Page 1

使用说明书
氧化锆氧变换器
通信功能(MODBUS)
型号:ZKM
INZ-TN5A0506-C
Page 2

- 目录 -
1. 关于通信功能 ·········································································································································· 1
1.1 概述 ··················································································································································· 1
2. 规格 ························································································································································ 2
2.1 通信规格 ············································································································································ 2
3. 连接 ························································································································································ 3
3.1 跨接器的设定····································································································································· 3
3.2 端子配置(TM2)······························································································································ 3
3.3 连接 ··················································································································································· 4
4. 通信条件设定 ·········································································································································· 6
4.1 设定项目 ············································································································································ 6
4.2 设定操作 ············································································································································ 6
5. MODBUS通信协议 ·································································································································· 7
5.1 概述 ··················································································································································· 7
5.2 信息的结构········································································································································· 8
5.3 从机的应答········································································································································· 9
5.4 功能代码 ·········································································································································· 10
5.5 错误校验代码(CRC-16)的计算 ·····································································································11
5.6 传送控制步骤··································································································································· 13
6. 信息的详细内容····································································································································· 15
6.1 读取字数据[功能代码:03H] ········································································································ 15
6.2 读取只读字数据[功能代码:04H]································································································· 16
6.3 写入字(1 个字)数据[功能代码:06H]······················································································· 17
6.4 写入连续字数据[功能代码:10H]································································································· 18
7. 地址映射表和数据格式·························································································································· 19
7.1 数据格式 ·········································································································································· 19
7.1.1 传输数据格式····························································································································· 19
7.1.2 小数点位置及测量单位的设定方法····························································································· 19
7.1.3 测量数据超过范围时的操作········································································································ 19
7.2 地址映射表······································································································································· 20
-i-
Page 3

1.1 概述
• 本装置具有通过 RS-232C 及 RS-485 接口进行通信的功能,能够与上位计算机等进行双向数据传
输。
• 通信系统由主机和从机构成。
使用 RS-485 接口时,1 台主机(上位计算机、可编程控制器、图示面板等)最多可以连接 31 台
从机(本装置)。
使用 RS-232C 接口时,1 台主机只可以连接 1 台从机(本装置)。
• 由于每次可以通信的从机仅限 1 台,所以主机要通过设定至各从机的“站号”确定通信对象。
• 为了主机和从机之间进行通信,接收、发送数据的格式必须相同。本装置根据 MODBUS 通信协
议确定了通信数据的格式。
1. 关于通信功能
-1-
Page 4

2.1 通信规格
① RS-232C 接口
项目 规格
电气规格 符合 EIA RS-232C 标准
通信方式 2线式 半双工
同步方式 起止式同步
连接状态 1:1
最多连接装置数 1台
传输速率
数据格式
传输代码 HEX 值(MODBUS RTU 模式)
错误校验
隔离 通信部位与其它部位为非隔离
② RS-485 接口
2. 规格
38400 bps
数据长度 8位
停止位 1位
奇偶校验 无
X 流程控制 无
CRC-16
项目 规格
电气规格 符合 EIA RS-485 标准
通信方式 2线式 半双工
同步方式 起止式同步
连接状态 1:N
最多连接装置数 31 台
通信距离 最长 500m(最长)
通信速率
数据格式
传输代码 HEX 值(MODBUS RTU 模式)
错误校验
隔离 与内部回路为隔离
38400 bps
数据长度 8位
停止位 1位
奇偶校验 无
X 流程控制 无
CRC-16
信号线与接地之间为功能隔离
-2-
Page 5

ࡻ
为了防止触电及故障,在所有配线连接完成之前,请勿通电。
3.1 跨接器的设定
按照通信类型,请切换跨接器。
3. 连接
通信类型
RS232C
RS485
JP1 JP2 JP5
3-5
4-6
1-3
2-4
3-5
4-6
1-3
2-4
3.2 端子配置(TM2)
① RS-232C 接口
端子号 信号名称 引脚连接图
1
2
3
② RS-485 接口
端子号 信号名称 引脚连接图
1 RTxD+
2 RTxD–
3
Transmit Data (TXD)
Recive Data (RXD)
信号接地
信号接地
2-3
1-2
123
123
-3-
Page 6

3.3 连接
① RS-232C 接口
᎐
ൺב༺ღ
۰˄ZKM˅
RXD
TXD
TXD
RXD
SG SG
FG
② RS-485 接口
• 请使用屏蔽双绞线连接。
(我们推荐的电线为:KPEV-SB(古河电工制造))
• 电缆最大长度为 500m。在 1 线路上可以连接 1 台主机和最多 31 台从机(本装置)。
• 请将终端电阻 100Ω(1/2W 以上)接在线路两端。
• 请将屏蔽线在靠主机的一点接地。
፮ޤݢᏟ
᎐
+
100Ω˄1/2W˅
ൺב༺ღ
۰˄ZKM˅
RTxD+
-
RTxD
-
SG
+
-
+
፮ޤݢᏟ
100Ω˄1/2W˅
-
• 并不一定要连接 SG,但是连接 SG 可以减少因噪声引起的通信错误。
-4-
Page 7

• 在对本装置施加的噪声等级超过 500V 的环境下使用本装置时,我们建议您在靠主机的线路上
使用噪声过滤器(如下图)。
[噪声过滤器](推荐产品):ZRAC2203-11(TDK 制造)
᎐
ወ໌ࣰఌධ
RS-485
ׁᎧ
• 通信期间如果 EMC 发生问题,请在通信电缆上装上铁氧体芯,用来降噪。
铁氧体芯(推荐产品): ZCAT 系列(TDK 制造)
MSFC 系列(森宫电机制造)
-5-
Page 8

4. 通信条件设定
为了主机与本装置之间能够正常通信,需要进行下列设定:
• 主机与本装置的所有通信条件应设定为同一条件。
• 连接在同一线路上的所有本装置需设定各不相同的“站号”。(请确保没有相同站号的装置)
4.1 设定项目
下表为设定项目。请通过正面的操作键进行设定。
项目 出厂值 设定范围 备注
通信速度
数据长度 8位 固定(不能变更)
停止位 1位 固定(不能变更)
奇偶校验设定 无 固定(不能变更)
站号
4.2 设定操作
请在参数设定画面上进行站号设定。(参见使用说明书)
38400bps
1
固定(不能变更)
0~31
(0:停止通信功能)
请将主机及所有站号的
从机设定为同一条件。
请对各从机设定各不相
同的站号。
-6-
Page 9

