wet bulb probe for WBGT index measurement Sensor type: Pt100
(*)
Measurement uncertainty:
Class A
Measurement range: +4°C … +80°C Connection: 4 wires plus SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters Cotton wick length: 10 cm approx. Tank capacity: 500 cc Tank autonomy: 15 days @ t = 40 °C
(**)
Response time T95 15 minutes
For the start up go on as indicated below:
-
Remove the sensor cap (the cap is not screwed).
-
Insert the cotton wick, previously dipped with distilled water, into the temperature probe. The
cotton wick must protrude from the probe for about 20 mm.
-
Replace the cap.
-
Fill the bottle with 500 cc of distilled water.
-
Turn the probe over and firmly screw the bottle to the probe reservoir.
-
Turn the probe quickly (to avoid water spillage).
-
Secure the probe stem to the clamp of the support rod (the two screws at the bottom of the probe
are not used with clamp fixing).
Screws
Reservoir
Cap
Cotton wic
k
Bottle
Protection fro
m
solar radiations
HD32.1 - 59 - V2.1
HP3217R
Relative humidity and temperature combined probe. Used for environment comfort indexes measurement
Sensors type: - Thin film Pt100 for temperature
-
Capacity sensor for relative
humidity
(*)
Measurement uncertainty:
Temperature: 1/3 DIN
Relative humidity: ± 2.5% Measurement range:
Temperature: -40°C … +100°C
Relative humidity: 0…100%RH Connection: 7 wires plus SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
(**)
%RH T95 1 minute Temperature T
95
15 minutes
HD32.1 - 60 - V2.1
HP3217DM
Two-sensor probe for natural ventilation wet bulb temperature and dry bulb temperature measurement. Used for environment comfort indexes measurement. Sensor type: Pt100
(*)
Measurement uncertainty:
Class A
Measurement range:
Natural wet Dry temperature
+4°C … +80°C
-30°C … +100°C
Connection: 7 wires plus 2-input
SICRAM module Connector: 8-pole female DIN45326 Cable length: 2 meters
Tank capacity: 15 cc Tank autonomy: 96 hours with RH=50%,
t=23°C
(**)
Dry bulb T95 4 minutes Wet bulb T95 30 minutes
The TP3217DM has priority on the: HD3201 and TP3207, if connected
.
HD32.1 - 61 - V2.1
HD320B2
CO2 Carbon Dioxide probe
CO2 measurement principle:
Infrared technology (NDIR) with double source
(*)
Measurement uncertainty:
±(50ppm+3% of the measurement) at 20°C, 50%RH and 1013hPa
Measurement range: 0 … 5000ppm
Resolution: 1ppm
Connector: 8 female poles DIN45326
Cable length: 2 metres
(**)
Response Time T63: 2 minutes
Temperature Effect: 0.2%/°C CO2 (Typical value)
Atmospheric pressure effect:
Compensated with the atmospheric pressure, inside the instrument
Long-term stability: 5% of the range/5 years (Typical
value)
Calibration: At one point on 0ppm or 400ppm
Work Relative Temperature/Humidity :
-5 …+50°C, 0 ... 95%RH no condensing
HD32.1 - 62 - V2.1
HD320A2
CO Carbon Monoxide probe
CO measurement principle:
Electro chemical cell with two electrodes
(*)
Measurement uncertainty:
±(3ppm+3% of the measurement) at 20°C, 50%RH and 1013hPa
Measurement range: 0 … 500ppm
Resolution: 0.1ppm
Connector: 8 female poles DIN45326
Cable length: 2 metres
(**)
Response Time T63: 1 minute
Expected life of the probe:
Usually, 5 years in normal environment conditions
Long-term stability: 5% of the measurement/year (Typical
value)
Calibration: At one point on 0ppm
Work Relative Temperature/Humidity:
-5 …+50°C, 0 ... 95%RH no condensing
The CO HD320A2 probe fixes on the HD320B2 probe through the suitable magnetic support code “
HD320A2S”.
(*) The probe is calibrated. Calibration data are stored in the SICRAM module.
(**) Response time T
95
is the time required to reach 95% of the final value. The response time measurement is
performed at neglectable air speed (still air).
HD320B2
HD320A2
HD32.1 - 63 - V2.1
HD3218K
Rod complete with clamp and fastening screw to support the probes.
AM32
Rod complete with 2 clamps and fastening screws to support the probes.
LP 32 F/R
Support bracket for photometric-radiometric probes for Light measurement LP471...
HD32.1 - 64 - V2.1
VTRAP32K
A tripod code VTRAP32 is available for the measurements. Adjustable height up to 1.50 meters, complete with head that can host up to 6 measurement probes. The same tripod can be used to support the measurement instrument during data capture.
The arms fitted with suitable clamps for the measurement probes can be inserted in the head, code
HD3218K.
The
VTRAP32K kit is composed of a tripod
code
VTRAP32 and 4 arms code HD3218K.
HD32.1 - 65 - V2.1
Performing the measurement
The tripod and required probes are assembled where you wish to perform the measurement. Then you need to setup the instrument and start the measurement. If you have to carry out the measurement s in another location, you need to move everything in that new location.
At the end of the measurement session, or later, the data are transferred to a PC for processing and reports.
6.1.1 Warnings, care and maintenance of the probes
Do not expose the probes to gases or liquids that could corrode the material of the probe. Clean
the probes carefully after each measurement.
Do not bend the probe connectors or force them upward or downward.
Comply with the correct polarity of the probes.
Do not bend or force the contacts when inserting the probe connector into the instrument.
Do not bend, deform or drop the probes, as this could cause irreparable damage.
Always select the most suitable probe for your application.
To obtain reliable measurements, temperature variations that are too rapid must be avoided.
Some probes are not insulated from their external casing; be very careful not to come
into contact with live parts (above 48V). This could be extremely dangerous for the instrument as well as for the operator, who could be electrocuted.
Avoid taking measurements in presence of high frequency sources, microwave ovens or large
magnetic fields; results may not be very reliable.
The instrument is water resistant, but should not be immersed in water. Should the instrument
fall into the water, check for any water infiltration.
Wind speed hot-wire sensor AP3203
The wind speed hot-wire sensor AP3203 is heated and, in the presence of gas
vapors, could trigger a fire or explosion. Do not use the probe in the presence of inflammable gases. Ensure that no potentially explosive gas or vapor leakage is present in the measurement environments.
The probe is delicate and should be handled with extreme care. As the sensor is only
partially protected during use, a simple collision could render the probe unusable.
