User’s Manual for
Standard TEMPpoint, VOLTpoint,
and MEASURpoint USB
Instruments
DT9871, DT9871U, DT9872, DT9873, DT9874
Page 2
Copyright Page
Notice
Measurement Computing Corporation does not authorize any Measurement Computing
Corporation product for use in life support systems and/or devices without prior written consent
from Measurement Computing Corporation. Life support devices/systems are devices or systems
that, a) are intended for surgical implantation into the body, or b) support or sustain life and whose
failure to perform can be reasonably expected to result in injury. Measurement Computing
Corporation products are not designed with the components required, and are not subject to the
testing required to ensure a level of reliability suitable for the treatment and diagnosis of people.
Trademark and Copyright Information
Measurement Computing Corporation, InstaCal, Universal Library, and the Measurement
Computing logo are either trademarks or registered trademarks of Measurement Computing
Corporation. Refer to the Copyrights & Trademarks section on mccdaq.com/legal for more
information about Measurement Computing trademarks.
Other product and company names mentioned herein are trademarks or trade names of their
respective companies.
This equipment has been tested and found to comply with CISPR EN55011 Class A and
EN61326-1 requirements and also with the limits for a Class A digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at his own
expense.
Changes or modifications to this equipment not expressly approved by Data Translation could
void your authority to operate the equipment under Part 15 of the FCC Rules.
Note: This product was verified to meet FCC requirements under test conditions that
included use of shielded cables and connectors between system components. It is important
that you use shielded cables and connectors to reduce the possibility of causing interference
to radio, television, and other electronic devices.
FCC
Page
Canadian Department of Communications Statement
This digital apparatus does not exceed the Class A limits for radio noise emissions from
digital apparatus set out in the Radio Interference Regulations of the Canadian Department of
Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites
applicables aux appareils numériques de la class A prescrites dans le Règlement sur le
brouillage radioélectrique édicté par le Ministère des Communications du Canada.
TEMPpoint™ is a family of temperature measurement instruments that includes the DT9871,
DT9871U, DT8871, DT8871U, DT9872, and DT8872. This manual describes the DT9871,
DT9871U, and DT9872 USB models.
VOLTpoint™ is a family of voltage measurement instruments that includes the DT9873 and
DT8873. This manual describes the DT9873 USB model.
MEASURpoint™ a family of mixed temperature and voltage measurement instruments that
includes the DT9874 andDT8874. This manual describes the DT9874 USB model.
Note: For information on the DT8871U, DT8871, DT8872, DT8873, and DT8874 LXI models
of TEMPpoint, VOLTpoint, and MEASURpoint, refer to the User’s Manual for Standard TEMPpoint, VOLTpoint, and MEASURpoint LXI Instruments.
The first part of this manual describes how to install and set up your instrument, and verify
that the instrument is working properly.
The second part of this manual describes the features and capabilities of your instrument
using the IVI-COM instrument driver software. Troubleshooting information is also provided.
Note: If you are programming the instrument using the IVI-COM driver, refer to the
DtxMeasurement IVI-COM driver online help for more information.
Intended Audience
This document is intended for engineers, scientists, technicians, or others responsible for
using and/or programming a TEMPpoint, VOLTpoint, or MEASURpoint instrument in the
Microsoft® Windows® XP, Windows Vista®, Windows 7, or Windows 8 operating system. It is
assumed that you have some familiarity with thermocouples, RTDs, and/or voltages and that
you understand your application.
9
Page 10
About this Manual
How this Manual is Organized
This manual is organized as follows:
• Chapter 1, “Overview,” summarizes the major features of the TEMPpoint, VOLTpoint,
and MEASURpoint instruments, as well as the supported software and accessories.
• Chapter 2, “Preparing to Use the Instrument,” describes how to unpack the instrument,
check the system requirements, install the software, and view the documentation online.
• Chapter 3, “Setting Up and Installing the Instrument,” describes how to apply power to
the instrument and connect the instrument to your computer.
• Chapter 4, “Wiring Signals,” describes how to wire signals to the instrument.
• Chapter 5, “Verifying the Operation of Your Instrument,” describes how to verify the
operation of the instrument.
• Chapter 6, “Principles of Operation,” describes the analog input and digital I/O features
of the TEMPpoint, VOLTpoint, and MEASURpoint instruments in detail.
• Chapter 7, “Troubleshooting,” provides information that you can use to resolve problems
with your instrument, should they occur.
• Appendix A, “Specifications,” lists the specifications of the TEMPpoint, VOLTpoint, and
MEASURpoint instruments.
• Appendix B, “Connector Pin Assignments,” describes the pin assignments of the digital
I/O connector on the TEMPpoint, VOLTpoint, and MEASURpoint instruments.
• An index completes this manual.
Conventions Used in this Manual
The following conventions are used in this manual:
• Notes provide useful information or information that requires special emphasis, cautions
provide information to help you avoid losing data or damaging your equipment, and
warnings provide information to help you avoid catastrophic damage to yourself or your
equipment.
• Items that you select or type are shown in bold.
• CAUTION – This icon denotes a caution, which advises you to consult the
documentation where this symbol is marked.
CAUTION – Do not operate this product in a manner not specified in this document.
Product misuse can result in a hazard. You can compromise the safety protection built into
the product if the product is damaged in any way.
10
Page 11
Related Information
Refer to the following documents for more information on using a TEMPpoint, VOLTpoint, or
MEASURpoint USB instrument:
• DtxMeasurement IVI-COM Driver online help. This document describes how to use the
IVI-COM driver to access the capabilities of a TEMPpoint, VOLTpoint, or MEASURpoint
instrument.
The IVI-COM driver works with any development environment that supports COM
programming, including MATLAB® from The MathWorks
or Visual Basic®.NET, Agilent® VEE Pro, National Instruments™ LabVIEW™ or
LabWindows™, and so on.
• QuickDAQ User’s Manual (UM-25242). This manual describes how to create a QuickDAQ
application to acquire and analyze data from a TEMPpoint, VOLTpoint, or MEASURpoint
instrument.
• IVI foundation (www.ivifoundation.org)
• Omega Complete Temperature Measurement Handbook and Encyclopedia® or the Omega
Engineering web site: http://www.omega.com. Both resources provide valuable
information on thermocouple types, RTD types, standards, and linearization.
About this Manual
TM
, Microsoft® Visual C#®.NET
Where To Get Help
Should you run into problems installing or using a TEMPpoint, VOLTpoint, or MEASURpoint
instrument, the Data Translation Technical Support Department is available to provide
technical assistance. Refer to Chapter 7 for more information. If you are outside the United
States or Canada, call your local distributor, whose number is listed on our web site
(www.mccdaq.com).
Data Translation provides a number of USB instruments to meet your measurement needs,
including the following:
• TEMPpoint – a family of temperature measurement instruments
• VOLTpoint – a family of voltage measurement instruments
• MEASURpoint – a family of mixed temperature and voltage measurement instruments
All of these instruments support Version 2.0 and 1.1 of the USB bus.
The following sections summarize the features of the TEMPpoint, VOLTpoint, and
MEASURpoint USB instruments.
TEMPpoint Features
TEMPpoint instruments include the following models: DT9871U, DT9871, and DT9872.
Figure 1 shows a DT9871U instrument.
14
Figure 1: TEMPpoint Instrument
The key features of TEMPpoint instruments are as follows:
• DT9871U and DT9871:
Configurable analog input channels for thermocouple or differential voltage inputs;
easy-access jacks for each channel for quick wiring
One CJC (cold junction compensation) input for each thermocouple channel
Page 15
B, E, J, K, N, R, S, and T thermocouple types supported; the instrument automatically
linearizes the measurements and returns the data as a 32-bit, floating-point
temperature values
Input range of ±0.075 V for the DT9871U (with 0.25 V RMS A/D noise using no
software filtering) and ±1.25 V for the DT9871 (with 5 V RMS A/D noise using no
software filtering)
Break-detection circuitry to detect open thermocouple inputs
• DT9872:
Configurable analog input channels for RTDs and differential voltage inputs;
easy-access jacks for each channel for quick wiring
100 , 500 , and 1000 platinum RTD types supported using alpha curves of 0.00385
(European) or 0.00392 (American)
4-wire, 3-wire, or 2-wire configurations; the DT9872 automatically linearizes the
measurements and returns the data as 32-bit, floating-point temperature, resistance, or
voltage values
Input range of ±1.25 V
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution
measurements
Overview
• Throughput rate of up to 10 Samples/s for all channels.
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the
instrument supports anytime calibration, performing an auto-calibration function on
software command
• Measurement Calibration Utility allows you to calibrate the instrument in the field (see
page 20 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog
input data stream for correlating analog and digital measurements
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from
±30 V at currents up to 400 mA (peak) AC or DC
15
Page 16
Chapter 1
VOLTpoint Features
Figure 2 shows a VOLTpoint instrument.
Figure 2: VOLTpoint Instrument
The key features of VOLTpoint (DT9873) instruments are as follows:
• Direct connection of analog input channels for differential voltage inputs; removable
screw terminal blocks for each channel for quick wiring
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution
measurements
• Software-selectable input range of ±10 V or ±60 V per channel (note that for the ±60 V
range, no more than 30 Vrms, 42.4 Vpk, 60 VDC is allowed)
• Throughput rate of up to 10 Samples/s for all channels
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the
instrument supports anytime calibration, performing an auto-calibration function on
software command
• Measurement Calibration Utility allows you to calibrate the instrument in the field (see
page 20 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog
input data stream for correlating analog and digital measurements
16
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from
±30 V at currents up to 400 mA (peak) AC or DC
Page 17
MEASURpoint Features
The standard MEASURpoint (DT9874) instrument provides 16 thermocouple channels, 16
RTD channels, and 16 voltage channels. Figure 3 shows a MEASURpoint instrument.
