Dataq Instruments’ hardware and software products are not designed to be used in the diagnosis and treatment of
humans, nor are they to be used as critical components in any life-support systems whose failure to perform can
reasonably be expected to cause significant injury to humans.
Designed and manufactured in the
United States of America
ii
Warranty and Service Policy
Product Warranty
DATAQ Instruments, Inc. warrants that its hardware will be free from defects in materials and workmanship
under normal use and service for a period of one year from the date of shipment. DATAQ Instruments’ obligations
under this warranty shall not arise until the defective material is shipped freight prepaid to DATAQ Instruments.
The only responsibility of DATAQ Instruments under this warranty is to repair or replace, at its discretion and on
a free of charge basis, the defective material.
This warranty does not extend to products that have been repaired or altered by persons other than DATAQ
Instruments employees, or products that have been subjected to misuse, neglect, improper installation, or accident.
DATAQ Instruments shall have no liability for incidental or consequential damages of any kind arising out of the
sale, installation, or use of its products.
Service Policy
1. All products returned to DATAQ Instruments for service, regardless of warranty status, must be on a freightprepaid basis.
2. DATAQ Instruments will repair or replace any defective product within five days of its receipt.
3. For in-warranty repairs, DATAQ Instruments will return repaired items to the buyer freight prepaid. Out of
warranty repairs will be returned with freight prepaid and added to the service invoice.
Open Reference File................................................................................................................................... 17
Current Data...............................................................................................................................................17
Data Display .............................................................................................................................................. 17
User Annotation......................................................................................................................................... 17
User Annotation… .....................................................................................................................................19
Time Base..................................................................................................................................................20
Beep on File Full........................................................................................................................................ 20
Save Default Setup.....................................................................................................................................21
Next Palette................................................................................................................................................24
User Palette................................................................................................................................................24
Export (Using Save As…).......................................................................................................................... 31
Data Cursor................................................................................................................................................ 35
Peak on Screen........................................................................................................................................... 35
Valley on Screen ........................................................................................................................................ 35
Center Cursor............................................................................................................................................. 35
%EOF (Distance to End of File)................................................................................................................. 36
Enable Live Display................................................................................................................................... 36
Next Mark.................................................................................................................................................. 36
Insert Mark ................................................................................................................................................ 36
Clear Marks to TM .................................................................................................................................... 36
Extract Channels (Using Save As…)............................................................................................................. 37
Jump to Beginning/End of File................................................................................................................... 39
Go to TM Position ..................................................................................................................................... 39
Go to Range…........................................................................................................................................... 39
Go to Time…............................................................................................................................................. 39
Select Live Display.................................................................................................................................... 45
Enable Time Marker.................................................................................................................................. 49
Start Time .................................................................................................................................................. 50
End Time................................................................................................................................................... 50
Time per Division...................................................................................................................................... 50
Data Display.............................................................................................................................................. 50
Limit & Frequency Display........................................................................................................................ 50
User Annotation ......................................................................................................................................... 50
User Annotation…..................................................................................................................................... 50
Data Cursor................................................................................................................................................ 53
More Bottom Y.......................................................................................................................................... 55
iii
Less Bottom Y............................................................................................................................................ 55
Left Limit...................................................................................................................................................55
Right Limit ................................................................................................................................................ 55
Using ActiveX Controls...............................................................................................................75
1
Introduction
This manual contains information designed to
familiarize you with the features and functions of
the following serial port data recording modules:
DI-151RS
DI-195B
and also with W
software and W
and analysis software.
DI-151RS
The DI-151RS is a portable, two-channel data
recording module that communicates through your
computer’s RS-232 (or serial) port. It can record
data at rates up to 240 samples per second with 12bits of measurement accuracy.
The DI-151RS is powered directly from the
computer to which it is connected (through the RS232 cable).
The DI-151RS is equipped with a dedicated
thermistor input, two digital inputs, and two singleended, bipolar analog input channels.
DI-195B
The DI-195B is a two-channel data recording
module that communicates through your computer’s
RS-232 (or serial) port. It can record data at rates up
to 240 samples per second with 12-bits of
measurement accuracy.
