Profile Measuring System Z500 Manual 2: Operation Manual
Profile Measuring System
Z500
Operation Manual
Authorized Distributor:
Cat. No. Z159-E1-2 Note: Specifications subject to change without notice. Printed in Japan
0103-0.5C
Manual
Cat. No. Z159-E1-2
Page 2
About this Manual
Please read the following manuals carefully and be sure you understand the information provided before attempting to install or operate the Z500.
Manual 1: Setup Manual
Multi-Dimensional Controller
Multi-Dimensional Controller
Z500
Setup Manual
Manual
Cat. No. Z000-E1-1
Cat. No. Z158-E1-2
Manual 2: Operation Manual
Installation
This manual describes the hardware for the Z500
(Multi-Dimensional Controller) and how to install the
components.
Be sure to read this manual first.
Wiring
Multi-Dimensional Controller
Z500
Operation Manual
Manual
Cat. No. Z000-E1-1
Cat. No. Z159-E1-2
Starting the Z500 and Positioning the Workpiece
Adjusting Images
Setting Measurement Conditions and Executing
Outputting to External Devices
SECTION 2 Basic Operations
This section describes how to start the Z500 and how to display images on the monitor.
SECTION 3 Menus for Conversational Menu
• Basic settings required for measurement
• Measurement of consistent workpieces
Menus for Conversational Menu
Settings can be made easily by entering information as requested - just as though you are having a conversation with the Z500.
SECTION 4 Menus for Expert Menu
• Customized settings
• Detailed settings
• Measurement of workpieces placed at different positions
Menus for Expert Menu
More detailed setting, such as position compensation for workpieces placed at different positions, can be made.
SECTION 7 I/O Format
This section provides details on the inputs and outputs used for communications with external devices via terminal blocks or RS-232C.
• Communications settings
• I/O format
• Timing for communications
OMRON Corporation
Industrial Automation Company
Application Sensors Division Sensing Devices and Components Division H.Q. Shiokoji Horikawa, Shimogyo-ku, Kyoto, 600-8530 Japan Tel:(81)75-344-7068 / Fax:(81)75-344-7107
Regional Headquarters
OMRON EUROPE B.V.
Sensor Business Unit Carl-Benz Str. 4, D-71154 Nufringen, Germany Tel:(49)7032-811-0 / Fax:(49)7032-811-199
OMRON ELECTRONICS LLC
1 East Commerce Drive, Schaumburg, IL 60173 U.S.A. Tel:(1)847-843-7900 / Fax:(1)847-843-8568
OMRON ASIA PACIFIC PTE.LTD.
83 Clemenceau Avenue, #11-01, UE Square, 239920 Singapore Tel:(65)835-3011 / Fax:(65)835-2711
OMRON CHINA CO.,LTD.
BEIJING OFFICE Room 1028, Office Building, Beijing Capital Times Square, No.88 West Chang'an Road, Beijing, 100031 China Tel:(86)10-8391-3005 / Fax:(86)10-8391-3688
Page 3
Multi-Dimensional Controller
Z500
Z500-MC10E/MC15E
Operation Manual
INTRODUCTION 2
SECTION 1
Features
SECTION 2
Basic Operations
SECTION 3
Menus for Conversational Menu
SECTION 4
Menus for Expert Menu
INTRODUCTION
SECTION 1 SECTION 2 SECTION 3 SECTION 4 SECTION 5 SECTION 6 SECTION 7 SECTION 8
9
19
39
53
SECTION 5
Other Functions
SECTION 6
System Settings
SECTION 7
I/O Format
SECTION 8
Troubleshooting
109
143
153
183
Manual
1
Page 4
INTRODUCTION
Special or Critical Applications
INTRODUCTION
Special or Critical Applications
When the Z500 will be used in one of the conditions or applications listed below, allow extra safety mar-
gins in ratings and functions, add extra safety feature such as fail-safe systems, and consult your
OMRON representative.
• Operating conditions or environments which are not described in the manual
• Nuclear power control systems, railroad systems, aviation systems, vehicles, combustion systems,
medical equipment, amusement equipment, or safety equipment
• Other systems, machines, and equipment that may have a serious influence on lives and property and
require extra safety features
Product Availability
Some of the products listed may not be available in some countries. Please contact your nearest
OMRON sales office by referring to the addresses provided at the back of this manual.
Regulations and Standards
The Z500 complies with the international regulations and standards below:
• EC Regulations
EMC Directive: No.89/336/EEC
• EN Standards (European Standards)
EN61326:1997 + A1:1998 + A2:2001 (EMI:Class A)
© OMRON, 2003
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form, or by any means, mechanical, electronic, photocopying, recording, or otherwise, without the prior written permission of OMRON. No patent liability is assumed with respect to the use of the information contained herein. Moreover, because OMRON is constantly striving to improve its high-quality products, the information contained in this manual is subject to change without notice. Every precaution has been taken in the preparation of this manual. Nevertheless, OMRON assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of the information contained in this publication.
Z500
Operation Manual
2
Page 5
INTRODUCTION
Table of Contents
INTRODUCTION.................................................................................... 3
INTRODUCTION
Table of Contents......................................................................................................... 3
Precaution on Safety ................................................................................................... 5
Menus............................................................................................................................ 6
Editor's Note................................................................................................................. 8
SECTION 1
Features ............................................................................................... 9
1-1 Features of the Z500............................................................................................10
1-2 Operational Flow.................................................................................................. 14
SECTION 2
Basic Operations................................................................................ 19
2-1 Starting the Z500 and Displaying Images.......................................................... 20
2-2 Menu Operations ................................................................................................ 23
2-2-1 Input Device ...............................................................................................23
2-2-2 Screen Displays......................................................................................... 24
2-2-3 Menu Tree ..................................................................................................25
2-2-4 Inputting Values ........................................................................................ 27
2-2-5 Adjustment Menu ...................................................................................... 28
2-3 Screen Types and Display Methods ..................................................................29
2-3-1 Image Monitor............................................................................................30
2-3-2 Profile Monitor ...........................................................................................33
2-3-3 Digital Monitor ...........................................................................................35
2-3-4 Trend Monitor ............................................................................................ 36
2-4 Saving Settings and Exiting the Z500................................................................ 37
Table of Contents
SECTION 3
Menus for Conversational Menu....................................................... 39
3-1 Setting Measurement Conditions Using the Conversational Menus............... 40
Step
1 Starting Conversational Menus ................................................................40
Step
2 Setting Basic Measurement Conditions ..................................................41
Step
3 Setting Measurement Contents ................................................................48
Step
4 Setting Measurement Conditions .............................................................49
Step
5 Setting Output Contents............................................................................50
Step
6 Starting Measurements .............................................................................51
3-2 Saving Settings and Exiting the Z500................................................................ 52
Operation Manual
Z500
3
Page 6
INTRODUCTION
Table of Contents
INTRODUCTION
SECTION 4
Menus for Expert Menu...................................................................... 53
4-1 Setting Measurement Conditions Using the Expert Menus.............................54
Step
1 Starting Up Expert Menu and Entering Set Mode ..................................54
Step
2 Adjusting Images ....................................................................................... 56
Step
3 Position Compensation .............................................................................64
Step
4 Measurement Settings............................................................................... 70
Step
5 Measurement Conditions ..........................................................................88
Step
6 Output Settings ........................................................................................ 100
Step
7 Performing Measurement........................................................................107
4-2 Saving Settings and Exiting the Z500.............................................................. 108
SECTION 5
Other Functions ............................................................................... 109
5-1 Executing Force-zero ........................................................................................ 110
5-2 Entering Display Mode ......................................................................................112
5-3 Changing the Screen Display ...........................................................................113
5-4 Changing Scenes .............................................................................................. 125
5-5 Entering Tools Mode ......................................................................................... 128
5-6 Testing Measurement Performance (Test) ......................................................129
5-7 Checking the Line Beam Position (Surrounding image) ...............................135
5-8 Backing Up Data to a Computer.......................................................................136
SECTION 6
System Settings ............................................................................... 143
6-1 Entering System Mode ......................................................................................144
6-2 Setting RS-232C Communications Specifications (Comm) .......................... 145
6-3 Environment Settings........................................................................................ 146
6-4 Downloading the Sensor Information to the Controller................................ 148
6-5 Compensating Installation Error (Slant correct).............................................149
6-6 Initializing the Z500 (Initialize) ..........................................................................151
6-7 Checking the System Version (Version) ........................................................ 152
SECTION 7
I/O Format ......................................................................................... 153
7-1 Terminal Blocks ................................................................................................. 154
7-2 RS-232C..............................................................................................................159
7-3 Analog Output....................................................................................................180
SECTION 8
Troubleshooting ............................................................................... 183
8-1 Troubleshooting ................................................................................................184
Index .................................................................................................. 187
4
Z500
Operation Manual
Page 7
INTRODUCTION
Precaution on Safety
OMRON products are manufactured for use according to proper procedures by a qualified operator and only for the purposes described in this manual. The following conventions are used to indicate and classify precautions in this manual. Always heed the information provided with them. Failure to heed precautions can result in injury to people or damage to property.
INTRODUCTION
Precaution on Safety
DANGER
WARNING
CAUTION
WARNING Failure to read and understand the information provided in this manual may result in
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury, or property damage.
personal injury or death, damage to the product, or product failure. Please read
each section in its entirety and be sure you understand the information provided in
the section and related sections before attempting any of the procedures or opera-
tions given.
Operation Manual
Z500
5
Page 8
INTRODUCTION
Menus
INTRODUCTION
Menus
Menus for Conversational Menu
Scn
Set Start setting
P.125
P.40
(0 to 15)
P.40
Run
Tools
System
Save
P.51
P.128
P.144
P.37
Meas method
Measurement region
Sensitivity
Meas set
(OUT0 to OUT7)
Conditions
Output contents
End setting
Test
Surrounding
image
Backup
Comm
Slant correct
Initialize
Version
P.41
P.43
P.44
P.48
P.49
P.50
P.52
P.129
P.135
P.136
P.145
P.146Environment
P.148Download
P.149
P.151
P.152
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
Item and hierarchy that can be set through only the adjustment menu
Sensitivity
Details
Conditions
Output
P.70
P.72
*1
P.70
*1
P.74
*1
P.70
*1
P.70
*1
P.70
*1
P.76
*1
*
1:The same construction
as that for
P.59
P.63
P.88
P.100
When
Define
setting of
Height
has been selected, however,
Calibrate
Change
Calibrate
Output
Delete
is disabled.
# to avg
Trigger
Details
Analog
P.86
P.78
P.84
P.87
P.88
P.90
P.96
P.100
6
Z500
Operation Manual
Display
P.112
Terminals
RS-232C P.105
*2:The hierarchy under
*2
Display Mode varies depending on the type of screen being displayed.
Adjustment Menu p.28
P.103
Page 9
Menus for Expert Menu
INTRODUCTION
INTRODUCTION
Scn
Set
P.125
P.54
(0 to15)
Image
Compensatn
Meas set
(OUT0 to OUT7)
Conditions
P.56
P.64
P.70
P.88
Meas method
Region
Sensitivity
Details
Reference
Compensation
method
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
# to avg
P.56
P.58
P.59
P.63
P.65
P.69
P.70
P.72
P.70
P.74
P.70
P.70
P.70
P.76
P.88
Change
*1
*1
*1
*1
*1
*1
*1
Calibrate
Output
Delete
*1:The same construction
as that for
Define
When selected, however, setting of disabled.
P.86
P.78
P.84
P.87
Height
has been
Calibrate
Menus
is
Run
Tools
System
Save
P.51
P.128
P.144
P.37
Output
Test
Surrounding
image
Backup
Comm
Slant correct
Initialize
Version
P.100
P.129
P.135
P.136
P.145
P.146Environment
P.148Download
P.149
P.151
P.152
Trigger
Details
Analog
Terminals
RS-232C
Item and hierarchy that can be set through only the adjustment menu
Display
P.90
P.96
P.100
P.103
P.105
P.112
*2
*2:The hierarchy under Display Mode
varies depending on the type of screen being displayed.
Adjustment Menu p.28
Operation Manual
Z500
7
Page 10
INTRODUCTION
Editor’s Note
INTRODUCTION
Editor’s Note
Visual Aids
The following headings appear in the left column of the manual to help you locate different types of information.
NOTICE
CHECK
REFERENCE
TwoSensor
HELP Indicates information helpful in operation.
Indicates information required to take full advantage of the functions and performance of the product. Incorrect application methods may result in the loss of damage or damage to the product. Read and follow all precautionary information.
Indicates points that are important in using product functions or in application procedures.
Indicates where to find related information.
Indicates information required when using 2 sensors.
Notation
Screen Messages
In this manual, screen message are given in bold/italic. E.g.: System System/Backup means the Menu is hierarchical.
OMRON Product References
All OMRON products are capitalized in this manual. The world "Unit" is also capitalized when it refers to an OMRON product, regardless of whether or not it appears in the proper name of the product.
8
Z500
Operation Manual
Page 11
Z500
Operation Manual
SECTION 1
Features
SECTION 1 explains the features of the Z500 and the flow of operation using the
Conversational and Expert Menus.