Overview
Figure 3: MEASURpoint Instrument
The key features of MEASURpoint instruments are as follows:
• Analog Input Channels 0 to 15:
Configurable channels for thermocouple or differential voltage inputs; easy-access
jacks for each channel for quick wiring
One CJC (cold junction compensation) input for each thermocouple channel
B, E, J, K, N, R, S, and T thermocouple types supported; the instrument automatically
linearizes the measurements and returns the data as a 32-bit, floating-point
temperature values
Input range of ±0.075 V (with 0.25 V RMS A/D noise using no software filtering)
Break-detection circuitry to detect open thermocouple inputs
• Analog Input Channels 16 to 31:
Configurable analog input channels for RTDs and differential voltage inputs;
easy-access jacks for each channel for quick wiring
100 , 500 , and 1000 platinum RTD types supported using alpha curves of 0.00385
(European) or 0.00392 (American)
17
Page 18
Chapter 1
4-wire, 3-wire, or 2-wire configurations; the DT9872 automatically linearizes the
measurements and returns the data as 32-bit, floating-point temperature, resistance, or
voltage values
Input range of ±1.25 V
• Analog Input Channels 31 to 48:
Direct connection of analog input channels for differential voltage inputs; removable
screw terminal blocks for each channel for quick wiring
Software-selectable input range of ±10 V or ±60 V (note that for the ±60 V range, no
more than 30 Vrms, 42.4 Vpk, 60 VDC is allowed)
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution
measurements
• 30 VAC, 60 VDC continuous functional isolation ch-ch and ch-gnd, verified by a 500 Vpk
withstand
• Throughput rate of up to 10 Samples/s for all channels
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the
instrument supports anytime calibration, performing an auto-calibration function on
software command
• Measurement Calibration Utility allows you to calibrate the instrument in the field (see
page 20 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog
input data stream for correlating analog and digital measurements
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from
±30 V at currents up to 400 mA (peak) AC or DC
18
Page 19
Supported Software
The following software is available for use with the TEMPpoint, VOLTpoint, and
MEASURpoint USB instruments:
• QuickDAQ Base Version – The base version of QuickDAQ is free-of-charge and allows
you to acquire and analyze data from all Data Translation USB and Ethernet devices,
except the DT9841 Series, DT9817, DT9835, and DT9853/54. Using the base version of
QuickDAQ, you can perform the following functions:
Discover and select your devices.
Configure all input channel settings for the attached sensors.
Load/save multiple hardware configurations.
Generate output stimuli (fixed waveforms, swept sine waves, or noise signals).
On each supported data acquisition device, acquire data from all channels supported
in the input channel list.
Choose to acquire data continuously or for a specified duration.
Choose software or triggered acquisition.
Overview
Log acquired data to disk in an .hpf file.
Display acquired data during acquisition in either a digital display using the Channel
Display window or as a waveform in the Channel Plot window.
Choose linear or logarithmic scaling for the horizontal and vertical axes.
View statistics about the acquired data, including the minimum, maximum, delta, and
mean values and the standard deviation in the Statistics window.
Export time data to a .csv or .txt file; you can open the recorded data in Microsoft
Excel® for further analysis.
Read a previously recorded .hpf data file.
Customize many aspects of the acquisition, display, and recording functions to suit
your needs, including the acquisition duration, sampling frequency, trigger settings,
filter type, and temperature units to use.
• QuickDAQ FFT Analysis Option – When enabled with a purchased license key, the
QuickDAQ FFT Analysis option includes all the features of the QuickDAQ Base version
plus these features:
The ability to switch between the Data Logger time-based interface and the FFT
Analyzer block/average-based interface.
Supports software, freerun, or triggered acquisition with accept and reject controls for
impact testing applications.
Allows you to perform single-channel FFT (Fast Fourier Transform) operations,
including AutoSpectrum, Spectrum, and Power Spectral Density, on the acquired
analog input data. You can configure a number of parameters for the FFT, including
the FFT size, windowing type, averaging type, integration type, and so on.
Allows you to display frequency-domain data as amplitude or phase.
19
Page 20
Chapter 1
Supports dB or linear scaling with RMS (root mean squared), peak, and peak-to-peak
scaling options
Supports linear or exponential averaging with RMS, vector, and peak hold averaging
options.
Supports windowed time channels.
Supports the following response window types: Hanning, Hamming, Bartlett,
Blackman, Blackman Harris, and Flat top.
Supports the ability to lock the waveform output to the analysis frame time.
Allows you to configure and view dynamic performance statistics, including the input
below full-scale (IBF), total harmonic distortion (THD), spurious free dynamic range
(SFDR), signal-to-noise and distortion ratio (SINAD), signal-to-noise ratio (SNR), and
the effective number of bits (ENOB), for selected time-domain channels in the Statistics
window.
Supports digital IIR (infinite impulse response) filters.
• QuickDAQ Advanced FFT Analysis Option – When enabled with a purchased software
license, the QuickDAQ Advanced FFT Analysis option includes all the features of the
QuickDAQ Base version with the FFT Analysis option plus these features:
Allows you to designate a channel as a Reference or Response channel.
Allows you to perform two-channel FFT analysis functions, including Frequency
Response Functions (Inertance, Mobility, Compliance, Apparent Mass, Impedance,
Dynamic Stiffness, or custom FRF) with H1, H2, or H3 estimator types,
Cross-Spectrum, Cross Power Spectral Density, Coherence, and Coherent Output
Power.
Supports the Exponential response window type.
Supports the following reference window types: Hanning, Hamming, Bartlett,
Blackman, Blackman Harris, FlatTop, Exponential, Force, and Cosine Taper windows.
Supports real, imaginary, and Nyquist display functions.
Allows you to save data in the .uff file format.
• DtxMeasurement IVI-COM driver – This driver provides access to the instrument’s
functions through a COM server. The IVI-COM driver works in any 32-bit or 64-bit
development environment that supports COM programming, including MATLAB, Visual
Basic.NET, Visual C#.NET, Agilent VEE Pro, LabVIEW, LabWindows, and others.
• Measurement Calibration Utility – Users can calibrate a TEMPpoint, VOLTpoint, or
MEASURpoint instrument in the field using precise calibration equipment and the
Measurement Calibration Utility. Since each instrument consists of up to 48 individual
channels, great care must be taken to ensure that proper warm-up times are followed and
precise calibration equipment is used.
20
The calibration utility ships with a comprehensive help file that describes the required
equipment and calibration procedure, including warm-up times, for each instrument.
The calibration utility allows you to revert to the factory calibration for any or all
channels, or revert back to the last user calibration values, if desired. In addition, this
utility generates a report that lists the starting and ending calibration values for each
channel, allowing traceability.
Page 21
Accessories
The following optional accessories are available for TEMPpoint, VOLTpoint, or
MEASURpoint instruments:
• STP37 screw terminal panel – The STP37, shown in Figure 4, permits easy screw terminal
connections for accessing the digital I/O signals of a TEMPpoint, VOLTpoint, or
MEASURpoint instrument.
Overview
Figure 4: STP37 Screw Terminal Panel
• EP333 cable–The EP333, shown in Figure 5, is a 2-meter shielded cable with two 37-pin
connectors that connects the STP37 screw terminal panel to the digital I/O connector of
the instrument.
Figure 5: EP333 Cable
• EP373 Single Rack-mountKit – Mounts one TEMPpoint, VOLTpoint, or MEASURpoint
instrument in a rack.
• EP374 Dual Rack-mountKit – Mounts two TEMPpoint, VOLTpoint, or MEASURpoint
instruments side-by-side in a rack.
21
Page 22
Chapter 1
Set Up and Install the Instrument
(see Chapter 3 starting on page 31)
Wire Signals
(see Chapter 4 starting on page 41)
Verify the Operation of the Instrument
(see Chapter 5 starting on page 63)
Prepare to Use the Instrument
(see Chapter 2 starting on page 25)
Getting Started Procedure
The flow diagram shown in Figure 6 illustrates the steps needed to get started using a
TEMPpoint, VOLTpoint, or MEASURpoint instrument. This diagram is repeated in each
Getting Started chapter; the shaded area in the diagram shows you where you are in the
getting started procedure.
Open the shipping box and verify that the following items are present:
• TEMPpoint, VOLTpoint, or MEASURpoint instrument
•USB cable
• EP361 +5V power supply and cable
• For DT9872, DT9874, and DT9874 instruments, a bag of pluggable screw terminable
blocks
If an item is missing or damaged, contact Data Translation. If you are in the United States, call
the Customer Service Department at (508) 946-5100. An application engineer will guide you
through the appropriate steps for replacing missing or damaged items. If you are located
outside the United States, call your local distributor, listed on Data Translation’s web site
(www.mccdaq.com).
Once you have unpacked your instrument, check the system requirements, as described in the
next section.
