COMMUNICATIONS & POWER INDUSTRIES CANADA INC.
45 RIVER DRIVE
GEORGETOWN, ONTARIO, CANADA L7G 2J4
TELEPHONE: (905) 877-0161
SERVICE MANUAL P/N 740986-00
CPI Canada Inc.
PRE-INSTALLATION
CMP 200
INSTALLATION
SERVICE AND
INSTALLATION
INTERFACING, PROGRAMMING, AND
CALIBRATION
1¾
2¾
3¾
MANUAL
P/N 740986-00
TROUBLESHOOTING
ACCEPTANCE TESTING
4¾
5¾
REGULAR MAINTENANCE
THEORY OF OPERATION
SPARES
SCHEMATICS
The original version of this manual (June 17, 2001)
has been drafted in the English language by:
Communications & Power Industries
communications & medical products division.
6¾
7¾
8¾
9¾
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-00 Rev. B Page 0-1
CPI Canada Inc.
PROPRIETARY AND CONFIDENTIAL
The design and information contained in this document is proprietary to Communications & Power Industries Canada
Inc. (CPI), has been designed and developed at private expense, and is the exclusive property of CPI. The
information contained herein is loaned in confidence solely for the installation, operation, and maintenance of CPI
equipment. This document may not be duplicated in whole or in part, or be used for re-engineering, reverse
engineering, or otherwise reproducing in any form or creating or attempting to create or permitting, allowing or
assisting others to create or manufacture CPI’s product or products derived therefrom, without the express written
permission of CPI.
1.1.2 Features ....................................................................................................................................................1-2
1.2.1 Safety and Warning Symbols....................................................................................................................1-8
1.2.2 Safety Notices and Warnings....................................................................................................................1-8
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-1
1 Pre-installation CPI Canada Inc.
1.1.0 INTRODUCTION
This chapter summarizes the main features of the CMP 200 X-ray generator (performance, regulatory
and compatibility). Safety information is provided, along with environmental, room, and installation
requirements. This chapter concludes with a pre-installation checklist and a diagram showing the major
component layout.
The information in this chapter is provided in order for the installer to be able to plan the site
layout prior to installation of the generator.
1.1.1 Generator Description
The CMP 200 X-ray generator is intended for use in stationary radiographic X-ray systems, including
tomography. The X-ray generator consists of a main power cabinet, and a control console. The main
power cabinet contains the HT tank and control circuits, the filament drivers, the low speed starter, and
interface connections to the room equipment.
The control console allows the operator to select the technique factors, image receptors, etc.,
and to initiate an X-ray exposure.
1.1.2 Features
The following are the main features of and the options available for the generator:
• Integral low speed starter, compatible with X-ray tubes with type “R” stator. Optional compatibility with
GE 23/23 Ω equal impedance “E” stator.
• 24 VDC, 110, or 220 VAC power source for Buckys.
• 24 VAC 150 watts power source for collimator lamp.
• 24 VDC 45 watts power source for system locks.
• Optional AEC.
1.1.3 Radiographic Performance
kVp range: 40 to 125 kV or 40 to 150 kV,
depending on model.
kVp steps: variable in 1 kV steps.
kVp accuracy:
Risetime (10-90%): <1.5 ms.
Time range: 1.0 to 6300 milliseconds.
mAs range 0.1 to 500 mAs (30/32/40 kW), 0.1 to
mAs accuracy: ± (10 % + 0.2) mAs.
mA range 10 to 400 mA (30/32 kW), 10 to 500
Coefficient of linearity: 0.05 (station to station) mAs.
Coefficient of reproducibility:
± 5 %.
630 mAs (50 kW).
mA (40 kW), 10 to 630 mA (50 kW).
kV, mAs ≤ 0.05.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-2 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.1.4 Environmental Specifications
OPERATING
Ambient temperature range
10 to 40 °C (50 to 104 °F).
Relative humidity 20 to 80%, non-condensing.
Altitude Sea-level to 2440 meters (700 to 1100 hPa).
TRANSPORT AND STORAGE
Ambient temperature range
-25 to 70 °C (-13 to 158 °F)..
