| Force | Pressure | Temperature | Switch
PT Compact
Models: TEP11
TES11, TES12, TES13, TES14
PT Compact plus
Screw - in resistance thermometer
Description
The resistance thermometer „PT Compact“ is designed for
temperature measurement in processes with medium and
low pressures. Measurement ranges between -200°C and
+600°C cover the majority of your temperature measuring
tasks.
The thermometers are available with platinum sensors
PT100 or PT1000 in 2 -, 3 -, 4 or 2x2-wire circuit. If the
measuring signal has to be transmitted over longer
distances, the PT Compact plus is available with an
integrated transmitter. This transmitter converts the
resistance of the platinum sensor into a linear temperature
signal 4-20 mA or 0-10 V.
Different process connections, as well as adjustable
compression fittings underline the variability of this
measuring instrument. To achieve fast reaction times a
version with a tapered stem is also available.
The electrical connection is realized with an L-plug
according to DIN EN 175301-803A. Optionally an M12x1
connection is available.
All thermometer parts which get in contact to the process
medium are made of stainless steel. The housing and the
replaceable measuring insert are bolted together by a
knurled nut. This allows the exchange of the measuring
insert without removing the thermometer from the process.
Even an upgrade is possible from PT Compact to
PT Compact plus or to TC400, a compact RTD with LEDdisplay and switching output, without opening the process
connection.
Features
O compact dimensions
O easy handling
O cost-efficient
O short delivery times
O optional with transmitter 4-20mA or 0-10V
O simple exchange
O customized solutions
Models
O -50°C up to +200°C (-60...400°F)
O -50°C up to +400°C (-60...750°F)
O -50°C up to +600°C (-60...1100°F)
O -200°C up to +600°C (-300...1100°F)
Measuring ranges (with transmitter)
O 0...50°C; 0...100°C; 0...120°C
O 0...200°F; 0...500°F; -50...300°F
O according to customers requirements
Applications
O engineering
O heating and cooling circuits,
air condition technology
O plant construction
O environment engineering

TES11 -> 2-wire
TES12 -> 3-wire
TES13 -> 4-wire
TES14 -> 2 x 2-wire
TEP11 -> 4-20 mA
TEP11 -> 0-10 V
Sensor
PT100 class B
optional PT100 class A
PT1000 class B
optional PT1000 class A
PT100 class B
optional PT100 class A
Output signal
and supply
voltage
4-20 mA, 2-wire
supply voltage: 10 – 30 V DC
ripple < 10%
0-10 V, 3-wire
supply voltage: 12 – 30 V DC,
ripple < 10%
sensor burnout: 23mA
sensor short circuit: 3,3 mA
standard: -50°C up to +200°C (-60...400°F)
-50°C up to +400°C (-60...750°F)
-50°C up to +600°C (-60...1100°F)
-200°C up to +600°C (-300...1100°F)
selectable measuring range:
standard version:
minimum range 50K
maximum range 250K
high temperature versions:
minimum range 150K
maximum range 800K
fixed thread: G ½ A, G ¼ A, G ⅜ A, G ¾ A, ½“NPT, ¼“NPT, M14x1,5
adjustable compression fitting: G ½ A, G ⅜ A, G ¼ A, ½“NPT
other connections on request
stainless steel 1.4571 (316 Ti)
other materials or coatings on request
Stem length
and
pressure ranges1)
3mm fast reaction version with tapered stem up to 12 bar1):
stem length 25mm: 3 x 0,3mm
stem length 50mm up to 100mm: 6 x 0,3mm with tapered stem 3 x 0,3mm
from stem 150mm: 8 x 1,75mm with tapered stem to 6 x 0,3mm with tapered stem 3 x 0,3mm
6 x 0,75mm from stem 50mm to 500mm: up to 40bar1)
8 x 1,75mm from stem 50mm to 1000mm: up to 100bar1)
special parts manufactured from full material for pressures up to 600 bar1)
transmitter: <0,5% of measuring range
at connector max. 125°C
with transmitter max. 85°C
L-plug acc. to DIN EN 175301-803 form A
optional: round connector, 4-pin, M12x1
emitted interference acc. DIN EN 61326
breakdown effect acc. to DIN EN 61326
IP65 acc. to DIN EN 60529 / IEC 529
Technical data
1)
Pressure ranges refer to static pressure; Rating depends on:
- process medium
- process pressure and temperature
- flow rate
- Stem design (length, diameter, wall thickness)
DE 1127 i
p. 2 / 4

tapered stem 3
for fast response-time
L-plug acc. to
DIN EN 175301-803A
housing and thread
( 1.4571 / 316 Ti)
fixed thread acc.
to DIN 3852
adjustable
compression
fitting
Dimensions
Wiring Diagrams
4..20mA: 1=Ub+/S+ 3=0V/S-
0..10V: 1=Ub+ 3=0V/Ub- 4=S+
DE 1127 i
p. 3 / 4

Temperature range, stem and process connection
-200°C ... +600°C (-300…1100°F) 1
-50°C ... +200°C (-60…400°F) 2
-50°C ... +400°C (-60…750°F) 3
-50°C ... +600°C (-60…1100°F) 4
1
2
3
4
5
6
7
Sensor Class A
Round connector M12x1, 4-pin
customer specific ……………..
Without transmitter
Diameter
3 mm - tapered, fast reaction stem
PT 1000 (only for models TES, without transmitter)
others (please add plain text)
50 mm (~2") only with fixed thread
75 mm (~3") only with fixed thread
Type of process
connection
Standard for temperature-ranges up to 400°C
Standard for temperature-ranges up to 600°C
Configuration PT-Compact
Configuration example: PT Compact plus with transmitter 4-20mA
Stem diameter 6mm
Temperature range -50…200°C (-60…400°F)
Fixed thread G1/2 A
Stem length 200mm
Measuring range 0…200°F
TEP11X221xxx (the last three digits= xxx are assigned from tecsis)
Subject to technical modifications
DE 1127 i