Preparing to Use the Instrument
27
Page 28
Chapter 2
Checking the System Requirements
For reliable operation, ensure that your computer meets the following system requirements:
• Processor: Pentium 4/M or equivalent
•RAM: 1 GB
• Screen Resolution: 1024 x 768 pixels
• Operating System: Windows 8, Windows 7, Windows Vista (32- and 64-bit)
Windows XP SP3 (32-bit)
• Disk Space: 4 GB
• Acrobat Reader 5.0 or later for viewing documentation
28
Page 29
Installing the Software
Install the software for your instrument from the web at
https://www.mccdaq.com/downloads/DTSoftware/MEASURpoint.
The installation program guides you through the installation process.
Preparing to Use the Instrument
29
Page 30
Chapter 2
Viewing the Documentation
Note: To view the documentation, you must have Adobe Reader 5.0 or greater installed on
your system.
You can access the documentation for your instrument from the Windows Start menu as
follows:
• For documentation about the TEMPpoint, VOLTpoint, or MEASURpoint instrument, click
Programs -> Data Translation, Inc -> Hardware Documentation -> User’s Manual for
Standard TEMPoint, VOLTpoint, and MEASURpoint USB Instruments.
• For documentation on QuickDAQ, click Programs -> Data Translation, Inc ->
QuickDAQ -> QuickDAQ User’s Manual
• For documentation on the DtxMeasurement IVI-COM driver, click Programs -> Data
Translation, Inc -> DtxMeasurement -> DtxMeasurement IVI Driver 1.1.8
Documentation.
The following may be helpful when using Adobe Reader:
• To navigate to a specific section of the document, click a heading from the table of contents
on the left side of the document.
• Within the document, click the text shown in blue to jump to the appropriate reference
(the pointer changes from a hand to an index finger).
• To go back to the page from which the jump was made, click the right mouse button and
Go Back, or from the main menu, click Document, and then Go Back.
• To increase or decrease the size of the displayed document, from the main menu, click
View, and then Zoom.
• By default, Adobe Reader smooths text and monochrome images, sometimes resulting in
blurry images. If you wish, you can turn smoothing off by clicking File, and then
Preferences/General, and unchecking Smooth Text and Images.
Note: Your TEMPpoint, VOLTpoint, and MEASURpoint instruments are factory-calibrated.
Thereafter, yearly recalibration is recommended. Refer to page 94 for more information on
calibration.
32
Page 33
Applying Power
EP361 +5 V
Power Supply
Power
Switch
TEMPpoint, VOLTpoint, and MEASURpoint instruments are shipped with an EP361 +5V
power supply and cable. To apply power to the instrument, do the following:
1. Connect the +5 V power supply to the power connector on the rear panel of the
instrument. Refer to Figure 7.
Setting Up and Installing the Instrument
Figure 7: Attaching a +5 V Power Supply to the Instrument
2. Plug the power supply into a wall outlet.
IMPORTANT: For proper grounding of your measurement instrument, ensure that you
use the power supply and cable (EP361) that is provided with the instrument and that you
use all three prongs of the cable when connecting it to your wall outlet.
3. Press the Power Switch to turn on the instrument.
The Power LED on the front panel lights to indicate that power is on.
33
Page 34
Chapter 3
Powe r L ED
Figure 8 shows the location of the Power LED on the front panel of the instrument; a
MEASURpoint instrument is shown in this example.
Figure 8: Power LED on the Front Panel of the Instrument
34
Page 35
Setting Up and Installing the Instrument
Attaching the Instrument to the Computer
This section describes how to attach a TEMPpoint, VOLTpoint, or MEASURpoint instrument
to the host computer.
Notes: Most computers have several USB ports that allow direct connection to USB devices.
If your application requires more TEMPpoint, VOLTpoint, or MEASURpoint instruments
than you have USB ports for, you can expand the number of USB devices attached to a single
USB port by using expansion hubs. For more information, refer to page 38.
You can unplug a USB instrument, then plug it in again, if you wish, without causing
damage. This process is called hot-swapping. Your application may take a few seconds to
recognize an instrument once it is plugged back in.
You must install the device driver before connecting your instrument to the host computer.
See page 29 for more information.
Connecting Directly to the USB Ports
To connect a TEMPpoint, VOLTpoint, or MEASURpoint instrument directly to a USB port on
your computer, do the following:
1. Make sure that you have attached a power supply to the instrument.
2. Attach one end of the USB cable to the USB port on the rear panel of the instrument, as
shown in Figure 9.
35
Page 36
Chapter 3
Attach USB cable
to USB port of host
computer
USB port of
instrument
Figure 9: Attaching the Instrument to the Host Computer
3. Attach the other end of the USB cable to one of the USB ports on the host computer.
The operating system automatically detects the instrument and starts the Found New Hardware
wizard.
4. For Windows Vista
:
a. Click Locate and install driver software (recommended).
The popup message "Windows needs your permission to continue" appears.
b. Click Continue.
The Windows Security dialog box appears.
c.Click Install this driver software anyway.
For Windows XP:
a. Click Next and/or Finish as required in the wizard.
Once the firmware is loaded, the wizard restarts to initiate the firmware to accept commands.
b. Click Next and/or Finish again.
36
Page 37
Setting Up and Installing the Instrument
USB LED
LMT LED
OPN LED
Note: Windows 7 and Windows 8 find the device automatically.
If the power supply and the instrument are attached correctly, the USB LED on the rear
panel, shown in Figure 10, turns green.
Figure 10: LEDs on the Rear Panel of the Instrument
5. Repeat the steps to attach another TEMPpoint, VOLTpoint, or MEASURpoint instrument
to the host computer, if desired.
37
Page 38
Chapter 3
Connecting to an Expansion Hub
Expansion hubs are powered by their own external power supply. Theoretically, you can
connect up to five expansion hubs to a USB port on the host computer. However, the practical
number of instruments that you can connect to a single USB port depends on the throughput
you want to achieve. Each of the hubs supports up to four instruments.
To connect multiple TEMPpoint, VOLTpoint, or MEASURpoint instruments to an expansion
hub, do the following:
1. Make sure that you have attached a power supply to the instrument.
2. Attach one end of the USB cable to the instrument and the other end of the USB cable to
an expansion hub.
3. Connect the power supply for the expansion hub to an external power supply.
4. Connect the expansion hub to the USB port on the host computer using another USB
cable.
The operating system automatically detects the instrument and starts the Found New Hardware
wizard.
5. For Windows Vista
a. Click Locate and install driver software (recommended).
The popup message "Windows needs your permission to continue" appears.
b. Click Continue.
The Windows Security dialog box appears.
c.Click Install this driver software anyway.
For Windows XP:
a. Click Next and/or Finish as required in the wizard.
Once the firmware is loaded, the wizard restarts to initiate the firmware to accept commands.
b. Click Next and/or Finish again.
Note: Windows 7 and Windows 8 find the device automatically.
If the power supply and the instrument are attached correctly, the USB LED on the rear panel,
shown in Figure 10, turns green.
6. Repeat these steps until you have attached the number of expansion hubs (up to five) and
instruments (up to four per hub) that you require. Refer to Figure 11.
The operating system automatically detects the instruments as they are installed.
:
38
Page 39
Setting Up and Installing the Instrument
+5 V Power Supply
+5 V Power Supply
USB Cable
Expansion Hubs
Host Computer
Power Supply
for Hub
USB Cables
USB Cable
Power Supply
for Hub
+5 V Power Supply
+5 V Power Supply
USB
Cables
TEMPpoint,
VOLTpoint, or
MEASURpoint
TEMPpoint,
VOLTpoint, or
MEASURpoint
TEMPpoint,
VOLTpoint, or
MEASURpoint
TEMPpoint,
VOLTpoint, or
MEASURpoint
Figure 11: Attaching Multiple Instruments Using Expansion Hubs
39
Page 40
Chapter 3
Configuring the Device Driver
Note: In Windows 7, Windows 8, and Vista, you must have administrator privileges to run
the Open Layers Control Panel. When you double-click the Open Layers Control Panel icon,
you may see the Program Compatibility Assistant. If you do, select Open the control panel using recommended settings. You may also see a Windows message asking you if you want
to run the Open Layers Control Panel as a "legacy CPL elevated." If you get this message,
click Yes.
If you do not get this message and have trouble making changes in the Open Layers Control
Panel, right click the DTOLCPL.CPL file and select Run as administrator. By default, this file
is installed in the following location:
Windows 7, Windows 8, and Vista (32-bit)
C:\Windows\System32\Dtolcpl.cpl
Windows 7, Windows 8, and Vista (64-bit)
C:\Windows\SysWOW64\Dtolcpl.cpl
To configure the device driver for a TEMPpoint, VOLTpoint, or MEASURpoint instrument, do
the following:
1. If you have not already done so, power up the host computer and all peripherals.
2. From the Windows Start menu, select Settings -> Control Panel.
3. From the Control Panel, double-click Open Layers Control Panel.
The Data Acquisition Control Panel dialog box appears.
4. Click the name of the instrument that you want to configure, and then click Advanced.
The Configurable Board Options dialog box appears.
5. For each channel, set the Channel Configuration as voltage (the default) or one of the
supported sensor types for that channel.
Note: If you wish, you can overwrite these channel input types programmatically using
your software development environment or application.
6. Click OK.
7. If you want to rename the instrument, click Edit Name, enter a new name for the
instrument, and then click OK. The name is used to identify the instrument in all
subsequent applications.
40
8. Repeat steps 4 to 7 for the other instrument that you want to configure.
9. When you are finished configuring the instrument, click Close.
Continue with the instructions on wiring in Chapter 4 starting on page 41.