Relative humidity 5 to 95%, non-condensing.
Atmospheric pressure range 500 to 1060 hPa (375 to 795 mm Hg).
1.1.5 Applicable Standards
The CMP 200 series of X-ray generators comply with the regulatory requirements and design standards in
this section as follows:
• VZW2555 series: Only the standards marked with an asterisk
* under SAFETY.
• VZW2556 series: All standards in this section.
A) SAFETY
* FDA Center for Devices & Radiological Health (CDRH) - 21 CFR title 21 subchapter J (USA).
•
* Radiation Emitting Devices Act - C34 (Canada).
•
• Medical Device Regulations (Canada).
• EC Directive 93/42/EEC concerning Medical Devices (European Community).
* EN 60601-1/IEC 60601-1, EN 60601-2-7/IEC 60601-2-7, CSA 601.1, UL60601.1.
•
-Type of protection against electric shock: Class I equipment.
-Degree of protection against electric shock: Not classified.
-Degree of protection against harmful ingress of water: Ordinary equipment.
-Mode of operation: Continuous operation with intermittent loading (standby - exposure).
-Equipment not suitable for use in presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
• EN 60601-1-4/IEC 60601-1-4, EN ISO 14971.
NOTE: All referenced standards are considered to be at the latest revision.
The CE Mark is a declaration by the manufacturer that the product complies with
the requirements of the applicable European Union (EU) medical device directive
and that the product has been subject to conformity assessment procedures as
provided in that directive.
A CSA mark with the indicators “C” and “US” means that product is certified for
both the U.S. and Canadian markets, to the applicable U.S. and Canadian
standards.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-3
1 Pre-installation CPI Canada Inc.
B) EMC (EN 60601-1-2:2001/IEC 60601-1-2:2001)
Guidance and manufacturer’s declaration – electromagnetic emissions
The VZW2556 series of X-ray generators is intended for use in the electromagnetic environment specified
below. The customer or the user of the VZW2556 series should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
NOTE It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be
verified to assure that they meet the minimum specifications.
Group 1 The VZW2556 series of X-ray generators use RF energy only for
their internal functions. Therefore, the RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
Class A
(The VZW2556
series of X-ray
generators in
combination with
shielded location)
Not Applicable
Not Applicable
The VZW2556 series of X-ray generators must be used only in a
shielded location with a minimum RF shielding effectiveness and,
for each cable that exits the shielded location, a minimum RF filter
attenuation of 40dB from 30 MHz to 230 MHz and 47dB from 230
MHz to 1 GHz. (The minimum at 30 MHz is 40dB and the
minimum at 230 MHz is 47dB).
The VZW2556 series is suitable for use in all establishments other
than domestic and those directly connected to the public lowvoltage power supply network that supplies buildings used for
domestic purposes.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-4 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.1.5 Applicable Standards (Cont)
Guidance and manufacturer’s declaration – electromagnetic immunity
The VZW2556 series of X-ray generators is intended for use in the electromagnetic environment specified
below. The customer or the user of the VZW2556 series should assure that it is used in such an environment.
Immunity
test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruption,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
IEC 61000-4-8
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV differential
mode
± 2 kV common
mode
< 5 % U
(> 95 % dip in UT)
for 0.5 cycle
40 % U
(60 % dip in UT)
for 5 cycles
70 % U
(30 % dip in UT)
< 5 % U
(> 95 % dip in UT)
for 5 s
T
T
T
T
Compliance
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV differential
mode
± 2 kV common
mode
< 5 % U
(> 95 % dip in UT)
for 0.5 cycle
40 % U
(60 % dip in UT)
for 5 cycles
70 % U
(30 % dip in UT)
< 5 % U
(> 95 % dip in UT)
T
T
T
T
for 5 s
Electromagnetic environment –
guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the VZW2556 series X-ray generator requires
continued operation during power mains
interruptions, it is recommended that the X-ray
generator be powered from an uninterruptible
power supply or battery.
