6. TROUBLESHOOTING
Error Indicator
Error IndicatorError Indicator
Error Indicator Cause
CauseCause
Cause Correction
CorrectionCorrection
Correction
WEIGHT ERROR
CODE 01
Error weight reading - Failure in load cell
- Failure in A/D
- Failure in load cell wires
OVERFLOW
CODE 02
Weight over 225 kg - Failure in load cell
- Failure in A/D
- Failure in load cell wires
UNDERFLOW
CODE 03
Weight under 0 kg - Failure in load cell
- Failure in A/D
- Failure in load cell wires
- Failure in load cell mounting
ZERO WEIGHT ERROR
CODE 04
Error in Zero reading at
machine start.
- Failure in load cell
- Failure in A/D
- Failure in load cell wires
- Failure in load cell mounting
- Knocks in load cell
HEIGHT ERROR
CODE 05
Height reading error. - Failure in height cell.
- Failure in height circuit.
- Failure in height wires.
- Failure in power supply +12V
PNEUMATIC ERROR
CODE 10
Error on the E2Prom
memory.
- Check if the
Pressure
electronic board is
well inserted.
- Replace the
Pressure
electronic board.
PNEUMATIC ERROR
CODE 11
Cuff over-inflated. More than
300 mmHg.
- Failure in air pressure.
- Failure in pressure measuring.
PNEUMATIC ERROR
CODE 12
The inflating process time
has exceeded.
- Failure in air pressure circuit.
- Failure in pressure measuring.
PNEUMATIC ERROR
CODE 13
Problem caused by abnormal
function during the detection
process.
- Failure in air pressure circuit.
- Failure in pressure measuring.
PNEUMATIC ERROR
CODE 14
Problem caused by abnormal
memory function.
- Failure en la E2PROM de la PressV3.
- Failure in I2C circuit.
PNEUMATIC ERROR
CODE 15
The process has been
cancelled by STOP button.
- User push STOP keypad to abort
pressure measurement.
PNEUMATIC ERROR
CODE 16
The measuring process time
has exceeded.
- Failure in air pressure circuit.
- Failure in pressure measuring.
PNEUMATIC ERROR
CODE 17
No weight detection on the
platform.
- To measure the blood pressure you
must stand on the platform.
PNEUMATIC ERROR
CODE 18
Blood pressure measuring
error.
- Data type very difficult to analyze : low
signal, abnormal pressure with
arrythmics, etc
BODYFAT ERROR
CODE 20
Reading error in Bodyfat - Comms error in I2C
- Reset failure in Bodyfat PCB
- Power supply in +5V Bodyfat
E2PROM BASE
CODE 30
Error data in E2PROM BASE - Error in E2PROM BASE
E2PROM PRESSV3
CODE 31
Error data in E2PROM
PRESSV3
- Error in E2PROM PRESSV3
E2PROM BASE & PRESSV3
CODE 32
Error data in E2PROM BASE
and PRESSV3
- Error in E2PROM BASE Y PRESSV3
ERROR PRINTER
CODE 40
Failure in PRN - Comms error in PRN
- Power failure in PRN