1-1 Features of the Z500 10
1-2 Operational Flow 14
SECTION 1 Features
9
Page 12
SECTION 1
Features
1-1 Features of the Z500
1-1-1 Eight Measurement Items for Various Detection Needs
SECTION 1
Select the measurement item according to application.
Complicated detection is also possible, by combining multiple measurement
items and with detailed settings.
Simultaneous measurement can be performed at 8 points.
1-1 Features of the Z500
REFERENCE
Refer to page 70.
Height
Edge position
Peak/bottom
1-1-2 Simple Setup Using Menus
The Z500 has 2 types of menu. Set the measurement contents according to the
menus displayed on the monitor screen.
Menus for Conversational Menu
Settings can be made easily by entering information as requested - just as
though you are having a conversation with the Z500.
REFERENCE
Refer to page 39.
Menus for Expert Menu
More detailed setting, such as position compensation for workpieces placed at
different positions, can be made.
Step : 2 pts
Width
Define
Step : 3 pts
Edge center
REFERENCE
Refer to page 53.
1-1-3 A Variety of Output Formats
Not only is analog output available, but data output to external devices can also
be performed via terminal blocks or RS-232C.
REFERENCE
Refer to page 100.
Z500
Operation Manual
10
Page 13
1-1-4 Monitoring While Viewing the Measurement Status
The Z500 has 4 types of monitor. Monitoring suited to the application is possible
at every stage from installation and measurement settings to operation and
maintenance.
REFERENCE
Refer to page 29.
SECTION 1
Features
SECTION 1
1-1 Features of the Z500
Image Monitor
This is the main monitor of the Z500. Three pieces of information; measurement status, sensitivity, and measurement result, can be checked together with their target images.
Scn 0 Run
NEAR
OUT 0
BRIGHT
+0005.18800
S+ / :Display S+ENT:Adjust
SHIFT
DARK
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
LV
PEAK
Zero's OFF
PASSmm
SHIFT
FAR
[ .] [ ]
+
Profile Monitor
Chronological change of height distribution of 126 points (profile) can be checked on a shaded 3-D image.
Scn 0 Run
1div x: 1.000 [s] y:+10.0000[mm]
Thru
+0005.18800
Sen 0
S+ / :Display S+ENT:Adjust
mm
9.96ms
PASS
SHIFT
SHIFT
SHIFT
SHIFT
Digital Monitor
Two or more measurement results can be checked at the same time.
Scn 0 Run
+0002.65432mm
0
+
+
PASS
+0004.65432mm
1
PASS
+0000.00000mm
2
ERROR
-----------mm
3
S+ / :Display S+ENT:Adjust
+
SHIFT
9.96ms
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
SHIFT
++
Trend Monitor
Chronological change of measured values can be checked.
Scn 0 Run
+
Thru
+
OUT 0
S+ / :Display S+ENT:Adjust
9.96ms
1div x: 1.000 [s] y:+10.0000[mm]
Zero's OFF
mm+0005.18800
PA S S
Operation Manual
Z500
11
Page 14
SECTION 1
Features
1-1-5 Flexible Settings for Data Output to External Devices
In addition to simply outputting measurement results, equations can be set to
output to external devices using measurement and judgment results.
SECTION 1
1-1 Features of the Z500
1-1-6 A Wide Variety of Useful Functions
Any Type of Workpiece can be Measured by Using the Image Adjustment Function
REFERENCE
Refer to page 70.
Z500
Higher measurement accuracy can be assured by using the sensitivity
adjustment function, even if the workpieces to be measured have different
brightness.
REFERENCE
Refer to pages 44 and 59.
Adjustment Menu
Six types of adjustments are available on shortcut keys by displaying the
adjustment menu.
REFERENCE
Refer to page 28.
Checking Surrounding Images of the Area that the Line Beam Strikes
Displays images of the area surrounding the part that the line beam strikes.
This function is useful when the Sensor is mounted inside an installation and the
measurement area cannot be checked directly.
REFERENCE
Refer to page 135.
Z500
Operation Manual
12
Page 15
Specifying Measurement Timing by Using the Trigger Function
Measurement timing can be specified arbitrarily by using the self-trigger and
external trigger functions. Characteristic quantity can be held while the
measurement is carried out.
(Measurement status at a set timing can be checked on the monitor screen.)
REFERENCE
SECTION 1
Features
SECTION 1
Refer to page 90.
Smooth Setup Changes
Scenes can be used to change between up to 16 types of measurement setups.
Simply change the scene to change to a different measurement setup.
Refer to page 125.
1-1 Features of the Z500
REFERENCE
Operation Manual
Z500
13
Page 16
SECTION 1
Features
1-2 Operational Flow
1-2-1 Menus for Conversational Menu
SECTION 1
1-2 Operational Flow
Turn ON the power.
Displaying Images on the Monitor
Refer to page 20,
Starting the Z500 and Displaying Images.
2-1
Preparations for Detection
Refer to this section to learn about the basic operational flow from setting measurement conditions to executing measurements.
CHECK
Refer to page 39 for information on settings using Conversational Menu.
Starting Up Conversational Menu and Entering Set Mode
Refer to page 40, See
Step 1 : Starting Conversational Menus.
Setting Basic Measurement Conditions
Refer to page 41, See
Step 2 :
Setting Basic Measurement Conditions.
Selecting and Setting Measurement Items
Refer to page 48, See
Step 3 : Setting Measurement Contents.
Refer to page 70 for details.
Screen Types and Display Methods
Refer to page 29.
Selecting Measurement Item Appropriate to the Measurement Goal
Refer to page 70.
Height Step: 2 pts Step: 3 pts Edge position Width Edge center Define
14
Setting Measurement Conditions (if necessary)
Refer to page 49, See
Step 4 :
Setting Measurement Conditions.
Refer to page 88 for details.
Setting Output Contents
Refer to page 50, See
Setting Detection Conditions and Executing
Step 5 : Setting Output Contents.
Refer to page 100 for details.
Starting Measurements
Refer to page 51, See
Step 6 : Starting Measurements.
Performing Test
Refer to page 129,
Testing Measurement Performance (Test).
5-6
Confirming Settings
Z500
Operation Manual
Checking the Line Beam Position
Refer to page 135,
Beam Position (Surrounding image).
5-7Checking the Line
Page 17
Troubleshooting
When an error message has been displayed on the screen,
Refer to page 184,
8-1Troubleshooting.
SECTION 1
Features
SECTION 1
1-2 Operational Flow
Changing Measurement Contents
Refer to page 86.
Clearing Measurement Items
Refer to page 87.
Changing the Screen Display
Refer to page 113.
Changing and Deleting Settings
Saving Detection Conditions
Refer to page 52,
Saving Settings and Exiting the Z500.
3-2
Executing Force-zero
Refer to page 110.
Communications with External Devices
Refer to page 145,
Setting RS-232C Communications Specifications (Comm).
6-2
I/O Format
Refer to page 153.
Setting Conditions by Model Type
Refer to page 125,
Changing Scenes.
5-4
Setting System Environment
Application Setting Operations
Conditions
Refer to page 146,
Environment Settings.
6-3
Initializing the Z500
Refer to page 151.
Saving Settings
Backing up Data
Refer to page 136,
to a Computer (Backup).
5-8 Backing Up Data
Operation Manual
Z500
15
Page 18
SECTION 1
Features
1-2-2 Menus for Expert Menu
SECTION 1
1-2 Operational Flow
Turn ON the power.
Displaying Images on the Monitor
Refer to page 20,
Starting the Z500 and Displaying Images.
2-1
Preparations for Detection
Refer to this section to learn about the basic operational flow from setting measurement conditions to executing measurements.
CHECK
Refer to page 54 for information on settings using Expert Menu.
Starting Up Expert Menu and Entering Set Mode
Refer to page 54,
Expert Menu and Entering Set Mode.
Step 1 : Starting Up
Screen Types and Display Methods
Refer to page 29.
Adjusting Images
Refer to page 56,
Step 2 : Adjusting Images.
Setting Position Compensation
Refer to page 64,
Step 3 : Position Compensation.
(if necessary)
Selecting and Setting Measurement Items
Refer to page 70,
Step 4: Measurement Settings.
Setting Measurement Conditions
Refer to page 88,
Step 5 : Measurement Conditions.
Setting Detection Conditions and Executing
Setting Output Contents
Refer to page 100,
Step 6 : Output Settings.
(if necessary)
Starting Measurements
Refer to page 107,
Step 7 : Performing Measurement.
Selecting Measurement Item Appropriate to the Measurement Goal
Refer to page 70.
Height Step: 2 pts Step: 3 pts Edge position Width Edge center Define
16
Z500
Operation Manual
Page 19
Troubleshooting
When an error message has been displayed on the screen,
Refer to page 184,
8-1Troubleshooting.
SECTION 1
Features
SECTION 1
1-2 Operational Flow
Performing Test
Refer to page 129,
Testing Measurement Performance (Test).
5-6
Checking the Line Beam Position
Refer to page 135,
Confirming Settings
Checking the Line Beam Position (Surrounding image).
5-7
Changing Measurement Contents
Refer to page 86.
Clearing Measurement Items
Refer to page 87.
Changing the Screen Display
Refer to page 113.
Changing and Deleting Settings
Saving Detection Conditions
Refer to page 108,
Saving Settings and Exiting the Z500.
4-2
Executing Force-zero
Refer to page 110.
Communications with External Devices
Refer to page 145,
Setting RS-232C Communications Specifications (Comm).
6-2
I/O Format
Refer to page 153.
Setting Conditions by Model Type
Refer to page 125,
Changing Scenes.
5-4
Setting System Environment
Application Setting Operations
Conditions
Refer to page 146,
Environment Settings.
6-3
Initializing the Z500
Refer to page 151.
Backing up Data
Saving Settings
Refer to page 136,
Backing Up Data to a Computer (Backup).
5-8
Operation Manual
Z500
17
Page 20
SECTION 1
Features
MEMO
SECTION 1
1-2 Operational Flow
18
Z500
Operation Manual
Page 21
Z500
Operation Manual
SECTION 2
Basic Operations
SECTION 2 shows basic menu operations of the Z500.
2-1 Starting the Z500 and Displaying Images 20
2-2 Menu Operations 23
2-2-1 Input Device 23
2-2-2 Screen Displays 24
2-2-3 Menu Tree 25
2-2-4 Inputting Values 27
2-2-5 Adjustment Menu 28
2-3 Screen Types and Display Methods 29
SECTION 2 Basic Operations
2-3-1 Image Monitor 30
2-3-2 Profile Monitor 33
2-3-3 Digital Monitor 35
2-3-4 Trend Monitor 36
2-4 Saving Settings and Exiting the Z500 37
19
Page 22
SECTION 2
Basic Operations
2-1 Starting the Z500 and Displaying Images
This section describes how to start the Z500 and how to display images on the monitor.
Place the workpiece and adjust the Sensor position while monitoring the displayed image.
Starting the Z500
#1
SECTION 2
2-1 Starting the Z500 and Displaying Images
1.
CHECK
2.
3.
Be sure that the basic Z500 components have been connected correctly.
Power Supply
Console
ENT
ESC TRIG
SHIFT
CONSOLE
Sensor
Z500
×× × × × × × × × × × ×
×× × × × × ××
× × × × × × ×
× × × × × × × × × ×
× × ×
× ×
Monitor
Before connecting components or wiring power supply lines and grounding wires, be sure
to refer to the relevant section in the
REFERENCE
Setup Manual
.
Refer to page 16 in the Setup Manual.
Turn ON the power supply to the monitor.
Turn ON the power supply to the Z500.
The Image Monitor will be displayed.
CHECK
Z500
Operation Manual
20
Scn 0 Run
NEAR
OUT 0
+0005.18800
S+ / :Display S+ENT:Adjust
If the Image Monitor is not displayed, use the
SHIFT
on the console to change the Monitors.
REFERENCE
Refer to page 29, Screen Types and Display Methods.
+
Right
Keys or
BRIGHT
DARK
SHIFT
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PA S Smm
Left
+
Keys
FAR
Page 23
Position the Workpiece and Adjust the Image
#2
Position the workpiece to be measured and adjust the displayed image in the
measurement range.
The measurement range varies depending on the Sensor models connected.
REFERENCE
Refer to page 26 in the Setup Manual.
Example: Z500-SW17
Sensor
100mm
SECTION 2
Basic Operations
SECTION 2
+20mm
2-1 Starting the Z500 and Displaying Images
0mm
Height
1.
Measurement range
+20mm
0mm
-20mm
Position
-20mm
Position the workpiece. In this example, the workpiece shown in the right figure will be described.
14.5mm
(±7.25mm)
17.3mm
(±8.65mm)
20.1mm
(±10.05mm)
Operation Manual
Z500
21
Page 24
SECTION 2
Basic Operations
A profile graph showing the level differences in the workpiece will be displayed.
SECTION 2
2-1 Starting the Z500 and Displaying Images
2.
CHECK
Adjust the position of the workpiece if necessary. Adjust the workpiece position if the Profile graph is not appeared in the center of the monitor.