3 A/m 3 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment
NOTE: UT is the A.C. mains voltage prior to application of the test level.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-5
1 Pre-installation CPI Canada Inc.
1.1.5 Applicable Standards (Cont)
Guidance and manufacturer’s declaration – electromagnetic immunity
The VZW2556 series of X-ray generators is intended for use in the electromagnetic environment specified
below. The customer or the user of the VZW2556 series should assure that it is used in such an environment.
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
test level
3 V
rms
150 kHz to
80MHz
3 V/m
80MHz to 2.5
GHz
Compliance
level
3 V
rms
150 kHz to
80MHz
3 V/m
80MHz to 2.5
GHz
Electromagnetic environment - guidance
The VZW2556 series of X-ray generators must be
used only in a shielded location with a minimum RF
shielding effectiveness and, for each cable that
enters the shielded location, a minimum RF filter
attenuation of 40dB from 30 MHz to 230 MHz and
47dB from 230 MHz to 1 GHz. (The minimum at 30
MHz is 40dB and the minimum at 230 MHz is
47dB.)
Field strengths outside the shielded location from
fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than 3
a
V/m.
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 These guidelines may not apply all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
NOTE 2 It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be
verified to assure that they meet the minimum specification.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
VZW2556 series of X-ray generators is used exceeds the applicable RF compliance level above, the X-ray
generator should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the X-ray generator.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-6 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.1.6 Electromagnetic Compatibility (EMC)
In accordance with the intended use, some models of this series of X-ray generators comply with the
European Council Directive concerning Medical Devices. The CE marking affixed to compliant products
signifies this. One of the harmonized standards of this Directive defines the permitted levels of
electromagnetic emission from this equipment and its required immunity from the electromagnetic
emissions of other devices.
It is not possible, however, to exclude with absolute certainty the possibility that other high frequency
electronic equipment, which is fully compliant to the EMC regulations, will not adversely affect the
operation of this generator. If the other equipment has a comparatively high level of transmission power
and is in close proximity to the generator, these EMC concerns (the risk of interference) may be more
pronounced. It is therefore recommended that the operation of equipment of this type such as mobile
telephones, cordless microphones and other similar mobile radio equipment be restricted from the vicinity
of this X-ray generator.
1.1.7 RoHS Compliance
CMP 200
产品中有毒有害物质或元素的名称及含量
Table of hazardous substances’ name and concentration.
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for
this part is below the limit requirement in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogeneous
materials used for this part is above the limit requirement in SJ/T11363-2006.
• Data listed in the table represents best information available at the time of publication
• Applications of hazardous substances in this medical device are required to achieve its intended
clinical uses, and/or to provide better protection to human beings and/or to environment, due to
lack of reasonably (economically or technically) available substitutes.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-7
1 Pre-installation CPI Canada Inc.
1.2.0 SAFETY
1.2.1 Safety and Warning Symbols
The following advisory symbols are used on the safety warning labels, and/or on circuit boards, and/or on
the operator console.
WARNING THIS X-RAY UNIT MAY BE
DANGEROUS TO PATIENT
AND OPERATOR UNLESS
SAFE EXPOSURE
FACTORS, OPERATING
INSTRUCTIONS AND
MAINTENANCE SCHEDULES ARE
OBSERVED.
1.2.2 Safety Notices and Warnings
High voltage symbol used to indicate the presence of
high voltage.
Warning symbol used to indicate a potential hazard
to operators, service personnel or to the equipment.
It indicates a requirement to refer to the
accompanying documentation for details.
Radiation exposure symbol used on operator
console. Lights to indicate that an exposure is in
progress. This is accompanied by an audible tone
from the console.
Radiation warning label on operator console.
Never allow unqualified personnel to operate the X-
ray generator.
WARNING:
WARNING:
THE MANUFACTURER ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR
THE MANUFACTURER ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-8 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS
SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.
PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY
GENERATORS ARE THE RESPONSIBILITY OF USERS OF SUCH GENERATORS.
CPI CANADA INC. PROVIDES INFORMATION ON ITS PRODUCTS AND
ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTERSALE OPERATING AND SAFETY PRACTICES.
