Decagon WP4-T, WP4 Operator's Manual

WP
4
Dewpoint PotentiaMeter
for models WP4 and WP4-T
Operator’s Manual
Ve r s i o n 2 . 1
Decagon Devices, Inc.
Decagon Devices, Inc.
950 NE Nelson Court Pullman WA 99163 tel: (509) 332-2756 fax: (509) 332-5158 www.decagon.com/wp4 wp4@decagon.com
Trademarks
“WP4 Dewpoint PotentiaMeter” is a registered trademark of Decagon Devices, Inc.
©
1998-2003 Decagon Devices, Inc., All rights reserved.
WP
4
PotentiaMeter
Table of Contents
i
Contents
1. Introduction
. . . . . . . . . . . . . . . . . . . . . . . . 1
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Note to our WP4 Users . . . . . . . . . . . . . . . . . . . . . . . . . .3
Seller’s Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2. About the WP4
. . . . . . . . . . . . . . . . . . . . 5
WP4 Models and Options . . . . . . . . . . . . . . . . . . . . . .5
WP4 and Water Potential . . . . . . . . . . . . . . . . . . . . . . .5
How the WP4 works . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
WP4 and Temperature . . . . . . . . . . . . . . . . . . . . . . . . . .7
3. Getting Started
. . . . . . . . . . . . . . . . . . . 10
Components of your WP4 . . . . . . . . . . . . . . . . . . . . . . .10
Choosing a Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Preparing the WP4 for Operation . . . . . . . . . . . . . .12
Portability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
4. The Menus
. . . . . . . . . . . . . . . . . . . . . . . . . 15
The Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Changing Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Normal sampling mode . . . . . . . . . . . . . . . . . . . . . . . . . .16
Continuous mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Sample Equilibration Screen . . . . . . . . . . . . . . . . . . . . .23
5. Calibration and Verification
. . . . 25
Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Verification Standards . . . . . . . . . . . . . . . . . . . . . . . . . .25
WP
4
PotentiaMeter
Table of Contents
ii
When to Verify Calibration . . . . . . . . . . . . . . . . . . . . .26
How to Verify and Calibrate the WP4 . . . . . . . . .27
6. Sample Preparation
. . . . . . . . . . . . . . 31
Preparing the Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Dry Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Samples and Temperature . . . . . . . . . . . . . . . . . . . . . . .33
7. Taking a Reading
. . . . . . . . . . . . . . . . 36
How WP4 takes Readings . . . . . . . . . . . . . . . . . . . . . . .37
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Increasing your Accuracy . . . . . . . . . . . . . . . . . . . . . . .39
8. Computer Interface
. . . . . . . . . . . . . . . 41
Using Hyperterminal in Windows 98
and Windows 2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
9. Theory: Water potential
. . . . . . . . 44
Water potential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Measuring Water Potential . . . . . . . . . . . . . . . . . . . . .45
Effect of Temperature on Water Potential . . . . . .46
10. Cleaning and Maintenance
. . . . . 48
Tools Needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Cleaning the Block and Sensors . . . . . . . . . . . . . . . . .49
Checking Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
11. Repair Instructions
. . . . . . . . . . . . . . . 53
Shipping Directions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Repair Costs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Loaner Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
12. Troubleshooting
. . . . . . . . . . . . . . . . . . 56
Problems and Solutions . . . . . . . . . . . . . . . . . . . . . . . . . .56
WP
4
PotentiaMeter
Table of Contents
iii
Component Performance Screen . . . . . . . . . . . . . . . .63
13. Further Reading
. . . . . . . . . . . . . . . . . . 65
References: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Application Notes: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Appendix A
. . . . . . . . . . . . . . . . . . . . . . . . . . 67
Preparing Salt Solution . . . . . . . . . . . . . . . . . . . . . . . . . .67
WP
4
PotentiaMeter
Table of Contents
iv
WP
4
PotentiaMeter
Introduction
1
1. Introduction
Welcome to Decagon’s WP4 Dewpoint PotentiaMe­ter, the research standard for measuring water potential. WP4 is the fastest, most accurate, and most reliable instrument available for measuring water potential using the chilled-mirror dewpoint technique. Whether you are a scientist or a student, WP4 will suit your needs. It is easy to use and pro­vides fast, accurate results. We hope you find this manual informative and helpful in understanding how to maximize the capabilities of your WP4.
About this Manual
This manual is for use with the standard model WP4 and the temperature-controlled model WP4-T. Included in this manual are instructions for setting up your WP4, calibration, and maintaining and car­ing for your instrument. Please read these instruc­tions before operating WP4 to ensure that the instrument performs to its full potential.
