User’s Manual for
Standard TEMPpoint, VOLTpoint,
and MEASURpoint LXI
Instruments
DT8871, DT8871U, DT8872, DT8873, DT8874
Copyright Page
Notice
Measurement Computing Corporation does not authorize any Measurement Computing
Corporation product for use in life support systems and/or devices without prior written consent
from Measurement Computing Corporation. Life support devices/systems are devices or systems
that, a) are intended for surgical implantation into the body, or b) support or sustain life and whose
failure to perform can be reasonably expected to result in injury. Measurement Computing
Corporation products are not designed with the components required, and are not subject to the
testing required to ensure a level of reliability suitable for the treatment and diagnosis of people.
Trademark and Copyright Information
Measurement Computing Corporation, InstaCal, Universal Library, and the Measurement
Computing logo are either trademarks or registered trademarks of Measurement Computing
Corporation. Refer to the Copyrights & Trademarks section on mccdaq.com/legal for more
information about Measurement Computing trademarks.
Other product and company names mentioned herein are trademarks or trade names of their
respective companies.
This equipment has been tested and found to comply with CISPR EN55011 Class A and
EN61326-1 requirements and also with the limits for a Class A digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at his own
expense.
Changes or modifications to this equipment not expressly approved by Data Translation could
void your authority to operate the equipment under Part 15 of the FCC Rules.
Note: This product was verified to meet FCC requirements under test conditions that
included use of shielded cables and connectors between system components. It is important
that you use shielded cables and connectors to reduce the possibility of causing interference
to radio, television, and other electronic devices.
FCC
Page
Canadian Department of Communications Statement
This digital apparatus does not exceed the Class A limits for radio noise emissions from
digital apparatus set out in the Radio Interference Regulations of the Canadian Department of
Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites
applicables aux appareils numériques de la class A prescrites dans le Règlement sur le
brouillage radioélectrique édicté par le Ministère des Communications du Canada.
Specifying a Static IP Address for your Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Reconfiguring the PC to Use a Static IP Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
8
About this Manual
TEMPpoint is a family of temperature measurement instruments that includes the DT9871,
DT9871U, DT8871, DT8871U, DT9872, and DT8872. This manual describes the DT8871,
DT8871U, and DT8872 Ethernet LXI (LAN eXtensions for Instrumentation) models.
VOLTpoint is a family of voltage measurement instruments that includes the DT9873 and
DT8873. This manual describes the DT8873 LXI model.
MEASURpoint™ a family of mixed temperature and voltage measurement instruments that
includes the DT9874 and DT8874. This manual describes the DT8874 LXI model.
Note: For information on the DT9871, DT9871U, DT9872, DT9873, and DT9874 USB models
of TEMPpoint, VOLTpoint, and MEASURpoint, refer to the User’s Manual for Standard TEMPpoint, VOLTpoint, and MEASURpoint USBInstruments.
The first part of this manual describes how to install and set up your instrument, and verify
that the instrument is working properly.
The second part of this manual describes the features and capabilities of your instrument
using the IVI-COM instrument driver software. Troubleshooting information is also provided.
Note: If you are programming the instrument using the IVI-COM driver, refer to the
DtxMeasurement IVI-COM driver online help for more information.
If you are using Standard Commands for Programmable Instruments (SCPI) to program your
instrument, refer to the SCPI Programmer’s Manual for MEASURpoint for more information.
Intended Audience
This document is intended for engineers, scientists, technicians, or others responsible for
using and/or programming a TEMPpoint, VOLTpoint, or MEASURpoint instrument. It is
assumed that you have some familiarity with thermocouples, RTDs, and/or voltages and that
you understand your application.
9
About this Manual
How this Manual is Organized
This manual is organized as follows:
• Chapter 1, “Overview,” summarizes the major features of the TEMPpoint, VOLTpoint,
and MEASURpoint instruments, as well as the supported software and accessories.
• Chapter 2, “Preparing to Use the Instrument,” describes how to unpack the instrument,
check the system requirements, install the software, and view the documentation online.
• Chapter 3, “Setting Up and Installing the Instrument,” describes how to apply power to
the instrument and connect the instrument to the network.
• Chapter 4, “Wiring Signals,” describes how to wire signals to the instrument.