5.1 概述
基于 MODBUS 通信协议的通信系统,通常采用“从主机开始通信,从机分别应答”的形式。
传输的顺序如下:
1)主机向从机发送指令信息。
2)子机确认接收的信息中的站号是否与本站的站号相同。
3)如果一致,则执行指令,并返回应答信息。
4)如果不一致,则丢弃接收的指令信息,等待下一个指令信息。
(a) 指令信息中的的站号与本站的站号相同时,
主机 → 从机
主机 ← 从机
(b) 指令信息中的站号与本站的站号不同时,
主机 → 从机
主机 ← 从机
主机通过在指令信息中指定从机的某一个站号,可以与连接在同一线路上的多台从机分别进行通
5. MODBUS 通信协议
指令信息
应答信息
指令信息
(无应答)
通信线上的数据
通信线上的数据
信。
-7-
Page 10

5.2 信息的结构
指令信息及应答信息由站号、功能代码、数据区以及错误校验代码 4 个部分构成,并按此顺序发
送。
下面就是各部分的介绍:
(1) 站号
站号(1 字节)
功能代码(1 字节)
数据区(2~133 字节)
错误校验代码(CRC-16)(2 字节)
图 5-1 信息的结构
这是从机的指定编号。仅可由在分析仪参数设定画面上设定为相同“站号”的从机来执行指令。
(2) 功能代码
这是让从机执行的功能的代码。
详细内容请参见
(3) 数据区
这是执行功能代码所需的数据。数据的结构根据功能代码而不同。详细内容请参见第 6 章。
线圈号和寄存器号配置到本装置内部的各个数据。请指定线圈号和寄存器号来执行该数据的通
信。
相对地址 =
例如,由某功能代码指定的寄存器号为 40003 时,
相对地址 = (40003 的低 4 位)- 1
在信息上使用此地址。
(4) 错误校验代码
5.4 节。
⎛
⎜
⎝
= 0002
线圈号 或 寄存器号
的低4位
⎞
-1
⎟
⎠
这是检测信号传输过程中出现的信息错误(位的变化)所需的代码。
MODBUS 通信协议(RTU 模式)使用 CRC-16(Cycric Redundancy Cheek:循环冗余校验)。
CRC计算方法请参见
5.5 节。
-8-
Page 11

5.3 从机的应答
(1) 正常时的应答
从机生成与各信息对应的应答信息,并回复。此时的信息结构与 5.2 节所解释的相同。
数据区的内容根据功能代码而不同。详细内容请参见第 6 章。
(2) 异常时的应答
指令信息的内容中出现传输错误之外的差错(指定了不存在的功能代码等)时,从机不执行此指
令,并生成错误应答信息。
错误代码 内容 说明
01H
02H
03H
(3) 无应答
错误应答信息的结构如
错误代码如
表 5-1 所示。
功能代码出错
线圈、寄存器的地址出错
线圈、寄存器的个数错误
图 5-2 所示,功能代码为指令信息的功能代码加上 80H的数值。
站号(1 字节)
功能代码 + 80H(1 字节)
错误代码(1 字节)
错误校验代码(CRC-16)(2 字节)
图 5-2 错误应答信息
表 5-1 错误代码
指定了不存在的功能代码。
请确认功能代码。
指定了该功能代码无法使用的线圈号、寄存
器的相对地址。
由于指定的个数过多,指定了不存在的线圈
号、寄存器号。
从机在下列情况下,从机不执行指令信息,也不做应答。
• 指令信息中指定的站号与给从机设定的站号不一致时
• 错误校验代码不一致时,或检测出传输错误(奇偶校验错误等)时。
• 结构信息的数据间隔(空白状态)超过 24 位时间时。(请参见
-9-
5.6 传送控制步骤)
Page 12

5.4 功能代码
MODBUS 通信协议中,根据功能代码分配寄存器号,各功能代码只对特定的寄存器号起作用。
其对应情况如表 5-2、按功能分类的信息字长如表 5-3 所示。
表 5-2 功能代码和目标地址的对应情况
功能代码
No.
03H
04H
06H
10H
功能 目标
读取(连续) 保持寄存器
读取(连续) 输入寄存器
写入 保持寄存器
写入(连续) 保持寄存器
号码 内容
4xxxx
3xxxx
4xxxx
4xxxx
寄存器号
读取/写入 字数据
读取 字数据
读取/写入 字数据
读取/写入 字数据
表 5-3 功能代码和信息字长
[单位:字节]
功能代码 内容 可指定数据数
指令信息 应答信息
最小 最大 最小 最大
03H
04H
06H
10H
读取字数据 64 个字
读取字数据(只读) 64 个字
写入字数据 1个字
写入连续字数据 64 个字
8 8 7 133
8 8 7 133
8 8 8 8
11 137 8 8
-10-
Page 13

5.5 错误校验代码(CRC-16)的计算
CRC-16 为 2 字节(16 位)的错误校验代码。计算范围为从信息的开头(站号)至数据区的最末
尾。
从机计算接收信息的 CRC 后,如果与接收的 CRC 代码不一致,则不做应答。
CRC-16 的计算步骤如下所示:
① 将 16 位寄存器置位 FFFF
。(CRC 寄存器)
H
② 将发送信息的头 1 个字节(8 位)和 CRC 寄存器进行“异或”(XOR),并将结果输入至 CRC
寄存器中。
③ 将 CRC 寄存器的内容向右移 1 位。向 MSB 填补 0。
④ 如果移位前的 LSD 为 0 时,则不做任何动作。
如果移位前的 LSB 为 1 时,则与 A001H 进行 XOR,并将结果输入至 CRC 寄存器。
⑤ 重复步骤③、④8 次。(移动 8 位)
⑥ 对于发送信息的下一个字节,也同样执行②~⑤的步骤。
同样地依次处理发送信息的每个字节。
⑦ 发送信息的最后 1 个字节(除错误代码以外的最后的数据)处理结束时的 CRC 寄存器的值为
CRC 代码。
⑧ 将此 CRC 值按低 8 位、高 8 位的顺序存储到发送信息的错误校验代码中。
发送信息(例)
01H
06H
00H
按顺序计算此间的数据,算出 CRC。
此时 CRC=75
99H
05H
03H
E8H
99H
75H
将调换了 CRC 高位和低位的错误
校验代码加入信息中。
CRC-16 的计算流程如下页的 图 5-3 所示。
-11-
Page 14