The DI-195B is powered by the included wall outlettype adapter, which can be plugged into any singlephase, 120 volt, 60 Hz. AC source. It features two,
signal-conditioned, differential, bipolar analog
inputs (screw terminal access, maximum
measurement range is defined by the DI-5B
modules). The DI-195B allows you to plug any two
DI-5B signal conditioning modules into its
backplane. The DI-5B signal conditioning modules
then filter, isolate, amplify, and/or convert virtually
any industrial signal that is applied into a high level
analog voltage, perfectly suited for recording.
INDAQ/Lite Waveform Recording Software and
W
WINDAQ Waveform Browser Playback and
Analysis Software
The WINDAQ software shipped with all of the serial
port instruments consists of W
INDAQ/Lite waveform recording
INDAQ Waveform Browser playback
INDAQ/Lite waveform
recording software and W
playback and analysis software.
INDAQ/Lite waveform recording software can be
W
used with any Dataq Instruments hardware product.
It allows you to record waveforms directly and
continuously to disk while monitoring a real time
display of the waveforms on your computer screen.
It operates at the full sample rate of the instrument
being used (allowing you to see a real time display
of your waveform signals on your computer’s
monitor at the instruments full sample rate), but is
limited to 240 Hz maximum throughput when you
enter the record mode.
INDAQ Waveform Browser allows you to review,
W
measure, analyze, compress, cut-and-paste,
export/import, and otherwise manipulate the
recorded waveform information. It is available for
download free-of-charge from our web site
(www.dataq.com).
INDAQ Waveform Browser
Conventions Used in the Documentation
Before you start using the DI-151RS, DI-195B,
INDAQ/Lite, or WINDAQ Waveform Browser
W
software, it’s important to understand the terms and
notational conventions used in this documentation.
General Conventions
Commands you choose are given with the menu
name preceding the command name. For example,
the phrase “Choose File Open” tells you to choose
the Open command from the File menu. This
naming convention describes the sequence you
follow in choosing a command — first you select
the menu, then you choose the command.
Mouse Conventions
In general, most mouse actions require only the left
mouse button. For example, carrying out a menu
command or working in a dialog box requires only
the left mouse button. However, the right mouse
button is not totally neglected. Among other things,
the right mouse button is used for copying waveform
data to the clipboard, waveform scaling, selecting a
waveform channel, etc.
Since the majority of mouse procedures are done
with the left mouse button, we will not specify
which mouse button to click, drag, or double-click
with in the procedures unless it is the RIGHT
mouse button. When the right mouse button is
required, it will be specified as such. For example,
“Double-click the right mouse button anywhere in
the bottom annotation line to move the cursor to the
2
lowest displayed waveform valley.” When not
specified, the left mouse button is assumed for the
procedure.
• “Point” means to position the mouse pointer
until the tip of the pointer rests on what you
want to point to on the screen. For example,
“Point to the View menu.”
• “Click” means to press and immediately
release the mouse button without moving the
mouse. For example, “To display the menu
that contains the command you want, click the
menu name in the menu bar.”
• “Double-click” means to click the mouse
button twice in rapid succession. For example,
“Double-click the icon to start the program.”
• “Drag” means to press the mouse button and
hold it down while you move the mouse; then
release the button. For example, “Drag down
to Data Cursor to enabled the cursor for onscreen display.”
WINDAQ/Lite Operating Modes
WINDAQ/Lite waveform recording software has
three operating modes. They are, in logical order of
occurrence; Setup, Record, and Standby. Each
operating mode shares many features with the other
three, but there are some features accessible only
from one given mode. The three modes are
discussed in detail as follows:
selected from the File menu. RECORD is used to
stream waveform information to disk. While in this
mode, all waveform recording features and functions
are available (with some restrictions) except the
following: all channel-specific operations (i.e.,
channel number, gain, offset); and sample rate
adjustments. In this mode, waveform information is
being continuously streamed to disk while the real
time display remains active. In this mode, sample
rate is limited to 240 Hz maximum throughput,
regardless of the instrument being used. When in
this mode, the Status: area of the bottom annotation
line displays RECORD.