• Chapter 5, “Configuring the Instrument Using the Web Interface,” describes how to
configure the instrument using the instrument’s web interface.
• Chapter 6, “Principles of Operation,” describes the analog input and digital I/O features
of the TEMPpoint, VOLTpoint, and MEASURpoint instruments in detail.
• Chapter 7, “Troubleshooting,” provides information that you can use to resolve problems
with your instrument, should they occur.
• Appendix A, “Specifications,” lists the specifications of the TEMPpoint, VOLTpoint, and
MEASURpoint instruments.
• Appendix B, “Connector Pin Assignments,” describes the pin assignments of the digital
I/O connector on the TEMPpoint, VOLTpoint, and MEASURpoint instruments.
• Appendix C, “Configuring Network Settings on Your PC,” describes how to configure the
network settings of your PC to use Auto-IP or a static IP address.
• An index completes this manual.
Conventions Used in this Manual
The following conventions are used in this manual:
• Notes provide useful information or information that requires special emphasis, cautions
provide information to help you avoid losing data or damaging your equipment, and
warnings provide information to help you avoid catastrophic damage to yourself or your
equipment.
• Items that you select or type are shown in bold.
• CAUTION – This icon denotes a caution, which advises you to consult the
documentation where this symbol is marked.
CAUTION – Do not operate this product in a manner not specified in this document.
Product misuse can result in a hazard. You can compromise the safety protection built into
the product if the product is damaged in any way.
10
Related Information
Refer to the following documents for more information on using a TEMPpoint, VOLTpoint, or
MEASURpoint LXI instrument:
• DtxMeasurement IVI-COM Driver online help. This document describes how to use the
IVI-COM driver to access the capabilities of your instrument.
The IVI-COM driver works with any development environment that supports COM
programming, including MATLAB® from The MathWorks
or Visual Basic®.NET, Agilent® VEE Pro, National Instruments™ LabVIEW™ or
LabWindows™, and so on.
• QuickDAQ User’s Manual (UM-25242). This manual describes how to create a QuickDAQ
application to acquire and analyze data from a TEMPpoint, VOLTpoint, or MEASURpoint
instrument.
• SCPI Programmer’s Manual for LXI Measurement Instruments. For programmers who want
to use the SCPI interface to program a TEMPpoint, VOLTpoint, or MEASURpoint
instrument, this document describes the supported SCPI commands and example
programs for these instruments.
• IVI foundation (www.ivifoundation.org)
About this Manual
TM
, Microsoft® Visual C#®.NET
• Omega Complete Temperature Measurement Handbook and Encyclopedia® or the Omega
Engineering web site: http://www.omega.com. Both resources provide valuable information on
thermocouple types, RTD types, standards, and linearization.
Where To Get Help
Should you run into problems installing or using a TEMPpoint, VOLTpoint, or MEASURpoint
instrument, the Data Translation Technical Support Department is available to provide
technical assistance. Refer to Chapter 7 for more information. If you are outside the United
States or Canada, call your local distributor, whose number is listed on our web site
(www.mccdaq.com).
Data Translation provides a number of LXI (Ethernet) instruments to meet your measurement
needs, including the following:
• TEMPpoint – a family of temperature measurement instruments
• VOLTpoint – a family of voltage measurement instruments
• MEASURpoint – a family of mixed temperature and voltage measurement instruments
All of these Ethernet instruments are class C devices that comply with LXI version 1.1.
The following sections summarize the features of the TEMPpoint, VOLTpoint, and
MEASURpoint LXI instruments.
TEMPpoint Features
TEMPpoint instruments include the following models: DT8871U, DT8871, and DT8872.
Figure 1 shows a DT8871U instrument.
14
Figure 1: TEMPpoint Instrument
The key features of TEMPpoint instruments are as follows:
• DT8871U and DT8871:
Configurable analog input channels for thermocouple or differential voltage inputs;
easy-access jacks for each channel for quick wiring
One CJC (cold junction compensation) input for each thermocouple channel
B, E, J, K, N, R, S, and T thermocouple types supported; the instrument automatically
linearizes the measurements and returns the data as a 32-bit, floating-point
temperature values
Input range of ±0.075 V for the DT8871U (with 0.25 V RMS A/D noise using no
software filtering) and
±1.25 V for the DT8871 (with 5 V RMS A/D noise using no
software filtering)
Break-detection circuitry to detect open thermocouple inputs
• DT8872:
Configurable analog input channels for RTDs and differential voltage inputs;
easy-access jacks for each channel for quick wiring
100 , 500 , and 1000 platinum RTD types supported using alpha curves of 0.00385
(European) or 0.00392 (American)
4-wire, 3-wire, or 2-wire configurations; the DT8872 automatically linearizes the
measurements and returns the data as 32-bit, floating-point temperature, resistance, or
voltage values
Input range of ±1.25 V
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution
measurements
Overview
• Throughput rate of up to 10 Samples/s for all channels.
