Decagon Devices WP4C Operator's Manual

WP
4C
Dewpoint PotentiaMeter
Operator’s Manual
Version 1
Decagon Devices, Inc.
Pullman WA 99163
tel: (509) 332-2756
fax: (509) 332-5158
www.decagon.com
support@decagon.com
Trademarks
“WP4C PotentiaMeter” is a
registered trademark of Decagon Devices, Inc.
©2010 Decagon Devices, Inc., All rights reserved.
WP4C PotentiaMeter
Table of Contents
i
Contents
ContentsContents
Contents
1. Introduction . . . . . . . . . . . . . . . . . .1
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Note to our WP4C Users . . . . . . . . . . . . . . . . . . . . . . . 2
Seller’s Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. About the WP4C . . . . . . . . . . . . . .4
Instrument Specifications . . . . . . . . . . . . . . . . . . . . . . . 4
WP4C and Water Potential . . . . . . . . . . . . . . . . . . . . 5
How the WP4C works . . . . . . . . . . . . . . . . . . . . . . . . . . 5
WP4C and Temperature . . . . . . . . . . . . . . . . . . . . . . . . 6
3. Getting Started . . . . . . . . . . . . . . .7
Components of your WP4C . . . . . . . . . . . . . . . . . . . . 7
Choosing a Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Preparing the WP4C for Operation . . . . . . . . . . . . 9
Portability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4. The Menus . . . . . . . . . . . . . . . . . . .12
The Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Changing Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Reading Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Sample Equilibration Screen . . . . . . . . . . . . . . . . . . . . 20
WP4C PotentiaMeter
Table of Contents
ii
5. Calibration and Verification . . 21
Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Verification Standards . . . . . . . . . . . . . . . . . . . . . . . . . . 21
When to Verify Calibration . . . . . . . . . . . . . . . . . . . . 22
How to Verify and Calibrate the WP4C . . . . . . . 22
6. Sample Preparation . . . . . . . . . . 25
Choosing a Sample Cup . . . . . . . . . . . . . . . . . . . . . . . . . 25
Preparing the Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Dry Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Samples and Temperature . . . . . . . . . . . . . . . . . . . . . . 27
Measuring Plant Samples . . . . . . . . . . . . . . . . . . . . . . . 29
7. Taking a Reading . . . . . . . . . . . . 30
How WP4C takes Readings . . . . . . . . . . . . . . . . . . . . 31
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8. Computer Interface . . . . . . . . . . . 33
9. Theory: Water Potential . . . . . . 35
Water Potential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring Water Potential . . . . . . . . . . . . . . . . . . . . . 35
Effect of Temperature on Water Potential . . . . . 37
Estimating Osmotic Potential . . . . . . . . . . . . . . . . . . . 38
10. Cleaning and Maintenance . . . 39
Tools Needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Accessing the Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Cleaning Procedure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Checking Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
WP4C PotentiaMeter
Table of Contents
iii
11. Repair Instructions . . . . . . . . . . . .45
Shipping Directions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Repair Costs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Loaner Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
12. Troubleshooting . . . . . . . . . . . . .48
Problems and Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Sensor Performance Screen . . . . . . . . . . . . . . . . . . . . . 54
13. Further Reading . . . . . . . . . . . . . .56
References: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Application Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Appendix A . . . . . . . . . . . . . . . . . . . .58
Preparing Salt Solutions . . . . . . . . . . . . . . . . . . . . . . . . . 58
Index . . . . . . . . . . . . . . . . . . . . . . . . . .61
WP4C PotentiaMeter
Table of Contents
iv
WP4C PotentiaMeter
1. Introduction
1
1. Introduction
Welcome to Decagon’s WP4C Dewpoint PotentiaMeter,
the research standard for measuring water potential. WP4C is the fastest, most accurate, and most reliable instrument available for measuring water potential using the chilled-mirror dewpoint technique. Whether you are a scientist or a student, WP4C will suit your needs. It is easy to use and provides fast, accurate results. We hope you find this manual informative and helpful in understanding how to maximize the capabilities of your WP4C.
About this Manual
Included in this manual are instructions for setting up your WP4C, calibration, and maintaining and caring for your instrument. Please read these instructions before operating WP4C to ensure that the instrument performs to its full potential.
Customer Service
If you ever need assistance with your WP4C, or if you just have questions, there are several ways to contact us:
NOTE: If you purchased your WP4C through a distributor, please contact them for assistance.
