Decagon Devices AquaLab 4TE, AquaLab DUO, AquaLab 4TEV Operator's Manual

AquaLab
Water Activity Meter
Operator’s Manual
For Series 4TE, 4TEV, DUO
Version 7
Decagon Devices, Inc.
2365 NE Hopkins Court
Pullman WA 99163
(509)332-2756
fax: (509)332-5158
www.aqualab.com
support@decagon.com
sales@decagon.com
Trademarks
AquaLab is a registered trademark of
Decagon Devices, Inc.
©2008-2012 Decagon Devices, Inc.
i
Contents
1. Introduction .................................1
About this Manual ....................................1
Customer Support .....................................1
Warranty ...................................................2
Seller’s Liability ..........................................2
2. About AquaLab ...........................4
AquaLab Model and Options ....................4
AquaLab 4 Instrument Speci cations ........4
AquaLab 4 DUO Speci cations .................5
How AquaLab Works ................................6
AquaLab and Temperature ........................6
Chilled Mirror Dewpoint Limitations .......8
3. Water Activity  eory ...................9
Moisture Content ......................................9
Water Activity............................................9
Water Potential ........................................12
Sorption Isotherms ..................................13
4. Getting Started ........................... 15
Components of your AquaLab ................. 15
Choosing a Location ............................... 15
Preparing AquaLab for Operation ............16
5. Menus ......................................... 18
Measurement Tab ....................................18
ii
Con guration Tab ................................... 19
Admin Settings ........................................25
Data Tab .................................................30
6. Cleaning and Maintenance ........32
Cleaning the Block and Sensors ...............33
Cleaning a Series 4TEV: .......................34
Cleaning Procedure: ...............................34
Veri cation of Calibration .......................36
7. Ve r i cation and Calibration ........37
Water Activity Veri cation .......................37
Veri cation of Calibration .......................39
8. Sample Preparation ....................46
Preparing the Sample ...............................46
Samples Needing Special Preparation ......47
Slow Water-Emitting Samples ..................48
Volatile Samples .......................................49
Low Water Activity ..................................50
Samples Not at Room Temperature .........50
9. Taking a Reading ....................... 52
Measurement Steps ..................................52
How AquaLab Takes Readings ................52
10. Duo Operation (Optional) ....... 55
Obtaining Product Isotherm Models .......56
Loading and Organizing Product Models 56
Moisture Content Adjustment ...............60
Restore Original Moisture Content Model
Setting s .................................................... 6 4
iii
How to Delete Models .............................66
11. Computer Interface ...................68
AquaLink RG ..........................................68
Using Windows Hyperterminal .............68
12. Troubleshooting .......................70
13. Support and Repair ..................83
Repair Costs ............................................84
Loaner Service .........................................84
14. Further Reading .......................85
Water Activity  eory & Measurement ...85
Food Safety and Microbiology .................89
Appendix A ................................... 111
Preparing Salt Solution .......................... 111
Appendix B ................................... 113
Temperature Correction ........................ 113
Appendix C ................................... 114
AquaLab Veri cation Standards ............ 114
Declaration of Conformity ........... 119
Certi cate of Traceability .............120
iv
AquaLab
1. Introduction
1
1. INTRODUCTION
Welcome to Decagon’s AquaLab Series 4, 4TE, 4TEV, and DUO, the industry standard for measuring water activity (aw). AquaLab is the quickest, most accurate, and most reliable instrument available for measuring water activity. Whether you are researching or work­ing on the production line, AquaLab will suit your needs. It is easy to use and provides accurate and timely results.
About this Manual
Included in this manual are instructions for setting up your AquaLab, verifying the calibration of the instrument, preparing samples, and maintaining and caring for your instrument. Please read these in­structions before operating AquaLab to ensure that the instrument performs to its full potential.
Customer Support
If you ever need assistance with your AquaLab, or if you just have questions or feedback, there are several ways to contact us:
NOTE: If you purchased your AquaLab through a distributor, please contact them for assistance.
E-mail support@aqualab.com
Please include your name, contact information, instrument serial number(s), and a description of your problem or question.
sales@aqualab.com
Please include your name, address, phone number, the items you wish to order and a purchase order number. Credit card numbers should always be called in.
AquaLab
1. Introduction
2
Phone
1-800-755-2751 (USA and Canada Only) 1-509-332-2756 International
Our Customer Support and Sales Representatives are available Monday thru Friday.
Fax
1-509-332-5158
Warranty
AquaLab has a 30-day satisfaction guarantee and a three-year war­ranty on parts and labor. Your warranty is automatically validated upon receipt of the instrument. We will contact you within the  rst 90 days of your purchase to see how the AquaLab is working for you.
Seller’s Liability
Seller warrants new equipment of its own manufacture against de­fective workmanship and materials for a period of three years from date of receipt of equipment (the results of ordinary wear and tear, neglect, misuse, accident and excessive deterioration due to corro­sion from any cause are not to be considered a defect); but Seller’s liability for defective parts shall in no event exceed the furnishing of replacement parts Freight On Board the factory where originally manufactured. Material and equipment covered hereby which is not manufactured by Seller shall be covered only by the warranty of its manufacturer. Seller shall not be liable to Buyer for loss, damage or injuries to persons (including death), or to property or things of whatsoever kind (including, but not without limitation, loss of anticipated pro ts), occasioned by or arising out of the installation,
AquaLab
1. Introduction
3
operation, use, misuse, nonuse, repair, or replacement of said ma­terial and equipment, or out of the use of any method or process for which the same may be employed.  e use of this equipment constitutes Buyer’s acceptance of the terms set forth in this warranty.  ere are no understandings, representations, or warranties of any kind, express, implied, statutory or otherwise (including, but with­out limitation, the implied warranties of merchantability and  tness for a particular purpose), not expressly set forth herein.
AquaLab
2. About AquaLab
4
2. About AquaLab
AquaLab is the fastest and most accurate instrument for measuring water activity, giving readings in  ve minutes or less. Its readings are reliable, providing ±0.003 aw accuracy.  e instrument is easy to clean and checking calibration is simple.
AquaLab Model and Options
Series 4: Uses chilled-mirror dewpoint sensor, but lacks tempera-
ture control features of premium models.
Series 4TE: User-selectable internal temperature control model, uses thermoelectric (Peltier) components to maintain internal tem­perature.
Series 4TEV: Uses both a chilled-mirror dewpoint sensor and a ca­pacitance sensor for measuring non-volatile and volatile substances, respectively. Either sensor is easily selected using the instrument’s menu system.
Series 4TE DUO: Uses chilled-mirror dewpoint and programmed models obtained from isotherm data to give the user both water ac­tivity and moisture content simultaneously in  ve minutes or less.
AquaLab 4 Instrument Speci cations
Water Activity Range: 0.050 to 1.000 a
w
Water Activity Accuracy: ±0.003 (4TE Dew Point Mode) Water Activity Accuracy: ±0.015 (4TEV Capacitance Mode) Water Activity Resolution: 0.0001
AquaLab
2. About AquaLab
5
Read Time1: ≤5 min. Sample Temperature Range: 15 to 50° C Sample Temperature Accuracy: ±0.2° C Sample Temperature Resolution: 0.01° C Sample Dish Capacity: 15ml Full Operating Environment: 5 to 50° C 20 to 80% Humidity Case Dimensions: 26.7 x 17.8 x 12.7cm Weight: 3.1 Kg Case Material: Lustran 433 (ABS) with  re retardant Display: 64 x 128 Graphical Data Communications: RS232A Serial, 9600 to 115200 baud Power: 110 to 220 VAC, 50/60Hz Warranty: 3 year parts and labor
1
On samples with no signi cant impedance to vapor loss
AquaLab 4 DUO Speci cations
Moisture Content Repeatability: 0.02% Accuracy to Moisture Content Ref.: 0.1% to 0.5%
AquaLab and Water Activity
Water activity (aw) is a measurement of the energy status of the water in a system. It indicates how tightly water is “bound”, structurally or chemically, within a substance. Water activity is the relative hu­midity of air in equilibrium with a sample in a sealed measurement chamber.  e concept of water activity is of particular importance in determining product quality and safety. Water activity in uences color, odor,  avor, texture and shelf-life of many products. It predicts safety and stability with respect to microbial growth, chemical and biochemical reaction rates, and physical properties. For a more de­tailed description of water activity as it pertains to products, please refer to Chapter 3 of this manual, titled “ Water Activity  eory”.
AquaLab
2. About AquaLab
6
How AquaLab Works
AquaLab uses the chilled-mirror dewpoint technique to measure the water activity of a sample. In an instrument that uses the dewpoint technique, the sample is equilibrated with the head-space of a sealed chamber that contains a mirror and a means of detecting condensa­tion on the mirror. At equilibrium, the relative humidity of the air in the chamber is the same as the water activity of the sample. In the AquaLab, the mirror temperature is precisely controlled by a thermoelectric (Peltier) cooler. Detection of the exact point at which condensation  rst appears on the mirror is observed with a photo­electric cell. A beam of light is directed onto the mirror and re ected into a photo detector cell.  e photo detector senses the change in re ectance when condensation occurs on the mirror. A thermo­couple attached to the mirror then records the temperature at which condensation occurs. AquaLab then signals you by beeping and dis­plays the  nal water activity and temperature.
In addition to the technique described above, AquaLab uses an in­ternal fan that circulates the air within the sample chamber to reduce equilibrium time. Since both dewpoint and sample surface tempera­tures are simultaneously measured, the need for complete thermal equilibrium is eliminated, which reduces measurement times to less than  ve minutes.
AquaLab and Temperature
Samples not read at room temperature during the read cycle will equilibrate with the AquaLab’s temperature before the water activity is displayed. Large temperature di erences will cause longer reading times, since a complete and accurate reading will not be made until the sample and the instrument are within 2°C of each other.
AquaLab
2. About AquaLab
7
 ere are several advantages in having a temperature-controlled wa­ter activity meter. A few major reasons are:
1. Research purposes. Temperature control can be used to study the e ects of temperature on the water activity
of a sample, make a comparison of the water activity of di erent samples independent of temperature, and conduct accelerated shelf-life studies or other water activity studies where temperature control is critical.  ere are many shelf-life, packaging, and isotherm studies in which tempera­ture control would be very bene cial.
2. To comply with government or internal regulations for speci c products.  ough the water activity of most products varies by less than ± 0.002 per °C, some regulations require measurement at a spe­ci c temperature.  e most common speci cation is 25°C, though 20°C is sometimes indicated.
3. To minimize extreme ambient temperature  uctuations. If the environmental and AquaLab temperatures  uctuate by as much as ± 5°C daily, water activity readings will vary by ± 0.01 aw. Temperature control eliminates variations due to changes in ambient conditions.
Series 4TE/4TEV/4TE-DUO
 e AquaLab Series 4TE models have thermoelectric components installed to allow the instrument to maintain a set chamber tem­perature.  e temperature is set using the con guration menu of any of the Series 4 models.
AquaLab
2. About AquaLab
8
Chilled Mirror Dewpoint Limitations
AquaLab’s limitation is its ability to accurately measure samples with high concentrations (typically >1%) of certain volatiles such as etha­nol or propylene glycol, which can condense on the surface of the chilled mirror.  e extent of the e ect is determined by how readily the material volatilizes, which is both concentration- and matrix­dependent.  erefore, even if your sample contains materials that could volatilize, it may still be possible to make accurate readings using the chilled mirror dewpoint sensor.