Customer Service
If you ever need assistance with your WP4, or if you just have questions, there are several ways to con-
WP
4
PotentiaMeter
Introduction
2
tact us:
Phone:
Our toll-free customer service number is available to our customers in the US and Canada, Monday through Friday, between 8 a.m. and 5 p.m. Pacific time at 1-800-755-2751. For our customers outside of the US and Canada, our regular telephone number is (509) 332-2756.
Fax:
Our fax number is (509) 332-5158. When you fax us, please include your WP4’s serial number, your name, address, phone and fax number along with a description of your problem.
E-mail:
If you need technical support, you can send us mes­sages via e-mail at support@decagon.com. Again, please include as part of your message your WP4’s serial number, your name, address, phone, fax num­ber, and return e-mail address.
If you have a question about your application with WP4, please send your message with the above information to wp4@decagon.com.
Warranty
WP4 has a 30-day satisfaction guarantee and a one-
WP
4
PotentiaMeter
Introduction
3
year warranty on parts and labor. To validate your warranty, simply contact us with the information on the warranty card included with this manual. You can return your warranty information by fax, e-mail, phone or by mailing the postage-paid card. Please include all the information requested on the card. It is necessary for Decagon to have your current mail­ing address and telephone number in case we need to send updated product information to you.
Note to our WP4 Users
This manual is written to aid the end user in under­standing the basic concepts of water potential, enabling them to use our instrument with confi­dence. Every effort has been made to ensure that the content of this manual is correct and scientifi­cally sound.
Seller’s Liability
Seller warrants new equipment of its own manufac­ture against defective workmanship and materials for a period of one year from date of receipt of equipment (the results of ordinary wear and tear, neglect, misuse, accident and excessive deteriora­tion due to corrosion from any cause are not to be considered a defect); but Seller’s liability for defec­tive parts shall in no event exceed the furnishing of replacement parts F.O.B. the factory where origi­nally manufactured. Material and equipment cov-
WP
4
PotentiaMeter
Introduction
4
ered hereby which is not manufactured by Seller shall be covered only by the warranty of its manu­facturer. Seller shall not be liable to Buyer for loss, damage or injuries to persons (including death), or to property or things of whatsoever kind (including, but not without limitation, loss of anticipated prof­its), occasioned by or arising out of the installation, operation, use, misuse, nonuse, repair, or replace­ment of said material and equipment, or out of the use of any method or process for which the same may be employed. The use of this equipment con­stitutes Buyer’s acceptance of the terms set forth in this warranty. There are no understandings, repre­sentations, or warranties of any kind, express, implied, statutory or otherwise (including, but with­out limitation, the implied warranties of merchant­ability and fitness for a particular purpose), not expressly set forth herein.
WP
4
PotentiaMeter
About the WP4
5
2. About the WP4
WP4 is the fastest instrument for measuring water potential, giving readings directly in MegaPascals within five minutes. It measures water potential from 0 to -60 MPa with an accuracy of ±0.1 MPa from 0 to -10MPa and ±1% from -10 to -60 MPa. The instrument is easy to use and maintain, with simple checking of calibration.
WP4 Models and Options
The WP4 comes in two models to suit the needs of different users. Here is a brief description of each:
WP4: Our standard model, adequate for most water potential needs.
WP4-T: User-selectable internal temperature control model, uses thermoelectric (Peltier) components to maintain a constant internal temperature.
WP4 and Water Potential
Water potential is a measurement of the energy sta­tus of the water in a system. It indicates how tightly
WP
4
PotentiaMeter
About the WP4
6
water is bound, structurally or chemically, within a substance. Water potential is the vapor pressure of air in equilibrium with a sample in a sealed mea­surement chamber. For a more detailed description of water potential, please refer to Chapter 9, titled “Theory: Water Potential” of this manual.
How the WP4 works
WP4 uses the chilled-mirror dewpoint technique to measure the water potential of a sample. In this type of instrument, the sample is equilibrated with the headspace of a sealed chamber that contains a mirror and a means of detecting condensation on the mirror. At equilibrium, the water potential of the air in the chamber is the same as the water potential of the sample. In the WP4, the mirror temperature is precisely controlled by a thermoelectric (Peltier) cooler. Detection of the exact point at which con­densation first appears on the mirror is observed with a photoelectric cell. A beam of light is directed onto the mirror and reflected into a photodetector cell. The photodetector senses the change in reflec­tance when condensation occurs on the mirror. A thermocouple attached to the mirror then records the temperature at which condensation occurs. WP4 then signals you by flashing a green LED and/or beeping. The final water potential and temperature of the sample is then displayed.
