Data Translation DT9871, DT9871U, DT9872, DT9873, DT9874 User Manual

Title Page

UM-23654-Z
User’s Manual for Standard TEMPpoint, VOLTpoint, and MEASURpoint USB Instruments
DT9871, DT9871U, DT9872, DT9873, DT9874
Copyright Page
Notice
Measurement Computing Corporation does not authorize any Measurement Computing Corporation product for use in life support systems and/or devices without prior written consent from Measurement Computing Corporation. Life support devices/systems are devices or systems that, a) are intended for surgical implantation into the body, or b) support or sustain life and whose failure to perform can be reasonably expected to result in injury. Measurement Computing Corporation products are not designed with the components required, and are not subject to the testing required to ensure a level of reliability suitable for the treatment and diagnosis of people.
Measurement Computing Corporation, InstaCal, Universal Library, and the Measurement Computing logo are either trademarks or registered trademarks of Measurement Computing Corporation. Refer to the Copyrights & Trademarks section on mccdaq.com/legal for more information about Measurement Computing trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies.
© 2018 Measurement Computing Corporation. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form by any means, electronic, mechanical, by photocopying, recording, or otherwise without the prior written permission of Measurement Computing Corporation.
Radio and Television Interference
This equipment has been tested and found to comply with CISPR EN55011 Class A and EN61326-1 requirements and also with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense.
Changes or modifications to this equipment not expressly approved by Data Translation could void your authority to operate the equipment under Part 15 of the FCC Rules.
Note: This product was verified to meet FCC requirements under test conditions that included use of shielded cables and connectors between system components. It is important that you use shielded cables and connectors to reduce the possibility of causing interference to radio, television, and other electronic devices.

FCC Page

Canadian Department of Communications Statement
This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites applicables aux appareils numériques de la class A prescrites dans le Règlement sur le brouillage radioélectrique édicté par le Ministère des Communications du Canada.

Table of Contents

Table of Contents
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Intended Audience. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
How this Manual is Organized . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Conventions Used in this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Related Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Where To Get Help. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Chapter 1: Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Hardware Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
TEMPpoint Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
VOLTpoint Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
MEASURpoint Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Supported Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Getting Started Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Part 1: Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Chapter 2: Preparing to Use the
Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Checking the System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Installing the Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Viewing the Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Chapter 3: Setting Up and Installing
the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Attaching the Instrument to the Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Connecting Directly to the USB Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Connecting to an Expansion Hub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Configuring the Device Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Chapter 4: Wiring Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
General Wiring Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Warm-Up Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Connecting Thermocouple Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Connecting RTD Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4-Wire RTD Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
3-Wire RTD Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
2-Wire RTD Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5
Contents
Connecting Voltage Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Connecting Voltage Inputs to Thermocouple Channels . . . . . . . . . . . . . . . . . . . . . . . . . 51
Connecting Voltage Inputs to RTD Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Connecting Voltage Inputs to Voltage Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Connecting Current Loop Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Connecting Current Loop Inputs to Thermocouple Channels. . . . . . . . . . . . . . . . . . . . 56
Connecting Current Loop Inputs to RTD Channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Connecting Current Loop Inputs to Voltage Channels . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Connecting Digital I/O Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Connecting Digital Input Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Connecting Digital Output Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Chapter 5: Verifying the Operation
of Your Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Select the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Configure the Channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Configure the Parameters of the Acquisition Config Window . . . . . . . . . . . . . . . . . . . . . . . 70
Configure the Appearance of the Channel Display Window . . . . . . . . . . . . . . . . . . . . . . . . 72
Configure the Appearance of the Channel Plot Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Configure the Appearance of the Statistics Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Position the Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Start the Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Part 2: Using Your Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Chapter 6: Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Block Diagrams. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
DT9871U Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
DT9871 Block Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
DT9872 Block Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
DT9873 Block Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
DT9874 Block Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Analog Input Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Analog Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Thermocouple Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Cold Junction Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Open Thermocouple Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
RTD Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Input Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Out of Range Data for Thermocouple Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Out of Range Data for RTD Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
6
Out of Range Data for Voltage Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Resolution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Sample Clock Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Trigger Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Conversion Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Specifying Analog Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
How Continuous Scan Works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Filtering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Data Format for Thermocouple Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Data Format for RTD Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Data Format for Voltage Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Error Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Digital I/O Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Digital Input Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Digital Output Lines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Channel-to-Channel Functional Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Resolution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Operation Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Contents
Chapter 7: Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
General Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
If Your Instrument Needs Factory Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Appendix A: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Basic Instrument Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Thermocouple Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
System Temperature Error for the DT9871U and DT9874 . . . . . . . . . . . . . . . . . . . . . . 112
System Temperature Error for the DT9871 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
RTD Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Temperature Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Voltage Measurement Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Isolation and Protection Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Memory Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Digital I/O Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Power, Physical, and Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Regulatory Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Connector Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Thermocouple Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
RTD Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
7
Contents
Voltage Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
External Power Supply Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Appendix B: Connector Pin Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
8

