The PosiTector CMM IS concrete moisture meter measures the
relative humidity and temperature in concrete slabs using in situ
probes, prior to the application of flooring materials or coatings.
When applying flooring materials or coatings to concrete, it is critical
to ensure that the moisture level inside the concrete has decreased
to acceptable levels. If not, moisture can accumulate below the
flooring or coating, causing discoloration, cupping, buckling,
blistering, adhesive failure and mold growth.
Designed in accordance with ASTM F2170, the PosiTector CMMIS provides a complete measurement solution for monitoring the
drying process of concrete slabs in compliance with industry
standards. The unique design of the PosiTector CMM IS probes
allow the user to place them into concrete immediately after the
holes are drilled and leave them in place until they reach
equilibrium, eliminating the need to reopen the hole at a later date.
PosiTector CMM IS probes broadcast measurements to the
PosiTector App (pg. 9) or PosiTector Advanced instruments
(pg. 11) using Bluetooth wireless technology.
The PosiTector CMM IS is available in 3 kit configurations:
PosiTector CMM IS Expansion Packs are available when
additional probes are required. A single expansion pack consists
of the following:
l PosiTector CMM IS probel Cap
l Calibration Check Chamberl (2) A-76 / LR-44 batteries
l Saturated Salt Solution (NaCl) l (2) Stackable Probe
Extensions
Calibration
Check
Chamber
PosiTector CMM IS probe
Stackable Probe
Extensions
Batteries
NOTE:
For instructions on how to use the PosiTector CMM IS
Fins
Cap
and the PosiTector App in accordance with ASTM F2170, refer
to Setting Up a New ASTM F2170 Project (pg. 6).
Quick Start
PosiTector CMM IS probes are designed for use with the
PosiTector App or a PosiTector Advanced Gage.
Power-up PosiTector CMM IS probes by unscrewing the battery
cover and inserting the battery (positive “+” facing up). Replace
the battery cover. The probe will begin broadcasting and is now
visible to the PosiTector App and PosiTector Advanced
instruments.
To power down the PosiTector CMMIS probe, unscrew battery cover and
remove battery using a pointed tool,
such as the probe extraction tool.
Viewing Readings from PosiTector CMM IS probes
PosiTector CMM IS measurement values can be viewed using
the PosiTector App or a PosiTector Advanced instrument. To
record measurement values, see How to Measure (pg. 7)
Using the PosiTector App:
Install the PosiTector App from the App Store (Apple iOS) or
Google play (Android).
Open the PosiTector App. Available probes within 10 m (30 feet)
will appear as “CMMIS” with the corresponding serial number, as
shown below:
The small circle icons to the left of each PosiTector CMM IS
probe display the most recent Relative Humidity reading. This
value is updated every five minutes.
To view measurements from the
PosiTector CMM IS probe, select it from
the search screen. Once connected, a
blue light is emitted from the bottom of the
probe and values are updated every five
seconds and displayed as shown to the
right. The battery indicator in the top left
corner shows the battery level of the
PosiTector CMM IS probe.
To view available probes, select Bluetooth Smart from the gage’s
Connect menu. Then select Smart Probes (shown below):
Available probes within 10 m (30 feet) will appear as “CMMIS”
with the corresponding serial number and latest relative humidity
reading, as shown below:
Checking the
Shortcut box
on the Smart
Probes screen
will enable
access to the
Smart Probes
screen from
the gage’s
main menu:
Select a probe to display its Relative Humidity and Temperature
values. Once connected, a blue light is emitted from the bottom
of the probe and values are updated every five seconds.
NOTE:
When a PosiTector DPM environmental probe is
attached to the PosiTector body, PosiTector CMM IS
measurement data can be recorded into the gage’s memory.
When any other probe is attached, such as a PosiTector 6000
probe, PosiTector CMM IS readings can be viewed only.
Calibration & Verification of Accuracy
Calibration
Calibration is typically performed by the manufacturer or a
qualified lab. All PosiTector CMM IS probes include a Certificate
of Calibration. In accordance with ASTM F2170, probes must be
recalibrated annually. DeFelsko recommends beginning the
calibration interval from either the date of calibration, date of
purchase, or date of receipt.
Verification of Accuracy
It is recommended that PosiTector CMM IS probes be stored
within the Calibration Check Chamber when not in use. The
Calibration Check Chamber and Saturated Salt Solution are
designed in accordance with ASTM E104 to maintain a constant
relative humidity. Per ASTM F2170, the Calibration Check
Chamber and Saturated Salt Solution maintain a constant relative
humidity of 75% (at 25° C) and keeps the probe in a ready-to-use
condition.
