Datasheet L8601-01 Datasheet (HAMAMATSU)

Page 1
90 kV
NEW
MICROFOCUS X-RAY SOURCE
L8601-01
FEATURES APPLICATIONS
● Focal Spot Size: 5 µm (at 4 W)
Providing 5µm focal spot size, for the sealed-off type X-ray
● Easy Handling
Convenient handling (with no high voltage wiring) since the
air cooled X-ray source and the high voltage power supply are in one package.
● Continuous Operation
It allows continuous maximum output operation.
● Built-in External Control Function
It is automatically operable from external control since it
has a built-in RS-232C interface.
● X-ray Non-destructive Test
• Printed Circuits • Electronic Components
• Semiconductor Devices • Ceramics
• Soldered Parts • Rubbers
• Combinated Materials • Plastics
• New materials • Die casting
● Multilayer Printed Circuit Board X-ray Inspection•Drilling Instrument
● Industrial On-line Process Control and Measurement
● Industrial X-ray Computer Tomography System
Page 2
SPECIFICATIONS
●GENERAL
Parameter
Target Voltage Target Current Maximum Output Power Focal Spot Size Beam Angle Minimum Distance Focus/Object Environmental X-ray Dose Leakage Operation
NOTE: a
In the range of output (= Tube voltage × Tube current) less than 10 W
Value/Description
20 to 90
0 to 250
10 (isowatts)
7 (5 µm at 4 W)
Continuous
a
39
9.5 1
Unit
kV µA
W
µm
degree
mm
mSv/h Max.
—
■X-RAY TUBE UNIT
Parameter Description
X-ray Tube Cooling Method Window Material Target Material Cathode Material Window Position High Voltage Power Supply Operation /Storage Temperature Operation/Storage Humidity Weight
NOTE: a
Without moisture condensation.
+10 to +40 °C / 0 to +50 °C
Sealed Type Convection Cooling Beryllium (150 µmt)
Tungsten
Dispenser Cathode
End-window
Built-in Type
85 % RH Max.
Approx. 6 kg
a
■X-RAY CONTROL UNIT
Parameter Description
Function
Protection External Control Operation /Storage Temperature Operation/Storage Humidity Power Consumption Input Voltage
Installation Method
Weight
Target Voltage/Current Pre-set, Auto Warm-up
Safety Interlock
RS-232C
0 to +40 °C
85 % RH Max.
200 W Max.
24 Vdc
Designed for both bench-top
and rack mounting
Approx. 5 kg
a
Page 3
DIMENSIONAL OUTLINE (Unit: mm)
X-RAY TUBE UNIT
X-RAY FOCAL SPOT
39°
9.5
24
90 KV MICROFOCUS X-RAY SOURCE
320.5 ± 2.0
135 ± 2
107 ± 2
70 5
X-RAY CONTROL UNIT
36.5
4-M3 L = 4
34
66
76
31
110.0 ± 0.5 16.5
40
290.0 ± 0.517
X-RAY UNIT INSTALLATION SCREW HOLES 4-M4 D5
MIN. 100
52 30
TLSOA0063EA
100 MIN.
X-RAY MODE
X-RAY
STANDBY
AGING
INTER LOCK
IDENTIFICATION PLATE
OFF
X-RAY
ON
INTER LOCK 2INTER LOCK 1AGING SIGNAL
LABEL
VOLTAGE
kV
CONTROL
PRE-SET
426 ± 29940
X-RAY90 CONTROL UNIT
CURRENT
µA
CONTROL
PRE-SET
OVER ERROR
OPERATE
ON
REMOTEOFF
50
99 ± 212
300 ± 2 30
482 ± 2
INTERCONNECTION CABLE CONNECTOR
OUTPUT
X-RAY REMOTE
RS-232C
DIA 33 mm MAX.
TLSOA0076EA
Page 4
RELATED PRODUCTS
X-RAY IMAGE INTENSIFIER CAMERA UNIT C7716, C7716-10
The C7716 series X-ray image intensifier (I. I.) camera unit now offers greatly improved X-ray detection efficiency even at low energy levels. This improve­ment stems primarily from a built-in X-ray image intensifier having an extreme­ly thin aluminum input window. Its thickness is only 0.3mm or less, virtually at the limit of present technology providing excellent X-ray transmit-tance and low scattering. The results are sharp, clear, high-quality images taken at low energy X-ray lev­els down to several keV which penetrate plastic (PET) materials.
X-RAY IMAGE INTENSIFIER CAMERA UNIT C7876, C7876-10
The C7876 is an X-ray image intensifier camera unit ideal for non-destructive inspection of light-element materials and radiation imaging at low-energy X-ray levels. The C7876 remarkably improves X-ray transmittance at low-energy X-ray lev­els by using a beryllium window instead of the aluminum window currently used for most X-ray image intensifiers. The result is a sharp and clear, high-contrast image taken in real time even at low energy X-ray levels down to a few keV.
90 KV MICROFOCUS X-RAY SOURCE
!
PRE-CAUTION TO USE
1. X-ray emitted from this device is harmful for human body. And it should be necessary for the operator to protect himself/herself
from it.
2. During an operation, the X-ray tube unit should be installed in the X-ray shielded cabinet in order to avoid any X-ray leakage.
Also the safety interlock system in X-ray control unit should be always used in order to avoid any misoperation.
# WARRANTY PERIOD
This device is guaranteed for one year from data of purchase or 3000 hours of operation, whichever comes first.
The warranty extends only to replacement of the products. The warranty does not cover damage due to misuse or natural galamity.
* This product(s) conforms to the EMC directive (89/336/EEC) of the European Union.
* To meet FDA standards, the Interlock-1 circuit incorporates a relay that shuts off the X-ray generation power supply circuit if the circuit is open.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2001 Hamamatsu Photonics K.K
HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.:
Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected]
Germany:
Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected]
France:
Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected]
United Kingdom:
North Europe: Italy:
Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [email protected]
Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected]
Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [email protected]
HOMEPAGE URL http://www.hamamatsu.com
TLSO1045E01 MAY. 2001 IP
Loading...