About this Manual

TEMPpoint™ is a family of temperature measurement instruments that includes the DT9871, DT9871U, DT8871, DT8871U, DT9872, and DT8872. This manual describes the DT9871, DT9871U, and DT9872 USB models.
VOLTpoint™ is a family of voltage measurement instruments that includes the DT9873 and DT8873. This manual describes the DT9873 USB model.
MEASURpoint™ a family of mixed temperature and voltage measurement instruments that includes the DT9874 andDT8874. This manual describes the DT9874 USB model.
Note: For information on the DT8871U, DT8871, DT8872, DT8873, and DT8874 LXI models of TEMPpoint, VOLTpoint, and MEASURpoint, refer to the User’s Manual for Standard TEMPpoint, VOLTpoint, and MEASURpoint LXI Instruments.
The first part of this manual describes how to install and set up your instrument, and verify that the instrument is working properly.
The second part of this manual describes the features and capabilities of your instrument using the IVI-COM instrument driver software. Troubleshooting information is also provided.
Note: If you are programming the instrument using the IVI-COM driver, refer to the DtxMeasurement IVI-COM driver online help for more information.

Intended Audience

This document is intended for engineers, scientists, technicians, or others responsible for using and/or programming a TEMPpoint, VOLTpoint, or MEASURpoint instrument in the Microsoft® Windows® XP, Windows Vista®, Windows 7, or Windows 8 operating system. It is assumed that you have some familiarity with thermocouples, RTDs, and/or voltages and that you understand your application.
9
About this Manual

How this Manual is Organized

This manual is organized as follows:
Chapter 1, “Overview,” summarizes the major features of the TEMPpoint, VOLTpoint, and MEASURpoint instruments, as well as the supported software and accessories.
Chapter 2, “Preparing to Use the Instrument,” describes how to unpack the instrument, check the system requirements, install the software, and view the documentation online.
Chapter 3, “Setting Up and Installing the Instrument,” describes how to apply power to the instrument and connect the instrument to your computer.
Chapter 4, “Wiring Signals,” describes how to wire signals to the instrument.
Chapter 5, “Verifying the Operation of Your Instrument,” describes how to verify the operation of the instrument.
Chapter 6, “Principles of Operation,” describes the analog input and digital I/O features of the TEMPpoint, VOLTpoint, and MEASURpoint instruments in detail.
Chapter 7, “Troubleshooting,” provides information that you can use to resolve problems with your instrument, should they occur.
Appendix A, “Specifications,” lists the specifications of the TEMPpoint, VOLTpoint, and MEASURpoint instruments.
Appendix B, “Connector Pin Assignments,” describes the pin assignments of the digital I/O connector on the TEMPpoint, VOLTpoint, and MEASURpoint instruments.
• An index completes this manual.