ASTM F2170 requires calibration to be verified within 30 days
prior to use. Power-up the probe by inserting a battery, positive
(+) side up. Probe calibration is verified by inserting a probe into
the Calibration Check Chamber, connecting to either the
PosiTector App or a PosiTector Advanced gage body and
taking a measurement (after allowing the probe at least 1 hour to
acclimate to the present environment). Probes should report a
relative humidity within 2% of the humidity value listed in Table 1
for the corresponding temperature. To simplify the verification
process, we recommend using the Cal Check function of the
PosiTector App or PosiTector Advanced. The Cal Check
function automatically determines whether a probe is reading
within tolerance.
Probes should be allowed to acclimate for at least 24 hours within
the Calibration Check Chamber and Soluble Salt Solution before
calibration is checked.
It is important to maintain a constant temperature within the
Calibration Check Chamber. Relative humidity variation due to
temperature fluctuation for the Calibration Check Chamber is
shown in Table 1.
Table 1
Extracted from Table A1.1,
Equilibrium Relative Humidity Values
for Selected Saturated Aqueous Salt
Solutions (ASTM E104)
At 25° C, the Soluble Salt Solution relative humidity range is
75.3% ± 0.2. Including the ±2% relative humidity tolerance
specified in ASTM F2170, a probe would be considered to be in
tolerance if it measured between 73.1 – 77.5 %.
NOTE:
When the Saturated Salt Solution (Boveda) is completely
dry, it will solidify. They should be replaced while there is still
moisture present and before they are completely solid.
PosiTector CMM IS probes, coupled with the PosiTector App or
PosiTector DPM Advanced simplify the process of viewing,
recording and reporting in accordance with ASTM F2170.
Step 1. Evaluate the location to determine how many tests must
be performed.
ASTM F2170 specifies that 3 tests must be performed
for the first 100 m2 (1000 ft2) of concrete and one
additional test for each 100 m2 (1000 ft2) of concrete.
One test must be performed within 3 ft. (1 m) of each
exterior wall.
Step 1b (if using the PosiTector App). The PosiTector App
makes it easy to view, record and report in accordance with ASTM
F2170. The app is available within the App Store (Apple iOS) and
Google play (Android).
Open the PosiTector App and select F2170 from the top
menu bar. Next, select New Project and enter the job
details.
A PosiTector DPM Advanced instrument or spare
PosiTector CMM IS probe can be used to record ambient
temperature as required by ASTM F2170. Under Ambient
Instrument, select an available PosiTector CMM IS probe
or manually enter data from a PosiTector DPM
or.To enable broadcasting, ensure the PosiTector Advanced is powered on (Bluetooth Smart enabled) or
the battery in the PosiTector CMM IS probe is installed.
Step 1c (PosiTector App only). If desired, insert a picture or
blueprint of the area to be measured. Use the to take a
photo of the blueprint, or use the to insert an image from
the photo gallery. This background image will make it easy to
locate each probe within the area.
Step 1d (PosiTector App only). Press the (+) icon at the bottom
right of the Project screen and then touch the screen in the
desired location to place the hole. The new hole location will be
overlayed on the image and will be identified as a circle containing
--- 0. Repeat this for each measurement location within the Area.
Create additional Areas as required.
Step 2 – Drill the Hole
ASTM F2170 sets out strict requirements for hole depth and
diameter. Using a rotary hammer drill with a carbide-tipped
masonry drill bit (¾” diameter), drill the hole to the required depth.
Do not use water for cooling or lubrication.
Ensure any dust and debris are removed from the hole. The
vacuum attachment included with the Complete Kit and Pro Kit
is ideal for this purpose. It can also be purchased separately.
Per ASTM F2170, check the diameter and depth of the hole using
the included tape measure. If either dimension is incorrect, the
hole must be re-drilled and cleaned. Incorrect hole diameter may
cause difficulty inserting or removing a PosiTector CMM IS
probe.
If using the PosiTector App, touch the circle corresponding to the
hole, and record the location, depth and date/time the hole was
drilled. Select Set Now to input the current time and date, or input
it manually.
Step 4 – Insert the Probe
PosiTector CMM IS probes are designed to be placed into the
hole as soon as the hole has been prepared. Remove the probe
from the Calibration Check Chamber and Saturated Salt Solution,
unscrew the battery cover and insert the battery, (+) positive side
up. Replace the cover.
The probe will immediately begin broadcasting readings via
Bluetooth, and can be viewed from the PosiTector App (pg. 9)
or a PosiTector Advanced (pg. 11).
The probe is now ready to be inserted into a hole. For a 4” slab
drying from one side a 1.6” hole is drilled and the PosiTectorCMM IS probe can be inserted without an extension. For deeper
holes, the included stackable ‘probe’ extensions (pg. 2) should
be used. Use one extension for each additional 1 inch (25 mm)
in concrete slab thickness (slabs drying from top only).