Keep the following recommendations in mind when wiring signals to a TEMPpoint,
VOLTpoint, or MEASURpoint instrument:
• Separate power and signal lines by using physically different wiring paths or conduits.
• To avoid noise, do not locate the instrument and cabling next to sources that produce high
electromagnetic fields, such as large electric motors, power lines, solenoids, and electric
arcs, unless the signals are enclosed in a mumetal shield.
• Locate the instrument’s front panel as far away as possible from sources of high or low
temperatures or strong air currents, such as fans.
• Prevent electrostatic discharge to the I/O while the instrument is operational.
• When wiring thermocouples, select an appropriate wire length and gauge for each
thermocouple; in general, use the shortest wire length and largest gauge for the
application to yield best results.
• Use shielded wire for maximum rejection of electrical interference.
Wiring Signals
43
Page 44
Chapter 4
Warm-Up Time
For accurate thermocouple measurements, MEASURpoint instruments require a warm-up
time of 1 hour for the analog circuitry to stabilize.
For accurate RTD measurements, ensure that your RTD sensors and external calibration
resistors warm up for 1 minute after the MEASURpoint instrument has been warmed up for 1
hour.
44
Page 45
Connecting Thermocouple Inputs
The DT9871U, DT9871, and DT9874 instruments contain thermocouple jacks for connecting
thermocouple inputs.
Note: On the standard DT9874 instrument, channels 0 to 15 correspond to the thermocouple
input channels.
Internally, these signals are connected in differential mode. You can mix and match the
following thermocouple types across channels: B, E, J, K, N, R, S, and/or T.
Each thermocouple input jack is polarized and accepts a mating plug in the appropriate
orientation. Tabl e 1 lists the color designations for the + and – polarities of the supported
thermocouple types for both the ANSI (American) and IEC (International) standards.
Table 1: Thermocouple Color Designation Standards
Wiring Signals
Thermocouple
Standard
ANSIType JWhiteRed
IECType JBlackWhite
Thermocouple
Type
Type KYellowRed
Type TBlueRed
Type EVioletRed
Type SBlackRed
Type RBlackRed
Type BGrayRed
Type NOrangeRed
Type KGreenWhite
Type TBrownWhite
Type EVioletWhite
Type SOrangeWhite
Type ROrangeWhite
Wire Color Coding
+ Polarity
Wire Color Coding
– Polarity
Type BGrayWhite
Type NPinkWhite
For more information on thermocouple standards, refer to the following web site:
http://www.omega.com/thermocouples.html.
45
Page 46
Chapter 4
Thermocouple
Channels
Thermocouple Input
–
+
–
+
Omega Plug (SMPW)
CAUTION:
When connecting inputs to the thermocouple connectors on a
MEASURpoint or TEMPpoint instrument, it is highly recommended
that you use only original Omega thermocouple plugs (SMPW), as
connectors from other suppliers may not be equivalent mechanically.
Refer to page 126 for more information on the connectors.
If you use connectors from suppliers other than Omega, there is a
risk that you may mechanically damage the thermocouple
connectors on the MEASURpoint or TEMPpoint instrument.
Figure 12 shows how to connect a thermocouple input to a thermocouple channel.
46
Figure 12: Connecting Thermocouple Inputs
Page 47
Connecting RTD Inputs
RTD
Channels
1
Current
3
– Sense2+Sense
4
Return
Each DT9872 and DT9874 contains pluggable screw terminals for connecting RTD inputs.
Internally, these signals are connected in differential mode.
Note: On the standard DT9874 instrument, channels 16 to 31 correspond to the RTD input
channels.
Figure 13 shows the numbering of the screw terminal blocks for RTD connections.
Wiring Signals
Figure 13: Screw Terminal Block Numbering for RTD Connections
The DT9872 supplies each RTD channel with 425 A of excitation current to prevent
self-heating. The resistance of the RTD circuit increases gradually, repeatably, and linearly
with temperature. As the resistance increases, the voltage drop across the RTD also increases.
47
Page 48
Chapter 4
RTD Channel
RTD
425 A
Current
– Sense+Sense
Return
1243
R
L
R
L
R
L
R
L
The DT9872 reads this voltage drop and automatically converts the voltage to the appropriate
temperature based on the RTD type.
The DT9872 and DT9874 support Pt100 (100 Platinum), Pt500 (500 Platinum), and Pt1000
(1000 Platinum) RTD types using Alpha coefficients of 0.00385 and 0.00392; you can mix and
match RTD types across RTD channels. Refer to the following web site for more information
on RTD types: http://www.omega.com.
To connect an RTD input, you can use a 4-wire, 3-wire, or 2-wire connection scheme, described
in the following subsections. For the best accuracy, use 4-wire RTD connections; this
connection scheme enables Kelvin sensing to minimize errors due to lead wire resistance.
4-Wire RTD Connections
The 4-wire configuration offers the best accuracy with long connection wires, compared to the
3- and 2-wire configurations. The 4-wire connection scheme eliminates errors due to lead wire
resistance (R
is automatically cancelled as long as the sense wires are connected.
Figure 14 shows a 4-wire RTD connection.
) and thermal heating. Wire impedance of up to 100 anywhere in the hookup
L
Figure 14: 4-Wire RTD Connection
48
3-Wire RTD Connections
The 3-wire configuration eliminates one wire from the 4-wire RTD connection. Lead wire
resistance (R
drop is essentially equal and opposite to the voltage drop across +Sense.
Figure 15 shows a 3-wire RTD connection.
) errors in the return wire from –Sense may be introduced unless the voltage
L
Page 49
Wiring Signals
RTD Channel
RTD
425 A
Current
– Sense+Sense
Return
R
L
1243
*RL is lead wire resistance.
Figure 15: 3-Wire RTD Connection
2-Wire RTD Connections
The 2-wire configuration is the least accurate of the RTD wiring configurations because the
lead wire resistance (R
measurement errors, particularly if the lead wire is long. If you decide to use the 2-wire
connection scheme, ensure that you use short lead wire connections.
For example, if the lead resistance is 0.5 in each wire, the lead resistance adds a 1 of error
to the resistance measurement. Using a 100 RTD (Pt100) with a 0.00385/°C European curve
coefficient, the resistance represents an initial error of 1 /(0.385 /°C) or 2.6°C. Since the
lead wire resistance changes with ambient temperature, additional errors are also introduced
in the measurement.
Figure 16 shows a 2-wire RTD connection.
) and its variation with temperature contribute significant
L
49
Page 50
Chapter 4
RTD Channel
RTD
425 A
Current
– Sense+Sense
Return
R
L
R
L
1243
Figure 16: 2-Wire RTD Connection
50
Page 51
Connecting Voltage Inputs
The way you connect voltage inputs depends on the channel type you are using. This section
describes how to connect voltage inputs to thermocouple input channels, RTD input channels,
and voltage input channels.
Connecting Voltage Inputs to Thermocouple Channels
Figure 17 shows how to connect a differential voltage input to a thermocouple input channel
on the DT9871U, DT9871, or DT9874 instrument.
Note: On the standard DT9874 instrument, channels 0 to 15 correspond to the thermocouple
input channels.
Wiring Signals
51
Page 52
Chapter 4
Voltage Input
–
+
Omega Plug (SMPW-U-M)
Analog Input 0Analog Input 0
Return
Signal Source
+–
Thermocouple
Channels
52
Figure 17: Connecting Voltage Inputs to a Thermocouple Channel
Page 53
Connecting Voltage Inputs to RTD Channels
Vin +Vin –
RTD Channel
– Sense+Sense
1243
Shield
Figure 18 shows how to connect a voltage input to an RTD channel on a DT9872 or DT9874
instrument.
Note: On the standard DT9874 instrument, channels 16 to 31 correspond to the RTD input
channels.
Wiring Signals
Figure 18: Connecting Voltage Inputs to an RTD Channel
The input impedance is well over 100 M using the voltage –Sense and +Sense inputs.
For best accuracy when connecting voltage inputs, use twisted-pair wires with a dead-ended
shield connected to pin 4 of the screw terminal block.
53
Page 54
Chapter 4
Voltage Input
Channels
1
+Sense
3
–Sense2No
Connect
4
Shield
Connecting Voltage Inputs to Voltage Channels
Each DT9873 and DT9874 contains pluggable screw terminals for connecting voltage inputs.
Note: On the standard DT9874 instrument, channels 32 to 47 correspond to the voltage input
channels.
Figure 19 shows the numbering of the screw terminal blocks for voltage input connections.
54
Figure 19: Screw Terminal Block Numbering for Voltage Connections
Page 55
Figure 20 shows how to connect voltage inputs to the DT9873 and DT9874.
Vin +Vin –
Voltage Input Channel
– Sense+Sense
1243
Shield
*Pin 2 is no connect
Wiring Signals
The input impedance is well over 100 M using the voltage –Sense and +Sense inputs.
Note: For best accuracy when connecting voltage inputs, use twisted-pair wires with a
dead-ended shield connected to pin 4 of the screw terminal block.
Figure 20: Connecting Voltage Inputs
55
Page 56
Chapter 4
Connecting Current Loop Inputs
In some applications, such as solar cell, fuel cell, and car battery testing applications, you may
want to accurately sense and measure current in a high voltage loop.
TEMPpoint, VOLTpoint, and MEASURpoint instruments provide channel-to-channel
functional isolation of 60 VDC maximum continuous working voltage, meaning that each
input can be referenced to 60 VDC maximum, with a 500 Vpk withstand rating, verified by a
5 s dielectric withstand test.