The STANDBY operating mode becomes active
when waveform recording to disk is temporarily
suspended by choosing the Stop command from the
File menu. While in this operating mode, all
waveform recording features and functions are
available except: channel-specific operations (i.e.,
channel number, offset, etc.) and sample rate
adjustments. In this mode only the real time display
is active, waveform recording to disk has been
stopped. When in this mode, the Status: area of the
bottom annotation line displays STBY. It is possible
to start and stop the recording process as many times
as you like during a data acquisition session.
The SET UP operating mode is active when
INDAQ/Lite is first started. SET UP is used to
W
configure data acquisition parameters such as the
number of acquired channels, channel gain, and
channel offset. In addition, this mode is used to
adjust the real time display for a visually pleasing
presentation. Virtually all other data acquisition
functions and adjustments are available from the
SET UP operating mode including control of the real
time display screen’s scaling and offset functions. In
this mode only the real time display is active, data is
not yet being streamed (or stored) to disk. In this
mode, sample rate is limited only by the capabilities
of the hardware. In other words, you can sample as
fast as your instrument will allow. When in this
mode, the Status: area of the bottom annotation line
displays SET-UP.
The RECORD operating mode is active when a file
has been opened and the Record command has been
3
Getting Started
Step 1
Installing the Hardware
Step 2
The following items are included with each
instrument purchase. Verify that you have the
following:
• DI-151RS, or DI-195B instrument.
• Communications cable designed to connect the
instrument to your computer’s serial port (with the
DI-195B, the AC power adapter is integrated into
the serial cable).
INDAQ software diskette(s). We are currently
• W
shipping both W
Waveform Browser software all on one disk.
However that may change at any time, in which
case you will have two diskettes.
• This user’s manual, which covers the hardware
aspects of the DI-151RS, and DI-195B serial port
modules and also documents W
waveform recording software and W
Waveform Browser playback and analysis
software.
• If you purchased a DI-195B, you should also have
one or two DI-5B plug-in modules. Note that
these are extra-cost items not included with the
purchase of the DI-195B kit, but are required for
operation.
If an item is missing or damaged, call Dataq
Instruments at (330) 668-1444. We will guide you
through the appropriate steps for replacing missing
or damaged items. Save the original packing
material in the unlikely event that your unit must,
for any reason, be sent back to Dataq Instruments.
INDAQ/Lite and WINDAQ
INDAQ/Lite
INDAQ
If you have an instrument other
than the DI-151RS or DI-195B and
you have not installed it yet, do so
now. Ignore the following steps
and proceed instead with the
installation instructions contained
in your hardware User’s Manual.
DI-151RS and DI-195B Only
1.Connect the male end of the communications
cable to the 9-pin female connector on the DI151RS or DI-195B module.
2.Connect the other end of the communications
cable to your computer’s serial or COM port.
3.If you purchased the DI-195B, plug the AC
power adapter into any standard 120V, 60Hz
wall outlet.
Installing WINDAQ
Software
NOTE
INDAQ/Lite waveform recording
W
software and W
Browser playback and analysis
software are both on the W
Resource CD, but they must be
installed individually with separate
installation procedures. The
following procedure will install
INDAQ/Lite recording software
W
first, then W
Browser playback and analysis
software.
1.Start Windows™.
INDAQ Waveform
INDAQ
INDAQ Waveform
NOTE
Before installing W
software, you should already have
the hardware installed or
connected. If you have not done so
already, install or connect your
hardware at this time.
Use the following procedure to
connect the DI-151RS, or DI-195B
module to your computer.
INDAQ
2.Insert the W
drive and close the drive tray.
For most user’s, the CD’s auto-run feature will
automatically display a list of options for you to
pick from. If you do not see this list of options
after a reasonable period of time, double-click
the My Computer icon on your desktop and then
double-click your CD-ROM icon to manually
display the list of options.
3.Choose the Install Software button and click
OK.
INDAQ Resource CD into your CD
4
4.Now choose the WINDAQ/Lite and Software
Step 3
Signals to the Instrument
Development Kit button and click OK.
5.Specify the instrument series (broad category)
that will be used with W
INDAQ/Lite software.