NOT MAINTAINED OR SERVICED ACCORDING TO THIS SERVICE AND
INSTALLATION MANUAL, OR FOR ANY GENERATOR THAT HAS BEEN MODIFIED
IN ANY WAY.
RADIATION OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM
POOR OPERATING TECHNIQUES OR PROCEDURES.
CPI Canada Inc. Pre-installation 1
1.2.2 Safety Notices and Warnings (Cont)
X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending
over periods of many months or even years. Operators and service personnel should avoid any exposure to the primary beam and take protective measures to safeguard against scatter radiation.
Scatter radiation is caused by any object in the path of the primary beam and may be of equal or less
intensity than the primary beam that exposes the film.
No practical design can incorporate complete protection for operators or service personnel who do not
take adequate safety precautions. Only authorized and properly trained service and operating personnel should be allowed to work with this X-ray generator equipment. The appropriate
personnel must be made aware of the inherent dangers associated with the servicing of high voltage
equipment and the danger of excessive exposure to X-ray radiation during system operation.
DO NOT CONNECT UNAPPROVED EQUIPMENT TO THE REAR OF THE
CONSOLE.
For the membrane console, J3 is for connection of an external hand switch, J4 is a
serial port for use by an external computer, and J8 is for the interconnect cable to the
main cabinet.
For the touch screen console, J2 on the touch screen interface board is for the
interconnect cable to the generator, J3 is for connection of an external hand switch, J5
connects to the membrane switch assembly with the on / off and prep / expose
switches, and J4 is a serial port for use by an external computer.
INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY
RESULT IN INJURY OR EQUIPMENT DAMAGE.
CAUTION
: DO NOT EXCEED THE TUBE MAXIMUM OPERATING LIMITS. INTENDED LIFE AND
RELIABILITY WILL NOT BE OBTAINED UNLESS GENERATORS ARE OPERATED
WITHIN PUBLISHED SPECIFICATIONS.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-9
1 Pre-installation CPI Canada Inc.
1.2.2 Safety Notices and Warnings (Cont)
WARNING
THE CONSOLE ON/OFF SWITCH DOES NOT DISCONNECT THE MAIN POWER
THE DC BUS CAPACITORS, LOCATED IN THE MAIN CABINET PRESENT A
AN LED CONNECTED ACROSS THE DC BUS INDICATES THE PRESENCE OF HIGH
: HAZARDOUS VOLTAGES EXIST INSIDE THE GENERATOR WHENEVER THE MAIN
POWER DISCONNECT IS SWITCHED ON. THESE AREAS INCLUDE, BUT ARE NOT
LIMITED TO, THE MAIN FUSE HOLDER AND ASSOCIATED CIRCUITS ON THE H.V.
AUXILIARY BOARD, THE AUXILIARY AND / OR TOUCHSCREEN TRANSFORMERS,
AND THE MAIN POWER CONTACTOR.
LED DS1 ON THE H.V. AUXILIARY BOARD INDICATES THE PRESENCE OF THE
+24 VDC SUPPLY.
FROM THE ABOVE AREAS INSIDE THE GENERATOR.
SAFETY HAZARD FOR AT LEAST 5 MINUTES AFTER THE POWER HAS BEEN
REMOVED FROM THE UNIT. CHECK THAT THESE CAPACITORS ARE
DISCHARGED BEFORE SERVICING THE GENERATOR.
VOLTAGE. THIS LED IS MOUNTED ON THE EMC CAPACITOR BOARD (ON SOME
MODELS, THE EMC CAPACITOR BOARD DOES NOT CONTAIN ANY
COMPONENTS OTHER THAN THE LED AND THE SERIES RESISTORS).
DO NOT RELY SOLELY ON BLEEDER CIRCUITS AND HIGH-VOLTAGE ON
INDICATORS IN THE GENERATOR TO PROTECT YOU. DUE TO THE
POSSIBILITY OF COMPONENT FAILURE, IT MUST NEVER BE ASSUMED
THAT AN UNLIT LED ENSURES THAT NO HIGH VOLTAGE IS PRESENT.
USING A VOLTMETER, CONFIRM THAT NO HIGH VOLTAGE IS PRESENT
BEFORE ATTEMPTING ANY SERVICE.
