Defibtech RMU-1000
Automated Chest
Compression System
User Manual
ELECTRONIC
DISTRIBUTION
RA C- E1510 EN - DN
Notices
Defibtech, L.L.C. shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of
this material.
Information in this document is subject to change without notice. Names and data
used in the examples are fictitious unless otherwise noted.
Limited Warranty
The “Limited Warranty” shipped with Defibtech ACC products serves as the sole
and exclusive warranty provided by Defibtech, L.L.C., with respect to the products
contained herein.
11 Warranty Information .............................................65
RA C- E1510 EN - DN
5
6
RA C- E1510 EN - DN
1 Introduction to the RMU-1000 ACC
This User Manual provides information to guide trained operators in the use and maintenance of
the Defibtech RMU -1000 Automated Chest Compressor (ACC) and its accessories.
This chapter includes an overview of the ACC, intended use information, a discussion of when it
should and should not be used, and information on operator training.
1.1 Overview
The RMU-1000 ACC is an automated, portable, battery-powered device that provides chest
compressions on adult patients who have cardiac arrest.
The ACC, when applied to a patient who is unconscious and not breathing, is designed to:
• Provide consistent depth and rate chest compressions.
• Allow for automated chest compressions in both the in-hospital and out- of-hospital settings
including during patient transport.
• Be applied to the patient with minimal interruption of CPR.
The major components of the RMU-1000 ACC are the Backboard, the Frame and the Compression
Module. The Backboard is placed under the patient to provide a base for the ACC system.
The Frame is placed over the patient and snaps into the Backboard with self-locking latches.
The Compression Module mounts into the Frame and contains the user interface, a replaceable
Battery Pack and the piston drive used to generate the chest compressions.
Compressions are initiated using a simple three-step operational sequence once the RMU-1000
ACC has been applied to a patient: the unit is turned on, the piston height is adjusted for the patient’s
chest size, and the compressions button is pushed. Additional user interface features include a
pause function, a warning indicator to notify the operator for possible misuse or malfunction, an
audible warning mute, and a Battery Pack capacity gauge.
The RMU-1000 ACC can be operated using a replaceable, rechargeable Battery Pack or with an
external power adapter. A fully-charged, new Battery Pack can provide continuous operation for
over an hour and can be recharged in the Compression Module.
A USB port on the Compression Module allows the Module to be connected to a personal
computer and for ACC data retrieval and event reporting when used in conjunction with utility
software available at www.defibtech.com.
1. Introduction
7
RA C- E1510 EN - DN
1.2 The Defibtech RMU-1000 ACC
1
9
10
11
12
13
2
3
4
5
15
14
6
8
RA C- E1510 EN - DN
8
7
The Defibtech RMU-1000 ACC (continued)
1. Introduction
1. User Control Panel. The User Control
Panel contains the user interface for the
ACC system.
2. Compression Module. The Compression
Module contains all the therapeutic
components of the ACC system, including
the piston drive mechanism, the control
electronics, the User Control Panel, and the
Battery Pack interface.
3. Serial Number. The serial number can be
found on the Compression Module.
4. External Power Input. The External
Power Input jack is used to connect an
external power adapter to operate the unit
or to charge the Battery Pack.
5. USB Port. The USB Port is located on
the bottom of the Compression Module.
It allows the Module to be connected to
a personal computer and for ACC data
retrieval and event reporting when used in
conjunction with utility software available at
www.defibtech.com. It is not intended to
be used during rescue operation.
6. Frame. The Frame attaches to the
Backboard and holds the Compression
Module. It is used to maintain the
Compression Module’s position over the
patient.
7. Backboard. The Backboard is the base
for the ACC system. It is placed under the
patient and provides an interface that the
Frame attaches to.
8. Backboard Latches. One on each side
of the Frame, the Backboard Latch is the
mechanism that secures the Frame to the
Backboard.
9. Stabilization Strap Connectors. One on
each side of the Frame, the Stabilization
Strap Connectors allow the Stabilization
Strap to be secured to the Frame to
maintain the ACC position over the patient's
chest.
10. Backboard Release Lever. One on each
side of the Frame, the Backboard Release
Levers are used to release the Frame from
the Backboard.
11. Handles. One on each side of the Frame,
the Handles provide a secure way to grasp
the Frame during ACC assembly and patient
transport.
12. Battery Pack. The Battery Pack provides
a replaceable primary power source for the
Compression Module.
13. Battery Pack Release. The Battery Pack
Release ejects the Battery Pack from the
Compression Module.
14. Patient Interface Pad. The Patient
Interface Pad is a user-replaceable, single
use component that provides the interface
between the piston and the patient’s chest.
15. Compression Piston. The Compression
Piston is driven by a motor housed inside
the Compression Module and, with the
Patient Interface Pad attached to the
distal end of the Piston, provides the
compressions to the patient's chest.
9
RA C- E1510 EN - DN
1.3 Indications for Use
The RMU-1000 Automated Chest Compressor (ACC) is to be used for performing external cardiac
compressions, as an alternative to manual CPR, on adult patients who have acute circulatory
arrest, defined as the absence of spontaneous breathing and loss of consciousness.
The RMU-1000 should only be used in situations where chest compressions are likely to help the
patient.
Federal Law (USA) restricts this device to sale by or on the order of a physician.
1.4 Contraindications
Do not use the RMU-1000 ACC in the following cases:
• It is not possible to position the ACC safely or correctly on the patient’s chest
• The patient is too small for the starting piston height to reach the patient’s chest
• The patient is too large for the Frame to attach to the Backboard or if the Compression
Module/Piston cannot be mounted without compressing the patient’s chest
Always follow local and/or recognized resuscitation guidelines for CPR when using the RMU1000 ACC.
1.5 Side Effects
The International Liaison Committee on Resuscitation (ILCOR) states the following side effects
of CPR:
“Rib fractures and other injuries are common but acceptable consequences of CPR given the
alternative of death from cardiac arrest. After resuscitation, all patients should be reassessed
and re-evaluated for resuscitation-related injuries.” (From the 2005 International Consensus
Conference on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science
With Treatment Recommendations, hosted by the American Heart Association in Dallas, Texas,
January 23–30, 2005. Published in Circulation. 2005; 112: III-5-III-16.)
The above side effects, as well as bruising and soreness of the chest, may commonly occur after
the use of the RMU-1000 ACC. (Black CJ, Busuttil A, Robertson C. Chest wall injuries following
The RMU-1000 ACC, when applied to a patient who is unconscious and not breathing, is intended
to:
• Provide consistent depth and rate chest compressions.
• Allow for automated chest compressions in both the in-hospital and out-of-hospital settings
including patient transport.
• Be applied to the patient with minimal interruption of CPR.
1.7 Operator Training Requirements
In order to safely and effectively operate the RMU-1000 ACC, it is the responsibility of the
operator to obtain the following training:
• RMU-1000 ACC training in accordance with the User Manual including handling of the
actual device
• CPR training in accordance with resuscitation guidelines as required by local, state,
provincial, and/or national regulations, e.g. American Heart Association, European Council of
Resuscitation
• Thorough knowledge and understanding of the material presented in this User Manual
The RMU-1000 ACC is intended for use by qualified medical personnel certified to administer
CPR (e.g. first responders, ambulatory personnel, nurses, physicians or medical staff).
1. Introduction
11
RA C- E1510 EN - DN
12
RA C- E1510 EN - DN
2 Warnings and Cautions
This chapter includes a list of warnings and cautions that relate to the RMU -1000 ACC and its
accessories. Many of these messages are repeated elsewhere in this User Manual and on the
RMU-1000 ACC or accessories.
2.2 WARNINGS:
Immediate hazards that will result in serious personal injury or death.
None known.
Conditions, hazards, or unsafe practices that may result in serious personal injury or death.
• Improper use can cause injury to patient. Use the RMU-1000 ACC only as instructed in the
User Manual.
• Improper use can cause injury to operator or bystander. Keep fingers and hands away from
Piston during operation.
• Improper maintenance can cause the RMU -1000 ACC not to function. Maintain the RMU1000 ACC and rechargeable Battery Pack only as described in the User Manual. Failure to
maintain the Battery Pack per the instructions outlined in this User Manual will result in the
RMU-1000 ACC becoming inoperable.
• No modification of this equipment is allowed. The RMU-1000 ACC contains no userserviceable parts. Do not disassemble, repair or alter the RMU-1000 ACC or any of its
components.
• Do not immerse Compression Module or Battery Pack in water or other liquids. Immersion
in fluids may result in fire or explosion.
• Do not sterilize the RMU-1000 ACC or its accessories.
• Do not let fluids get into the RMU-1000 Compression Module. Avoid spilling fluids on the
ACC or its accessories. Spilling fluids into the RMU -1000 ACC may damage it or cause a
fire or shock hazard.
• If the patient is too large for the Frame, remove Frame and continue manual CPR
compressions. Do not use the ACC if the Frame cannot be latched to the backboard.
• If the Piston cannot be adjusted to reach the patient's chest, the patient is too small.
Remove Frame and continue with manual CPR compressions.
• Do not initiate ACC compressions if the Piston is not in the proper position. An incorrect
start position may compromise the patient’s blood circulation.
• Do not use the ACC if the Frame cannot be latched in place (for any reason).
• Incorrect position over chest can result in injury or lack of effectiveness.
• Incorrect Piston start height can result in injury or lack of effectiveness.
• Carefully monitor the position of the Piston on the patient's chest to ensure that it has not
moved from the appropriate target area.