开始
将 CR 置为 FFFFH(16 进制数)
将 1 置为 I
对 CR 和指定信息的 I 字符的每个字符(1 个
字节)执行“异或”(XOR)逻辑运算,并将
其结果赋给 CR。
将 J 值为 1
CR 右端位是否为
NO
1?
YES
将 CR 仅右移 1 位后,与 A001H进行
“异或”(XOR)逻辑运算,并将其结果
赋给 CR。
*变量的说明
CR:CRC 错误校验数据(2 字节)
I:指令信息计算字符的位数
J:CR 计算次数的确认
将 CR 仅右移 1 位
J 加 1
NO
8 次计算已结束了吗?
J>8
YES
I 加 1
NO
所有字符的计算已结
束了吗?
I>所有字符的总数
(按指令信息站号、功能代码及数据的顺序进
行计算。)
按低位及高位的顺序,在指令信息的末尾加
入 CR 的计算结果。
YES
结束
图 5-3 CRC-16 的计算流程图
-12-
Page 15

5.6 传送控制步骤
(1) 主机的通信方法
主机应遵守以下项目进行通信:
(1-1) 在发送指令信息之前,设置 48 位时间以上的空白状态。
(1-2) 1 个指令信息的各字节的发送间隔应小于 24 位时间。
(1-3) 指令信息发送后,在 24 位时间以内进入接收待机状态。
(1-4) 接收到应答信息后,设置 48 位时间以上的空白状态,然后再发送下一个指令信息。[与
(1-1)意思相同]
(1-5) 为了安全,应建立确认应答信息后,发现无应答或出现错误时,至少重试 3 次的结构。
注释) 上述的定义为最低需求值。为了安全,建议在编写主机的程序时留 2~3 倍的余量。具体推
荐按如下编程:
38400bps 时,空白状态(1-1)为 5ms 以上、字节间隔(1-2)、从发送到接收的
转换(1-3)为 1ms 以内。
-13-
Page 16

(2) 说明
1)帧检测
本通信系统在线路上有下列 2 种状态。
(a) 空白状态(线路上无数据的状态)
(b) 通信状态(线路上有数据在传输的状态)
连接在线路上的设备最初处于接收状态,监视线路。当线路上出现 24 位时间以上的空白状态
时,检测到上 1 帧结束,在下 1 个 24 位时间以内,进入接收待机状态。线路上一旦出现数据,就
开始接收,再次检测到 24 位时间以上的空白状态时,检测到该帧的结束。即从最初的 24 位时间
以上的空白状态到下一次出现 24 位时间以上的空白状态为止,线路上出现的数据作为 1 帧(汇集
的数据块)接收下来。
因此,1 帧(指令信息)需要遵守下列规则发送。
(1-1)指令信息发送之前,设置 48 位时间以上的空白状态。
(1-2)1 个指令信息的各字节的发送间隔,应在 24 位时间以内。
2)从机应答
从机在检测到帧(检测到 24 位时间以上空白状态)后,将该帧作为指令信息进行处理。指令信
息针对本站时,须返回应答信息,该处理时间约 1~30ms。(该时间随指令信息的内容而变化。)
因此,主机发送指令信息后,须遵守以下规则:
(1-3)指令信息发送后,在 24 位时间以内进入接收待机状态。
需要 2.5ms 以上的时间间隔
(推荐 5ms 以上)
控制站 → 本装置
POL1 POL2
1~30msec
控制站 ← 本装置
POL1 应答数据
线路上的数据
POL1 POL2
POL1 应答数据
-14-
Page 17

6. 信息的详细内容
6.1 读取字数据[功能代码:03H]
功能代码 1个信息可以读取的最大字数 相对地址 寄存器号 内容
03H
(1) 信息的结构
指令信息的结构(字节) 应答信息的结构(字节)
功能代码
(相对地址)
读取字数
CRC 数据
读取字数据的排列
MSB LSB
第一个字数据内容的高位字节
第一个字数据内容的低位字节
下一个字数据内容的高位字节
下一个字数据内容的低位字节
~ ~
最后的字数据内容的高位字节
最后的字数据内容的低位字节
(2) 功能说明
64 个字 0000~0081 40001~40082 设定数据
站号
高位
低位
高位
低位
低位
高位
⎫
⎬
1~60
⎭
第一个字数据内容
下一个字数据内容
~ ~
最后的字数据内容
CRC 数据
站号
功能代码
读取字节数 读取字数×2 读取开始编号
高位
低位
高位
低位
高位
低位
低位
高位
从开始读取的编号起,按读取字数读取连续的字数据。从机按高位字节、低位字节的顺序发送读
取字数据。
-15-
Page 18

6.2 读取只读字数据[功能代码:04H]
功能代码 1个信息可以读取的最大字数 相对地址 寄存器号 内容
0000~0081 30001~30079 内部数据
04H
(1) 信息的结构
指令信息的结构(字节) 应答信息的结构(字节)
读取字数据的排列
MSB LSB
~ ~
(2) 功能说明
64 个字
站号
功能代码
高位
(相对地址)
读取字数
CRC 数据
低位
高位
低位
低位
高位
第一个字数据内容的高位字节
第一个字数据内容的低位字节
下一个字数据内容的高位字节
下一个字数据内容的低位字节
最后的字数据内容的高位字节
最后的字数据内容的低位字节
⎫
⎬
1~60
⎭
1000~1049 31001~31044 错误履历数据
2000~2049 32001~32044 警报履历数据
3000~3049 33001~33044 操作履历数据
第一个字数据内容
下一个字数据内容
~ ~
最后一个字数据
CRC 数据
站号
功能代码
读取字节数 读取字数×2 读取开始编号
内容
高位
低位
高位
低位
高位
低位
低位
高位
从开始读取的编号起,按读取字数读取连续的字数据。从机按高位字节、低位字节的顺序发送读
取字数据。
-16-
Page 19

