BEIJING OFFICE
Room 1028, Office Building,
Beijing Capital Times Square,
No.88 West Chang'an Road,
Beijing, 100031 China
Tel:(86)10-8391-3005 / Fax:(86)10-8391-3688
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in
any form, or by any means, mechanical, electronic, photocopying, recording, or otherwise, without the prior written
permission of OMRON.
No patent liability is assumed with respect to the use of the information contained herein. Moreover, because
OMRON is constantly striving to improve its high-quality products, the information contained in this manual is subject
to change without notice. Every precaution has been taken in the preparation of this manual. Nevertheless, OMRON
assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use
of the information contained in this publication.
Index .................................................................................................. 187
4
Z500
Operation Manual
Page 7
INTRODUCTION
Precaution on Safety
OMRON products are manufactured for use according to proper procedures by a qualified operator and
only for the purposes described in this manual.
The following conventions are used to indicate and classify precautions in this manual. Always heed the
information provided with them. Failure to heed precautions can result in injury to people or damage to
property.
INTRODUCTION
Precaution on Safety
DANGER
WARNING
CAUTION
WARNING Failure to read and understand the information provided in this manual may result in
Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury, or property damage.
personal injury or death, damage to the product, or product failure. Please read
each section in its entirety and be sure you understand the information provided in
the section and related sections before attempting any of the procedures or opera-
tions given.
Operation Manual
Z500
5
Page 8
INTRODUCTION
Menus
INTRODUCTION
Menus
Menus for Conversational Menu
Scn
SetStart setting
P.125
P.40
(0 to 15)
P.40
Run
Tools
System
Save
P.51
P.128
P.144
P.37
Meas method
Measurement
region
Sensitivity
Meas set
(OUT0 to OUT7)
Conditions
Output contents
End setting
Test
Surrounding
image
Backup
Comm
Slant correct
Initialize
Version
P.41
P.43
P.44
P.48
P.49
P.50
P.52
P.129
P.135
P.136
P.145
P.146Environment
P.148Download
P.149
P.151
P.152
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
Item and hierarchy that can be set
through only the adjustment menu
Sensitivity
Details
Conditions
Output
P.70
P.72
*1
P.70
*1
P.74
*1
P.70
*1
P.70
*1
P.70
*1
P.76
*1
*
1:The same construction
as that for
P.59
P.63
P.88
P.100
When
Define
setting of
Height
has been selected, however,
Calibrate
Change
Calibrate
Output
Delete
is disabled.
# to avg
Trigger
Details
Analog
P.86
P.78
P.84
P.87
P.88
P.90
P.96
P.100
6
Z500
Operation Manual
Display
P.112
Terminals
RS-232CP.105
*2:The hierarchy under
*2
Display Mode varies
depending on the type
of screen being displayed.
Adjustment Menu p.28
P.103
Page 9
Menus for Expert Menu
INTRODUCTION
INTRODUCTION
Scn
Set
P.125
P.54
(0 to15)
Image
Compensatn
Meas set
(OUT0 to OUT7)
Conditions
P.56
P.64
P.70
P.88
Meas method
Region
Sensitivity
Details
Reference
Compensation
method
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
# to avg
P.56
P.58
P.59
P.63
P.65
P.69
P.70
P.72
P.70
P.74
P.70
P.70
P.70
P.76
P.88
Change
*1
*1
*1
*1
*1
*1
*1
Calibrate
Output
Delete
*1:The same construction
as that for
Define
When
selected, however,
setting of
disabled.
P.86
P.78
P.84
P.87
Height
has been
Calibrate
Menus
is
Run
Tools
System
Save
P.51
P.128
P.144
P.37
Output
Test
Surrounding
image
Backup
Comm
Slant correct
Initialize
Version
P.100
P.129
P.135
P.136
P.145
P.146Environment
P.148Download
P.149
P.151
P.152
Trigger
Details
Analog
Terminals
RS-232C
Item and hierarchy that can be set
through only the adjustment menu
Display
P.90
P.96
P.100
P.103
P.105
P.112
*2
*2:The hierarchy under Display Mode
varies depending on the type of screen
being displayed.
Adjustment Menu p.28
Operation Manual
Z500
7
Page 10
INTRODUCTION
Editor’s Note
INTRODUCTION
Editor’s Note
Visual Aids
The following headings appear in the left column of the manual to help you
locate different types of information.
NOTICE
CHECK
REFERENCE
TwoSensor
HELPIndicates information helpful in operation.
Indicates information required to take full advantage of the functions and
performance of the product. Incorrect application methods may result in the loss
of damage or damage to the product. Read and follow all precautionary
information.
Indicates points that are important in using product functions or in application
procedures.
Indicates where to find related information.
Indicates information required when using 2 sensors.
Notation
Screen Messages
In this manual, screen message are given in bold/italic.
E.g.: SystemSystem/Backup means the Menu is hierarchical.
OMRON Product References
All OMRON products are capitalized in this manual. The world "Unit" is also
capitalized when it refers to an OMRON product, regardless of whether or not it
appears in the proper name of the product.
8
Z500
Operation Manual
Page 11
Z500
Operation Manual
SECTION 1
Features
SECTION 1 explains the features of the Z500 and the flow of operation using the
Conversational and Expert Menus.
1-1 Features of the Z50010
1-2 Operational Flow14
SECTION 1Features
9
Page 12
SECTION 1
Features
1-1 Features of the Z500
1-1-1 Eight Measurement Items for Various Detection Needs
SECTION 1
Select the measurement item according to application.
Complicated detection is also possible, by combining multiple measurement
items and with detailed settings.
Simultaneous measurement can be performed at 8 points.
1-1 Features of the Z500
REFERENCE
Refer to page 70.
•
Height
•
Edge position
•
Peak/bottom
1-1-2 Simple Setup Using Menus
The Z500 has 2 types of menu. Set the measurement contents according to the
menus displayed on the monitor screen.
Menus for Conversational Menu
Settings can be made easily by entering information as requested - just as
though you are having a conversation with the Z500.
REFERENCE
Refer to page 39.
Menus for Expert Menu
More detailed setting, such as position compensation for workpieces placed at
different positions, can be made.
•
Step : 2 pts
•
Width
•
Define
•
Step : 3 pts
•
Edge center
REFERENCE
Refer to page 53.
1-1-3 A Variety of Output Formats
Not only is analog output available, but data output to external devices can also
be performed via terminal blocks or RS-232C.
REFERENCE
Refer to page 100.
Z500
Operation Manual
10
Page 13
1-1-4 Monitoring While Viewing the Measurement Status
The Z500 has 4 types of monitor. Monitoring suited to the application is possible
at every stage from installation and measurement settings to operation and
maintenance.
REFERENCE
Refer to page 29.
SECTION 1
Features
SECTION 1
1-1 Features of the Z500
Image Monitor
This is the main monitor of the Z500.
Three pieces of information;
measurement status, sensitivity, and
measurement result, can be checked
together with their target images.
Scn 0 Run
NEAR
OUT 0
BRIGHT
+0005.18800
S+ / :Display S+ENT:Adjust
SHIFT
DARK
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
LV
PEAK
Zero's OFF
PASSmm
SHIFT
FAR
[ .]
[ ]
+
Profile Monitor
Chronological change of height
distribution of 126 points (profile)
can be checked on a shaded 3-D
image.
Scn 0 Run
1div x: 1.000 [s]
y:+10.0000[mm]
Thru
+0005.18800
Sen 0
S+ / :Display S+ENT:Adjust
mm
9.96ms
PASS
SHIFT
SHIFT
SHIFT
SHIFT
Digital Monitor
Two or more measurement results
can be checked at the same time.
Scn 0 Run
+0002.65432mm
0
+
+
PASS
+0004.65432mm
1
PASS
+0000.00000mm
2
ERROR
-----------mm
3
S+ / :Display S+ENT:Adjust
+
SHIFT
9.96ms
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
SHIFT
++
Trend Monitor
Chronological change of measured
values can be checked.
Scn 0 Run
+
Thru
+
OUT 0
S+ / :Display S+ENT:Adjust
9.96ms
1div x: 1.000 [s]
y:+10.0000[mm]
Zero's OFF
mm+0005.18800
PA S S
Operation Manual
Z500
11
Page 14
SECTION 1
Features
1-1-5 Flexible Settings for Data Output to External Devices
In addition to simply outputting measurement results, equations can be set to
output to external devices using measurement and judgment results.
SECTION 1
1-1 Features of the Z500
1-1-6 A Wide Variety of Useful Functions
Any Type of Workpiece can be Measured by Using the Image Adjustment Function
REFERENCE
Refer to page 70.
Z500
Higher measurement accuracy can be assured by using the sensitivity
adjustment function, even if the workpieces to be measured have different
brightness.
REFERENCE
Refer to pages 44 and 59.
Adjustment Menu
Six types of adjustments are available on shortcut keys by displaying the
adjustment menu.
REFERENCE
Refer to page 28.
Checking Surrounding Images of the Area that the Line Beam Strikes
Displays images of the area surrounding the part that the line beam strikes.
This function is useful when the Sensor is mounted inside an installation and the
measurement area cannot be checked directly.
REFERENCE
Refer to page 135.
Z500
Operation Manual
12
Page 15
Specifying Measurement Timing by Using the Trigger Function
Measurement timing can be specified arbitrarily by using the self-trigger and
external trigger functions. Characteristic quantity can be held while the
measurement is carried out.
(Measurement status at a set timing can be checked on the monitor screen.)
REFERENCE
SECTION 1
Features
SECTION 1
Refer to page 90.
Smooth Setup Changes
Scenes can be used to change between up to 16 types of measurement setups.
Simply change the scene to change to a different measurement setup.
Refer to page 125.
1-1 Features of the Z500
REFERENCE
Operation Manual
Z500
13
Page 16
SECTION 1
Features
1-2 Operational Flow
1-2-1 Menus for Conversational Menu
SECTION 1
1-2 Operational Flow
Turn ON the power.
Displaying Images on the Monitor
Refer to page 20,
Starting the Z500 and Displaying Images.
2-1
Preparations for Detection
Refer to this section to learn about the basic
operational flow from setting measurement
conditions to executing measurements.
CHECK
Refer to page 39 for information on
settings using Conversational Menu.
Starting Up Conversational Menu and
Entering Set Mode
Refer to page 40, See
Step 1 : Starting Conversational Menus.
Setting Basic Measurement Conditions
Refer to page 41, See
Step 2 :
Setting Basic Measurement Conditions.
Selecting and Setting Measurement Items
Refer to page 48, See
Step 3 : Setting Measurement Contents.
Refer to page 70 for details.
Screen Types and Display Methods
Refer to page 29.
Selecting Measurement Item
Appropriate to the
Measurement Goal
Refer to page 70.
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Define
14
Setting Measurement Conditions (if necessary)
Refer to page 49, See
Step 4 :
Setting Measurement Conditions.
Refer to page 88 for details.
Setting Output Contents
Refer to page 50, See
Setting Detection Conditions and Executing
Step 5 : Setting Output Contents.
Refer to page 100 for details.
Starting Measurements
Refer to page 51, See
Step 6 : Starting Measurements.