AquaLab Series 4TEV which incorporates both a chilled mirror sen­sor and a capacitance sensor for measuring volatile substances is Decagon’s solution for products containing volatile materials. If you are unsure if you need the TEV model, please call and discuss your product with a Decagon Representative. Refer to Chapter 8’s section titled ”Volatile Samples” or call Decagon for more details.
AquaLab
3. Water Activity  eory
9
3. Water Activity  eory
Water is a major component of foods, pharmaceuticals, and cosmet­ics. Water in uences the texture, appearance, taste and spoilage of these products.  ere are two basic types of water analysis: moisture content and water activity.
Moisture Content
 e meaning of the term moisture content is familiar to most peo­ple. It implies a quantitative analysis to determine the total amount of water present in a sample. Primary methods for determining moisture content are loss on drying and Karl Fisher titration, but secondary methods such as infrared and NMR are also used. Mois­ture content determination is essential in meeting product nutri­tional labeling regulations, specifying recipes and monitoring pro­cesses. However, moisture content alone is not a reliable indicator for predicting microbial responses and chemical reactions in materi­als.  e limitations of moisture content measurement are attributed to di erences in the intensity with which water associates with other components.
Water Activity
Water activity is a measure of the energy status of the water in a system, and thus is a far better indicator of perishability than water content. Figure 1 shows how the relative activity of microorganisms, lipids and enzymes relate to water activity. While other factors, such as nutrient availability and temperature, can a ect the relationships, water activity is the best single measure of how water a ects these processes.
AquaLab
3. Water Activity  eory
10
Fig. 1: Water Activity Diagram—adapted from Labuza
Water activity of a system is measured by equilibrating the liquid phase water in the sample with the vapor phase water in the head­space and measuring the relative humidity of the head-space. In the AquaLab, a sample is placed in a sample cup which is sealed inside a sample chamber. Inside the sample chamber is a fan, a dew point sensor, a temperature sensor, and an infrared thermometer.  e dewpoint sensor measures the dewpoint temperature of the air in the chamber, and the infrared thermometer measures the sample temperature. From these measurements, the relative humidity of the head-space is computed as the ratio of dewpoint temperature saturation vapor pressure to saturation vapor pressure at the sample temperature. When the water activity of the sample and the rela­tive humidity of the air are in equilibrium, the measurement of the head-space humidity gives the water activity of the sample.  e pur­pose of the fan is to speed equilibrium and to control the boundary layer conductance of the dewpoint sensor.
AquaLab
3. Water Activity  eory
11
In addition to equilibrium between the liquid phase water in the sample and the vapor phase, the internal equilibrium of the sample is important. If a system is not at internal equilibrium, one might measure a steady vapor pressure (over the period of measurement) which is not the true water activity of the system. An example of this might be a baked good or a multi-component food. Initially out of the oven, a baked good is not at internal equilibrium; the outer surface is at a lower water activity than the center of the baked good. One must wait a period of time in order for the water to migrate and the system to come to internal equilibrium. It is important to remember the restriction of the de nition of water activity to equi­librium.
Temperature E ects
Temperature plays a critical role in water activity determination. Most critical is the measurement of the di erence between sample and dewpoint temperature. If this temperature di erence were in er­ror by 1°C, an error of up to 0.06 aw could result. In order for water activity measurements to be accurate to 0.001, temperature di er­ence measurements need to be accurate to 0.017°C. AquaLab’s in­frared thermometer measures the di erence in temperature between the sample and the block. It is carefully calibrated to minimize tem­perature errors, but achieving 0.017°C accuracy is di cult when temperature di erences are large. Best accuracy is therefore obtained when the sample is near chamber temperature.
Another e ect of temperature on water activity occurs when sam­ples are near saturation. A sample that is close to 1.0 aw and is only slightly warmer than the sensor block will condense water within the block.  is will cause errors in the measurement, and in subsequent measurements until the condensation disappears. A sample at 0.75 aw needs to be approximately 4°C above the chamber temperature to
AquaLab
3. Water Activity  eory
12
cause condensation.  e AquaLab warns the user if a sample is more than 4°C above the chamber temperature, but for high water activity samples the operator needs to be aware that condensation can occur if a sample that is warmer than the block is put in the AquaLab.
Water Potential
Some additional information may be useful for understanding what water activity is and why it is such a useful measure of moisture status in products. Water activity is closely related to a thermody­namic property called the water potential, or chemical potential (μ) of water, which is the change in Gibbs free energy (∆G) when water concentration changes. Equilibrium occurs in a system when (µ) is the same everywhere in the system. Equilibrium between the liquid and the vapor phases implies that () is the same in both phases. It is this fact that allows us to measure the water potential of the vapor phase and use that to determine the water potential of the liquid phase. Gradients in (µ) are driving forces for moisture movement.  us, in an isothermal system, water tends to move from regions of high water potential (high aw) to regions of low water potential (low aw). Water content is not a driving force for water movement, and therefore can not be used to predict the direction of water move­ment, except in homogeneous materials.
Factors In Determining Water Activity
 e water activity of the water in a system is in uenced by factors that e ect the binding of water.  ey include osmotic, matric, and pressure e ects. Typically water activity is measured at atmospheric pressure, so only the osmotic and matric e ects are important.
Osmotic E ects
Osmotic e ects are well known from biology and physical chemis­try. Water is diluted when a solute is added. If this diluted water is
AquaLab
3. Water Activity  eory
13
separated from pure water by a semi-permeable membrane, water tends to move from the pure water side through the membrane to the side with the added solute. If su cient pressure is applied to the solute-water mixture to just stop the  ow, this pressure is a measure of the osmotic potential of the solution. Addition of one mole of an ideal solute to a kilogram of water produces an osmotic pressure of
22.4 atm.  is lowers the water activity of the solution from 1.0 to
0.98 aw. For a given amount of solute, increasing the water content of the systems dilutes the solute, decreasing the osmotic pressure, and increasing the water activity. Since microbial cells are high con­centrations of solute surrounded by semi-permeable membranes, the osmotic e ect on the free energy of the water is important for determining microbial water relations and therefore their activity.
Matric E ects
 e sample matrix a ects water activity by physically binding water within its structure through adhesive and cohesive forces that hold water in pores and capillaries, and to particle surfaces. If cellulose or protein were added to water, the energy status of the water would be reduced. Work would need to be done to extract the water from this matrix.  is reduction in energy status of the water is not osmotic, because the cellu­lose or protein concentrations are far too low to produce any signi cant dilution of water.  e reduction in energy is the result of direct physical binding of water to the cellulose or protein matrix by hydrogen bonding and van der Waal forces. At higher water activity levels, capillary forces and surface tension can also play a role.
Sorption Isotherms
Relating Water Activity to Water Content
Changes in water content a ect both the osmotic and matric bind­ing of water in a product.  us a relationship exists between the water activity and water content of a product.  is relationship is
AquaLab
3. Water Activity  eory
14
called the sorption isotherm, and is unique for each product. Besides being unique to each product, the isotherm changes depending on whether it was obtained by drying or wetting the sample.  ese fac­tors need to be kept in mind if one tries to use water content to infer the stability or safety of a product. Typically, large safety margins are built into water content speci cations to allow for these uncertainties.
While the sorption isotherm is often used to infer water activity from water content, one could easily go the other direction and use the water activity to infer the water content.  is is particularly attrac­tive because water activity is much more quickly measured than wa­ter content.  is method gives particularly good precision in the center of the isotherm. In order to infer water content from water activity, one needs an isotherm for the particular product. Decagon sells an Isotherm Generator called the AquaSorp IG or you can also have Decagon run the isotherm for a fee.
For example, if one were using the AquaLab to monitor the water content of dried potato  akes, one would measure the water activity and water content of potato  akes dried to varying degrees using the standard drying process for those  akes. An isotherm would be con­structed using those data, and the water content would be inferred using the measured water activity of samples and that isotherm. We have an upgrade available to Series 4TE users that would allow you to determine moisture content and water activity simultaneously.  is instrument is called the Series 4TE DUO.
 e importance of the concept of water activity of foods, pharma­ceuticals, and cosmetics cannot be over emphasized. Water activity is a measure of the energy status of the water in a system. More impor­tantly, the usefulness of water activity in relation to microbial growth, chemical reactivity, and stability over water content has been shown.
AquaLab
4. Getting Started
15
4. Getting Started
Components of your AquaLab
Your AquaLab should have been shipped with the following items:
AquaLab water activity meter
• Calibration Certi cate
• Power cord
RS-232 interface cable
100 disposable sample cups
• Operator’s Manual
Quick Start guide
• Cleaning Kit
3 vials each of the following veri cation solutions:
1.000 a
w
Distilled Water
0.760 a
w
6.0 molal NaCl
0.500 a
w
8.57 molal LiCl
0.250 aw 13.41 molal LiCl
Choosing a Location
To ensure that your AquaLab operates correctly and consistently, place it on a level surface.  is reduces the chance that sample ma­terial will spill and contaminate the sample chamber. Also select a location where the temperature remains fairly stable to avoid tem­perature changes that can a ect accuracy.  is location should be well away from air conditioner and heater vents, open windows, etc. Place the AquaLab in a location where cleanliness can be maintained to prevent contamination of the sample chamber.
AquaLab
4. Getting Started
16
Preparing AquaLab for Operation
After  nding a good location for your AquaLab, plug the power cord into the back of the unit.  e ON/OFF switch is located on the lower left corner of the AquaLab’s back panel. When the AquaLab is turned on, you should see a model name/number screen and then the main screen as shown below.
 e main screen shows the water activity (aw) in the middle of the screen and the sample temperature right below. On the Series 4TEV model you will also see either DEW or CAP indicating whether you are using the dewpoint or capacitance sensor respectively.
NOTE: In order to provide the most accurate readings, your AquaLab should be allowed a 15 minute warm-up period.
AquaLab
4. Getting Started
17
If users have been setup on the instrument, the following screen will appear instead of the main screen. (See Chapter 5 for more informa­tion on administrative settings and user setup).
Select the appropriate user and login to begin.
AquaLab
5. Menus
18
5. Menus
At the top of the display screen there are three tabs: Measurement, Con guration, and Data.  ese tabs indicate the three menus you can access. To change between the tabs press the right most button below the document icon.
 e enter icon is the read or enter button. Once the latch is set to the read position, the document icon will switch to an “X” icon, which allows the user to stop the current reading. During a reading, press­ing enter again will restart the reading.
Measurement Tab
 e measurement tab, as seen above, is the main screen which dis­plays each time you turn on your AquaLab. If this screen doesn’t appear, refer to Chapter 12 for troubleshooting instructions. As mentioned earlier, the water activity and sample temperature are displayed on the screen.
AquaLab
5. Menus
19
Pushing the right or left arrow keys will change the display to a temperature equilibration screen shown below.  is screen shows the temperature di erence between the sample temperature and the lid temperature.
Con guration Tab
When at the con guration screen, pressing the up and down arrow keys moves the cursor through the various con guration options Press the left and right arrows to page through the options.  e enter button will allow you to change the highlighted setting.
AquaLab
5. Menus
20
Calibration:
Pressing the Enter button when Calibration is highlighted starts the veri cation process. For more details on the water activity veri ca­tion procedure refer to Chapter 7. Refer to Chapter 10 for moisture content veri cation information (Duo model only). You may also reset the calibration to the factory defaults by highlighting the De­faults option and pressing Enter.  is will reset all options to the way they were when the instrument arrived at your location.
Temperature:
 e default temperature is 25°C. Press the enter button to change the temperature setting.  e AquaLab Series 4TE models may be set between 15 and 50°C by 0.1°C intervals. Using the up and down arrows, set the AquaLab to your desired temperature and press the save button.