1.2.3 Safety Warning Labels
This subsection defines the safety labels used inside and outside the generator cabinet.
NOTE:
WARNING:
WARNING:
This information is provided to help you establish safe operating conditions for both you and your X-ray
generator. Do not operate this X-ray generator except in accordance with these instructions, and any
additional information provided by the X-ray generator manufacturer and / or competent safety
authorities.
THESE LABELS AND WARNINGS ARE PROVIDED TO ALERT SERVICE
PERSONNEL THAT SERIOUS INJURY WILL RESULT IF THE HAZARD IDENTIFIED
IS IGNORED.
SWITCH OFF THE MAIN POWER DISCONNECT AND ALLOW SUFFICIENT TIME
FOR ALL CAPACITORS TO DISCHARGE BEFORE REMOVING ANY COVERS.
IF ANY COVERS MUST BE REMOVED FOR SERVICE, TAKE ALL REQUIRED
PRECAUTIONS WITH RESPECT TO THE HAZARD(S) AND IMMEDIATELY
REPLACE THE COVERS WHEN THE NEED FOR REMOVAL IS COMPLETED.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-10 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.2.3 Safety Warning Labels (Cont)
Weight Label
This label is attached to the main generator cabinet and to the HT tank, and states the approximate
weight of the main cabinet with the HT tank, and the weight of the HT tank separately. Do not attempt to
lift these items without proper assistance.
Caution HV/High Energy Warning Label
This label is attached to the generator cabinet and on the inside of the back cover above the HT tank.
The DC bus capacitors (approximately 300 to 670 VDC, depending on model) will remain charged for up
to 5 minutes after the AC mains is disconnected or the console is switched off.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-11
This label is attached to the outside of the generator cabinet, to the cover over the inverter assembly, and
to the cover in front of the touchscreen fusing board. Mains voltage is present inside the cabinet
whenever the main disconnect is switched on. Additionally, the DC bus capacitors will remain charged for
up to 5 minutes after the AC mains is disconnected or the console is switched off.
WARNING:
HT Tank - Transformer Terminals Notice
This notice is printed on the HT tank lid and cautions against over-tightening the nuts on the transformer
feedthrough terminals (for the primary of the HT transformers).
WAIT A MINIMUM OF 5 MINUTES AFTER THE INPUT MAINS POWER HAS BEEN
REMOVED BEFORE REMOVING ANY COVERS. ONCE THE COVER(S) ARE
REMOVED. CHECK THAT THE VOLTAGE ACROSS THE DC BUS CAPACITORS IS
NEAR ZERO BEFORE SERVICING. IF THIS VOLTAGE EXCEEDS 48 VDC, THE
CAPACITORS MUST BE MANUALLY DISCHARGED BY QUALIFIED SERVICE
PERSONNEL.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-12 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.2.3 Safety Warning Labels (Cont)
Danger High Tension Notice
This notice is printed on the HT tank lid. High voltage may be present at the primary terminals on the tank
lid board, at the output high voltage connectors, and at the mA/mAs measuring jacks if the shorting link is
opened for mA/mAs measurements.
HIGH VOLTAGE HAZARD: Be certain that you are aware of all potential high voltage
locations and hazards as detailed in this section before removing any covers, or attempting
any service on this X-ray generator.
FUSE RATINGS:
LOCATION
MAIN LINE FUSES: F1, F2: FRN-R-50 * F1, F2, F3: AG40 *
* Refer to chapter 8, spares list, for CPI part number for these fuses.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-13
1 Pre-installation CPI Canada Inc.
1.2.3 Safety Warning Labels (Cont)
HIGH VOLTAGE HAZARD: Approximately 400 VAC is present on the console board in the
area of T1, C36, and J5. This is a high voltage source for the fluorescent backlight on the LCD
display.
HIGH VOLTAGE HAZARD: AC mains voltage and / or DC bus voltage (approximately 325 to
670 VDC, depending on model) is present on the H.V. auxiliary board whenever the AC mains
is energized. Ensure that the AC mains is switched off and locked out before servicing this
board. See the note below regarding the DC bus voltage.