Performing Test
Refer to page 129,
Testing Measurement Performance (Test).
5-6
Confirming Settings
Z500
Operation Manual
Checking the Line Beam Position
Refer to page 135,
Beam Position (Surrounding image).
5-7Checking the Line
Page 17
Troubleshooting
When an error message has
been displayed on the screen,
SECTION 2 shows basic menu operations of the Z500.
2-1 Starting the Z500 and Displaying Images20
2-2 Menu Operations23
2-2-1 Input Device23
2-2-2 Screen Displays24
2-2-3 Menu Tree25
2-2-4 Inputting Values27
2-2-5 Adjustment Menu28
2-3 Screen Types and Display Methods29
SECTION 2Basic Operations
2-3-1 Image Monitor30
2-3-2 Profile Monitor33
2-3-3 Digital Monitor35
2-3-4 Trend Monitor36
2-4 Saving Settings and Exiting the Z50037
19
Page 22
SECTION 2
Basic Operations
2-1 Starting the Z500 and Displaying Images
This section describes how to start the Z500 and how to display images on the monitor.
Place the workpiece and adjust the Sensor position while monitoring the displayed image.
Starting the Z500
#1
SECTION 2
2-1 Starting the Z500 and Displaying Images
1.
CHECK
2.
3.
Be sure that the basic Z500 components have been connected correctly.
Power Supply
Console
ENT
ESC TRIG
SHIFT
CONSOLE
Sensor
Z500
××× × × × × × × × × ×
××× × × × ××
× × × × × × ×
× × × × × × × × × ×
× × ×
× ×
Monitor
Before connecting components or wiring power supply lines and grounding wires, be sure
to refer to the relevant section in the
REFERENCE
Setup Manual
.
Refer to page 16 in the Setup Manual.
Turn ON the power supply to the monitor.
Turn ON the power supply to the Z500.
The Image Monitor will be displayed.
CHECK
Z500
Operation Manual
20
Scn 0 Run
NEAR
OUT 0
+0005.18800
S+ / :Display S+ENT:Adjust
If the Image Monitor is not displayed, use the
SHIFT
on the console to change the Monitors.
REFERENCE
Refer to page 29, Screen Types and Display Methods.
+
Right
Keys or
BRIGHT
DARK
SHIFT
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PA S Smm
Left
+
Keys
FAR
Page 23
Position the Workpiece and Adjust the Image
#2
Position the workpiece to be measured and adjust the displayed image in the
measurement range.
The measurement range varies depending on the Sensor models connected.
REFERENCE
Refer to page 26 in the Setup Manual.
Example: Z500-SW17
Sensor
100mm
SECTION 2
Basic Operations
SECTION 2
+20mm
2-1 Starting the Z500 and Displaying Images
0mm
Height
1.
Measurement range
+20mm
0mm
-20mm
Position
-20mm
Position the workpiece.
In this example, the workpiece shown in
the right figure will be described.
14.5mm
(±7.25mm)
17.3mm
(±8.65mm)
20.1mm
(±10.05mm)
Operation Manual
Z500
21
Page 24
SECTION 2
Basic Operations
A profile graph showing the level differences in the workpiece will be displayed.
SECTION 2
2-1 Starting the Z500 and Displaying Images
2.
CHECK
Adjust the position of the workpiece if
necessary.
Adjust the workpiece position if the
Profile graph is not appeared in the
center of the monitor.
Scn 0 Run
NEAR
BRIGHT
DARK
OUT 0
+0005.18800
Profile
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PA S Smm
FAR
S+ / :Display S+ENT:Adjust
If the lines representing level differences have swollen elliptically, the light density is
excessive. If the lines are too thin to see, the light density is insufficient. In either case,
adjust the sensitivity.
REFERENCE
Refer to page 59 for information on how to adjust sensitivity.
22
Z500
Operation Manual
Page 25
2-2 Menu Operations
2-2-1 Input Device
Menu operations are performed on the Console.
SECTION 2
Basic Operations
Trigger Key
TRIG
Escape Key
ESC
SHIFT Key
SHIFT
KeyFunction
Escape KeyReturns the cursor to the previous menu display or operation.
Trigger KeyThe effect of the Trigger Key varies with the function.
Enter Key
SHIFT Key
Up, Down, Left, and
Right Keys
ESCTRIG
SHIFT
Executes a function or sets a value.
Note: On the F160-KP, also functions as a Cursor Key.
Must be pressed in combination with another key to have any
effect.
Specific functions are assigned to combinations of the
Key and other keys for specific screens.
The Up and
down and also to set values.
Use the Up Key to increase a value by 1.
Left
Down
and
Use the
Hold down the Up or
a value.
The
right.
ENT
Z300-KP
CONSOLE
Down
Keys are used to move the cursor up and
Key to decrease a value by 1.
Right
Enter Key
ENT
Up, Down, Left, and Right Keys
Down
Key to quickly increase or decrease
Keys are used to move the cursor left or
SHIFT
SECTION 2
2-2 Menu Operations
Operation Manual
Z500
23
Page 26
SECTION 2
2-2 Menu Operations
SECTION 2
Basic Operations
2-2-2 Screen Displays
The Z500 is operated by selecting functions displayed on the screen. Familiarize yourself with each
function before operating the Z500.
Scene Number
There are 16 scenes.
Setting different measurement
conditions for each one allows
easy switching between different
setups.
Cursor Keys
The cursor is moved to the
desired function by pressing
the Up, Down, Left,
and Right Keys.
Scn 0 Run
S+ / :Display S+ENT:Adjust
Key Operations
Displays special key combinations at the bottom of the screen
where available. S refers to the SHIFT Key. "S + ENT" indicates
that the ENT Key should be pressed while the SHIFT Key is pressed.
Mode (See the table below)
The current operating mode is displayed.
9.96ms
Measurement period (Sampling time)
It varies depending on the item
to be measured.
Mode
DisplayDescription
SetMode to set the inspection conditions.
Performs measurement.
Run
ToolsUsed to save settings and images to a computer as backup.
System
Save
The measurement results are output to an external device via terminal
block or RS-233C.
Used to set system conditions for the Z500. Select this mode to switch
menus and settings for communications with external devices.
Used to save data to flash memory in the Z500.
If new settings have been made, be sure to save the data before quitting.
24
Z500
Operation Manual
Page 27
SECTION 2
Basic Operations
2-2-3 Menu Tree
Z500 menus are hierarchical. The cursor is moved to the required functions to set measurement
conditions. Use the following procedures to move around the menu tree.
1.
2.
In this example, the cursor will move to the Image Menu.
3.
Move the cursor to the item to be moved.
(Move the cursor to the desired function
in Set Mode using the Up and Down
Keys.)
Press the ENT Key.
Repeat (1.) and (2.) to move the cursor
to lower levels.
Press the ESC Key once to move to the
upper level.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
SECTION 2
2-2 Menu Operations
Meas method
Region
Sensitivity
Details
Operation Manual
Z500
25
Page 28
SECTION 2
2-2 Menu Operations
SECTION 2
Basic Operations
Triangle Mark
Items with an inverted triangle after them have a list of alternatives.The method for selecting the
alternatives is given here.
Scn 0 System
:
1.
2.
Move the cursor to the item to be set.
In this example, move the cursor to
Baud rate.
Press the ENT Key.
Baud rate
Data Length
Parity
Stop bits
Delimiter
Flow
38400bps
:
:
:
:
:
End
8bit
None
1bit
CR
None
The selections will be displayed.
3.
Move the cursor to the desired baud rate
using the Up and Down Keys.
4.
Press the ENT Key.
The selections will be registered.
Scn 0 System
Baud rate
Data Length
Parity
Stop bits
Delimiter
Flow
Scn 0 System
Baud rate
Data Length
Parity
Stop bits
Delimiter
Flow
:
38400bps
:
2400bps
:
4800bps
:
9600bps
:
19200bps
:
38400bps
End
:
38400bps
:
:
:
:
:
End
8bit
None
1bit
CR
None
8bit
None
1bit
CR
None
26
Z500
Operation Manual
Page 29
SECTION 2
Basic Operations
2-2-4 Inputting Values
The section explains how to input values when setting judgment or measurement conditions.
1.
Move the cursor to the item for which a
value is to be changed.
2.
Press the ENT Key.
The cursor will change to a cursor the size of a single digit.
3.
Move the cursor to the digit to be
changed using the Left and Right Keys.
4.
Change the value.
Use the Up Key to increase the value.
Use the Down Key to decrease the
value.
Press the SHIFT + ENT Keys in Run Mode. The adjustment menu will be displayed.
The adjustment menu is useful to display the set screens of the following six items without passing
through the menu hierarchies.
SECTION 2
2-2 Menu Operations
CHECK
Scn 0 Run
ESCTRGENT
Display
Sensitvy
Output
Conditns
ESC:Back S+ / :Switch display
KeyAllocationReference
Trigger KeyDisplayP.112
Enter KeyTestP.129
Up KeySensitivityP.59
Down KeyConditionsP.88
Left KeyOutputP.100
Right KeyMeas setP.70
Escape KeyNone -
The adjustment menu can be performed from the Image Monitor, the Digital Monitor, the
Profile Monitor, or the Trend Monitor.
9.96ms
Test
Meas set
28
Z500
Operation Manual
Page 31
SECTION 2
Basic Operations
2-3 Screen Types and Display Methods
There are 4 Monitors to display images on the Z500; Image Monitor, Digital Monitor, Trend Monitor, and
Profile Monitor. These monitors are useful to check measurement information as required from various
viewpoints. Use the SHIFT + Right Keys or SHIFT + Left Keys on the console to change the Monitors.
The first time the power supply is turned ON after delivery, the Image Monitor is displayed.
Image Monitor
This is the main monitor of the Z500.
Three pieces of information;
measurement status, sensitivity, and
measurement result, can be checked
together with their target images.
Scn 0 Run
NEAR
OUT 0
BRIGHT
+0005.18800
S+ / :Display S+ENT:Adjust
SHIFT
DARK
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
LV
PEAK
Zero's OFF
PASSmm
SHIFT
FAR
[ .]
[ ]
+
Profile Monitor
Chronological change of height
distribution of 126 points (profile)
can be checked on a shaded 3-D
image.
Scn 0 Run
1div x: 1.000 [s]
Thru
Sen 0
S+ / :Display S+ENT:Adjust
y:+10.0000[mm]
+0005.18800
mm
9.96ms
PASS
SHIFT
SHIFT
SHIFT
SHIFT
Digital Monitor
Two or more measurement results
can be checked at the same time.
Scn 0 Run
+0002.65432mm
0
+
+
PASS
+0004.65432mm
1
PASS
+0000.00000mm
2
ERROR
-----------mm
3
S+ / :Display S+ENT:Adjust
+
SHIFT
9.96ms
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
SHIFT
++
SECTION 2
2-3 Screen Types and Display Methods
Trend Monitor
Chronological change of measured
values can be checked.
Scn 0 Run
+
Thru
+
OUT 0
S+ / :Display S+ENT:Adjust
9.96ms
1div x: 1.000 [s]
y:+10.0000[mm]
Zero's OFF
mm+0005.18800
PA S S
CHECK
CHECK
Enable to select which Monitor will be displayed when the Z500 turns ON.
