DATREND Systems vPad-XSC User Manual

vPad-XSC
Automation Interface for
Impulse 4000 Defibrillator Analyzer
User Manual
vPad-XSC
Automation Interface
User Manual
© 2019 Datrend Systems Inc.
130 - 4020 Viking Way
Richmond, BC • CANADA • V6V 2L4
Tel: 800.667.6557 (North America Only) or
604.291.7747 • Fax 604.294.2355
e-mail: customerservice@datrend.com
To order this manual, use Part Number 6100-024
Revision
A Initial Release 2019-Jan-24
Revision History
Description Date
Copyright
Datrend Systems Inc. (“DSI”) agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Datrend Systems Inc.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the freight carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.
Claims
Our routine method of shipment is via common carrier. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact your local sales representative or DSI immediately.
Standard Terms and Conditions
Refunds & Credits
Please note only serialized products (products labelled with a distinct serial number) and accessories are eligible for partial refund and/or credit. Non-serialized parts and accessory items (cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. In order to receive a partial refund/credit, the product must not have been damaged, and must be returned complete (meaning all manuals, cables, accessories, etc.) within 90 days of original
purchase and in “as new” and resalable condition. The Return Procedure must be followed.
Return Procedure
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number,
obtained from Datrend Customer Service. All items being returned must be sent prepaid (freight, duty, brokerage,
and taxes ) to our factory location.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15%. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20%. Additional charges for damage and/or missing parts and accessories will be applied to all returns. Products which are not in “as new” and resalable condition, are not eligible for credit return and will be returned to the customer at their expense.
Certification
This instrument was thoroughly tested and inspected and found to meet DSI’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Research Council of Canada (NRC) and/or the National Institute of Standards and Technology (NIST). Devices for which there are no NRC/NIST calibration standards are measured against in-house performance standards using accepted test procedures.
Page i
Warranty
Warranty and Product Support
Datrend Systems Inc. ("DSI") warrants this instrument to be free from defects in materials and workmanship under normal use and service for 90 days from the date of original purchase. During the warranty period DSI will, at our option, either repair or replace defects in materials and workmanship at no charge; provided the product is returned (shipping, duty, brokerage and taxes prepaid) to DSI. Any and all transportation charges incurred are the responsibility of the purchaser and are not included within this warranty. This warranty extends only to the original purchaser and does not cover damage from abuse, neglect, accident or misuse or as the result of service or modification by other than DSI. IN NO EVENT SHALL DATREND SYSTEMS INC. BE LIABLE FOR CONSEQUENTIAL DAMAGES.
No warranty shall apply when damage is caused by any of the following:
! Power failure, surges, or spikes, ! Damage in transit or when moving the instrument, ! Improper power supply such as low voltage, incorrect voltage, defective wiring or inadequate fuses, ! Accident, alteration, abuse or misuse of the instrument, ! Fire, water damage, theft, war, riot, hostility, acts of God, such as hurricanes, floods, etc.
Only serialized products (those items bearing a distinct DSI serial number tag) are covered under this warranty. PHYSICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED UNDER THE WARRANTY. Items such as cables and non-serialized modules are not covered under this warranty.
This warranty gives you specific legal rights and you may have other rights, which vary from province to province, state to state, or country to country. This warranty is limited to repairing the instrument to DSI's specifications.
When you return an instrument to DSI for service, repair or calibration, we recommend shipment using the original shipping foam and container. If the original packing materials are not available, we recommend the following guide for repackaging:
! Use a double-walled carton of sufficient strength for the weight being shipped. ! Use heavy paper or cardboard to protect all instrument surfaces. Use non-abrasive material around all projecting
parts.
! Use at least four inches of tightly packed, industrial-approved, shock-absorbent material all around the instrument.
DSI will not be responsible for lost shipments or instruments received in damaged condition due to improper packaging or handling. All warranty claim shipments must be made on a prepaid basis (freight, duty, brokerage, and taxes). No returns will be accepted without a Return Materials Authorization ("RMA”) number. Please contact Datrend (refer to Chapter 5 of this manual) to obtain an RMA number and receive help with shipping/customs documentation.
Page ii
Warranty Disclaimer
Should you elect to have your instrument serviced and/or calibrated by someone other than Datrend Systems, please be advised that the original warranty covering your product becomes void when the tamper-resistant Quality Seal is removed or broken without proper factory authorization. We strongly recommend, therefore, that you send your instrument to Datrend Systems for service and calibration, especially during the original warranty period. In all cases, breaking the tamper-resistant Quality Seal should be avoided at all cost, as this seal is the key to your original instrument warranty. In the event that the seal must be broken to gain internal access to the instrument (e.g., in the case of a customer-installed firmware upgrade), you must first contact Datrend Systems at 1-800-667-6557. You will be required to provide us with the serial number for your instrument as well as a valid reason for breaking the Quality Seal. You should break this seal only after you have received factory authorization. Do not break the Quality Seal before you have contacted us! Following these steps will help ensure that you will retain the original warranty on your instrument without interruption.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Datrend Systems will not be responsible for any injuries sustained due to unauthorized equipment modifications.
DSI DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION.
THIS PRODUCT CONTAINS NO USER-SERVICEABLE COMPONENTS. UNAUTHORIZED REMOVAL OF THE INSTRUMENT COVER SHALL VOID THIS AND ALL OTHER EXPRESSED OR IMPLIED WARRANTIES.
vPad™, vPad-XSC™, and vPad-Check™ are trademarks of Datrend Systems Inc.
Android
is trademark of Google Inc.