6.3 写入字(1 个字)数据[功能代码:06H]
功能代码 1个信息可以读取的最大字数 相对地址 寄存器号 内容
06H
(1) 信息的结构
指令信息的结构(字节) 应答信息的结构(字节)
站号
功能代码
写入指定编号
(相对地址)
写入字数据
CRC 数据
(2) 功能说明
将指定的数据写入到写入指定编号的字数据中。主机按高位字节、低位字节的顺序传输写入数
1 个字 0000~0081 40001~40082 设定数据
高位
低位
高位 高位
低位
低位
高位
写入指定编号
(相对地址)
写入字数据
CRC 数据
站号
功能代码
高位
低位
低位
低位
高位
据。
-17-
Page 20

6.4 写入连续字数据[功能代码:10H]
功能代码 1个信息可以读取的最大字数 相对地址 寄存器号 内容
10H
(1) 信息的结构
指令信息的结构(字节) 应答信息的结构(字节)
功能代码
写入开始编号
(相对地址)
写入字数
写入字节数
第一个写入字数据
下一个写入字数据
~ ~
最后的写入字数据
CRC 数据
写入字数据的排列
MSB LSB
第一个字数据内容的高位字节
第一个字数据内容的低位字节
下一个字数据内容的高位字节
下一个字数据内容的低位字节
~ ~
最后的字数据内容的高位字节
最后的字数据内容的低位字节
(2) 功能说明
64 个字 0000~0081 40001~40082 设定数据
站号
高位
低位
高位 高位
低位
高位
低位
高位
低位
⎫
⎬
1~60
⎭
⎫
⎬
写入字数×2
⎭
写入开始编号
(相对地址)
CRC 数据
站号
功能代码
写入字数
高位
低位
低位
高位
高位
低位
低位
低位
高位
从开始写入的编号起,按写入字数写入连续的字数据。主机按高位字节、低位字节的顺序发送写
入字数据。
-18-
Page 21

7. 地址映射表和数据格式
7.1 数据格式
7.1.1 传输数据格式
本装置所使用的 MODBUS 传输协议为 RTU(Remote Terminal Unit)模式。传输的数据为“数值”,
并不是 ASCII 码。
7.1.2 小数点位置及测量单位的设定方法
校对浓度设定值、警报的上下限值、测量浓度值、范围值的数据,在传输的数据上不附加小数点及
测量单位。
为了算出正确的值,请对各数据的小数点位置进行设定。详细内容请参见各数据的“读取/写入数
据”项目内容。
7.1.3 测量数据超过范围时的操作
即使测量数据超出范围,画面显示为“ ̄  ̄  ̄  ̄”,但装置仍然会发送当前的测量值。
-19-
Page 22

7.2 地址映射表
字数据[读取/写入]:功能代码[03H、06H、10H]
相对
地址
10 40011
11 40012
12 40013
13 40014
14 40015
15 40016
16 40017
17 40018
18 40019
19 40020
20 40021
寄存
器号
0 40001
1 40002
2 40003
3 40004
4 40005
5 40006
6 40007
7 40008
8 40009
9 40010
数据
类型
单位【高位】 0:vol%
字
范围的小数点位置【低位】
字 范围
单位【高位】 0:vol%
字
范围的小数点位置【低位】
字 范围
字 最大值/最小值计算时间[小时] 0~240
字 设定标志
自动校对开始时间的年【高位】 00~99
字
自动校对开始时间的月【低位】 1~12
自动校对开始时间的日【高位】 1~31
字
自动校对开始时间的点【低位】 0~23
自动校对开始时间的分【高位】 0~59
字
未使用
自动校对周期时间(天)【高位】 0~99
字
自动校对周期时间(小时)
【低位】
校对满量程气体浓度【高位】
(范围 1)
双字
双字
双字
双字
校对满量程气体浓度【低位】
(范围 1)
校对 0 量程气体浓度【高位】
(范围 1)
校对 0 量程气体浓度【低位】
(范围 1)
校对满量程气体浓度【高位】
(范围 2)
校对满量程气体浓度【低位】
(范围 2)
校对 0 量程气体浓度【高位】
(范围 2)
校对 0 量程气体浓度【低位】
(范围 2)
自动吹净开始时间的年【高位】 00~99
字
自动吹净开始时间的月【低位】 1~12
自动吹净开始时间的日【高位】 1~31
字
自动吹净开始时间的点【低位】 0~23
自动吹净开始时间的分【高位】 0~59
字
未使用
存储内容
0:「0.000」
1:「00.00」
0~9999
(小数点位置为“00.00”时 0200~
5000、“0.000”时 2000~9999)
0:「0.000」
1:「00.00」
0~9999
(小数点位置为“00.00”时 0200~
5000、“0.000”时 2000~9999)
第 2 位:禁止所有校对
0 0
0~99
00010~50000
(1digit = 0.001 vol%)
00010~25000
(1digit = 0.001 vol%)
00010~50000
(1digit = 0.001 vol%)
00010~25000
(1digit = 0.001 vol%)
0 0
读取/写入数据
(0:允许 1禁止)
默认值
0
1
2500
0
1
2500
24
0
99
1
1
0
0
7
0
206000
20000
206000
20000
99
1
1
0
0
-20-
Page 23