AquaLab
5. Menus
21
Temp Eq:
 e Temperature Equilibration option allows you to set the level of temperature equilibration desired before the water activity measure­ment begins.  e range is 0.5 to 4.0°C. A setting of 4.0°C begins the measurement immediately (assuming the sample is not >4.0°C above or below the block temperature). A setting of 0.5 °C will cause the instrument to wait until the sample temperature is within <0.5°C of the block temperature before starting the water activity measurement.
Sensor:
In the AquaLab Series 4TEV model only, this option indicates the selected sensor type, either dewpoint or capacitance ( e Series 4 and 4TE models will always be Dewpoint). Pressing Enter when the Sensor option is highlighted allows you to change between a capaci­tance sensor or chilled mirror dewpoint sensor for sampling with or without volatiles, respectively.
AquaLab
5. Menus
22
Mode:
Users may choose between single, continuous, or custom mode by pushing the save button.
Single Mode
Single mode reads the sample once, after which the instrument noti­ es you that it is  nished and the water activity and temperature are displayed on the screen.
Continuous Mode
Continuous mode reads your sample until you open the chamber lid or stop the test using the stop button. AquaLab reads the sample, displays the water activity and temperature, then begins another read cycle without further input from the user. Between samples, the ma­chine will signal you with beeps.  is mode eliminates the possibility of moisture exchange with the environment outside the chamber in between readings. A time on the bottom left of the screen tracks the cumulative read time. All readings taken during continuous mode are saved on the instrument’s memory if the autosave feature is se­lected (see Auto Save below). If AquaLab is connected to a computer using AquaLink RG (See Chapter 11), all readings taken during con­tinuous mode will be downloaded to the AquaLink RG software.
Custom Mode
Custom mode allows a sample to be read multiple times until a de­sired level of stability is achieved.  e user determines how many consecutive tests they want to be within a given water activity sta­bility setting. For instance, the customer can choose to have 4 con­secutive tests be within +/- 0.001aw.  e instrument will continue to run tests until it records 4 consecutive tests that are within +/-
0.001aw and then will stop and report the value of the  nal test. If autosave is turned on, all test readings will be saved to the instru-
AquaLab
5. Menus
23
ments memory, but only the  nal reading will appear on the main measurement screen. If AquaLab is connected to a computer using AquaLink RG (See Chapter 11), all readings taken during a cus­tom mode test will be downloaded to the AquaLink RG software.
On the mode screen, at the top of the page, will appear the current mode settings with the number of tests appearing  rst, followed by the stability value (∆aw). Pressing enter with the custom mode highlight­ed will allow the number of tests and stability settings to be changed.
To change the number of readings, use the right/left arrow but­tons to highlight the number under Readings, and then use the up and down buttons to change to any value between 2 and 9.
AquaLab
5. Menus
24
To change the stability setting, use the right/left arrow buttons to highlight the number under ∆a
w
, and then use the up and down but­tons to change to any value between 0.0005 and 0.0200. To save the settings and  nish, press the save button (to exit without updating, press the cancel button).  e mode screen will now appear with the updated custom settings appearing at the top of the screen. Press the save button to return to the con guration screen and begin using the custom mode (To exit without updating, press the cancel button).
Date:
AquaLab Series 4 models now have an internal calendar and clock.  e time and date are recorded with each water activity reading. Pressing Enter when the Date option is highlighted allows you to set the date in the instrument. Press the left and right arrows to change between the month, day and year. Press the up or down arrows to change any of the individual values.
Time: Pressing Enter when the Time option is highlighted allows you to set the current local time. Press the up or down arrows to change any of the individual values. Press the left or right buttons to change between hour and minutes.  e hour setting automatically changes between AM and PM.
AquaLab
5. Menus
25
Regional Formatting:
Allows you to con gure how all Series 4 models will display infor­mation. You may choose the temperature scale (Celsius vs Fahren­heit), the date display (mm/dd/yy vs. dd/mm/yy), the hour format (12 vs 24 hour) and the language.
Admin Settings
Allows you to create an administrator password as well as create, edit and delete additional users.
AquaLab
5. Menus
26
 e admin option allows the administrator to grant or block access to some or all of the con guration options in all Series 4 models. For example: If the administrator wanted to make sure that all samples were read at 25°C the administrator would set their temperature to 25°C and then would lock all other users out of that con guration screen.  is is accomplished by entering the Access function and se­lecting the desired option to toggle it on and o . Additionally you can lock and unlock all of them at once. (For example, if you do not want John Doe changing the instruments measurement temperature, the administrator can lock that function for John.)
 e areas that can be locked are calibration, temperature, tempera­ture equilibration, sensor selection, mode, date/time, region, pass­word, auto-save, number of beeps, contrast, and delete functions.
AquaLab
5. Menus
27
User Setup: Users can be added, edited or deleted from this screen. An alphabet
screen will appear where a name can be entered using lower case, upper case and accents.
NOTE: User setup is not required for instrument operation. It is in place for users wanting to be compliant with 21 CFR Part 11 or who want to maintain the settings they have selected.
Auto Save: AquaLab Series 4 models have the ability to store water activity read­ings within the instrument. By selecting Auto Save “On,” every wa­ter activity reading will be automatically stored in the instrument’s internal memory. AquaLab Series 4 can store up to 8,000 records before the memory is full. If you select Auto Save “o ” then no data is automatically stored, although any individual reading may be manually stored right after the test is completed, before the next test is started.
To manually store a water activity or append an annotation to the active reading that has been autosaved, press the save icon button after the water activity measurement is completed. Pressing the icon opens a “name” screen. You may give this reading a name by press-
AquaLab
5. Menus
28
ing the arrow buttons to highlight the letter and then pressing the “Check” icon button. Press the save icon to save this data record with the name you have speci ed.
NOTE: Pressing the save icon button without giving it a name will save the reading without a name. If the save icon is not pressed after a read­ing, and the reading is autosaved, it is not possible to give an annotation later
Beeps:
Allows you to set the reading  nished noti cation from 4 beeps to continuous beeps. You may also turn the audible noti cation o .
Contrast:
Allows you to set the contrast of the screen to your liking. Viewing the screen from a sitting versus a standing position may require con­trast adjustment for the best visibility in that position.
Diagnostics:
For the chilled-mirror dewpoint sensor it provides you with lid, base, sample and mirror temperatures, optical voltage as well as the chilled mirror dew point user calibration.
AquaLab
5. Menus
29
For the capacitance sensor (TEV Models only) it provides you lid, base, and sample temperatures, relative humidity, as well as the ca­pacitive sensor calibration.
About:  is screen provides important information including the serial number and code version of your instrument.
AquaLab
5. Menus
30
Data Tab
View:
 is selection will allow you to view your stored measurements.  e up/down arrows will move you through the stored data with the most recent measurements at the top of the table. You may also press the left and right arrows to page quickly through the data. See Chapter 11: Computer Interface for information about downloading these readings to a computer.
When you are viewing the summary screen, you may press the enter button on a highlighted reading to get detailed information on the reading as shown below.
AquaLab
5. Menus
31
 e information shown is the water activity of the sample, the tem­perature, the test time, the user who ran the test (if setup), the date of the reading, the sensor used (4TEV only), the time the reading was taken, and the sequence number of the stored reading.
Delete: Selecting this option will delete all of the information currently stored in the instrument. If you have not backed up this informa­tion with Aqualink RG, you will be reminded of this by the follow­ing message:
NOTE: You will NOT be able to recover deleted data.
AquaLab
6. Cleaning and Maintenance
32
6. Cleaning and Maintenance
Keeping your AquaLab clean is vital to maintaining the accuracy of your instrument. Dust and sampling debris can contaminate the sampling chamber and must therefore be regularly cleaned out. To clean your instrument, carefully follow these instructions and refer to the labeled diagram below.
AquaLab
6. Cleaning and Maintenance
33
Purpose
 e purpose for the cleaning procedure is to remove grease, dirt and other soluble substances which can absorb/release water during veri­ cation, calibration, and/or sample testing. For a smooth and even dew formation, it requires the mirror to be perfectly clean. If there are any contaminants (e.g.  ngerprints) on the mirror, the dew will form unevenly and thus a ect the accuracy of the reading.
When to Clean
 e instrument should be cleaned if visual inspection indicates the chamber is dirty or as instructed in the veri cation  owchart on page 41.
Materials Needed
• A thin plastic rod or other non-metal implement
• Distilled Water
• Isopropyl Alcohol (IPA) or Decagon Cleaning Solution
• Kimwipes®
You may also purchase the AquaLab Cleaning Kit which comes with all the above materials except the Isopropyl Alcohol and Distilled Water.
NOTE: Wash your hands with soap and water and/or use clean lab gloves before starting the cleaning procedure.  is will prevent oils from contam­inating the cleaning materials, the sample chamber and/or the sensors.
Cleaning the Block and Sensors
Accessing the Sample Chamber
Turn the power o on your AquaLab. If latched, move the lever over to the open position. Lift the chamber cover to expose the
AquaLab
6. Cleaning and Maintenance
34
sample chamber and sensors.  e sample chamber consists of all surfaces inside the red o-ring when the lid is closed.
Cleaning a Series 4TEV:
If cleaning an AquaLab Series 4TEV, follow the cleaning procedures listed below being careful not to get cleaning solution or alcohol on the capacitance sensor  lter (see illustration on previous page). Re­peated exposure of cleaning materials or contaminants to the  lter may cause inaccurate readings. If the  lter appears to be contami­nated, it may need to be replaced. (To replace the capacitance sensor  lter, use a tweezer or small knife blade to pry up the edge of the  lter, being careful not to disturb the sensor beneath. Discard the soiled  lter, then with clean hands, press a new  lter into place.)
Cleaning Procedure:
Cleaning your AquaLab is a multi-step procedure which in­volves washing, rinsing, and drying for each speci c area as out­lined below (refer to illustration at the beginning of this chap­ter to identify the location of the components to be cleaned):
1. Cleaning the Sample Chamber
Note: Be extremely careful not to damage the fan blades (see illustra­tion) when cleaning the chamber.
a. Remove any debris that may have collected within or around
the sample chanber.
b. Wrap a NEW Kimwipe around the end of the thin plastic
rod (spatula) and moisten it with isopropyl alcohol or Deca­gon Cleaning Solution. Note: Do NOT dip a used Kimwipe
into your container of IPA or cleaning solution (the IPA or cleaning solution will become contaminated).
c. WASHClean upper chamber, o-ring, and all surfaces of the
block within the o-ring. You may need to replace the Kim-
AquaLab
6. Cleaning and Maintenance
35
wipe if it becomes too dirty during this process.
d. Clean lower block with a fresh Kimwipe. Be sure to clean
the entire block surface.
e. RINSERepeat steps b-d using new Kimwipes with distilled
water.
f. DRYRepeat steps b-d using new, dry Kimwipes to help
remove any moisture remaining from the cleaning.
g. Visually inspect the sample chamber for cleanliness. Re-clean
if necessary. Note: Do not reuse Kimwipes.
2. Clean the Mirror
a. Wrap a new Kimwipe around the end of the thin plastic rod
(spatula) and moisten it with isopropyl alcohol or Decagon Cleaning Solution.
b. WASHSwipe the moistened Kimwipe across the mirror
once. (A single swipe is usually su cient to remove con­taminants.)
c. RINSERepeat steps a-b using new Kimwipes moisted with
distilled water instead of cleaning solution.
d. DRYRepeat steps a-b using new, dry Kimwipes to help
remove any moisture remaining from the cleaning.
e. Visually inspect the mirror for cleanliness. Re-clean if neces-
sary.