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the
instrument supports anytime calibration, performing an auto-calibration function on
software command
• Measurement Instrument Calibration Utility allows you to calibrate the instrument in the
field (see page 21 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog
input data stream for correlating analog and digital measurements
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from
±30 V at currents up to 400 mA (peak) AC or DC
15
Chapter 1
VOLTpoint Features
Figure 2 shows a VOLTpoint instrument.
Figure 2: VOLTpoint Instrument
The key features of VOLTpoint (DT8873) instruments are as follows:
• Direct connection of analog input channels for differential voltage inputs; removable
screw terminal blocks for each channel for quick wiring
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution
measurements
• Software-selectable input range of ±10 V or ±60 V per channel (note that for the ±60 V
range, no more than 30 Vrms, 42.4 Vpk, 60 VDC is allowed)
• Throughput rate of up to 10 Samples/s for all channels
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the
instrument supports anytime calibration, performing an auto-calibration function on
software command
• Measurement Instrument Calibration Utility allows you to calibrate the instrument in the
field (see page 21 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog
input data stream for correlating analog and digital measurements
16
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from
±30 V at currents up to 400 mA (peak) AC or DC
MEASURpoint Features
The standard MEASURpoint (DT8874) instrument provides 16 thermocouple channels, 16
RTD channels, and 16 voltage channels. Figure 3 shows a MEASURpoint instrument.
Overview
Figure 3: MEASURpoint Instrument
The key features of MEASURpoint (DT8874) instruments are as follows:
• Analog Input Channels 0 to 15:
Configurable channels for thermocouple or differential voltage inputs; easy-access
jacks for each channel for quick wiring
One CJC (cold junction compensation) input for each thermocouple channel
B, E, J, K, N, R, S, and T thermocouple types supported; the instrument automatically
linearizes the measurements and returns the data as a 32-bit, floating-point
temperature values
Input range of ±0.075 V (with 0.25 V RMS A/D noise using no software filtering)
Break-detection circuitry to detect open thermocouple inputs
17
Chapter 1
• Analog Input Channels 16 to 31:
Configurable analog input channels for RTDs and differential voltage inputs;
easy-access jacks for each channel for quick wiring
100 , 500 , and 1000 platinum RTD types supported using alpha curves of 0.00385
(European) or 0.00392 (American)
4-wire, 3-wire, or 2-wire configurations; the DT8872 automatically linearizes the
measurements and returns the data as 32-bit, floating-point temperature, resistance, or
voltage values
Input range of ±1.25 V
• Analog Input Channels 31 to 48:
Direct connection of analog input channels for differential voltage inputs; removable
screw terminal blocks for each channel for quick wiring
Software-selectable input range of ±10 V or ±60 V per channel (note that for the ±60 V
range, no more than 30 Vrms, 42.4 Vpk, 60 VDC is allowed)
• One 24-bit, Delta-Sigma A/D converter per channel for simultaneous, high-resolution
measurements
• 30 VAC, 60 VDC continuous functional isolation ch-ch and ch-gnd, verified by a 500 Vpk
withstand
• Throughput rate of up to 10 Samples/s for all channels
• Software or external, digital trigger on digital input line 0 starts acquisition
• Auto-calibrating front-end resets the zero point on each power-up; in addition, the
instrument supports anytime calibration, performing an auto-calibration function on
software command
• Measurement Instrument Calibration Utility allows you to calibrate the instrument in the
field (see page 21 for more information on this utility)
• 8 opto-isolated digital input lines; you can read the digital input port through the analog
input data stream for correlating analog and digital measurements
• 8 opto-isolated digital output lines; the outputs are solid-state relays that operate from
±30 V at currents up to 400 mA (peak) AC or DC
18
Supported Software
The following software is available for use with the TEMPpoint, VOLTpoint, and
MEASURpoint instruments:
• Eureka Discovery Utility – This utility helps you locate or "discover" all LXI (Ethernet)
instruments that are connected to your system and provides the following information
about your instrument: the IP address, manufacturer, model number, serial number, and
version of the firmware that is running on your instrument. In addition, you can use this
utility to configure Windows firewall settings and update the firmware for your Data
Translation LXI instrument.