相对
地址
寄存
器号
21 40022
22 40023
23 40024
24 40025
25 40026
26 40027
27 40028
28 40029
29 40030
30 40031
31 40032
32 40033
33 40034
34 40035
35 40036
36 40037
37 40038
数据
类型
字
自动吹净周期时间(小时)
【高位】
存储内容
0~99
自动吹净周期时间(分)【低位】 0~99
字 吹净时间[秒] 0~999
校对时的传感器诊断执行标志
【高位】
字
校对时的传感器复活处理执行标志
【低位】
1:有效
0:无效
1:有效
0:无效
字 传感器复活类型 0:施加 AC
字
未使用
未使用
0 0
0 0
0:未使用
1:吹净
2:加热器停止
数字输入 1 设定值
字
3:校对禁止
4:远程校对
5:远程 AO 固定
6:最大•最小计算复位
7:范围切换
数字输入 2 设定值 与数字输入 1 设定值相同
数字输入 3 设定值 与数字输入 1 设定值相同
字
未使用
0 0
0:未使用
1:上限警报
2:下限警报
字 警报接点输出设定值
3:上上限警报
4:下下限警报
5:上下限警报
6:上上下下限警报
加热器温度错误条件[℃]
【高位】
字
加热器温度错误条件滞后宽度
[%]【低位】
浓度警报上限值[vol%]
O
2
【高位】(范围 1)
双字
浓度警报上限值[vol%]
O
2
0~100
0~20
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 1)
浓度警报下限值[vol%]
O
2
【高位】(范围 1)
双字
浓度警报下限值[vol%]
O
2
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 1)
浓度警报上上限值[vol%]
O
2
【高位】(范围 1)
双字
浓度警报上上限值[vol%]
O
2
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 1)
浓度警报下下限值[vol%]
O
2
【高位】(范围 1)
双字
浓度警报下下限值[vol%]
O
2
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 1)
读取/写入数据
默认值
24
0
30
0
0
0
0
0
0
0
70
1
550000
500000
200
100
-21-
Page 24

相对
地址
寄存
器号
38 40039
39 40040
40 40041
41 40042
42 40043
43 40044
44 40045
45 40046
46 40047
47 40048
48 40049
49 40050
50 40051
51 40052
52 40053
53 40054
54 40055
55 40056
56 40057
57 40058
58 40059
59 40060
60 40061
61 40062
62 40063
63 40064
64 40065
65 40066
66 40067
67 40068
68 40069
69 40070
70 40071
71 40072
数据
类型
浓度警报条件滞后宽度[%]
O
字
双字
2
(范围 1)
浓度警报上限值[vol%]
O
2
【高位】(范围 2)
浓度警报上限值[vol%]
O
2
存储内容
读取/写入数据
0~20
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 2)
浓度警报下限值[vol%]
O
2
【高位】(范围 2)
双字
浓度警报下限值[vol%]
O
2
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 2)
浓度警报上上限值[vol%]
O
2
【高位】(范围 2)
双字
浓度警报上上限值[vol%]
O
2
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 2)
浓度警报下下限值[vol%]
O
2
【高位】(范围 2)
双字
浓度警报下下限值[vol%]
O
2
1~550000
(1digit = 0.0001 vol%)
【低位】(范围 2)
浓度警报条件滞后宽度[%]
O
字
2
(范围 2)
0~20
AO 固定设定类型【高位】 1:有效, 0:无效
0:0%
字
AO 固定选择【低位】
1:100%
2:上次计算值
3:用户指定
字 AO 固定设定值 0~100
字 测量回复时间[秒] 0~300
字 工厂设定标志
第 8 位:键盘锁定
(0:无效、1:有效)
字 LCD 背照光自动熄灯时间[分] 0~99
字 暖机操作时间监视时间[分] 0~60
字 移动平均处理时间[秒] 0~60
字 加热器控制设定(控制温度) 700~900
字 加热器控制设定(低温温度) 0~300
字 加热器控制设定(低温温度时间) 0~60
字 未使用
字 4mA(0V)调整值(模拟输出) 0~65535
字 20mA(1V)调整值(模拟输出) 0~65535
双字
双字
双字
双字
双字
0 量程系数【高位】(范围 1)
0 量程系数【低位】(范围 1)
满量程系数【高位】(范围 1)
满量程系数【低位】(范围 1)
0 量程系数【高位】(范围 2)
0 量程系数【低位】(范围 2)
满量程系数【高位】(范围 2)
满量程系数【低位】(范围 2)
AD1 调整值(Low)【高位】
AD1 调整值(Low)【低位】
0~4294967295
(1digit = 0.001)
0~4294967295
0~4294967295
(1digit = 0.001)
0~4294967295
0~4294967295
双字 AD1 调整值(High)【高位】 0~4294967295
默认值
10
550000
500000
200
100
10
0
0
0
10
0
10
45
2
800
200
0
5250
27000
1000
0
1000
0
7700
25000
-22-
Page 25

相对
地址
寄存
器号
72 40073
73 40074
74 40075
75 40076
76 40077
77 40078
78 40079
79 40080
80 40081
81 40082
82 40083
83 40084
84 40085
85 40086
86 40087
87 40088
88 40089
89 40090
90 40091
91 40092
92 40093
93 40094
94 40095
95 40096
96 40097
97 40098
98 40099
99 40100
100 40101
101 40102
102 40103
数据
类型
存储内容
AD1 调整值(High)【低位】
双字
双字
双字
双字
双字
双字
AD2 调整值(Low)【高位】
AD2 调整值(Low)【低位】
AD2 调整值(High)【高位】
AD2 调整值(High)【低位】
AD3 调整值(Low)【高位】
AD3 调整值(Low)【低位】
AD3 调整值(High)【高位】
AD3 调整值(High)【低位】
AD4 调整值(Low)【高位】
AD4 调整值(Low)【低位】
AD4 调整值(High)【高位】
AD4 调整值(High)【低位】
0~4294967295
0~4294967295
0~4294967295
0~4294967295
0~4294967295
0~4294967295
字 辉度值 0~65535
第 0 位:0 量程气体气瓶阀门接点
第 1 位:满量程气体气瓶阀门接点
第 2 位:吹净接点(DO)
接点的 AB 设定【高位】
第 3 位:保养接点(DO)
第 4 位:异常接点(FAULT)
字
第 5 位:警报接点(ALARM)
(0:A、1:B)
第 0 位:DI1 接点(DI)
接点的 AB 设定【低位】
第 1 位:DI2 接点(DI)
第 2 位:DI3 接点(DI)
(0:A、1:B)
字 校对等待(白费)时间[秒] 0~60
字 PID 控制参数:P 0~65535
字 PID 控制参数:I 0~65535
字 PID 控制参数:D 0~65535
字 AC 电压施加时间(分) 0~65535
字 未使用
字 未使用
字 传感器诊断阻抗值(Ω) 0~999
字 传感器阻抗 0~65535
站号【高位】
字
【低位】未使用
字 通信设定
0~99
0:RS232C
1:RS485
字 燃烧效率系数 0~199
字 稳定判断时间 (秒) 0~600
字 校对系数决定监视时间 (秒) 0~600
字 范围标志
字 校对范围 连锁标志
0:范围 1
1:范围 2
0:范围连锁
1:显示范围
读取/写入数据
(DO)
(DO)
(DO)
(DO)
默认值
2000
32000
2000
32000
2000
32000
150
0
10
20
3000
600
20
0
70
420
32
0
0
-23-
Page 26