Scn 0 Run
NEAR
BRIGHT
DARK
OUT 0
+0005.18800
Profile
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PA S Smm
FAR
S+ / :Display S+ENT:Adjust
If the lines representing level differences have swollen elliptically, the light density is
excessive. If the lines are too thin to see, the light density is insufficient. In either case,
adjust the sensitivity.
REFERENCE
Refer to page 59 for information on how to adjust sensitivity.
22
Z500
Operation Manual
Page 25
2-2 Menu Operations
2-2-1 Input Device
Menu operations are performed on the Console.
SECTION 2
Basic Operations
Trigger Key
TRIG
Escape Key
ESC
SHIFT Key
SHIFT
Key Function
Escape Key Returns the cursor to the previous menu display or operation.
Trigger Key The effect of the Trigger Key varies with the function.
Enter Key
SHIFT Key
Up, Down, Left, and Right Keys
ESC TRIG
SHIFT
Executes a function or sets a value. Note: On the F160-KP, also functions as a Cursor Key.
Must be pressed in combination with another key to have any effect. Specific functions are assigned to combinations of the Key and other keys for specific screens.
The Up and down and also to set values. Use the Up Key to increase a value by 1.
Left
Down
and
Use the Hold down the Up or a value. The right.
ENT
Z300-KP
CONSOLE
Down
Keys are used to move the cursor up and
Key to decrease a value by 1.
Right
Enter Key
ENT
Up, Down, Left, and Right Keys
Down
Key to quickly increase or decrease
Keys are used to move the cursor left or
SHIFT
SECTION 2
2-2 Menu Operations
Operation Manual
Z500
23
Page 26
SECTION 2
2-2 Menu Operations
SECTION 2
Basic Operations
2-2-2 Screen Displays
The Z500 is operated by selecting functions displayed on the screen. Familiarize yourself with each
function before operating the Z500.
Scene Number
There are 16 scenes. Setting different measurement conditions for each one allows easy switching between different setups.
Cursor Keys
The cursor is moved to the desired function by pressing the Up, Down, Left, and Right Keys.
Scn 0 Run
S+ / :Display S+ENT:Adjust
Key Operations
Displays special key combinations at the bottom of the screen where available. S refers to the SHIFT Key. "S + ENT" indicates that the ENT Key should be pressed while the SHIFT Key is pressed.
Mode (See the table below)
The current operating mode is displayed.
9.96ms
Measurement period (Sampling time)
It varies depending on the item to be measured.
Mode
Display Description
Set Mode to set the inspection conditions.
Performs measurement.
Run
Tools Used to save settings and images to a computer as backup.
System
Save
The measurement results are output to an external device via terminal block or RS-233C.
Used to set system conditions for the Z500. Select this mode to switch menus and settings for communications with external devices.
Used to save data to flash memory in the Z500. If new settings have been made, be sure to save the data before quit­ting.
24
Z500
Operation Manual
Page 27
SECTION 2
Basic Operations
2-2-3 Menu Tree
Z500 menus are hierarchical. The cursor is moved to the required functions to set measurement
conditions. Use the following procedures to move around the menu tree.
1.
2.
In this example, the cursor will move to the Image Menu.
3.
Move the cursor to the item to be moved. (Move the cursor to the desired function in Set Mode using the Up and Down Keys.)
Press the ENT Key.
Repeat (1.) and (2.) to move the cursor to lower levels.
Press the ESC Key once to move to the upper level.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Scn 0 Set
Image
Compensatn Meas set Conditions Output
SECTION 2
2-2 Menu Operations
Meas method
Region Sensitivity Details
Operation Manual
Z500
25
Page 28
SECTION 2
2-2 Menu Operations
SECTION 2
Basic Operations
Triangle Mark
Items with an inverted triangle after them have a list of alternatives.The method for selecting the
alternatives is given here.
Scn 0 System
:
1.
2.
Move the cursor to the item to be set. In this example, move the cursor to Baud rate.
Press the ENT Key.
Baud rate Data Length Parity Stop bits Delimiter Flow
38400bps
: : : : :
End
8bit
None
1bit
CR
None
The selections will be displayed.
3.
Move the cursor to the desired baud rate using the Up and Down Keys.
4.
Press the ENT Key.
The selections will be registered.
Scn 0 System
Baud rate Data Length Parity Stop bits Delimiter Flow
Scn 0 System
Baud rate Data Length Parity Stop bits Delimiter Flow
:
38400bps
:
2400bps
:
4800bps
:
9600bps
:
19200bps
:
38400bps
End
:
38400bps
: : : : :
End
8bit
None
1bit
CR
None
8bit
None
1bit
CR
None
26
Z500
Operation Manual
Page 29
SECTION 2
Basic Operations
2-2-4 Inputting Values
The section explains how to input values when setting judgment or measurement conditions.
1.
Move the cursor to the item for which a value is to be changed.
2.
Press the ENT Key.
The cursor will change to a cursor the size of a single digit.
3.
Move the cursor to the digit to be changed using the Left and Right Keys.
4.
Change the value. Use the Up Key to increase the value. Use the Down Key to decrease the value.
5.
6.
Repeat (4.) to change other values.
Press the ENT Key.
Scn 0 Set
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
Scn 0 Set
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
Scn 0 Set
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
:
[ 0.01000]
:
[ 0.00000] : :
50.0%
:
Clamp
: : :
End
:
[0000.01000] :
[ 0.00000] : :
50.0%
:
Clamp
: : :
End
:
[0000.02000] :
[ 0.00000] : :
50.0%
:
Clamp
: : :
End
None
2pixel
None
5pixel
None
2pixel
None
5pixel
None
2pixel
None
5pixel
SECTION 2
2-2 Menu Operations
The values will be set.
Scn 0 Set
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
:
[ 0.02000]
:
[ 0.00000] : : : : : :
End
Operation Manual
None
2pixel
50.0% Clamp
None
5pixel
Z500
27
Page 30
SECTION 2
Basic Operations
2-2-5 Adjustment Menu
Press the SHIFT + ENT Keys in Run Mode. The adjustment menu will be displayed.
The adjustment menu is useful to display the set screens of the following six items without passing
through the menu hierarchies.
SECTION 2
2-2 Menu Operations
CHECK
Scn 0 Run
ESC TRG ENT
Display
Sensitvy
Output
Conditns
ESC:Back S+ / :Switch display
Key Allocation Reference
Trigger Key Display P.112
Enter Key Test P.129
Up Key Sensitivity P.59
Down Key Conditions P.88
Left Key Output P.100
Right Key Meas set P.70
Escape Key None -
The adjustment menu can be performed from the Image Monitor, the Digital Monitor, the
Profile Monitor, or the Trend Monitor.
9.96ms
Test
Meas set
28
Z500
Operation Manual
Page 31
SECTION 2
Basic Operations
2-3 Screen Types and Display Methods
There are 4 Monitors to display images on the Z500; Image Monitor, Digital Monitor, Trend Monitor, and
Profile Monitor. These monitors are useful to check measurement information as required from various
viewpoints. Use the SHIFT + Right Keys or SHIFT + Left Keys on the console to change the Monitors.
The first time the power supply is turned ON after delivery, the Image Monitor is displayed.
Image Monitor
This is the main monitor of the Z500. Three pieces of information; measurement status, sensitivity, and measurement result, can be checked together with their target images.
Scn 0 Run
NEAR
OUT 0
BRIGHT
+0005.18800
S+ / :Display S+ENT:Adjust
SHIFT
DARK
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
LV
PEAK
Zero's OFF
PASSmm
SHIFT
FAR
[ .] [ ]
+
Profile Monitor
Chronological change of height distribution of 126 points (profile) can be checked on a shaded 3-D image.
Scn 0 Run
1div x: 1.000 [s]
Thru
Sen 0
S+ / :Display S+ENT:Adjust
y:+10.0000[mm]
+0005.18800
mm
9.96ms
PASS
SHIFT
SHIFT
SHIFT
SHIFT
Digital Monitor
Two or more measurement results can be checked at the same time.
Scn 0 Run
+0002.65432mm
0
+
+
PASS
+0004.65432mm
1
PASS
+0000.00000mm
2
ERROR
-----------mm
3
S+ / :Display S+ENT:Adjust
+
SHIFT
9.96ms
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
SHIFT
++
SECTION 2
2-3 Screen Types and Display Methods
Trend Monitor
Chronological change of measured values can be checked.
Scn 0 Run
+
Thru
+
OUT 0
S+ / :Display S+ENT:Adjust
9.96ms
1div x: 1.000 [s] y:+10.0000[mm]
Zero's OFF
mm+0005.18800
PA S S
CHECK
CHECK
Enable to select which Monitor will be displayed when the Z500 turns ON.
REFERENCE
Refer to page 146.
Measurement continues even while the screen is switched.
Operation Manual
Z500
29
Page 32
SECTION 2
Basic Operations
2-3-1 Image Monitor
This is the main monitor of the Z500. The first time the power supply is turned ON after delivery, this
screen is displayed. Mostly used to check the measurement status, sensitivity information, and
measurement results.
Image Monitor consists of 2 displays; Measurement status display and Measurement sensitivity display.
Use the TRIG Key to change the displays.
SECTION 2
2-3 Screen Types and Display Methods
Measurement Status Display
Scn 0 Run
NEAR
OUT 0
BRIGHT
DARK
Sen0
Sen1
Zero's OFF
+0005.18800
S+ / :Display S+ENT:Adjust
Note:
Measurement Status Display
Common to Measurement status display and Measurement sensitivity display.
The height information within the measurement range can be checked.
The height information between 0 and 125 lines is displayed with the Profile
graph.
Sensor
Line beam
9.96ms
LV
PEAK
LV
PEAK
PA S Smm
[ 15] [ 170]
Measurement Sensitivity Display
FAR
Screen Displays
NEAR
BRIGHT
TRIG
[ .]
Scn 0 Run
[ ]
Measurement
Result Display
(See note)
OUT 0
+0005.18800
DARK
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PA S Smm
FAR
S+ / :Display S+ENT:Adjust
Workpiece
Measurement range
Measurement center
30
Z500
Operation Manual
Line 0
Position
Line 125
Height
Scn 0 Run
BRIGHT
DARK
OUT 0
+0005.18800
9.96ms
Sen0
LV
PEAK
Sen1
LV
PEAK
Zero's OFF
PA S Smm
S+ / :Display S+ENT:Adjust
LowHigh
Measurement region (green)
Profile (yellow)
FARNEAR
[ 15]
[ 170]
[ .] [ ]
Page 33
SECTION 2
Basic Operations
CHECK
The Profile graph can be enlarged.
REFERENCE
Refer to page 113.
CHECK
The Profile graph can be output as a measurement value to an external device.
REFERENCE
Refer to page 101.
CHECK
Whether or not the lines interfering with the measurement should be interpolated can be
determined.
REFERENCE
Refer to page 99.
CHECK
The measurement region can be changed.
REFERENCE
Refer to page 58.
Measurement Sensitivity Display
The Measurement sensitivity level (LV) and received level (PEAK) can be
checked. The light density and the status of the workpiece surface at a cursor
position are displayed with the line brightness graph.
SECTION 2
2-3 Screen Types and Display Methods
Bright
Brightness
Dark
CHECK
Cursor (Red)
Refer to page 116 for information on how to change cursor position.
High
Scn 0 Run
OUT 0
Height
BRIGHT
DARK
+0005.18800
S+ / :Display S+ENT:Adjust
Line Brightness (Brightness Distribution)
The line brightness graph at the cursor position is displayed. It is possible to know at a glance whether or not, for example, the light density is insufficient, or is at the saturation level.
Low
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
LV
PEAK
Zero's OFF
PASSmm
Measurement period (Sampling time)
Line beam image The shape of workpieces and
FARNEAR
[ .] [ ]
the way they are detected can be checked.
LV [ 15] PEAK[170]
LV: Displays current sensitivity level. (1 to 31) The value increases with increase in darkness of the workpiece, while it decreases with increase in brightness of the workpiece. PEAK: Displays current received level. Use this value as reference when setting the sensitivity level. 120 max.: Insufficient light density 255 min.: Excessive light density
For details on sensitivity adjustment method, refer to page 59.
The line brightness graph for the specified range can be enlarged.
REFERENCE
Refer to page 113.
Operation Manual
Z500
31
Page 34
SECTION 2
2-3 Screen Types and Display Methods
SECTION 2
Basic Operations
Measurement Result Display
Check whether the measurements are correctly performed.
The measurement points and the measurement values are displayed.
Scn 0 Run
BRIGHT
DARK
OUT 0
+0005.18800
S+ / :Display S+ENT:Adjust
Measurement value The measurement value is displayed in green if the judgment result is PASS. It is displayed in red if the judgment result is HIGH, LOW, or ERROR.
Output number The output number for the currently displayed measurement result is displayed.
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PASSmm
Measurement points (red)
The measured points are displayed. No point is displayed when the measurement is in error. The level and range necessary for the determination of
FARNEAR
measurement points are represented by thin lines. (In this example, the distance between the 2 points is measured.)