After measurement, the sensor set on the probe head must be protected with the supplied
threaded cylinder.
During use, the AP3203 omni directional probe must be protected with the special metallic
sphere.
The sphere protection must be removed after use, and the sensor must be closed into the special
protection cylinder.
Do not let fingers touch the sensors.
Use only distilled water to clean the probe.
HD32.1 - 66 - V2.1
The
AP3203 probe is fitted with a spherical
protection screen. To reduce the space occupied when not used, the
AP3203 is supplied with a
protection cylinder that can be screwed on the probe's head.
Relative humidity (RH) and temperature probe HP3217R
- Do not let fingers touch the sensors. Avoid staining them with oil, grease, resins.
-
The sensor base is in alumina so it could easily break.
-
The sensors can be cleaned from dust and smog using distilled water and a very soft brush (e.g.
badger);
-
If the measurements are not consistent, check that the sensors are not dirty, corroded, splintered
or broken.
-
In order to check the RH measurement consistency you can use the standard saturated salt
solutions:
HD75 (75% RH) and HD33 (33% RH).
Natural ventilation wet bulb probe HP3201 Dual probe for natural ventilation wet bulb and dry bulb temperature measurement HP3217DM
The commissioning is carried out as follows:
-
Remove the cover, it is not screwed-on.
-
Insert the braid in the temperature probe; the braid has been previously wet with distilled water.
-
Fill a container up to ¾ with distilled water.
-
Close the container’s cover.
-
Warning: Do not turn the probe vertically as water will spill out.
- The braid should protrude about 20 mm from the temperature probe.
-
The braid will calcify with time (harden); it should be replaced periodically.
TP3227K probe composed of 2 standalone temperature probes, Pt100 sensor and TP3227PC probe composed of 2 standalone temperature probes, Pt100 sensor:
Adjustment of the sensors at 1.70 m, 1.10 m and 0.10 m:
Screw the telescopic rod code TP3227.2 L=450mm to the probe TP3227. Once the rod has been fastened on the clamp, adjust the height to 1.70 m for the fixed sensor. The sliding probe should be placed at 1.10 m from the floor. You should use the combined ankle/floor probe
TP3227PC
to perform the measurements at 0.10 m from the floor.
HD32.1 - 67 - V2.1
TP3227K probe composed of 2 standalone temperature probes, Pt100 sensor and TP3227PC probe composed of 2 standalone temperature probes, Pt100 sensor:
Adjustment of the sensors at 1.10 m, 0.60 m and 0.10 m:
Screw the telescopic rod code TP3227.2 L=450mm to the probe TP3227. Once the rod has been fastened on the clamp, adjust the height to 1.10m for the fixed sensor. The sliding probe should be placed at 0.60 m from the floor. You should use the combined ankle/floor probe
TP3227PC
to perform the measurements at 0.10 m from the floor.
Combined probe for radiant temperature measurement TP3207TR
The face of the probe marked by the symbol is the air flow hot side. It should be oriented toward the hot source (wall/wall ceiling/floor or floor/ceiling).
6.2 PROBES FOR THE OPERATING PROGRAM C: PHYSICAL QUANTITIES
6.2.1 Temperature measurement using the probe Pt100 complete with SICRAM module
The instrument works with temperature probes fitted with the SICRAM module (with a Platinum Pt100 sensor with 100 resistance at 0°C). The excitation current was chosen in order to minimize the sensor self-heating effects. The SICRAM module acts as an interface between the sensor on the probe and the instrument. There is a microprocessor circuit with a permanent memory inside the module that enables the instrument to recognize the type of probe connected and to read its calibration information.
Upon turning on the instrument automatically detects the probes fitted with SICRAM module:
The probes are detected during turn on, and this cannot be performed when the instrument is already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it off and on.
In all versions the temperature sensor is housed in the end part of the probe.
The response time for the measurement of the temperature in air is greatly reduced if the air is moving. If the air is still, stir the probe back and forth. The response times are longer than those for liquid measurements.
The temperature measurement by
immersion is carried out by inserting the probe in the liquid for at
least 60mm; the sensor is housed in the end part of the probe.
In the temperature measurement by
penetration the probe tip must be inserted to a depth of at least
60mm, the sensor is housed in the end part of the probe. when measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe.
In order to perform a correct
contact measurement, the measurement surface must be even and
smooth, and the probe must be perpendicular to the measurement plane. A contact measurement is hard to perform due to various factors: The operator must be experienced in handling the probe and consider all the factors influencing it.
So as to obtain the correct measurement, the insertion of a drop of oil or heat-conductive paste is useful (do not use water or solvents). This method improves the response time, in addition to accuracy.
The °C or °F unit of measurement can be chosen for display, printing, and logging.
The sensor is calibrated in the factory, and the Callendar Van Dusen parameters are recorded in the SICRAM module.
HD32.1 - 68 - V2.1
6.2.2 Technical information on temperature probes Pt100 using SICRAM module
Model Type Application range Accuracy
TP472I
Immersion -196°C…+500°C
TP472I.O
Immersion -50°C…+300°C
TP473P.I
Penetration -50°C…+400°C
TP473P.O
Penetration -50°C…+300°C
TP474C.O
Contact -50°C…+300°C
TP475A.O
Air -50°C…+250°C
TP472I.5
Penetration -50°C…+400°C
TP472I.10
Penetration -50°C…+400°C
±0.1 °C (t = 0 °C) ±0.2 °C (-50 °C ≤ t 250°C)
±0.3 °C (t < -50 °C; t > 250 °C)
Common characteristics
Resolution 0.01°C in the range ±199.99°C,
0.1°C in the remaining range
Temperature drift @20°C 0.003%/°C
HD32.1 - 69 - V2.1
6.2.3 Measurement of relative humidity using the combined humidity/temperature probe
The instrument works by using combined humidity/temperature probes (temperature with Pt100 sensor). The combined humidity/temperature probes are fitted with SICRAM module that acts as an interface between the sensor on the probe and the instrument. There is a microprocessor circuit with a permanent memory inside the module that enables the instrument to recognize the type of probe connected and to read its calibration information.
The probes are detected during turn on, and this cannot be performed when the instrument is already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it off and on
.