只读字数据:功能代码[04H]
相对
地址
寄存
器号
(H)
0 30001
1 30002
2 30003
3 30004
4 30005
5 30006
6 30007
7 30008
8 30009
9 30010
10 30011
11 30012
12 30013
13 30014
14 30015
15 30016
16 30017
17 30018
18 30019
19 30020
20 30021
21 30022
22 30023
23 30024
数据
类型
双字
氧浓度(%)移动平均后【高位】
存储内容 读取数据
氧浓度(%)移动平均后【低位】
双字
氧气 mV 换算值【高位】
氧气 mV 换算值【低位】
字 加热器温度(°C)
字 燃烧效率(%)
双字
双字
氧浓度最大值【高位】
氧浓度最大值【低位】
氧浓度最小值【高位】
氧浓度最小值【低位】
现在时间的秒【高位】 0~59
字
现在时间的分【低位】 0~59
现在时间的点【高位】 0~23
字
现在时间的星期【低位】
现在时间的日【高位】 1~31
字
现在时间的月【低位】 1~12
现在时间的年【高位】 00~99
字
未使用
下次的自动校对开始时间的年【高位】 00~99
字
下次的自动校对开始时间的月【低位】 1~12
下次的自动校对开始时间的日【高位】 1~31
字
下次的自动校对开始时间的点【低位】 0~23
下次的自动校对开始时间的分【高位】 0~59
字
未使用
下次的自动吹净开始时间的年【高位】 00~99
字
下次的自动吹净开始时间的月【低位】 1~12
下次的自动吹净开始时间的日【高位】 1~31
字
下次的自动吹净开始时间的点【低位】 0~23
下次的自动吹净开始时间的分【高位】 0~59
字
未使用
字 未使用
字 未使用
字 未使用
字 事件变量【高位】
0~0xFFFFFFFF
(1digit = 0.0001 vol%)
0~0xFFFFFFFF
(1digit = 0.001 mV)
0~0xFFFF
(1digit = 0.1 °C)
0~0xFFFF
(1digit = 0.1 %)
0~0xFFFFFFFF
(1digit = 0.0001 vol%)
0~0xFFFFFFFF
(1digit = 0.0001 vol%)
1:天、2:一、4:二、8:三、
16:四、32:五、64:六
0
0
0
第 0 位:0 量程输入调整
第 1 位:满量程输入调整
第 2 位:温度 0 量程输入调整
第 3 位:温度满量程输入调整
第 4 位:补助 0 量程输入调整
第 5 位:补助满量程输入调整
第 6 位: 4mA 调整期间标志
第 7 位:20mA 调整期间标志
(0:无、1:正在执行)
-24-
Page 27

相对
地址
23 30024
24 30025
25 30026
寄存
器号
(H)
数据
类型
字 事件变量【低位】
事件变量【高位】
字
事件变量【低位】
事件变量【高位】
字
事件变量【低位】
存储内容 读取数据
第 0 位:手动满量程校对
第 1 位:自动满量程校对
第 2 位:手动 0 量程校对
第 3 位:自动 0 量程校对
第 4 位:手动一并校对
第 5 位:自动校对开始
第 6 位:自动吹净
第 7 位:手动吹净
(0:无、1:正在执行)
第 0 位:手动传感器诊断
第 1 位:传感器诊断(校对时)
第 3 位:手动传感器复活
第 4 位:传感器复活(校对时)
第 6 位:施加 AC
第 7 位:传感器高温
(0:无、1:正在执行)
第 0 位:将校对系数写入 EEPROM
第 1 位:满量程气体气瓶阀门开关
(1:开、0:关)
第 2 位:零点气体气瓶阀门开关
(1:开、0:关)
第 4 位:计算复位
第 5 位:背照光熄灯时间监视标志
第 6 位:键盘锁定状态
第 7 位:校对禁止状态
(0:无、1:正在执行)
第 0 位:加热器控制
第 1 位:固定信号
第 2 位:暖机操作固定
第 4 位:背照光熄灯用按键输入
第 5 位:传感器复活等待高温
第 6 位:自动吹净开始
第 7 位:手动一并校对开始
(0:无、1:正在执行)
第 0 位:远程加热器处于 OFF 状态
第 1 位:远程模拟输出处于固定状态
第 2 位: 手动 0 量程校对时执行传感器
诊断
第 3 位:自动校对时执行传感器诊断
第 4 位: 自动校对时执行传感器复活执
行数次
第 5 位:低温暖机运行
第 6 位:远程校对
第 7 位:远程吹净
(0:无 1:正在执行)
-25-
Page 28