The way you connect current loop inputs depends on the channel type you are using. This
section describes how to connect current loop inputs to thermocouple input channels, RTD
input channels, and voltage input channels.
Connecting Current Loop Inputs to Thermocouple Channels
Thermocouple input channels on the DT9871U and DT9874 have an input range of ±0.075 V.
Therefore, you can use a 1 series resistor to measure ±0.075 A. Similarly, you can use a 0.1
series resistor to measure ±0.75 A.
Thermocouple input channels on the DT9871 have an input range of ±1.25 V. Therefore, you
can use a 1 series resistor to measure ±1.25 A. Similarly, you can use a 0.1 series resistor to
measure ±12.5 A or a 10 series resistor to measure ±0.125 A.
Figure 21 shows how to wire your signals to measure a current loop. In this example, the input
is referenced to ±60 V.
Note: On the standard DT9874 instrument, channels 0 to 15 correspond to the thermocouple
input channels.
56
Page 57
Wiring Signals
1 series resistor
60 V
Circuit+
+
–
+
–
Use a 1
series resistor to convert current to voltage.
For thermocouple channels on the DT9871U and DT9874, 1
= 0.075 A = 0.075 V.
For thermocouple channels on the DT9871, 1
= 1.25 A = 1.25 V.
Thermocouple
Channels
Figure 21: Connecting Current Loop Inputs to Thermocouple Channels
57
Page 58
Chapter 4
1 shunt resistor
Use a 1
shunt resistor to convert current to voltage: 1 = 1.25 A = 1.25 V.
60 V
Circuit+
+
–
RTD Channel
– Sense+Sense
1243
Shield
Connecting Current Loop Inputs to RTD Channels
RTD channels on the DT9872 and DT9874 instruments have an input range of ±1.25 V.
Therefore, you can use a 1 shunt resistor to measure ±1.25 A. Similarly, you can use a 0.1
shunt resistor to measure ±12.5 A or a 10 shunt resistor to measure ±0.125 A.
Figure 22 shows how to wire your signals to measure a current loop. In this example, the input
is referenced to ±60 V.
Note: On the standard DT9874 instrument, channels 16 to 31 correspond to the RTD input
channels.
58
Figure 22: Connecting Current Loop Inputs to RTD Channels
Page 59
Connecting Current Loop Inputs to Voltage Channels
250
shunt resistor
28 V (can be up to ±60 V)
±20 mA
Load
+
–
Voltage Channel
– Sense+Sense
1243
Shield
In this example, the input range is ±10 V.
Voltage channels on the DT9873 and DT9874 instruments have an input range of ±10 V or
±60 V. You select the input range for each channel using software.
With the 24-bit A/D converter, high current, high side current shunts can be used for
resolutions of less than 0.01 A on a 100 A range.
Typi c al Sh unts :
• Vishay WSMS5515
.2 m - 2W - 100A - 20mV
• Vishay CSM2512S
10 m - 1W - 10A - 100mV
• Deltec MUB-500-50
.1 m - 25W - 500A - 50mV
Notes: On the DT9874 instrument, channels 32 to 47 correspond to the voltage input
channels.
Wiring Signals
Figure 23 shows an example of wiring signals to measure ±20 mA using the ±10 V input
range.
Figure 23: Connecting a Current Loop Input to a Voltage Channel to Measure ±20 mA
59
Page 60
Chapter 4
EP333 Cable
STP37 Screw Terminal Panel
Digital I/O
Connector
Connecting Digital I/O Signals
To make digital I/O connections easier, you can use the optional STP37 screw terminal panel
and EP333 cable with your TEMPpoint, VOLTpoint, or MEASURpoint instrument. Connect
the STP37 to the digital I/O connector of the instrument as shown in Figure 24:
Figure 24: Connecting the Instrument to the STP37
Figure 25 shows the layout of the STP37 screw terminal panel and lists the assignments of
each screw terminal.
60
Page 61
Wiring Signals
J1
2
21
3
22
4
23
5
24
6
25
7
26
8
27
9
28
36
17
35
16
34
15
33
14
32
13
31
12
30
11
29
10
120181937
TB2
TB4
TB3
TB5
TB1
Digital Input 1+
Digital Output 7
Digital Input 1
Digital Input 2+
Digital Input 2
Digital Input 3+
Digital Input 3
Digital Input 4+
Digital Input 4
Digital Input 5+
Digital Input 5
Digital Input 6+
Digital Input 6
Digital Input 7+
Digital Input 7
Not Connected
Not Connected
Digital Output 7
Digital Output 6
Digital Output 6
Digital Output 5
Digital Output 5
Digital Output 4
Digital Output 4
Digital Output 3
Digital Output 3
Digital Output 2
Digital Output 2
Digital Output 1
Digital Output 1
Digital Output 0
Digital Output 0
Digital Input 0+
Digital Input 0
Not Connected
Not Connected
Not Connected
DIN 0 +
DIN 1 +
TTL Outputs
Instrument
Digital I/O Connector
DIN 0 –
DIN 1 –
1 k*
+5 V*
*1 k pull-up to +5 V required for TTL outputs.
pin 21
pin 1
pin 20
pin 2
Figure 25: STP37 Screw Terminal Panel
Connecting Digital Input Signals
Figure 26 shows how to connect digital input signals (lines 0 and 1, in this case) to the digital
I/O connector on the TEMPpoint, VOLTpoint, or MEASURpoint instrument.
Figure 26: Connecting Digital Inputs
61
Page 62
Chapter 4
Digital Out 0
Digital Out 0pin 10
120 V AC
or 240 V AC
Relay
Motor
10 A
Fuse
NeutralHot
30V DC @ 400 mA
+ *
–
pin 29
Instrument
Digital I/O Connector
*Output can switch AC or DC.
Controlled by
Software
Note: Relay, motor, and circuits must be properly rated. Reinforced isolation must be
maintained between the digital output circuits and the relay/motor
switching circuits.
Connecting Digital Output Signals
The digital output lines of a TEMPpoint, VOLTpoint, or MEASURpoint instrument act as
solid-state relays. The customer-supplied signal can be ±30 V at up to 400 mA (peak) AC or
DC.
You can use the digital output lines of the instrument to control solid-state or mechanical
relays or high-current electric motors. Figure 27 shows how to connect digital output signals
to line 0 of the instrument to control a motor relay.
You can verify the operation of your TEMPpoint, VOLTpoint, or MEASURpoint instrument
using the base version of the QuickDAQ application that is provided with the instrument.
(Refer to page 29 for information on installing the QuickDAQ application.)
This chapter describes how to use QuickDAQ to measure and display temperature and
voltage data from a data acquisition device. For this example, the following sensors are
connected to a DT9874 MEASURpoint instrument:
• T thermocouple is connected to analog input channel 0
• 4-wire, Euro PT1000 RTD is connected to analog input channel 8
• ±10 V input signal is connected to analog input channel 16
Verifying the Operation of Your Instrument
65
Page 66
Chapter 5
Select the Instrument
To get started with QuickDAQ, follow these steps:
1. Connect your instrument to the USB port of your computer, and connect your sensors to
the instrument.
2. Start QuickDAQ.
The Device Selection window appears.
3. Select MEASURpoint_USB for the Device Family name.
4. By default, the application "discovers" all devices that are available for the specified
device family and displays the module name of the USB devices in the drop-down list. If
you want to refresh this list to determine if other devices are available, click Refresh.
5. Select the module name that corresponds to your instrument, and click Add.
Information about the device, including the model number, serial number, firmware version, driver
version, and scanning status is displayed.
66
Page 67
6. If you want to rename your device, do the following:
Verifying the Operation of Your Instrument
a. Click the Row Selector button for the device.
b. Click the module name in the Name column to highlight it and enter a meaningful
name to represent each available device.
In this example, Deskunit is used as the device name.
7. Click OK.
The QuickDAQ interface is displayed.
67
Page 68
Chapter 5
Configure the Channels
Configure the channels as follows:
1. Configure each analog input channel by clicking the Configuration menu, and clicking
Input Channel Configuration, or by clicking the Input Channel Configuration toolbar
button ( ).
2. Enable analog input channels 0, 8, and 16 by clicking the checkbox under the Enable
column.
68
3. Under the Channel Name column, enter a meaningful name for each channel. For this
example, enter the following names:
For analog input channel 0, enter T Therm as the channel name.
For analog input channel 8, enter PT1000 RTD as the channel name.
For analog input channel 16, enter Voltage In as the channel name.
4. Under the Sensor column, select the following sensor types for the analog input channels:
For analog input channel 0, select Ty pe T as the thermocouple type.
For analog input channel 8, select Euro PT1000 as the RTD type.
For analog input channel 16, select +/-10V as the input voltage range.
Page 69
Verifying the Operation of Your Instrument
5. Under the Engineering Units column, Deg C is selected by default. If you want to change
this setting, change the temperature units under the Acquisition Config - Acquisition
tab.
In this example, Deg C is used.
6. If you know the offset for your calibrated thermocouple and RTD, enter it under the EU
Offset column. For this example, leave 0 as the EU offset for the thermocouple and RTD channels and 1000 as the EU offset for the voltage input channel.
7. Leave the test point values for each channel unchanged.
8. Click Close to close the Channel Configuration dialog box.
69
Page 70
Chapter 5
Configure the Parameters of the Acquisition Config
Window
For this example, set the Acquisition Config parameters as follows:
1. Click the Recording tab.
70
2. For Filename generation, select Filename.
3. Enter a meaningful name for the data file.
In this example, QuickDAQ Data.hpf is used.