HIGH VOLTAGE HAZARD: High voltage is present on all components connected to the AC
mains (line fuses, main power contactor, H.V. auxiliary board, auxiliary and / or touchscreen
transformer, etc) whenever the AC mains is switched on. Additionally, DC bus voltage is
present on certain components (mains rectifier assembly, DC bus capacitors, inverter
assembly, HT tank, H.V. auxiliary board, etc) whenever the generator is switched on, and will
remain on for up to 5 minutes after the console is switched off or the AC mains is switched off
or disconnected.
High voltage (approximately 325 to 670 VDC, depending on model) is present on the inverter
assembly and associated components whenever the AC mains is energized and the console
is switched on, and for up to 5 minutes after the console is switched off or the AC mains is
disconnected. THIS COMBINATION OF HIGH VOLTAGE AND HIGH CURRENT IS
POTENTIALLY LETHAL. USE EXTREME CAUTION WHEN SERVICING THIS UNIT.
DANGEROUS
VOLTAGE
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-14 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.3.0 PREPARING FOR INSTALLATION
1.3.1 Generator Heat Output
The maximum heat output of the main generator cabinet is less than 1000 BTU / hour in normal clinical
use, with a maximum of 70 BTU / hour heat output for the console. The console is primarily conduction
cooled, and the main cabinet is fan cooled. The console and main cabinet should never be covered when
the generator is switched on, as any covering may interfere with the cooling.
1.3.2 Generator Power Requirements
The tables in this section show mains power requirements for various configurations of CMP 200 X-ray
generators. The installer must ensure that the generator is connected to the proper mains voltage as per
the nameplate on the generator.
The table below depicts the power requirements for the 30 kW CMP 200 X-ray generators.
Line Voltage 208 VAC - 5% to 230 VAC + 10%, 1 phase.
Line Frequency 50/60 Hz.
Momentary Current 195 Amps at 230 VAC.
Nominal Current
Momentary Power Consumption 43 kVA.
The table below depicts the power requirements for the 32 kW CMP 200 X-ray generators.
*
≤5 Amps.
Line Voltage 208 VAC - 5% to 230 VAC + 10%, 3 phase.
400 VAC ± 10%, 3 phase.
480 VAC ± 10%, 3 phase.
Line Frequency 50/60 Hz.
Momentary Current 112 Amps/phase at 230 VAC.
65 Amps/phase at 400 VAC.
Nominal Current
Momentary Power Consumption 45 kVA.
*
≤5 Amps.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-15
1 Pre-installation CPI Canada Inc.
1.3.2 Generator Power Requirements (Cont)
The table below depicts the power requirements for the 40 kW CMP 200 X-ray generators.
Line Voltage 208 VAC - 5% to 230 VAC + 10%, 1 phase.
480 VAC ± 10%, 3 phase.
Line Frequency 50/60 Hz.
Momentary Current 250 Amps at 230 VAC (1 phase).
140 Amps/phase at 230 VAC (3 phase).
80 Amps/phase at 400 VAC.
65 Amps/phase at 480 VAC.
Nominal Current
Momentary Power Consumption 55 kVA.
The table below depicts the power requirements for the 50 kW CMP 200 X-ray generators.
*
≤5 Amps.
Line Voltage 208 VAC - 5% to 230 VAC + 10%, 3 phase.
400 VAC ± 10%, 3 phase.
480 VAC ± 10%, 3 phase.
Line Frequency 50/60 Hz.
Momentary Current 175 Amps/phase at 230 VAC.
100 Amps/phase at 400 VAC.
80 Amps/phase at 480 VAC.
Nominal Current
Momentary Power Consumption 69 kVA.
*
≤5 Amps.
*Nominal Current = Generator standby current only. External or installer-supplied equipment
connected to the generator may increase the nominal current beyond the values shown.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-16 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.3.2 Generator Power Requirements (Cont)
The following table defines the power line requirements for the generators.