相对
地址
寄存
器号
(H)
26 30027
27 30028
28 30029
29 30030
30 30031
31 30032
32 30033
33 30034
34 30035
35 30036
数据
类型
字
字
字
字
字
字
双字
双字
存储内容 读取数据
用于警报控制的变量【高位】
用于警报控制的变量【低位】
用于警报控制的变量【高位】
用于警报控制的变量【低位】
数字接点输入 远程校对开始【高位】
数字接点输入 校对禁止【低位】
数字接点输入 吹净【高位】
数字接点输入 计算复位【低位】
数字接点输入 (1)状态【高位】
数字接点输入 (2)状态【低位】
数字接点输入 (3)状态【高位】
后备
AO 的 110%输出值【高位】
AO 的 110%输出值【低位】
AD1(O
AD1(O
浓度)【高位】
2
浓度)【低位】
2
第 0 位:传感器断线
第 1 位: 用于温度控制的热电偶线路断
线
第 2 位: 用于燃烧管理的热电偶线路断
线
第 3 位:阻抗设定错误
第 4 位:传感器异常
第 5 位:超出 O
范围
2
(0:正常、1:发生错误)
第 0 位:满量程气体校对稳定化错误
第 1 位:满量程校对错误
第 2 位:0 量程气体校对稳定化错误
第 3 位:0 量程校对错误
第 4 位:校对发生异常
第 6 位:加热器高温设定超时
(0:正常、1:发生错误)
第 0 位:暖机操作错误
第 1 位:加热器温度错误
第 2 位:A/D 饱和错误
第 5 位:O
异常
2
(0:正常、1:发生错误)
第 0 位:O
第 1 位:O
第 2 位:O
第 3 位:O
上上限值警报
2
上限值警报
2
下限值警报
2
下下限值警报
2
第 4 位:警报信息
第 5 位:错误信息
第 7 位:少氧模式
(0:正常、1:发生错误)
0:未执行远程校对
1:执行远程校对
0:校对有效
1:处于禁止校对状态
0:未执行远程吹净
1:执行远程吹净
0:未执行计算复位
1:执行计算复位
0:数字输入 1 为 ON
1:数字输入 1 为 OFF
0:数字输入 2 为 ON
1:数字输入 2 为 OFF
0:数字输入 3 为 ON
1:数字输入 3 为 OFF
-
0~0xFFFFFFFF
0~0xFFFFFFFF
-26-
Page 29

相对
地址
寄存
器号
(H)
36 30037
37 30038
38 30039
39 30040
40 30041
41 30042
42 30043
43 30044
44 30045
45 30046
46 30047
47 30048
48 30049
49 30050
50 30051
51 30052
52 30053
53 30054
54 30055
55 30056
56 30057
57 30058
58 30059
59 30060
60 30061
61 30062
62 30063
63 30064
64 30065
65 30066
66 30067
67 30068
68 30069
69 30070
70 30071
71 30072
72 30073
73 30074
74 30075
75 30076
76 30077
77 30078
78 30079
数据
类型
双字
AD2(加热器温度控制)【高位】
存储内容 读取数据
0~0xFFFFFFFF
AD2(加热器温度控制)【低位】
双字
双字
双字
AD3(燃烧管理)【高位】
AD3(燃烧管理)【低位】
AD4(热电偶环境温度)【高位】
AD4(热电偶环境温度)【低位】
氧浓度(%)mv -> % 变换后【高位】
氧浓度(%)mv -> % 变换后【低位】
0~0xFFFFFFFF
0~0xFFFFFFFF
0~0xFFFFFFFF
(1digit = 0.0001 vol%)
字 氧浓度(%)用于模拟输出 0~0xFFFF
双字
双字
双字
双字
双字
双字
双字
双字
双字
传感器温度(计数值)【高位】
传感器温度(计数值)【低位】
加热器温度的 mV 换算值【高位】
加热器温度的 mV 换算值【低位】
加热器温度 mV 的平均值【高位】
加热器温度 mV 的平均值【低位】
加热器温度(°C)
字
(温度补偿后、线性化后)
补助输入温度的 mV 换算值【高位】
补助输入温度 mV 换算值【低位】
燃烧管理 mV 的平均值【高位】
燃烧管理 mV 的平均值【低位】
字 燃烧效率温度
热电偶(RCJ)(AD 值)【高位】
热电偶(RCJ)(AD 值)【低位】
热电偶(RCJ)平均值【高位】
热电偶(RCJ)平均值【低位】
加热器温度输入 0 量程【高位】
加热器温度输入 0 量程【低位】
加热器温度输入满量程【高位】
加热器温度输入满量程【低位】
0~0xFFFFFFFF
0~0xFFFFFFFF
(1digit = 0.001 mv)
0~0xFFFFFFFF
(1digit = 0.001 mv)
0~0xFFFF
(1digit = 1 °C)
0~0xFFFFFFFF
(1digit = 0.001 mv)
0~0xFFFFFFFF
(1digit = 0.001 mv)
0~0xFFFF
(1digit = 0.1 °C)
0~0xFFFFFFFF
0~0xFFFFFFFF
0~0xFFFFFFFF
0~0xFFFFFFFF
字 PWM ON 计数器 0~50
双字
双字
双字
双字
加热器温度的偏差【高位】
加热器温度的偏差【低位】
加热器温度的上次偏差【高位】
加热器温度的上次偏差【低位】
加热器温度的上上次偏差【高位】
加热器温度的上上次偏差【低位】
操作量(%)【高位】
操作量(%)【低位】
0~0xFFFFFFFF
(1digit = 0.0001 °C)
0~0xFFFFFFFF
(1digit = 0.0001 °C)
0~0xFFFFFFFF
(1digit = 0.0001 °C)
0~10000
MAX MIN 计算时间(倒数计时)
【高位】
双字
MAX MIN 算出时间(倒数计时)
0~0xFFFFFFFF
【低位】
字 用于阻抗的 EO 0~0xFFFF
字 用于阻抗的 EC 0~0xFFFF
字 用于阻抗的 偏移 0~0xFFFF
-27-
Page 30

