Digits 1 row of 4 digits, large 20mm
(0.8") size, 7-segment LED
Construction
Casing Panel mount case, 5 buttons
Ingress protection rating IP65 dust/
splash proof (face only)
Dimensions (H x W x D)
48 x 96 x 120mm (1.89 x 3.78 x 4.72")
Cutout area (H x W)
45 x 92mm (1.77 x 3.62")
3
2
FRONT PANEL & DISPLAY
2.1 - Front panel
SPX
The SP LED's are used to
indicate active setpoints.
This button is used to
access the Input Setup & Calibration menu (Section 6).
This button is used to save
your settings and advance
to the next step in the setup process.
This button is typically used to scroll through options or increase values in the
setup menu. Pressing this button from the main display allows you to view/reset the
PEAK value (see 2.4).
This button is typically used to scroll through options or decrease values in the
setup menu. Pressing this button from the main display allows you to view/reset the
VA LY (valley) value (see 2.4).
48mm
(1.89")
96mm (3.78")
This button is used to access the Setpoint Setup menu (Section 7) and the
Setpoint Open Access menu (Section 8).
2.2 - Display brightness
To adjust the display brightness, press the and buttons together from the
main display. BRI appears and toggles with the current setting. Use the and
buttons to adjust the LED backlight, and then press to nish.
Note that you will also need to change the Input Mode in soware from the front
panel - see 6.2C.
G Slide the input module back into the meter case.
Make sure that it is sitting in the tracks on the le and right. Press rmly until the input module
is fully inserted and sits ush with the other boards that are visible from the back of the meter.
H Replace the back plate.
Begin by inserting the two lower tabs into the slots, and then position the upper tabs so that
they will not catch on the top lip of the meter case. Apply rm pressure until the back plate
clicks into place.
I Reconnect the plugs and return the meter to the panel installation.
Don't forget that once the unit is connected and powered up, you will need to
enter the Input Setup menu (6.2) and conrm that the correct Input Mode is
A Enter the calibration mode by pressing the button.
_ _ ENTER CAL PIN NUMBER scrolls across the display and toggles with 0. Use
the
press . If the correct PIN is entered, setup is started at 6.2.
If an incorrect PIN number is entered, _ _ _ INCORRECT PIN NUMBER – ACCESS
DENIED scrolls across the display and it returns to normal operating mode.
You will have the opportunity to change your PIN number at the end of this section
(6.6). If you have forgotten your PIN number, see Section 9.
and buttons to enter your security code (factory default '1'). Then
6.2 - Input setup
A _ _ INPUT SETUP scrolls across the display and toggles with SKIP. Press to
skip to 6.3, or the button and then to ENTR (enter) input setup.
B _ _ MAINS FREQUENCY scrolls across the display. Use the
to select 50HZ or 60HZ, and then press .
C _ _ INPUT MODE scrolls across the display and toggles with the current selec-
tion. Use the
2V (0–2V) or 10V (0–10V). Then press .
Note that if you change the INPUT MODE in this step, then the input header on the analog
input module may also need to be changed. See Section 5 for more information.
D _ _ DECIMAL POINT POSITION scrolls across the display and toggles with
the current selection. Use the and buttons to select NODP (no decimal
point), 0.1, 0 .12 , or 0.123, and then press to accept and continue.
E _ _ DISPLAY ROUNDING scrolls across the display and toggles with the selected
rounding. Using the
and buttons to select: 0–20 (0–20mA), 4–20 (4–20mA),
Rounding is quoted in display counts and is not inuenced by decimal point position. For example, if your input signal is 5.3mA, the display will show: 5.3 (for rounding=NONE), 5.4 (for
rounding=2), 5.5 (for rounding=5), or 5.0 (for rounding=10).
6.3 - Calibration
A _ _ CALIBRATION TECHNIQUE scrolls across the display and toggles with SKIP.
Press to skip to 6.4, or use the and buttons to select a calibration
method: either AUTO or MAN (manual). Then press .
¨ If you selected AUTO, complete steps 6.3B–D now.
¨ If you selected MAN, complete steps 6.3E–F now.
¨ If you selected SKIP, skip to 6.4 now.
AUTO (key-in) calibration uses zero and span values to calculate the scale and oset. This is the
most accurate calibration method, but requires known low and high input signals (or the use
of a calibrator).
MAN The manual calibration procedure uses low and high display values, and is intended for
a pre-calibrated sensor with a known output range. (For example 4mA=0 and 20mA=1000.) It
does not require any input signals to be applied to the controller during calibration.
Auto calibration
B _ _ APPLY LOW INPUT SIGNAL – – – ENTER LOW DISPLAY VALUE scrolls
across and toggles with the current selection. Apply the required low input sig-
nal, and wait a moment for the signal to stabilise. Then, using the
and
buttons, enter your low display value, and press to accept and continue.
C _ _ APPLY HIGH INPUT SIGNAL – – – ENTER HIGH DISPLAY VALUE scrolls
across and toggles with the current selection. Apply the required high input
signal, and wait a moment for the signal to stabilise. Then, using the
and
buttons, enter your high display value, and press .
D If Auto calibration was successful, you will be directed out of the calibration
menu to the operational display without viewing any further scrolling messag-
es. (To proceed to step 6.4, you must select SKIP at 6.3A.)