: THE FOLLOWING TABLE CONTAINS RECOMMENDED VALUES FOR THE WIRE SIZES BETWEEN
NOTE
THE MAINS DISCONNECT AND THE GENERATOR. THE ACTUAL VALUES USED AT AN
INSTALLATION ARE DEPENDENT ON THE QUALITY OF THE INPUT LINE (VOLTAGE LEVEL), THE
CURRENT REQUIREMENTS, AND THE LENGTH OF THE CABLE RUN, AND MUST BE CONFIRMED
BY THE INSTALLER.
FINAL SELECTION OF GENERATOR INPUT WIRE AND DISCONNECTS AS WELL AS THE
CABLING FROM THE DISTRIBUTION TRANSFORMER TO THE MAINS DISCONNECT MUST MEET
THE REQUIREMENTS OF THE LOCAL ELECTRICAL CODES, AND IS USUALLY DETERMINED BY
HOSPITAL / CONTRACTOR ENGINEERING.
THE RATINGS LISTED CONSIDER THE GENERATOR REQUIREMENTS ONLY. THE INSTALLER
MUST MAKE THE NECESSARY COMPENSATION FOR ADDITIONAL LOAD REQUIREMENTS.
A POOR QUALITY INPUT LINE MAY RESULT IN THE INSTALLER HAVING TO
* Refer to 1.3.3 for general grounding information.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-17
1 Pre-installation CPI Canada Inc.
1.3.2 Generator Power Requirements (Cont)
Recommended Service Disconnect: As per the above table
• All wiring and grounding should comply with the national electrical code or equivalent.
• All wiring must be copper.
• The disconnect switch shall be located within reach of the operator.
1.3.3 Generator Ground Requirements
• A suitable ground must be connected from the disconnect switch to the main ground of the generator,
located to the right of the main fuse block. The ground wire is typically part of the line cord, and the
current capacity of the ground conductor must normally be equal to or greater than that of the line
conductors.
2
• A copper ground cable, #10 AWG (6 mm
to the H.T. tank ground stud (located at the top of the HT tank).
• If a neutral line is provided with the system, under no circumstances is it to be used for ground
purposes. The ground must carry fault currents only.
) or larger should be connected from the X-ray tube housing
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-18 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.3.4 Locating and Mounting the Generator
The main generator cabinet is self-standing and does not need to be supported. However, the installation
should meet the following requirements:
• The floor must be flat and level.
• The generator installation area must be clean and free of dirt or debris.
• The installer must supply generator hold-down brackets, if required. Alternately, mounting holes may
be drilled in the base of the generator. The generator may then be anchored to the floor via these
holes.
• Sufficient room must be provided to allow access to the rear and side of the generator for installation
and service. See figure 1-1 for recommended clearances.
• A cable trough, conduit, or raceway (1 in; 2.5 mm, diameter) should be provided from the control
console to the main cabinet to allow routing of the control cable if required.
• The control console is normally freestanding on a desk or shelf. It may be anchored if necessary.
Figure 1-1: Generator clearances
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-19
1 Pre-installation CPI Canada Inc.
1.3.5 Cable Entrance and Seismic Center Location
Figure 1-2 shows the locations of the cable access slots, the AC mains cable entry, and the seismic
center location for the CMP 200 X-ray generator.
Figure 1-2: CMP 200 cable entry locations and seismic center
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-20 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.3.6 Tools and test Equipment Required
The following is a checklist of recommended tools and test equipment for installation and calibration of
the generator.
CHECK √
DESCRIPTION
General hand tools for installation: Wrenches, nut drivers, assortment of
screwdrivers, pliers, etc.
If the generator is to be anchored to the floor, suitable tools (i.e. drill, drill bits,
etc) and mounting hardware must be available.
A supply of connectors for wiring: terminal lugs, caps, splices etc.
A calibrated DVM that indicates true RMS voltages.
Dual trace memory oscilloscope with a minimum 20 MHz bandwidth;
appropriate leads, probes, etc.
Device for measuring true kVp. This may be a Dynalyzer equivalent or a non-
invasive meter such as the Keithley TRIAD system.
A calibrated radiation meter with detectors that will allow for R/min and uR
type measurements (or uGy and Gy/min).
A suitable mA / mAs meter.