Measurement of relative humidity
The humidity probes are humidity/temperature combined probes: The humidity sensor is a capacitive type sensor, the temperature sensor is a Pt100 (100Ω at 0°C). The instrument measures relative humidity %RH and temperature, and starting from the fixed barometric pressure value of 1013.25mbar it calculates the following resulting quantities:
g/kg Grams of vapor in a kilogram of dry air
g/m
3
Grams of vapor in a cubic meter of dry air
hPa Partial vapor pressure (hPa)
J/g Enthalpy
Td Dew point (°C or °F)
Tw Wet bulb temperature (°C or °F)
A measurement is performed by placing the probe in the area of whose parameters you wish to measure. Keep the probe far from elements that might interfere with measurement such as: heat or sources of cooling, walls, air-streams, etc. Avoid temperature drops that might cause condensation. A reading taken when no heat drop occurs is practically immediate. In contrast, in conditions involving heat drops, it is necessary to wait until the sensors and their housing have reached a thermal equilibrium in order to prevent heat irradiation or absorption on the relative humidity sensor, which would cause a faulty measurement. Since temperature affects relative humidity; move the probe like a fan in order to speed the response time in the presence of heat drops.
The calibration of the humidity/temperature sensor by the user is not required. The humidity sensor is calibrated in our laboratory at 23°C at the points of 75%RH, 33%RH and 11.4%RH. On request, the probes can be checked at different isotherms. The temperature sensor is calibrated in the factory and the Callendar Van Dusen parameters are recorded in the SICRAM module.
HD32.1 - 70 - V2.1
6.2.4 Technical information on relative humidity and temperature probes using SICRAM module
Application range Accuracy
Model
Temperature
sensor
%RH Temperature %RH Temp.
HP472ACR
Pt100 0…100%RH -20…+80 °C ±0.3 °C
HP473ACR
Pt100 0…100%RH -20…+80 °C ±0.3 °C
HP474ACR
Pt100 0…100%RH -40…+150 °C ±0.3 °C
HP475ACR
Pt100 0…100%RH -40…+150 °C ±0.3 °C
HP475AC1R
Pt100 0…100%RH -40…+180 °C ±0.3 °C
HP477DCR
Pt100 0…100%RH -40…+100 °C ±0.3 °C
HP478ACR
Pt100 0…100%RH -40…+150 °C
±1.5% (0…85%RH)
±2.5% (85…100%RH)
@ T=15…35 °C
(2 + 1.5% measure)%
@ T= remaining range
±0.3 °C
Common characteristics
Relative humidity
Sensor Capacitive Resolution 0.1%RH Temperature drift @20°C 0.02%RH/°C Response time %RH at constant temperature 10sec (10
Æ80%RH; air speed=2m/s)
Temperature
Resolution 0.1°C Temperature drift @20°C 0.003%/°C
Important notes:
1) Do not let hands touch the RH sensor.
2) The sensor base is in alumina so it could easily break
3) Storage of the saturated solutions: The saturated solutions must be stored in a dark environment
at a constant temperature of about 20°C with the container well closed inside a dry room.
HD32.1 - 71 - V2.1
Relative humidity of saturated salts at different temperatures
Temp. Lithium Magnesium Sodiu m
°C Chloride Chloride Chloride
0 11.23 ± 0.54 33.66 ± 0.33 75.51 ± 0.34
5 11.26 ± 0.47 33.60 ± 0.28 75.65 ± 0.27
10 11.29 ± 0.41 33.47 ± 0.24 75.67 ± 0.22
15 11.30 ± 0.35 33.30 ± 0.21 75.61 ± 0.18
20 11.31 ± 0.31 33.07 ± 0.18 75.47 ± 0.14
25 11.30 ± 0.27 32.78 ± 0.16 75.29 ± 0.12
30 11.28 ± 0.24 32.44 ± 0.14 75.09 ± 0.11
35 11.25 ± 0.22 32.05 ± 0.13 74.87 ± 0.12
40 11.21 ± 0.21 31.60 ± 0.13 74.68 ± 0.13
45 11.16 ± 0.21 31.10 ± 0.13 74.52 ± 0.16
50 11.10 ± 0.22 30.54 ± 0.14 74.43 ± 0.19
55 11.03 ± 0.23 29.93 ± 0.16 74.41 ± 0.24
60 10.95 ± 0.26 29.26 ± 0.18 74.50 ± 0.30
65 10.86 ± 0.29 28.54 ± 0.21 74.71 ± 0.37
70 10.75 ± 0.33 27.77 ± 0.25 75.06 ± 0.45
75 10.64 ± 0.38 26.94 ± 0.29 75.58 ± 0.55
80 10.51 ± 0.44 26.05 ± 0.34 76.29 ± 0.65
85 10.38 ± 0.51 25.11 ± 0.39
90 10.23 ± 0.59 24.12 ± 0.46
95 10.07 ± 0.67 23.07 ± 0.52
100 9.90 ± 0.77 21.97 ± 0.60
HD32.1 - 72 - V2.1
6.2.5 Wind speed measurement
The instrument works with hot-wire and vane probes fitted with the SICRAM module.
The SICRAM module acts as an interface between the sensor on the probe and the instrument. There is a microprocessor circuit with a permanent memory inside the module that enables the instrument to recognize the type of probe connected and to read its calibration information.
Note: The vane probes can be exclusively connected to input 8.
The probe is detected during turn on, and this cannot be performed when the instrument is already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it off and on
.
The AP471 and AP472 series probes measure the incident wind speed and flow rate; some also measure air temperature. The hot-wire measurement principle is used for the AP471 series and the vane principle for the AP472 series. On request, the probes of the AP471 series can be fitted with a telescopic rod that eases measurements in areas difficult to reach (for example vents). The typical applications are wind speed and flow rate checks in air conditioning, heating and cooling systems, or environmental comfort determination, etc. In addition, the measured fluid temperature must be considered: the probes measure air flows at 80°C maximum temperature.
The measurements provided by the instrument using the probes are: wind speed, flow rate, and air temperature. The following units of measurement are available:
for wind speed: m/s - km/h - ft/min - mph (miles/hour) - knots;
for air temperature: °C and °F;
for flow rate: l/s (liters/s) - m
3
/s - m3/min - m3/h - ft3/s - ft3/min
The flow rate measurement requires knowledge of the duct or vent area orthogonal to the flow: the menu item "
SECT" define the section area m
2
or inch2. In order to set this section, you have to
open the Shortcut menu by pressing
SHIFT FNC. The Shortcut menu will appear:
Use the ▲▼ navigation keys to select section and press Enter. The following screen will appear:
HD32.1 - 73 - V2.1
Use the
◄► navigation keys to highlight the digits in the section. Use the ▲▼ navigation keys to
modify the highlighted digit.
The area comprised must be between 0.0001 m2 (1 cm2) and 1.9999 m2.