If calibration fails, _ _ CALIBRATION FAILED will scroll across the display and
you will be directed back to the operational display. The most likely cause of this
samples that the controller will average, and then press .
Increasing the number of samples will stabilise measurement, but it will also slow down response rates. A typical value is 4.
C _ _ AVE WINDOW scrolls across the display and toggles with the currently se-
lected averaging window value. Using the and buttons, alter the signal
averaging window. Then press .
If your input signal contains large noise spikes, you can increase the size of the averaging window to ensure that these are still averaged. However, increasing the window size too far will
reduce the ability of the controller to respond quickly to real changes in input signal. Setting
AVE WINDOW to 0 will give continuous averaging as per the selected averaging samples. A
typical value is 10% of your system capacity.
6.5 - Analog output setup
N.B. All new units are calibrated before shipping. Recalibration is only necessary if settings
are wiped or the unit's accuracy requires verication aer a long period of use. e.g. 1 year.
A _ _ ANALOG OUTPUT SETUP scrolls across the display and toggles with SKIP.
If your controller does not have analog output installed, (or you do not wish to
congure your analog output now), press to skip to 6.6. Otherwise, press
the button and then to ENTR (enter) analog output setup.
B _ _ LOW SCALE VALUE FOR ANALOG OUTPUT scrolls across the display and
toggles with the currently selected low scale display value. Use the
and
buttons to enter your cal low position, and then press .
C _ _ HIGH SCALE VALUE FOR ANALOG OUTPUT scrolls across the display and
toggles with the currently selected high scale display value. Use the
and
buttons to enter your cal high position, and then press .
D _ _ CALIBRATE ANALOG OUTPUT? scrolls across the display and toggles with
SKIP. Press
now to skip analog output calibration and continue to 6.6, or the
button and then to ENTR (enter).
Factory analog output calibration is precisely set before shipping this instru-
ment, and should not be adjusted unless advised by the manufacturer.
The soware in your controller will allow you to congure up to 4 setpoints,
however full functionality is only supported when relay output hardware is
installed.
(Setpoints with no corresponding relay output hardware may be used as simple LED
indicators, if desired. In this case, features requiring relay output functionality will
continue to appear in the setup menu, but will be ignored by the controller.)
SETPOINT SETUP
7.1 - Enter Setpoint PIN number
A Enter setpoint setup mode by pressing and holding the button for 3 seconds.
_ _ ENTER SP PIN NUMBER scrolls across the display and toggles with 0.
Use the
Then press . If the correct PIN is entered, setup is started at 7.2.
If an incorrect PIN number is entered, _ _ _ INCORRECT PIN NUMBER – ACCESS
DENIED scrolls across the display and it returns to normal operating mode.
You will have the opportunity to change your PIN number at the end of this section
(7.3). If you have forgotten your PIN number, see Section 9.
and buttons to enter your security code (factory default '1').
7.2 - Setpoint setup
A _ _ EDIT SETPOINT scrolls across the display and toggles with SKIP. Press
now to skip to 7.3, or use the and buttons to select a setpoint to edit,
and then press .
B _ _ SP VALUE scrolls across the display and toggles with the current value for
the selected setpoint. Using the
which the selected setpoint will activate, and then press .
C _ _ SP ACTIVATION scrolls across the display and toggles with the current acti-
activation to operate ABVE (above) or BLW (below) the setpoint value, and
then press .
ABVE: Relay turns on above the setpoint value and o below it.
BLW: Relay turns on below the setpoint value and o above it.
D _ _ HYSTERESIS TYPE scrolls across the display and toggles with the current set-
ting for the selected setpoint. Using the and buttons, select either ALM
(alarm) or CTRL (control), and then press .
ALM - SETPOINT VALUE controls setpoint
activation point. HYSTERESIS VALUE con-
trols setpoint deactivation point.
CTRL - SETPOINT VALUE controls setpoint
deactivation point. HYSTERESIS VALUE con-
trols setpoint reactivation point.
Hysteresis
band
Hysteresis
band
Energised Below
Hysteresis
band
Hysteresis
band
Energised Below
E _ _ HYSTERESIS VALUE scrolls across the display and toggles with the hysteresis
value for the selected setpoint. Use the and buttons to adjust this value if
required, and then press .
The HYSTERESIS VALUE denes the separation band between setpoint activation and deacti-
vation, and will operate as per the HYSTERESIS TYPE setting selected in 7.2D.
F _ _ MAKE DELAY scrolls across the display and toggles with the current make
delay time for the selected setpoint. This is the time delay between setpoint ac-
tivation, and when the relay turns on. Adjust this value in 0.1 second increments
using the and buttons, and then press .
G _ _ OPEN ACCESS TO SP VALUE scrolls across the display and toggles with the
open access permission setting for the selected setpoint. Use the
and to
select either YES or NO, and then press .
When enabled, this option allows the selected setpoint's value to be edited directly aer pressing the
button, without needing to enter a PIN number or go through all of the other op-
tions. Each setpoint can individually have this option enabled or disabled. See Section 8.
H The step that you proceed to now will depend on which setpoint you are editing
(selected in 7.2A):
¨ If you are currently editing SP 1, skip to 7.2J now.
¨ If you are currently editing SP 2–4, continue to 7.2I now.