3. Clean the  ermopile and Optical Sensor
a. Wrap a new Kimwipe around the end of the thin plastic rod
(spatula) and moisten it with isopropyl alcohol or Decagon Cleaning Solution.
b. WASHSwipe the moistened Kimwipe across thermopile
and optical sensor. (A single swipe across the sensor is usu­ally su cient to remove contaminants.)
c. RINSERepeat steps a-b using new Kimwipes moistened
with distilled water instead of cleaning solution.
d. DRYRepeat steps a-b but use a new, dry Kimwipe to help
remove any moisture remaining from the cleaning.
AquaLab
6. Cleaning and Maintenance
36
e. Visually inspect the thermopile and optical sensor for cleanli-
ness. Re-clean if necessary.
4. Additional Drying Time
a. Visually inspect the sample chamber and sensors for contami-
nants, including moisture. If necessary, repeat the cleaning process using new Kimwipes.
b. Let stand for about 5 minutes to ensure the sample chamber
is dry.
Veri cation of Calibration
After you have cleaned the chamber and other parts of your AquaLab, it is important to check the instrument’s performance in order to correct for any linear o set that may have occurred during the clean­ing process.
Before you check the instrument we recommend that you run a sample of the activated charcoal pellets provided in your AquaLab cleaning kit.  is cleans the air inside the chamber, helping it come back to a stable sampling environment.
Verify the linear o set against known veri cation standards accord­ing to the procedure described in the next chapter. If a linear o set has occurred, refer to “adjust for linear o set” section in Chapter 7 for directions on how to correct for linear o set. If, after adjusting for linear o set, your instrument is still not reading samples cor­rectly, please contact Decagon for support.
AquaLab
7. Veri cation and Calibration
37
7. Veri cation and Calibration
It is important to verify AquaLab’s water activity calibration against known standards to guarantee optimal performance and accuracy. Decagon recommends veri cation daily, once per shift or before each use.
Water Activity Veri cation
AquaLab uses the chilled-mirror dewpoint technique to determine water activity. Because this is a primary measurement of relative hu­midity, no calibration is necessary; however, it is important to verify for linear o set periodically.  e components used by the instru­ment to measure water activity are subject to contamination which may a ect the AquaLab’s performance. When this occurs, it changes the accuracy of the instrument.  is is what is called a “linear o set.”  erefore, frequent veri cation assures you that your AquaLab is performing correctly. Linear o set is checked by using two di erent veri cation standards.
Veri cation Standards
Veri cation standards are specially prepared unsatuarated salt solu­tions having a speci c molality and water activity value which are accurately measurable.  e veri cation standards that were sent with your initial shipment are very accurate and readily available from Decagon. Using veri cation standards to verify accuracy can greatly reduce preparation errors. For these reasons, we recommend using standards available through Decagon for the most accurate veri ca­tion of your AquaLab’s performance.
Performance Veri cation Standards come in  ve water activity lev­els: 1.000, 0.984, 0.760, 0.500, and 0.250 aw.  e standards are
AquaLab
7. Veri cation and Calibration
38
produced under a strict quality assurance regime. Please contact Decagon Devices to order additional standards via sales@decagon. com or 1-800-755-2751.
Verifi cation Standard
@ 25°C
Water Activity
13.41m LiCl 0.250 ±0.003
8.57m LiCl 0.500 ±0.003
6.0m NaCl 0.760 ±0.003
0.5m KCl 0.984 ±0.003
Distilled Water 1.000 ±0.003
NOTE: If you need to obtain a Material Safety Data Sheet (MSDS) for any of these standards, a printable version is available on our web­site at www.decagon.com/msds.
To use a veri cation standard, remove the twist top and pour the contents into an AquaLab sample cup. Information about the stan­dard’s value and molality can be found printed on the outside of the plastic vial. If for some reason you cannot obtain Decagon’s veri cation standards and need to make a saturated salt solution for veri cation, refer to Appendix A.
In TEV models, the capacitance sensor can hold a memory of high water activity samples such as distilled water or the 0.984 a
w
stan­dard. If you verify calibration with one of these high water activity standards, you will need to wait an hour to allow the capacitance sensor to dry before testing samples of lower water activity or the results may be slightly high.
AquaLab
7. Veri cation and Calibration
39
Veri cation of Calibration
When to Verify for Linear O set
Linear o set should be checked against two known veri cation standards daily, either once per shift or before each use. Linear o set should never be veri ed solely against distilled water, since it does not give an accurate representation of the linear o set. For batch pro­cessing, the instrument should be checked regularly against a known standard of similar water activity. It is also a good idea to check the o set with a standard of similar water activity when the general water activity range of your sample is changing. Checking the water activity of a standard solution will alert you to the possibility of unit contamination or shifts in the linear o set from other causes.
NOTE:  e veri cation process is the same for both the dewpoint and capacitance sensors in TEV models except that the accuracy for the ca­pacitance sensor is ± 0.015 a
w
. Also, if using the 1.000 aw or 0.984 a
w
veri cation standards, you must wait an hour before testing other stan­dards or samples in capacitance mode.
Veri cation
To verify for linear o set of your AquaLab do the following: (refer to the  owchart at the end of this section)
1. Choose a veri cation standard that is close to the water activity of the sample you are measuring. Note:  e AquaLab needs to warm up for approximately 15 minutes to make accurate readings.
2. Empty a vial of solution into a sample cup and place it in the AquaLab’s testing chamber. Make sure that your standard is as close to the instrument temperature as possible.
Note: Make sure the rim and outside of the sample cup are clean.
AquaLab
7. Veri cation and Calibration
40
3. Carefully close the lid and move the lever to the READ position.
4. Take two readings.  e water activity readings should be within ± 0.003 a
w
of the given value for the veri cation standard. See Ap- pendix B for the correct water activity value of Decagon’s standards at temperatures other than 25°C.
5. If your AquaLab is reading within ±0.003 aw of the veri cation standard, chose a second veri cation standard that would border the range of water activity you plan to test. For example, if you plan to test for water activity readings ranging between 0.713 and 0.621 you should use the 6.0M, NaCl (0.76aw)standard for your  rst veri­ cation and the 8.57M LiCl (0.50aw) for the second veri cation.
6. Prepare a sample cup of the second veri cation standard and make two readings.  e second water activity reading for the second veri cation standard should be within ±0.003 aw.
7. If either of the veri cation standards is not correct, it is probably due to contamination of the sensor chamber. For cleaning instruc­tions, see Chapter 6. After cleaning, repeat veri cation from step two.
8. If you are consistently getting readings outside the water activity of your  rst veri cation standard by more than ±0.003 aw, a linear o set has probably occurred. In this case, adjust the reading to match the veri cation standard’s correct value as outlined in the next section.
AquaLab
7. Veri cation and Calibration
41
Correct
Not Correct
Next
Correct
Not Correct
Measure Verification Standard
Correct
Not Correct
Repeat Process
Go to Linear Offset Procedure
Go to Sampling Procedure
Clean Sample Chamber
Re-Read Standard
Measure 2nd Standard
Clean Sample Chamber
 is  owchart is a graphical representation of the directions given above for checking for linear o set.
AquaLab
7. Veri cation and Calibration
42
Adjust for Linear O set
1. Once you are certain a linear o set has occurred, toggle to the Con guration tab by pressing the Menu icon button. Calibration is the  rst option highlighted in the con guration tab. Press the Enter icon button to begin the veri cation process. You will be guided through the linear o set routine through on screen commands.  e following screen will appear:
NOTE:  e DUO model will show both water activity and moisture
content on this screen. For TEV Models, make sure you have the correct sensor selected. .
2. Press the Enter button to start the linear offset process. To return to the main menu, press the cancel button. After pressing the enter button, the following screen will appear:
AquaLab
7. Veri cation and Calibration
43
3. Empty the whole vial of solution into a sample cup. We recom­mend using the 6.0 NaCl (0.76a
w
). Do not adjust for the o set us­ing distilled water. Ensure the rim and outside of the cup are clean. Place the sample cup in the AquaLab’s sample chamber.
NOTE:  e same veri cation standard may be used to verify and adjust the linear o set.
4. Carefully close the lid and move the lever to the READ position. Press the Check icon button to begin testing.
NOTE: If you decide at this point not to continue with the linear o set program, just return the lever to the OPEN position or press the cancel button and you will be returned to the previous screen.
5. After your AquaLab has  nished measuring the veri cation stan­dard, it will display the following screen:
6. Press the up and down arrows to adjust the water activity reading to its proper value for the particular veri cation standard you are measuring. When the correct value is displayed, press the Save icon button to store this new value. To cancel and return to the main menu, press the cancel button and no changes will be made.
AquaLab
7. Veri cation and Calibration
44
7. Re-measure the veri cation standard again in normal sampling mode. It should read the proper value (within ±0.003 a
w
) at a given temperature for your particular standard (see Appendix B for tem­peratures other than 25°C ).
Measure the water activity of a second veri cation standard accord­ing to the veri cation procedure described above. If both veri ca­tion readings are within ±0.003 a
w
then the instrument is ready to
begin testing.
If you still have incorrect veri cation standard readings after clean­ing the chamber and adjusting for linear o set, contact Decagon by email at support@decagon.com or by phone at 509-332-2756 (800-755-2751 in US and Canada) for further instructions. If you purchased your Decagon instrument from one of our international distributors, please contact them for local service and support.
How to Restore Factory Defaults
To restore original calibration settings, do the following:
1. Toggle to the Con guration tab by pressing the Menu icon but­ton. Select Calibration and press the Enter button (Select water activity for DUO models.).
AquaLab
7. Veri cation and Calibration
45
2. Scroll down to Defaults and press the Enter icon button to access the Restore Factory Defaults routine. To cancel and return to the main menu, press the Cancel icon button. After pushing the Enter icon button, the following screen will appear:
NOTE: For TEV models make sure you have the correct sensor selected.
3. To restore the factory calibration values, press the Check icon button. To cancel and return to the main menu, press the cancel button. After pressing the Check icon button, the following screen will appear:
4. To return to the main menu screen, press the Check icon button.
AquaLab
8. Sample Preparation
46
8. Sample Preparation
Proper sample preparation is an important step in keeping your AquaLab clean and achieving repeatable results. Careful preparation and loading of samples will lengthen time between cleanings and help you avoid downtime.
Preparing the Sample
1. Make sure the sample to be measured is homogeneous. Multi- component samples (e.g., mu ns with raisins) or samples that have outside coatings (like deep-fried, breaded foods) can be measured, but may take longer to equilibrate. For samples like these, AquaLab may take more than  ve minutes to give an accurate reading, or may require multiple readings of the same sample. Measuring the water activity of these types of products is discussed in detail later in this chapter (see Samples Needing Special Preparation).
2. Place the sample in a disposable sample cup, completely cov- ering the bottom of the cup, if possible. AquaLab is able to accu­rately measure a sample that does not (or cannot) cover the bottom of the cup. For example, raisins only need to be placed in the cup and not  attened to cover the bottom. A larger sample surface area increases instrument e ciency by providing more stable infrared sample temperatures. It also speeds up the reading by shortening the time needed to reach vapor equilibrium.
3. Do not  ll the sample cup more than half full. Over lled cups will contaminate the sensors in the sensor chamber. Filling the sample cup will not make the readings faster or more accurate.  ere only needs to be enough sample in the cup to allow the water in the sample to equilibrate with the water in the vapor phase and not
AquaLab
8. Sample Preparation
47
change the moisture content of the sample. Covering the bottom of the sample cup provides enough sample to get an accurate reading.