• Changed position over the chest during operation can result in injury or lack of effectiveness.
• Do not leave the ACC running while unattended. Patient injury may result if the unit is left
unattended.
RA C- E1510 EN - DN
2. Warnings and Cautions
13
WARNINGS (continued)
14
RA C- E1510 EN - DN
• The ACC can become hot during extended use. The applied part of the piston may rise 5°C
above ambient.
• If the position of the Piston changes as a result of defibrillation or other therapies,
immediately stop compressions and re-adjust the position of the ACC.
• Mechanical chest compressions may cause artifact and interfere with ECG analysis.
Always pause compressions before performing ECG analysis with other equipment.
• When the Battery Pack indicator shows one red segment, replace the Battery Pack as
soon as possible with a sufficiently charged Battery Pack or apply external power.
• To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
• If a spare Battery Pack or external power source are not available and the ACC stops
compressions, remove the unit from the patient and begin manual compressions
immediately.
• If there is a malfunction, the compressions are not sufficient, or something unusual
occurs during operation, then push the ON/OFF Button for one second to stop the ACC
from delivering compressions and remove the unit from the patient. Start manual chest
compressions as soon as possible.
• If there is a malfunction and the unit will not turn OFF, remove Battery Pack to stop
compressions. Remove the unit from the patient. Start manual chest compressions as
soon as possible.
• The Patient Wrist Straps are designed only for use with Defibtech RMU -1000 Automated
Chest Compression Systems. Do not apply to other items such as a stretcher or bedside
rails.
• Do not use the Patient Wrist Straps if damaged, soiled, or if Velcro does not bond
sufficiently to securely fasten the patient’s wrists to the RMU-1000 ACC.
• Do not over-tighten the Patient Wrist Straps to prevent potential injury. If the Patient
Wrist Straps are too loose, they may not securely fasten the patient’s arms to the ACC
appropriately.
• Do not use the Patient Wrist Straps to lift the patient. The straps are only intended to
attach the patient’s arms to the RMU-1000 ACC during transport.
• Observe the patient’s arms during use of the Patient Wrist Straps. Do not leave a patient
unattended when using the ACC or its accessories (e.g. Patient Wrist Straps).
• Do not obstruct intravenous (IV) access when deploying the RMU-1000 or when attaching
the Patient Wrist Straps.
• Use of damaged equipment or accessories may cause the device to perform improperly
and/or result in injury to the patient or operator.
• The ACC system components must be cleaned and the Patient Interface Pad replaced
between patients to avoid cross contamination.
• The use of accessories other than those specified may result in increased emissions or
decreased immunity of the RMU-1000 ACC.
• Portable and mobile RF communications equipment should be used no closer to any part
of the RMU-1000 ACC, including cables, than necessary.
WARNINGS (continued)
• The RMU-1000 ACC should not be used adjacent to other equipment and that if adjacent,
the RMU-1000 ACC should be observed to verify normal operation in the configuration in
which it will be used.
• Do not use in the presence of flammable gases or anesthetics. Use care when operating
this device close to oxygen sources (such as bag-valve-mask devices or ventilator tubing).
Turn off gas source or move away from patient, if necessary.
2.3 CAUTIONS:
Conditions, hazards, or unsafe practices that may result in minor personal injur y, damage
to the RMU-1000 ACC, or loss of data.
• The Compression Module must be assembled and locked to the Frame for proper operation.
• The Frame must be latched to the Backboard for proper operation.
• A new battery is set to ship mode. Must be activated before use! Activate battery by
charging in unit or charger.
• Follow all Battery Pack labeling instructions. Do not use a Battery Pack after its expiration date.
• Use only Defibtech approved batteries and accessories.
• The ACC must be paused in order to replace a Battery Pack. Failure to do so will require the
user to power up the ACC and reset the start position in order to resume compressions.
• Only use Defibtech accessories to power the ACC from an external power source.
• The Battery Pack must always be installed to operate the ACC from external power.
Without the Battery Pack, the ACC will flash the warning indicator and will not perform
compressions.
• The mains power quality must be that of a typical commercial or hospital environment.
• Defibrillation electrodes and pads should not be in contact with and be clear of the Piston
and other ACC components.
• Make sure other equipment and/or drugs are applicable for use with the ACC. Consult the
equipment's operating instructions.
• Avoid gel on chest. Gel on the chest (e.g. from defibrillation pads or ultrasound) in the
Patient Interface Pad target area may result in movement of the piston. Be sure to remove
any gel before use.
• Defibrillation pads or electrodes must be removed or moved away from the Patient
Interface Pad target area.
• Do not use on open wounds or if visible signs of existing injuries.
• The ACC should be applied to the patient’s bare chest. Remove clothing, undergarments
and jewelry before use.
• If the RMU-1000 ACC becomes hot, prolonged contact with the patient’s skin could result
in skin burns. If necessary, remove the patient’s arms from the Patient Wrist Straps.
• Do not allow the use of Stabilization Strap to delay or prevent other treatment of the
patient. Apply strap as soon as possible after beginning compressions and always before
patient movement.
RA C- E1510 EN - DN
2. Warnings and Cautions
15
CAUTIONS (continued)
• Recycle or dispose of lithium-ion batteries in accordance with local, state, provincial, and/
or national regulations. To avoid fire and explosion hazard, do not burn or incinerate the
Battery Pack. Do not crush.
• Use and store the RMU-1000 ACC only within the range of environmental conditions
specified in the technical specifications.
• Always store the ACC so it is ready to go for use. Store the Compression Module with a
fully charged Battery Pack installed and a Patient Interface Pad attached to the Piston. It
is recommended to maintain a charged spare Battery Pack and have the external power
adapter available with the unit at all times.
• Although the RMU-1000 ACC is designed for a wide variety of field use conditions, rough
handling beyond specifications may result in damage to the unit.
• Federal Law (USA) restricts this device to sale by or on the order of a physician.
16
RA C- E1510 EN - DN
3 Setting Up the RMU-1000 ACC
RMU-1000 Quick Reference Guide
This chapter describes the steps required to make your Defi btech RMU-1000 ACC operational.
The RMU-1000 ACC is designed to be stored in a Carrying Case with a small number of easy to
assemble components or stored in a fully assembled “ready” state. This chapter explains how to
set up the RMU-1000 ACC device.
3.1 Overview
The following components and accessories are included with the RMU-1000 ACC. Replacement
and other accessories are detailed in the “RMU-1000 ACC Accessories” section. Before getting
started, identify each component and ensure that the package is complete.
Backboard
Frame
Battery Pack
Compression
Module
Carrying CaseAC Adapter
Patient
Interface Pad
(package of 3)
Stabilization
Strap
TopBottom
3. Setting Up the
RMU-1000 ACC
Patient Wrist
Straps
User Manual
and Quick
Reference Guide
Before and during deployment of the RMU-1000:
• Confirm patient is not breathing and unresponsive.
• Start manual compressions.
• Minimize compression interruptions.
• Perform manual compressions whenever possible.
1
2
3
4
Run
Continuous
5
Run with
Breaths
Emergo Europe
Defibtech, L.L.C.
Molenstraat 15
741 Boston Post Road, Suite 201
EC REP
2513 BH The Hague, The Netherlands
Guilford, CT 06437 USA
Defi btech RMU-1000
Place Backboard under patient
and expose chest.
Automated Chest
Compression System
Attach Frame to Backboard, with
patient’s arms outside the Frame.
Insert Compression Module with
Patient Interface Pad installed into
Frame (down and twist).
Turn on (press and hold
for at least 1 second).
Adjust Piston height until it touches
the patient’s chest.
Start compressions.
Apply Stabilization Strap.
User Manual
17
RA C- E1510 EN - DN
3.2 Complete Initial Assembly of the ACC
Before being placed into service, the RMU-1000 ACC unit should be completely assembled and
operationally checked to ensure that all components are present and functional.
3.3 The Backboard
The Backboard is the base for the ACC system. It is placed under the patient and has attachment
points to which the Frame latches. There are no moving parts on the Backboard.
Attachment
Points
Attachment
Points
3.4 The Frame
The Frame holds the Compression Module. It attaches to the Backboard and is used to maintain
the Compression Module’s position over the patient.
To attach the Frame to the Backboard:
• Place the Backboard on a flat surface, such that the patient position diagram is visible.
• Attach the Frame to the Backboard by aligning the Frame latches over the Backboard pins
and pushing down until the latches snap into place. The latches may be clicked into place
one at a time or simultaneously.
18
RA C- E1510 EN - DN
The Frame (continued)
To remove the Frame, push in on the two backboard release levers and lift the Frame off of the
Backboard. The latches may be released together or one at a time.
3.5 Attaching the Patient Wrist Straps to the Frame
Forearm
Strap
Frame Strap
Frame
Strap
Loop
The Patient Wrist Straps are designed to attach a patient’s arms to the ACC for ease of transporting
the patient and the ACC. As shown in the above left illustration, each Patient Wrist Strap is
comprised of two main strap sections: a Frame Strap that attaches to the ACC’s Frame and an
Forearm Strap that wraps around the patient’s wrist.
Note: As shown in the above right illustration, when attaching the ACC’s Frame to the Backboard,
the Patient Wrist Straps can be oriented towards the patient’s head to maintain access to the ACC
Frame’s Backboard Release Levers or on the opposite side of the ACC Frame as appropriate.