• Instrument Web Interface – This built-in interface, described in Chapter 5, allows you to
verify the operation of your instrument and perform basic functions with Internet
Explorer and no additional software. Using it, you can configure your instrument, control
output signals, measure input signals, and save results to disk.
• QuickDAQ Base Version – The base version of QuickDAQ is free-of-charge and allows
you to acquire and analyze data from all Data Translation USB and Ethernet devices,
except the DT9841 Series, DT9817, DT9835, and DT9853/54. Using the base version of
QuickDAQ, you can perform the following functions:
Discover and select your devices.
Overview
Configure all input channel settings for the attached sensors.
Load/save multiple hardware configurations.
Generate output stimuli (fixed waveforms, swept sine waves, or noise signals).
On each supported data acquisition device, acquire data from all channels supported
in the input channel list.
Choose to acquire data continuously or for a specified duration.
Choose software or triggered acquisition.
Log acquired data to disk in an .hpf file.
Display acquired data during acquisition in either a digital display using the Channel
Display window or as a waveform in the Channel Plot window.
Choose linear or logarithmic scaling for the horizontal and vertical axes.
View statistics about the acquired data, including the minimum, maximum, delta, and
mean values and the standard deviation in the Statistics window.
Export time data to a .csv or .txt file; you can open the recorded data in Microsoft
Excel® for further analysis.
Read a previously recorded .hpf data file.
Customize many aspects of the acquisition, display, and recording functions to suit
your needs, including the acquisition duration, sampling frequency, trigger settings,
filter type, and temperature units to use.
19
Chapter 1
• QuickDAQ FFT Analysis Option – When enabled with a purchased license key, the
QuickDAQ FFT Analysis option includes all the features of the QuickDAQ Base version
plus basic FFT analysis features, including the following:
The ability to switch between the Data Logger time-based interface and the FFT
Analyzer block/average-based interface.
Supports software, freerun, or triggered acquisition with accept and reject controls for
impact testing applications.
Allows you to perform single-channel FFT (Fast Fourier Transform) operations,
including AutoSpectrum, Spectrum, and Power Spectral Density, on the acquired
analog input data. You can configure a number of parameters for the FFT, including
the FFT size, windowing type, averaging type, integration type, and so on.
Allows you to display frequency-domain data as amplitude or phase.
Supports dB or linear scaling with RMS (root mean squared), peak, and peak-to-peak
scaling options
Supports linear or exponential averaging with RMS, vector, and peak hold averaging
options.
Supports windowed time channels.
Supports the following response window types: Hanning, Hamming, Bartlett,
Blackman, Blackman Harris, and Flat top.
Supports the ability to lock the waveform output to the analysis frame time.
Allows you to configure and view dynamic performance statistics, including the input
below full-scale (IBF), total harmonic distortion (THD), spurious free dynamic range
(SFDR), signal-to-noise and distortion ratio (SINAD), signal-to-noise ratio (SNR), and
the effective number of bits (ENOB), for selected time-domain channels in the Statistics
window.
Supports digital IIR (infinite impulse response) filters.
• QuickDAQ Advanced FFT Analysis Option – When enabled with a purchased software
license, the QuickDAQ Advanced FFT Analysis option includes all the features of the
QuickDAQ Base version with the FFT Analysis option plus advanced FFT analysis
features, including the following:
Allows you to designate a channel as a Reference or Response channel.
Allows you to perform two-channel FFT analysis functions, including Frequency
Response Functions (Inertance, Mobility, Compliance, Apparent Mass, Impedance,
Dynamic Stiffness, or custom FRF) with H1, H2, or H3 estimator types,
Cross-Spectrum, Cross Power Spectral Density, Coherence, and Coherent Output
Power.
Supports the Exponential response window type.