4. Leave the Enable Continuous Acquisition checkbox unchecked.
5. For Acquisition Duration, select 1 minute as the time to acquire the measurement data.
The amount of available disk space is shown; in addition, the number of scans in the Acquisition
Info area is updated based on the acquisition duration that is selected.
6. For X Axis Span, enter 5 seconds as the span for the x-axis.
7. Click the Acquisition tab.
Page 71
Verifying the Operation of Your Instrument
8. For this example, ensure that the following default settings are used:
Sampling Frequency: 10 Hz
Trigger Source: Software
Temperature Unit: Celsius
Filter Type: Moving Average
9. If desired, hide the Acquisition Config window by clicking the Auto-Hide pin ( ) in
the top, right corner of the window.
71
Page 72
Chapter 5
Configure the Appearance of the Channel Display
Window
Configure the appearance of the Channel Display window as follows:
1. Ensure that the Visible Display column in the Plot and Data Config window is checked
for all three enabled channels.
2. Click the Channel Display - Style tab, and select the color scheme that you want for the
Channel Display window.
In this example, black is used for the background color, green is used for the foreground color, and
white is used for the label color.
72
3. Leave the indicator style as 7 segment.
Page 73
Verifying the Operation of Your Instrument
Configure the Appearance of the Channel Plot Window
Configure the appearance of the Channel Plot window as follows:
1. In the Plot and Data Config window, set up the following parameters:
a. Ensure that the Visible Plot column is checked for all three enabled channels.
b. Leave the Show Cursor column unchecked for all three enabled channels.
c.Under Plot Column, use the default plot column setting of 1 for all three enabled
channels.
d. Under the Signal Group column, select A for the thermocouple and RTD channels,
and select B for the voltage input channel.
e. Under the Color column, assign a unique color to each trace.
2. Click the Channel Plot - Control tab, and slide the Plot height bar to the left until you can
see both plots on the screen at once.
73
Page 74
Chapter 5
3. Leave the following settings unchanged:
Plot update rate: 1
Smooth scrolling: unchecked
Disable during acquisition: unchecked
Time domain X axis units: Relative time
Display Milliseconds: unchecked
Display Days: unchecked
4. In the display area, click the tab for the Channel Plot window.
5. Click the Show Legend control (
6. For the top plot in this window, change the text for the label on the x-axis, by doing the
following:
a. Right-click on the label.
b. Select Edit Label.
c.Enter the following text: Thermocouple and RTD Channels
7. For the bottom plot in this window, change the text for the label on the x-axis, by doing
the following:
) on the toolbar.
74
a. Right-click on the label.
b. Select Edit Label.
Page 75
c.Enter the following text: Voltage Channels
The Channel Plot window should appear as follows:
Verifying the Operation of Your Instrument
75
Page 76
Chapter 5
Configure the Appearance of the Statistics Window
Configure the appearance of the Statistics window as follows:
1. Ensure that the Visible Statistics column in the Plot and Data Config window is checked
for all three enabled channels:
2. If desired, hide the Plot and Data Config window by clicking the Auto-Hide pin ( ) in
the top, right corner of the window.
76
Page 77
Position the Windows
If you want see the data that is displayed in the Channel Display, Channel Plot, and Statistics
windows at once, you need to move the windows to different locations in the display area.
In the following example, the Channel Display window is located at the top of the display
area, the Channel Plot window is located in the middle of the display area, and the Statistics
window is located at the bottom of the display area:
Verifying the Operation of Your Instrument
Perform the following steps to position the Channel Display window at the top of the display
area, the Channel Plot window in the middle of the display area, and the Statistics window at
the bottom of the display area:
1. Click the tab for the Statistics window, drag the window toward the middle of the display
area, move the mouse over the guide on the bottom of the guide diamond, and then
release the mouse button.
The Statistics window is now placed at the bottom of the display area.
77
Page 78
Chapter 5
2. Click the tab for the Channel Plot window, drag the window toward the middle of the
display area, move the mouse over the guide on the bottom of the guide diamond, and
then release the mouse button.
The Channel Plot window is now placed in the middle of the display area, revealing the Channel
Display window at the top of the display area.
3. Resize each window, as desired.
78
Page 79
Start the Measurement
Once you have configured the channels and the display area, start acquisition and log data to
disk by clicking the Record toolbar button ().
Results similar to the following are displayed in the display area.
Verifying the Operation of Your Instrument
If desired, you can view the data in Excel by clicking the Open Current Data in Excel toolbar
button ( ).
This section includes the block diagrams for the DT9871U, DT9871, and DT9872 TEMPpoint
instruments, DT9873 VOLTpoint instrument, and DT9874 MEASURpoint instruments.
DT9871U Block Diagram
Figure 28 shows the block diagram of the DT9871U TEMPpoint instrument.
84
Figure 28: Block Diagram of the DT9871U TEMPpoint Instrument
Page 85
DT9871 Block Diagram
Isolation Barrier
Control
FPGA
64 kB
SRAM
1 of 8
Digital Input
Isolators
1 of 8
Digital Output
Isolators
37-Pin D-Shell
Connector
Calibration
ROM
+100 nA Break
Detection
24-Bit
A/D
Isolated
DC-DC
1 of up to 48 Channels
CJC Per
Point
USB 2.0
Processor
ID ROM
Limit LED
USB LED
Open TC LED
Powe r L ED
USB 2.0
Figure 29 shows the block diagram of the DT9871 instrument.
Principles of Operation
Figure 29: Block Diagram of the DT9871 TEMPpoint Instrument
85
Page 86
Chapter 6
Isolation Barrier
Control
FPGA
64 kB
SRAM
1 of 8
Digital Input
Isolators
1 of 8
Digital Output
Isolators
37-Pin D-Shell
Connector
Calibration
ROM
*The Open TC LED is not used
on this instrument.
1
2
3
4
+
–
+425 A
Current Source
2.5 Hz Filter
Sense
Return
24-Bit
A/D
Isolated
DC-DC
1 of up to 48 Channels
USB 2.0
Processor
ID ROM
Limit LED
USB LED
Open TC LED
Power LED
USB 2.0
DT9872 Block Diagram
Figure 30 shows the block diagram of the DT9872 TEMPpoint instrument.
Figure 30: Block Diagram of the DT9872 TEMPpoint Instrument
86
Page 87
DT9873 Block Diagram
Isolation Barrier
Control
FPGA
64 kB
SRAM
1 of 8
Digital Input
Isolators
1 of 8
Digital Output
Isolators
37-Pin D-Shell
Connector
Calibration
ROM
*The Open TC LED is not used
on this instrument.
1 of up to 48 Channels
USB 2.0
Processor
ID ROM
Limit LED
USB LED
Open TC LED
Power LED
USB 2.0
2
1
3
4
+
–
1 M
2.5 Hz Filter
Sense
Return
24-Bit
A/D
Isolated
DC-DC
Gain Select
60
10
Figure 31 shows the block diagram of the DT9873 VOLTpoint instrument.
Principles of Operation
Figure 31: Block Diagram of the DT9873 VOLTpoint Instrument
87
Page 88
Chapter 6
Isolation Barrier
Control
FPGA
64 kB
SRAM
1 of 8
Digital Input
Isolators
1 of 8
Digital Output
Isolators
37-Pin D-Shell
Connector
Calibration
ROM
1
2
3
4
+
–
+425 A
Current Source
2.5 Hz Filter
Sense
Return
24-Bit
A/D
Isolated
DC-DC
Channels 16 to 31
USB 2.0
Processor
ID ROM
Limit LED
USB LED
Open TC LED
Power LED
USB 2.0
Channels 32 to 47
2
1
3
4
+
–
1 M
2.5 Hz Filter
Sense
Return
24-Bit
A/D
Isolated
DC-DC
Gain Select
60
10
+10 nA Break
Detection
24-Bit
A/D
Isolated
DC-DC
Channels 0 to 15
x20
+
–
CJC Per
Point
DT9874 Block Diagram
Figure 32 shows the block diagram of the DT9874 MEASURpoint instrument.
88
Figure 32: Block Diagram of the DT9874 MEASURpoint Instrument
Page 89
Analog Input Features
This section describes the following features of the analog input (A/D) subsystem on
TEMPpoint, VOLTpoint, and MEASURpoint instruments:
• Analog input channels, described on this page
• Input ranges, described on page 92
• Resolution, described on page 94
• Calibration, described on page 94
• Sample clock, described on page 95
• Trigger source, described on page 95
• Conversion modes, described on page 96
•Filtering, described on page 97
• Data format, described on page 97
• Error conditions, described on page 98
Principles of Operation
Analog Input Channels
TEMPpoint, VOLTpoint, and MEASURpoint instruments provide up to 48 analog input
channels. The number of channels and the channel types supported (thermocouple, RTD, or
voltage input) depend on the specific instrument model you purchased, as shown in Tabl e 2.