Page iii
Page iv
vPad-XSC for Impulse 4000 USER MANUAL
Table of Contents
TABLE OF CONTENTS .............................................. V
1OVERVIEW .................................................... 1
1.1 Standard Accessories ............................................. 1
1.2 Optional Accessories ............................................. 1
1.3 Checklists and vPL ............................................... 2
2CONNECTION AND SETUP ......................................... 3
3CHECKLIST PROGRAMMING ........................................ 5
3.1 Initialization .................................................... 5
3.2 Configuration ................................................... 6
3.2.1 Setting Limits and Tolerances .................................. 6
3.2.1.1 Example 1: Setting a Defibrillator Energy Limit ................. 8
3.2.1.2 Example 2: Setting Defibrillator Charge Time Limit .............. 8
3.2.1.3 Example 3: Setting Pacemaker Pulse Current Limit ............. 9
3.2.1.4 Example 4: Disabling a Limit ............................... 9
3.2.2 Resetting Limits to Defaults ................................... 10
3.2.3 ECG Simulation ............................................ 11
3.2.3.1 Example 1: Configuring ECG Simulation ..................... 11
3.3 Defibrillator Tests ............................................... 13
3.3.1 Defibrillator Energy Test ...................................... 14
3.3.1.1 Example 8: Measuring Defibrillator Energy ................... 14
3.3.2 Defibrillator Charge Time Test ................................. 16
3.3.2.1 Example 9: Measuring Defibrillator Charge Time .............. 16
3.3.3 Defibrillator Sync Delay (Cardioversion) Test ..................... 17
3.3.3.1 Example 10: Measuring Defibrillator Sync Delay .............. 17
3.4 Pacemaker Tests ............................................... 19
3.4.1 Pacemaker Pulse Test ....................................... 19
3.4.1.1 Example 11: Measuring Pacemaker Pulse ...................19
3.4.2 Pacemaker Refractory Period Test ............................. 20
3.4.2.1 Example 12: Measuring Pacemaker Refractory Period .......... 20
3.5 Direct Commands ............................................... 22
4APPENDIX A - VPL PROCEDURES .................................. 23
5CALIBRATION AND MAINTENANCE ................................... 27
Table of Contents # Page v
vPad-XSC for Impulse 4000 USER MANUAL
1 Overview
vPad-XSC™ for Impulse 4000 (referred to as "the Interface" in this manual) is an intelligent hardware interface which supports vPad electrical safety analyzers having the vPad-Check™ software application (App) installed. The Interface allows vPad-Check to conduct automated maintenance and inspection sequences on defibrillators using the Impulse 4000 analyzer.
In addition to automated control of Impulse 4000 operation, the vPad-XSC provides automatic assessment of measurement results based on user-defined limits
This manual provides guidance for users integrating vPad-Check™ and vPad-XSC™ for Impulse 4000 into new and existing maintenance processes.
1.1 Standard Accessories
- XSC Interface unit (P/N 7400-109 or 7400-110)
- RJ-12 cable (P/N 3140-069) for connection to XBUS of vPad safety analyzer
- User Manual (P/N 6100-024)
1.2 Optional Accessories
- DB9F to DB25F null modem cable (P/N 3140-442) for connection to Impulse 4000 (Assman AK124-3-R)
For a complete list of available accessories, visit www.datrend.com or contact Datrend Customer Service (see Chapter 5 for contact details)
Overview/Chapter 1 # Page 1
vPad-XSC for Impulse 4000 USER MANUAL
1.3 Checklists and vPL
In vPad-Check, equipment maintenance and inspection procedures are called
Checklists
. A Checklist can contain a sequence of prompts for the operator as well as machine instructions to configure the vPad safety analyzer and to conduct other tests through a remote instrument. In this case, the remote instrument is an Impulse 4000 defibrillator and transcutaneous pacemaker analyzer.
In a Checklist, prompts and machine instructions are written as statements in
Language
(vPL). A statement generally consists of a specific keyword followed by one or more
vPad Programming
parameters. For example, the statement
xopen "Impulse 4000 XSC" | D | 5
initializes a data connection between the Interface and Impulse 4000 using the keyword “xopen”. Other statements are responsible for configuring and initiating tests through Impulse 4000 .
Checklists can be written on the Android tablet, or on a personal computer (PC), using a text editor, or using the Datrend vPad-IDE. Once a new Checklist has been written, it can be imported to vPad-Check through a USB flash drive, via Bluetooth or a cloud service (Dropbox, Google Drive or OneDrive). Further information on Checklists and vPL, including a full list of vPL keywords, is provided in the vPad-
Check Operating Manual which can be opened with the Datrend Docs app: .
Overview/Chapter 1 # Page 2
vPad-XSC for Impulse 4000 USER MANUAL
2 Connection and Setup
Verify DIP switches on the Interface unit are set appropriately for your application: for a wired setup, DIP switch 3 (BTENA) is set to the
For a wired setup, connect the DB9 serial port of the Interface to the DB9 serial port of Impulse 4000 using the null modem serial cable (P/N 3140-400) provided with the Interface.
Connect the XBUS port on the Interface to the XBUS port on the vPad safety analyzer using the RJ12 cable (P/N 3140-069) provided with the Interface. Note that the two XBUS ports on the Interface are interchangeable. This enables several vPad-XSC interfaces of different types to be chained in series if necessary.
OFF
position, and the remaining two are set to
ON
.
Once connected to the vPad safety analyzer, the “Power” indicator on the Interface should light up to show that it is now ready for operation.
Power up Impulse 4000 .
The system is now ready for operation.
Installation/Chapter 2 # Page 3