20
Supports the following reference window types: Hanning, Hamming, Bartlett,
Blackman, Blackman Harris, FlatTop, Exponential, Force, and Cosine Taper windows.
Supports real, imaginary, and Nyquist display functions.
Allows you to save data in the .uff file format.
• DtxMeasurement IVI-COM driver – This driver provides access to the instrument’s
functions through a COM server. The IVI-COM driver works in any 32- or 64-bit
development environment that supports COM programming, including MATLAB, Visual
Basic.NET, Visual C#.NET, Agilent VEE Pro, LabVIEW, LabWindows, and others.
• SCPI commands – Use SCPI commands to program TEMPpoint, VOLTpoint, or
MEASURpoint LXI instruments. Refer to the SCPI Programmer’s Manual for LXI Measurement Instruments for information on the supported SCPI commands and example
programs.
• Measurement Instrument Calibration Utility – Users can calibrate a TEMPpoint,
VOLTpoint, or MEASURpoint instrument in the field using precise calibration equipment
and the Measurement Instrument Calibration Utility. Since each instrument consists of up
to 48 individual channels, great care must be taken to ensure that proper warm-up times
are followed and precise calibration equipment is used.
The calibration utility ships with a comprehensive help file that describes the required
equipment and calibration procedure, including warm-up times, for each instrument.
The calibration utility allows you to revert to the factory calibration for any or all
channels, or revert back to the last user calibration values, if desired. In addition, this
utility generates a report that lists the starting and ending calibration values for each
channel, allowing traceability.
Overview
Refer to the Data Translation web site (www.mccdaq.com) for information about selecting the
right software package for your needs.
21
Chapter 1
Accessories
The following optional accessories are available for TEMPpoint, VOLTpoint, or
MEASURpoint instruments:
• STP37 screw terminal panel – The STP37, shown in Figure 4, permits easy screw terminal
connections for accessing the digital I/O signals of a TEMPpoint, VOLTpoint, or
MEASURpoint instrument.
Figure 4: STP37 Screw Terminal Panel
• EP333 cable–The EP333, shown in Figure 5, is a 2-meter shielded cable with two 37-pin
connectors that connects the STP37 screw terminal panel to the digital I/O connector of
the instrument.
Figure 5: EP333 Cable
• EP373 Single Rack-mountKit – Mounts one TEMPpoint, VOLTpoint, or MEASURpoint
instrument in a rack.
• EP374 Dual Rack-mountKit – Mounts two TEMPpoint, VOLTpoint, or MEASURpoint
instruments side-by-side in a rack.
22
Getting Started Procedure
Set Up and Install the Instrument
(see Chapter 3 starting on page 33)
Wire Signals
(see Chapter 4 starting on page 45)
Verify the Operation of the Instrument
(see Chapter 5 starting on page 67)
Prepare to Use the Instrument
(see Chapter 2 starting on page 27)
The flow diagram shown in Figure 6 illustrates the steps needed to get started using a
TEMPpoint, VOLTpoint, or MEASURpoint instrument. This diagram is repeated in each
Getting Started chapter; the shaded area in the diagram shows you where you are in the
getting started procedure.
Open the shipping box and verify that the following items are present:
• TEMPpoint, VOLTpoint, or MEASURpoint instrument
• EP372 Ethernet cable
• EP361 +5V power supply and cable
• For DT8872, DT8873, and DT8874 instruments, a bag of pluggable screw terminable
blocks.
If an item is missing or damaged, contact Data Translation. If you are in the United States, call
the Customer Service Department at 508-946-5100. An application engineer will guide you
through the appropriate steps for replacing missing or damaged items. If you are located
outside the United States, call your local distributor, listed on Data Translation’s web site
(www.mccdaq.com).
Preparing to Use the Instrument
29
Chapter 2
Checking the System Requirements
For reliable operation, ensure that your computer meets the following system requirements:
• Processor: Pentium 4/M or equivalent
•RAM: 1 GB
• Screen Resolution: 1024 x 768 pixels
• Disk Space: 4 GB
•Ethernet port
• Administrator privileges for software installation
• For access to the instrument web interface:
Java Version 6, Update 5 or greater
Internet Explorer 6.0 or 7.0 web browser
Refer to page 69 for more information on installing Java and configuring your browser
settings
• Acrobat Reader 5.0 or later for viewing documentation