Pressing
F1 toggles the display between m2 and inch2;
Press
Enter to confirm the information and exit from the section setting.
Press ◄► to select the digit
Press ▲▼ to chan
g
e the digit
HD32.1 - 74 - V2.1
6.2.6 AP471S… Hot-wire wind speed measurement probes complete with SICRAM module
The AP471 S1 and AP471 S3 probes measure incident air flows up to 40m/s. The AP471 S2, AP471 S4 and AP471 S5 probes are fitted with an omni directional sensor allowing measurement of speeds up to 5m/s in any direction of the air flow incident on the probe. The AP471 S4 probe is fitted with support base and sensor protection. The wind speed measurement is temperature compensated within the range 0°C…+80°C. The AP471 S1, AP471 S2 and AP471 S3 probes measure the environment temperature in the range
-25°C…+80°C; the AP471 S4 probe in the range 0°C…+80°C. The AP472 S... modules are calibrated in the factory; no calibration is required by the user.
A
P471 S1
A
P471 S3
The AP471 S1, S2 and S3 probes are fitted with a cylindrical protection screen that can slide longitudinally over a groove. The screen has two end-of-travel positions that block it in measurement condition (completely low) or rest condition (completely high). To reduce the space occupied when not used, the AP471 S4 and AP471 S5 are supplied with a protection cylinder that can be screwed on the probe's head.
AP471 S4
SPEED SENSOR
LINE OF S TREAM DIRECTION
COMPENSATION SENSOR
AMBIENT
SENSOR
TEMPERATURE
CYLINDRICAL SLEEVE
AP471 S1 - AP471 S3
AP471 S2
HD32.1 - 75 - V2.1
Operation
Extend the telescopic rod to the necessary length paying attention to the cable so that it can slide
freely and without strain.
Uncover the sensor and introduce the probe in the air flow being measured, maintaining the arrow at the top of the probe parallel to the flow, as indicated in the figures.
The probe should be maintained orthogonal to the flow and not tilted in relation to it:
OK NO
Proceed with measurement following the instructions provided in this chapter.
Top of the probe
Flow direction indicator
HD32.1 - 76 - V2.1
6.2.7 Technical information on Hot-wire wind speed measurement and temperature probes using SICRAM module
AP471 S1 - AP471 S2 - AP471 S3 - AP471 S4
AP471 S1 - AP471 S3 AP471 S2 AP471 S4
Type of measurements Wind speed, calculated flow rate, air temperature
Type of sensor
Speed NTC thermistor Omni directional NTC
thermistor
Temperature NTC thermistor NTC thermistor
Measurement range
Speed 0.1…40m/s 0.1…5m/s
Temperature -25…+80°C -25…+80°C 0…80°C
Measurement resolution:
Speed
0.01 m/s
0.1 km/h 1 ft/min
0.1 mph
0.1 knot
Temperature 0.1°C
Measurement accuracy
Speed ±0.2 m/s (0.10…0.99 m/s) ±0.05 m/s (0.10…0.99 m/s) ±0.4 m/s (1.00…9.99 m/s) ±0.15 m/s (1.00…5.00 m/s) ±0.8 m/s (10.00…40.00 m/s)
Temperature ±0.8 °C (-10…+80°C) ±0.8 °C (-10…+80 °C)
Minimum speed 0.1 m/s
Air temperature compensation
0…80°C
Care and maintenance of the probes
The speed sensor of the AP471 S… probes is heated and,
in the presence of gas vapors, could trigger a fire or explosion. Do not use the probe in the presence of inflammable gases. Ensure that no gas or explosive vapor leakage is present in the measurement environments.
The probe is very delicate and should be handled with extreme care. Even a simple collision, especially of the omni directional probes that have an uncovered sensor, could render the probe unusable. After measurement, the sensor set on the probe head must be protected with the supplied metallic screen or threaded cylinder. During use, the AP471 S4 omni directional probe must be protected with the special metallic grid. During transportation, the sensor must be closed into a cylinder screwed on the end part of the probe. Do not let fingers touch the sensors. Use only distilled water to clean the probe.
HD32.1 - 77 - V2.1
Dimensions
85 87200 min - 880 max
Ø13
Ø8
AP471 S1
110
Ø13
Ø8
AP471 S2
85 87
102
200 min 880 max
Ø13
Ø8
AP471 S3
Flexible Part
HD32.1 - 78 - V2.1
MIN=350 - MAX=1020
=120
AP471 S4
HD32.1 - 79 - V2.1
6.2.8 AP472S… Vane wind speed measurement probes complete with SICRAM module
The AP472 S1 and AP472 S2 vane probes measure the incident wind speed and flow rate. The probe AP472 S1 measures also the temperature using a thermocouple of type K. On request, they can be fitted with a telescopic rod that eases measurements in areas difficult to reach (for example vents). The probes' speed and temperature measurement ranges are outlined in the table below:
Speed (m/s) Temperature (°C)
Temperature
sensor
Diameter
(mm)
AP472 S1
0.6…25 -25…+80
Thermocouple
K
100
AP472 S2
0.5…20 -25…+80 ---- 60
Greater diameters are suitable for flow measurements in the presence of turbulence with medium­low air speeds (i.e. at the exit of the ducts). Lower diameters are suitable in applications where the probe surface must be much slower than the duct cross section within which the measurement is carried out, i.e. ventilation ducts.
Calibrations
The AP472 S1 and AP472 S2 probes are calibrated in the factory; no calibration is required by the user.
Operation
Where present, extend the telescopic rod to the necessary length
paying attention to the cable so
that it can slide freely and without strain.
Introduce the probe in the air flow being measured, maintaining the arrow at the top of the probe parallel to the flow as indicated in the following figure.
HD32.1 - 80 - V2.1
The probe should be maintained orthogonal to the flow and not tilted in relation to it:
OK NO
The probe is correctly positioned in relation to the air-flow when the value measured is the maximum possible. Proceed with measurement following the instructions provided in this chapter.
Care and maintenance of the probes
The probe performance, mainly at low speeds, largely depends on the very slow friction of the vane on its own axis. In order not to compromise this characteristic, it is recommended that forcing is avoided, as well as blocking or rotating the vane with the fingers, and if possible, avoid inserting it in air flows that could soil the probe.
HD32.1 - 81 - V2.1
Dimensions
AP472 S1 AP472 S2
Unscrew the handle (3) holding the probe body still in the point (1).
