Datasheet ECM130PT32, ECM130PT31, ECM130PS48, ECM130PS28, ECM130PS24 Datasheet (IPSI)

...
Page 1
PRELIMINARY
Q3 2003
Specification
AC/DC Medical Open Frame
Small Size 3.00” x 5.00 x 1.28”
Low Leakage Current
Meets Class B, EN55011 Conducted
Medical Safety Agency Approvals
Zero Voltage Switching Technology
130 Watts
ECM130 Series
Input
AC Input Voltage 90-264 VAC Input Frequency 47-63 Hz Power Factor 0.98 at 115 VAC or 0.90 at 230 VAC Inrush Current 30 A max at 115 VAC
60 A max at 230 VAC cold start at 25°C
Input Current 2.6 A (rms) max at 115 VAC,
1.3 A (rms) max at 230 VAC Safety Ground 120 µA max at 115 VAC, 60 Hz Leakage Current 250 µA max at 240 VAC, 60 Hz Input Protection Dual fuse at 3.15 A each
Output
Output Voltage Single trim part for all outputs Output Power 130 Watts Output Voltage Adj. Fixed, not user adjustable Minimum Load 0 min load for single output models
10% for multiple output models Line Regulation ±0.5% maximum Load Regulation ±1%, ±5% for all secondary outputs
on multiple output models Ripple & Noise ±1% pk-pk maximum Transient Response 4% max deviation, 500 µs recovery
time for a 25% load change Temperature 0.05%/°C
Coefficient Hold Up Time >15 ms minimum Overvoltage 115% to 140% V1 only
Protection recycle input to reset Overcurrent 120% - 150% with auto recovery
Protection Remote Sense Output 1 only, compensates for up
to 0.5 V drop
General
Efficiency (Typical) >80% minimum nominal line
rated load, dependant on model
Power Density 6.7 W/in
3
single output models MTBF 100,000 hrs minimum Switching Frequency • 120 KHz ±10 KHz Isolation Voltage 4000 VAC Input to Output
1500 VAC Input to Ground 500 VAC Output to Ground
Size 5.00” x 3.00” x 1.28”
(1.35” for multi output models)
Weight 100 grams
Environmental
Operating 0 °C to 70 °C see derating chart Temperature Full power to 50 °C
Cooling 130 W with 18 CFM airflow
120 W with 10 CFM airflow
100 W with convection cooling Humidity 5 to 95% RH, non-condensing Storage -20 °C to +85 °C
Temperature
EMC & Safety
EMC EN61000-3 -2, EN60601-1 -2,
FCC 20780 Level B
EN50082-2 performance A,
no degradation of performance Safety Approvals EN60601-1, UL2601-1, CSA C22.2
No. 601.1
CE Mark to applicable sections of
MDD
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA
www.XPiQ.com
Intelligent Design Quality Product
XPiQ inc.
* Preliminary datasheet - see website for current specs
ECM130.qxd 15/5/03 9:10 am Page 1
Page 2
Rev 1.0
Mechanical Details - 130 Watts
1. 0 minimum load required for single output models. For multiple output models 5V requires 0.5A and +12V or +15V requires 0.05A minimum load.
Notes
OUTPUT VOLTAGE & CURRENT RATINGS ECM130
Output Output Output Current Ripple & Noise Model
Power Voltage Convection Cooled Max 18 CFM Pk-Pk
(
Number 130 W 5.0 V 20.0 A 26.0 A 50 mV ECM130PS05 130 W 12.0 V 9.0 A 11.0 A 120 mV ECM130PS12 130 W 15.0 V 7.0 A 8.7 A 150 mV ECM130PS15 130 W 24.0 V 4.5 A 5.4 A 240 mV ECM130PS24 130 W 28.0 V 3.9 A 4.6 A 280 mV ECM130PS28 130 W 48.0 V 2.3 A 2.7 A 480 mV ECM130PS48
5.0 V 10.0 A 15.0 A 50 mV
130 W 12.0 V 3.0 A 4.1 A 120 mV
ECM130PT31
-12.0 V 0.8 A 1.1 A 120 mV
5.0 V 10.0 A 15.0 A 50 mV ECM130PT32
130 W 15.0 V 2.5 A 4.1 A 150 mV
-15.0 V 0.8 A 1.1 A 150 mV
Derating Curve - 130 Watts
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA
www.XPiQ.com
Notes:
1. Dimensions in mm. Tolerance is ± 1mm.
2. AC input connector (TB1): Molex 5277-02A or equivalent.
3. DC output connector (TB2): Molex 5273 or equivalent.
4. Remote sense connector (TB3): Molex 5045-02A or equivalent.
Models
Pins PS05 PS12, 15, 24, 48 PT31 PT32
1 +V +V +5V +5.0 V 2 +V +V +5V +5.0 V 3 +V +V +5V +5.0 V 4 +V RET RET RET 5 RET RET RET RET 6 RET RET RET RET 7 RET N/A RET RET 8 RET N/A +12V +15.0 V 9 N/A N/A -12V -15.0 V
10 N/A N/A N/A N/A
* Preliminary datasheet - see website for current specs
*
* Total height of unit 1.28” (32.4 mm) for single output models, 1.35” (34.4 mm) for triple output models.
ECM130.qxd 15/5/03 9:10 am Page 2
Loading...