A From the operational display, press the SEL/MENU button.
B D.POINT will appear on the display. Scroll through the menu items using the
and
back to the operational display.
buttons. Press
to enter the selected menu, or the
button to go
4.2 - D.POINT (Decimal Point)
A Enter the D.POINT menu by following the instructions in 4.1.
B Your current decimal point position will appear on the display - either: X.XXXX,
X.XXX, X.XX, X.X or X. Use the
le or right, and then press
Note that adjusting your decimal point position will not scale your calibration. 4 decimal places
(X.XXXX) are only available when CUSTOM UNITS (see 4.4B) are not used.
and buttons to shi the decimal point
to save and return to the operational display.
4.3 - ROUND (Rounding)
A Enter the ROUND menu by following the instructions in 4.1.
B The currently selected display rounding will appear. Use the
Rounding is quoted in display counts and is not in uenced by decimal point position.
4.4 - UNITS
A Enter the UNITS menu by following the instructions in 4.1.
B The currently selected display units will appear. Use the
select: OFF, K.G., LB., TON., or CUSTOM, and then press
If you selected CUSTOM, continue to 4.4C now.
If you selected something else, there is nothing further to con gure. A er
pressing
OFF: No units will be displayed. K.G./LB./TON.: Units will be displayed in the units area at the
right edge of the screen, a er the six display digits. CUSTOM: Units will be displayed in the
sixth display digit - your usable display counts will be reduced to ve.
the display will return to normal operating mode.
and buttons to
to save.
C Your currently selected custom unit will appear in the sixth position on the
display. Use the
to save and return to the operational display.
and buttons to select a letter from A-Z, and then press
4.5 - CAL. (Calibration)
A Enter the CAL. menu by following the instructions in 4.1.
B The currently selected calibration method will appear on the display. Use the
and buttons to select a calibration method: KEY IN, 2POINT, or RESET,
and then press
If you selected KEY IN, continue to 4.5C now.
If you selected 2POINT, skip to 4.5E now.
If you selected RESET, there is nothing further to con gure. The LPI610
to continue.
will restore the factory calibration and return to normal operating mode.
KEY IN scales the display using high and low display values, and relies on the factory set
calibration. 2POINT calibrates the unit using live input signals. A calibrator or other external
signal source is required for this procedure. RESET will restore the factory settings.
A Enter the B.LIGHT menu by following the instructions in 4.1.
B The currently selected backlight time will appear. This is the amount of time
that the backlight will remain on for a er the
and
return to the operational display.
Note that the backlight charges o the input signal and is not intended to be used for extended periods of time. It may appear dimmer if used repeatedly, or if used immediately a er
the instrument has been connected.
buttons to adjust the time from 1-9 SEC, and then press
button is pressed. Use the
to save and
4.7 - CNTRST (Contrast)
A Enter the CNTRST menu by following the instructions in 4.1.
B The currently selected screen contrast will appear on the display. Use the
and
press
buttons to adjust the contrast between CNT. 1 and CNT. 6, and then