DATASHEET
www.perkinelmer.com/opto
Description
The Channel Photomultiplier DCModule series is designed for all low
light applications e.g. in industries,
life science or medical analysis, in
high energy physics or pure research.
It is an easy to use module, containing
the Channel Photomultiplier,
a high voltage power supply, an
amplifier with I/U conversion
and an active quenching circuit for
high light protection. It is also
possible to apply an external gate
function. Strong variations in light
levels are possible due to the high
dynamic range of the installed CPM.
The exceptional low noise and high
sensitivity facilitates detection of
extremely low light levels.
Features
· High dynamic range
· No cooling required
· Very high stability in noise level
· Adjustable gain
· Active quenching
· Circuit for high light protection
· Gateable CPM input (only Bialkali types)
· Optical fiber read-out possible
· 5 volts operating voltage
· Monitor voltage output
MD Series CPM Modules
Ultra High Sensitivity Gateable Channel
Photomultiplier DC-Modules
Lighting Imaging Telecom Sensors
Channel Photomultipliers and Modules
.

MD 900 Series
1/3” DC Photosensor Module
Technical Specifications
Supply voltage 5 V dc
Supply current < 280 mA
Settling time < 1s (time to stabilize HV after supply voltage applied)
Over-illumination protection: active quenching control (internal)
Current to voltage conversion factor 1 V/ 20nA (other values on request)
Output signal 0 ... 10 V dc **)
Output offset ± 3mV
Gain up to 1 x 10
8
, via internal potentiometer (or by external control voltage
0...3 V via ext. OpAmp circuit)
Active Quenching Control TTL-Pulse, active high, RESET: internal via timer, typ. 1 s, external
via 5 V-pulse
Frequency bandwidth DC to 1 kHz (other values on request, max. 20 kHz)
GATE voltage V
gate
5 V: h to l set time Vcato V
ch-ent
+100 V : ~ 150 µs
5 V: l to h set time V
ca
to V
ch-ent
- 100 V : ~ 150 µs
High voltage pre-adjusted 1500 volts
Maximum ratings
Input voltage +5.5 V
Operating temperature 5 to 40 °C
Storage temperature -20 to 50 °C
Weight ~ 350 g
Operating Conditions
Performance Characteristics
0,1
1
10
100
100 200 300 400 500 600 700 800 900
Current Amplification vs Control Voltage
1,E+02
1,E+03
1,E+04
1,E+05
1,E+06
1,E+07
1,E+08
1,E+09
0,9 1,4 1,9 2,4 2,9
*) Additional models on request
**) for long term operation: max. average output voltage of <
5 V (anode current of <100 nA) is recommended

MD 1300 Series
1/2” DC Photosensor Module
Technical Specifications
Supply voltage 5 V dc
Supply current < 280 mA
Settling time < 1s (time to stabilize HV after supply voltage applied)
Over-illumination protection: active quenching control (internal)
Current to voltage conversion factor 1 V/ 20nA (other values on request)
Output signal 0 ... 10 V dc **)
Output offset ± 3mV
Gain up to 1 x 10
8
, via internal potentiometer (or by external control voltage
0...3 V via ext. OpAmp circuit)
Active Quenching Control TTL-Pulse, active high, RESET: internal via timer, typ. 1 s, external via
5 V-pulse
Frequency bandwidth DC to 1 kHz (other values on request, max. 20 kHz)
GATE voltage V
gate
5 V: h to l set time Vcato V
ch-ent
+100 V : ~ 150 µs
5 V: l to h set time V
ca
to V
ch-ent
- 100 V : ~ 150 µs
High voltage pre-adjusted 1500 volts
Maximum ratings
Input voltage +5.5 V
Operating temperature 5 to 40 °C
Storage temperature -20 to 50 °C
Weight ~ 420 g
Operating Conditions
Performance Characteristics
0.1
1
10
10 0
100 200 300 400 500 600 700 800 900
Current Amplification vs Control Voltage
1,E+02
1,E+03
1,E+04
1,E+05
1,E+06
1,E+07
1,E+08
1,E+09
0,9 1,4 1,9 2,4 2,9
*) Additional models on request
**) for long term operation: max. average output voltage of <
5 V (anode current of <100 nA) is recommended

MD 1900 Series
3/4” DC Photosensor Module
Technical Specifications
Operating Conditions
Performance Characteristics
0.1
1
10
10 0
100 200 300 400 500 600 700 800 900
Current Amplification vs Control Voltage
1,E+02
1,E+03
1,E+04
1,E+05
1,E+06
1,E+07
1,E+08
1,E+09
0,9 1,4 1,9 2,4 2,9
*) Additional models on request
Supply voltage 5 V dc
Supply current < 280 mA
Settling time < 1s (time to stabilize HV after supply voltage applied)
Over-illumination protection: active quenching control (internal)
Current to voltage conversion factor 1 V/ 20nA (other values on request)
Output signal 0 ... 10 V dc **)
Output offset ± 3mV
Gain up to 1 x 108, via internal potentiometer (or by external control voltage 0...3 V
via ext. OpAmp circuit)
Active Quenching Control TTL-Pulse, active high, RESET: internal via timer, typ.1s, external via
5 V-pulse
Frequency bandwidth DC to 1 kHz (other values on request, max. 20 kHz)
GATE voltage V
gate
5 V: h to l set time Vcato V
ch-ent
+100 V : ~ 150 µs
5 V: l to h set time Vca to V
ch-ent
- 100 V : ~ 150 µs
High voltage pre-adjusted 1500 volts
Maximum ratings
Input voltage +5.5 V
Operating temperature 5 to 40 °C
Storage temperature -20 to 50 °C
Weight ~ 450 g
**) for long term operation: max. average output voltage of <
5 V (anode current of <100 nA) is recommended

www.perkinelmer.com/opto
© 2001 PerkinElmer, Inc. All rights reserved.
USA:
PerkinElmer Optoelectronics
2175 Mission College Blvd.
Santa Clara, CA 95054
Phone: (408) 565-0850
Fax: (408) 565-0793
EUROPE:
PerkinElmer Optoelectronics GmbH
Wenzel-Jaksch-Str. 31
65199 Wiesbaden
Germany
Phone: +49 611 492 534
Fax: +49 611 492 578
ASIA:
PerkinElmer Optoelectronics
47 Ayer Rajah Crescent #06-12
Singapore 139947
Phone: +65 775 2022
Fax: +65 775 1008
DS-287 Rev A1101
Dimensions (mm)
MD 900 Series
MD 1300 Series
MD 1900 Series
Red: Vcc, input (+5 V to +5.5 V DC)
Black: GND
Coax: External Gate in (TTL,active high)
Yellow: Monitor voltage output (|V
ch-ent
/1000|)
Green: Control voltage input via ext.
OpAmp (ref. to manual)
BNC Connector:
- DC 0 - 10 V Signal Output
- GND
Connections
· CAUTION: HIGH VOLTAGE WARNING
This product operates at high voltage. Extreme care must be taken to ensure operator safety and to avoid damage to other
instruments. Avoid direct contact with the entrance window of the built in CPM when high voltage is applied. Avoid placing
conductive material close to the cathode.
Ensure that no light levels are applied, generating higher anode currents than specified.
All given values are nominal/typical @ 20 °C ambient temperature; specification subject to change without notice
.