Forced-zero
Zero's OFF: Forced-zero not executed Zero's ON: Forced-zero executed
Judgment results
HIGH : Greater than the judgment upper limit PASS : Within the judgment limits LOW : Less than the judgment lower limit ERROR : Measurement error
-The workpiece is out of the measurement range
-The light density is insufficient
-Information on the measurement point
cannot be obtained, or other cases RESET: No-measurement state LD-OFF: LD-OFF input
CHECK
Offset can be set for Force-zero.
REFERENCE
Refer to page 84.
32
Z500
Operation Manual
Page 35
SECTION 2
Basic Operations
2-3-2 Profile Monitor
Use to check the status of the workpiece surface.
Surface condition of a workpiece can be monitored on a shaded 3-D image screen. Height information
and its chronological change can also be monitored.
9.96ms
PA S S
Profile Display
Measurement Result Display
Trend Display
Shaded 3-D image display
Scn 0 Run
1div x: 1.000 [s] y:+10.0000[mm]
Thru
Sen 0
S+ / :Display S+ENT:Adjust
+0005.18800
mm
By moving the workpiece, the profile graph (height information for 126 points) can be monitored in
chronological order for up to 10 seconds.
TRIG Key can be used for switching from waveform display to waveform display stop and vise versa.
The trigger function enables shaded 3-D image to be displayed at specified timing.
Line beam
Movement orientation
SECTION 2
2-3 Screen Types and Display Methods
CHECK
Display range and display time can be changed.
REFERENCE
Refer to page 121.
Operation Manual
Z500
33
Page 36
SECTION 2
Basic Operations
Shaded 3-D Image Display and Profile Display
Surface condition of a workpiece is displayed by a shaded 3-D image with 0 to
255 gradations. A low section is displayed darkly, while a higher section is
displayed brightly.
SECTION 2
2-3 Screen Types and Display Methods
Shaded 3-D Image
Scn 0 Run
Thru
Sen 0
S+ / :Display S+ENT:Adjust
Profile Cursor (Red)
1div x: 1.000 [s] y:+10.0000[mm]
+0005.18800
mm
9.96ms
PA S S
Waveform display status
During waveform display, (green) is displayed. While the waveform is stopped, (red) is displayed.
Trend Display and Measurement Result Display
Trend Cursor
Trend Display
The waveform displayed in chronological
order at the Trend cursor position is displayed.
High
Height
Low
Time(s)
Screen Mode
Thru: continuous display
Trig: displayed in trigger mode
Sensor number
The current sensor number is displayed.
Scn 0 Run
Thru
Sen 0
S+ / :Display S+ENT:Adjust
+0005.18800
Profile Display
The height information at the Profile cursor position is displayed.
Line 0
Position
Line 125
Height
1div x: 1.000 [s] y:+10.0000[mm]
Measurement value
High
mm
Low
9.96ms
PA S S
Conversion values for X and Y axes
The units corresponding to 1 division (1 div) on the waveform display are displayed. If the setting is changed, the display also changes. Use these values as reference when displaying the waveform.
y
1div
x
Judgment result
1div x: 1.000 y:+10.000
34
CHECK
Z500
Operation Manual
The cursor position can be changed.
REFERENCE
Refer to page 121.
Page 37
SECTION 2
Basic Operations
2-3-3 Digital Monitor
Mostly used during line operation.
Two or more measurement results that have been set for the output numbers of the displayed scenes will
be displayed.
Output number
Press the
Key to switch the number.
TRIG
Scn 0 Run
+0002.65432mm
0
PASS
+0004.65432mm
1
PASS
+0000.00000mm
2
ERROR
-----------mm
3
S+ / :Display S+ENT:Adjust
Scn 0 Run
+0002.65432mm
4
PASS
+0004.65432mm
5
PASS
+0000.00000mm
6
ERROR
-----------mm
7
S+ / :Display S+ENT:Adjust
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
HI=+0010.00000 LO=-0010.00000
9.96ms
TRIGTRIG
9.96ms
Judgment value
The judgment value for each
output number will be displayed.
HI: Judgment upper limit
LO: Judgment lower limit
Measurement value
Judgment result
---- will be displayed for output numbers that have not been set.
SECTION 2
2-3 Screen Types and Display Methods
CHECK
The measurement items can also be displayed instead of the judgment values.
REFERENCE
Refer to page 117.
Scn 0 Run
0
+0002.65432mm PASS
1
+0004.65432mm PASS
2
+0000.00000mm ERROR
3
-----------mm
S+ / :Display S+ENT:Adjust
9.96ms
Step: 2 pts
Width
Define
Measurement items
Operation Manual
Z500
35
Page 38
SECTION 2
2-3 Screen Types and Display Methods
SECTION 2
Basic Operations
2-3-4 Trend Monitor
Mostly used for setting and checking the Trigger function.
By using the Trend Monitor, the measurement results to be monitored can be checked in chronological
order for up to 30 seconds. Press the TRIG Key to switch from waveform display to waveform display
stop and vise versa. The trigger function enables waveform to be displayed at specified timing.
Conversion values for X and Y axes
Scn 0 Run
1div x: 1.000 [s] y:+10.0000[mm]
Thru
OUT 0
S+ / :Display S+ENT:Adjust
Measurement value
9.96ms
Zero's OFF
mm+0005.18800
PASS
Judgment result
The units corresponding to 1 division (1 div) on the waveform display are displayed. If the setting is changed, the display also changes. Use these values as reference when displaying the waveform.
y
1div
x
Measurement period (Sampling time)
Waveform display status
During waveform display, (green) is displayed.
While the waveform is stopped, (red) is displayed.
Trigger waveform
The trigger status is displayed.
Force-zero Zero's OFF: Force-zero not executed Zero's ON: Force-zero executed
1div x: 1.000 y:+10.000
36
Output number The current output number is displayed.
Screen Mode Thru: The waveform obtained by the Sensor is displayed continuously. Trig: A still image of the waveform based on the trigger settings is displayed.
Thin lines are used to represent the measurement start and measurement end timings.
CHECK
Display range and display time can be changed. (Only the display range can be changed
for each output number.)
REFERENCE
Refer to page 118.
CHECK
Characteristic quantities (such as average value and peak value being held) at specified
timing can be output.
REFERENCE
Refer to page 84.
Z500
Operation Manual
Page 39
SECTION 2
Basic Operations
2-4 Saving Settings and Exiting the Z500
When setting have been changed, be sure to save to flash memory before turning the power OFF.
NOTICE
1.
2.
A list of modes will be displayed.
3.
Flash memory data is loaded each time the Z500 is started up. Therefore, when setting
have been changed, be sure to save to flash memory before turning the power OFF. If the
power is turned OFF without saving, all of the setting changes will be lost.
Display the screen for Run Mode.
Move the cursor to Run and press the ENT Key.
Select Save.
SECTION 2
Scn 0 Run
2-4 Saving Settings and Exiting the Z500
Scn 0 Run
Set Run Tools System
Save
A confirmation message will be displayed.
4.
NOTICE
When the saving has been completed, the screen in (3.) will return.
5.
Select Execute.
Do not turn OFF the power or input a RESET signal while a message is being displayed in
any save or load operation. Data in memory will be destroyed, and the Z500 may not
operate correctly the next time it is started.
Turn OFF the power supply and exit the Z500.
The data will be saved.
OK?
Execute Cancel
Operation Manual
Z500
37
Page 40
SECTION 2
2-4 Saving Settings and Exiting the Z500
SECTION 2
Basic Operations
MEMO
38
Z500
Operation Manual
Page 41
Z500
Operation Manual
SECTION 3
Menus for Conversational Menu
SECTION 3 explains how to set measurement conditions using the Conversational
Menu.
3-1
Setting Measurement Conditions Using the Conversational Menus
1 Starting Conversational Menus 40
Step
2 Setting Basic Measurement Conditions 41
Step
3 Setting Measurement Contents 48
Step
4 Setting Measurement Conditions 49
Step
5 Setting Output Contents 50
Step
6 Starting Measurements 51
Step
3-2 Saving Settings and Exiting the Z500 52
SECTION 3 Menus for Conversational Menu
40
39
Page 42
SECTION 3
Menus for Conversational Menu
3-1
Setting Measurement Conditions Using the Conversational Menus
This section explains the procedures for using Conversational Menus, from setting measurement
conditions through to starting measurement.
The operating guides displayed on the screen can be used just like a conversation to help perform
procedures.
Step
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
The mode selections will be displayed.
The Z500 will enter Set Mode.
Starting Conversational Menus
1
1.
The screen shown right will be displayed when the Conversational Menus are started for the first time.
2.
Move the cursor to Run and press the
ENT Key.
3.
Use the Up and Down Keys to move the cursor and select Set.
4.
Select Next to start making the settings.
Scn 0 Run
BRIGHT
DARK
OUT 0
9.96ms
Sen0
LV
PEAK
Sen1
LV
PEAK
Zero's OFF
+0005.18800
S+ / :Display S+ENT:Adjust
Scn 0 Run
Set
Run Tools System
Save
Perform settings in the following order.
STEP 1:Measurement method STEP 2:Measurement region STEP 3:Sensitivty adj method
Next
Back
[ 15] [ 170]
PASSmm
FARNEAR
[ .] [ ]
Follow the instructions given by the operation guide displayed on the screen for the rest of the procedure.
Two Sensor
The selections for the sensors to be used for measurement will be displayed.
Select the Sensor number.
Z500
Operation Manual
40
Page 43
SECTION 3
Menus for Conversational Menu
Step
Basic settings required for measurement are described below.
#1
Specify the region to be measured.
Select Zoom to assure higher position measurement accuracy.
Normally, use with the default setting (Wide).
Setting Basic Measurement Conditions
2
#1
Page 41
Select the Measurement Method
#2
Select the Measurement Method
Selection Description
Wide*
Zoom
Measures the whole measurement range of the connected sensor.
Enlarges and measures a part of the measurement range of the connected sensor. Select this function to assure higher position measurement accuracy.
Page 43
Set the Measurement Region
#3
Page 44
Select the Sensitivity Adjustment Method
The asterisk (*) indicates the default setting.
SECTION 3
Step
1
Step
2
Step
3
Operation Manual6
Z500
Step
4
Step
5
Step
6
41
Page 44
SECTION 3
d
Menus for Conversational Menu
Measurement range
Example: Z500-SW17
+20mm
Wide
+20mm
Zoom
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
CHECK
1.
2.
0mm
-20mm
The measurement range varies depending on the Sensor models connected.
REFERENCE
14.5mm
( 7.25mm)
17.3mm
( 8.65mm)
20.1mm
( 10.05mm)
0mm
-20mm
4.8mm
5.7mm
6.7mm
This range is enlarged three times and displaye on the monitor screen.
Refer to page 28 in the Setup Manual.
Select the measurement method while monitoring the image to be measured.
Scn 0 Set
Wide
Zoom
Press the ENT Key.
Method: Select Zoom to increase measurement accuracy.
ENT:Next ESC:Back
Go to
#2
Setting the Measurement Region.
.
Go to #2. Setting the Measurement Region.
Z500
Operation Manual
42
Page 45
SECTION 3
Menus for Conversational Menu
#2
Setting the Measurement Region
Specify the region to be measured.
The Z500 adjusts sensitivity and performs measurements in a region that has been previously set.
Though the whole image can be used for the measurement region, measurement accuracy can be
raised by specifying a restricted measurement region.
Position the workpiece correctly and specify the measurement region.
CHECK
Correct measurement may not be possible at the boundaries of the measurement regions.
Leave space at the boundaries when setting the regions.
Profile
1. Position the workpiece correctly while
monitoring the profile graph.
SECTION 3
2.
The screen for specifying the region will be displayed.
3.
The cursor will be moved to the ending line.
4.
Press the ENT Key.
Specify the starting line of the measurement region.
Repeat the procedure described in (3.) to specify an ending line and press the ENT Key.
Measurement image displayed. Check workpiece position.
ENT:Next ESC:Back
Cursor
Sen0's measurement region Strt+0023.597mm Stp -0023.983mm Strt 0lines Stp 125lines ENT:Next ESC:Back
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Go to #3. Select the Sensitivity Adjustment Method.
Operation Manual6
Z500
43
Page 46
SECTION 3
)
ght
y
e
ow
Menus for Conversational Menu
#3
Select the sensitivity adjustment method for the measurement range.
If the measurement results fluctuate, change the adjustment method.
Normally, these conditions can be left at the default settings.
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
Select the Sensitivity Adjustment Method
Sensitivity Adjustment
Selection
Adjust the sensitivity according to sensitivity information in the measurement region.
The sensitivity can be set to one of 31 levels (Setting range: 0 to
31). Select when it is difficult to assure accurate measurement in Auto. Such cases may appear on a production line where workpieces of different colors are processed.
Adjust the sensitivity for each line in the measurement region. Select when it is difficult to assure accurate measurement in Auto due to a widely distributed brightness of the workpiece surfaces. However, sampling time will be longer. Position a reference workpiece and execute Multi. The Z500 will set automatically the upper and lower limits of sensitivity on the basis of the brightness of the reference workpiece.
CHECK
Auto*
Fixed
Multi
Once the sensitivity has been set in Multi, Sensitvy can be performed only through the
adjustment menu.
Description
The asterisk (*) indicates the default setting.
REFERENCE
Refer to page 28 for details on the adjustment menu, and page 63 for details on
Details.