REFERENCE
Refer to page 146.
Measurement continues even while the screen is switched.
Operation Manual
Z500
29
Page 32
SECTION 2
Basic Operations
2-3-1 Image Monitor
This is the main monitor of the Z500. The first time the power supply is turned ON after delivery, this
screen is displayed. Mostly used to check the measurement status, sensitivity information, and
measurement results.
Image Monitor consists of 2 displays; Measurement status display and Measurement sensitivity display.
Use the TRIG Key to change the displays.
SECTION 2
2-3 Screen Types and Display Methods
Measurement Status Display
Scn 0 Run
NEAR
OUT 0
BRIGHT
DARK
Sen0
Sen1
Zero's OFF
+0005.18800
S+ / :Display S+ENT:Adjust
Note:
Measurement Status Display
Common to Measurement status display and Measurement sensitivity display.
The height information within the measurement range can be checked.
The height information between 0 and 125 lines is displayed with the Profile
graph.
Sensor
Line beam
9.96ms
LV
PEAK
LV
PEAK
PA S Smm
[ 15]
[ 170]
Measurement Sensitivity Display
FAR
Screen Displays
NEAR
BRIGHT
TRIG
[ .]
Scn 0 Run
[ ]
Measurement
Result Display
(See note)
OUT 0
+0005.18800
DARK
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PA S Smm
FAR
S+ / :Display S+ENT:Adjust
Workpiece
Measurement range
Measurement center
30
Z500
Operation Manual
Line 0
Position
Line 125
Height
Scn 0 Run
BRIGHT
DARK
OUT 0
+0005.18800
9.96ms
Sen0
LV
PEAK
Sen1
LV
PEAK
Zero's OFF
PA S Smm
S+ / :Display S+ENT:Adjust
LowHigh
Measurement region (green)
Profile (yellow)
FARNEAR
[ 15]
[ 170]
[ .]
[ ]
Page 33
SECTION 2
Basic Operations
CHECK
The Profile graph can be enlarged.
REFERENCE
Refer to page 113.
CHECK
The Profile graph can be output as a measurement value to an external device.
REFERENCE
Refer to page 101.
CHECK
Whether or not the lines interfering with the measurement should be interpolated can be
determined.
REFERENCE
Refer to page 99.
CHECK
The measurement region can be changed.
REFERENCE
Refer to page 58.
Measurement Sensitivity Display
The Measurement sensitivity level (LV) and received level (PEAK) can be
checked. The light density and the status of the workpiece surface at a cursor
position are displayed with the line brightness graph.
SECTION 2
2-3 Screen Types and Display Methods
Bright
Brightness
Dark
CHECK
Cursor (Red)
Refer to page 116 for information on how to
change cursor position.
High
Scn 0 Run
OUT 0
Height
BRIGHT
DARK
+0005.18800
S+ / :Display S+ENT:Adjust
Line Brightness (Brightness Distribution)
The line brightness graph at the cursor position is displayed.
It is possible to know at a glance whether or not, for example,
the light density is insufficient, or is at the saturation level.
Low
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
LV
PEAK
Zero's OFF
PASSmm
Measurement period (Sampling time)
Line beam image
The shape of workpieces and
FARNEAR
[ .]
[ ]
the way they are detected
can be checked.
LV [ 15]
PEAK[170]
LV: Displays current sensitivity level. (1 to 31)
The value increases with increase in darkness of the
workpiece, while it decreases with increase in
brightness of the workpiece.
PEAK: Displays current received level.
Use this value as reference when setting
the sensitivity level.
120 max.: Insufficient light density
255 min.: Excessive light density
For details on sensitivity adjustment method,
refer to page 59.
The line brightness graph for the specified range can be enlarged.
REFERENCE
Refer to page 113.
Operation Manual
Z500
31
Page 34
SECTION 2
2-3 Screen Types and Display Methods
SECTION 2
Basic Operations
Measurement Result Display
Check whether the measurements are correctly performed.
The measurement points and the measurement values are displayed.
Scn 0 Run
BRIGHT
DARK
OUT 0
+0005.18800
S+ / :Display S+ENT:Adjust
Measurement value
The measurement value is displayed
in green if the judgment result is PASS.
It is displayed in red if the judgment
result is HIGH, LOW, or ERROR.
Output number
The output number for the currently displayed
measurement result is displayed.
9.96ms
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Zero's OFF
PASSmm
Measurement points (red)
The measured points are displayed.
No point is displayed when the measurement is in error.
The level and range necessary for the determination of
FARNEAR
measurement points are represented by thin lines.
(In this example, the distance between the 2 points is
measured.)
Forced-zero
Zero's OFF: Forced-zero not executed
Zero's ON: Forced-zero executed
Judgment results
HIGH : Greater than the judgment upper limit
PASS : Within the judgment limits
LOW : Less than the judgment lower limit
ERROR : Measurement error
-The workpiece is out of the measurement range
-The light density is insufficient
-Information on the measurement point
cannot be obtained, or other cases
RESET: No-measurement state
LD-OFF: LD-OFF input
CHECK
Offset can be set for Force-zero.
REFERENCE
Refer to page 84.
32
Z500
Operation Manual
Page 35
SECTION 2
Basic Operations
2-3-2 Profile Monitor
Use to check the status of the workpiece surface.
Surface condition of a workpiece can be monitored on a shaded 3-D image screen. Height information
and its chronological change can also be monitored.
9.96ms
PA S S
Profile Display
Measurement Result Display
Trend Display
Shaded 3-D image display
Scn 0 Run
1div x: 1.000 [s]
y:+10.0000[mm]
Thru
Sen 0
S+ / :Display S+ENT:Adjust
+0005.18800
mm
By moving the workpiece, the profile graph (height information for 126 points) can be monitored in
chronological order for up to 10 seconds.
TRIG Key can be used for switching from waveform display to waveform display stop and vise versa.
The trigger function enables shaded 3-D image to be displayed at specified timing.
Line beam
Movement orientation
SECTION 2
2-3 Screen Types and Display Methods
CHECK
Display range and display time can be changed.
REFERENCE
Refer to page 121.
Operation Manual
Z500
33
Page 36
SECTION 2
Basic Operations
Shaded 3-D Image Display and Profile Display
Surface condition of a workpiece is displayed by a shaded 3-D image with 0 to
255 gradations. A low section is displayed darkly, while a higher section is
displayed brightly.
SECTION 2
2-3 Screen Types and Display Methods
Shaded 3-D Image
Scn 0 Run
Thru
Sen 0
S+ / :Display S+ENT:Adjust
Profile Cursor (Red)
1div x: 1.000 [s]
y:+10.0000[mm]
+0005.18800
mm
9.96ms
PA S S
Waveform display status
During waveform display, (green) is displayed.
While the waveform is stopped, (red) is displayed.
Trend Display and Measurement Result Display
Trend Cursor
Trend Display
The waveform displayed in chronological
order at the Trend cursor position is displayed.
High
Height
Low
Time(s)
Screen Mode
Thru: continuous display
Trig: displayed in trigger mode
Sensor number
The current sensor number is displayed.
Scn 0 Run
Thru
Sen 0
S+ / :Display S+ENT:Adjust
+0005.18800
Profile Display
The height information at the
Profile cursor position is displayed.
Line 0
Position
Line 125
Height
1div x: 1.000 [s]
y:+10.0000[mm]
Measurement
value
High
mm
Low
9.96ms
PA S S
Conversion values for X and Y axes
The units corresponding to 1
division (1 div) on the waveform
display are displayed. If the
setting is changed, the display
also changes. Use these
values as reference when
displaying the waveform.
y
1div
x
Judgment result
1div x: 1.000
y:+10.000
34
CHECK
Z500
Operation Manual
The cursor position can be changed.
REFERENCE
Refer to page 121.
Page 37
SECTION 2
Basic Operations
2-3-3 Digital Monitor
Mostly used during line operation.
Two or more measurement results that have been set for the output numbers of the displayed scenes will
be displayed.
Output number
Press the
Key to switch the number.
TRIG
Scn 0 Run
+0002.65432mm
0
PASS
+0004.65432mm
1
PASS
+0000.00000mm
2
ERROR
-----------mm
3
S+ / :Display S+ENT:Adjust
Scn 0 Run
+0002.65432mm
4
PASS
+0004.65432mm
5
PASS
+0000.00000mm
6
ERROR
-----------mm
7
S+ / :Display S+ENT:Adjust
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
HI=+0010.00000
LO=-0010.00000
9.96ms
TRIGTRIG
9.96ms
Judgment value
The judgment value for each
output number will be displayed.
HI: Judgment upper limit
LO: Judgment lower limit
Measurement value
Judgment result
---- will be displayed for output numbers
that have not been set.
SECTION 2
2-3 Screen Types and Display Methods
CHECK
The measurement items can also be displayed instead of the judgment values.
REFERENCE
Refer to page 117.
Scn 0 Run
0
+0002.65432mm
PASS
1
+0004.65432mm
PASS
2
+0000.00000mm
ERROR
3
-----------mm
S+ / :Display S+ENT:Adjust
9.96ms
Step: 2 pts
Width
Define
Measurement items
Operation Manual
Z500
35
Page 38
SECTION 2
2-3 Screen Types and Display Methods
SECTION 2
Basic Operations
2-3-4 Trend Monitor
Mostly used for setting and checking the Trigger function.
By using the Trend Monitor, the measurement results to be monitored can be checked in chronological
order for up to 30 seconds. Press the TRIG Key to switch from waveform display to waveform display
stop and vise versa. The trigger function enables waveform to be displayed at specified timing.
Conversion values
for X and Y axes
Scn 0 Run
1div x: 1.000 [s]
y:+10.0000[mm]
Thru
OUT 0
S+ / :Display S+ENT:Adjust
Measurement
value
9.96ms
Zero's OFF
mm+0005.18800
PASS
Judgment result
The units corresponding to 1 division (1 div) on the
waveform display are displayed. If the setting is
changed, the display also changes. Use these
values as reference when displaying the waveform.
y
1div
x
Measurement period (Sampling time)
Waveform display status
During waveform display, (green) is displayed.
While the waveform is stopped, (red) is displayed.
Trigger waveform
The trigger status is displayed.
Force-zero
Zero's OFF: Force-zero not executed
Zero's ON: Force-zero executed
1div x: 1.000
y:+10.000
36
Output number
The current output number is displayed.
Screen Mode
Thru: The waveform obtained by the Sensor is displayed continuously.
Trig: A still image of the waveform based on the trigger settings is displayed.
Thin lines are used to represent the measurement start and measurement end timings.
CHECK
Display range and display time can be changed. (Only the display range can be changed
for each output number.)
REFERENCE
Refer to page 118.
CHECK
Characteristic quantities (such as average value and peak value being held) at specified
timing can be output.
REFERENCE
Refer to page 84.
Z500
Operation Manual
Page 39
SECTION 2
Basic Operations
2-4 Saving Settings and Exiting the Z500
When setting have been changed, be sure to save to flash memory before turning the power OFF.
NOTICE
1.
2.
A list of modes will be displayed.
3.