4. Make sure the rim and outside of the sample cup are clean. Wipe any excess sample material from the rim of the cup with a clean Kimwipe. Material left on the rim or the outside of the cup can contaminate the sensor chamber and be transferred to subse­quent samples.
5. If a sample will be read at some other time, put the sample cup’s disposable lid on the cup to restrict water transfer. For long­term storage, seal the lid by placing tape or Para lm® completely around the cup/lid junction.
6. Be consistent in sample preparation practices. If you crush, grind, or slice your sample, be consistent in the method you use in order to obtain reproducible results.
Samples Needing Special Preparation
AquaLab reads most materials in  ve minutes or less. Some sam­ples, however, may require longer reading times, due to their na­ture.  ese materials need additional preparation to ensure quick, accurate readings. To  nd out whether special sample preparation is necessary, take several readings to see if readings (a
w
and time) stabi­lize. If continued readings take longer than six minutes, remove the sample and take a reading of a veri cation standard.  is will ensure the sample itself is causing the long read time, and that there is not a problem with your instrument. If the veri cation standard also takes longer than six minutes to test, the chamber may be dirty. Refer to Chapter 6 for cleaning procedures.
AquaLab
8. Sample Preparation
48
Coated and Dried Samples
Samples with high sugar or fat coatings often require multiple read­ings, because it takes longer for them to equilibrate. If this is the case for your samples, it is not a problem with your instrument; it simply means that your particular sample takes longer than most to equilibrate.
To reduce the time needed to take an water activity reading for coat­ed or dried samples, you can crush or slice the sample before sam­pling.  is increases the surface area of the sample, thus decreasing reading times. Keep in mind, however, that modifying some samples may alter their water activity readings.
For example, a candy may have a soft chocolate center and a hard outer coating.  e water activity reading for the center and the outer coating are di erent, so one would need to evaluate which part of the sample needed to be measured before crushing it. When the candy is crushed, the water activity will represent the average water activity of the entire sample; whereas leaving the candy whole will give a read­ing for the coating, which may act as a barrier to the center.
Slow Water-Emitting Samples
Some extremely dry, dehydrated, highly viscous water-in-oil (but­ter), high fat, or glassy compositions may require multiple tests due to their slow water-emitting properties.  is is because the slow emission of water decreases the change in water activity su ciently that the instrument determines the test to be complete, even though changes in water activity are still occuring.  e most e ective way to test these types of samples is to run them in the AquaLab using the continous or custom mode and wait for the water activity readings to stablize.
AquaLab
8. Sample Preparation
49
For faster reading, it is important to have the water activity of the chamber at or below the water activity of these type of samples.  is causes the sample to release water to the vapor phase and equilibrate with the chamber. If the water activity of the head-space is greater than this type of sample, a long period of time will be required to reach equilibrium and the water activity of the sample may be a ected.
Volatile Samples
AquaLab will give accurate readings on most samples. However, samples with certain volatiles in high enough concentrations may give inaccurate water activity values.  is is because the volatiles condense on the mirror during the reading process, but do not evaporate from the mirror as water does. As a result, the reading on samples with volatiles will not be accurate.  e concentration of vol­atiles that will cause interference is variable and matrix dependent.  e most e ective method to determine if volatiles are a problem is to compare dewpoint readings to capacitance readings. If the dew­point readings are more than 0.0153 higher than the capacitance readings, volatiles are likely a problem.
Decagon’s Series 4TEV is designed for measuring volatiles such as propylene glycol and ethanol.  e Series 4TEV contains both a chilled-mirror dewpoint and a capacitance sensor. Simply choose the sensor you want to use from the menu in the instrument.  e only di erence in operation is a lower accuracy of ±0.015 aw for the capacitance sensor. All other operations and features will be the same, including measurement times and adjusting for linear o set. After measuring volatiles with the volatiles sensor, it is a good idea to clean the chamber and run charcoal before switching to the dew point sensor.
AquaLab
8. Sample Preparation
50
Low Water Activity
When a sample’s water activity value is below the cooling capacity of the Series 4, your AquaLab will display an error message indicating the lowest reading it attained on that particular sample. See Chapter 12’s troubleshooting problem #5 for possible solutions.
If your sample is not below 0.03 a
w
but is still getting the error mes-
sage, refer to Chapter 12 for other possible explanations.
Samples Not at Room Temperature
Samples that are 4°C colder or warmer than the instrument (cham­ber) temperature will need to equilibrate to instrument tempera­ture before a fast, accurate reading can be made. Rapid changes in temperature over short periods of time will cause the water activ­ity readings to rise or fall until the temperature stabilizes. When the temperature stabilizes within one or two degrees of the chamber temperature, you can proceed with normal measurements.
High-water activity samples that are warmer than the chamber tem­perature can cause condensation inside the measuring chamber, which will adversely a ect subsequent readings. A warning message appears (Sample too hot) if the sample temperature is more than 4°C above chamber temperature. If this message appears, immediately remove the sample from the instrument, place a lid on the cup, and allow the sample to cool to within 4°C of the instrument before measuring.
Samples that are lower than 4°C of the instrument’s temperature will cause long read times.  e sample temperature must be within one or two degrees of the chamber temperature before fast, accurate readings can be made.
AquaLab
8. Sample Preparation
51
NOTE: Powdery substances can be blown by the fan so be sure not to over ll the sample cup and verify the cleanliness of the sample chamber before reading a new sample.
AquaLab
9. Taking a Reading
52
9. Taking a Reading
Measurement Steps
Once you have veri ed for cleanliness, calibration and prepared your sample, you are ready to take readings.  e process is simple:
• Move the chamber lever to the Open position and lift the cham­ber lid.
• Check the top lip and outside of the sample cup to make sure they are free from sample residue and that sample cup isn’t over­filled. (remember, an over-filled sample cup may contaminate the chamber’s sensors).
• Place your prepared sample cup in the chamber. The sample cup lid must be removed while in the testing chamber for correct functionality.
• Close the chamber lid and move the lever to the Read position. This will seal the chamber and start the reading.
In 1 to 2 minutes, the  rst water activity measurement will be dis­played on the LCD (this is an intermediate reading and not the  nal water activity). Length of read times may vary depending on temperature di erences between the chamber and your sample, and other properties of your sample.
How AquaLab Takes Readings
AquaLab’s reading cycle continues until the rate of change of three consecutive readings are less than 0.0005 a
w
of each other.  e in-
strument crosses the dew threshold numerous times to ensure equi-
AquaLab
9. Taking a Reading
53
librium and the accuracy of readings. When the instrument has  n­ished its read cycle, the water activity is displayed, the read time is displayed, the spinning measurement icon is replaced by the Store icon and, if enabled, you will hear a series of beeps.
Cautions!
• Never leave a sample in your AquaLab after a reading has
been taken. The sample may spill and contaminate the in­strument’s chamber if the instrument is accidentally moved or jolted.
• Never try to move your instrument after a sample has been
loaded. Movement may cause the sample material to spill and contaminate the sample chamber.
• If a sample has a temperature that is 4°C higher (or more)
than the AquaLab’s chamber, the instrument will beep and display a warning as shown below. Remove the sample until it is at room temperature.
Although the instrument will measure warmer samples, the readings may be inaccurate. Warm samples can cause condensation in the
AquaLab
9. Taking a Reading
54
chamber if they have a high water activity. It is best to remove the sample from the instrument, place a lid on the cup and allow the sample to cool before reading.
• The physical temperature of the instrument should be between 15 - 50°C. Between these ambient temperatures, AquaLab will measure samples of similar temperature quickly and accurately. The AquaLab Series 4TE and 4TEV have temperature control capabilities that enable them to read samples at temperatures different from ambient temperature, but no higher than 50°C.
• If a sample has a water activity lower than about 0.03, AquaLab will display the < symbol (see below) notifying you that your sample is too dry to be accurately measured by the AquaLab.
• If you know that your sample’s water activity is above what the screen is telling you, your instrument’s sensors may have been contaminated and will need to be cleaned (see Chapter 6) or serviced (see Chapter 13).
AquaLab
10. Duo Operation (Optional)
55
10. Duo Operation (Optional)
Previously, measuring moisture content and water activity required di erent instruments. Now it is possible to determine both moisture content and water activity with one machine.  e Series 4TE can be upgraded to Series 4TE DUO which can display moisture content simultaneously with water activity.
To calculate moisture content using water activity requires an under­standing of the relationship between the two parameters.  is rela­tionship, referred to as the moisture sorption isotherm, is complex and unique to each product type. A product’s isotherm can be used to calculate moisture content based on a water activity measurement.  is is most easily accomplished using a model that characterizes the isotherm. For additional information about sorption isotherms and models, please refer to Chapter 3.  e DUO generates water activ­ity values just as a Series 4TE, but then it uses preloaded product speci c isotherm models to calculate moisture content and present it on the screen with the water activity. For information about upgrad­ing your Series 4TE to a Series 4TE DUO, please contact Decagon Devices.
AquaLab
10. Duo Operation (Optional)
56
Obtaining Product Isotherm Models
Since the isotherm relationship for each product is unique, each product’s isotherm model must be determined ex­perimentally.  is only needs to be done once, but must be done prior to testing moisture content with the DUO.  ere are several strategies that can be used to generate models. Please contact Decagon for information on model development.
Loading and Organizing Product Models
A Product’s model must be loaded into the Series 4TE DUO before it can calculate moisture content. Each product must have its own model and the model can either be loaded at the factory by Decagon or by using the AquaLink RG software program.  is software is included with each DUO purchase or upgrade. Product model  les generated by Decagon are sent to customers via email and can then be loaded into the instrument by connecting to the instrument us­ing the AquaLink RG software.  e software uses a model loading tool to add and remove product models from the Series 4TE DUO, allowing the user to control and organize their product models. Up to 100 models can be stored on the instrument.  e AquaLink RG software can also download data (including moisture content) from the instrument, present the data in table form,  lter the data, and print reports.
AquaLab
10. Duo Operation (Optional)
57
Measuring Moisture Content
With the product models loaded into the instrument, the Series 4TE DUO can generate moisture content and water activity simul­taneously.
Selecting a Product for Analysis
• With the Series 4TE DUO turned on, toggle to the configura-
tion screen by pressing the menu button.
• At the configuration screen, scroll down and select moisture
content.
AquaLab
10. Duo Operation (Optional)
58
• A list of available models will be listed by name.
• Select the model for the product to be analyzed. Selecting “None” will not select any model.
Taking a Reading
• Readings are taken with the DUO the same as outlined in Chap­ter 9. First, return to the main screen.
• The product chosen for analysis will be shown in the tab at the top of the screen. If a different product is desired to be analyzed, it is possible to scroll through all of the available product models on the screen by pressing the up and down buttons. This eliminates
AquaLab
10. Duo Operation (Optional)
59
the need to return to the configuration screen to change products. When the tab at the top shows “Measurement”, no model is se­lected and only water activity will be displayed on the screen.
• Place a sample in the chamber and begin testing by sliding the
lever left to the read position. For information about sample preparation, please see Chapter 8 and for additional information about running a test, please see Chapter 9.
• When the test is complete, the screen will display the water activ-
ity and moisture content for the product selected. If the wrong model was selected by accident, the up and down buttons can be used to toggle through all of the product models and the mois­ture content value will adjust based on the model selected.
AquaLab
10. Duo Operation (Optional)
60
• The test can be saved to the instrument’s memory by pressing the button under the save icon. An annotation can be added if desired. If autosave has been selected, the data will already be saved but without any annotation.