HD32.1 - 82 - V2.1
The AP472 S1 - AP472 S2 probes, in addition to the telescopic rod with swivel head can use the rigid telescopic rod 16 mm. Unscrew the handle (3) holding the probe body st ill in the point (1).
Screw the rod end
AP471S1.23.6 (4) on the screw (2). You can add more telescopic rods
AP471S1.23.6. The last element can be the handle (3) or the telescopic rod AP471S1.23.7 (5).
6.2.9 Technical information on Vane wind speed measurement probes using SICRAM module
AP472 S1 AP472 S2
Type of Measurements
Air velocity,
calculated flow,
air temperature
Air velocity,
calculated flow
Diameter 100 mm 60 mm
Type of Measurement
Velocity Vane Vane
Temperature Tc K ----
Measuring Range
Velocity (m/s) 0.6…25 0.5…20
Temperature -25…+80 -25…+80
Resolution
Velocity 0.01 m/s - 0.1 km/h - 1 ft/min - 0.1 mph - 0.1 knots
Temperature 0.1°C ----
Accuracy
Velocity ±(0.4m/s +1.5%f.s.) ±(0.4m/s +1.5%f.s.)
Temperature ±0.8°C ----
Min. Velocity 0.6m/s 0.5m/s
HD32.1 - 83 - V2.1
6.2.10 Light measurement
The instrument works with probes of the LP471… series: These are photometric and radiometric probes that measure
illumination (LP471 PHOT), irradiance (LP471 RAD, LP471 UVA, LP471
UVB and LP471 UVC),
PAR (LP471 PAR) and luminance (LP471 LUM 2). All the probes,
except the LUM 2, are provided with a diffuser for cosine correction.
Upon turning on the instrument automatically detects the probe connected to the input: It is
sufficient to connect it. If the instrument is already on, turn it off and back on again in order for the probe to be detected. The unit of measurement is determined according to the probe connected to the input: In cases where more than one unit of measurement is provided for the same probe, use the UNIT key to select the one desired. All probes are calibrated in the factory; no calibration is required by the user.
The probe is detected during turn on, and this cannot be performed when the instrument is already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it off and on.
HD32.1 - 84 - V2.1
6.2.11 Technical characteristics of photometric and radiometric probes complete with SICRAM module to be connected with the instruments on line
ILLUMINANCE measurement probe LP 471 PHOT complete with SICRAM module and equipped with the instrument
Measurement range (lux): 0.10…199.99 …1999.9 …19999 …199.99⋅103 Resolution (lux): 0.01 0.1 1 0.01⋅103 Spectral range:
in agreement with standard photopic curve V(λ)
Class C Calibration uncertainty:
<4%
f'1 (in agreement with photonic response V(λ)):
<6%
f
2
(response according to the cosine law): <3%
f
3
(linearity): <1%
f
4
(instrument reading error): <0.5%
f
5
(fatigue): <0.5%
f
6
(T) (α temperature coefficient)
<0.05%/K
Drift after 1 year:
<1%
Working temperature:
0…50°C
Reference Standard
CIE No. 69 – UNI 11142
LUMINANCE measurement probe LP 471 LUM 2 complete with SICRAM module and equipped with the instrument
Measurement range (cd/m2):
0.1…1999.9 …19999 …199.9910
3
…1999.9103
Resolution (cd/m2): 0.1 1
0.0110
3
0.1⋅103
Optical angle:
2°
Spectral range:
in agreement with standard photonic curve V(λ)
Class C Calibration uncertainty: <5% f'1 (in agreement with photonic response
V(λ)):
<8%
f
3
(linearity): <1%
f
4
(instrument reading error): <0.5%
f
5
(fatigue): <0.5%
f
6
(T) (α temperature coefficient)
<0.05% K
Drift after 1 year: <1% Working temperature:
0…50°C
Reference Standard CIE No. 69 – UNI 11142
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
380 430 480 530 580 630 680 730 780 830
λ
(nm)
Relative spectral response
HD32.1 - 85 - V2.1
Quantum radiometric probe for the measurement of the photon flow across the chlorophyll range PAR LP 471 PAR complete with SICRAM module and equipped with the instrument
Measurement range (μmol/m2s):
0.01… 199.99 200.0…1999.9 2000…10000
Resolution (μmol/m2s):
0.01 0.1 1
Spectral range: 400nm…700nm Calibration uncertainty: <5% f
2
(response according to the cosine law): <6%
f
3
(linearity): <1%
f
4
(instrument reading error):
±1digit
f5 (fatigue): <0.5% Drift after 1 year:
<1%
Working temperature:
0…50°C
Typical normalized response curve
0,00
0,20
0,40
0,60
0,80
1,00
350 400 450 500 550 600 650 700 750
λ
(nm)
N
ormalized response curve of
LP 471 PAR probe
Theoretical normalized response curve
HD32.1 - 86 - V2.1
IRRADIANCE measurement probe LP 471 RAD complete with SICRAM module and equipped with the instrument
Measurement range (W/m2):
0.110-3… 999.9⋅10-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution (W/m2):
0.110-3
0.001 0.01 0.1
Spectral range: 400nm…1050nm Calibration uncertainty: <5% f
2
(response according to the cosine law): <6%
f
3
(linearity): <1%
f
4
(instrument reading error):
±1digit
f5 (fatigue): <0.5% Drift after 1 year:
<1%
Working temperature:
0…50°C
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
350 450 550 650 750 850 950 1050
λ
(nm)
Relative spectral response
HD32.1 - 87 - V2.1
IRRADIANCE measurement probe LP 471 UVA complete with SICRAM module and equipped with the instrument
Measurement range (W/m2):
110-3… 999.9⋅10-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution (W/m2):
0.110-3
0.001 0.01 0.1
Spectral range: 315nm…400nm (Peak 360nm) Calibration uncertainty: <5% f
2
(response according to the cosine law): <6%
f
3
(linearity): <1%
f
4
(instrument reading error):
±1digit
f5 (fatigue): <0.5% Drift after 1 year:
<2%
Working temperature:
0…50°C
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
280 300 320 340 360 38 0 400 420
λ
(nm)
Relative spectral response
HD32.1 - 88 - V2.1
IRRADIANCE measurement probe LP 471 UVB complete with SICRAM module and equipped with the instrument
Measurement range (W/m2):
110-3… 999.9⋅10-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution (W/m2):
0.110-3
0.001 0.01 0.1
Spectral range: 280nm…315nm (Peak 305nm) Calibration uncertainty: <5% f
2
(response according to the cosine law): <6%
f
3
(linearity): <2%
f
4
(instrument reading error):
±1digit
f5 (fatigue): <0.5% Drift after 1 year:
<2%
Working temperature:
0…50°C
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
250 260 270 280 290 300 310 320 330 340 350
λ
(nm)
Relative spectral response
HD32.1 - 89 - V2.1
IRRADIANCE measurement probe LP 471 UVC complete with SICRAM module and equipped with the instrument
Measurement range (W/m2):
110-3… 999.9⋅10
-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution (W/m2):
0.110-3
0.001 0.01 0.1
Spectral range: 220nm…280nm (Peak 260nm) Calibration uncertainty: <5% f
2
(response according to the cosine law): <6%
f
3
(linearity): <1%
f
4
(instrument reading error):
±1digit
f5 (fatigue): <0.5% Drift after 1 year:
<2%
Working temperature:
0…50°C
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
200 210 220 230 240 250 260 270 280 290 300
λ
(nm)
Relative spectral response
HD32.1 - 90 - V2.1
6.2.12 HD320A2 probe for the measurement of CO Carbon monoxide
HD320A2 probe measures the carbon monoxide concentration in air. It’s a colorless, odorless gas, lighter than the air and it can cause explosions or fires. It is poisonous even in low quantities: indeed, it’s sufficient a concentration of 10-30ppm of carbon monoxide in air to produce symptoms of poisoning and about 2000ppm are fatal in less than 30 minutes. Carbon monoxide is formed when substances containing carbon are burned in absence of oxygen, or when, although the amount of oxygen is sufficient, the combustion occurs at high temperature, e.g. in car engines. Carbon monoxide is one of the major pollutant agents in urban areas. Moreover, being odorless, is an insidious poison. Together with the HD320B2 probe, the HD320A2 probe allows analyzing and monitoring the air quality in internal environments and detecting any loss of CO.