Sensitivity level (Displayed only when Fixed is selected from the Sensitivity Menu)
Setting range
LV0 (See note
LV31 High
Note: The laser will turn OFF when selecting [LV0].
Sensitivit
L
Color of workpiec
Bri
Dark
44
Z500
Operation Manual
Page 47
When selecting Auto
SECTION 3
Menus for Conversational Menu
1.
A confirmation message will be displayed.
2.
Select Auto.
Select Execute when the basic settings have completed. Select Settings when continuing the measurement contents settings.
Step 3.
Go to
Scn 0 Set
Auto
Fixed Multi
Sen0's sensitivity method: Select Fixed or Multi if measurement unstable with Auto.
ENT:Set ESC:Back
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute Cancel Settings
SECTION 3
Step
1
Step
2
Step
3
Operation Manual6
Z500
Step
4
Step
5
Step
6
45
Page 48
SECTION 3
Menus for Conversational Menu
When selecting Fixed
SECTION 3
The screen for setting the sensitivity level will be displayed.
3-1 Setting Measurement Conditions Using the Conversational Menus
1.
2.
Select Fixed.
Select the level using the Up and Down Keys while monitoring the image on the monitor. The level displayed first is the level selected by the Z500 for the workpiece being measured. Adjust the level while monitoring the line brightness graph and the received level on the monitor. 120 max.: Insufficient light density 255 min.: Excessive light density
Scn 0 Set
Auto
Fixed
Multi
Sen0's sensitivity method: Select Fixed or Multi if measurement unstable with Auto.
ENT:Set ESC:Back
BRIGHT
DARK
Received level
Sen0
LV
PEAK
Sen1
LV
PEAK
Line Brightness
LV 15
Sen0's sensitivity method: Select Fixed or Multi if measurement unstable with Auto.
ENT:Set ESC:Back
[ 15] [ 170]
FARNEAR
[ .] [ ]
3.
Press the ENT Key.
A confirmation message will be displayed.
4.
Select Execute when the basic settings have completed. Select Settings when continuing the measurement contents settings.
Go to
Z500
Operation Manual
46
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute Cancel Settings
Step 3.
Page 49
When selecting Multi
SECTION 3
Menus for Conversational Menu
1.
The sensitivity multiple setting screen will be displayed.
2.
3.
Select Multi.
Check whether or not the workpiece is positioned inside the measurement range.
Press the ENT Key.
Scn 0 Set
Auto Fixed
Multi
Sen0's sensitivity method: Select Fixed or Multi if measurement unstable with Auto.
ENT:Set ESC:Back
Line beam image
Scn 0 Set
Sensitivity set for each line. Set the workpiece correctly and press the ENT Key.
ENT:Set ESC:Back
SECTION 3
Step
1
Step
2
Step
3
A confirmation message will be displayed.
4.
Select Execute when the basic settings have completed. Select Settings when continuing the measurement contents settings.
Step 3.
Go to
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute Cancel Settings
Operation Manual6
Step
4
Step
5
Step
6
Z500
47
Page 50
SECTION 3
Menus for Conversational Menu
Step
The Z500 provides eight measurement items. Up to eight measurement items can be set for each scene.
Select the measurement item according to application.
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
3
Setting Measurement Contents
This section gives an outline only. Refer to SECTION 4 (page 70) for details on
measurement contents.
REFERENCE
Refer to page 70.
48
Z500
Operation Manual
Page 51
Menus for Conversational Menu
4
Step
Set the measurement conditions, such as average number of times and trigger setting.
Setting can be performed only through the adjustment menu.
Setting Measurement Conditions
REFERENCE
Refer to page 28, Adjustment Menu.
This section gives an outline only. Refer to SECTION 4 (page 88) for details on
measurement conditions.
REFERENCE
SECTION 3
Refer to page 88.
SECTION 3
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Operation Manual6
Z500
49
Page 52
SECTION 3
Menus for Conversational Menu
Step
The Z500 can transmit data to an external device in three formats; analog, terminal, and RS-232C. The
operation procedures for setting the output contents to the external device will be described in SECTION
4. Setting can be performed only through the adjustment menu.
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
5
Setting Output Contents
REFERENCE
Refer to page 28, Adjustment Menu.
This section gives an outline only. Refer to SECTION 4 (page 100) for details on
output contents.
REFERENCE
Refer to page 100.
50
Z500
Operation Manual
Page 53
SECTION 3
Menus for Conversational Menu
Step 6
Starting Measurements
Perform the measurement under the conditions that have been previously set.
Use the following procedure to execute measurement according to the conditions set for the scene
currently displayed.
The data will be output to external devices as well.
1. Confirm that Run
If Run
is not selected, press the ESC Key several times to switch the
is selected.
selection.
Measurement will be executed and the measurement results will
be displayed on the screen. Check that measurement has been
performed correctly.
The output number for the currently displayed measurement result is displayed.
Scn 0 Run
BRIGHT
DARK
OUT 0
9.96ms
Sen0
LV
PEAK
Sen1
LV
PEAK
Zero's OFF
+0005.18800
S+ / :Display S+ENT:Adjust
Judgment resultMeasurement value
[ 15]
[ 170]
PASSmm
SECTION 3
FARNEAR
Step
[ .] [ ]
1
Step
2
Step
3
CHECK
Conditions can be changed simply without leaving Run Mode using the
adjustment menu.
Measurement will be interrupted while adjustments are made.
REFERENCE
Refer to page 28.
CHECK Change the Monitors using the SHIFT + Right Keys or SHIFT + Left Keys if
necessary.
REFERENCE
Refer to page 29.
CHECK For measurements through the execution of Force-zero
REFERENCE
Refer to page 110.
Z500
Operation Manual6
Step
4
Step
5
Step
6
51
Page 54
SECTION 3
Menus for Conversational Menu
3-2 Saving Settings and Exiting the Z500
When settings have been completed, be sure to save to flash memory before turning the power OFF.
Refer to SECTION 2 (page 37) for details on saving settings.
REFERENCE
Refer to page 37.
SECTION 3
3-2 Saving Settings and Exiting the Z500
NOTICE
Flash memory data is loaded each time the Z500 is started up. Therefore, when
setting have been changed, be sure to save to flash memory before turning the
power OFF. If the power is turned OFF without saving, all of the setting changes
will be lost.
52
Z500
Operation Manual
Page 55
Z500
Operation Manual
SECTION 4
Menus for Expert Menu
This section explains how to set measurement conditions using the Expert Menu.
4-1 Setting Measurement Conditions Using the Expert Menus 54
1 Starting Up Expert Menu and Entering Set Mode 54
Step
2 Adjusting Images 56
Step
3 Position Compensation 64
Step
4 Measurement Settings 70
Step
5 Measurement Conditions 88
Step
6 Output Settings 100
Step
7 Performing Measurement 107
Step
SECTION 4 Menus for Expert Menu
4-2 Saving Settings and Exiting the Z500 108
53
Page 56
SECTION 4
Menus for Expert Menu
4-1 Setting Measurement Conditions Using the Expert Menus
This section explains how to set measurement conditions using the Expert Menus and the basic steps
before starting measurements.
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
Step 1
Starting Up Expert Menu and Entering Set Mode
To set the measurement conditions, it is first necessary to enter Set Mode.
1.
Move the cursor to Run and press the
Scn 0 Run
ENT Key.
A list of modes will be displayed.
2.
Select System.
The selections will be displayed.
3.
Select Environment.
Scn 0 Run
Scn 0 System
Comm
Environment
Slant correct Initialize Version
Set Run Tools
System
Save
The screen for setting environment will be displayed.
4.
Move the cursor to Conversatn and press the ENT Key.
Z500
Operation Manual
54
Scn 0 System
Settings Sensor Scene Monitor 1st scene Dec. pts
:
Conversatn
:
Mirror
:
Console
:
Image
:
Scn 0
:
5 digits
End
Page 57
SECTION 4
Menus for Expert Menu
The selections will be displayed.
5.
6.
The setting will be registered and the screen in (3.) will return.
7.
The screen in (2.) will return.
8.
Move the cursor to Expert and press the ENT Key.
Select End.
Press the ESC Key.
Select Set.
Scn 0 System
Settings Sensor Scene Monitor 1st scene Dec. pts
Scn 0 System
Set
Run Tools System
Save
:
Conversatn
Conversatn
: :
Console
: : :
5 digits
End
Mirror
Expert
Image
Scn 0
SECTION 4
Step
1
The selections available with the Expert Menu's Set Mode will be
displayed.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
55
Page 58
SECTION 4
Menus for Expert Menu
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
Step 2
Adjusting Images
To facilitate accurate measurements, images are adjusted for clearer display of the section to be
measured.
Measurement Methods
Specify the region to be measured.
Select Zoom to assure higher position measurement accuracy.
Normally, use with the default setting (Wide).
Selection Description
Measures the whole measurement range of the connected sensor.
Enlarges and measures a part of the measurement range of the connected sensor. Select this function to assure higher position measurement accuracy.
The asterisk (*) indicates the default setting.
Wide
+20mm
Zoom
CHECK
Measurement Range
Wide*
Zoom
When the measurement method is changed, all settings for scene data wil be cleared.
Example: Z500-SW17
+20mm
56
CHECK
Z500
Operation Manual
0mm
-20mm
The measurement range varies depending on the Sensor models connected.
REFERENCE
14.5mm
( 7.25mm)
17.3mm
( 8.65mm)
20.1mm
( 10.05mm)
0mm
-20mm
4.8mm
5.7mm
6.7mm
This range is enlarged three times and displayed on the monitor screen.
Refer to page 28 in the Setup Manual.
Page 59
SECTION 4
Menus for Expert Menu
1.
The measurement method selections will be displayed.
Two Sensor
2.
3.
Select Image/Meas Method.
The selections for the sensors to be used for
measurement will be displayed.
Select the Sensor number.
Go to (2.).
Select the measurement method while monitoring the image to be adjusted.
Press the ENT Key.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Scn 0 Set
Wide
Zoom
Meas method
Region Sensitivity Details
SECTION 4
The setting will be registered and the screen in (1.) will return.
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
57
Page 60
SECTION 4
Menus for Expert Menu
Setting Measurement Region
Specify the region to be measured. The Z500 adjusts sensitivity and performs measurements in a region that has been previously set. Though the whole image can be used for the measurement region, measurement accuracy can be raised by specifying a restricted measurement region.
CHECK
Correct measurement may not be possible at the boundaries of the measurement regions.
Leave space at the boundaries when setting the regions.
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
1.
The screen for specifying the region will be displayed.
2.
The cursor will be moved to the ending line.
Select Image/Region.
Specify the starting line of the measurement region.
Move the cursor with the Up and Down Keys. (Use these keys together with the SHIFT Key to move the cursor quickly.)
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Cursor
Meas method
Region
Sensitivity Details
3.
A confirmation message will be displayed.
4.
The setting will be registered and the screen in (1.) will return.
Repeat the procedure described in (2.) to specify the ending line of the measurement region.
Select Execute.
Z500
Operation Manual
58
The measurement region
will be set. OK?
Execute
Cancel
Page 61
Sensitivity Adjustment
)
ght
y
e
ow
Select the sensitivity adjustment method for the measurement range.
If the measurement results fluctuate, change the adjustment method.
Normally, these conditions can be left at the default settings.
CHECK
The setting can be performed through the adjustment menu.
Refer to page 28.
SECTION 4
Menus for Expert Menu
REFERENCE
Two Sensor
Select the Sensor number to be used for measurement.
Sensitivity Adjustment
Selection Description
Auto*
Fixed
Multi
Adjust the sensitivity according to sensitivity information in the measurement region.
The sensitivity can be set to one of 31 levels (Setting range: 0 to 31). Select when it is difficult to assure accurate measurement in Auto. Such cases may appear on a production line where workpieces of different colors are processed.
Adjust the sensitivity for each line in the measurement region. Select when it is difficult to assure accurate measurement in Auto due to a widely distributed brightness of the workpiece surfaces. However, sampling time will be longer. Position a reference workpiece and execute Multi. The Z500 will set automatically the upper and lower limits of sensitivity on the basis of the brightness of the reference workpiece. After the settings have been completed, fine adjustment can be made on
Details.
REFERENCE
Refer to page 63.
The asterisk (*) indicates the default setting.
Sensitivity level (Displayed only when Fixed is selected from the Sensitivity Menu)
SECTION 4
Step
1
Step
2
Step
3
Step
4
Setting range
LV0 (See note
LV31 High
Note: The laser will turn OFF when selecting [LV0].
Sensitivit
L
Color of workpiec
Bri
Dark
Step
5
Step
6
Step
7
Operation Manual
Z500
59
Page 62
SECTION 4
Menus for Expert Menu
When selecting Auto
Select Image/Sensitivity.
1.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Meas method Region
Sensitivity
Details
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
The selections for sensitivity adjustment will be displayed.
2.
Select Auto.
The setting will be registered and the screen in (1.) will return.
Auto
Fixed Multi
60
Z500
Operation Manual
Page 63
When selecting Fixed
Select Image/Sensitivity.
1.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
SECTION 4
Menus for Expert Menu
Meas method Region
Sensitivity
Details
The selections for sensitivity adjustment will be displayed.