Conventions Used in this Manual

The following conventions are used in this manual:
• Notes provide useful information or information that requires special emphasis, cautions provide information to help you avoid losing data or damaging your equipment, and warnings provide information to help you avoid catastrophic damage to yourself or your equipment.
• Items that you select or type are shown in bold.
CAUTION – This icon denotes a caution, which advises you to consult the documentation where this symbol is marked.
CAUTION – Do not operate this product in a manner not specified in this document. Product misuse can result in a hazard. You can compromise the safety protection built into the product if the product is damaged in any way.
10

Related Information

Refer to the following documents for more information on using a TEMPpoint, VOLTpoint, or MEASURpoint USB instrument:
• DtxMeasurement IVI-COM Driver online help. This document describes how to use the IVI-COM driver to access the capabilities of a TEMPpoint, VOLTpoint, or MEASURpoint instrument.
The IVI-COM driver works with any development environment that supports COM programming, including MATLAB® from The MathWorks or Visual Basic®.NET, Agilent® VEE Pro, National Instruments™ LabVIEW™ or LabWindows™, and so on.
QuickDAQ User’s Manual (UM-25242). This manual describes how to create a QuickDAQ application to acquire and analyze data from a TEMPpoint, VOLTpoint, or MEASURpoint instrument.
• IVI foundation (www.ivifoundation.org)
• Omega Complete Temperature Measurement Handbook and Encyclopedia® or the Omega Engineering web site: http://www.omega.com. Both resources provide valuable information on thermocouple types, RTD types, standards, and linearization.
About this Manual
TM
, Microsoft® Visual C#®.NET

Where To Get Help

Should you run into problems installing or using a TEMPpoint, VOLTpoint, or MEASURpoint instrument, the Data Translation Technical Support Department is available to provide technical assistance. Refer to Chapter 7 for more information. If you are outside the United States or Canada, call your local distributor, whose number is listed on our web site (www.mccdaq.com).
11
About this Manual
12
1

Overview

Hardware Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Supported Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Getting Started Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
13
Chapter 1

Hardware Features

Data Translation provides a number of USB instruments to meet your measurement needs, including the following:
• TEMPpoint – a family of temperature measurement instruments
• VOLTpoint – a family of voltage measurement instruments
• MEASURpoint – a family of mixed temperature and voltage measurement instruments
All of these instruments support Version 2.0 and 1.1 of the USB bus.
The following sections summarize the features of the TEMPpoint, VOLTpoint, and MEASURpoint USB instruments.

TEMPpoint Features

TEMPpoint instruments include the following models: DT9871U, DT9871, and DT9872.
Figure 1 shows a DT9871U instrument.
14
Figure 1: TEMPpoint Instrument
The key features of TEMPpoint instruments are as follows:
• DT9871U and DT9871:
Configurable analog input channels for thermocouple or differential voltage inputs;
easy-access jacks for each channel for quick wiring
One CJC (cold junction compensation) input for each thermocouple channel
B, E, J, K, N, R, S, and T thermocouple types supported; the instrument automatically
linearizes the measurements and returns the data as a 32-bit, floating-point temperature values
Input range of ±0.075 V for the DT9871U (with 0.25 V RMS A/D noise using no
software filtering) and ±1.25 V for the DT9871 (with 5 V RMS A/D noise using no software filtering)
Break-detection circuitry to detect open thermocouple inputs
• DT9872:
Configurable analog input channels for RTDs and differential voltage inputs;
easy-access jacks for each channel for quick wiring
100 , 500 , and 1000 platinum RTD types supported using alpha curves of 0.00385
(European) or 0.00392 (American)
4-wire, 3-wire, or 2-wire configurations; the DT9872 automatically linearizes the
measurements and returns the data as 32-bit, floating-point temperature, resistance, or voltage values
Input range of ±1.25 V
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution measurements
Overview
• Throughput rate of up to 10 Samples/s for all channels.
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the instrument supports anytime calibration, performing an auto-calibration function on software command
• Measurement Calibration Utility allows you to calibrate the instrument in the field (see
page 20 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog input data stream for correlating analog and digital measurements
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from ±30 V at currents up to 400 mA (peak) AC or DC
15
Chapter 1