WP4C PotentiaMeter
1. Introduction
2
Phone:
Our toll-free customer service number is available to our customers in the US, Monday through Friday, between 8 a.m. and 5 p.m. Pacific time at 1-800-755-2751. For our customers outside of the US, our regular telephone num-
ber is (509) 332-2756.
Fax:
Our fax number is (509) 332-5158. When you fax us,
please include your WP4C’s serial number, your name, address, phone and fax number along with a description of your problem.
E-mail:
If you need technical support or have questions about your application with WP4C, you can send us messages via
e-mail at support@decagon.com. Again, please include
as part of your message your WP4C’s serial number, your name, address, phone, fax number, and return e-mail address.
Warranty
WP4C has a 30-day satisfaction guarantee and a one-year warranty on parts and labor.
Note to our WP4C Users
This manual is written to aid the end user in understand­ing the basic concepts of water potential, enabling them to use our instrument with confidence. Every effort has been
WP4C PotentiaMeter
1. Introduction
3
made to ensure that the content of this manual is correct and scientifically sound.
Seller’s Liability
Seller warrants new equipment of its own manufacture against defective workmanship and materials for a period of one year from date of receipt of equipment (the results of ordinary wear and tear, neglect, misuse, accident and excessive deterioration due to corrosion from any cause are not to be considered a defect); but Seller’s liability for defective parts shall in no event exceed the furnishing of replacement parts F.O.B. the factory where originally man­ufactured. Material and equipment covered hereby which is not manufactured by Seller shall be covered only by the warranty of its manufacturer. Seller shall not be liable to Buyer for loss, damage or injuries to persons (including death), or to property or things of whatsoever kind (including, but not without limitation, loss of anticipated profits), occasioned by or arising out of the installation, operation, use, misuse, nonuse, repair, or replacement of said material and equipment, or out of the use of any method or process for which the same may be employed. The use of this equipment constitutes Buyer’s acceptance of the terms set forth in this warranty. There are no under­standings, representations, or warranties of any kind, express, implied, statutory or otherwise (including, but without limitation, the implied warranties of merchantabil­ity and fitness for a particular purpose), not expressly set forth herein.
WP4C PotentiaMeter
2. About the WP4C
4
2. About the WP4C
Instrument Specifications
Range: 0 to -300MPa* Accuracy: ±0.05MPa* from 0 to -5 MPa
1% from -5 to -300 MPa
Measurement time:
~10-15 min. for most soil samples in precise mode ~20 min. for plant tissue samples <5 minutes in fast mode (reduced accuracy)
Temperature Control: 15 to 40° C (±0.2° C) Operating Environment: 5 to 40° C (41 to 104° F)
Sensor Type: 1) Chilled-mirror dewpoint sensor
2) Infrared temperature sensor
Sample Cup Capacity: 7 ml recommended (15 ml full) Dimensions: 24.1 x 22.9 x 8.9 cm (9.5 x 9.0 x 3.5 in) Weight: 3.2 Kg (7.1 lbs) Case Material: Powder Painted Aluminum Display: 20 x 2 alphanumeric LCD with backlighting Data Communication: RS232A compatible, 8-data bit
ASCII code, 9600 baud, no parity, 1 stop bit
Power: 110VAC to 220 VAC, 50/60 Hz Interface Cable: Standard RS232 serial cable (included) Compatible Standards: ASTM D6836-07 Warranty: 1 year parts and labor
* Note that the WP4C (and all vapor pressure instruments) are
WP4C PotentiaMeter
2. About the WP4C
5
limited by accuracy in the wet end of the water potential range. The range of 0 to -5 MPa has an accuracy of 0.05 MPa. For example, a measurement at -0.1MPa has an accuracy of ±50% of the read­ing, and -1 MPa has an accuracy of ±5% of the reading. Function­ally this means that the WP4C will not measure water potential accurately near field capacity.
WP4C and Water Potential
Water potential is a measurement of the energy status of the water in a system. It indicates how tightly water is bound, structurally or chemically, within a substance. Water potential can be computed from the vapor pressure of air in equilibrium with a sample in a sealed measure­ment chamber. For a more detailed description of water potential, please refer to Chapter 9, titled “Theory: Water Potential” of this manual.