2.
Select Fixed.
The screen for setting the sensitivity level will be displayed.
3.
Select the level using the Up and Down Keys while monitoring the
image on the monitor.
The level displayed first is the level selected by the Z500 for the work­piece being measured. Adjust the level while monitoring the line bright­ness graph and the received level on the monitor. 120 max. : Insufficient light density 255 min. : Excessive light density
4.
Press the ENT Key.
Auto
Fixed
Multi
Line Brightness Received level
LV 15
BRIGHT
DARK
Sen0
LV
PEAK
Sen1
LV PEAK
[ 15]
[ 170]
SECTION 4
Step
1
FARNEAR
Step
2
[ .] [ ]
Step
3
Step
4
Step
5
Step
6
The setting will be registered and the screen in (1.) will return.
Operation Manual
Z500
Step
7
61
Page 64
SECTION 4
Menus for Expert Menu
When selecting Multi
Select Image/Sensitivity.
1.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Meas method Region
Sensitivity
Details
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
The selections for sensitivity adjustment will be displayed.
2.
The sensitivity multiple setting screen will be displayed.
3.
Select Multi.
Check whether or not the workpiece is positioned inside the measurement range and press the ENT Key.
Auto Fixed
Multi
Line beam image
Scn 0 Set
The setting will be registered and the screen in (1.) will return.
Z500
Operation Manual
62
Page 65
Details
SECTION 4
Menus for Expert Menu
Use this function to measure transparent workpieces. This function is also
useful to shorten the response time in Sensitivity/Multi. Normally, these
conditions can be left at the default settings. After changing the settings, perform
a workpiece measurement to check that measurement can still be performed
correctly.
CHECK
1.
Item
Surface
Upper 0 to 31(31*) Specify the range to be adjusted the sensitivity.
Lower 0 to 31(1*)
Gradations 120 to 220(170*)
The setting can be performed through the adjustment menu.
REFERENCE
Seleciton/
Setting range
Frnt* Back
Description
The reflecting surface for which sensitivity adjustment is performed is selected. To measure the back side of a transparent workpiece, select Back. Normally (i.e., when measuring opaque workpieces), this setting can be left as Frnt.
This function is effective when sensitivity adjustment method is set to Auto or Multi. If the adjustment range is small, the response time for switching sensitivity will be short.
Set the target gradation for sensor sensitivity adjustment.
The asterisk (*) indicates the default setting.
Refer to page 28.
Select Image/Details.
Scn 0 Set
Image
Compensatn Meas set Conditions Output
Meas method Region
Sensitivity
Details
SECTION 4
Step
1
Step
2
Step
3
Step
4
The setting screen for sensitivity details will be displayed.
2.
3.
Make the settings for each item.
Select End.
The setting will be registered and the screen in (1.) will return.
Scn 0 Set
Surface Upper Lower Gradations
:
Frnt
:
[31]
:
[ 1]
:
[170]
End
Operation Manual
Step
5
Step
6
Step
7
Z500
63
Page 66
SECTION 4
Menus for Expert Menu
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
Step 3
Position Compensation
The position compensation function is used when the position of the measurement workpieces are not
consistent.
By using this function, the displacement between the reference position and the current position is
obtained, and this displacement is compensated for in measurements. Compensation can be made for
both height and position.
Height
Position
• Reference Position
The measurement range and measurement workpiece are in the correct
positions.
Workpiece
Sensor
Height
Scn 0 Run
BRIGHT
DARK
Measurement range
Measurement center
LowHigh
Measurement point
9.96ms
FARNEAR
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
64
When the Position of the Workpiece is Lower
By making position displacement compensation settings ...
For measurement, compensate the height based on the displacement of the reference position.
Z500
Operation Manual
Scn 0 Run
BRIGHT
DARK
Measurement point
9.96ms
FARNEAR
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Page 67
Operational Flow
1
2
#
Page 65
Registering the Reference Position
#
Page 69
Set the Compensation Method
SECTION 4
Menus for Expert Menu
CHECK
#1
Registering the Reference Position
To use the position compensation function, set Compensatn and Meas set in this order.
Register the position to be used for a reference for position compensation.
Before setting the necessary items, place the workpiece at the correct position.
In the example provided below, the procedure for compensating under the
following conditions is explained.
Measurement point (Average)
Reference height measurement point
Selection Description
Average
Peak
Bottom
Peak Bottom
Reference height Reference position
Region
Measurement point (Edge position)
Edge level
Define the average profile position within a specified region as the measurement point.
Define the peak profile position within a specified region as the measurement point.
Define the bottom profile position within a specified region as the measurement point.
SECTION 4
Step
1
Step
2
Step
3
Step
4
Step
5
Region
Step
6
Step
7
Operation Manual
Z500
65
Page 68
SECTION 4
Menus for Expert Menu
SECTION 4
1.
The selection for position compensation will be displayed.
2.
Select Compensatn.
Select Reference.
Scn 0 Set
Image
Compensatn
Meas set Conditions Output
Scn 0 Set
Image
Compensatn
Meas set Conditions Output
None
Reference
4-1 Setting Measurement Conditions Using the Expert Menus
A profile graph will be displayed.
3.
The screen for setting the reference height will be displayed.
4.
Check whether or not the workpiece is positioned correctly and press the ENT Key.
Specify the starting line of the reference height region.
Use the Up and Down Keys to move the cursor. (Use these keys together with the SHIFT Key to move the cursor quickly.)
Profile
Cursor
Measurement region
66
Z500
Operation Manual
Page 69
The ending-line setting screen will be displayed.
SECTION 4
Menus for Expert Menu
5.
The characteristic quantity setting screen will be displayed.
6.
Specify the ending line of the reference height region.
Select a characteristic quantity that can be used as the measurement point of the reference height.
The cross cursor of the measurement point will move in response to the selections. Select the item while monitoring the image. In this example, select Average.
Average
Peak Bottom
Cursor
Cursor
SECTION 4
Step
1
Step
2
The setting screen for reference position will be displayed.
7.
Specify the edge level to be used for the reference position.
Use the Left and Right Keys to move the cursor. (Use these keys together with the SHIFT Key to move the cursor quickly.)
Cursor
Operation Manual
Step
3
Step
4
Step
5
Step
6
Step
7
Z500
67
Page 70
SECTION 4
Menus for Expert Menu
SECTION 4
A cross cursor to be used for specifying the reference height will
be displayed.
8.
A confirmation message will be displayed.
9.
Use the Up and Down Keys to select the measurement point of the reference position.
The cursor of the measurement point is moved. Select the measurement point while monitoring the image.
Select Execute.
Cursor
Edge level
Reference will be
registered.
OK?
Execute Cancel Back
4-1 Setting Measurement Conditions Using the Expert Menus
The setting will be registered and the screen in (2.) will return.
68
Z500
Operation Manual
Page 71
#2
Set the Compensation Method
SECTION 4
Menus for Expert Menu
CHECK
1.
Select the position compensation method.
Height
Position
Unless Reference has been set, the compensation method selections will not be
displayed.
Move the cursor to None and press the ENT Key.
.
Selection Description
None* No compensation is required.
Height Compensate the height.
Position Compensate the position.
Height&Pos Compensate both the height and position.
The asterisk (*) indicates the default setting.
Scn 0 Set
Image
Compensatn
Meas set Conditions Output
None
Reference
SECTION 4
Step
1
The selection for compensation method will be displayed.
2.
The setting will be registered and the screen in (1.) will return.
Select the compensation method. In this example, select Height&Pos.
None Height
Position
Height&Pos
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
69
Page 72
SECTION 4
Menus for Expert Menu
SECTION 4
Step 4
Measurement Settings
Set the measurement contents. Operational flow varies depending on the measurement item.
Setting Step: 2 pts, Edge position, and Define are taken as examples for description.
Operational Flow
#2
#1
Page 70
Set Measurement Items
When selecting
#2
Compensate the measurements under the output conditions given in
CHECK
CHECK
When not selecting
Define
Define
Calibration setting is disabled.
To use the position compensation function, set Compensatn and Meas set in this order.
The settings can be performed through the adjustment menu.
REFERENCE
Refer to page 28.
Page 78 (if necessary)#3 Page 84 (if necessary)
Set Calibration
:
:
Set Output Conditions
Changing Measurement Contents
Clearing Measurement Items
#3.
Page 86
Page 87
4-1 Setting Measurement Conditions Using the Expert Menus
#1
Set Measurement Items
Up to eight measurement contents can be set for each scene.
There are 8 types of the measurement item. Different measurement item can be
set for each output number.
Types of Measurement Item
Selection Description Symbol
Height* Measure the height within the specified region.
Specify two measurement regions. After taking the measurement point within the 1st region
Step: 2 pts
Step: 3 pts
Edge position
as the reference, measure the distance from the mea­surement point in the 2nd region.
REFERENCE
Specify three measurement regions. After taking the average of the measurement points within the two regions specified above, measure the difference from the measurement point in the 3rd region.
Measures the specified edge position.
REFERENCE
Refer to page 72.
Refer to page 74.
70
*Height is set in OUT0 as the default setting.
Z500
Operation Manual
Page 73
Types of Measurement Item (Continued)
Selection Description Symbol
SECTION 4
Menus for Expert Menu
Width
Edge center
Peak/bottom
Define
Measure the difference between the specified two edge positions.
Based on the specified two edge positions, measure the center.
Measure the peak or bottom point of the height within the specified region.
Using the equation, set the measurement items. The measurement result for an output number can be assigned to the equation.
REFERENCE
Refer to page 76.
Characteristic Quantity of Measurement Points
Selection Description
Average
Peak
Bottom
Back
Define the average profile position within a specified region as the measurement point.
Define the peak profile position within a specified region as the measurement point.
Define the bottom profile position within a specified region as the measurement point.
Set the back side of the workpiece for the measurement point. This procedure is effective when glass sheets and other transparent workpieces are measured.
*Height is set in OUT0 as the default setting.
SECTION 4
Step
1
Step
2
Step
3
Peak and Bottom
Peak Bottom
Region
Back
Workpiece
Operation Manual
Surface
Back
Z500
Step
4
Step
5
Step
6
Step
7
71
Page 74
SECTION 4
Menus for Expert Menu
Step: 2 pts
Level differences, grooves, etc. of workpieces are measured.
The following conditions will be taken as an example for describing the setting procedures.
SECTION 4
Step: 2 pts
Measurement point (Average)
Measurement point (Average)
1.
Select Meas set.
A list of output numbers will be displayed.
2.
Select the output number for which Step: 2 pts is set. In this example, Select OUT1.
CHECK
Height is set in OUT0 as the default set- ting. To set an item other than Height in OUT0, delete the existing measurement item.
Top face region
Bottom face region
Scn 0 Set
Image
Compensatn
Meas set
Conditions Output
Scn 0 Set
OUT0:Height
OUT1:
OUT2: OUT3: OUT4: OUT5: OUT6: OUT7:
REFERENCE
Refer to page 87.
The selection for measurement item will be displayed.
3.
Select the measurement item.
In this example, select Step: 2 pts.
Z500
Operation Manual
72
Measrmnt item Height
Step: 2 pts
Step: 3 pts Edge position Width Edge center Peak/bottom Define
Page 75
SECTION 4
Menus for Expert Menu
The setting screen for bottom face measurement region will be
displayed.
4.
The ending-line setting screen will be displayed.
5.
The characteristic quantity setting screen will be displayed.
Specify the starting line of the bottom face measurement region. Move the cursor with the Up and Down Keys. (Use these keys together with the SHIFT Key to move the cursor quickly.)
Repeat the procedure described in (4.) to specify the ending line of the bottom face measurement region.
Cursor
Cursor
SECTION 4
Cursor
6.
The setting screen for top face measurement region will be displayed.
7.
A confirmation message will be displayed.
8.
Select a characteristic quantity that can be used as the measurement point of the bottom face. The cross cursor of the measurement point will move in response to the selections. Select the item while monitoring the image. In this example, select Average.
Repeat (4.) through (6.) to set the characteristic quantities of the top face measurement region and measurement point.
Select Execute.
Average
Peak Bottom Back
Measrmnt items will be set.
OK?
Execute Cancel Back
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
The setting will be registered and the screen in (2.) will return.
Operation Manual
Z500
73
Page 76
SECTION 4
SECTION 4
Menus for Expert Menu
Edge position
Measures the edge position of the workpiece.
The following conditions will be taken as an example for describing the setting procedures.
Edge level
Measurement point (Edge position)
In this example, the intersection point of the edge level and the profile becomes
the measurement point.
Set the edge level as follows.
Edge level
Profile
The intersection is the measurement point.
Edge position
CHECK
The setting edge level and the measurement point can be checked on the Image Monitor.
REFERENCE
Refer to page 32.
1.
Select Meas set.
A list of output numbers will be displayed.
2.
Select the output number for which Edge position is set. In this example, select OUT2.
CHECK
Height is set in OUT0 as the default
setting. To set an item other than Height in OUT0, delete the existing measure­ment item.