VOLTpoint Features

Figure 2 shows a VOLTpoint instrument.
Figure 2: VOLTpoint Instrument
The key features of VOLTpoint (DT9873) instruments are as follows:
• Direct connection of analog input channels for differential voltage inputs; removable screw terminal blocks for each channel for quick wiring
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution measurements
• Software-selectable input range of ±10 V or ±60 V per channel (note that for the ±60 V range, no more than 30 Vrms, 42.4 Vpk, 60 VDC is allowed)
• Throughput rate of up to 10 Samples/s for all channels
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the instrument supports anytime calibration, performing an auto-calibration function on software command
• Measurement Calibration Utility allows you to calibrate the instrument in the field (see
page 20 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog input data stream for correlating analog and digital measurements
16
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from ±30 V at currents up to 400 mA (peak) AC or DC

MEASURpoint Features

The standard MEASURpoint (DT9874) instrument provides 16 thermocouple channels, 16 RTD channels, and 16 voltage channels. Figure 3 shows a MEASURpoint instrument.
Overview
Figure 3: MEASURpoint Instrument
The key features of MEASURpoint instruments are as follows:
• Analog Input Channels 0 to 15:
Configurable channels for thermocouple or differential voltage inputs; easy-access
jacks for each channel for quick wiring
One CJC (cold junction compensation) input for each thermocouple channel
B, E, J, K, N, R, S, and T thermocouple types supported; the instrument automatically
linearizes the measurements and returns the data as a 32-bit, floating-point temperature values
Input range of ±0.075 V (with 0.25 V RMS A/D noise using no software filtering)
Break-detection circuitry to detect open thermocouple inputs
• Analog Input Channels 16 to 31:
Configurable analog input channels for RTDs and differential voltage inputs;
easy-access jacks for each channel for quick wiring
100 , 500 , and 1000 platinum RTD types supported using alpha curves of 0.00385
(European) or 0.00392 (American)
17
Chapter 1
4-wire, 3-wire, or 2-wire configurations; the DT9872 automatically linearizes the
measurements and returns the data as 32-bit, floating-point temperature, resistance, or voltage values
Input range of ±1.25 V
• Analog Input Channels 31 to 48:
Direct connection of analog input channels for differential voltage inputs; removable
screw terminal blocks for each channel for quick wiring
Software-selectable input range of ±10 V or ±60 V (note that for the ±60 V range, no
more than 30 Vrms, 42.4 Vpk, 60 VDC is allowed)
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution measurements
• 30 VAC, 60 VDC continuous functional isolation ch-ch and ch-gnd, verified by a 500 Vpk withstand
• Throughput rate of up to 10 Samples/s for all channels
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the instrument supports anytime calibration, performing an auto-calibration function on software command
• Measurement Calibration Utility allows you to calibrate the instrument in the field (see
page 20 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog input data stream for correlating analog and digital measurements
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from ±30 V at currents up to 400 mA (peak) AC or DC
18