Flash memory data is loaded each time the Z500 is started up. Therefore, when setting
have been changed, be sure to save to flash memory before turning the power OFF. If the
power is turned OFF without saving, all of the setting changes will be lost.
Display the screen for Run Mode.
Move the cursor to Run and press the
ENT Key.
Select Save.
SECTION 2
Scn 0 Run
2-4 Saving Settings and Exiting the Z500
Scn 0 Run
Set
Run
Tools
System
Save
A confirmation message will be displayed.
4.
NOTICE
When the saving has been completed, the screen in (3.) will return.
5.
Select Execute.
Do not turn OFF the power or input a RESET signal while a message is being displayed in
any save or load operation. Data in memory will be destroyed, and the Z500 may not
operate correctly the next time it is started.
Turn OFF the power supply and exit the Z500.
The data will be saved.
OK?
ExecuteCancel
Operation Manual
Z500
37
Page 40
SECTION 2
2-4 Saving Settings and Exiting the Z500
SECTION 2
Basic Operations
MEMO
38
Z500
Operation Manual
Page 41
Z500
Operation Manual
SECTION 3
Menus for Conversational Menu
SECTION 3 explains how to set measurement conditions using the Conversational
Menu.
3-1
Setting Measurement Conditions Using the Conversational Menus
1 Starting Conversational Menus40
Step
2 Setting Basic Measurement Conditions41
Step
3 Setting Measurement Contents48
Step
4 Setting Measurement Conditions49
Step
5 Setting Output Contents50
Step
6 Starting Measurements51
Step
3-2 Saving Settings and Exiting the Z50052
SECTION 3Menus for Conversational Menu
40
39
Page 42
SECTION 3
Menus for Conversational Menu
3-1
Setting Measurement Conditions Using the Conversational Menus
This section explains the procedures for using Conversational Menus, from setting measurement
conditions through to starting measurement.
The operating guides displayed on the screen can be used just like a conversation to help perform
procedures.
Step
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
The mode selections will be displayed.
The Z500 will enter Set Mode.
Starting Conversational Menus
1
1.
The screen shown right will be displayed
when the Conversational Menus are
started for the first time.
2.
Move the cursor to Run and press the
ENT Key.
3.
Use the Up and Down Keys to move the
cursor and select Set.
4.
Select Next to start making the settings.
Scn 0 Run
BRIGHT
DARK
OUT 0
9.96ms
Sen0
LV
PEAK
Sen1
LV
PEAK
Zero's OFF
+0005.18800
S+ / :Display S+ENT:Adjust
Scn 0 Run
Set
Run
Tools
System
Save
Perform settings in
the following order.
STEP 1:Measurement method
STEP 2:Measurement region
STEP 3:Sensitivty adj method
Next
Back
[ 15]
[ 170]
PASSmm
FARNEAR
[ .]
[ ]
Follow the instructions given by the operation guide displayed on the screen for the rest of the procedure.
Two Sensor
The selections for the sensors to be used for measurement will be displayed.
Select the Sensor number.
Z500
Operation Manual
40
Page 43
SECTION 3
Menus for Conversational Menu
Step
Basic settings required for measurement are described below.
#1
Specify the region to be measured.
Select Zoom to assure higher position measurement accuracy.
Normally, use with the default setting (Wide).
Setting Basic Measurement Conditions
2
#1
Page 41
Select the Measurement
Method
#2
Select the Measurement Method
SelectionDescription
Wide*
Zoom
Measures the whole measurement range of the connected
sensor.
Enlarges and measures a part of the measurement range of the
connected sensor.
Select this function to assure higher position measurement
accuracy.
Page 43
Set the Measurement
Region
#3
Page 44
Select the Sensitivity
Adjustment Method
The asterisk (*) indicates the default setting.
SECTION 3
Step
1
Step
2
Step
3
Operation Manual6
Z500
Step
4
Step
5
Step
6
41
Page 44
SECTION 3
d
Menus for Conversational Menu
Measurement range
Example: Z500-SW17
+20mm
Wide
+20mm
Zoom
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
CHECK
1.
2.
0mm
-20mm
The measurement range varies depending on the Sensor models connected.
REFERENCE
14.5mm
( 7.25mm)
17.3mm
( 8.65mm)
20.1mm
( 10.05mm)
0mm
-20mm
4.8mm
5.7mm
6.7mm
This range is enlarged
three times and displaye
on the monitor screen.
Refer to page 28 in the Setup Manual.
Select the measurement method while
monitoring the image to be measured.
Scn 0 Set
Wide
Zoom
Press the ENT Key.
Method:
Select Zoom to increase
measurement accuracy.
ENT:Next ESC:Back
Go to
#2
Setting the Measurement Region.
.
Go to #2. Setting the Measurement Region.
Z500
Operation Manual
42
Page 45
SECTION 3
Menus for Conversational Menu
#2
Setting the Measurement Region
Specify the region to be measured.
The Z500 adjusts sensitivity and performs measurements in a region that has been previously set.
Though the whole image can be used for the measurement region, measurement accuracy can be
raised by specifying a restricted measurement region.
Position the workpiece correctly and specify the measurement region.
CHECK
Correct measurement may not be possible at the boundaries of the measurement regions.
Leave space at the boundaries when setting the regions.
Profile
1.Position the workpiece correctly while
monitoring the profile graph.
SECTION 3
2.
The screen for specifying the region will be displayed.
3.
The cursor will be moved to the ending line.
4.
Press the ENT Key.
Specify the starting line of the
measurement region.
Repeat the procedure described in (3.) to specify an ending line and press the
ENT Key.
Go to #3. Select the Sensitivity Adjustment Method.
Operation Manual6
Z500
43
Page 46
SECTION 3
)
ght
y
e
ow
Menus for Conversational Menu
#3
Select the sensitivity adjustment method for the measurement range.
If the measurement results fluctuate, change the adjustment method.
Normally, these conditions can be left at the default settings.
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
Select the Sensitivity Adjustment Method
Sensitivity Adjustment
Selection
Adjust the sensitivity according to sensitivity information in the
measurement region.
The sensitivity can be set to one of 31 levels (Setting range: 0 to
31).
Select when it is difficult to assure accurate measurement in
Auto. Such cases may appear on a production line where
workpieces of different colors are processed.
Adjust the sensitivity for each line in the measurement region.
Select when it is difficult to assure accurate measurement in
Auto due to a widely distributed brightness of the workpiece
surfaces.
However, sampling time will be longer.
Position a reference workpiece and execute Multi. The Z500
will set automatically the upper and lower limits of sensitivity on
the basis of the brightness of the reference workpiece.
CHECK
Auto*
Fixed
Multi
Once the sensitivity has been set in Multi, Sensitvy can be performed only through the
adjustment menu.
Description
The asterisk (*) indicates the default setting.
REFERENCE
Refer to page 28 for details on the adjustment menu, and page 63 for details on
Details.
Sensitivity level (Displayed only when Fixed is selected from the
Sensitivity Menu)
Setting range
LV0 (See note
LV31High
Note:The laser will turn OFF when selecting [LV0].
Sensitivit
L
Color of workpiec
Bri
Dark
44
Z500
Operation Manual
Page 47
When selecting Auto
SECTION 3
Menus for Conversational Menu
1.
A confirmation message will be displayed.
2.
Select Auto.
Select Execute when the basic settings
have completed.
Select Settings when continuing the
measurement contents settings.
Step 3.
Go to
Scn 0 Set
Auto
Fixed
Multi
Sen0's sensitivity method:
Select Fixed or Multi if
measurement unstable with Auto.
ENT:Set ESC:Back
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute Cancel Settings
SECTION 3
Step
1
Step
2
Step
3
Operation Manual6
Z500
Step
4
Step
5
Step
6
45
Page 48
SECTION 3
Menus for Conversational Menu
When selecting Fixed
SECTION 3
The screen for setting the sensitivity level will be displayed.
3-1 Setting Measurement Conditions Using the Conversational Menus
1.
2.
Select Fixed.
Select the level using the Up and Down
Keys while monitoring the image on the
monitor.
The level displayed first is the level
selected by the Z500 for the workpiece
being measured. Adjust the level while
monitoring the line brightness graph and
the received level on the monitor.
120 max.: Insufficient light density
255 min.: Excessive light density
Scn 0 Set
Auto
Fixed
Multi
Sen0's sensitivity method:
Select Fixed or Multi if
measurement unstable with Auto.
ENT:Set ESC:Back
BRIGHT
DARK
Received level
Sen0
LV
PEAK
Sen1
LV
PEAK
Line Brightness
LV 15
Sen0's sensitivity method:
Select Fixed or Multi if
measurement unstable with Auto.
ENT:Set ESC:Back
[ 15]
[ 170]
FARNEAR
[ .]
[ ]
3.
Press the ENT Key.
A confirmation message will be displayed.
4.
Select Execute when the basic settings
have completed.
Select Settings when continuing the
measurement contents settings.
Go to
Z500
Operation Manual
46
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute Cancel Settings
Step 3.
Page 49
When selecting Multi
SECTION 3
Menus for Conversational Menu
1.
The sensitivity multiple setting screen will be displayed.
2.
3.
Select Multi.
Check whether or not the workpiece is
positioned inside the measurement
range.
Press the ENT Key.
Scn 0 Set
Auto
Fixed
Multi
Sen0's sensitivity method:
Select Fixed or Multi if
measurement unstable with Auto.
ENT:Set ESC:Back
Line beam image
Scn 0 Set
Sensitivity set for each line.
Set the workpiece correctly
and press the ENT Key.
ENT:Set ESC:Back
SECTION 3
Step
1
Step
2
Step
3
A confirmation message will be displayed.
4.
Select Execute when the basic settings
have completed.
Select Settings when continuing the
measurement contents settings.
Step 3.
Go to
Settings will be registered.
Allocate measurement items
to OUT0 to OUT7
in the measurement settings.
Settings can be accessed
from the Adjustment Menu.
Execute Cancel Settings
Operation Manual6
Step
4
Step
5
Step
6
Z500
47
Page 50
SECTION 3
Menus for Conversational Menu
Step
The Z500 provides eight measurement items. Up to eight measurement items can be set for each scene.
Select the measurement item according to application.
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
3
Setting Measurement Contents
This section gives an outline only. Refer to SECTION 4 (page 70) for details on
measurement contents.
REFERENCE
Refer to page 70.
48
Z500
Operation Manual
Page 51
Menus for Conversational Menu
4
Step
Set the measurement conditions, such as average number of times and trigger setting.
Setting can be performed only through the adjustment menu.
Setting Measurement Conditions
REFERENCE
Refer to page 28, Adjustment Menu.
This section gives an outline only. Refer to SECTION 4 (page 88) for details on
measurement conditions.
REFERENCE
SECTION 3
Refer to page 88.
SECTION 3
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Operation Manual6
Z500
49
Page 52
SECTION 3
Menus for Conversational Menu
Step
The Z500 can transmit data to an external device in three formats; analog, terminal, and RS-232C. The
operation procedures for setting the output contents to the external device will be described in SECTION
4. Setting can be performed only through the adjustment menu.
SECTION 3
3-1 Setting Measurement Conditions Using the Conversational Menus
5
Setting Output Contents
REFERENCE
Refer to page 28, Adjustment Menu.