Scn 0 Set
Image
Compensatn
Meas set
Conditions Output
Scn 0 Set
OUT0:Height OUT1:Step: 2 pts
OUT2:
OUT3: OUT4: OUT5: OUT6: OUT7:
74
REFERENCE
Refer to page 87.
Z500
Operation Manual
Page 77
SECTION 4
Menus for Expert Menu
The selection for measurement item will be displayed.
3.
The setting screen for edge position will be displayed.
4.
Select the measurement item. In this example, Select Edge position.
Specify the level to be measured as the edge. Move the cursor with the Left and Right Keys. (Use these keys together with the SHIFT Key to move the cursor quickly.)
Measrmnt item Height
Step: 2 pts
Step: 3 pts
Edge position
Width Edge center Peak/bottom Define
SECTION 4
The setting screen for edge position will be displayed.
5.
CHECK
A confirmation message will be displayed.
6.
Specify the edge position regarded as the measurement point. Move the cursor with the Up and Down Keys. In some cases, the cursor will not be displayed on the intersection point of the edge level and the edge position. However, selection of an arbitrary cursor will give the same measurement result (the intersection point becomes the measurement point), as long as the selected cursor exists near the intersection point.
Select Execute.
Cursor
Measrmnt items will be set.
OK?
Execute Cancel Back
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
The setting will be registered and the screen in (2.) will return.
Operation Manual
Z500
75
Page 78
SECTION 4
Menus for Expert Menu
Define
Set the equation by using the measurement results for the output number.
Output number for which
OUT3
OUT1 OUT2
equation is set
Equation
SECTION 4
Define
List of Items
Note 1
• Output number
CHECK
OUT0
OUT1 OUT2 OUT3 OUT4
See note 1.
OUT0
OUT1 OUT2 OUT3 OUT4
OUT5 OUT6 OUT7 ( )
See note 2. See note 3.
OUT5 OUT6 OUT7 ( )
ABS MAX MIN AVE Cons
ABS MAX MIN
DEL
AVE
BS
Cons
Set
Move the cursor.
DEL BS Set
Numerals between 0 and 9999999.999 can be set.
List of Items
Select the items displayed here for which equation is set.
Delete the items.
DEL : Deletes the item to the right of the cursor. BS : Deletes the item to the left of the cursor.
Sets the equation.
When this item is selected, equation can be performed by using the
measurement results (results obtained after calibration) that have been set for
the output number.
Any output number cannot be selected, unless its measurement items have been set.
76
Note 2
• Arithmetic Operators
Note 3
•Functions
Z500
Operation Manual
Operator Meaning
+ Addition
- Subtraction
* Multiplication
/ Real number division
Function Description
ABS Calculates the absolute value. ABS (argument)
MAX
MIN
AVE
Gives the largest value of arguments. MAX (Argument 1, Argument 2, Argument 3,…)
Gives the smallest value of arguments. MIN (Argument 1, Argument 2, Argument 3,…)
Calculates the average value of arguments. AVE (Argument 1, Argu­ment 2, Argument 3,…)
Page 79
SECTION 4
Menus for Expert Menu
1.
Select Meas set.
A list of output numbers will be displayed.
2.
Select the output number for which Define is set. In this example, select OUT3.
CHECK
Height is set in OUT0 as the default set- ting. To set an item other than Height in OUT0, delete the existing measurement
item.
Scn 0 Set
Image
Compensatn
Meas set
Conditions Output
Scn 0 Set
OUT0:Height OUT1:Step: 2 pts OUT2:Edge position
OUT3:
OUT4: OUT5: OUT6: OUT7:
SECTION 4
REFERENCE
Refer to page 87.
The setting screen for equation will be displayed.
3.
Select the item from the list of items and set the equation.
4.
Select Set when the equation has been set.
A confirmation message will be displayed.
5.
Select Execute.
OUT3
OUT0
OUT5
OUT1
OUT6
OUT2
OUT7
OUT3
(
OUT4
)
Measrmnt items will be set.
Execute Cancel Back
OK?
ABS
MAX
MIN
AVE
Cons
DEL BS Set
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
The setting will be registered and the screen in (2.) will return.
Operation Manual
Z500
77
Page 80
SECTION 4
Menus for Expert Menu
#2
Set Calibration
Set calibrations to compensate the measurement values. Set calibrations to compensate for differences
between measurement values and actual dimensions due to the color or material of the measurement
workpiece. The setting method varies depending on the measurement items set for the output number.
Compensation Measurement items Description
Compensate the measured height and position.
Compensate the measured level difference and width.
Height, Edge position, Edge center, Peak/bottom
Step: 2 pts, Step: 3 pts, Width
For setting, repeat the measurement twice.
REFERENCE
For setting, repeat the measurement once.
REFERENCE
Refer to page 78.
Refer to page 81.
SECTION 4
Define
CHECK
Output can be used for fine adjustment after calibration setting. Use Output also to
compensate the measurement values when Define has been selected for the
measurement item.
REFERENCE
Refer to page 84.
CHECK
Refer to page 83 for information on how to clear calibration settings.
Compensate the Measured Height and Position
To compensate for the error, perform the measurement twice. Applicable measurement items: Height, Edge position, Edge center, Peak/bottom
Example 1: When selecting
Workpiece height: 10 mm Use the method shown below to calibrate workpieces so that a displacement value of 9.95 mm is measured as 10 mm.
9.95mm
Calibrate to 10 mm.
Measurement of point 1
Height
Measurement value
10mm
Measurement of point 2
Compensate for this error.
10mm
Displacement of workpiece
78
Measure without the workpiece.
Z500
Operation Manual
Input the height
of point 1.
0mm
Input the height
of point 2.
10mm
Measure with the workpiece in position.
Page 81
SECTION 4
Menus for Expert Menu
Example 2: When selecting
Movement distance from the edge position: 5 mm Use the method shown below to calibrate workpieces so that a displacement value of 4.95 mm is measured as 5 mm.
4.95mm
Calibrate to 5 mm.
Measurement of point 1
Input the
position
of point 1.
0mm
Measure with the workpiece at the reference position.
Edge position
Measurement value
5mm
Measurement of point 2
Move the workpiece by 5 mm for the measurement.
5mm
Compensate for this error.
Workpiece Movement
Input the
position
of point 2.
5mm
Height in Example 1 is taken here to describe the setting method.
Make the necessary adjustments when setting a different item.
SECTION 4
1.
Move the cursor to Height and press the ENT Key.
The selections will be displayed.
2.
Select Calibrate.
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts OUT2:Edge position OUT3:Define OUT4: OUT5: OUT6: OUT7:
Scn 0 Set
Change
Calibrate
Output Delete
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
79
Page 82
SECTION 4
Menus for Expert Menu
SECTION 4
A confirmation message will be displayed.
3.
The measurement screen for point 1 will be displayed.
4.
The screen for inputting the position of point 1 will be displayed.
5.
Select Execute.
Press the ENT Key without a workpiece in position.
Set the value to 0.00000 and press the ENT Key.
System will be calibrated.
Clear settings now if
required.
Execute Cancel
00000.00000
Clear
Define
The measurement screen for point 2 will be displayed.
6.
The screen for inputting the position of point 2 will be displayed.
7.
A confirmation message will be displayed.
8.
The setting will be registered and the screen in (2.) will return.
Position the workpiece correctly and press the ENT Key.
Set the value to 5.00000 and press the ENT Key.
Select Execute.
00005.00000
System will be calibrated.
Execute Cancel
80
Z500
Operation Manual
Page 83
Compensate the Measured Level Difference and Width
To compensate for the error, perform the measurement once.
Applicable measurement items: Step: 2 pts, Step: 3 pts, Width
SECTION 4
Menus for Expert Menu
Example : When selecting
Level difference: 5 mm Use the method shown below to calibrate workpieces so that a displacement value of 4.95 mm is measured as 5 mm.
Point 1
Point 2
To compensate for this error, perform the measurement once.
Measure with the workpiece in position.
1.
Move the cursor to Step: 2 pts and
Calibrate to 5 mm.
Input the value of point 1.
press the ENT Key.
0mm
Step: 2 pts
Measurement value
5mm
4.95mm
5mm
Input the value of point 2.
5mm
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts OUT2:Edge position OUT3:Define OUT4: OUT5: OUT6: OUT7:
Compensate for this error.
Level difference of workpiece
SECTION 4
Step
1
Step
2
Step
3
Step
4
The selections will be displayed.
2.
Select Calibrate.
Scn 0 Set
Change
Calibrate
Output Delete
Operation Manual
Z500
Step
5
Step
6
Step
7
81
Page 84
SECTION 4
Menus for Expert Menu
SECTION 4
A confirmation message will be displayed.
3.
The measurement screen for point will be displayed.
4.
The screen for inputting the position of point 1 will be displayed.
5.
Select Execute.
Position the workpiece correctly and press the ENT Key.
Set the value to 0.00000 and press the ENT Key.
System will be calibrated.
Clear settings now if
required.
Execute Cancel
00000.00000
Clear
Define
The screen for inputting the position of point 2 will be displayed.
6.
A confirmation message will be displayed.
7.
The setting will be registered and the screen in (2.) will return.
Set the value to 5.00000 and press the ENT Key.
Select Execute.
00005.00000
System will be calibrated.
Execute Cancel
82
Z500
Operation Manual
Page 85
Clearing Calibration Settings
SECTION 4
Menus for Expert Menu
1.
The selections will be displayed.
2.
Move the cursor to the measurement item to be cleared and press the ENT Key.
Select Calibrate.
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts OUT2:Edge position OUT3:Define OUT4: OUT5: OUT6: OUT7:
Scn 0 Set
Change
Calibrate
Output Delete
SECTION 4
Step
1
A confirmation message will be displayed.
3.
The setting will be cleared and the screen in (2.) will return.
Select Clear.
System will be calibrated.
Clear settings now if
required.
Execute Cancel
Clear
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
83
Page 86
SECTION 4
SECTION 4
Menus for Expert Menu
#3
Set Output Conditions
Set the output conditions of measurement results as necessary.
Item Selection Description
Upper
Lower
Hold
Zero
-9999.99999 to
9999.99999
Set when the Trigger function is used. Select the contents of the data to be held.
REFERENCE
None* Hold measurement is not performed.
Sampling The value at the start of measurement is held.
Peak
Bottom The lowest value measured during measurement is held.
Peak-Peak
Average
-9999.99999 to
9999.99999
Set the judgment conditions. Set the range for a PASS judgment.
For details on the Trigger function, refer to page 90.
The highest value measured during measurement is held.
The difference between the highest and lowest values measured during measurement is held.
The average value between the highest and lowest values measured during measurement is held.
Set the offset value for force-zero execution. Set the value to be added to 0.
REFERENCE
Refer to page 110, Force-zero.
Define
Span and Offset
Span
A coefficient (the span) is set to compensate for incline in the Sensor characteristics.
Measurement value
(mm)
Span
Offset
Unit
4.0
0.20000 to
4.00000
(1.00000*)
-9999.99999 to
9999.99999
(0.00000*)
mm*
2
mm None
0.2
Displacement of workpiece
Set to make fine adjustments to measurement values after calibration. Set for the incline in the Sensor characteristics.
Set to make fine adjustments to measurement values after calibration. Added to or subtracted from measurement values that have been set.
Select the unit of the measured values to be displayed on the monitor screen.
The asterisk (*) indicates the default setting.
Offset
A fixed value (the offset) is added to or subtracted from measurement values.
Measurement value
(mm)
Displacement of workpiece
84
CHECK
Z500
Operation Manual
When calibration is executed, the span and the offset will be automatically set.
Page 87
SECTION 4
Menus for Expert Menu
1.
Move the cursor to the output number to be set the output conditions and press the ENT Key.
The selections will be displayed.
2.
Select Output .
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts OUT2:Edge position OUT3:Define OUT4: OUT5: OUT6: OUT7:
Scn 0 Set
Change Calibrate
Output
Delete
SECTION 4
The setting screen for output conditions will be displayed.
3.
4.
Set the necessary items.
Select End.
The setting will be registered and the screen in (2.) will return.
Scn 0 Set
Upper Lower Hold Zero Span Offset Unit
:
[ 10.00000]
:
[ -10.00000]
:
None
:
[ 0.00000]
:
[1.00000]
:
[ 0.00000]
:
mm
End
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
85
Page 88
SECTION 4
Menus for Expert Menu
Changing Measurement Contents
Change the measurement contents that have been set for the output number.
SECTION 4
Changing Measurement Contents
CHECK
To change measurement items, clear the existing measurement items and then set new items.
Refer to page 87.
1.
Move the cursor to the output number to be changed the output conditions and press the ENT Key.
The selections will be displayed.
2.
Select Change.
REFERENCE
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts OUT2:Edge position OUT3: OUT4: OUT5: OUT6: OUT7:
Scn 0 Set
Change
Calibrate Output Delete
The setting screen for measurement range will be displayed.
The remaining procedures are the same as those for measurement item setting.
Z500
Operation Manual
86
Page 89
Clearing Measurement Items
Clear the measurement items that have been set for the output numbers.
REFERENCE
Refer to page 127 for details on clearing separately for each scene.
SECTION 4
Menus for Expert Menu
1.
The selections will be displayed.
2.