WP
4
PotentiaMeter
About the WP4
7
In addition to the technique described above, WP4 uses an internal fan that circulates the air within the sample chamber to reduce time to equilibrium. Since both dewpoint and sample surface tempera­tures are simultaneously measured, the need for complete thermal equilibrium is eliminated, which reduces measurement times to less than five min­utes.
The WP4-T gives you the option of controlling the sample temperature, by means of an internal thermo-electrically controlled module that monitors and stabilizes the sample block temperature accord­ing to how it is programmed. This can also help in making more rapid measurements.
WP4 and Temperature
Temperature gradients can often be of concern for making stable water potential measurements. Since water potential is influenced by temperature, best results are obtained by controlling the temperature. Samples that are not at room temperature during the read cycle will need to equilibrate to the tempera­ture of the WP4 before accurate measurements of water potential can be made. Large temperature dif­ferences will cause longer reading times, since a complete and accurate reading will not be made until the sample and the instrument are within 0.1 degree of each other. To help you monitor the tem-
WP
4
PotentiaMeter
About the WP4
8
perature difference between your sample and the instrument, you can access a sample equilibration screen at the main menu.
Though temperature control is usually unnecessary for most applications, there are some instances where it is desired. For this reason, Decagon offers a temperature-controlled WP4 model: the WP4-T. There are several advantages in having a tempera­ture-controlled model. Here are a few major rea­sons:
1. Research purposes. To study the effects of
temperature on the water potential of a sample, comparison of the water potential of different samples independent of temperature, or other water potential studies where temperature con­trol is critical.
2. To minimize extreme ambient temperature fluctuations. If the laboratory and WP4 temper­atures fluctuate by as much as ± 5°C daily, water potential readings can vary as much as ± 0.5 MPa on a dry soil sample.
WP4-T
The WP4-T has thermoelectric components installed that allow the instrument to maintain a set chamber temperature without the need of an external water bath. The WP4-T has an extra option in the System
WP
4
PotentiaMeter
About the WP4
9
configuration menu (see page 21) that allows the user to set the temperature.
WP
4
PotentiaMeter
Getting Started
10
3. Getting Started
Components of your WP4
Your WP4 should have been shipped with the fol­lowing items:
WP4 main unit
Quickstart guide
Certificate of Calibration
Power cord
RS-232 interface cable
100 disposable sample cups
Operator’s Manual
12 vials of 0.5 Molal KCl
Cleaning Kit
Choosing a Location
To ensure that your WP4 operates correctly and
WP
4
PotentiaMeter
Getting Started
11
consistently, place it on a level surface. This reduces the chance that sample material will spill and con­taminate the inside of the instrument. To protect the internal electrical components, and to avoid inaccu­rate readings, place your WP4 in a location where the temperature remains fairly stable. This location should be well away from air conditioner and heater vents, open windows, outside doors, refriger­ator exhausts, or other items that may cause rapid temperature fluctuation.
Features
Front view of WP4
LED indicator
light
LCD
Function Keys
Sample drawer
WP
4
PotentiaMeter
Getting Started
12
Back view of WP4
Preparing the WP4 for Operation
After finding a good location for your WP4, plug the power cord to the back of the unit. Before turning it on, pull open the sample drawer (turn the knob to the “OPEN/LOAD” position). An empty disposable sample cup is usually placed upside-down in the drawer to protect it during shipment. Remove this sample cup and turn the instrument on. The ON/ OFF switch is located on the lower right corner of the WP4’s back panel. The following screens will appear on the LCD:
RS-232 interface
fuse well
ON/OFF Switch
Power cord
Plug
Fan
Lid thumb-screw
Model WP4
PotentiaMeter
v3.0
WP
4
PotentiaMeter
Getting Started
13
Then:
This is the main menu, displaying the water poten­tial in both MegaPascals (MPa) and pF, and the sam­ple temperature in °C.
In order to provide the most accurate readings, WP4 should ideally be allowed a warm-up period of 30 minutes after turning it on. When you insert a sam­ple into the chamber drawer and turn the drawer knob to the “READ” position, the instrument will begin the read cycle to measure the water potential of your sample.
Portability
On occasion you may want to take water potential measurements in the field where it is not feasible to take samples and return to the lab. The following is a procedure for powering your WP4 using your vehicle as a power source at sites where AC power is not readily available.