The results can be viewed by moving to the data screen (press the right most button, which is below the document icon, to toggle between tabs) as shown in Chapter 5 under the Data Tab section. The only difference will be that moisture content data will now ap­pear in the upper right column on the detailed information screen.
Moisture Content Adjustment
AquaLab Duo moisture analyzers calculate moisture content val­ues based on water activity readings by utilizing models stored within the instrument. Because moisture content results vary be­tween reference methods, it is important to ensure that the mod­el in the instrument correlates well with your reference method moisture contents (i.e. Karl Fischer titration, oven loss on dry­ing, etc.). Moisture content di erences among various methods are usually linear and can be easily corrected with a linear o ­set.  erefore, if moisture contents calculated with the AquaLab Duo instrument are not agreeing with your reference method, the problem can likely be addressed by adjusting a linear o set.
AquaLab
10. Duo Operation (Optional)
61
When to Adjust for Linear O set
Reference methods can di er between labs, so it is a good idea to check for a linear o set upon receipt of a new isotherm model from Decagon. In addition, the linear o set should be adjusted if moisture contents being calculated by the AquaLab Duo instru­ment are consistently higher or lower for a product than your refer­ence method values over several samples.
How to Adjust for Linear O set or Create a New Model Based O an Old Model
1. For the product whose model is to be o set, collect 3 subsam­ples for analysis.
2. Use 2 of the subsamples to run duplicate reference method moisture contents and determine the average moisture content.
3. Once you’ve obtained a reference moisture content, navigate to the calibration screen in the Con guration menu of the AquaLab Duo Moisture Analyzer and select %Moisture from the list of calibration types.
4. Select Edit to edit and replace an existing model. Select New if you would like to create a new o set model with this calibra-
AquaLab
10. Duo Operation (Optional)
62
tion instead of replacing the existing model. Pressing Enter opens a model screen listing all models currently loaded on the instrument.
5. Scroll down to  nd the model for the product to be o set and press Enter. If you selected New, choose a reference model to use as a basis for your new model.
6.  e screen will instruct you to place a sample of the product in the testing chamber. Place the 3rd subsample from step 1 in a sample cup, then put the sample cup in the testing chamber of the AquaLab Duo instrument and close the lid.
7. Press Enter to begin a reading.
AquaLab
10. Duo Operation (Optional)
63
8. Once the reading is complete, a screen will display the water activity measured as well as the moisture content based on the target model. Adjust the moisture content reading using the up and down arrows until it matches the moisture content value obtained from your reference method and click Save.
Note: If you chose to edit an existing model, pressing save updates the model but keeps the same name. If you chose to create a new model, pressing save will bring up an annotation screen where you will enter the new name for the model. Pressing the cancel button will return you to the Con guration menu and cancel the moisture content adjustment.
9. Re-measure the sample again in normal sampling mode. It should now read the corrected moisture content value you provided in the previous step.
If your moisture content readings are still inconsistent, contact Decagon by email at support@decagon.com or by phone at 509­332-2756 (800-755-2751 in US and Canada) for further instruc­tions. If you purchased your AquaLab Instrument from one of our international distributors, please contact them for local service and support.
AquaLab
10. Duo Operation (Optional)
64
Restore Original Moisture Content Model Settings
To restore the original model settings, do the following:
1. Navigate to the calibration screen in the Con guration menu of the AquaLab Duo Moisture Analyzer and select %Moisture from the list of calibration types.
Note: If you don’t see %Moisture as an option you may not have a Duo model or you may not have any models installed.
2. Scroll down to Edit and press the Enter Button.
3. Select the model that you would like to reset to its original set­ting and press the Enter button.
AquaLab
10. Duo Operation (Optional)
65
4. Scroll down to Defaults and press the Enter icon button to restore to defaults. To cancel and return to the main menu, press the Cancel icon button. After pushing the Enter icon button, the following screen will appear:
5. To restore the original model settings, press the Check icon button. To cancel and return to the main menu, press the Cancel button. After pressing the Check icon button, the fol­lowing screen will appear:
AquaLab
10. Duo Operation (Optional)
66
6. To return to the main menu screen, press the Check icon button.
How to Delete Models
If you  nd that a model is no longer needed, you have the option of deleting the model directly from the instrument. If the model is deleted, other users will no longer be able to use it. Also, if the model hasn’t been backed up with AquaLink RG, you will not be able to recover the model at a later time.
1. Navigate to the calibration screen in the Con g menu of the AquaLab Moisture Analyzer and select %Moisture from the list of calibration types.
Note: If you don’t see %Moisture as an option you may not have a Duo
model or you may not have any models installed.
AquaLab
10. Duo Operation (Optional)
67
Scroll down to Delete and press the Enter Button.
2. Select the model you would like to delete and press the Enter icon button to continue or the Cancel icon button to cancel.
3. Upon pressing Enter, the following screen should appear indi­cating the model to be deleted. Press the Check icon to delete the model or press the Cancel icon to cancel the deletion.
AquaLab
11. Computer Interface
68
11. Computer Interface
Your AquaLab was shipped to you with a standard RS-232 interface cable. Using this, you can load data to a computer using Decagon’s AquaLink RG program or your computer’s terminal program for further analysis and storage.
AquaLink RG
An optional software program, AquaLink Report Generator (RG), is available for use with your AquaLab. AquaLink RG is a Windows based program designed for data collection and customized report generation for all Series 4 AquaLab models. AquaLink RG logs water activity, temperature, time of measurement, and date stamps along with other information. AquaLink RG also has sample iden­ti cation and comment  elds that you can use to help annotate the data your AquaLab is gathering.
A 30 day trial cd of this program is attached to the front cover of this manual. If you are interested in purchasing the full version of AquaLink RG, contact Decagon or your local distributor. If you have purchased the AquaLab 4TE DUO you will automatically receive the full version of AquaLink RG with your manual.
Using Windows Hyperterminal
To use Hyperterminal with your AquaLab, follow these steps:
• On your computer, press the Start button and select Programs > Accessories > Hyperterminal and click on the Hyperterminal icon.
AquaLab
11. Computer Interface
69
• At the prompt, choose a name for this program (AquaLab is a
good one) and choose an arbitrary icon above to represent it. In future downloads, you will be able to click on this icon in have it already set up for you to download. Click the OK button.
• A pop-up menu labeled “Connect To” will appear. Click on the
scroll bar on the bottom of the screen labeled “Connect Using” and select the COM Port your RS-232 cable is connected to.
• A pop-up menu labeled “ COM Properties” will appear, showing
the port settings for the COM port you selected. Make sure the settings are the following: Bits per second, 9600; 8 databits, no parity, 1 stop bit, and flow control set to “none”. Click OK.
• Plug your RS-232 cable to the COM port you selected and con-
nect it to your AquaLab. Begin sampling. AquaLab’s data will be displayed on screen as it samples.
• When you are finished sampling, you can print the data in the
terminal session, or cut and paste it to a spreadsheet or text editor. To save the data, go into the Transfers menu and select “Capture text,” and designate where it should be saved.
AquaLab
12. Troubleshooting
70
12. Troubleshooting
AquaLab is a high performance, low maintenance instrument, de­signed to have few problems if used with care. Unfortunately, some­times even the best operators using the best instruments encounter technical di culties. Below is quick reference guide that will direct you to detailed solutions of some problems that may occur. If these remedies still don’t resolve your problem, then please contact Deca­gon for help (see Customer Support in Chapter 1). Here is a list of some problems that may occur.
NOTE: If you purchased your Decagon instrument from one of our inter­national distributors, please contact them for local service and support.
Troubleshooting Quick Guide
If this problem occurs: Refer to:
AquaLab won’t turn on ................................................ Problem #1
Readings are slow or inconsistent ................................ Problem #2
Water activity readings on solutions are ....................... Problem #3
too high/low to adjust
Screen displays “Sample too hot” ............................... Problem #4
Screen displays “aw < 0.0” ........................................... Problem #5
Dew point sensor failure ............................................. Problem #6
Veri cation is not correct............................................. Problem #7
AquaLab
12. Troubleshooting
71
Troubleshooting Quick Guide (Continued)
If this problem occurs: Refer to:
Screen displays “Crystal failure”................................... Problem #8
Screen displays “Contaminated Mirror” ...................... Problem #9
Screen displays “Firmware is corrupted” .................... Problem #10
Screen displays “Readings are disabled” ..................... Problem #11
How do I activate my Demo? .................................... Problem #12
DUO Model--Test was run with wrong model. ......... Problem #13
DUO Model--%MC displayed is not correct. ........... Problem #14
DUO Model--%MC is not shown on screen ............. Problem #15
DUO Model--Moisture Content is not correct. ........ Problem #16
1. PROBLEM:
AquaLab won’t turn on.
SOLUTIONS:
1) Check to make sure your power cord is securely attached to the back of the instrument and it is plugged into the power outlet.
2) A power surge may have caused a fuse to blow. To change the fuses, follow these instructions:
a. Unplug the power cord.
AquaLab
12. Troubleshooting
72
b. Locate the panel where the power cord plugs in.  e fuse box
is on the right side of that panel. Press in on the release tab and pull the fuse-holder out. Pull the broken fuse(s) out and replace with a 1.25 Amp 250V fuse.
Caution: Do not use any other kind of fuse or you will risk dam­age to your instrument as well as void your warranty.
c. Replace the fuse-holder and push it into the fuse-well until
the release tab snaps in place.
d. Re-connect the power cord and turn your instrument on. If
the fuse blows again, a failed component may be causing the problem. Contact Decagon to make arrangements for repairs.
2. PROBLEM:
Readings are slow or inconsistent.
SOLUTIONS:
1)  e sample chamber may be dirty. Refer to Chapter 6 for di­rections on cleaning the sample chamber.
2)  e temperature di erence between the sample and the block chamber may be too great.  e sample will need to equili­brate to instrument temperature before a fast, accurate read­ing can be made. (Refer to Chapter 8, Samples Not at Room Temperature.)
3) Some products absorb or desorb moisture very slowly, caus­ing measurements to take longer than usual, and nothing can
AquaLab
12. Troubleshooting
73
be done to speed up the process. Refer to Chapter 8 for fur­ther explanation.
4) Your sample may contain volatiles. Volatiles are known to cause unstable readings, because they condense on the sur­face of the chilled mirror and alter readings. Please refer to the volatiles section in Chapter 8 for hints on reducing di ­culties with measuring samples with propylene glycol. If you have further questions regarding the measurement of vola­tiles contact Decagon.
5) A fan blade in the block chamber may be broken or bent. If even salt standards take a long time to read, and the sample chamber is clean, you may have a broken chamber fan blade.  is is especially likely if you have just cleaned the chamber. If you suspect this may have happened, contact Decagon for details on replacement.
3. PROBLEM:
Water activity readings on veri cation standards are too high/low and a linear o set adjustment cannot be made any higher/lower.
SOLUTIONS:
1)  e thermopile in your chamber, which measures sample temperature, may have become contaminated. Refer to Chap­ter 6 for directions on cleaning.
2)  e chamber mirror may be dirty. Refer to Chapter 6 for di­rections on cleaning.
4. PROBLEM:
Message on screen displays the following:
AquaLab
12. Troubleshooting
74
SOLUTION:
Your sample’s temperature is too high for the instrument to equili­brate with it in a reasonable amount of time.  e instrument and sample need to be in temperature equilibrium before accurate measurements can be made.  erefore, very cold samples will take a very long time to measure for the same reason. To avoid this problem, make sure to only measure samples that are at the same temperature as the instrument.