How the WP4C works
WP4C uses the chilled-mirror dewpoint technique to mea­sure the water potential of a sample. In this type of instru­ment, the sample is equilibrated with the headspace of a sealed chamber that contains a mirror and a means of detecting condensation on the mirror. At equilibrium, the water potential of the air in the chamber is the same as the water potential of the sample. In the WP4C, the mirror temperature is precisely controlled by a thermoelectric (Peltier) cooler. Detection of the exact point at which con­densation first appears on the mirror is observed with a photoelectric cell. A beam of light is directed onto the
WP4C PotentiaMeter
2. About the WP4C
6
mirror and reflected into a photodetector. The photode­tector senses the change in reflectance when condensation occurs on the mirror. A thermocouple attached to the mirror then records the temperature at which condensa­tion occurs. Values begin to be displayed indicating that initial measurements are being taken. WP4C then signals you by flashing a green LED and/or beeping when final values are reached. The final water potential and tempera­ture of the sample is displayed.
In addition to the technique described above, WP4C uses an internal fan that circulates the air within the sample chamber to reduce time to equilibrium. Since both dewpoint and sample surface temperatures are simultaneously measured, the need for complete thermal equilibrium is eliminated.
The WP4C controls the sample temperature by means of an internal thermo-electrical module that monitors and stabilizes the sample block temperature according to how it is set.
WP4C and Temperature
Large temperature differences, between sample and block, will cause longer reading times, since a complete and accu­rate reading will not be made until the difference between the sample tempeature and the block temperature is less than
1.0 degree. To help you monitor the temperature difference between your sample and the block, you can access a sample equilibration screen at the main menu.
WP4C PotentiaMeter
3. Getting Started
7
3. Getting Started
Components of your WP4C
Your WP4C should have been shipped with the following items:
WP4C main unit
Quickstart guide
Certificate of Calibration
Power cord
RS-232 interface cable
25 plastic sample cups and lids
10 Stainless Steel sample cups
Operator’s Manual
12 vials of 0.5 Molal KCl
Cleaning Kit
WP4C PotentiaMeter
3. Getting Started
8
Choosing a Location
To ensure that your WP4C operates correctly and consis­tently, place it on a level surface. This reduces the chance that sample material will spill and contaminate the inside of the instrument. To protect the internal electrical com­ponents, and to avoid inaccurate readings, place your WP4C in a location where the temperature remains fairly stable. This location should be well away from air condi­tioner and heater vents, open windows, outside doors, refrigerator exhausts, or other items that may cause rapid temperature fluctuation.
Features
Front view of WP4C
LED indicator
light
LCD
Function Keys
Sample drawer
WP4C PotentiaMeter
3. Getting Started
9
Back view of WP4C
Preparing the WP4C for Operation
After finding a good location for your WP4C, plug the power cord to the back of the unit. Before turning it on, pull open the sample drawer (turn the knob to the “OPEN/LOAD” position). An empty disposable sample cup is usually placed upside-down in the drawer to protect it during shipment. Remove this sample cup and turn the instrument on. The ON/OFF switch is located on the lower right corner of the WP4C’s back panel. The follow­ing screens will appear on the LCD:
RS-232 interface
fuse well
ON/OFF Switch
Power cord
Plug
Fan
Lid thumb-screw
Model WP4C
PotentiaMeter
v4.03
WP4C PotentiaMeter
3. Getting Started
10
Then:
This is the main menu, displaying the water potential in both MegaPascals (MPa) and pF, and the sample tempera­ture in °C.
In order to provide the most accurate readings, WP4C should ideally be allowed a warm-up period of 15-30 min­utes after turning it on. When you insert a sample into the chamber drawer and turn the drawer knob to the “READ” position, the instrument will begin the read cycle to measure the water potential of your sample.
Portability
On occasion you may want to take water potential mea­surements in the field where it is not feasible to take sam­ples and return to the lab. The following is a procedure for powering your WP4C using your vehicle as a power source at sites where AC power is not readily available.
1. Purchase a portable power inverter that plugs into the 12V output (cigarette lighter) of your vehicle. The inverter should have a continuous output of at least 140 Watts.
2. Place the WP4C on a level surface. Care should be
0.0°C
0.00 MPa
pF 0.00
p
WP4C PotentiaMeter
3. Getting Started
11
taken to minimize temperature gradients that may affect the instrument while in the field. A Styrofoam box, for example, will help minimize temperature effects.