1. Purchase a portable power inverter that plugs
into the 12V output (cigarette lighter) of your vehicle. We strongly recommend that this
0.0°C
0.00 MPa
pF 0.00
WP
4
PotentiaMeter
Getting Started
14
inverter have a continuous output of at least 140 Watts.
2. Place the WP4 on a level surface. Care should be taken to minimize temperature gradients that may affect the instrument while in the field. A Styrofoam box, for example, will help minimize temperature effects.
3. Plug the 12-volt inverter to the 12-volt output of the vehicle, or directly to the battery itself.
4. Plug the WP4 to the inverter, and turn it on. When the instrument is on, it draws 1 amp. Check the rating of your battery if you want to know how long it will power the instrument (for example, if your battery is rated for 60 amp hours, it will work for 60 hours when the vehicle isn’t running.
5. Allow the instrument to warm up for 30 minutes as you would in the lab. Check the calibration of the instrument before proceeding with sampling.
WP
4
PotentiaMeter
Menus
15
4. The Menus
The Main Menu
Every time you turn on your WP4, it will come to this screen. If this screen doesn’t appear, refer to Chapter 12 for troubleshooting instructions. As men­tioned earlier, the water potential and sample tem­perature are displayed on the screen. On each side of the LCD there are buttons. Each button performs a different function depending on which mode you want. Following is a description of the modes and options you may use, and the buttons used to set them.
0.0°C
0.00 MPa
pF 0.00
sample equilibration screen system setup
continuous/normal
language
selection
mode
0.0°C
0.00 MPa
pF 0.00
WP
4
PotentiaMeter
Menus
16
Changing Languages
The WP-4 comes to you with English as the default on-screen user language. If you prefer not to use English, you can change it to one of a variety of other languages: German, French, Spanish, Italian, Swedish, Danish, Norwegian, Czech, Portuguese, or Japanese. This is done simply by pressing the upper right button of the instrument while it is not reading a sample. You will see the following screen:
Press the upper right key again, and the next lan­guage option (German) will appear:
Each time you press the right button, the display will scroll to the next language option. Select the desired language, then press the lower left button to exit.
Normal sampling mode
The first time you turn on the WP4, it will be in nor­mal sampling mode. In this mode, a single measure­ment is made each time you read a sample. Here is
English
-exit-
Deutsch
-zurück-
WP
4
PotentiaMeter
Menus
17
a brief summary of the normal sampling mode fea­tures:
single sampling (not continuous).
instrument may or may not beep after mea-
surement is made, depending on the system setting.
green LED blinks until you turn the drawer
knob to the OPEN/LOAD position.
diagram of button options from main menu
Continuous mode
Continuous mode measures the water potential of your sample continuously until you turn the drawer knob to the OPEN/LOAD position. This can be use­ful in doing long term monitoring of samples that take an especially long time to come to vapor equi­librium, such as plants and leaves. In this mode, it
sample temperature menu
system setup
continuous/normal mode
0.0°C
0.00 MPa
pF 0.00
WP
4
PotentiaMeter
Menus
18
will measure the sample, stop to display the water potential and sample temperature, then begin another read cycle. Between samples, it will signal you with the green LED flash, accompanied by the beeper, if it is enabled. Some find it helpful to con­nect their WP4 to a computer while in continuous mode in order to log and store data over time. For instructions on how to do this, see Chapter 8.
To toggle between the normal and continuous modes, press the top left button. The display will show a small “c” to the left of the water potential readings:
main menu with continuous mode enabled
If you press the upper left button again, the “c” will disappear and you will be back in normal sampling mode.
System configuration
If you press the bottom left button while in the main menu, it will bring you to the system configu-
c
“c” for continuous mode
0.0°C
0.00 MPa
pF 0.00
WP
4
PotentiaMeter
Menus
19
ration menu. This menu allows you to make minor
system changes. You can change how the beeper signals after each sample and enter the calibration menu as well.
System configuration menu
Changing the beeper
When you are sampling, the WP4 has two ways of notifying you that the water potential reading is completed for your sample: the beeper and a flash­ing green LED, located on the left front corner of the WP4’s case. In normal sampling mode, the LED will flash once when a sample is started. When it is
system configuration menu button
0.0°C
0.00 MPa
pF 0.00
-Exit-
+
-
set T
Beeper
Beeper Mode indicator
Calibration
menu
Icon
(continuous shown)
temp. set (WP4-T model only)
Loading...
+ 58 hidden pages