3. Setting Up the
RMU-1000 ACC
19
RA C- E1510 EN - DN
Attaching the Patient Wrist Straps to the Frame (continued)
To maximize the available time to perform a rescue, Defi btech recommends that the Patient Wrist
Straps be affi xed to the ACC Frame using the instructions shown below prior to a rescue and that
the Frame be stored in the ACC’s Carrying Case with the Patient Wrist Straps already attached
to the Frame.
1. Attach the Frame Strap to the ACC Frame by inserting the rounded end of the Frame Strap
through the Frame Strap Loop, as shown in the above left illustration.
2. After the Frame Strap has been inserted through the Frame Strap Loop, pull the end of the
Frame Strap in the opposite direction so that the Frame Strap is tightly wrapped around the
ACC Frame, as shown in the above right illustration. Secure in place using the Velcro
®
on the
underside of the Frame Strap.
3. If necessary, rotate the attached Patient Wrist Strap so that the Forearm Strap component
faces up. This allows the procedure described in Section 4.5 to be performed.
4. Repeat the above three steps to attach the second Patient Frame Strap’s Frame Strap to the
ACC Frame.
20
RA C- E1510 EN - DN
3.6 The Compression Module
The Compression Module contains all the active components of the ACC system, including the
User Control Panel, the Battery Pack and the Compression Piston. It easily attaches to the Frame
and locks into place for operation.
User Control Panel
Battery Pack
Patient Interface Pad
3. Setting Up the
RMU-1000 ACC
Compression Piston
21
RA C- E1510 EN - DN
The Compression Module (continued)
To attach the Compression Module to the Frame, perform the following steps:
• Attach the Frame to the Backboard (described in Section 3.4).
• Insert the locking sleeve of the Compression Module into the module receptacle of the
Frame as shown below. The Module should be inserted at approximately 90 degrees to
the Frame at which point it should rest in the Frame. Rotate the Module in either direction
until it is in line with the Frame and snaps to lock into place. When properly attached, the
Compression Module should be securely attached to the Frame.
Locking
Sleeve
Module
Receptacle
22
RA C- E1510 EN - DN
To remove the Compression Module from the Frame, push down on the Compression Module
and then rotate it approximately 90 degrees to either direction. The Compression Module can
then be lifted out of the Frame. Be careful not to drop the Module.
3.7 Installing and Removing the Patient Interface Pad
Attach a Patient Interface Pad to the distal end of the Piston by pressing the pad onto the piston
until it snaps into place, rotating pad if necessary. To remove the Patient Interface Pad, grasp the
pad and pull, as shown below.
Note: Prior to installation, check to make sure that the Patient Interface Pad is clean and is
undamaged. If the Patient Interface Pad is damaged (e.g. one or more of the pins on underside of
the pad is crushed and/or missing), discard it and install an undamaged pad.
Installing
Removing
3.8 Installing and Removing the Battery Pack
The Battery Pack provides power to the RMU-1000 ACC. Do not install the Battery Pack after the
expiration date printed on the label.
Before inserting the Battery Pack into the RMU-1000 ACC, ensure
that the Battery Pack opening in the side of the Compression Module
is clean and clear of any foreign objects. Insert the Battery Pack into
the opening on the side of the ACC Compression Module. Push the
Battery Pack all the way in until the latch clicks. The Battery Pack will
operate in either orientation (with the contacts toward the unit).
When the Battery Pack is inserted, the RMU-1000 ACC will display the status of the Battery Pack
on the Compression Module's Battery Pack indicator for approximately three seconds.
Ship Mode: Battery Packs are shipped from the factory at less than 30% charge capacity
in a low-power state known as ship mode to minimize energy loss during shipment. Any
Battery Pack in ship mode cannot power the ACC until it is taken out of ship mode and
should be transferred from ship mode to active mode as soon as possible upon receipt.
To transfer a Battery Pack from ship mode to active mode, insert the Battery Pack into the
Compression Module as described earlier in this section. Then connect the AC Adapter to
the Compression Module, as described in Section 3.9, “Charging the Battery Pack.” Upon
detecting external power (allow at least 10 seconds for this to occur), the Battery Pack will
automatically switch from ship mode to active mode. The optional Battery Pack Charging
Station (see Section 6.7) can also be used to switch Battery Packs from ship mode to
active mode. Defibtech recommends charging any Battery Pack that has been brought out
of ship mode to full capacity (see Section 3.9 for details). In addition, a Battery Pack should
not be stored in ship mode for any longer than 12 months before charging it to full capacity.
3. Setting Up the
RMU-1000 ACC
CAUTION
A new battery is set to ship mode. Must be activated before use!
Activate battery by charging in unit or charger.
23
RA C- E1510 EN - DN
Installing and Removing the Battery Pack (continued)
To remove the Battery Pack, squeeze the Battery Pack eject release latches on either side of the
Battery Pack opening. After the Battery Pack is partially ejected, pull the Battery Pack out.
The Battery Pack must always be installed in the unit in order to operate the
CAUTION
RMU-1000, even when powered by the AC adapter.
3.9 Charging the Battery Pack
The ACC battery is a proprietary rechargeable Battery Pack. Make sure to fully charge a Battery
Pack as part of preparing the ACC for service. To charge the Battery Pack, install it into the
Compression Module and connect the AC Adapter to the external power input jack on the
Compression Module. The Battery Pack can also be charged using an optional Battery Pack
Charging Station (see Section 6.7 for details).
24
RA C- E1510 EN - DN
To connect the AC Adapter to the ACC, insert the AC adapter plug into the Compression Module's
external power input jack as shown in the center illustration above. As the jack is keyed, the raised
notch on the plug must align with the notch on the jack in order for the plug to seat properly and
lock into place. To remove the plug, grasp the top-most portion of the plug barrel and pull, as
shown in the right-most illustration above.
A fully charged Battery Pack condition is indicated by all- green LEDs on the User Control Panel's
Battery Pack status indicator. (To determine Battery Pack status, see the "Battery Pack Indications
and Alerts" charts that follow.)
Note: The Battery Pack may become unresponsive and non-functional if improperly maintained and
left fully depleted. See Section 5.1, “Routine Unit Maintenance”, for ACC maintenance information.
Charging the Battery Pack (continued)
Only use approved accessories with the RMU-1000. The Battery
Pack, Battery Charger, and AC adapter are specifically designed for
CAUTION
use with the ACC. Using other accessories can cause permanent
damage and void the warranty.
Battery Pack Indications and Alerts (User Control Panel)
LED IndicationsVisual IndicationDescriptionAction
All green
Battery Pack fullycharged (>80%)
None
Partial green
Red bar (solid)
Red bar
(slow flashing)
Red bar
(fast flashing)
Partial green
(slow flashing)
Number of bars show
percent of Battery Pack
charge (20% per bar)
Battery Pack low
(<20% charge
remaining)
Battery Pack low
(<10% charge remaining)
Battery Pack not
installed
Battery Pack is in ship
mode
Battery Pack malfunction
Battery Pack depleted
Battery Pack chargingNone
None (if possible, charge to full capacity)
If performing a rescue, replace Battery
Pack with a charged Battery Pack or
connect external power. Otherwise,
charge Battery Pack as soon as possible.
If performing a rescue, replace Battery
Pack with a charged Battery Pack or
connect external power. Otherwise,
charge Battery Pack as soon as possible.
Install a charged Battery Pack
Transfer Battery Pack from ship mode
to active mode. See“Ship Mode” in
Section 3.8,“Installing and Removing the Battery Pack”, for more information.
Replace Battery Pack with a charged
Battery Pack
If performing a rescue, replace Battery
Pack with a charged Battery Pack or
connect external power. Otherwise,
charge Battery Pack as soon as possible.
3. Setting Up the
RMU-1000 ACC
Left LED is
amber color
All off
Battery Pack has reached
end of life
Battery Pack depleted
Battery Pack malfunction
Replace with a new Battery Pack
If performing a rescue, replace Battery
Pack with a charged Battery Pack or
connect external power. Otherwise,
charge Battery Pack as soon as possible.
Replace Battery Pack with a charged
Battery Pack
25
RA C- E1510 EN - DN
Charging the Battery Pack (continued)
Battery Pack Indications and Alerts (Battery Pack)
To check the charge of a Battery Pack while it is not installed in the ACC, press the button on the
bottom of the Battery Pack for about a second. The Battery Pack charge indicator will show the
amount of remaining charge:
>80%<80%<60%<40%<20%
*NOTE: If the Battery Pack indicators do not illuminate after the button on the Battery Pack is
pressed, see Section 5.4, “Troubleshooting”.
Depleted or
Malfunction*
3.10 Completing and Testing the RMU-1000 ACC
Once the previous steps have been completed to set up your RMU-1000 ACC, follow this
procedure to perform an initial test of the unit:
1. Turn the unit on by pressing the ON/OFF button for at least one second.
2. Verify that piston is fully retracted into the Compression Module, the Warning Indicator is not illuminated and the Battery Pack Indicator is green.
3. Press the Adjust Down button to drive the piston to the bottom of its stroke.
4. Press the Adjust Up button to bring the piston back to the home position.
5. Turn unit off by pressing and holding the ON/OFF button for more than one second.
Adjust Up/
Down
Pause
Battery Pack
Indicator
26
RA C- E1510 EN - DN
USER
CONTROL
PANEL
ON/OFF
Run
Compressions
(Top: continuous;
Bottom: with breaths)
Warning
Indicator
Warning
Mute
Service
Indicator
3.11 Disassembling and Storing the RMU-1000 ACC
The RMU-1000 ACC should be stored in the Carrying Case. Use the following steps to
disassemble and store the unit in the case:
• Remove the Compression Module from the Frame by pushing down and rotating the
Compression Module approximately 90 degrees in either direction. Lift the Compression
Module from Frame and place in the appropriate storage section of the case. Always leave
a fully charged Battery Pack inserted in the Compression Module when storing.