5. PROBLEM:
Message on screen displays the following (example):
AquaLab
12. Troubleshooting
75
SOLUTIONS:
1)  e sample is too dry for the instrument to read accurately. If your sample has a water activity that is less than below the detection limits of the instrument, this message will come up. Essentially, it means that there is not enough sample mois­ture to condense on the mirror and provide a reading.
2)  e mirror may be dirty. Try cleaning the mirror and cham­ber and measuring the sample again.
6. PROBLEM:
Message on screen displaying dew point sensor failure.
SOLUTION:
 e Cooler is damaged and will need to be serviced by Decagon. See Chapter 12 for detailed instructions.
AquaLab
12. Troubleshooting
76
7. PROBLEM
Veri cation is not correct.
SOLUTIONS:
1)  e sample chamber and mirror need to be cleaned. See Chapter 6 for detailed cleaning instructions. If veri cation is still not correct, then linear o set has occurred.
2) Verify and Adjust for Linear O set. After you have cleaned the sample chamber and mirror (Chpt. 7) you will need to use a Veri cation Standard to verify and adjust for Linear O set as described in Chapter 7.
8. PROBLEM:
Message on screen displays the following:
SOLUTION:
 e crystal that runs the  rmware is having trouble starting. Oc­casionally, cycling the power will solve the problem. If this mes­sage continues to appear, the instrument will need to be ser­viced by Decagon. See Chapter 13 for detailed instructions.
AquaLab
12. Troubleshooting
77
9. PROBLEM:
Message on screen displays the following:
SOLUTION:
 e mirror used for dewpoint measurements requires cleaning. Fol­low the instructions outlined in Chapter 6: Cleaning and Main­tenance before trying to run your sample again. If this message continues to appear, contact Decagon for further options.
10. PROBLEM:
Message on screen displays the following:
AquaLab
12. Troubleshooting
78
SOLUTION:
 e  rmware on the instrument is corrupted and needs to be re­loaded. To download new  rmware to the Series 4 models, the in­strument must be serviced by Decagon.
11. PROBLEM:
Message on screen displays the following:
SOLUTION:
 e trial period for your Demo Unit has expired. Contact Decagon Devices for additional options.
12. PROBLEM:
Message on screen displays the following:
AquaLab
12. Troubleshooting
79
SOLUTION:
In order to begin your trial period for your AquaLab Series 4 instru­ment, you will need to contact Decagon Devices for instructions on how to activate your demo.
13. DUO PROBLEM:
Test was run with wrong model.
SOLUTION:
1) On the measurement screen, toggle to the correct model us­ing the up and down arrow keys.  e moisture content value will be updated to correspond with the model selected.
2) If the correct model is not available, the model may not be loaded on the instrument.
a. To determine which models are loaded on the instrument,
cycle to the menu tab, select Moisture Content and then the loaded models will appear.
3) If the correct model is not available, load the appropriate model using Aqualink RG Software.  e AquaLab DUO can hold a total of 100 models at any one time. You may need to remove a model using the RG Software or use the delete op­tion in the % moisture calibration menu before you can add a new one. Any model that is removed from the instrument with AquaLink RG will be stored and may be reloaded again later.
AquaLab
12. Troubleshooting
80
14. DUO PROBLEM:
Moisture Content displayed is not correct.
SOLUTION:
1) Model selected may not be correct for the product being test­ed.
a. Toggle through the available models to  nd a more appropri-
ate model.
b. If the model is correct but not giving correct moisture con-
tent values it may be necessary to generate a new model for the product or update an existing model. For information about generating a model, contact Decagon Devices for up­dating a model, refer to Chapter 10: Duo Operation.
15. DUO PROBLEM:
Moisture content does not show up on the screen.
SOLUTION:
Moisture content has not been activated.
1) Toggle to menu tab, select moisture content, and select the appropriate model.
a. If no models appear in moisture content screen, models will
need to be reloaded using AquaLink RG software.
b. If moisture content is not an active selection, the DUO fea-
ture may not be active. Content Decagon Devices to learn how to activate the DUO feature.
AquaLab
12. Troubleshooting
81
16. DUO PROBLEM:
Message on the screen displays the following:
SOLUTION:
1) When a moisture content reading is not shown, the water ac­tivity or temperature for that reading is beyond the scope of the moisture sorption isotherm.  is can happen under the following conditions:
a.  e isotherm equation calculates a moisture content that
is less than 0% or greater than 100% with the given water activity.
b.  e control temperature is signi cantly di erent than the
isotherm temperature.
Make sure that the sample’s water activity and the instrument’s controlling temperature are within the scope of the selected mois­ture sorption isotherm model.
AquaLab
12. Troubleshooting
82
Diagnostic Screen
If, after cleaning your instrument and reading the other trouble­shooting hints, you have reason to believe that one of the compo­nents of your AquaLab may be causing measurement error, you may access a screen that will display values for component perfor­mance.  is is done by navigating to the Con guration tab and then by scrolling down to the diagnostics option. Press enter and you will be given a list of components and their values.
 is screen shows typical values for the dew point method. Lid, base and sample temperatures may  uctuate but should not change more than 0.03 degrees. Typical ranges for the lid, base and sample temperatures is between 24.5 and 25.5 degrees.
If the mirror temperature is at lid temperature, the cooler has failed and must be replaced. If the mirror is below the lid temperature or appears to be random, the thermocouple wire is broken and must be repaired.
A typical optical range is between 500 mV and 2900 mV .
For capacitance mode, not shown here, the RH percentage should always be between 0 and 100%.
AquaLab
13. Support and Repair
83
13. Support and Repair
NOTE: If you purchased your AquaLab from one of our international distributors, please contact them.  ey will be able to provide you with local support and service.
When encountering problems with your AquaLab (that can’t be re­solved with the help of this manual), please contact Decagon Cus­tomer Support at support@decagon.com, 800-755-2751 (US and Canada), 509-332-2756 (International) or fax us at (509) 332-5158. Please have the serial number and model of the instrument ready.
All AquaLabs returning to Decagon for servicing must be accompa­nied with a Return Material Authorization (RMA) form. Prior to shipping the instrument, please contact a Decagon customer sup­port representative to obtain an RMA.
Shipping Directions:  e following steps will help to ensure the safe shipping and process­ing of your AquaLab.
1) Ship your AquaLab in its original cardboard box with sus­pension packaging. If this is not possible, use a box that has at least 4 inches of space between your instrument and each wall of the box.
2) Place the AquaLab in a plastic bag to avoid dis guring marks from the packaging.
3) Don’t ship the power cord or serial cable.
4) If the original packaging is not available, pack the box mod­erately tight with packing material (e.g. styrofoam peanuts or
AquaLab
13. Support and Repair
84
bubble wrap), ensuring the instrument is suspended in the pack­ing material.
5) On the RMA form, please verify the ship to and bill to infor­mation, contact name, and problem description. If anything is incorrect please contact a Decagon representative.
6) Tape the box in both directions for added support.
7) Include the RMA number in the attention line on the ship­ping label.
Ship to: Decagon Devices Inc. ATTN: RMA (insert your RMA #) 2365 NE Hopkins Court Pullman, WA 99163
Repair Costs
Manufacturer’s defects and instruments within the three-year warran­ty will be repaired at no charge. Non-warranty repair charges for parts, labor and shipping will be billed to you. An extra fee may be charged for rush work. Decagon will provide an estimated repair cost, if requested.
Loaner Service
Decagon has loaner instruments to keep you measuring water activity while your instrument is being serviced. If your AquaL­ab is still under calibration warranty or you have a service plan with your instrument, there is no charge for the loaner service.
AquaLab
14. Further Reading
85
14. Further Reading
Water Activity  eory & Measurement
Bousquet-Ricard, M., G. Qualyle, T. Pharm, and J. C. Cheftel.
1980. Comparative study of three methods of determining water activity in intermediate moisture foods. Lebensm Wiss Technol 13:169-173.
Cazier, J.B., and V. Gekas. 2001. Water activity and its prediction: a review. International Journal of Food properties 4(1):35-43.
Chirife, J., G. Favetto, C. Ferro-Fontán, and S.L.Resnik. 1983.  e water activity of standard saturated salt solutions in the range of intermediate moisture foods. Lebensm Wiss Technol 16:36-38.
Duckworth, R. 1975. Water relations of foods. Academic Press, New York.
Gómez, R., and J. Fernandez-Salguero. 1992. Water activity and chemical composition of some food emulsions. Food Chem 45:91-93.
Greenspan, L. 1977. Humidity  xed points of binary saturated aqueous solutions. J Res Nat Bur Stand - A Phys Chem 81A:89-96.
Karmas, E. 1981. Measurement of moisture content. Cereal Foods World 26:332-334.
Kitic, D., D.C. Pereira-Jardim, G.J. Favetto, S.L. Resnik, and J. Chirife. 1986.  eoretical prediction of the water activity of stan­dard saturated salt solutions at various temperatures. Journal of Food Science 51:1037-1042.
AquaLab
14. Further Reading
86
abuza, T.P., and R. Contreras-Medellin. 1981. Prediction of moisture protection requirements for foods. Cereal Foods World 26:335-343.
Labuza, T.P., K. Acott, S.R.Tatini, R.Y. Lee, J. Flink, and W. Mc­Call. 1976. Water activity determination: A collaborative study of di erent methods. Journal of Food Science 41:910-917.
Marcolli, C., and  . Peter. 2005. Water activity in polyol/water systems: new UNIFAC parameterization. Atmospheric Chemistry and Physics 5:1545-1555.
Ninni, L., M.S. Camargo, and A.J.A. Meirelles. 2000. Water activity in polyol systems. Journal of Chemical and Engineering Data 45:654-660.
Prior, B.A. 1979. Measurement of water activity in foods: A review. Journal of Food Protection 42:668-674.
Rahman, M.S. and S.S. Sablani. 2001. Measurement of water ac­tivity by electronic sensors. P. A2.5.1-A2.5.4 In R.E.Wrolstad (ed.) Current Protocols In Food Analytical Chemistry. John Wiley & Sons, Inc., New York.
Rahman, M.S., S.S. Sablani, N. Guizani, T.P. Labuza, and P.P. Le­wicki. 2001. Direct manometic determination of vapor pressure. P. A2.4.1-A2.4.6. In R.E. Wrolstad (ed.) Current Protocols In Food Analytical Chemistry. John Wiley & Sons, Inc., New York.
Reid, D.S., A.J. Fontana, M.S. Rahman, S.S. Sablani, T.P. Labuza, N. Guizani, and P.P. Lewicki. 2001. Vapor pressure measurements of water p. A2.1.1-A2.5.4. In R.E. Wrolstad (ed.) Current Protocols In Food Analytical Chemistry. John Wiley & Sons, Inc., New York.
AquaLab
14. Further Reading
87
Reid, D.S. 1976. Water activity concepts in intermediate moisture foods. p. 54-65. In R.Davies, G.G.Birch, and K.J.Parker (ed.) Inter­mediate Moisture Foods. Applied Science Publishers, London.
Richard, J., and T.P. Labuza. 1990. Rapid determination of the wa­ter activity of some reference solutions, culture media and cheese using a dew point method. Sci. des Aliments 10:57-64.
Roa,V., and M.S.Tapia de Daza. 1991. Evaluation of water activ­ity measurements with a dew point electronic humidity meter. Leb­ensm Wiss Technol 24:208-213.
Rodel, W. 2001. Water activity and its measurement in food. P. 453-
483. In E. Kress-Rogers, and C.B. Brimelow (ed.) Instrumentation and sensors for the food industry. CRC Press LLC, Boca Raton, FL.
Roos, K.D. 1975. Estimation of water activity in intermediate mois­ture foods. Food Tech 29:26-30.