3. Plug the 12-volt inverter to the 12-volt output of the
vehicle, or directly to the battery itself.
4. Plug the WP4C to the inverter, and turn it on. When
the instrument is on, it draws up to 1 amp. Check the rating of your battery if you want to know how long it will power the instrument (for example, if your battery is rated for 60 amp hours, it will work for 60 hours when the vehicle isn’t running.
5. Allow the instrument to warm up for 15-30 minutes as
you would in the lab. Check the calibration of the instrument before proceeding with reading.
WP4C PotentiaMeter
4. The Menus
12
4. The Menus
The Main Menu
Every time you turn on your WP4C, it will come to this screen. If this screen doesn’t appear, refer to Chapter 12 for troubleshooting instructions. As mentioned earlier, the water potential and sample temperature are displayed on the screen. On each side of the LCD there are buttons. Each button performs a different function depending on which mode you want. Following is a description of the modes and options you may use, and the buttons used to set them.
0.0°C
0.00 MPa
pF 0.00
p
sample temperature menu system setup
language
selection
0.0°C
0.00 MPa
pF 0.00
p
precise/continuous/fast mode
WP4C PotentiaMeter
4. The Menus
13
Changing Languages
The WP4C comes to you with English as the default on­screen user language. If you prefer not to use English, you can change it to one of a variety of other languages: Ger­man, French, Spanish, Italian, Swedish, Danish, Norwe­gian, Czech, Portuguese, Japanese, Polish or Finnish. This is done simply by pressing the upper right button of the instrument while it is not reading a sample. You will see the following screen:
Press the upper right key again, and the next language option (German) will appear:
Each time you press the right button, the display will scroll to the next language option. Select the desired language, then press the lower left button to exit.
Reading Modes
Precise Mode
The first time you turn on the WP4C, it will be in precise mode. In this mode, measurements on a sample are
English
-Exit-
Deutsch
-zurück-
WP4C PotentiaMeter
4. The Menus
14
repeated until successive readings agree within a preset
tolerance (0.03MPa for Ψ>-40MPa; otherwise 0.3MPa) .
The WP4C always starts in precise mode. To toggle between the precise, continuous and fast modes, press the top left button. The display will show a small p, c or f to the left of the water potential readings:
main menu with continuous mode enabled
Here is a brief summary of the precise mode features:
Precise water potential value is insured by repeat­ing measurements on a sample.
Read time is typically 10-15 minutes.
Green LED blinks until you turn the drawer knob to the OPEN/LOAD position.
Continuous Mode
Continuous mode measures the water potential of your sample continuously until you turn the drawer knob to the OPEN/LOAD position. This can be useful in doing long term monitoring of samples that take an especially long
c
“c” for continuous mode
0.0°C
0.00 MPa
pF 0.00
WP4C PotentiaMeter
4. The Menus
15
time to come to vapor equilibrium, such as plant samples and moist soil samples with water potential > -0.5MPa. In this mode the WP4C will measure the sample, stop to dis­play the water potential and sample temperature, then begin another read cycle. Between samples, it will signal you with the green LED flash, accompanied by the beeper, if it is enabled. Some find it helpful to connect their WP4C to a computer while in continuous mode in order to log and store data over time. For instructions on how to do this, see Chapter 8.
Fast Mode
In fast mode, the sample is measured only once. Read time is typically 3-5 minutes. Readings are less precise, particu­larly in the wet range. However, fast mode is recom­mended for dry soil samples with water potential <-40 MPa.
System Configuration
If you press the bottom left button while in the main menu, it will bring you to the system configuration menu. This menu allows you to make minor system changes.
WP4C PotentiaMeter
4. The Menus
16
You can change how the beeper signals after each reading and enter the calibration menu as well.
System configuration menu
Changing the beeper
When you are reading, the WP4C has two ways of notify­ing you that the water potential reading is completed for your sample: the beeper and a flashing green LED, located on the left front corner of the WP4C’s case. In fast and precise reading modes, the LED will flash once when a sample is started. When it is finished the LED will flash
system configuration menu button
0.0°C
pF 0.00
p
0.00 MPa
Beeper
Beeper Mode indicator
Calibration
menu
Icon
(4 beeps shown)
temperature set
+
-
-Exit-
4x
set T
Loading...
+ 50 hidden pages