• Release the Frame from the Backboard by pressing the Backboard release levers and lifting
the Frame from the Backboard. Note that the Frame can be released one side at a time.
• Place the components, User Manual, and Quick Reference Guide in the Carrying Case.
• Place at least one Patient Interface Pad in the Carrying Case.
Store the RMU-1000 ACC in environmental conditions within range of the specifi cations (see
“Environmental” section in Chapter 7 of this manual).
3. Setting Up the
RMU-1000 ACC
27
RA C- E1510 EN - DN
28
RA C- E1510 EN - DN
4 Using the RMU-1000 ACC
This chapter describes how to use the RMU-1000 ACC during an event. The RMU -1000 ACC
was designed for simple operation allowing the operator to focus on providing care to the patient.
The following sections describe in detail how to use the RMU -1000 ACC. The basic steps for use are:
• Place Backboard under patient
• Attach Frame to Backboard
• Attach Compression Module to Frame
• Press ON/OFF button for at least one second to turn ON
• Adjust piston height
• Press Run button to perform compressions
4.1 Overview
4. Using the
RMU-1000 ACC
Battery Pack
Battery Pack
Release
Backboard
Release
Lever
Stabilization
Strap
Connector
Backboard
Latch
Compression
Module
Handle
User Control
Panel
External Power
Input
Compression
Piston
Patient
Interface Pad
Frame
Backboard
29
RA C- E1510 EN - DN
Overview (continued)
Control Panel
Adjust Up/Down
Pause
Battery Pack Indicator
Warning
Indicator
Warning
Mute
30
RA C- E1510 EN - DN
ON/OFF
Run
Compressions
(Top: continuous;
Bottom: with breaths)
ON/OFF Button – Press the ON/OFF button for at least one second to turn the ACC on or off.
Adjust Up/Down Buttons – These buttons are used to move the Piston up or down relative
to the patient’s chest.
Adjust Down moves the piston toward the patient.
Adjust Up retracts the piston into the Compression Module.
Run Compressions Buttons – These buttons start chest compressions.
The Run Continuous button performs compressions until the Pause or Off
button is pressed. The adjacent LED flashes at the rate the AHA/ERC/ILCOR
suggests for giving rescue breaths without the ACC stopping for delivery of
breaths (approximately 1 breath every 6 seconds).
The Run With Breaths button performs compressions according to the
compressions-with-breaths protocol and pauses for the operator to give rescue
breaths. A reminder chirp and flashing LED occur during the 3 compressions
prior to the ventilation pause.
Pause Button – The Pause button stops compressions when running. Push Pause a second
time to resume compressions.
Battery Pack Indicator – Indicates the approximate remaining Battery Pack capacity. When
the Battery Pack is getting low, only one indicator segment will be visible and it will turn red –
replace Battery Pack as soon as possible or apply external power (see Section 3.9 for details).
Warning Indicator – The Warning Indicator illuminates to notify the user that the ACC has
determined that there is a problem (see Section 5.4,“Troubleshooting”).
Warning Mute Button – The Warning Mute button silences the audible sound associated with
a warning. The Warning Mute will automatically disable after 1 minute.
Service Indicator – The Service Indicator illuminates when the ACC requires servicing. See
Section 5.4,“Troubleshooting,” for more details.
Service
Indicator
4. Using the
4.2 Arrival and Setup
This section details the steps required to use the RMU-1000 ACC during an emergency. The
following instructions are for a two-rescuer scenario.
Note: To minimize CPR interruptions and to most effectively use the RMU-1000, two
rescuers are recommended.
Note: Providing manual chest compressions takes precedence over setting up and
initiating use of the RMU-1000.
The following instructions are for a two-rescuer scenario.
STEP 1) Confi rm that the patient is unresponsive and not breathing. Remove clothing from the
patient’s chest and begin manual CPR immediately.
RMU-1000 ACC
STEP 2) Open the Carrying Case and remove the ACC Backboard.
STEP 3) Place the Backboard under patient just below armpits. Lift patient body slightly and
slide the Backboard under patient or roll patient from side to side, as needed. The center of
the Backboard should be in line with the nipple line of the patient. Accurate placement of the
Backboard will help with the alignment step later.
STEP 4) Resume manual CPR.
RA C- E1510 EN - DN
31
Arrival and Setup (continued)
STEP 5) Place Frame over patient such that the Backboard is aligned with the mounting pins
on the Frame. Push down fi rmly until the Frame latches to the Backboard. Alternatively, attach
the side of the Frame nearest to you to the Backboard and then rotate the Frame to latch to the
opposite side of the Frame. Continue manual CPR compressions while attaching the Frame to
the Backboard.
STEP 6) Pull up on the Frame to make sure that Frame is securely latched to the Backboard.
If patient is too large for the Frame, remove Frame and continue manual CPR
compressions. Do not use the ACC if the Frame cannot be latched to the
WARNING
backboard.
32
RA C- E1510 EN - DN
STEP 7) Remove Compression Module from case. Check to make sure a Patient Interface Pad
is installed. If not, install a Patient Interface Pad per the instructions in Section 3.7, “Installing and Removing the Patient Interface Pad." Also make sure that a Battery Pack is installed. If not, install
a Battery Pack per the instructions in Section 3.8, “Installing and Removing the Battery Pack."
STEP 8) Mount the Compression Module in the Frame by inserting it at approximately 90 degrees
to the Frame and twisting the Module in either direction until it is aligned with the Frame and
locks in place as shown:
4. Using the
Arrival and Setup (continued)
Compression Module must be locked to Frame for proper operation.
CAUTION
STEP 9) If needed, adjust the Frame and Backboard assembly so that the Compression Module
piston is positioned over the chest and directly in line with the nipples. Note that the compression
target point is the same location as that used for manual compressions according to resuscitation
guidelines (Rajab,T et al. Technique for chest compressions in adult CPR. World J Emerg Surg.
2011; 6:41).
RMU-1000 ACC
WARNING
Do not initiate ACC compressions if the piston is not in the proper position.
Too low Too high
33
RA C- E1510 EN - DN
4.3 Operation and Adjustment
STEP 1) Press the ON/OFF button for at least one second to turn the unit on. If the Battery Pack
indicator shows red (low battery), or the ACC does not turn on, replace the Battery Pack or connect
external power. See Section 4.7 “Power“ for details.
STEP 2) The Piston must be adjusted to the height appropriate for the specific patient to ensure
that compressions are delivered to the proper depth. Adjust the height of the Piston by pressing
the Adjust Down and Adjust Up buttons until the Piston is touching the patient’s chest as shown.
Not Touching
CorrectToo High
Note: The ACC will automatically stop if the Piston encounters excessive resistance.
If the Piston cannot be adjusted to reach the patient’s chest, the patient is too
WARNING
small. Remove Frame and continue with manual CPR compressions.
34
RA C- E1510 EN - DN
STEP 3) Once the Piston is properly adjusted, start compressions, in accordance with your
emergency response protocol, by pushing the Run Continuous button OR the Run with
Breaths button:
Run Continuous
Run with Breaths
For any reason to temporarily stop compressions, press the pause button. To resume
compressions, press the pause button again or the appropriate run compressions button.
Note: While in use, the Compression Module’s internal fan may activate in order to maintain an
optimal operational temperature level inside the Module. The fan will deactivate when that level
has been achieved and may subsequently reactivate, if needed. This behavior is normal.
Note: If there is no user action after the ACC is powered on and the device is not performing
compressions, the Compression Module will automatically power itself off after 10 continuous
minutes of inactivity have elapsed.
4. Using the
Operation and Adjustment (continued)
Incorrect position of the Piston over chest can result in injury or lack of
WARNING
WARNING
WARNING
WARNING
WARNING
effectiveness.
Incorrect Piston start height can result in injury or lack of effectiveness.
Carefully monitor the position of the Piston on the patient’s chest to ensure
that it has not moved from the appropriate target area.
Changed position over the chest during operation can result in injury or lack
of effectiveness.
ACC compressions may interfere with ECG analysis. Pause compressions
during ECG analysis.
RMU-1000 ACC
WARNING
WARNING
WARNING
CAUTION
When Battery Pack indicator shows one red segment, replace Battery Pack
as soon as possible or apply external power.
Do not leave the ACC running while unattended. Patient injury may result if
the unit is left unattended.
If there are malfunctions, interruptions, the compressions are not sufficient,
or something unusual occurs during operation, then push the ON/OFF Button
for one second to stop the ACC from delivering compressions and remove the
unit from the patient. Start manual chest compressions as soon as possible.
Improper operation may injure operator or bystander. Keep fingers and hands
away from Piston during operation.
35
RA C- E1510 EN - DN
4.4 Stabilization
Once compressions have started, to help ensure the ACC remains appropriately positioned, apply
the Stabilization Strap as described below:
• Remove the Stabilization Strap from Carrying Case, if not already at the patient's side.
• Lift patient’s head and place strap behind patient’s neck. Note: Use other accepted patient
handling techniques if the patient has or may be suspected to have head, neck, spine or
other bone-structure compromising injuries.
• Connect the Stabilization Strap to the Frame by pushing the strap clips into the Frame's
Stabilization Strap Connectors until they click into place.