Scott, V.N., and D.T. Bernard. 1983. In uence of temperature on the measurement of water activity of food and salt systems. Journal of Food Science 48:552-554.
Snavely, M.J., J.C. Price, and H.W. Jun. 1990. A comparison of three equilibrium relative humidity measuring devices. Drug Dev. Ind. Pharm. 16:1399-1409.
Stamp, J.A., S. Linscott, C. Lomauro, and T.P. Labuza. 1984. Mea­surement of water activity of salt solutions and foods by several elec­tronic methods as compared to direct vapor pressure measurement. Journal of Food Science 49:1139-1142.
AquaLab
14. Further Reading
88
Stolo , L. 1978. Calibration of water activity measuring instru­ments and devices: Collaborative study. Journal of the Association of O cial Analytical Chemists 61:1166-1178.
Troller, J.A. 1983. Methods to measure water activity. Journal of Food Protection 46:129-134.
Troller, J.A., and J.H.B Christian. 1978. Water activity and food. Academic Press, New York.
Troller, J.A., and V.N. Scott. 1992. Measurement of water activity (Aw) and acidity. p. 135-151. In C. Vanderzant, and D.F. Splittstoesser (ed.) Compendium of Methods for the Microbiological Examination of Foods. American Public Health Association, Washington, D.C.
van den Berg, C. 1986. Water activity. p. 11-36. In D. MacCarthy (ed.) Concentration and drying of foods. Elsevier Applied Science Publishers, London.
van den Berg, C. 1991. Food-water relations: Progress and integra­tion, comments and thoughts. In H. Levine, and L. Slade (ed.) Wa­ter Relationships in Foods. Plenum Press, New York.
van den Berg, C., and S. Bruin. 1981. Water activity and its estima­tion in food systems:  eoretical aspects. p. 1-61. In L.B. Rockland, and G.F. Stewart (ed.) Water Activity: In uences on Food Quality. Academic Press, New York.
Vega-Mercado, H., andG.V. Barbosa-Canovas. 1994. Prediction of water activity in food systems: A review on theoretical models. Re­vista Espanola De Ciencia Y Tecnologia De Alimentos 34:368-388.
AquaLab
14. Further Reading
89
Vega-Mercado, H., B. Romanach, and G.V. Barbosa-Canovas.
1994. Prediction of water activity in food systems: A computer pro­gram for predicting water activity in multicomponent foods. Revista Espanola De Ciencia Y Tecnologia De Alimentos 34:427-440.
Vos, P.T., andT.P. Labuza. 1974. Technique for measurements of wa­ter activity in the high aw range. J. Agric. Food Chem. 22:326-327.
Voysey, P. 1993. An evaluation of the AquaLab CX-2 system for measuring water activity. F. M. B. R. A. Digest No. 124 24-25.
Food Safety and Microbiology
Bei, Z.H., and R.-M.J. Nout. 2000. E ects of temperature, wa­ter activity and gas atmosphere on mycelial growth of tempe fungi Rhizopus microsporus var. microcporus and R. microsporus var. oligosporus. World Journal of Microbiology and Biotechnology 16:853-858.
Beuchat, L.R. 1981. Microbial stability as a ected by water activity. Cereal Foods World 26:345-349.
Brandt, L. 1996. Bound for success. Controlling water activity gives technologists the edge in developing safe, shelf-stable foods. Food Formulating 2:41-48.
Chirife, J., andM.P. Buera. 1994. Water activity, glass transition and microbial stability in concentrated/semimoist food systems. Journal of Food Science 59:921-927.
Chirife, J., andM.P. Buera. 1995. A critical review of some non­equilibrium situations and glass transitions on water activity values
AquaLab
14. Further Reading
90
of foods in the microbiological growth range. Journal of Food Engi­neering 25:531-552.
Chirife, J., andM.P. Buera. 1996. Water activity, water glass dynam­ics, and the control of microbiological growth in foods. Critical Rev. in Food Sci. Nutr. 36:465-513.
Farberm, J.M., F. Coates, and E. Daley. 1992. Minimum water ac­tivity requirements for the growth of Listeria monocytogenes. Lett Appl Microbiol 15:103-105.
Franks, F. 1991. Water activity: a credible measure of food safety and quality? Trends Food Sci Technol March:68-72.
Garcia de Fernando, G.D., O. Diaz, M. Fernandez, and J.A. Or­donez. 1992. Changes in water activity of selected solid culture me­dia throughout incubation. Food Microbiology 9:77-82.
Gibson, A.M., J. Baranyi, J.I. Pitt, M.J. Eyles, and T.A. Roberts. 1994. Pre­dicting fungal growth:  e e ect of water activity on Aspergillus  avus and related species. International Journal of Food Microbiology 23:419-431.
Goaleni, N., J.E. Smith, J. Lacey, and G. Gettinby. 1997. E ects of temperature, water activity, and incubation time on production of a atoxins and cyclopiazonic acid by an isolate of Aspergillus  avus in surface agar culture. Appl Environ Microbiol 63:1048-1053.
Hardman, T.M. 1988. Water and food quality. Elseiver Press, London.
Hocking, A.D., and B.F. Miscamble. 1995. Water relations of some Zygomycetes isolated from food. Mycological Research 99:1113-1118.
AquaLab
14. Further Reading
91
Hocking, A.D., B.F. Miscamble, and J.I. Pitt. 1994. Water relations of Alternaria alternata, Cladosporium cladosporioides, Cladospori­um sphaerospermum, Curvulario lunata and Curvulario pallescens. Mycological Research 98:91-94.
Houtsma, P.C., A. Heuvelink, J. Dufrenne, and S. Notermans. 1994. E ect of sodium lactate on toxin production, spore germination and heat resistance of proteolytic Clostridium botulinum strains. Journal of Food Protection 57:327-330.
Kress-Rogers, E. 1993. Food quality measurement. Food Industry News September:23-26.
Kuntz, L.A. 1992. Keeping microorganisms in control. Food Prod­uct Design August:44-51.
Levine, H., and L. Slade. 1991. Water Relationships in Foods. Plenum Press, New York.
Li, K.Y., and J.A. Torres. 1993. Water activity relationships for se­lected mesophiles and psychrotrophs at refrigeration temperature. Journal of Food Protection 56:612-615.
Lopez-Malo, A., S. Guerrero, and S.M. Alzamora. 2000. Probabi­listic modeling of Saccharomyces cerevisiae inhibition under the ef­fects of water activity, pH, and potassium sorbate concentration. Journal of Food Protection 63:91-95.
Mannheim,C.H., J.X. Liu, and S.G. Gilbert. 1994. Control of water in foods during storage. Journal of Food Engineering 22:509-532.
Marauska, M., A. Vigants, A. Klincare, D. Upite, E. Kaminska, and
AquaLab
14. Further Reading
92
M. Bekers. 1996. In uence of water activity and medium osmolal­ity on the growth and acid production of Lactobacillus casei var. alactosus. Proceedings of the Latvian Academy of Sciences Section B Natural Exact and Applied Sciences 50:144-146.
Masana, M.O., and J. Baranyi. 2000. Growth/no growth interface of Brochothrix thermosphacta as a function of pH and water activ­ity. Food Microbiology 17:485-858.
Mattick, K. L., F. Jorgensen, J.D. Legan, M.B. Cole, J. Porter, H.M. Lappin-Scott, and T.J. Humphrey. 2000. Survival and  lamenta­tion of Salmonella enterica serovar Enteritidis PT4 and Salmonella enterica serovar Typhimurium DT104 at low water activity. Appl Environ Microbiol 66:1274-1279.
Mattick, K.L., F. Jorgensen, J.D. Legan, H.M. Lappin-Scott, and T.J. Humphrey. 2000. Habituation of Salmonella spp. at reduced water activity and its e ect on heat tolerance. Appl Environ Micro­biol 66:4921-4925.
Mattick, K.L., F. Jorgensen, J.D. Legan, H.M. Lappin-Scott, and T.J. Humphrey. 2001. Improving recovery of Salmonella enterica Serovar Typhimurium DT104 cells injured by heating at di erent water activity values. Journal of Food Protection 64:1472-1476.
McMeekin, T.A., and T. Ross. 1996. Shelf life prediction: Status and fu­ture possibilities. International Journal of Food Microbiology 33:65-83.
Miller, A.J. 1992. Combined water activity and solute e ects on growth and survival of Listeria monocytogenes. Journal of Food Protection 55:414-418.
AquaLab
14. Further Reading
93
Nakajo, M., and Y. Moriyama. 1993. E ect of pH and water activ­ity on heat resistance of spores of Bacillus coagulans. Journal of the Japanese Society for Food Science and Technology 40:268-271.
Nelson, K.A., and T.P. Labuza. 1994. Water activity and food poly­mer science: Implications of state on arrhenius and WLF models in predicting shelf life. Journal of Food Engineering 22:271-289.
Nesci, A., M. Rodrigues, and M. Etcheverry. 2003. Control of As­pergillus growth and a atoxin production using antioxidants at dif­ferent conditions of water activity and pH. Journal of Applied Mi­crobiology 95:279-287.
Nolan, D.A., D.C .Chamblin, and J.A. Troller. 1992. Minimal wa­ter activity levels for growth and survival of Listeria monocytogenes and Listeria innocua. International Journal of Food Microbiology 16:323-335.
Noorlidah, A., A. Nawawi, and I. Othman. 2000. Fungal spoilage of starch-based foods in relation to its water activity (aw). Journal of Stored Products Research 36:47-54.
Park,C.M., and L.R.Beuchat. 2000. Survival of Escherichia coli O157:H7 in potato starch as a ected by water activity, pH and tem­perature. Lett Appl Microbiol 31(5):364-367.
Petersson, S., and J. Schnuerer. 1995. Biocontrol of mold growth in high-moisture wheat stored under airtight conditions by Pichia anomala, Pichia guilliermondii, and Saccharomyces cerevisiae. Appl Environ Microbiol 61:1027-1032.
AquaLab
14. Further Reading
94
Pitt, J.I., and B.F. Miscamble. 1995. Water relations of Aspergillus  avus and closely related species. Journal of Food Protection 58:86-90.
Plaza, P., J. Usall, N. Teixido, and I. Vinas. 2003 E ect of water activity and temperature on germination and growth of Penicillium digitatum, P. italicum and Geoteichum candidum. Journal of Ap­plied Microbiology 94:549-554.
Quintavalla, S., and G. Parolari. 1993. E ects of temperature, water activity and pH on the growth of Bacillus cells and spore: A response surface methodology study. International Journal of Food Microbi­ology 19:207-216.
Rockland, L.B., and G.F. Stewart. 1981. Water activity: In uences on food quality. Academic Press, New York.
Rockland, L.B., and S.K. Nishi. 1980. In uence of water activity on food product quality and stability. Food Tech 34:42-59.
Saad, R.R. 1992. E ect of water activity on growth and lipids of xerophilic fungi, Aspergillus repens and Aspergillus amstelodami. Zentralblatt Fuer Mikrobiologie 147:61-64.
Salter, M.A., D.A. Ratkowsky, T. Ross, and T.A. McMeekin. 2000. Modelling the combined temperature and salt (NaCl) limits for growth of a pathogenic Escherichia coli strain using nonlinear logistic regression. International Journal of Food Microbiology 61:159-167.
Santos,J., T.M.Lopez-Diaz, M.C.Garcia-Lopez, M.C.Garcia-Fer­nandez, and A.Otero. 1994. Minimum water activity for the growth of Aeromonas hydrophila as a ected by strain, temperature and hu­mectant. Lett Appl Microbiol 19:76-78.
Loading...