This section gives an outline only. Refer to SECTION 4 (page 100) for details on
output contents.
REFERENCE
Refer to page 100.
50
Z500
Operation Manual
Page 53
SECTION 3
Menus for Conversational Menu
Step 6
Starting Measurements
Perform the measurement under the conditions that have been previously set.
Use the following procedure to execute measurement according to the conditions set for the scene
currently displayed.
The data will be output to external devices as well.
1.Confirm that Run
If Run
is not selected, press the ESC Key several times to switch the
is selected.
selection.
Measurement will be executed and the measurement results will
be displayed on the screen. Check that measurement has been
performed correctly.
The output number for the currently displayed
measurement result is displayed.
Scn 0 Run
BRIGHT
DARK
OUT 0
9.96ms
Sen0
LV
PEAK
Sen1
LV
PEAK
Zero's OFF
+0005.18800
S+ / :Display S+ENT:Adjust
Judgment resultMeasurement value
[ 15]
[ 170]
PASSmm
SECTION 3
FARNEAR
Step
[ .]
[ ]
1
Step
2
Step
3
CHECK
Conditions can be changed simply without leaving Run Mode using the
adjustment menu.
Measurement will be interrupted while adjustments are made.
REFERENCE
Refer to page 28.
CHECKChange the Monitors using the SHIFT + Right Keys or SHIFT + Left Keys if
necessary.
REFERENCE
Refer to page 29.
CHECKFor measurements through the execution of Force-zero
REFERENCE
Refer to page 110.
Z500
Operation Manual6
Step
4
Step
5
Step
6
51
Page 54
SECTION 3
Menus for Conversational Menu
3-2 Saving Settings and Exiting the Z500
When settings have been completed, be sure to save to flash memory before turning the power OFF.
Refer to SECTION 2 (page 37) for details on saving settings.
REFERENCE
Refer to page 37.
SECTION 3
3-2 Saving Settings and Exiting the Z500
NOTICE
Flash memory data is loaded each time the Z500 is started up. Therefore, when
setting have been changed, be sure to save to flash memory before turning the
power OFF. If the power is turned OFF without saving, all of the setting changes
will be lost.
52
Z500
Operation Manual
Page 55
Z500
Operation Manual
SECTION 4
Menus for Expert Menu
This section explains how to set measurement conditions using the Expert Menu.
4-1 Setting Measurement Conditions Using the Expert Menus54
1 Starting Up Expert Menu and Entering Set Mode54
Step
2 Adjusting Images56
Step
3 Position Compensation64
Step
4 Measurement Settings70
Step
5 Measurement Conditions88
Step
6 Output Settings100
Step
7 Performing Measurement107
Step
SECTION 4Menus for Expert Menu
4-2 Saving Settings and Exiting the Z500108
53
Page 56
SECTION 4
Menus for Expert Menu
4-1 Setting Measurement Conditions Using the Expert Menus
This section explains how to set measurement conditions using the Expert Menus and the basic steps
before starting measurements.
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
Step 1
Starting Up Expert Menu and Entering Set Mode
To set the measurement conditions, it is first necessary to enter Set Mode.
1.
Move the cursor to Run and press the
Scn 0 Run
ENT Key.
A list of modes will be displayed.
2.
Select System.
The selections will be displayed.
3.
Select Environment.
Scn 0 Run
Scn 0 System
Comm
Environment
Slant correct
Initialize
Version
Set
Run
Tools
System
Save
The screen for setting environment will be displayed.
4.
Move the cursor to Conversatn and
press the ENT Key.
Z500
Operation Manual
54
Scn 0 System
Settings
Sensor
Scene
Monitor
1st scene
Dec. pts
:
Conversatn
:
Mirror
:
Console
:
Image
:
Scn 0
:
5 digits
End
Page 57
SECTION 4
Menus for Expert Menu
The selections will be displayed.
5.
6.
The setting will be registered and the screen in (3.) will return.
7.
The screen in (2.) will return.
8.
Move the cursor to Expert and press the
ENT Key.
Select End.
Press the ESC Key.
Select Set.
Scn 0 System
Settings
Sensor
Scene
Monitor
1st scene
Dec. pts
Scn 0 System
Set
Run
Tools
System
Save
:
Conversatn
Conversatn
:
:
Console
:
:
:
5 digits
End
Mirror
Expert
Image
Scn 0
SECTION 4
Step
1
The selections available with the Expert Menu's Set Mode will be
displayed.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
55
Page 58
SECTION 4
Menus for Expert Menu
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
Step 2
Adjusting Images
To facilitate accurate measurements, images are adjusted for clearer display of the section to be
measured.
Measurement Methods
Specify the region to be measured.
Select Zoom to assure higher position measurement accuracy.
Normally, use with the default setting (Wide).
SelectionDescription
Measures the whole measurement range of the connected
sensor.
Enlarges and measures a part of the measurement range of the
connected sensor.
Select this function to assure higher position measurement
accuracy.
The asterisk (*) indicates the default setting.
Wide
+20mm
Zoom
CHECK
Measurement Range
Wide*
Zoom
When the measurement method is changed, all settings for scene data wil be cleared.
Example: Z500-SW17
+20mm
56
CHECK
Z500
Operation Manual
0mm
-20mm
The measurement range varies depending on the Sensor models connected.
REFERENCE
14.5mm
( 7.25mm)
17.3mm
( 8.65mm)
20.1mm
( 10.05mm)
0mm
-20mm
4.8mm
5.7mm
6.7mm
This range is enlarged three
times and displayed on the
monitor screen.
Refer to page 28 in the Setup Manual.
Page 59
SECTION 4
Menus for Expert Menu
1.
The measurement method selections will be displayed.
Two Sensor
2.
3.
Select Image/Meas Method.
The selections for the sensors to be used for
measurement will be displayed.
Select the Sensor number.
Go to (2.).
Select the measurement method while
monitoring the image to be adjusted.
Press the ENT Key.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set
Wide
Zoom
Meas method
Region
Sensitivity
Details
SECTION 4
The setting will be registered and the screen in (1.) will return.
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
57
Page 60
SECTION 4
Menus for Expert Menu
Setting Measurement Region
Specify the region to be measured.
The Z500 adjusts sensitivity and performs measurements in a region that has
been previously set. Though the whole image can be used for the measurement
region, measurement accuracy can be raised by specifying a restricted
measurement region.
CHECK
Correct measurement may not be possible at the boundaries of the measurement regions.
Leave space at the boundaries when setting the regions.
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
1.
The screen for specifying the region will be displayed.
2.
The cursor will be moved to the ending line.
Select Image/Region.
Specify the starting line of the
measurement region.
Move the cursor with the Up and Down
Keys.
(Use these keys together with the SHIFT
Key to move the cursor quickly.)
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Cursor
Meas method
Region
Sensitivity
Details
3.
A confirmation message will be displayed.
4.
The setting will be registered and the screen in (1.) will return.
Repeat the procedure described in (2.) to specify the ending line of the
measurement region.
Select Execute.
Z500
Operation Manual
58
The measurement region
will be set. OK?
Execute
Cancel
Page 61
Sensitivity Adjustment
)
ght
y
e
ow
Select the sensitivity adjustment method for the measurement range.
If the measurement results fluctuate, change the adjustment method.
Normally, these conditions can be left at the default settings.
CHECK
The setting can be performed through the adjustment menu.
Refer to page 28.
SECTION 4
Menus for Expert Menu
REFERENCE
Two Sensor
Select the Sensor number to be used for measurement.
Sensitivity Adjustment
SelectionDescription
Auto*
Fixed
Multi
Adjust the sensitivity according to sensitivity information in the
measurement region.
The sensitivity can be set to one of 31 levels (Setting range: 0 to 31).
Select when it is difficult to assure accurate measurement in Auto.
Such cases may appear on a production line where workpieces of
different colors are processed.
Adjust the sensitivity for each line in the measurement region.
Select when it is difficult to assure accurate measurement in Auto due to
a widely distributed brightness of the workpiece surfaces.
However, sampling time will be longer.
Position a reference workpiece and execute Multi. The Z500 will set
automatically the upper and lower limits of sensitivity on the basis of the
brightness of the reference workpiece.
After the settings have been completed, fine adjustment can be made on
Details.
REFERENCE
Refer to page 63.
The asterisk (*) indicates the default setting.
Sensitivity level (Displayed only when Fixed is selected from the
Sensitivity Menu)
SECTION 4
Step
1
Step
2
Step
3
Step
4
Setting range
LV0 (See note
LV31High
Note: The laser will turn OFF when selecting [LV0].
Sensitivit
L
Color of workpiec
Bri
Dark
Step
5
Step
6
Step
7
Operation Manual
Z500
59
Page 62
SECTION 4
Menus for Expert Menu
When selecting Auto
Select Image/Sensitivity.
1.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Meas method
Region
Sensitivity
Details
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
The selections for sensitivity adjustment will be displayed.
2.
Select Auto.
The setting will be registered and the screen in (1.) will return.
Auto
Fixed
Multi
60
Z500
Operation Manual
Page 63
When selecting Fixed
Select Image/Sensitivity.
1.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
SECTION 4
Menus for Expert Menu
Meas method
Region
Sensitivity
Details
The selections for sensitivity adjustment will be displayed.
2.
Select Fixed.
The screen for setting the sensitivity level will be displayed.
3.
Select the level using the Up and
Down Keys while monitoring the
image on the monitor.
The level displayed first is the level
selected by the Z500 for the workpiece being measured. Adjust the
level while monitoring the line brightness graph and the received level on
the monitor.
120 max. : Insufficient light density
255 min. : Excessive light density
4.
Press the ENT Key.
Auto
Fixed
Multi
Line Brightness Received level
LV 15
BRIGHT
DARK
Sen0
LV
PEAK
Sen1
LV
PEAK
[ 15]
[ 170]
SECTION 4
Step
1
FARNEAR
Step
2
[ .]
[ ]
Step
3
Step
4
Step
5
Step
6
The setting will be registered and the screen in (1.) will return.
Operation Manual
Z500
Step
7
61
Page 64
SECTION 4
Menus for Expert Menu
When selecting Multi
Select Image/Sensitivity.
1.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Meas method
Region
Sensitivity
Details
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
The selections for sensitivity adjustment will be displayed.
2.
The sensitivity multiple setting screen will be displayed.
3.
Select Multi.
Check whether or not the workpiece
is positioned inside the measurement
range and press the ENT Key.
Auto
Fixed
Multi
Line beam image
Scn 0 Set
The setting will be registered and the screen in (1.) will return.
Z500
Operation Manual
62
Page 65
Details
SECTION 4
Menus for Expert Menu
Use this function to measure transparent workpieces. This function is also
useful to shorten the response time in Sensitivity/Multi. Normally, these
conditions can be left at the default settings. After changing the settings, perform
a workpiece measurement to check that measurement can still be performed
correctly.
CHECK
1.
Item
Surface
Upper0 to 31(31*)Specify the range to be adjusted the sensitivity.
Lower0 to 31(1*)
Gradations 120 to 220(170*)
The setting can be performed through the adjustment menu.