• The Stabilization Strap's length can be adjusted using the self-adhesive Velcro that holds
both strap clips to the Strap. Be sure that the Strap is suffi ciently tightened so that the
ACC's correct position over the patient's chest can be maintained.
36
RA C- E1510 EN - DN
• Make sure the Piston is correctly positioned on patient’s chest. If not, stop compressions,
loosen the Stabilization Strap and readjust as instructed above. Note: As with any step in
using the ACC, it is important to minimize the time CPR is not being performed. If the ACC
is not performing compressions for any reason, always consider performing manual CPR.
Stabilization (continued)
• To remove the Stabilization Strap, grip the top of the strap clip and pull the clip away from
the Frame in a angled motion, as shown in the images below.
4.5 Securing the Patient Wrist Straps to the Patient
Note: To maximize the available time to perform a rescue, Defi btech recommends that the
Patient Wrist Straps be affi xed to the ACC Frame using the instructions shown in Section 3.5
prior to a rescue and that the Frame be stored in the ACC’s Carrying Case with the Patient Wrist
Straps already attached to the Frame.
4. Using the
RMU-1000 ACC
As shown in the illustration above, when attaching the ACC’s Frame to the Backboard, the Patient
Wrist Straps can be oriented towards the patient’s head to maintain access to the ACC Frame’s
Backboard Release Levers or on the opposite side of the ACC Frame as appropriate.
37
RA C- E1510 EN - DN
Securing the Patient Wrist Straps to the Patient (continued)
To secure the Patient Wrist Straps to the patient’s wrists:
• Secure the Forearm Strap of the Patient Wrist Strap by placing the patient’s wrist on top of
the Forearm Strap and placing the soft Velcro part of the strap across the patient’s wrist,
as shown in the above left illustration.
• Secure the patient’s wrist to the ACC’s Frame by wrapping the hooked Velcro onto the soft
Velcro section, as shown in the above right illustration. Make sure that the strap securely
holds the patient’s arm in place.
Do not over-tighten the Patient Wrist Straps to prevent potential injury.
WARNING
38
RA C- E1510 EN - DN
• Repeat the above two steps to secure patient’s other arm to the ACC’s Frame.
Observe the patient’s arms during use. Do not leave a patient
unattended when using the ACC or its accessories (e.g. Patient Wrist
WARNING
WARNING
WARNING
CAUTION
Straps).
Make sure that intravenous (IV) access is not obstructed in any way
due to use of the Patient Wrist Straps.
Do not use the Patient Wrist Straps to lift the patient. The straps are
only intended to attach the patient’s arms to the ACC during transport.
If the ACC becomes hot, prolonged contact with the patient’s skin
could result in skin burns. If necessary, remove the patient’s arms from
the Patient Wrist Straps.
4.6 Transport
Note: Prior to transporting a patient with an affixed RMU-1000 ACC, make sure that the
Stabilization Strap has been attached to the unit and that a stretcher or other transportation
equipment is nearby.
To move the patient to a stretcher or another piece of transportation equipment:
• Prepare the stretcher/transport equipment near the patient.
• Position two people on either side of the patient. Other personnel may be needed to
stabilize the patient’s head and limbs, as necessary.
• When ready to move the patient, push Pause to temporarily stop compressions.
• Lift the patient by grabbing the Handle with one hand and use the other hand to support
the lower torso by grasping the patient’s leg, belt or pants.
• After the patient is safely on the stretcher/transport equipment, check that the RMU-1000
ACC and the Piston have not changed their position or readjust them to the target area, if
necess ar y.
• Push Pause again or the appropriate Run Compressions button to resume compressions.
During transport, the RMU-1000 ACC can be active if the RMU-1000 ACC and patient are safely
and securely positioned on the stretcher/transport equipment and the RMU-1000 ACC remains in
the target area and angle on the patient’s chest.
4. Using the
RMU-1000 ACC
WARNING
Carefully monitor the position of the piston on the patient’s chest to ensure
that it has not moved from the appropriate target area. Pause compressions
and readjust position if needed.
39
RA C- E1510 EN - DN
4.7 Power
If the Battery Pack charge becomes low during use, the warning and Battery Pack indicators will
flash. The Battery Pack status indicator will show only one red indicator bar.
OP TIO N 1) If a charged spare Battery Pack is available:
• Obtain the spare Battery Pack and have it ready for installation.
• Push Pause to temporarily stop compressions.
• Quickly eject the Battery Pack from the ACC by pressing the Battery Pack Release and
remove the depleted Battery Pack.
• With minimal interruption, install the spare Battery Pack.
• Wait for the Pause LED indicator to illuminate.
• Re-start compressions by pushing the Pause button again or one of the Run buttons.
40
RA C- E1510 EN - DN
Note: If the Battery Pack change takes more than approximately 15 seconds, the unit will power
off with the piston in place. Upon spare Battery Pack insertion, the Compression Module must
be powered on and the piston will automatically retract to set its start position. The piston should
then be re-adjusted to patient’s chest.
OPTION 2) At any time, the ACC can be connected to an external power source by connecting
the AC adapter to the external input jack of the Compression Module. See Section 3.9 for details.
Power (continued)
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
4. Using the
RMU-1000 ACC
The Battery Pack must always be installed to operate the ACC from external
power.
Only use Defibtech accessories to power the ACC from an external power
source.
The ACC must be paused in order to replace a Battery Pack. Failure to do so
will require the user to power up the ACC and reset the start position in order
to resume compressions.
If there is a malfunction during emergency use and the RMU-1000 ACC cannot
be paused or powered off, remove the Battery Pack from the Compression
Module. When the ACC has stopped, remove the Frame from the patient (as
the Piston does not automatically retract when the Battery Pack is removed).
Start manual compressions as soon as possible.
If a spare Battery Pack or external power source are not available and the
ACC stops compressions, remove the unit from the patient and begin manual
compressions immediately.
4.8 Other Therapies
The ACC may be used in conjunction with other therapies such as defibrillation and other patient
procedures, as appropriate.
Make sure other equipment and/or drugs are applicable for use with the ACC.
CAUTION
CAUTION
WARNING
WARNING
Consult the equipment’s operating instructions.
Defibrillation electrodes and pads should not be in contact with and be clear
of the Piston and other ACC components.
If the position of the Piston changes as a result of defibrillation or other
therapies, immediately stop compressions and re-adjust the position of the
ACC.
Chest compressions may interfere with shock delivery and ECG analysis.
Always pause compressions when providing a shock and during ECG analysis
with other equipment.
41
RA C- E1510 EN - DN
4.9 Removal from Patient
To remove the ACC from the patient:
• Turn the ACC off by pressing and holding the ON/OFF button for at least one second.
• To release a patient’s arm from the Patient Wrist Strap, pull up on the fabric tab on the end
of the Forearm Strap until the Velcro is no longer holding the patient’s arm in place. Repeat
for the patient’s other arm.
• Remove the Stabilization Strap (see Section 4.4).
• Press down and rotate the Compression Module approximately 90 degrees in either
direction. Lift Compression Module from Frame. (Note: This step may be performed after
Frame is removed from patient.)
• Press the latch release levers to disconnect the Frame from the Backboard.
• Remove the Backboard from underneath the patient.
4.10 Post-Use Procedures
After the RMU-1000 ACC has been used on a patient, the unit should be cleaned following
procedures in the “Cleaning” section in Chapter 5 of this manual and prepared for the next use.
The following steps should be performed:
• Remove and dispose of the used Patient Interface Pad.
• Remove and clean the Stabilization Strap.
• The Patient Wrist Straps may be removed for cleaning purposes or if they need to be
replaced (see Section 5.2). To remove the Patient Wrist Straps from the ACC’s Frame, pull
up on the rounded end of the Frame Strap until it can be slid through the Frame Strap Loop.
Note: To maximize the available time to perform a rescue, Defibtech recommends that the
Patient Wrist Straps remain affixed to the ACC’s Frame and that the Frame be stored in the
ACC’s Carrying Case with the Patient Wrist Straps already attached to the Frame.
• Clean all the components that have been in contact with the patient and let them dry (see
Section 5.2).
• Replace the Battery Pack with a fully charged Battery Pack or fully charge the Battery Pack
in the ACC unit.
• Install a new Patient Interface Pad.
• Repack the ACC components and spares in the Carrying Case.
4.11 Operational Environment
42
RA C- E1510 EN - DN
The Defibtech ACC is designed to operate in a wide range of environmental conditions. To ensure
the reliability and safety of the ACC in a given environment, refer to the “Environmental” section
in Chapter 7 of this manual for a detailed list of specified environmental conditions.
5 Maintenance and Troubleshooting
This chapter describes the maintenance and troubleshooting procedures for the RMU-1000 ACC.
The unit warnings and alerts are described along with recommended routine maintenance. A
troubleshooting guide is provided to help diagnose user serviceable problems.
The RMU-1000 ACC contains no user serviceable parts.
5.1 Routine Unit Maintenance
The RMU-1000 ACC is designed to be very low maintenance. Simple maintenance tasks are to
be performed regularly to ensure its readiness (see sample maintenance table below). Different
maintenance intervals may be appropriate depending on the environment where the RMU-1000
ACC is deployed, and ultimately the maintenance program is at the discretion of the emergency
response program's medical director. As with all rechargeable batteries, the rechargeable Battery
Pack self-discharges over time, and therefore, must be maintained in accordance with the User
Manual in order to be ready for use.