Supported Software

The following software is available for use with the TEMPpoint, VOLTpoint, and MEASURpoint USB instruments:
QuickDAQ Base Version – The base version of QuickDAQ is free-of-charge and allows you to acquire and analyze data from all Data Translation USB and Ethernet devices, except the DT9841 Series, DT9817, DT9835, and DT9853/54. Using the base version of QuickDAQ, you can perform the following functions:
Discover and select your devices.
Configure all input channel settings for the attached sensors.
Load/save multiple hardware configurations.
Generate output stimuli (fixed waveforms, swept sine waves, or noise signals).
On each supported data acquisition device, acquire data from all channels supported
in the input channel list.
Choose to acquire data continuously or for a specified duration.
Choose software or triggered acquisition.
Overview
Log acquired data to disk in an .hpf file.
Display acquired data during acquisition in either a digital display using the Channel
Display window or as a waveform in the Channel Plot window.
Choose linear or logarithmic scaling for the horizontal and vertical axes.
View statistics about the acquired data, including the minimum, maximum, delta, and
mean values and the standard deviation in the Statistics window.
Export time data to a .csv or .txt file; you can open the recorded data in Microsoft
Excel® for further analysis.
Read a previously recorded .hpf data file.
Customize many aspects of the acquisition, display, and recording functions to suit
your needs, including the acquisition duration, sampling frequency, trigger settings, filter type, and temperature units to use.
QuickDAQ FFT Analysis Option – When enabled with a purchased license key, the QuickDAQ FFT Analysis option includes all the features of the QuickDAQ Base version plus these features:
The ability to switch between the Data Logger time-based interface and the FFT
Analyzer block/average-based interface.
Supports software, freerun, or triggered acquisition with accept and reject controls for
impact testing applications.
Allows you to perform single-channel FFT (Fast Fourier Transform) operations,
including AutoSpectrum, Spectrum, and Power Spectral Density, on the acquired analog input data. You can configure a number of parameters for the FFT, including the FFT size, windowing type, averaging type, integration type, and so on.
Allows you to display frequency-domain data as amplitude or phase.
19
Chapter 1
Supports dB or linear scaling with RMS (root mean squared), peak, and peak-to-peak
scaling options
Supports linear or exponential averaging with RMS, vector, and peak hold averaging
options.
Supports windowed time channels.
Supports the following response window types: Hanning, Hamming, Bartlett,
Blackman, Blackman Harris, and Flat top.
Supports the ability to lock the waveform output to the analysis frame time.
Allows you to configure and view dynamic performance statistics, including the input
below full-scale (IBF), total harmonic distortion (THD), spurious free dynamic range (SFDR), signal-to-noise and distortion ratio (SINAD), signal-to-noise ratio (SNR), and the effective number of bits (ENOB), for selected time-domain channels in the Statistics window.
Supports digital IIR (infinite impulse response) filters.
QuickDAQ Advanced FFT Analysis Option – When enabled with a purchased software license, the QuickDAQ Advanced FFT Analysis option includes all the features of the QuickDAQ Base version with the FFT Analysis option plus these features:
Allows you to designate a channel as a Reference or Response channel.
Allows you to perform two-channel FFT analysis functions, including Frequency
Response Functions (Inertance, Mobility, Compliance, Apparent Mass, Impedance, Dynamic Stiffness, or custom FRF) with H1, H2, or H3 estimator types, Cross-Spectrum, Cross Power Spectral Density, Coherence, and Coherent Output Power.
Supports the Exponential response window type.
Supports the following reference window types: Hanning, Hamming, Bartlett,
Blackman, Blackman Harris, FlatTop, Exponential, Force, and Cosine Taper windows.
Supports real, imaginary, and Nyquist display functions.
Allows you to save data in the .uff file format.
DtxMeasurement IVI-COM driver – This driver provides access to the instrument’s functions through a COM server. The IVI-COM driver works in any 32-bit or 64-bit development environment that supports COM programming, including MATLAB, Visual Basic.NET, Visual C#.NET, Agilent VEE Pro, LabVIEW, LabWindows, and others.
Measurement Calibration Utility – Users can calibrate a TEMPpoint, VOLTpoint, or MEASURpoint instrument in the field using precise calibration equipment and the Measurement Calibration Utility. Since each instrument consists of up to 48 individual channels, great care must be taken to ensure that proper warm-up times are followed and precise calibration equipment is used.
20
The calibration utility ships with a comprehensive help file that describes the required equipment and calibration procedure, including warm-up times, for each instrument.
The calibration utility allows you to revert to the factory calibration for any or all channels, or revert back to the last user calibration values, if desired. In addition, this utility generates a report that lists the starting and ending calibration values for each channel, allowing traceability.

Accessories

The following optional accessories are available for TEMPpoint, VOLTpoint, or MEASURpoint instruments:
STP37 screw terminal panel – The STP37, shown in Figure 4, permits easy screw terminal connections for accessing the digital I/O signals of a TEMPpoint, VOLTpoint, or MEASURpoint instrument.
Overview
Figure 4: STP37 Screw Terminal Panel
EP333 cableThe EP333, shown in Figure 5, is a 2-meter shielded cable with two 37-pin connectors that connects the STP37 screw terminal panel to the digital I/O connector of the instrument.
Figure 5: EP333 Cable
EP373 Single Rack-mount Kit – Mounts one TEMPpoint, VOLTpoint, or MEASURpoint instrument in a rack.
EP374 Dual Rack-mount Kit – Mounts two TEMPpoint, VOLTpoint, or MEASURpoint instruments side-by-side in a rack.
21
Chapter 1
Set Up and Install the Instrument
(see Chapter 3 starting on page 31)
Wire Signals
(see Chapter 4 starting on page 41)
Verify the Operation of the Instrument
(see Chapter 5 starting on page 63)
Prepare to Use the Instrument
(see Chapter 2 starting on page 25)