The sensor for the measurement of CO consists of an electro - chemical cell with two electrodes.
CO probe calibration
The probe is calibrated by the company and, usually, doesn’t request any intervention by the user.
However, there is the possibility to make a new calibration that corrects the sensor zero:
in clean air (outside the CO concentration is less than 0,1ppm)
With the help of nitrogen cylinders (code MINICAN.12A).
CO zero calibration in clean air:
1. Place the instrument in an environment with clean air (outside, far from the companies or
the streets, the CO concentration is less than 0.1ppm), switch the instrument on and wait at least 15 minutes till the measurements becomes stable.
2.
Press SHIFT FNC key: the shortcut window appears. With ▲▼ arrows select “cal CO”
and confirm with ENTER.
3.
The screen for the operation to do on the sensor appears (calibration or replacement):
4.
With Up and Down arrows, select “Cal zero” and confirm with ENTER. The screen for the
calibration of CO sensor appears.
HD32.1 - 91 - V2.1
5. Press F2= START to start the calibration:
Next to “CAL ZERO” writing is indicated the CO concentration value measured by the instrument.
6.
During the calibration “Zero CO in progress” message appears. Wait for some minutes to
execute the process without modifying the working conditions.
7.
At the end, the instrument gives an acoustic signal out and visualized “Calibration
completed” message. Press F2=Exit for coming to the measurement.
8.
The process is finished.
Zero CO calibration with nitrogen cylinder (code MINICAN.12A):
1. Switch the instrument on and wait at least 15 minutes till the measurements becomes stable.
2.
Connect the pipe coming from MINICAN.12A cylinder with the rubber cowling on the CO
sensor head.
3.
Press SHIFT FNC key: the shortcut window appears. With ▲▼ arrows select “cal CO”
and confirm with ENTER.
4.
The screen for the operation to execute on the sensor appear (calibration or replacement):
HD32.1 - 92 - V2.1
5. With Up and Down arrows select “Cal zero” and confirm with ENTER. The screen for CO
sensor calibration appears.
6.
Supply the gas adjusting the fluxmeter of the cylinder in order to have a constant fluid
between 0.1 and 0.2 l/min.
7.
Press F2= START to start calibrating:
Next to “CAL ZERO” writing is indicated the CO concentration value measured by the instrument.
8.
During the calibration “Zero CO in progress” message appears. Wait for some minutes to
execute the process without modifying the working conditions.
9.
At the end, the instrument gives an acoustic signal out and visualized “Calibration
completed” message. Close the cylinder tap and remove the CO sensor cowling.
9.
Press F2=Exit for coming back to calibrate.
10.
Insert the protection.
11.
The process is finished.
Replacement of CO sensor:
In normal conditions of use, CO sensor has an average expected life up to 5 years. If it’s necessary to replace CO sensor, proceed as indicated below:
1.
Disconnect the probe from the instrument.
2.
Unscrew the head of the probe and extract the sensor of void CO.
3.
Take note of the number written on the edge of the sensor that indicates the sensibility in
nA/ppm.
4.
Insert the new sensor electrodes into the contacts.
5.
Screw the cap with the probe filter.
6.
Connect the probe and switch the instrument on. Press Shift Fnc key: the shortcut window
appears. With ▲▼ arrows select “Cal CO” and confirm with ENTER.
HD32.1 - 93 - V2.1
7.
The screen for the operation to done on the sensor appears (calibration or replacement):
8.
With Up and Down arrows select “Set sensitivity” and confirm with ENTER. The screen for
the replacement of CO sensor appears.
9.
With Up and Down arrows set the sensor sensibility value. Press ENTER to confirm: the
instrument comes back to the previous screen.
10.
If necessary, calibrate the zero of the CO new sensor.
11.
Press ESC to come back in measurement.
12.
The process is finished.
HD32.1 - 94 - V2.1
6.2.13 HD320B2 probe for the measurement of CO2 carbon dioxide concentration
HD320B2 probe measures the carbon dioxide concentration in air. It’s indicated for checking and monitoring the indoor air quality.
Typical applications are the check of the air quality in all the buildings where there is a crowd of people (schools, hospitals, auditoria, canteens, etc.), in the working places to optimize the comfort.
CO
2
measurement is obtained with an infrared special sensor (NDIR technology: Non-Dispersive Infrared Technology) that, thanks to the use of a double filter and a special measurement technique, warranties precise, stable and long-term measurements. The air to check is spread inside the measurement chamber through the protection membrane placed at the top of the probe.
CO2 probe calibration
The probe is calibrated by the company and usually doesn't request any intervention by the user.
However, there is the possibility to execute a new calibration that corrects the sensor offset:
at 400ppm in clean air
at 0ppm with the help of nitrogen cylinder (code MINICAN.12A).
