Datasheet MP983, MP972, MP973, MP963, MP1983 Datasheet (Perkinelmer)

...
Page 1
DATASHEET
www.perkinelmer.com/opto
.
Description
The Photon Counting Head MP 900 series is designed for applications in all fields of single photon detection, e.g. chemoluminescence, bioluminescence, flourescence, in-vitro ssay,environmental measurements or pure research. It is an easy to use module, containing the Channel Photomultiplier, a high voltage power supply, a discrimination amplifier and a pulse shaper for fast output pulses. An installed active quenching system avoids over­illumination to the detector.
It is also possible to apply an external gate function for time correlated photon counting. Strong variations in light levels are possible due to the high dynamic range of the installed CPM. The exceptional low noise and high sensitivity facilitates detection of extremely low light levels.
Features
· High dynamic range
· No cooling required
· Very high stability in noise level
· Active quenching circuit for high light protection
· Gateable CPM input
· Optical fiber read-out possible
· 5 volts operating voltage
· Monitor voltage output
MP Series Photon Counting Modules
Ultra High Sensitivity Gateable Photon Counting Module
Lighting Imaging Telecom Sensors
Channel Photomultipliers and Modules
Page 2
MP 900 Series 1/3” Photoncounting Module
Technical Specifications
Supply voltage 5 V dc Supply voltage 5 V dc Input current at max. count rate < 280 mA Settling time < 1s (time to stabilize HV after supply voltage applied) Over-illumination protection: active quenching control (internal) Linear count rate: 5 MHz Output pulse: TTL, positive Active Quenching Control TTL-Pulse, active high,
RESET:internal via timer, typ. 2,5 s, external via 5 V-pulse Output pulse width 15 ns, (opt. 130 ns) GATE voltage V
gate
5 V: h to l set time Vcato V
ch-ent
+100 V : ~ 150 µs
5 V: l to h set time V
ca
to V
ch-ent
- 100 V : ~ 150 µs
Maximum ratings
Input voltage +5.5 V Operating temperature 5 to 40 °C Storage temperature -20 to 50 °C Weight ~ 350 g
Operating Conditions
Performance Characteristics
Spectral response
0.1
1
10
10 0
100 200 300 400 500 600 700 800 900
Wa velength (nm)
Quantumefficiency (%)
C 911
C 942
C 962
C 921
C 952
C 972
C 943
C 973 C 983
C 982
C 922
C 944
Count Rate Linearity
1E+0
1E+1
1E+2
1E+3
1E+4
1E+5
1E+6
1E+7
1E+8
1E-18 1E-17 1E-16 1E-15 1E-14 1E-13 1E-12 1E-11 1E-10
Light intensity [ W @ 410 nm ]
Counts
*) Additional models on request
Page 3
MD 1300 Series 1/2” Photon Counting Module
Technical Specifications
Supply voltage 5 V dc Input current at max. count rate < 280 mA Settling time < 1s (time to stabilize HV after supply voltage applied) Over-illumination protection: active quenching control (internal) Linear count rate: 5 MHz Output pulse: TTL, positive Active Quenching Control TTL-Pulse, active high,
RESET:internal via timer, typ. 2,5 s, external via 5 V-pulse Output pulse width 15 ns, (opt. 130 ns) GATE voltage V
gate
5 V: h to l set time Vcato V
ch-ent
+100 V : ~ 150 µs
5 V: l to h set time V
ca
to V
ch-ent
- 100 V : ~ 150 µs
Maximum ratings
Input voltage +5.5 V Operating temperature 5 to 40 °C Storage temperature -20 to 50 °C Weight ~ 420 g
Operating Conditions
Performance Characteristics
Spectral response
0.1
1
10
10 0
100 200 300 400 500 600 700 800 900
Wa velength (nm)
Quantumefficiency (%)
C 13 11
C 1342
C 1362
C 1321
C 1352
C 1372
C 1343
C 1373 C 1383
C 1382
C 1322
C 1344
Count Rate Linearity
1E+0
1E+1
1E+2
1E+3
1E+4
1E+5
1E+6
1E+7
1E+8
1E-18 1E-17 1E-16 1E-15 1E-14 1E-13 1E-12 1E-11 1E-10
Light intensity [ W @ 410 nm ]
Counts
*) Additional models on request
Page 4
Mp 1900 Series 3/4” Photon Counting Module
Technical Specifications
Operating Conditions
Performance Characteristics
Spectral response
0.1
1
10
10 0
100 200 300 400 500 600 700 800 900
Wa velength (nm)
Quantumefficiency (%)
C 19 11
C 1942
C 1962
C 1921
C 1952
C 1972
C 1943
C 1973 C 1983
C 1982
C 1322
C 1944
Count Rate Linearity
1E+0
1E+1
1E+2
1E+3
1E+4
1E+5
1E+6
1E+7
1E+8
1E-18 1E-17 1E-16 1E-15 1E-14 1E-13 1E-12 1E-11 1E-10
Light intensity [ W @ 410 nm ]
Counts
*) Additional models on request
Supply voltage 5 V dc Input current at max. count rate < 280 mA Settling time < 1s (time to stabilize HV after supply voltage applied) Over-illumination protection: active quenching control (internal) Linear count rate: 5 MHz Output pulse: TTL, positive Active Quenching Control TTL-Pulse, active high,
RESET:internal via timer, typ. 2,5 s, external via 5 V-pulse Output pulse width 15 ns, (opt. 130 ns) GATE voltage V
gate
5 V: h to l set time Vcato V
ch-ent
+100 V : ~ 150 µs
5 V: l to h set time Vcato V
ch-ent
- 100 V : ~ 150 µs Maximum ratings Input voltage +5.5 V Operating temperature 5 to 40 °C Storage temperature -20 to 50 °C Weight ~ 450 g
Page 5
www.perkinelmer.com/opto
© 2001 PerkinElmer, Inc. All rights reserved.
.
USA:
PerkinElmer Optoelectronics 2175 Mission College Blvd. Santa Clara, CA 95054 Phone: (408) 565-0850 Fax: (408) 565-0793
EUROPE:
PerkinElmer Optoelectronics GmbH Wenzel-Jaksch-Str. 31 65199 Wiesbaden Germany Phone: +49 611 492 534 Fax: +49 611 492 578
ASIA:
PerkinElmer Optoelectronics 47 Ayer Rajah Crescent #06-12 Singapore 139947 Phone: +65 775 2022 Fax: +65 775 1008
DS-289 Rev A1101
Dimensions (mm)
MP 900 Series
MP 1300 Series
MP 1900 Series
Red: Vcc, input (+5 V to +5.5 V DC) Black: GND Coax: External Gate in (TTL,active high) Yellow: Monitor voltage output (|V
ch-ent
/1000|)
Green: Control voltage input via ext.
OpAmp (ref. to manual)
BNC Connector:
- TTL Signal Output
- GND
Connections
· CAUTION: HIGH VOLTAGE WARNING
This product operates at high voltage. Extreme care must be taken to ensure operator safety and to avoid damage to other instruments. Avoid direct contact with the entrance window of the built in CPM when high voltage is applied. Avoid placing conductive material close to the cathode. Ensure that no light levels are applied, generating higher anode currents than specified. All given values are nominal/typical @ 20 °C ambient temperature; specification subject to change without notice
CPM
Loading...