Getting Started Procedure

The flow diagram shown in Figure 6 illustrates the steps needed to get started using a TEMPpoint, VOLTpoint, or MEASURpoint instrument. This diagram is repeated in each Getting Started chapter; the shaded area in the diagram shows you where you are in the getting started procedure.
Figure 6: Getting Started Flow Diagram
22

Part 1: Getting Started

2
Preparing to Use the
Instrument
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Checking the System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Installing the Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Viewing the Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
25
Chapter 2
Set Up and Install the Instrument
(see Chapter 3 starting on page 31)
Wire Signals
(see Chapter 4 starting on page 41)
Verify the Operation of the Instrument
(see Chapter 5 starting on page 63)
Prepare to Use the Instrument
(this chapter)
26

Unpacking

Open the shipping box and verify that the following items are present:
• TEMPpoint, VOLTpoint, or MEASURpoint instrument
•USB cable
• EP361 +5V power supply and cable
• For DT9872, DT9874, and DT9874 instruments, a bag of pluggable screw terminable blocks
If an item is missing or damaged, contact Data Translation. If you are in the United States, call the Customer Service Department at (508) 946-5100. An application engineer will guide you through the appropriate steps for replacing missing or damaged items. If you are located outside the United States, call your local distributor, listed on Data Translation’s web site (www.mccdaq.com).
Once you have unpacked your instrument, check the system requirements, as described in the next section.
Preparing to Use the Instrument
27
Chapter 2

Checking the System Requirements

For reliable operation, ensure that your computer meets the following system requirements:
• Processor: Pentium 4/M or equivalent
•RAM: 1 GB
• Screen Resolution: 1024 x 768 pixels
• Operating System: Windows 8, Windows 7, Windows Vista (32- and 64-bit)
Windows XP SP3 (32-bit)
• Disk Space: 4 GB
• Acrobat Reader 5.0 or later for viewing documentation
28

Installing the Software

Install the software for your instrument from the web at https://www.mccdaq.com/downloads/DTSoftware/MEASURpoint.
The installation program guides you through the installation process.
Preparing to Use the Instrument
29
Chapter 2

Viewing the Documentation

Note: To view the documentation, you must have Adobe Reader 5.0 or greater installed on your system.
You can access the documentation for your instrument from the Windows Start menu as follows:
• For documentation about the TEMPpoint, VOLTpoint, or MEASURpoint instrument, click
Programs -> Data Translation, Inc -> Hardware Documentation -> User’s Manual for Standard TEMPoint, VOLTpoint, and MEASURpoint USB Instruments.
• For documentation on QuickDAQ, click Programs -> Data Translation, Inc -> QuickDAQ -> QuickDAQ User’s Manual
• For documentation on the DtxMeasurement IVI-COM driver, click Programs -> Data Translation, Inc -> DtxMeasurement -> DtxMeasurement IVI Driver 1.1.8 Documentation.
The following may be helpful when using Adobe Reader:
• To navigate to a specific section of the document, click a heading from the table of contents on the left side of the document.
• Within the document, click the text shown in blue to jump to the appropriate reference (the pointer changes from a hand to an index finger).
• To go back to the page from which the jump was made, click the right mouse button and Go Back, or from the main menu, click Document, and then Go Back.
• To increase or decrease the size of the displayed document, from the main menu, click View, and then Zoom.
• By default, Adobe Reader smooths text and monochrome images, sometimes resulting in blurry images. If you wish, you can turn smoothing off by clicking File, and then Preferences/General, and unchecking Smooth Text and Images.
30
Loading...
+ 104 hidden pages