REFERENCE
Seleciton/
Setting range
Frnt*
Back
Description
The reflecting surface for which sensitivity adjustment
is performed is selected. To measure the back side of
a transparent workpiece, select Back. Normally (i.e.,
when measuring opaque workpieces), this setting can
be left as Frnt.
This function is effective when sensitivity adjustment
method is set to Auto or Multi. If the adjustment range
is small, the response time for switching sensitivity will
be short.
Set the target gradation for sensor sensitivity
adjustment.
The asterisk (*) indicates the default setting.
Refer to page 28.
Select Image/Details.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Meas method
Region
Sensitivity
Details
SECTION 4
Step
1
Step
2
Step
3
Step
4
The setting screen for sensitivity details will be displayed.
2.
3.
Make the settings for each item.
Select End.
The setting will be registered and the screen in (1.) will return.
Scn 0 Set
Surface
Upper
Lower
Gradations
:
Frnt
:
[31]
:
[ 1]
:
[170]
End
Operation Manual
Step
5
Step
6
Step
7
Z500
63
Page 66
SECTION 4
Menus for Expert Menu
SECTION 4
4-1 Setting Measurement Conditions Using the Expert Menus
Step 3
Position Compensation
The position compensation function is used when the position of the measurement workpieces are not
consistent.
By using this function, the displacement between the reference position and the current position is
obtained, and this displacement is compensated for in measurements. Compensation can be made for
both height and position.
Height
Position
• Reference Position
The measurement range and measurement workpiece are in the correct
positions.
Workpiece
Sensor
Height
Scn 0 Run
BRIGHT
DARK
Measurement range
Measurement center
LowHigh
Measurement point
9.96ms
FARNEAR
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
64
• When the Position of the Workpiece is Lower
By making position displacement compensation settings ...
For measurement, compensate the height based on the displacement of the reference position.
Z500
Operation Manual
Scn 0 Run
BRIGHT
DARK
Measurement point
9.96ms
FARNEAR
Sen0
[ 15]
LV
[ 170]
PEAK
Sen1
[ .]
LV
[ ]
PEAK
Page 67
Operational Flow
1
2
#
Page 65
Registering
the Reference
Position
#
Page 69
Set the
Compensation
Method
SECTION 4
Menus for Expert Menu
CHECK
#1
Registering the Reference Position
To use the position compensation function, set Compensatn and Meas set in this order.
Register the position to be used for a reference for position compensation.
Before setting the necessary items, place the workpiece at the correct position.
In the example provided below, the procedure for compensating under the
following conditions is explained.
Measurement
point (Average)
Reference height measurement point
SelectionDescription
Average
Peak
Bottom
PeakBottom
Reference heightReference position
Region
Measurement
point (Edge position)
Edge level
Define the average profile position within a specified region as the
measurement point.
Define the peak profile position within a specified region as the
measurement point.
Define the bottom profile position within a specified region as the
measurement point.
SECTION 4
Step
1
Step
2
Step
3
Step
4
Step
5
Region
Step
6
Step
7
Operation Manual
Z500
65
Page 68
SECTION 4
Menus for Expert Menu
SECTION 4
1.
The selection for position compensation will be displayed.
2.
Select Compensatn.
Select Reference.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
None
Reference
4-1 Setting Measurement Conditions Using the Expert Menus
A profile graph will be displayed.
3.
The screen for setting the reference height will be displayed.
4.
Check whether or not the workpiece
is positioned correctly and press the
ENT Key.
Specify the starting line of the
reference height region.
Use the Up and Down Keys to move
the cursor.
(Use these keys together with the
SHIFT Key to move the cursor
quickly.)
Profile
Cursor
Measurement
region
66
Z500
Operation Manual
Page 69
The ending-line setting screen will be displayed.
SECTION 4
Menus for Expert Menu
5.
The characteristic quantity setting screen will be displayed.
6.
Specify the ending line of the reference
height region.
Select a characteristic quantity that can
be used as the measurement point of
the reference height.
The cross cursor of the measurement
point will move in response to the
selections.
Select the item while monitoring the image.
In this example, select Average.
Average
Peak
Bottom
Cursor
Cursor
SECTION 4
Step
1
Step
2
The setting screen for reference position will be displayed.
7.
Specify the edge level to be used for the
reference position.
Use the Left and Right Keys to move
the cursor.
(Use these keys together with the SHIFT
Key to move the cursor quickly.)
Cursor
Operation Manual
Step
3
Step
4
Step
5
Step
6
Step
7
Z500
67
Page 70
SECTION 4
Menus for Expert Menu
SECTION 4
A cross cursor to be used for specifying the reference height will
be displayed.
8.
A confirmation message will be displayed.
9.
Use the Up and Down Keys to select the
measurement point of the reference
position.
The cursor of the measurement point is
moved.
Select the measurement point while
monitoring the image.
Select Execute.
Cursor
Edge level
Reference will be
registered.
OK?
ExecuteCancelBack
4-1 Setting Measurement Conditions Using the Expert Menus
The setting will be registered and the screen in (2.) will return.
68
Z500
Operation Manual
Page 71
#2
Set the Compensation Method
SECTION 4
Menus for Expert Menu
CHECK
1.
Select the position compensation method.
Height
Position
Unless Reference has been set, the compensation method selections will not be
displayed.
Move the cursor to None and press
the ENT Key.
.
SelectionDescription
None*No compensation is required.
HeightCompensate the height.
PositionCompensate the position.
Height&PosCompensate both the height and position.
The asterisk (*) indicates the default setting.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
None
Reference
SECTION 4
Step
1
The selection for compensation method will be displayed.
2.
The setting will be registered and the screen in (1.) will return.
Select the compensation method.
In this example, select Height&Pos.
None
Height
Position
Height&Pos
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
69
Page 72
SECTION 4
Menus for Expert Menu
SECTION 4
Step 4
Measurement Settings
Set the measurement contents. Operational flow varies depending on the measurement item.
Setting Step: 2 pts, Edge position, and Define are taken as examples for description.
Operational Flow
#2
#1
Page 70
Set Measurement
Items
When selecting
#2
Compensate the measurements under the output conditions given in
CHECK
CHECK
When not
selecting
Define
Define
Calibration setting is disabled.
To use the position compensation function, set Compensatn and Meas set in this order.
The settings can be performed through the adjustment menu.
REFERENCE
Refer to page 28.
Page 78 (if necessary)#3 Page 84 (if necessary)
Set Calibration
:
:
Set Output
Conditions
Changing
Measurement
Contents
Clearing
Measurement
Items
#3.
Page 86
Page 87
4-1 Setting Measurement Conditions Using the Expert Menus
#1
Set Measurement Items
Up to eight measurement contents can be set for each scene.
There are 8 types of the measurement item. Different measurement item can be
set for each output number.
Types of Measurement Item
SelectionDescriptionSymbol
Height*Measure the height within the specified region.
Specify two measurement regions.
After taking the measurement point within the 1st region
Step: 2 pts
Step: 3 pts
Edge position
as the reference, measure the distance from the measurement point in the 2nd region.
REFERENCE
Specify three measurement regions.
After taking the average of the measurement points
within the two regions specified above, measure the
difference from the measurement point in the 3rd region.
Measures the specified edge position.
REFERENCE
Refer to page 72.
Refer to page 74.
70
*Height is set in OUT0 as the default setting.
Z500
Operation Manual
Page 73
Types of Measurement Item (Continued)
SelectionDescriptionSymbol
SECTION 4
Menus for Expert Menu
Width
Edge center
Peak/bottom
Define
Measure the difference between the specified two edge
positions.
Based on the specified two edge positions, measure the
center.
Measure the peak or bottom point of the height within
the specified region.
Using the equation, set the measurement items.
The measurement result for an output number can be
assigned to the equation.
REFERENCE
Refer to page 76.
Characteristic Quantity of Measurement Points
SelectionDescription
Average
Peak
Bottom
Back
Define the average profile position within a specified region as the
measurement point.
Define the peak profile position within a specified region as the
measurement point.
Define the bottom profile position within a specified region as the
measurement point.
Set the back side of the workpiece for the measurement point.
This procedure is effective when glass sheets and other transparent
workpieces are measured.
*Height is set in OUT0 as the default setting.
SECTION 4
Step
1
Step
2
Step
3
Peak and Bottom
Peak Bottom
Region
Back
Workpiece
Operation Manual
Surface
Back
Z500
Step
4
Step
5
Step
6
Step
7
71
Page 74
SECTION 4
Menus for Expert Menu
Step: 2 pts
Level differences, grooves, etc. of workpieces are measured.
The following conditions will be taken as an example for describing the setting procedures.
SECTION 4
Step: 2 pts
Measurement
point (Average)
Measurement
point (Average)
1.
Select Meas set.
A list of output numbers will be displayed.
2.
Select the output number for which
Step: 2 pts is set.
In this example, Select OUT1.
CHECK
Height is set in OUT0 as the default set-ting. To set an item other than Height in
OUT0, delete the existing measurement item.
Top face region
Bottom face region
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set
OUT0:Height
OUT1:
OUT2:
OUT3:
OUT4:
OUT5:
OUT6:
OUT7:
REFERENCE
Refer to page 87.
The selection for measurement item will be displayed.
3.
Select the measurement item.
In this example, select Step: 2 pts.
Z500
Operation Manual
72
Measrmnt item
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
Page 75
SECTION 4
Menus for Expert Menu
The setting screen for bottom face measurement region will be
displayed.
4.
The ending-line setting screen will be displayed.
5.
The characteristic quantity setting screen will be displayed.
Specify the starting line of the bottom
face measurement region.
Move the cursor with the Up and Down
Keys.
(Use these keys together with the SHIFT
Key to move the cursor quickly.)
Repeat the procedure described in (4.)
to specify the ending line of the bottom
face measurement region.
Cursor
Cursor
SECTION 4
Cursor
6.
The setting screen for top face measurement region will be displayed.
7.
A confirmation message will be displayed.
8.
Select a characteristic quantity that can
be used as the measurement point of
the bottom face.
The cross cursor of the measurement
point will move in response to the
selections.
Select the item while monitoring the image.
In this example, select Average.
Repeat (4.) through (6.) to set the
characteristic quantities of the top face
measurement region and measurement
point.
Select Execute.
Average
Peak
Bottom
Back
Measrmnt items will be set.
OK?
ExecuteCancelBack
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
The setting will be registered and the screen in (2.) will return.
Operation Manual
Z500
73
Page 76
SECTION 4
SECTION 4
Menus for Expert Menu
Edge position
Measures the edge position of the workpiece.
The following conditions will be taken as an example for describing the setting procedures.
Edge level
Measurement
point (Edge position)
In this example, the intersection point of the edge level and the profile becomes
the measurement point.
Set the edge level as follows.
Edge level
Profile
The intersection is
the measurement point.
Edge position
CHECK
The setting edge level and the measurement point can be checked on the Image Monitor.
REFERENCE
Refer to page 32.
1.
Select Meas set.
A list of output numbers will be displayed.
2.
Select the output number for which
Edge position is set.
In this example, select OUT2.