Move the cursor to the output number to be cleared the output conditions and press the ENT Key.
Select Delete.
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts OUT2:Edge position OUT3:Define OUT4: OUT5: OUT6: OUT7:
Scn 0 Set
Change Calibrate Output
Delete
SECTION 4
Step
1
Step
2
Step
3
A confirmation message will be displayed.
3.
The measurement items for the selected output number are cleared off and the screen in (1.) will return.
Select Execute.
Delete OUT0 settings.
Execute Cancel
Z500
Operation Manual
Step
4
Step
5
Step
6
Step
7
87
Page 90
SECTION 4
Menus for Expert Menu
SECTION 4
Clearing Measurement Items
Step 5
Measurement Conditions
Set the measurement conditions. The following can be set in Conditions and its lower hierarchies.
Perform the setting as required.
Menu Description Reference
The average of the setting number of times is output as
# to avg
Trigger
Details
the measurement result. Set this function to disregard sudden changes in the waveform.
From the measurement results that have been obtained within a specified measurement period, the specified characteristic quantity is held and output.
Removing noises, outputting measurement failure, and other miscellaneous methods can be set in more detail.
Page 88.
Page 90.
Page 96.
Averaging number
Use this function to disregard sudden changes in the waveform. The average of
the setting number of times is output as the measurement result.
Selection
• The Relationship between the Measurement Response Time and the Resolution
Averaging number
1
1, 2, 4, 8*, 16, 32, 64, 128, 256
Measurement response time
Short
Resolution
Low
The asterisk (*) indicates the default setting.
CHECK
1.
256 Long
The setting can be performed through the adjustment menu.
REFERENCE
High
Refer to page 28.
Select Conditions.
Scn 0 Set
Image
Compensatn Meas set
Conditions
Output
88
Z500
Operation Manual
Page 91
SECTION 4
Menus for Expert Menu
The selections will be displayed.
2.
The setting screen for averaging number will be displayed.
3.
4.
Select # to avg.
Select averaging number.
Select End.
Scn 0 Set
# to avg
Trigger Details
# to avg
: 8
End
SECTION 4
The setting will be registered and the screen in (2.) will return.
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
89
Page 92
SECTION 4
Menus for Expert Menu
Trigger setting
Use the Trigger function to measure moving workpieces in chronological order.
From the measurement results that have been obtained within a specified
measurement period, the specified characteristic quantity is held and output.
Use the procedure explained below to set the trigger input method.
Workpiece movement direction
SECTION 4
Clearing Measurement Items
Height(mm)
Time(s)
Position(mm)
• Trigger Input Method
Trigger OFF
End
Start
Trigger ON
Selection Description
Free*
Measurement is performed continuously. Hold measurement is not performed.
Select when using a sync sensor.
External
The trigger turns ON at the same time as the TRIGGER terminal on the terminal block.
The trigger-ON timing is determined according to the measurement value. The trigger turns ON when the measurement value goes above (or below) the trigger level. If Self is selected, the screen for selecting the output number (for the trigger setting) and the trigger direction is displayed.
The characteristic quantities being held are output at this timing.
The characteristic quantity of the measurement during this period will be held.
terminal
OFF
ON
Trigger ON
TRIGGER
90
CHECK
Z500
Operation Manual
Trigger ON
Self
Up*
The trigger turns ON when the measurement value goes above the trigger level.
Trigger level
Measurement value waveform
The trigger turns ON
Down
when the measurement value goes below the trigger level.
Measurement value waveform
Trigger level
The asterisk (*) indicates the default setting.
Use Output to select the characteristic quantity that holds the measured values.
REFERENCE
Refer to page 84.
Page 93
SECTION 4
Menus for Expert Menu
The procedure for make trigger settings under the following details is explained
below.
The difference between the maximum value and minimum value of the workpiece height is output.
Measurement item Trigger input method Hold setting (see note)
OUT0
Note:
Set the Hold setting by using
Refer to page 84.
Operational Flow
#1
Page 91
Display Waveforms in the Trend Monitor (Through Mode)
Display Waveforms in the Trend Monitor
#1
Use the waveform to decide the trigger level and to set the measurement start
position.
CHECK
1.
2.
3.
Set the screen mode to Thru in advance.
REFERENCE
Refer to page 118.
Press the SHIFT + Right Keys or SHIFT + Left Keys several times to switch to Trend Monitor.
Pass the workpiece under the Sensor and display the waveform.
The instant the waveform representing the workpiece is displayed, press the
TRIG Key.
OUT0:Height
Self
Peak-Peak
Meas set
#2
Page 92
Set the Trigger Input Method (Trigger Mode)
in advance.
#3
Page 95
Confirm Trigger Settings
Scn 0 Run
1div x: 1.000 [s] y:+10.0000 [mm]
Thru
OUT 0
S+ / :Display S+ENT:Adjust
9.96ms
Zero's OFF
mm+0005.18800
PASS
SECTION 4
Step
1
Step
2
Step
3
Step
4
Step
5
The waveform display is fixed.
Scn 0 Set
Thru
The symbol changes.
1div x: 1.000 [s] y:+10.0000 [mm]
Z500
Operation Manual
Step
6
Step
7
91
Page 94
SECTION 4
Menus for Expert Menu
Set the Trigger Input Method
#2
Select the trigger input method, and set the trigger position and the trigger level.
CHECK
The setting can be performed through the adjustment menu.
REFERENCE
Refer to page 28.
SECTION 4
Clearing Measurement Items
1.
The selections will be displayed.
2.
The selections for trigger input method will be displayed.
3.
Enter Set Mode and select Conditions.
Select Trigger.
Select the trigger input method and press the ENT Key. In this example, select Self.
Scn 0 Set
Image
Compensatn Meas set
Conditions
Output
# to avg
Trigger
Details
Free External
Self
The setting screen for trigger contents will be displayed.
4.
5.
Set each item.
In this example, select Trigger to OUT0 and Direction to Up.
Select Execute.
The trigger setting screen will be displayed. The waveform set in
Z500
Operation Manual
92
is displayed.
#1
Scn 0 Set
Free External
Self
Trigger Direction
Execute
:OUT0 :Up
Cancel
Page 95
SECTION 4
Menus for Expert Menu
6.
Specify the trigger-ON position and trigger level, and press the ENT Key.
Use the Left and Right Keys to specify the trigger position. Use the Up and Down Keys to specify the trigger level. Use these Keys together with the SHIFT Key to move the cursor quickly.
CHECK
When the trigger turns ON, the waveform above the trigger level is displayed at a position
based on the position set here.
Trigger Position in the Center Trigger Position on the Left
The waveform before and after the trigger turns ON can be monitored.
The trigger-OFF setting screen will be displayed.
Mainly the waveform after the trigger turns ON can be monitored.
Trigger-ON position
Trigger
Thru
Trigger-OFF position
Trigger level
1div x: 1.000 [s] y:+10.0000 [mm]
SECTION 4
7.
Using the Left and Right Keys, specify the trigger-OFF position.
Use the Left and Right Keys together with the SHIFT Key to move the cursor quickly.
The setting screen for measurement start timing will be
displayed.
8.
Using the Left and Right Keys, specify the measurement starting position.
Use the Left and Right Keys together with the SHIFT Key to move the cursor quickly.
CHECK
The time for cursor position (time after the trigger is turned ON) is displayed on the lower part of the screen. Refer this time for setting the measurement starting position.
Trigger
Thru
Measurement start position
Trigger
Thru
Measure start 249.00 ms
Time for cursor position
1div x: 1.000 [s] y:+10.0000 [mm]
1div x: 1.000 [s] y:+10.0000 [mm]
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
93
Page 96
SECTION 4
Menus for Expert Menu
SECTION 4
Clearing Measurement Items
The cursor will be moved to the measurement ending position.
9.
Repeat the procedure described in (8.) to specify the measurement ending position.
CHECK
The time for cursor position (time after the trigger is turned ON) is displayed on the lower part of the screen. Refer this time for setting the measurement ending position.
A confirmation message will be displayed.
10.
Select Execute.
The setting will be registered and the screen in (2.) will return.
Measurement end position
Trigger
Thru
Measure start 249.00 ms Measure stop 2320.68 ms
Time for cursor position
The settings
will be changed.
Execute Cancel Back
OK?
1div x: 1.000 [s] y:+10.0000 [mm]
94
Z500
Operation Manual
Page 97
Menus for Expert Menu
Confirm Trigger Settings
#3
Check if measurements are being correctly performed with the set trigger contents.
CHECK
Set the screen mode to Trig in advance.
REFERENCE
Refer to page 118.
SECTION 4
1.
2.
The waveform is displayed from the trigger position set in #2.
3.
• When the measurement start and end positions have to be adjusted
Press the SHIFT + Right Keys or SHIFT + Left Keys several times to switch to Trend Monitor.
Pass the workpiece under the Sensor.
Perform a workpiece measurement several times to check that measurement can be performed correctly.
, with the waveform kept displayed, and adjust the
Return to the procedure in positions.
#2
Scn 0 Run
1div x: 1.000 [s] y:+10.0000 [mm]
Thru
OUT 0
S+ / :Display S+ENT:Adjust
Scn 0 Run
Trig
Zero's OFF
mm+0005.18800
PASS
9.96ms
1div x: 1.000 [s] y:+10.0000 [mm]
Trigger-OFF positionTrigger-ON position
9.96ms
SECTION 4
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Operation Manual
Z500
Step
7
95
Page 98
SECTION 4
Menus for Expert Menu
Details setting
Use the details setting when measurement cannot be performed correctly or when more accurate
measurement is required.
SECTION 4
Clearing Measurement Items
1.
Select Conditions.
The selections will be displayed.
2.
Select Details.
The setting screen for details will be displayed.
3.
4.
Make the settings for each item.
Select End.
Scn 0 Set
Image
Compensatn Meas set
Conditions
Output
Scn 0 Set
judgement width Edge width Smoothing Noise Level Error data Profile Peak/btm width
# to avg Trigger
Details
End
:
[ 0.01000]
:
[ 0.00000]
None
:
2pixel
:
50.0%
:
Clamp
:
Intrpolate
:
1pixel
:
The setting will be registered and the screen in (2.) will return.
Z500
Operation Manual
96
Page 99
Judgment value hysteresis width (judgement width)
Set the hysteresis width for Judgment limits.
SECTION 4
Menus for Expert Menu
Setting range
0.00000 to 9999.99999
The default setting varies depending on the Sensor connected.
Example: HI: 100.000 mm; LO: 80.000 mm Judgment value hysteresis width: 0.05 mm
HI
(100.0000)
LO
(80.0000)
OFF
HIGH
ON
Output
LOW
PASS
OFF ON
OFF ON
Edge Detection Hysteresis Width (Edge width)
Set the hysteresis width of edge detection.
Change the edge width when edge detection is unstable.
Setting range
Example: Edge level: 80.000 mm
Edge level hysteresis width: 0.0500 mm
0.00000 to 9999.99999 (0.00000*)
Edge level
(80.0000)
Edge position
Hysteresis(0.0500)
Resets at 99.9500 mm
The asterisk (*) indicates the default setting.
Hysteresis (0.0500)
Hysteresis(0.0500)
Resets at 80.0500 mm
SECTION 4
Step
1
Step
2
Smoothing
(Smoothing)
Noise removal
Use this function to smooth the measured images.
Selection Description
None No smoothing.
Weak* Smoothing is performed at a low level.
Strong Smoothing is performed at a high level.
The asterisk (*) indicates the default setting.
(Noise)
Other than images with a setting pixel will be regarded as noise and excluded from measurement.
Selection
0pixel, 1pixel, 2pixel, 3pixel, 4pixel*, 5pixel, 6pixel, 7pixel,
The asterisk (*) indicates the default setting.
Operation Manual
Step
3
Step
4
Step
5
Step
6
Step
7
Z500
97
Page 100
SECTION 4
Menus for Expert Menu
SECTION 4
Measurement Level
(Level)
Change this when workpiece material fluctuation is too extensive to assure
correct measurement. Set the measurement level of the measurement height (a
percentage of the peak level of the line brightness). The default setting is 50.0%.
Normally, use this setting.
Selection
The intermediate point between the measurement level edges provides the height measurement result.
The peak measurement position can be adjusted by raising the measurement level.
12.5%, 25.0%, 37.5%, 50.0%*, 62.5%, 75.0%, 87.5%,
The asterisk (*) indicates the default setting.
Line Brightness
Peak level
Measurement level (i.e., Peak level Measurement level (%))
Set this value.
Measurement
Measurement level (50 %)
level (75 %)
Clearing Measurement Items
Error data
(Error data)
In cases where the light density is temporarily excessive or insufficient (i.e.,
measurement is not possible) due to defects or holes in the workpiece, it is
possible to set the Z500 to hold the previous data.
Selection Description
Clamp* An error is displayed and is output.
Hold
The value measured immediately before measurement became impossible is displayed and output.
The asterisk (*) indicates the default setting.
Operation with Clamp Setting
The parts in the waveform corresponding to the holes indicate measurement failure.
Measurement value
Operation with Hold Setting
Time
98
Z500
Operation Manual
The previous data is held for parts where measurement is not possible.
Measurement value
Time