A strobe or reed type tachometer to verify that the anode is rotating up to
speed.
A sufficient selection of absorbers to allow AEC calibration if this option is
fitted. A suggested selection is Lexan in thickness of 5.0, 10.0, and 15.0 cm,
or water in plastic containers of homogenous density in thickness of 5.0, 10.0,
and 15.0 cm.
Vapor proof compound for the HT terminations.
1.3.7 Pre-Installation Checklist
Before starting the generator installation, review the following checklist.
CHECK √
DESCRIPTION
Is there an unloading area to transport the generator from the delivery truck to
the inside of the building?
If the installation is not on the same floor as the delivery entrance, is there an
elevator available?
Is there a transport dolly or similar device to move the generator?
Do any regulatory bodies need to be notified prior to installation?
If movers are required, have arrangements for time and equipment been
completed?
Are lifting straps or some other suitable device available to lift the generator off
the shipping pallet?
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-21
1 Pre-installation CPI Canada Inc.
1.4.0 GENERATOR LAYOUT AND MAJOR COMPONENTS
The dimensions and weight of the generator are shown in the table below.
ITEM LENGTH WIDTH HEIGHT WEIGHT
Main cabinet in shipping pack 30.5 (775) 21.5 (546) 38 (965) 129 (59)
Main cabinet unpacked 22.8 (580) 13.7 (348) 24.3 (617) 113 (51)
Control console 12.3 (313) 10.9 (277) 3.7 (94) 6 (2.72)
The above dimensions are inches (mm), weights are in pounds (Kg).
Figures 1-3 and 1-4 show the major components located inside the generator cabinet. Figure 1-5 is an
internal view of the console, showing the major components and cabling. Figure 1-3 does not show the
fan-mounting bracket, nor does it represent all models. This is meant to show major component layout
only.
Figure 1-3: Major generator subassemblies view 1
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-22 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
R
1.4.0 GENERATOR LAYOUT AND MAJOR COMPONENTS (Cont)
H.V . AUXILIAR Y
BOARD
GENERATOR
CONTROL BOARD
U6
AEC BOARD
EPROM
TOUCHSCREEN
BATTERY
TRANSFORME
(OPTIONAL)
FILAMENT
BOARD
“” POSITION
JW2
JW2
LOAD DEFAULTS
“” POSITION
NORMAL
CMP200_007A.CDR
Figure 1-4: Major generator subassemblies view 2
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-23
1 Pre-installation CPI Canada Inc.
1.4.0 GENERATOR LAYOUT AND MAJOR COMPONENTS (Cont)
Figure 1-5: Console internal view including EPROM location
1.5.0 COMPATIBILITY LISTING
This X-ray generator is compatible with the following equipment:
X-RAY TUBES:
Refer to the compatibility statement at the end of this chapter.
: THE LOW SPEED STARTER IS COMPATIBLE WITH “R” TYPE STATORS OR GE 23/23 Ω
NOTE
EQUAL IMPEDANCE STATORS. REFER TO THE COMPATIBILITY STATEMENT OR THE SECTION
“LOW SPEED STARTER TUBE COMPATIBILITY” IN CHAPTER 2 TO DETERMINE THE
COMPATIBILITY OF YOUR GENERATOR.
AEC DEVICES:
Refer to the compatibility statement at the end of this chapter.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-24 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
CPI Canada Inc. Pre-installation 1
1.6.0 CUSTOMER SUPPORT
Address any questions regarding X-ray generator operation to:
Mail: Customer Support Department
Communications and Power Industries Canada Inc.
45 River Drive
Georgetown, Ontario, Canada L7G 2J4
Telephone: (905) 877-0161
Fax: (905) 877-8320
Attention: Customer Support Department
E-mail: marketing@cmp.cpii.com
Attention: Customer Support Department
1.7.0 COMPATIBILITY STATEMENT
The compatibility statement for this generator follows this page.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y Page 1-25
1 Pre-installation CPI Canada Inc.
(This page intentionally left blank)
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-26 CMP 200 X-Ray Generator Service Manual Ch # 740987-01 Rev. Y
Loading...
+ 243 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.