WeeklyAfter Each UseAction
••
••
••
••
•
••
Make sure the Battery Pack is fully charged (see Section 3.9, “Charging
the Battery Pack”).
Check the condition of the system. Make sure the Carrying Case contains
accessories including Patient Interface Pad(s), AC Adapter, Stabilization
Strap, and Patient Wrist Straps.
Make sure a Patient Interface Pad is installed on the Compression
Module.
Make sure at least one unused Patient Interface Pad is stored in the
Carrying Case.
Check the Battery Pack expiration date.
Turn the ACC on to perform a self-test. Make sure the piston is retracted
and the PAUSE indicator comes on with no warning indicators.
5. Maintenance and
Troubleshooting
Note: If the unit has been dropped, mishandled, or abused, a thorough evaluation of operation
should be performed.
5.2 Cleaning
After each use, clean any dirt or contaminants from the RMU-1000 ACC’s Frame, Backboard,
and Compression Module. The following are important guidelines that must be adhered to when
cleaning the device:
•
To clean the RMU-1000 ACC’s Frame, Backboard, or Compression Module, use a soft
cloth dampened with one of the following recommended cleaning agents:
– Soapy water
– Ammonia-based cleaners
– Hydrogen peroxide
– Isopropyl alcohol (70 percent solution)
– Chlorine bleach (30 ml/liter water)
• Do not immerse the RMU-1000 ACC’s Frame, Backboard, or Compression Module,
components in fluids or allow fluids to enter the unit.
• Do not spray cleaning solutions directly on the unit or its connectors.
• Do not use abrasive materials or strong solvents such as acetone or acetone-based
cleaning agents.
43
RA C- E1510 EN - DN
Cleaning (continued)
• After cleaning, allow the unit to completely dry. Before returning it to service, always turn
the unit on for a few seconds. If the unit detects a problem, the Warning Indicator will be
illuminated. Otherwise, turn the unit off.
• When necessary, clean the Stabilization Strap by removing both strap clips by loosening
the Velcro that holds the clips in place. Machine wash the Stabilization Strap and air dry (do
not tumble dry). Replace both strap clips after the strap has been cleaned and return the
Stabilization Strap to the ACC’s Carrying Case. Otherwise, discard the Stabilization Strap
and replace it with a new one.
• When necessary, clean the Patient Wrist Straps by detaching them from the RMU-1000
ACC and loosening all Velcro. Machine wash and air dry (do not tumble dry). When dry,
Defibtech recommends re-affixing the Patient Wrist Straps to the ACC’s Frame (see
Section 3.5). If the Patient Wrist Straps cannot be cleaned, are damaged, or the Velcro is
worn out, dispose of them and replace with new ones.
Please note that none of the items provided with the RMU-1000 ACC are sterile or require sterilization.
Do not sterilize the RMU-1000 ACC or its accessories.
WARNING
5.3 Storage
The RMU-1000 ACC should be stored in its Carrying Case and placed in a readily accessible
location. In general, the unit should be stored in clean, dry and moderate temperature conditions.
Make sure that the environmental conditions of the storage location are within the ranges detailed
in the “Environmental” section in Chapter 7 of this manual.
5.4 Troubleshooting
44
RA C- E1510 EN - DN
Note: If stored outside of the Compression Module, a Battery Pack in active mode (see Section
3.8 for details about battery states) should be recharged to full capacity at least every 6 months.
If stored in the Compression Module, the Battery Pack should be recharged to full capacity at
least every 2 months.
The following table lists the symptoms, the possible causes, and the possible solutions for
common problems. Refer to the other sections of the user manual for detailed explanations on
how to address any issues. If the unit continues to be non-functional, refer the unit for servicing.
(Refer to Chapter 10 of this manual for contact information.)
Charge depleted Battery Pack (see Section 3.9) or
replace depleted Battery Pack with a charged Battery
Pack (see Section 3.8).
Remove ACC from patient (see Section 4.9) and start
manual chest compressions as soon as possible.
Adjust the Piston (see Section 4.3).
If performing a rescue, replace Battery Pack with a
charged Battery Pack or connect external power (see
Section 4.7). Otherwise, charge Battery Pack as soon
as possible (see Section 3.9).
If performing a rescue, replace Battery Pack with a
charged Battery Pack or connect external power (see
Section 4.7). Otherwise, charge Battery Pack as soon
as possible (see Section 3.9).
Transfer Battery Pack from ship mode to active
mode. See“Ship Mode” in Section 3.8,“Installing and Removing the Battery Pack”, for more information.
Replace Battery Pack with a charged Battery Pack (see
Section 3.8).
If performing a rescue, replace Battery Pack with a
charged Battery Pack or connect external power (see
Section 4.7). Otherwise, charge Battery Pack as soon
as possible (see Section 3.9).
Replace Battery Pack with a new Battery Pack (see
Section 6 for ordering information; see Section 3.8 for
installation instructions).
Transfer Battery Pack from ship mode to active mode
(see“Ship Mode” in Section 3.8,“Installing and Removing the Battery Pack”, for more information).
Check for proper Piston position and height. Press
Pause button to clear the condition and try again
(see Section 4.3). If condition persists, remove ACC
from patient (see Section 4.9) and start manual chest
compressions as soon as possible. Note: Pressing the
Warning Mute button (see Section 4.1) will silence the
audible alert for 1 minute.
Push the On/Off Button for one second to power
off the ACC. Retry Operation and Adjustment (see
Section 4.3). If condition persists, remove ACC from
patient (see Section 4.9) and start manual chest
compressions as soon as possible.
Push the On/Off Button for one second to power off the
ACC. Remove ACC from patient (see Section 4.9) and
start manual chest compressions as soon as possible.
ACC will perform a rescue properly (Note: If the
Warning Indicator is flashing while the Service
Indicator is lit, the ACC may not perform a rescue
properly). Have ACC serviced as soon as practical
(see Chapter 10 for contact information).
ACC immediately turns off
Adjust Up/Down Buttons
blinking
Battery Pack Indicator on
User Control Panel is red
Battery Pack Indicator on
User Control Panel is
slowly flashing red
Battery Pack Indicator on
User Control Panel is
quickly flashing red
First LED of Battery Pack
Indicator on User Control
Panel is amber
Battery Pack Indicators
on Battery Pack do not
illuminate after button on
Battery Pack is pressed
Warning Indicator flashing;
audible alert (beeping);
compressions stop (if
running)
ACC fails to perform
compressions
Compressions are not
sufficient, or something
unusual occurs during
operation
Service Indicator is lit and
Warning Indicator is NOT
flashing
Battery Pack is depleted
Battery Pack malfunctionReplace Battery Pack (see Section 3.8).
ACC malfunction
Piston position adjustment
required
Battery Pack has less than
20% charge remaining
Battery Pack has less than
10% charge remaining
Battery Pack is not installedInstall a charged Battery Pack (see Section 3.8).
Battery Pack is in ship
mode
Battery Pack malfunction
Battery Pack depleted
Battery Pack has reached
end of life
Battery Pack is in ship
mode
Battery Pack depletedCharge Battery Pack as soon as possible (see Section 3.8).
The ACC has detected a
problem
An operator or ACC error
has occurred
ACC malfunction
ACC requires periodic
service (see Section 5.7 for
details) and/or has detected
a problem that should be
serviced
5. Maintenance and
Troubleshooting
If the error condition persists or service is required, call your Authorized Distributor or Defibtech.
Refer to Chapter 10 of this manual for contact information.
RA C- E1510 EN - DN
45
5.5 USB Port
The USB Port is located on the bottom of the Compression Module. It allows the Module to be
connected to a personal computer and for ACC data retrieval and event reporting when used in
conjunction with utility software available at www.defibtech.com. It is not intended to be used
during rescue operation.
USB Port
Do not use the USB cable during emergency use
WARNING
as it interferes with patient care.
5.6 Event Data
As part of Defibtech’s on-going regulatory compliance activities, event data shared with Defibtech
may be used by Defibtech to fulfill regulatory obligations. Any identifying personal data or health
information received is considered confidential within Defibtech and will not be used for any other
purpose. Please contact Defibtech at support@defibtech.com should there be any further questions.
5.7 Service
5.8 Recycling Information
Recycling Assistance
Preparation For Recycling
Packaging For Recycling
46
RA C- E1510 EN - DN
The RMU-1000 ACC contains no user serviceable parts. Defibtech recommends periodic
maintenance every 18 months of use. After approximately 200 hours of operation (based upon
performing 100 compressions per minute), the Service Indicator will flash to indicate that the unit
requires maintenance. If the unit needs servicing, call your Authorized Distributor or Defibtech.
Refer to Chapter 10 of this manual for contact information.
At the end of useful life, recycle the ACC and its accessories.
For recycling assistance contact your local Defibtech distributor.
Recycle in accordance with local and national regulations.
Items should be clean and contaminant-free prior to being recycled.
When recycling used items, follow local clinical procedures.
Packaging should be recycled in accordance with local and national requirements.
6 RMU-1000 ACC Accessories
This chapter describes the component parts and the accessories that can be used with the
Defi btech RMU-1000 ACC. For contact information on obtaining replacement component parts
and accessories, refer to Chapter 10 in this manual.
6.1 Patient Interface Pad
The Patient Interface Pad is a user-replaceable, single-use component that provides the interface
between the piston and the patient’s chest.
Top Bottom
6.2 Battery Pack
The Battery Pack provides a replaceable primary power source for the Compression Module.
6.3 External AC Adapter
The external AC Adapter provides external power to run the ACC and charge the installed Battery Pack.