CHECK
Height is set in OUT0 as the default
setting. To set an item other than Height
in OUT0, delete the existing measurement item.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:
OUT3:
OUT4:
OUT5:
OUT6:
OUT7:
74
REFERENCE
Refer to page 87.
Z500
Operation Manual
Page 77
SECTION 4
Menus for Expert Menu
The selection for measurement item will be displayed.
3.
The setting screen for edge position will be displayed.
4.
Select the measurement item.
In this example, Select Edge position.
Specify the level to be measured as the
edge.
Move the cursor with the Left and Right
Keys.
(Use these keys together with the SHIFT
Key to move the cursor quickly.)
Measrmnt item
Height
Step: 2 pts
Step: 3 pts
Edge position
Width
Edge center
Peak/bottom
Define
SECTION 4
The setting screen for edge position will be displayed.
5.
CHECK
A confirmation message will be displayed.
6.
Specify the edge position regarded as
the measurement point.
Move the cursor with the Up and Down
Keys.
In some cases, the cursor will not be
displayed on the intersection point of the
edge level and the edge position.
However, selection of an arbitrary cursor
will give the same measurement result
(the intersection point becomes the
measurement point), as long as the
selected cursor exists near the
intersection point.
Select Execute.
Cursor
Measrmnt items will be set.
OK?
ExecuteCancelBack
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
The setting will be registered and the screen in (2.) will return.
Operation Manual
Z500
75
Page 78
SECTION 4
Menus for Expert Menu
Define
Set the equation by using the measurement results for the output number.
Output number for which
OUT3
OUT1 OUT2
equation is set
Equation
SECTION 4
Define
List of Items
Note 1
• Output number
CHECK
OUT0
OUT1
OUT2
OUT3
OUT4
See note 1.
OUT0
OUT1
OUT2
OUT3
OUT4
OUT5
OUT6
OUT7
(
)
See note 2. See note 3.
OUT5
OUT6
OUT7
(
)
ABS
MAX
MIN
AVE
Cons
ABS
MAX
MIN
DEL
AVE
BS
Cons
Set
Move the cursor.
DEL
BS
Set
Numerals between 0 and 9999999.999 can be set.
List of Items
Select the items displayed
here for which equation is set.
Delete the items.
DEL : Deletes the item to the right of the cursor.
BS : Deletes the item to the left of the cursor.
Sets the equation.
When this item is selected, equation can be performed by using the
measurement results (results obtained after calibration) that have been set for
the output number.
Any output number cannot be selected, unless its measurement items have been set.
76
Note 2
• Arithmetic Operators
Note 3
•Functions
Z500
Operation Manual
OperatorMeaning
+ Addition
-Subtraction
*Multiplication
/Real number division
FunctionDescription
ABS Calculates the absolute value. ABS (argument)
MAX
MIN
AVE
Gives the largest value of arguments. MAX (Argument 1, Argument 2,
Argument 3,…)
Gives the smallest value of arguments. MIN (Argument 1, Argument 2,
Argument 3,…)
Calculates the average value of arguments. AVE (Argument 1, Argument 2, Argument 3,…)
Page 79
SECTION 4
Menus for Expert Menu
1.
Select Meas set.
A list of output numbers will be displayed.
2.
Select the output number for which
Define is set.
In this example, select OUT3.
CHECK
Height is set in OUT0 as the default set-ting. To set an item other than Height in
OUT0, delete the existing measurement
item.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:
OUT4:
OUT5:
OUT6:
OUT7:
SECTION 4
REFERENCE
Refer to page 87.
The setting screen for equation will be displayed.
3.
Select the item from the list of items and
set the equation.
4.
Select Set when the equation has been
set.
A confirmation message will be displayed.
5.
Select Execute.
OUT3
OUT0
OUT5
OUT1
OUT6
OUT2
OUT7
OUT3
(
OUT4
)
Measrmnt items will be set.
ExecuteCancelBack
OK?
ABS
MAX
MIN
AVE
Cons
DEL
BS
Set
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
The setting will be registered and the screen in (2.) will return.
Operation Manual
Z500
77
Page 80
SECTION 4
Menus for Expert Menu
#2
Set Calibration
Set calibrations to compensate the measurement values. Set calibrations to compensate for differences
between measurement values and actual dimensions due to the color or material of the measurement
workpiece. The setting method varies depending on the measurement items set for the output number.
CompensationMeasurement itemsDescription
Compensate the
measured height
and position.
Compensate the
measured level
difference and
width.
Height, Edge position, Edge
center, Peak/bottom
Step: 2 pts, Step: 3 pts, Width
For setting, repeat the measurement
twice.
REFERENCE
For setting, repeat the measurement
once.
REFERENCE
Refer to page 78.
Refer to page 81.
SECTION 4
Define
CHECK
Output can be used for fine adjustment after calibration setting. Use Output also to
compensate the measurement values when Define has been selected for the
measurement item.
REFERENCE
Refer to page 84.
CHECK
Refer to page 83 for information on how to clear calibration settings.
Compensate the Measured Height and Position
To compensate for the error, perform the measurement twice.
Applicable measurement items: Height, Edge position, Edge center,
Peak/bottom
Example 1: When selecting
Workpiece height: 10 mm
Use the method shown below to calibrate workpieces so that a displacement value of 9.95 mm is
measured as 10 mm.
9.95mm
Calibrate to 10 mm.
Measurement of point 1
Height
Measurement value
10mm
Measurement of point 2
Compensate for this error.
10mm
Displacement of workpiece
78
Measure without
the workpiece.
Z500
Operation Manual
Input the height
of point 1.
0mm
Input the height
of point 2.
10mm
Measure with the
workpiece in position.
Page 81
SECTION 4
Menus for Expert Menu
Example 2: When selecting
Movement distance from the edge position: 5 mm
Use the method shown below to calibrate workpieces so that a displacement value of 4.95 mm is
measured as 5 mm.
4.95mm
Calibrate to 5 mm.
Measurement of point 1
Input the
position
of point 1.
0mm
Measure with the workpiece
at the reference position.
Edge position
Measurement value
5mm
Measurement of point 2
Move the workpiece by 5 mm
for the measurement.
5mm
Compensate for this error.
Workpiece Movement
Input the
position
of point 2.
5mm
Height in Example 1 is taken here to describe the setting method.
Make the necessary adjustments when setting a different item.
SECTION 4
1.
Move the cursor to Height and press the
ENT Key.
The selections will be displayed.
2.
Select Calibrate.
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0 Set
Change
Calibrate
Output
Delete
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
79
Page 82
SECTION 4
Menus for Expert Menu
SECTION 4
A confirmation message will be displayed.
3.
The measurement screen for point 1 will be displayed.
4.
The screen for inputting the position of point 1 will be displayed.
5.
Select Execute.
Press the ENT Key without a workpiece
in position.
Set the value to 0.00000 and press the
ENT Key.
System will be calibrated.
Clear settings now if
required.
ExecuteCancel
00000.00000
Clear
Define
The measurement screen for point 2 will be displayed.
6.
The screen for inputting the position of point 2 will be displayed.
7.
A confirmation message will be displayed.
8.
The setting will be registered and the screen in (2.) will return.
Position the workpiece correctly and
press the ENT Key.
Set the value to 5.00000 and press the
ENT Key.
Select Execute.
00005.00000
System will be calibrated.
ExecuteCancel
80
Z500
Operation Manual
Page 83
Compensate the Measured Level Difference and Width
To compensate for the error, perform the measurement once.
Level difference: 5 mm
Use the method shown below to calibrate workpieces so that a displacement value of 4.95 mm is
measured as 5 mm.
Point 1
Point 2
To compensate for this error, perform the measurement once.
Measure with the
workpiece in position.
1.
Move the cursor to Step: 2 pts and
Calibrate to 5 mm.
Input the value
of point 1.
press the ENT Key.
0mm
Step: 2 pts
Measurement value
5mm
4.95mm
5mm
Input the value
of point 2.
5mm
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Compensate for this error.
Level difference of workpiece
SECTION 4
Step
1
Step
2
Step
3
Step
4
The selections will be displayed.
2.
Select Calibrate.
Scn 0 Set
Change
Calibrate
Output
Delete
Operation Manual
Z500
Step
5
Step
6
Step
7
81
Page 84
SECTION 4
Menus for Expert Menu
SECTION 4
A confirmation message will be displayed.
3.
The measurement screen for point will be displayed.
4.
The screen for inputting the position of point 1 will be displayed.
5.
Select Execute.
Position the workpiece correctly and
press the ENT Key.
Set the value to 0.00000 and press the
ENT Key.
System will be calibrated.
Clear settings now if
required.
ExecuteCancel
00000.00000
Clear
Define
The screen for inputting the position of point 2 will be displayed.
6.
A confirmation message will be displayed.
7.
The setting will be registered and the screen in (2.) will return.
Set the value to 5.00000 and press the
ENT Key.
Select Execute.
00005.00000
System will be calibrated.
ExecuteCancel
82
Z500
Operation Manual
Page 85
Clearing Calibration Settings
SECTION 4
Menus for Expert Menu
1.
The selections will be displayed.
2.
Move the cursor to the measurement
item to be cleared and press the ENT
Key.
Select Calibrate.
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0 Set
Change
Calibrate
Output
Delete
SECTION 4
Step
1
A confirmation message will be displayed.
3.
The setting will be cleared and the screen in (2.) will return.
Select Clear.
System will be calibrated.
Clear settings now if
required.
ExecuteCancel
Clear
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
83
Page 86
SECTION 4
SECTION 4
Menus for Expert Menu
#3
Set Output Conditions
Set the output conditions of measurement results as necessary.
ItemSelectionDescription
Upper
Lower
Hold
Zero
-9999.99999 to
9999.99999
Set when the Trigger function is used. Select the contents of the data to
be held.
REFERENCE
None*Hold measurement is not performed.
SamplingThe value at the start of measurement is held.
Peak
BottomThe lowest value measured during measurement is held.
Peak-Peak
Average
-9999.99999 to
9999.99999
Set the judgment conditions. Set the range for a PASS
judgment.
For details on the Trigger function, refer to page 90.
The highest value measured during measurement is
held.
The difference between the highest and lowest values
measured during measurement is held.
The average value between the highest and lowest
values measured during measurement is held.
Set the offset value for force-zero execution.
Set the value to be added to 0.
REFERENCE
Refer to page 110, Force-zero.
Define
Span and Offset
Span
A coefficient (the span) is set to compensate
for incline in the Sensor characteristics.
Measurement value
(mm)
Span
Offset
Unit
4.0
0.20000 to
4.00000
(1.00000*)
-9999.99999 to
9999.99999
(0.00000*)
mm*
2
mm
None
0.2
Displacement
of workpiece
Set to make fine adjustments to measurement values
after calibration. Set for the incline in the Sensor
characteristics.
Set to make fine adjustments to measurement values
after calibration. Added to or subtracted from
measurement values that have been set.
Select the unit of the measured values to be displayed on
the monitor screen.
The asterisk (*) indicates the default setting.
Offset
A fixed value (the offset) is added to or
subtracted from measurement values.
Measurement value
(mm)
Displacement
of workpiece
84
CHECK
Z500
Operation Manual
When calibration is executed, the span and the offset will be automatically set.
Page 87
SECTION 4
Menus for Expert Menu
1.