Table 2: Number and Type of Analog Input Channels
# of
Instrument
Typ e
TEMPpointDT9871U-8 and DT9871-888 thermocouple inputs (numbered 0 to 7)
DT9871U-16 and DT9871-16 16 16 thermocouple inputs (numbered 0 to 15)
DT9871U-24 and DT9871-24 2424 thermocouple inputs (numbered 0 to 23)
DT9871U-32 and DT9871-32 3232 thermocouple inputs (numbered 0 to 31)
DT9871U-40 and DT9871-40 4040 thermocouple inputs (numbered 0 to 39)
DT9871U-48 and DT9871-48 4848 thermocouple inputs (numbered 0 to 47)
DT9872-888 RTD inputs (numbered 0 to 7)
DT9872-161616 RTD inputs (numbered 0 to 15)
Models
Analog Input
Channels
Channel Types
DT9872-24 2424 RTD inputs (numbered 0 to 23)
DT9872-323232 RTD inputs (numbered 0 to 31)
DT9872-404040 RTD inputs (numbered 0 to 39)
DT9872-484848 RTD inputs (numbered 0 to 47)
89
Page 90
Chapter 6
VOLTpointDT9873-888 voltage inputs (numbered 0 to 7)
MEASURpointDT9874-16T-16R-16V 4816 thermocouple inputs (numbered 0 to 15)
Instrument
Typ e
Thermocouple Input Channels
Table 2: Number and Type of Analog Input Channels (cont.)
# of
Analog Input
Models
DT9873-161616 voltage inputs (numbered 0 to 15)
DT9873-242424 voltage inputs (numbered 0 to 23)
DT9873-323232 voltage inputs (numbered 0 to 31)
DT9873-404040 voltage inputs (numbered 0 to 39)
DT9873-484848 voltage inputs (numbered 0 to 47)
Channels
Channel Types
16 RTD inputs (numbered 16 to 31)
16 voltage inputs (numbered 32 to 47)
For channels that support thermocouples, you can attach a voltage input or any of the
following thermocouple types in a mix and match fashion: B, E, J, K, N, R, S, and/or T.
By default, these channels are configured for voltage inputs. You can specify the thermocouple
types for channels using the DT-Open Layers Control Panel applet, described on page 40, the
QuickDAQ application, or by using an API call in your application program.
Note: In a mix-and-match system, it is easy to accidentally mismatch the software and
hardware configuration for a channel. Therefore, it is recommended that you pay particular
attention when configuring channels, since the resultant errors may be not large enough to
notice initially, but may be significantly larger than the accuracy specification for the
instrument.
Since each channel has its own 24-bit A/D, analog inputs are measured simultaneously. Refer
to page 96 for more information on specifying the channels for a scan.
Tabl e 3 lists the supported measurement range for each thermocouple type. (Refer to page 92
for information on the supported input range for voltage measurements.)
90
Page 91
Table 3: Supported Measurement Range for Each Thermocouple Type
Supported Measurement Range
Thermocouple
Typ e
B250 C (482 F)1820° C (3308° F)
E–200 C (–328 F)1000° C (1832° F)
Minimum Maximum
Principles of Operation
J–210 C (–346 F)1200
K–200 C (–328 F)1372° C (2502° F)
N–200 C (–328 F)1300° C (2372° F)
R–50 C (–58 F)1768° C (3214° F)
S–50 C (–58 F)1768° C (3214° F)
T–200 C (–328 F) 400° C (752° F)
C (2192° F)
Refer to Appendix A for the thermocouple accuracy of thermocouple channels over the
dynamic range of the instrument.
Cold Junction Compensation
Each thermocouple channel has its own cold-junction compensation (CJC) at the input. The
software reads the value of the CJC input along with the value of the analog input channel and
automatically corrects for errors based on the specified thermocouple type and the
thermocouple linearization data stored in onboard ROM.
A separate multiplexed A/D is used to acquire all the CJC input values. The software takes
care of correlating the CJC measurements with the analog input measurements.
Note: The software provides the option of returning CJC values in the data stream. This
option is seldom used, but is provided if you want to implement your own temperature
conversion algorithms in software when using continuous operations. Refer to page 97 for
more information on this feature.
Open Thermocouple Detection
Break detection circuitry (+10 nA on the DT9871U and DT9874; +100 nA on the DT9871) is
provided for thermocouple channels to ensure that open thermocouples are detected. The
Open (OPN) LED on the rear panel lights when this condition occurs; see Figure 10 on page 37
for the location of this LED.
91
Page 92
Chapter 6
In addition, the software returns the value SENSOR_IS_OPEN (99999 decimal) for any
channel that was configured for a thermocouple input and has either an open thermocouple
or no thermocouple connected to it. This value is returned anytime a voltage greater than 100
mV is measure on the input, since this value is greater than any legitimate thermocouple
voltage.
If the channel is configured for a voltage input (not a thermocouple type), the Open (OPN)
LED never lights and the SENSOR_IS_OPEN value is not returned. Instead, the voltage value
is returned. If no input is connected to the channel, the software returns a value of
approximately 0.7 V due to the open thermocouple detection pull-up circuit.
RTD Channels
For channels that support RTDs, you can attach a voltage input or any of the following RTD
types in a mix and match fashion: Platinum 100 (Pt100), Platinum 500 (Pt500), or Platinum
1000 (Pt1000) RTD using an European alpha curve of 0.00385 or an American alpha curve of
0.00392. The supported temperature measurement range for these RTD types is –200 C (–328
F) to 850 C (1562 F). You can also measure a resistance value between 0 and 4k Ohms, if
desired. (Refer to page 92 for information on the supported input range for voltage
measurements.)
By default, all channels are configured for voltage inputs. You can specify the RTD types for
channels using the DT-Open Layers Control Panel applet, described on page 40, the
QuickDAQ application, or by using an API call in your application program
Notes: In a mix-and-match system, it is easy to accidentally mismatch the software and
hardware configuration for a channel. Therefore, it is recommended that you pay particular
attention when configuring channels, since the resultant errors may be not large enough to
notice initially, but may be significantly larger than the accuracy specification for the
instrument.
Since each channel has its own 24-bit A/D, analog inputs are measured simultaneously. Refer
to page 96 for more information on specifying the channels for a scan.
Input Ranges
The input voltage range that is supported by your TEMPpoint, VOLTpoint, or MEASURpoint
instrument depends on the specific instrument model that you purchased, as shown in
Tabl e 4.
92
Page 93
Table 4: Supported Input Ranges
Instrument
Type
TEMPpointDT9871U±0.75 V for all channels
DT9871±1.25 V for all channels
DT9872±1.25 V for all channels
ModelsInput Range
Principles of Operation
VOLTpointDT9873±10 V or ±60 V
(software-selectable for each channel)
MEASURpointDT9874-16T-16R-16V±0.75 V for channels 0 to 15;
±1.25 V for channels 15 to 31;
±10 V or ±60 V for channels 32 to 47
(software-selectable for each channel)
a. Note that the maximum input voltage for the ±60 V range cannot be more than 30 Vrms, 42.4 Vpk,
60 VDC.
a
a
Out of Range Data for Thermocouple Channels
Each thermocouple type corresponds to an allowable voltage range. If a voltage is measured
on the input that is outside of the legal range for the selected thermocouple type, the channel
may be configured for the wrong type of thermocouple or something other than a
thermocouple may be connected to the channel.
For channels configured with a thermocouple type of None (voltage), the Limit (LMT) LED on
the rear panel of the instrument lights to alert you when the voltage is out of range; see Figure
10 on page 37 for the location of this LED.
For channels configured with a thermocouple type other than None (voltage), the LMT LED
lights when the temperature limit is out of range for the specified thermocouple type.
In addition, if the input voltage is less than the legal voltage range for the selected
thermocouple type, the software returns the value TEMP_OUT_OF_RANGE_LOW (–88888
decimal). If the input voltage is greater than the legal voltage range for the selected
thermocouple type, the software returns the value TEMP_OUT_OF_RANGE_HIGH (88888
decimal).
Note: If you are continuously measuring from a properly configured thermocouple input
channel and the thermocouple opens or becomes disconnected, the open thermocouple
pull-up circuit causes the input voltage to rise to approximately 0.7 V over a few seconds.
In this case, the temperature value rises very quickly, and you will receive the
TEMP_OUT_OF_RANGE_HIGH (88888 decimal) value followed by the OPEN_SENSOR
(99999 decimal) value. In this case, the OPN LED lights when the open thermocouple is
detected and the LMT LED lights when the temperature limit is out of range for the
thermocouple type.
93
Page 94
Chapter 6
Out of Range Data for RTD Channels
Each RTD type corresponds to an allowable voltage range. If a voltage is measured on the
input that is outside of the legal range for the selected RTD type, the channel may be
configured for the wrong type of RTD or something other than an RTD may be connected to
the channel.
For channels configured with a RTD type of voltage, the Limit (LMT) LED on the rear panel of
the instrument lights to alert you when the voltage is out of range (greater than +1.25 V or less
than –1.25 V); see page 37 for the location of this LED. For channels configured with an RTD
type other than voltage, the LMT LED lights when the temperature limit is out of range for the
specified RTD type.
In addition, if the input voltage is less than the legal voltage range for the selected RTD type,
the software returns the value TEMP_OUT_OF_RANGE_LOW (–88888.0 decimal). If the
input voltage is greater than the legal voltage range for the selected RTD type, the software
returns the value TEMP_OUT_OF_RANGE_HIGH (88888.0 decimal).
Out of Range Data for Voltage Channels
Each voltage input channel has an allowable voltage range (±10 V or ±60 V on MEASURpoint
instruments). You configure the input range for each channel using software. If a voltage is
measured on the input that is outside of the legal range for that channel, the Limit (LMT) LED
on the rear panel of the instrument lights to alert you; see page 37 for the location of this LED.
Resolution
TEMPpoint, VOLTpoint, and MEASURpoint instruments support a resolution of 24 bits for
the analog input subsystem; you cannot specify the resolution in software.