Note: A Battery Pack must be installed to operate the ACC using an external power source.
6. RMU-1000 ACC
Accessories
47
RA C- E1510 EN - DN
6.4 Stabilization Strap
The Stabilization Strap is used to stabilize the RMU-1000 ACC while compressions are being
applied by the unit to the patient (see Section 4.4 “Stabilization” for details).
6.5 Patient Wrist Straps
The Patient Wrist Straps attach a patient’s arms to the RMU-1000 ACC for ease of transporting
the patient and the ACC (see Sections 3.5 and 4.5 for details).
6.6 USB Cable
6.7 Battery Pack Charging Station
48
RA C- E1510 EN - DN
An optional USB Cable (USB-A to Mini-B) allows the RMU-1000 ACC’s Compression Module to
be connected to a personal computer and for ACC data retrieval and event reporting when used in
conjunction with utility software available at www.defibtech.com. The ACC's USB port is located
on the bottom of the Compression Module (see Section 5.5 “USB Port” for details). For more
information, contact Defibtech or your authorized distributor (see Chapter 10, “Contacts” ).
The Battery Pack Charging Station is an optional accessory that charges up to two Battery Packs
simultaneously. For more information, visit www.defibtech.com or contact Defibtech or your
authorized distributor (see Chapter 10, “Contacts”).
6.8 Tactical Case
A compact Tactical Case is available as an optional accessory. For more information, visit
www.defibtech.com or contact Defibtech or your authorized distributor (see Chapter 10,
“Contacts”).
Accessories
49
RA C- E1510 EN - DN
6. RMU-1000 ACC
50
RA C- E1510 EN - DN
7 Technical Specifications
7.1 Defibtech RMU-1000 ACC
General
CategorySpecification
Size (assembled)23.5 x 20.75 x 9 inches (59.7 x 52.7 x 22.9 cm)
Size (in carrying case)24 x 18 x 10 inches (61.0 x 45.7 x 25.4 cm)
Weight (with Battery Pack)15.9 lbs (7.1 kg)
PowerRechargeable Battery Pack or 24V DC input
Meets applicable requirements of
Design standards
Device classification
Patient and CPR
CategorySpecification
Patient
Compression depth
Compression frequency101 ±1 compressions per minute
Compression duty cycle50% ±5%
Compression modes
• IEC 60601-1
• UL 60601-1
• CAN/CSA C22.2 60601-1
• IEC 60601-1-2
Internally powered
Class II (with external power source)
Adult patients that fit into the ACC
• Chest width – 18 inches (45.7 cm) maximum
• Chest height – 6.5 to 11.8 inches (16.5 to 30 cm)
Use of the RMU-1000 ACC System is not restricted by patient weight.
2.1 inches ±0.1 inches (5.3 cm ±0.3 cm)
from Start Position (nominal patient)
• Continuous compressions
• Compressions with breathing (30:2, 30 compressions with 3-second pause
for ventilation) factory default; future protocols via field updates
Less than 3 hours in ACC*
Less than 2 hours if charging one Battery Pack in optional external battery
pack charging station (less than 3 hours if charging two Battery Packs;
see Section 6.7 for details)*
Recommended to replace Battery Pack every 3 years or if Battery
Pack indicator displays a replace Battery Pack condition (~300 charge/
discharge cycles**).
0 – 40°C (32 – 104°F);
-20 – 60°C (-4 – 140°F) short-term <1 month.
**one charge/discharge cycle is defined as charging
and discharging the full capacity of the Battery Pack
*typical, new battery, at 25°C
7.3 AC Power Adapter
RPM-1000 External AC Power Adapter
52
RA C- E1510 EN - DN
CategorySpecification
Model numberRPM-1000
Input voltage100 – 240VAC, 50/60Hz nominal
Input current≤2.3A
Rated output24.0VDC (±5%)
Operating / storage temperatureSee “Environmental” specifications chart in Section 7.1
Emissions and immunityRefer to Chapter 8 for details
7.4 Notice to European Union Customers
The crossed-out wheeled bin symbol indicates that this equipment has been
put on the market after 13 August 2005, and is included in the scope of the
directive 2002/96/EEC on Waste Electrical and Electronic Equipment (WEEE)
and of the national decree(s), which transpose provisions of such directive.
At the end of its lifetime, this equipment can only be disposed of in compliance with the provisions
of the above-mentioned European directive (and as amended) as well as with the corresponding
national regulations. Severe penalties are possible for unauthorized disposal.
Electrical and Electronic Equipment (EEE) may contain polluting components and hazardous
substances, the accumulation of which could pose serious risk for the environment and human
health. It is for this reason that local administrations provide regulations, which encourage reuse
and recycling, and prohibit the disposal of WEEE as unsorted municipal waste and require the
collection of such WEEE separately (at specifically authorized treatment facilities). Manufacturers
and authorized distributors are required to supply information about a safe treatment and
disposition of the specific equipment.
You may also return this equipment to your distributor when purchasing a new one. As for reuse
and recycling, notwithstanding the limits imposed by the nature and the use of this equipment, the
manufacturer will do its best to develop recovery processes. Please contact the local distributor
for information.
7. Technical Specifications
53
RA C- E1510 EN - DN
54
RA C- E1510 EN - DN
8 Electromagnetic Conformity
8.1 Guidance and Manufacturer’s Declaration
The essential performance of the RMU-1000 ACC and its accessories is to provide accurate chest
compression depths and rates.
The RMU-1000 ACC is intended for use within the electromagnetic environment specified below.
The customer or the user of the RMU-1000 ACC should assure that it is used within the stated
environmental specifications.
RF emissionsThe RMU-1000 uses RF energy only for its internal
CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Group 1
Class B
Class A
Complies
Electromagnetic Immunity
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The RMU-1000 is suitable for use in all buildings including
domestic homes and places directly connected to the
public low-voltage Power Supplies Network that supplies
buildings used for domestic purposes.
8. Electromagnetic
Conformity
Immunity testIEC 60601 test levelCompliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
±6 kV contact
±8 kV air
±2 kV for power line
supply lines
±1 kV for input/output
lines
±2 kV±2 kV
Ut = 0%, 0.5 cycle
Ut = 0%, 1 cycle
Ut = 70%, 25/30 cycles
Ut = 0%, 250/300 cycles
30 A/m30 A/m
±8 kV contact
±15 kV air
±2 kV for power line
supply lines
±1 kV for input/output
lines
Ut = 0%, 0.5 cycle
Ut = 0%, 1 cycle
Ut = 70%, 25/30 cycles
Ut = 0%, 250/300 cycles
Electromagnetic environment – guidance
There are no special requirements with
respect to electrostatic discharge.
The mains power quality must be that of a
typical commercial or hospital environment.
The mains power quality must be that of a
typical commercial or hospital environment.
The mains power quality must be that of a
typical commercial or hospital environment.
If the user of the [Equipment or System]
requires continued operation during
power mains interruptions, Defibtech
recommends that the [Equipment or
System] is energized from a Power Supply
or Battery Pack that cannot be interrupted.
Power frequency magnetic fields should
not be greater than levels characteristic of a
typical location in a commercial or hospital
environment.
55
RA C- E1510 EN - DN
Electromagnetic Immunity (continued)
Immunity testIEC 60601 test levelCompliance level
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz
to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the RMU-1000 is used exceeds the applicable RF compliance level above,
the RMU-1000 should be observed to verify normal operation. If abnormal performance is observed additional measures
may be necessary, such as reorienting or relocating the RMU-1000.
10 V/m
80 MHz to 2.7 GHz
10 Vrms
150 KHz to 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
20 V/m
10 Vrms
6 Vrms (ISM)
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer
to any part of the RMU-1000, including
cables, than necessary. The recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter is shown in the following table.
Interference may occur in the
vicinity of equipment marked
with this symbol.
56
RA C- E1510 EN - DN
Guidance and Manufacturer's Declaration (continued)
Separation Distances
The RMU-1000 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the RMU-1000 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the RMU -1000 as recommended below, according
to the maximum output of the communications equipment.
Recommended separation distances between portable
and mobile RF communications equipment and the RMU-1000
Separation distance according to frequency of transmitter (m)
Rated maximum
output power of
transmitter (W)
0.010.120.23
0.10.380.73
11.202.30
103.797.27
10012.0023.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: As 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;
13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the
ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease
the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P
8. Electromagnetic
Conformity
57
RA C- E1510 EN - DN
Guidance and Manufacturer's Declaration (continued)
The use of accessories other than those specified may result in
WARNING
WARNING
WARNING
CAUTION
increased emissions or decreased immunity of the RMU-1000 ACC.
Portable and mobile RF communications equipment should be used
no closer to any part of the RMU-1000 ACC, including cables, than
necess ar y.
The RMU-1000 ACC should not be used adjacent to other equipment
and that if adjacent, the RMU-1000 ACC should be observed to verify
normal operation in the configuration in which it will be used.
The mains power quality must be that of a typical commercial or
hospital environment.
58
RA C- E1510 EN - DN
9 Glossary of Symbols
SymbolMeaning
ON/OFF Button
• Turns the device on or off (hold button for at least one second).
Adjust Down Button
• Drives the piston down toward the patient.
Adjust Up Button
• Retracts the piston up away from the patient.
Run Continuous Compressions Button
• Performs compressions until the Pause or Off button is pressed.
Run Compressions with Breaths Button
• Performs compressions according to the compressions-with-breaths protocol and pauses for the
operator to give breaths.