Move the cursor to the output number to
be set the output conditions and press
the ENT Key.
The selections will be displayed.
2.
Select Output .
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0 Set
Change
Calibrate
Output
Delete
SECTION 4
The setting screen for output conditions will be displayed.
3.
4.
Set the necessary items.
Select End.
The setting will be registered and the screen in (2.) will return.
Scn 0 Set
Upper
Lower
Hold
Zero
Span
Offset
Unit
:
[ 10.00000]
:
[ -10.00000]
:
None
:
[ 0.00000]
:
[1.00000]
:
[ 0.00000]
:
mm
End
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
85
Page 88
SECTION 4
Menus for Expert Menu
Changing Measurement Contents
Change the measurement contents that have been set for the output number.
SECTION 4
Changing Measurement Contents
CHECK
To change measurement items, clear the existing measurement items and then
set new items.
Refer to page 87.
1.
Move the cursor to the output number to
be changed the output conditions and
press the ENT Key.
The selections will be displayed.
2.
Select Change.
REFERENCE
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0 Set
Change
Calibrate
Output
Delete
The setting screen for measurement range will be displayed.
The remaining procedures are the same as those for measurement item setting.
Z500
Operation Manual
86
Page 89
Clearing Measurement Items
Clear the measurement items that have been set for the output numbers.
REFERENCE
Refer to page 127 for details on clearing separately for each scene.
SECTION 4
Menus for Expert Menu
1.
The selections will be displayed.
2.
Move the cursor to the output number to
be cleared the output conditions and
press the ENT Key.
Select Delete.
Scn 0 Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0 Set
Change
Calibrate
Output
Delete
SECTION 4
Step
1
Step
2
Step
3
A confirmation message will be displayed.
3.
The measurement items for the selected output number are cleared off and the screen in (1.) will return.
Select Execute.
Delete OUT0 settings.
ExecuteCancel
Z500
Operation Manual
Step
4
Step
5
Step
6
Step
7
87
Page 90
SECTION 4
Menus for Expert Menu
SECTION 4
Clearing Measurement Items
Step 5
Measurement Conditions
Set the measurement conditions. The following can be set in Conditions and its lower hierarchies.
Perform the setting as required.
MenuDescriptionReference
The average of the setting number of times is output as
# to avg
Trigger
Details
the measurement result. Set this function to disregard
sudden changes in the waveform.
From the measurement results that have been obtained
within a specified measurement period, the specified
characteristic quantity is held and output.
Removing noises, outputting measurement failure, and
other miscellaneous methods can be set in more detail.
Page 88.
Page 90.
Page 96.
Averaging number
Use this function to disregard sudden changes in the waveform. The average of
the setting number of times is output as the measurement result.
Selection
• The Relationship between the Measurement Response Time and the Resolution
Averaging
number
1
1, 2, 4, 8*, 16, 32, 64, 128, 256
Measurement
response time
Short
Resolution
Low
The asterisk (*) indicates the default setting.
CHECK
1.
256Long
The setting can be performed through the adjustment menu.
REFERENCE
High
Refer to page 28.
Select Conditions.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
88
Z500
Operation Manual
Page 91
SECTION 4
Menus for Expert Menu
The selections will be displayed.
2.
The setting screen for averaging number will be displayed.
3.
4.
Select # to avg.
Select averaging number.
Select End.
Scn 0 Set
# to avg
Trigger
Details
# to avg
: 8
End
SECTION 4
The setting will be registered and the screen in (2.) will return.
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
89
Page 92
SECTION 4
Menus for Expert Menu
Trigger setting
Use the Trigger function to measure moving workpieces in chronological order.
From the measurement results that have been obtained within a specified
measurement period, the specified characteristic quantity is held and output.
Use the procedure explained below to set the trigger input method.
Workpiece movement direction
SECTION 4
Clearing Measurement Items
Height(mm)
Time(s)
Position(mm)
• Trigger Input Method
Trigger OFF
End
Start
Trigger ON
SelectionDescription
Free*
Measurement is performed continuously. Hold measurement is not
performed.
Select when using a sync sensor.
External
The trigger turns ON at the same time as
the TRIGGER terminal on the terminal
block.
The trigger-ON timing is determined according to the measurement
value. The trigger turns ON when the measurement value goes above
(or below) the trigger level. If Self is selected, the screen for selecting
the output number (for the trigger setting) and the trigger direction is
displayed.
The characteristic quantities being
held are output at this timing.
The characteristic quantity of
the measurement during
this period will be held.
terminal
OFF
ON
Trigger ON
TRIGGER
90
CHECK
Z500
Operation Manual
Trigger ON
Self
Up*
The trigger turns ON
when the measurement
value goes above the
trigger level.
Trigger level
Measurement
value waveform
The trigger turns ON
Down
when the measurement
value goes below the
trigger level.
Measurement
value waveform
Trigger level
The asterisk (*) indicates the default setting.
Use Output to select the characteristic quantity that holds the measured values.
REFERENCE
Refer to page 84.
Page 93
SECTION 4
Menus for Expert Menu
The procedure for make trigger settings under the following details is explained
below.
The difference between the maximum value and minimum value of the workpiece height is output.
Measurement item
Trigger input method
Hold setting (see note)
OUT0
Note:
Set the Hold setting by using
Refer to page 84.
Operational Flow
#1
Page 91
Display Waveforms
in the Trend Monitor
(Through Mode)
Display Waveforms in the Trend Monitor
#1
Use the waveform to decide the trigger level and to set the measurement start
position.
CHECK
1.
2.
3.
Set the screen mode to Thru in advance.
REFERENCE
Refer to page 118.
Press the SHIFT + Right Keys or SHIFT
+ Left Keys several times to switch to
Trend Monitor.
Pass the workpiece under the Sensor
and display the waveform.
The instant the waveform representing
the workpiece is displayed, press the
TRIG Key.
OUT0:Height
Self
Peak-Peak
Meas set
#2
Page 92
Set the Trigger Input
Method
(Trigger Mode)
in advance.
#3
Page 95
Confirm Trigger
Settings
Scn 0 Run
1div x: 1.000 [s]
y:+10.0000 [mm]
Thru
OUT 0
S+ / :Display S+ENT:Adjust
9.96ms
Zero's OFF
mm+0005.18800
PASS
SECTION 4
Step
1
Step
2
Step
3
Step
4
Step
5
The waveform display is fixed.
Scn 0 Set
Thru
The symbol changes.
1div x: 1.000 [s]
y:+10.0000 [mm]
Z500
Operation Manual
Step
6
Step
7
91
Page 94
SECTION 4
Menus for Expert Menu
Set the Trigger Input Method
#2
Select the trigger input method, and set the trigger position and the trigger level.
CHECK
The setting can be performed through the adjustment menu.
REFERENCE
Refer to page 28.
SECTION 4
Clearing Measurement Items
1.
The selections will be displayed.
2.
The selections for trigger input method will be displayed.
3.
Enter Set Mode and select Conditions.
Select Trigger.
Select the trigger input method and
press the ENT Key.
In this example, select Self.
Scn 0 Set
Image
Compensatn
Meas set
Conditions
Output
# to avg
Trigger
Details
Free
External
Self
The setting screen for trigger contents will be displayed.
4.
5.
Set each item.
In this example, select Trigger to OUT0
and Direction to Up.
Select Execute.
The trigger setting screen will be
displayed.
The waveform set in
Z500
Operation Manual
92
is displayed.
#1
Scn 0 Set
Free
External
Self
Trigger
Direction
Execute
:OUT0
:Up
Cancel
Page 95
SECTION 4
Menus for Expert Menu
6.
Specify the trigger-ON position and
trigger level, and press the ENT Key.
Use the Left and Right Keys to specify
the trigger position.
Use the Up and Down Keys to specify
the trigger level.
Use these Keys together with the SHIFT
Key to move the cursor quickly.
CHECK
When the trigger turns ON, the waveform above the trigger level is displayed at a position
based on the position set here.
Trigger Position in the Center Trigger Position on the Left
The waveform before and
after the trigger turns ON
can be monitored.
The trigger-OFF setting screen will be displayed.
Mainly the waveform after
the trigger turns ON can be
monitored.
Trigger-ON position
Trigger
Thru
Trigger-OFF position
Trigger level
1div x: 1.000 [s]
y:+10.0000 [mm]
SECTION 4
7.
Using the Left and Right Keys, specify
the trigger-OFF position.
Use the Left and Right Keys together
with the SHIFT Key to move the cursor
quickly.
The setting screen for measurement start timing will be
displayed.
8.
Using the Left and Right Keys, specify
the measurement starting position.
Use the Left and Right Keys together
with the SHIFT Key to move the cursor
quickly.
CHECK
The time for cursor position (time after
the trigger is turned ON) is displayed on
the lower part of the screen. Refer this
time for setting the measurement
starting position.
Trigger
Thru
Measurement start position
Trigger
Thru
Measure start249.00 ms
Time for cursor position
1div x: 1.000 [s]
y:+10.0000 [mm]
1div x: 1.000 [s]
y:+10.0000 [mm]
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Step
7
Operation Manual
Z500
93
Page 96
SECTION 4
Menus for Expert Menu
SECTION 4
Clearing Measurement Items
The cursor will be moved to the measurement ending position.
9.
Repeat the procedure described in (8.)
to specify the measurement ending
position.
CHECK
The time for cursor position (time after
the trigger is turned ON) is displayed on
the lower part of the screen. Refer this
time for setting the measurement ending
position.
A confirmation message will be displayed.
10.
Select Execute.
The setting will be registered and the screen in (2.) will return.
Measurement end position
Trigger
Thru
Measure start249.00 ms
Measure stop 2320.68 ms
Time for cursor position
The settings
will be changed.
ExecuteCancelBack
OK?
1div x: 1.000 [s]
y:+10.0000 [mm]
94
Z500
Operation Manual
Page 97
Menus for Expert Menu
Confirm Trigger Settings
#3
Check if measurements are being correctly performed with the set trigger contents.
CHECK
Set the screen mode to Trig in advance.
REFERENCE
Refer to page 118.
SECTION 4
1.
2.
The waveform is displayed from the trigger position set in #2.
3.
• When the measurement start and end positions have to be adjusted
Press the SHIFT + Right Keys or SHIFT
+ Left Keys several times to switch to
Trend Monitor.
Pass the workpiece under the Sensor.
Perform a workpiece measurement
several times to check that
measurement can be performed
correctly.
, with the waveform kept displayed, and adjust the
Return to the procedure in
positions.
#2
Scn 0 Run
1div x: 1.000 [s]
y:+10.0000 [mm]
Thru
OUT 0
S+ / :Display S+ENT:Adjust
Scn 0 Run
Trig
Zero's OFF
mm+0005.18800
PASS
9.96ms
1div x: 1.000 [s]
y:+10.0000 [mm]
Trigger-OFF positionTrigger-ON position
9.96ms
SECTION 4
Step
1
Step
2
Step
3
Step
4
Step
5
Step
6
Operation Manual
Z500
Step
7
95
Page 98
SECTION 4
Menus for Expert Menu
Details setting
Use the details setting when measurement cannot be performed correctly or when more accurate