Push the probe completely to the bottom of the hole. Cover the
hole using the included yellow plastic cap. Per ASTM F2170, the
probe must remain in the hole for 24 hours to achieve moisture
equilibrium before measurements are recorded.
If using the PosiTector App, touch the circle corresponding to the
hole, and select the serial number of the inserted probe to
associate that probe with the hole.
Record the date and time each probe is inserted by selecting SetNow to automatically set the current time and date, or input it
manually.
NOTE:
A new battery will continuously power the probe for up to
3 weeks.
Step 5 – Record Measurements
ASTM F2170 requires waiting 24 hours, before probe
measurements are recorded. There is no need to remove the
yellow cap to access the probes in the holes. The probes remain
powered-on and broadcasting while in situ.
To Record Measurements Using the PosiTector App:
1. Open the PosiTector App and select F2170 from the top
3. Record measurements for each hole in the project:
To record probe measurements, touch the circle
corresponding to the hole and touch the Measure button to
record the Relative Humidity and Temperature within the
corresponding hole.
Ambient Relative Humidity and Air Temperature (above the hole)
can be recorded using a spare PosiTector CMM IS probe or
PosiTector DPM Advanced instrument. The ambient instrument
being used is specified within the Project setup (pg. 6). Select
Measure, and then Ok to store the displayed values. Alternately,
a relative humidity and temperature value can be entered
manually.
Repeat Step 3 for each hole in the project. Once readings have
been recorded for all holes in the project, professional reports can
OTE:
be generated that conform to ASTM F2170. Select REPORT from
within the Project to automatically generate the report.
To Record Measurements Using a PosiTector DPMAdvanced:
1. Access the Connect Menu within the gage menu.
2. Under Bluetooth Smart, select Smart Probes (shown
below). Available PosiTector CMM IS probes will be visible.
Select the probe to be recorded.
3. The relative humidity and temperature measured by the
PosiTector CMM IS probe will be displayed below the ambient
relative humidity, air temperature and surface temperature
measured by the PosiTector DPM probe, as shown below:
To record, ensure a batch is open in memory and press the (+)
button on the gage. The displayed readings of the PosiTector DPM
probe and the PosiTector CMM IS probe will be recorded into the
gage’s memory. If a batch is not currently open, select New Batch
from the gage’s Memory menu. To add measurements to an
existing batch, select Open from the gage’s Memory menu.
Measurements from multiple PosiTector CMM IS probes can be
recorded within the same batch by repeating the above steps for
each probe in the project.
If any probe other than the PosiTector DPM probe is
attached to the PosiTector Advanced body while connected to
a PosiTector CMM IS probe, the gage body will ignore the
attached probe and only display the PosiTector CMM IS
measurements.
Recorded measurement values can be reported using PosiSoft
Solutions. See the PosiTector DPM Instruction Manual for more
information.
Step 6 – Remove the PosiTector CMM IS Probes
The PosiTector CMM IS probes are designed to be reusable.
Once measurements are complete, use the extraction tool to
remove the probes from the holes. Clean any visible debris or
concrete dust from the probes and inspect the silicone fins. If the
silicone fins are damaged or worn, they should be replaced. If no
further tests are required, remove the batteries by unscrewing the
battery covers and pushing the battery out using the probe
extraction tool and place each probe into a Calibration Check
Chambers for storage.
Once testing is complete, the holes should be filled with a
cementitious patching compound in accordance with ASTM
F2170.
The PosiTector CMM IS Probe does not appear in the
PosiTector App or in the PosiTector Advanced Gage Body
after the battery is inserted
Check the battery in the probe and replace if required. Ensure
Bluetooth is enabled on the Smart device running the PosiTector
App or Bluetooth Smart within the PosiTector body.
RH value is slow to stabilize
If the probe is exposed to very low humidity for an extended time
period and then placed in a high humidity environment, the
displayed Relative Humidity (RH) may appear to respond slowly.
Allow the probe to stabilize at least 60 minutes for a change of
RH greater than 50%. When placed into the Calibration Check
Chamber, allow at least 24 hours for the relative humidity to
stabilize.
Returning for Service
For issues related to the PosiTector CMM IS probes:
1. Install new batteries in the proper alignment as shown within
battery compartment.
2. Examine the sensor for debris or damage.
3. Re-attempt measurement.
For Issues related to the PosiTector Advanced Gage Body:
1. Perform a Hard Reset as follows:
A. Power down the Gage and wait 5 seconds.
B. Simultaneously press and hold the (+) and center
buttons until the Reset symbol appears.
WARNING:
This returns the Gage to a known, “out-of-the-box” condition.
All readings in memory will be erased.
2. Update your PosiTector gage body and re-attempt
measurements.
If these steps do not resolve the issue, visit our Service & Support
page for information on sending your Gage to us for a no charge
evaluation.