Pause Button
• Stops compressions when running (or resumes compressions when paused).
Warning Indicator
• Flashes to alert the user that the ACC has determined that there is a problem.
Warning Mute Button
• Silences audible sound associated with the Warning Indicator.
Service Indicator
• The Service Indicator illuminates when the ACC requires servicing.
Battery Pack Indicator
• Indicates the approximate remaining Battery Pack capacity.
Check Battery Pack Charge Button (on Battery Pack)
• Checks the charge of the Battery Pack when it is not installed in the ACC.
Connection point for AC Adapter. Only use the AC Adapter supplied with the Defibtech ACC.
USB Port.
9. Glossary of Symbols
59
RA C- E1510 EN - DN
YYYY
YYYY
YYYY-MM-DD
Glossary of Symbols (continued)
SymbolMeaning
Caution, consult accompanying documents.
Pinch point. Keep hands and fingers clear.
Do not expose to high heat or open flame. Do not incinerate.
Recyclable.
Consult operating instructions.
Refer to instruction manual / booklet.
Do not damage or crush.
60
RA C- E1510 EN - DN
YYYY-MM-DD
Follow proper disposal procedures.
Meets the requirements of the European Medical Device Directive.
Operational temperature limitation.
Use by yyyy-mm-dd.
Defibrillation proof - Can withstand the effects of an externally applied defibrillation shock. Internally
powered with defibrillator-proof BF-type patient applied parts (per EN 60601-1).
Manufacturer.
Date of manufacture.
Glossary of Symbols (continued)
YYYY
YYYY
YYYY
SymbolMeaning
Manufacturer and date of manufacture.
YYYY-MM-DD
Do not reuse.
3
EC REP
Quantity per box.
For USA users only.
Federal Law (USA) restricts this device to sale by or on the order of a physician.
Catalogue number.
Keep dry.
Keep away from sunlight.
Handle with care.
Transportation and storage requirements.
See environmental requirements on packaging.
Authorized European Representative:
EMERGO EUROPE
Prinsessegracht 20
2514 AP The Hague
The Netherlands
9. Glossary of Symbols
IP43
IP44
Not made with natural rubber latex.
Lot number.
Protected against solid objects over 1 mm; Protected against direct water sprays up to 60° from the
vertical. Refer to IEC 60529 for further information.
Protected against solid objects over 1 mm; Protected against water sprayed from any direction. Refer
to IEC 60529 for further information.
61
RA C- E1510 EN - DN
(XX)XXXXXXXXXXXXXX
Glossary of Symbols (continued)
SymbolMeaning
Conformity to product safety testing requirements.
Serial number.
(XX)XXXXXX
(XX)XXXXXXXXX
SAMPLE
18.25VDC, 96.5 Wh
5.3Ah, Li-Ion
5ICP20/38/65
Unique Device Identification (UDI) information.
(NOTE: Sample shown at left is for visual reference purposes only; actual UDI information specific to
this device appears on a physical label affixed to the unit’s components and/or its packaging.)
Lithium-ion battery.
Product is not sterile.
62
RA C- E1510 EN - DN
10 Contacts
Manufacturer
Defibtech, L.L.C.
741 Boston Post Road, Suite 201
Guilford, CT 06437 USA
Tel.: 1-(866) 333-4241 (Toll-free within North America)
Defibtech, L.L.C. provides a LIMITED WARRANTY that the
Automated Chest Compressor (ACC) (i.e., Backboard, Frame,
and Compression Module) and its associated components
(i.e., Battery Pack, AC Adapter, Stabilization Strap, Patient
Wrist Straps, and Patient Interface Pad (PIP)), and accessories
(e.g., Carrying Case, USB Cable) (collectively referred to as
“Products” and singularly as “Product”) whether purchased
concurrently or separately, shall be substantially free from
defects in material and workmanship appearing under normal
service and use. To qualify for WARRANTY SERVICE, the
Product must have been continuously owned by the original
purchaser and the original purchaser must have purchased the
Product from Defibtech or an authorized Defibtech retailer.
This LIMITED WARRANTY may not be assigned or transferred.
WARRANTY PERIOD
The Products shall have a WARRANTY PERIOD of one (1) year
beginning on the date of delivery. The WARRANTY PERIOD for
a single use Product (e.g., Patient Interface Pad ) and Products
having an expiration date shall end upon the earlier of use,
expiration (if applicable), or end of the WARRANTY PERIOD.
Any WARRANTY SERVICE, including but not limited to repair
or replacement shall not extend a Product’s WARRANTY
PERIOD.
LIMITED WARRANTY SCOPE
This LIMITED WARRANTY does not cover damage of any sort
resulting from, but not limited to, accidents, misuse, improper
storage, improper operation, alterations, unauthorized service,
tampering, abuse, neglect, fire, flood, war, or acts of God.
Misuse shall include but not be limited to: use of the ACC with
unapproved components; use of a Product with unapproved
devices; or use of the Product in uncertified environments or
settings. Defibtech does not warranty error-free or interruptionfree performance of any Product.
LIMITED WARRANTY VOIDED
The LIMITED WARRANTY is immediately voided if: the
Product is serviced or repaired by any entity, including persons,
not authorized by Defibtech; specified Product maintenance
is not performed; the Product is used with one, or more,
unauthorized components or devices; or the Product is not
used in accordance with Defibtech approved instructions.
WARRANTY SERVICE
At Defibtech’s sole discretion, Defibtech shall have the option
to repair, replace, or provide a credit. In the event of repair
or replacement, Defibtech shall have the right at its sole
discretion to replace the Product with a new, or refurbished,
same or similar Product, and all Products or parts replaced
shall become the property of Defibtech. Determination of a
similar Product shall be at the sole discretion of Defibtech. In
the case of replacement, the replacement at a minimum shall
reflect the prorated time remaining for the Product based on
the remaining WARRANTY PERIOD. In the case of a credit,
the credit shall be the prorated value of the Product based on
the lower of the original Product cost of the same or similar
Product and the remaining WARRANTY PERIOD. Repair or
replacement of a Product under this LIMITED WARRANTY
does not extend the Product’s WARRANTY PERIOD.
WARRANTY SERVICE (CONTINUED)
In order to obtain WARRANTY SERVICE, the original owner
must contact the Defibtech authorized retailer from whom the
Product was purchased, or Defibtech customer service. In the
event a Product must be returned, a Defibtech issued Return
Material Authorization (RMA) number is required. Products
returned without a Defibtech authorized RMA number will
not be accepted. The Product shall be shipped at the original
end user’s expense to a destination specified by the retailer
or Defibtech.
OBLIGATIONS AND WARRANTY LIMITS
THIS LIMITED WARRANTY IS THE SOLE AND EXCLUSIVE
WARRANTY FOR DEFIBTECH’S PRODUCTS AND IS
EXPRESSLY IN LIEU OF ANY OTHER EXPRESS OR IMPLIED
WARRANTIES INCLUDING WITHOUT LIMITATION ANY
WARRANTY AS TO MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. DEFIBTECH’S MAXIMUM LIABILITY
ARISING OUT OF THE SALE OF THE PRODUCT’S OR THEIR
USE, WHETHER BASED UPON WARRANTY, CONTRACT,
TORT OR OTHERWISE, SHALL NOT EXCEED THE ACTUAL
PAYMENTS RECEIVED BY DEFIBTECH IN CONNECTION
THEREWITH. DEFIBTECH SHALL NOT BE LIABLE FOR
ANY INCIDENTAL, SPECIAL OR CONSEQUENTIAL LOSS,
DAMAGE OR EXPENSE (INCLUDING WITHOUT LIMITATION
LOST PROFITS) DIRECTLY OR INDIRECTLY ARISING FROM
THE SALE, INABILITY TO SELL, USE OR LOSS OF USE OF
ANY PRODUCT (HOWEVER CAUSED AND ON ANY THEORY
OF LIABILITY), EVEN IF DEFIBTECH HAS BEEN ADVISED
OF THE POSSIBILITY OF SUCH LOSS. THE FOREGOING
LIMITATION SHALL NOT APPLY TO ANY CLAIMS FOR BODILY
INJURY OR DEATH TO THE EXTENT THAT LIMITATION
OF DAMAGES FOR SUCH CLAIMS IS UNENFORCEABLE
OR AGAINST PUBLIC POLICY UNDER ANY APPLICABLE
STATUTE OR RULE OF LAW.
ANY LEGAL ACTION ARISING FROM THE PURCHASE OR
USE OF A PRODUCT SHALL BE COMMENCED WITHIN ONE
YEAR FROM THE ACCRUAL OF THE CAUSE OF ACTION, OR
BE BARRED FOREVER. IN NO EVENT SHALL DEFIBTECH’S
LIABILITY UNDER THIS WARRANTY EXCEED THE PURCHASE
PRICE OF THE PRODUCT.
NO PERSON (INCLUDING ANY AGENT, DEALER, OR
REPRESENTATIVE OF DEFIBTECH) IS AUTHORIZED
TO MAKE ANY REPRESENTATION OR WARRANTY
CONCERNING THE PRODUCTS, EXCEPT TO REFER TO THIS
LIMITED WARRANTY.
If any part or term of this LIMITED WARRANTY is held to be
illegal, unenforceable or in conflict with applicable law by any
court of competent jurisdiction, the validity of the remaining
portions of the LIMITED WARRANTY shall not be affected.
The user may also have other rights that vary from state to
state or country to country.
11. Warranty Information
65
RA C- E1510 EN - DN
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.