只读字数据:功能代码[04H]
错误履历信息
相对地址
1000 31001
寄存器号
(H)
数据类型 存储内容 读取数据 默认值
字
错误履历数据存储数【高位】 0~12
下一个存储指针【低位】 0~11
1001 31002
1002 31003
字
字
数据存储起始指针(最新)【高位】 0~11
数据存储末尾指针(最旧)【低位】 0~11
数据存储位置 0 的错误代码【高位】 0~255
数据存储位置 1 的错误代码【低位】 0~255
~ ~ ~ ~ ~ ~
1007 31008
1008 31009
1009 31010
1010 31011
字
字
字
字
数据存储位置 10 的错误代码【高位】 0~255
数据存储位置 11 的错误代码【低位】 0~255
数据存储位置 0 的错误发生时间的年【高位】 00~99
数据存储位置 0 的错误发生时间的月【低位】 1~12
数据存储位置 0 的错误发生时间的日【高位】 1~31
数据存储位置 0 的错误发生时间的点【低位】 0~23
数据存储位置 0 的错误发生时间的分【高位】 0~59
数据存储位置 0 的错误发生时间的秒【低位】 0~99
~ ~ ~ ~ ~ ~
1041 31042
1042 31043
1043 31044
字
字
字
数据存储位置 11 的错误发生时间的年
【高位】
数据存储位置 11 的错误发生时间的月
【低位】
数据存储位置 11 的错误发生时间的日
【高位】
数据存储位置 11 的错误发生时间的点
【低位】
数据存储位置 11 的错误发生时间的分
【高位】
数据存储位置 11 的错误发生时间的秒
【低位】
00~99
1~12
1~31
0~23
0~59
0~99
0
0
0
0
0
0
0
0
0
1
1
0
0
0
0
1
1
0
0
0
-28-
Page 31

只读字数据:功能代码[04H]
警报履历信息
相对地址
2000 32001
寄存器号
(H)
数据类型 存储内容 读取数据 默认值
字
警报履历数据存储数【高位】 0~12
下一个存储指针【低位】 0~11
2001 32002
2002 32003
字
字
数据存储起始指针(最新)【高位】 0~11
数据存储末尾指针(最旧)【低位】 0~11
数据存储位置 0 的错误代码【高位】 0~255
数据存储位置 1 的错误代码【低位】 0~255
~ ~ ~ ~ ~ ~
2007 32008
2008 32009
2009 32010
2010 32011
字
字
字
字
数据存储位置 10 的错误代码【高位】 0~255
数据存储位置 11 的错误代码【低位】 0~255
数据存储位置 0 的警报发生时间的年【高位】 00~99
数据存储位置 0 的警报发生时间的月【低位】 1~12
数据存储位置 0 的警报发生时间的日【高位】 1~31
数据存储位置 0 的警报发生时间的点【低位】 0~23
数据存储位置 0 的警报发生时间的分【高位】 0~59
数据存储位置 0 的警报发生时间的秒【低位】 0~99
~ ~ ~ ~ ~ ~
2041 32042
2042 32043
2043 32044
字
字
字
数据存储位置 11 的警报发生时间的年
【高位】
数据存储位置 11 的警报发生时间的月
【低位】
数据存储位置 11 的警报发生时间的日
【高位】
数据存储位置 11 的警报发生时间的点
【低位】
数据存储位置 11 的警报发生时间的分
【高位】
数据存储位置 11 的警报发生时间的秒
【低位】
00~99
1~12
1~31
0~23
0~59
0~99
0
0
0
0
0
0
0
0
0
1
1
0
0
0
0
1
1
0
0
0
-29-
Page 32

只读字数据:功能代码[04H]
操作履历信息
相对地址
3000 33001
寄存器号
(H)
数据类型 存储内容 读取数据 默认值
字
操作履历数据存储数【高位】 0~12
下一个存储指针【低位】 0~11
3001 33002
3002 33003
字
字
数据存储起始指针(最新)【高位】 0~11
数据存储末尾指针(最旧)【低位】 0~11
数据存储位置 0 的错误代码【高位】 0~255
数据存储位置 1 的错误代码【低位】 0~255
~ ~ ~ ~ ~ ~
3007 33008
3008 33009
3009 33010
3010 33011
字
字
字
字
数据存储位置 10 的错误代码【高位】 0~255
数据存储位置 11 的错误代码【低位】 0~255
数据存储位置 0 的操作发生时间的年【高位】 00~99
数据存储位置 0 的操作发生时间的月【低位】 1~12
数据存储位置 0 的操作发生时间的日【高位】 1~31
数据存储位置 0 的操作发生时间的点【低位】 0~23
数据存储位置 0 的操作发生时间的分【高位】 0~59
数据存储位置 0 的操作发生时间的秒【低位】 0~99
~ ~ ~ ~ ~ ~
3041 33042
3042 33043
3043 33044
字
字
字
数据存储位置 11 的操作发生时间的年
【高位】
数据存储位置 11 的操作发生时间的月
【低位】
数据存储位置 11 的操作发生时间的日
【高位】
数据存储位置 11 的操作发生时间的点
【低位】
数据存储位置 11 的操作发生时间的分
【高位】
数据存储位置 11 的操作发生时间的秒
【低位】
00~99
1~12
1~31
0~23
0~59
0~99
0
0
0
0
0
0
0
0
0
1
1
0
0
0
0
1
1
0
0
0
-30-
Page 33

★ 说明书意见表 ★
尊敬的顾客
您好,如您对本说明书有宝贵的建议、要求及其它留意到的事项,或发现本书存在难以理解的内容等时,请具体地
记在本页中交给本公司的销售人员。
使用说明书号
INZ-TN5A0506-C
提交日期 年 月 日
公司名称
使用说明书名称
页 行 内 容
富士电机填写栏 负责人
氧化锆氧变换器通信功能(MODBUS)
使用说明书
请在建议、要求、内容难以理解(内容上有错误、解释不够、用词没有统一性、有错别字、其它)
……………中选一项并画圈“○”
受理 年 月 日 受理编号
提交人
部 门
姓 名
Page 34

ภ᎙ሃჵ
*ቂׁٛ൰಼፩ܿٛ൰ˈขႚכᏇႿኡޚቂངಖ༚ȃ
141-0032ׁ๊ބޕ൰ۅฏܐජᇜݺ಼22ߣ3(Gate City Ohsaki, East Tower)
http://www.fesys.co.jp
ᇨװ᎐ᇔ http://www.fic-net.jp
ຢࣴ༁ᅑฏጓᡷ௸789ૂᇁ࣭ৗࣖ٠29௦B3.C2༃
Te l˖+86-21-5496-2211(Ꮣ)Fax˖+86-21-6417-6672
ף˖200032
http://www.fics.com.cn
ሓٛ൰࡙ட݃ኊሓˈׁᏁல๒૰קˈขȃ
Ꮐᅲᇬˈขቪ༧ডᏪفܿ࢞ཌྷႚྈႼȃ
Printed in China