The instrument can automatically recognize the mode of the started calibration: if 400ppm or 0ppm. The calibration has to be done on one point: each new calibration cancels the previous one.
The concentration of carbon dioxide in air is influenced by different factors: the human activities (companies, pollution, combustion, etc.) cause an increase of this percentage in air. The calibration value is equal to 400ppm and it’s in clean air, for example in the country far from the more polluted areas.
Go on as indicated below:
1. If you want to calibrate around 400ppm, make sure to apply clean air to the instrument
through a membrane placed on the head of the probe.
2.
For the calibration at 0ppm, remove the cap placed at the base of the probe in order to
discover the plug of the calibration gas inlet and connect the tube coming from the nitrogen cylinder. Adjust the fluxmeter of the cylinder on the flow from 0.3 to 0.5l/min.
3.
Switch the instrument on and wait for at least 15 minutes before going on.
4.
Press SHIFT FNC key: the shortcut window appears. With ▲▼ arrows select “cal CO2
and confirm with ENTER.
5.
Supply CO
2
for at least 2 minutes in order that the measurement becomes stable.
6.
Going on supplying CO
2
to the probe, press F2 = CAL CO2 function key: the calibration,
which lasts three minutes, starts. In this phase the instrument measures CO
2
and calibrates itself to a value next to 0ppm if you are using the nitrogen cylinder, at 400ppm if you are calibrating it in clean air.
HD32.1 - 95 - V2.1
7. Wait for three minutes necessary for the calibration without modifying the working
conditions.
8.
If the timer reaches the zero, the instrument gives an acoustic signal out that confirms that
the calibration is finished.
Note: the instrument rejects the calibration values that exceed ±150ppm from the theoretic value.
HD32.1 - 96 - V2.1
7. SERIAL INTERFACE AND USB
The HD32.1 is fitted with an electrically isolated RS-232C serial interface, and an USB 2.0 interface. Optionally, they can be connected using a serial cable with sub D 9-pole female connectors (code
9CPRS232) and a cable with USB 2.0 connectors (code CP22).
The USB connection requires the previous installation of a driver included in the DeltaLog10 software package.
Install the driver before connecting the USB cable to the PC (follow the
instructions included in the software package).
Standard parameters of the instrument RS232 serial transmission are:
Baud rate 38400 baud
Parity None
N. bit 8
Stop bit 1
Protocol Xon/Xoff
It is possible to change the RS232C serial port baud rate by setting the "
Selection of the serial
transmission speed (Baud Rate)" parameter in the menu (please see the menu on chapter 5.3.1 The
Baud Rate). The possible values are: 38400, 19200, 9600, 4800, 2400, 1200. The other transmission parameters are fixed. The USB 2.0 connection does not require the setting of parameters (Baud rate = 460800 fixed).
The selection of the port is carried out directly by the instrument: If the USB port is connected to a PC, the RS232 serial port is automatically disabled, and vice versa.
The instruments are provided with a complete set of commands and data queries to be sent via the PC.
All the commands transferred to the instrument must have the following structure:
XXCR where: XX is the command code and CR is the Carriage Return (ASCII 0D)
The XX command characters are exclusively upper case characters. Once a correct command is entered, the instrument responds with "&"; when any wrong combination of characters is entered, the instrument responds with "?". The instrument response strings end with the sending of the CR command (Carriage Return) and LF (Line Feed). Before sending commands to the instrument via the serial port, locking the keyboard to avoid functioning conflicts is recommended: Use the P0 command. When complete, restore the keyboard with the P1 command.
7.1 THE OPERATING PROGRAM A: MICROCLIMATE ANALYSIS
Command Response Description
P0
&
Ping (locks the instrument keyboard for 70 seconds)
P1
&
Unlocks the instrument keyboard
S0
G0
Model HD32.1 prog.A
Instrument model
G1
M=THERMAL MICROCLIMATE
Model description
G2
SN=12345678
Instrument serial number
G3
Firm.Ver.=01.00
Firmware version
G4
Firm.Date=2005/10/12
Firmware date
G5
cal 0000/00/00 00:00:00
Calibration date and time
C1
Probe 1 type, serial number, calibration date
C2
Probe 2 type, serial number, calibration date
HD32.1 - 97 - V2.1
Command Response Description
C3
Probe 3 type, serial number, calibration date
C4
Probe 4 type, serial number, calibration date
C5
Probe 5 type, serial number, calibration date
C6
Probe 6 type, serial number, calibration date
C7
Probe 7 type, serial number, calibration date
C8
Probe 8 type, serial number, calibration date
GC
Print instrument's heading
GB
User ID=0000000000000000
User code (set with T2xxxxxxxxxxxxxxxx)
H0
Tw= 19.5 øC
Print wet bulb temperature
H1
Tg= 22.0 øC
Print globe thermometer temperature
H2
Ta= 21.6 øC
Print air temperature (dry bulb);
H3
Pr= 1018.1 hPa
Print atmospheric pressure
H4
RH= 50.5 %RH
Print relative humidity
H5
Va= 0.20 m/s
Print wind speed
H6
Tr= 18.5 øC
Print average radiation temperature
H7
WBGT(i)= 23.0 øC
Print indoor WGBT (without solar radiation)
H8
WBGT(o)= 24.0 øC
Print outdoor WGBT (with solar radiation)
H9
WCI=_ERROR_ øC
Print WCI
HA
Print date, time, Tw, Tg, Ta, Pr, RH, Va, Tr, WBGT(i), WBGT(o), WCI
LN
A00 -A01 -B02 -B03 - ..
- .. - .. - .. - .. - ..
- .. - .. - .. - .. - ..
- .. -
Print instrument memory map: If a section is allocated a number is displayed, if it is free 2 points (..) are displayed.
LFn
!Log n.= 0!started on:!2006/01/01 00:37:32
Print memory n section status. The number, the storage start date and time are displayed. (n= hexadecimal number 0-F). If the section is empty:“-->No Log Data<--“
LDn
Print data stored in section n. If the section is empty: “-->No Log Data<--“
LEn
&
Cancel data stored in section n.
LEX
&
Cancel data stored in all sections.
K1
&
Immediate printing of data
K0
&
Stop printing data
K4
&
Start logging data
K5
&
Stop logging data
KP
&
Auto-power-off function=ENABLE
KQ
&
Auto-power-off function=DISABLE
WC0
&
Setting SELF off
WC1
&
Setting SELF on
RA
Sample print = 0sec
Reading of PRINT interval set
RL
Sample log = 30sec
Reading of LOG interval set
WA#
&
Setting PRINT interval. # is a hexadecimal number 0…D that represents the position of the interval in the list 0, 1, 5, 10, …, 3600 seconds.