Calibration
Each TEMPpoint, VOLTpoint, and MEASURpoint instrument is factory-calibrated to meet or
exceed its published specifications using standards traceable to NIST.
The calibration process includes multiple steps. First, the A/D on each channel is calibrated
for offset and gain; these values (including the zero point) are stored in ROM. Then, each CJC
circuit is calibrated for thermocouple input channels, and the reference current is
characterized for RTD channels.
While each instrument was designed to preserve high accuracy measurements over time, it is
recommended that your instrument be recalibrated every year to ensure that it meets or
exceeds specifications.
You can calibrate your instrument in the field using precise calibration equipment and the
Measurement Calibration Utility, described on page 15.
94
Page 95
Optionally, you can return your instrument to Data Translation for recalibration. For
information on factory recalibration, contact Data Translation at 508-946-5100 (if you are in the
USA) or call your local distributor (if you are located outside the USA); see our web site
(www.mccdaq.com) for the name and telephone number of your nearest distributor.
In addition, each instrument auto-calibrates on each power-up cycle to guarantee
high-accuracy measurements. This process, also known as auto-zeroing, resets the zero point
of each A/D. You can also auto-calibrate the instrument at any time (as long as acquisition is
not in progress) using a software command. Refer to your software documentation for more
information on the auto-calibration feature.
Sample Clock Source
TEMPpoint, VOLTpoint, and MEASURpoint instruments support an internal clock with a
maximum sampling rate of 10 Samples/s.
Use software to specify an internal clock source and a clock frequency between
0.000152590219 Hz and 10.0 Hz.
Principles of Operation
Note: The clock frequency that you specify is rounded to the closest "correct" value that the
instrument can accept without error. Internally, the 10 Hz clock is divided by an integer in the
range of 1 to 65535 (the internal clock divider) to determine the closest value. Using software,
you can query this setting to determine the actual clock frequency that is used.
When the continuous operation is started, all the channels specified in the channel list are read
simultaneously at the specified clock frequency.
Trigger Sou rce
A trigger is an event that occurs based on a specified set of conditions. Acquisition starts when
the instrument detects the initial trigger event and stops when you stop the operation.
TEMPpoint, VOLTpoint, and MEASURpoint instruments support the following trigger
sources for starting analog input operations:
• Software trigger – A software trigger event occurs when you start the analog input
operation (the computer issues a write to the instrument to begin conversions).
• External digital trigger – An external digital trigger event occurs when the instrument
detects a voltage from +3 V to +28 V DC on digital input line 0. Initially, the external signal
must be low and then go high for at least 100 ms to be detected as a trigger. Once
triggered, the state of digital input 0 is ignored.
95
Page 96
Chapter 6
Conversion Modes
TEMPpoint, VOLTpoint, and MEASURpoint instruments support continuous scan conversion
modes for reading input measurements.
Continuous scan mode takes full advantage of the capabilities of the TEMPpoint, VOLTpoint,
and MEASURpoint instruments. Use continuous scan mode if you want to accurately control
the period between successive simultaneous conversions of specific channels.
In addition to the analog input channels, this conversion mode allows you to read the digital
input port (all 8 digital input lines) as part of the analog input data stream. This feature is
particularly useful when you want to correlate the timing of analog and digital events.
Specifying Analog Input Channels
Using software, enable the analog input channels that you want to sample by specifying the
channel numbers in the channel list. You can also read the value of the digital input port
through the analog input data stream by specifying the digital input channel in the channel
list; the number of the digital input channel depends on how many channels the TEMPpoint,
VOLTpoint, or MEASURpoint instrument provides, as shown in Tab le 5 .
Table 5: Supported Channels for Continuous Operations
Total Number of
Analog Input Channels
0 to 78
0 to 1516
0 to 2324
0 to 3132
0 to 3940
0 to 4748
Channel for Reading
the Digital Input Port
The channels are read in order from the lowest channel number to the highest channel number
in the list of enabled channels; this process is known as a scan.
How Continuous Scan Works
When you issue a command to start the scan, the instrument simultaneously samples all the
analog input channels, CJC inputs (if applicable), and the digital input port, and converts the
analog inputs to temperature, resistance, or voltage based on the sensor type. If the channel is
enabled, the sampled data is placed in the FIFO on the instrument.
96
Page 97
Principles of Operation
Operation starts
Chan 0
Chan 1
Chan 2
Input
Sample
Clock
Data is acquired continuously
Chan 0
Chan 1
Chan 2
Chan 0
Chan 1
Chan 2
Chan 0
Chan 1
Chan 2
The FIFO on the instrument is used as a circular buffer. Acquisition continues indefinitely
until you stop the operation. When the FIFO is full, the operation wraps to the beginning of
the FIFO; values are overwritten starting at the first location in the FIFO. It is up to your
application to retrieve the data from the FIFO; refer to your software documentation for more
information.
The conversion rate is determined by the frequency of the input sample clock; refer to page 95
for more information about the input sample clock. The sample rate, which is the rate at which
a single entry in the channel list is sampled, is the same as the conversion rate due to the
simultaneous nature of the MEASURpoint instrument.
Figure 33 illustrates scanning a list of three enabled channels: channel 0, channel 1, and
channel 2. In this example, analog input data is acquired simultaneously on each clock pulse
of the input sample clock. Data is acquired continuously.
Filtering
TEMPpoint, VOLTpoint, and MEASURpoint instruments use a Delta-Sigma analog-to-digital
converter (ADC) for each analog input channel to provide simultaneous sampling of all
inputs. The Delta-Sigma converter operates at 10 Hz effectively providing a filter that rejects
50 Hz and 60 Hz power line frequency components and that removes aliasing, a condition
where high frequency input components erroneously appear as lower frequencies after
sampling.
In addition to the filter provided in hardware, you can further reduce noise by selecting one of
the following filter options in software: Moving Average or Raw. Refer to your software
documentation for more information on selecting a filter type.
Data Format
TEMPpoint, VOLTpoint, and MEASURpoint instruments return data as 32-bit floating-point
values. How the data is returned depends on the channel type, as described in the following
subsections.
Figure 33: Continuous Scan Mode
97
Page 98
Chapter 6
Data Format for Thermocouple Channels
If you specify a thermocouple type of None for a thermocouple input channel, a voltage
measurement is selected and the instrument returns a voltage value. For the DT9871U and
DT9874 instruments, the value is in the range of ±0.075 V; for the DT9872 instrument, the
value is in the range of ±1.25 V. For all other thermocouple types, a temperature value, in
degrees C, or one of the error constants, described on page 99, is returned.
In normal operation, one floating-point value is returned for each enabled channel (including
the digital input port). If you enable the capability of returning CJC data in the data stream,
described on page 91, two floating-point values are returned in the data stream for each
enabled analog input channel. The first value in the pair represents the temperature (or
voltage) of the channel; the second value in the pair represents the CJC temperature (in
degrees C) for that channel.
Data Format for RTD Channels
If you specify an RTD type of None for an RTD input channel, a voltage measurement is
selected and the instrument returns a voltage value in the range of ±1.25 V. If you specify an
RTD type of Ohms, a resistance value is returned. For all other RTD types, a temperature
value, in degrees C, or one of the error constants, described on page 99, is returned.
One floating-point value is returned for each enabled channel (including the digital input
port).
Data Format for Voltage Channels
For voltage channels on a MEASURpoint instrument, a voltage value in the range of ±10 V or
±60 V is returned for each channel, depending on how the channel was configured. Note that
the maximum input voltage for the ±60 V range cannot be more than 30 Vrms, 42.4 Vpk,
60 VDC.
One floating-point value is returned for each enabled channel (including the digital input
port).
Error Conditions
TEMPpoint, VOLTpoint, and MEASURpoint instruments report overrun errors by sending an
overrun event to the application program. If this error condition occurs, the instrument stops
acquiring and transferring data to the host computer. To avoid this error, try one or more of
the following:
• Reduce the sample rate
• Close any other applications that are running
98
• Run the program on a faster computer
Page 99
Principles of Operation
Additionally, the following constants may be reported to the host:
• 99999.0 – SENSOR_IS_OPEN, described on page 91
• 88888.0 – TEMP_OUT_OF_RANGE_HIGH, described on page 93 and page 94
• –88888.0 – TEMP_OUT_OF_RANGE_LOW, described on page 93 and page 94
If any of these constants is reported, the A/D subsystem continues to acquire data; the error
condition is cleared when the data falls within range.
99
Page 100
Chapter 6
1 of 8 Opto-Isolators
Digital Input
+
-
2.2K
Digital I/O Features
TEMPpoint, VOLTpoint, and MEASURpoint instruments provide 8 digital input lines and 8
digital output lines that you can use to control external equipment, including solid-state or
mechanical relays.
This section describes the following digital I/O features:
• Digital input lines, described below
• Digital output lines, described on page 101
• Channel-to-channel functional isolation, described on page 101
• Resolution, described on page 101
• Operation modes, described on page 101
Digital Input Lines
TEMPpoint, VOLTpoint, and MEASURpoint instruments feature eight, isolated, digital input
lines.
Digital inputs operate from +3 to +28 V DC, with a switching time of 2 ms maximum. Figure
34 shows the digital input circuitry; a 2.2 k resistor is used in series with the LED in the
opto-isolator input.
Figure 34: Digital Input Circuitry
A digital line is high (switch is closed) if its value is 1; a digital line is low (switch is open) if its
value is 0.
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.