WL#
&
Setting LOG interval. # is a hexadecimal number 1…D that represents the position of the interval in the list 15, …, 3600 seconds.
HD32.1 - 98 - V2.1
7.2 THE OPERATING PROGRAM B: DISCOMFORT ANALYSIS
Command Response Description
P0
&
Ping (locks the instrument keyboard for 70 seconds)
P1
&
Unlocks the instrument keyboard
S0
G0
Model HD32.1 prog.B
Instrument model
G1
M=THERMAL MICROCLIMATE
Model description
G2
SN=12345678
Instrument serial number
G3
Firm.Ver.=01.00
Firmware version
G4
Firm.Date=2005/10/12
Firmware date
G5
cal 0000/00/00 00:00:00
Calibration date and time
C1
Probe 1 type, serial number, calibration date
C2
Probe 2 type, serial number, calibration date
C3
Probe 3 type, serial number, calibration date
C4
Probe 4 type, serial number, calibration date
C5
Probe 5 type, serial number, calibration date
C6
Probe 6 type, serial number, calibration date
C7
Probe 7 type, serial number, calibration date
C8
Probe 8 type, serial number, calibration date
GC
Print instrument's heading
GB
User ID=0000000000000000
User code (set with T2xxxxxxxxxxxxxxxx)
H0
Th= 19.5 øC
Print temperature at head height
H1
Tb= 22.0 øC
Print temperature at body height
H2
Tn= 21.6 øC
Print temperature of the net radiometer
H3
Tk= 19.5 øC
Print temperature at ankles height
H4
Tf= 19.5 øC
Print temperature of the floor
H5
Pt= 0.0 W/m2
Print power of net radiometer
H6
Dt= 0.0 øC
Print asymmetrical radiant temperature of the net radiometer
HA
Print date, time, Th, Tb, Tn,Tk, Tf, Pt, Dt
LN
A00 -A01 -B02 -B03 - ..
- .. - .. - .. - .. - ..
- .. - .. - .. - .. - ..
- .. -
Print instrument memory map: If a section is allocated a number is displayed, if it is free 2 points (..) are displayed.
LFn
!Log n.= 0!started on:!2006/01/01 00:37:32
Print memory n section status. The number, and storage start date and time are displayed. (n= hexadecimal number 0-F). If the section is empty:“-->No Log Data<--“
LDn
Print data stored in section n. If the section is empty: “-->No Log Data<--“
LEn
&
Cancel data stored in section n.
LEX
&
Cancel data stored in all sections.
K1
&
Immediate printing of data
K0
&
Stop printing data
K4
&
Start logging data
K5
&
Stop logging data
KP
&
Auto-power-off function=ENABLE
KQ
&
Auto-power-off function=DISABLE
WC0
&
Setting SELF off
WC1
&
Setting SELF on
RA
Sample print = 0sec
Reading of PRINT interval set
HD32.1 - 99 - V2.1
Command Response Description
RL
Sample log = 30sec
Reading of LOG interval set
WA#
&
Setting PRINT interval. # is a hexadecimal number 0…D that represents the position of the interval in the list 0, 1, 5, 10, …, 3600 seconds.
WL#
&
Setting LOG interval. # is a hexadecimal number 1…D that represents the position of the interval in the list 15, …, 3600 seconds.
7.3 THE OPERATING PROGRAM C: PHYSICAL QUANTITIES
Command Response Description
P0
&
Ping (locks the instrument keyboard for 70 seconds)
P1
&
Unlocks the instrument keyboard
S0
G0
Model HD32.1 prog.C
Instrument model
G1
M=THERMAL MICROCLIMATE
Model description
G2
SN=12345678
Instrument serial number
G3
Firm.Ver.=01.00
Firmware version
G4
Firm.Date=2005/10/12
Firmware date
G5
cal 0000/00/00 00:00:00
Calibration date and time
C1
Probe 1 type, serial number, calibration date
C2
Probe 2 type, serial number, calibration date
C3
Probe 3 type, serial number, calibration date
C4
Probe 4 type, serial number, calibration date
C5
Probe 5 type, serial number, calibration date
C6
Probe 6 type, serial number, calibration date
C7
Probe 7 type, serial number, calibration date
C8
Probe 8 type, serial number, calibration date
GC
Print instrument's heading
GB
User ID=0000000000000000
User code (set with T2xxxxxxxxxxxxxxxx)
H0
Tpt= 19.5 øC
Print Pt100 temperature
H1
RH= 50.0 %
Print %RH
H2
Trh= 21.6 øC
Print temperature of the RH probe
H3
Va= 0.25 m/s
Print air speed
H4
Fl= 1.5 l/s
Print air flux of air speed pr obe
H5
Tv= 20.5 øC
Print temperature of air speed probe
H6
Lux= 550.0 lux
Print lux
HA
Print date, time, Tpt, RH, Trh, Va, Fl, Tv, Lux1, Lux2, CO
2
, CO
LN
A00 -A01 -B02 -B03 - ..
- .. - .. - .. - .. - ..
- .. - .. - .. - .. - ..
- .. -
Print instrument memory map: If a section is allocated a number is displayed, if it is free 2 points (..) are displayed.
LFn
!Log n.= 0!started on:!2006/01/01 00:37:32
Print memory n section status. The number, and storage start date and time are displayed. (n= hexadecimal number 0-F). If the section is empty:“-->No Log Data<--“
LDn
Print data stored in section n. If the section is empty: “-->No Log Data<--“
LEn
&
Cancel data stored in section n.
HD32.1 - 100 - V2.1
Command Response Description
LEX
&
Cancel data stored in all sections.
K1
&
Immediate printing of data
K0
&
Stop printing data
K4
&
Start logging data
K5
&
Stop logging data
KP
&
Auto-power-off function=ENABLE
KQ
&
Auto-power-off function=DISABLE
WC0
&
Setting SELF off
WC1
&
Setting SELF on
RA
Sample print = 0sec
Reading of PRINT interval set
RL
Sample log = 30sec
Reading of LOG interval set
WA#
&
Setting PRINT interval. # is a hexadecimal number 0…D that represents the position of the interval in the list 0, 1, 5, 10, …, 3600 seconds.
WL#
&
Setting LOG interval. # is a hexadecimal number 1…D that represents the position of the interval in the list 15, …, 3600 seconds.
Loading...