For example, choose the DI-1xx SerialData Acquisition Units button if you
have either the DI-151RS module or
the DI-195B module connected to your
serial port (both are 100 Series
instruments), or choose the DI-4xxSeries Plug-in Cards button if you
have a DI-400, DI-401, or DI-410
board installed, etc.
When the instrument series button is
selected, click the OK button.
6.The remaining installation steps vary by
instrument. In most cases, the on-screen
prompts provide enough information to
successfully get you through the installation.
However, if you are unsure of what to do next
or if you need additional information, contact
Dataq Instruments technical support.
If you have an instrument other
than the DI-151RS or DI-195B,
ignore the following steps and
proceed instead with the input
signal connection instructions
contained in your hardware User’s
Manual.
To connect signals to the DI-151RS or DI-195B,
insert the stripped end of a signal lead into the
desired terminal directly under the screw. Tighten
the pressure flap by rotating the screw clockwise
with a small screwdriver. Make sure that the
pressure flap tightens only against the signal wire
and not the wire insulation. Do not overtighten. Tug
gently on the signal lead to ensure that it is firmly
secured.
DI-151RS
All user connections are made to the 8-port screw
terminal connector as follows:
7.When W
INDAQ/Lite software has successfully
been installed, double-click the My Computer
icon on your desktop and then double-click your
CD-ROM icon to manually display the list of
installation options.
8.Again choose the Install Software button and
click OK.
9.Select the W
and click OK to install W
INDAQ Waveform Browser button
INDAQ Waveform
Browser playback and analysis software. The
on-screen prompts should provide enough
information to successfully get you through the
installation. However, if you need additional
information, contact Dataq Instruments
technical support.
Connecting Your Input
NOTE
Use the following procedure to
connect your input signals to the
DI-151RS or DI-195B module.
Ch 1: Channel 1 signal input.
Gnd: Channel 1 signal ground.
Therm 1: Thermistor lead 1 input.
Therm 2: Thermistor lead 2 input.
Dig 0: Digital input 0.
Dig 1: Digital input 1.
Gnd: Channel 2 signal ground.
Ch 2: Channel 2 signal input.
Your DI-151RS comes with the on-board waveform
generator configured to provide a sample input
signal. The onboard waveform generator is enabled
by connecting a jumper wire between the Ch 1 and
Dig 1 inputs. A thermistor is also included for
temperature measurements. So it doesn’t get lost
during shipping, one leg of the thermistor is
5
connected to a screw terminal. Unless you would
source
ground
(common mode voltage)
gnd
input terminals
signal source
shield
amplifier
circulating currents
noise on signal of interest.
like to make temperature measurements with the
thermistor, remove it completely from the DI151RS. Do not make voltage measurements with the
thermistor installed. Any voltage measurement made
with the thermistor installed will be inaccurate.
When using the termistor to make temperature
measurements, insert one thermistor leg into input
Therm 1 and the other thermistor leg into Therm 2.
In order to get accurate temperature measurements,
the thermistor must be calibrated. The procedure for
doing this is included in step 8.
signal to be measured is ground-referenced, or the
signal to be measured is isolated from ground. In
either case, it is important to properly connect the
differential amplifier.
Ground Referenced Signal Sources — A ground
referenced signal source is one with a local ground
that may not be (and probably is not) at the same
potential as the computer’s ground. This potential
difference between signal ground and computer
ground is referred to as a common mode voltage and
is caused by a number of different factors.
When the DI-151RS is configured for voltage
measurements, Ch 1 and its Gnd make up one
single-ended analog input channel, while Ch 2 and
its Gnd make up the other.
DI-195B
All user connections are made to the two, four-port
screw terminal connectors.
IN+ and IN- are the signal input terminals for each
channel. EX+ and EX- are excitation voltage
outputs for those transducers that require excitation.
Obviously, you’ll want to tailor the DI-5B signal
conditioning modules used in the instrument to the
kind of input signals you wish to record.
signal
ground potential difference
computer
The most common of these is different physical
locations of the computer and signal ground points.
Since wire is not a perfect conductor (i.e., exhibiting
zero resistance regardless of length) a voltage drop,
however small, will always be present. The
differential amplifier is unique in its ability to
measure signals originating from sources with
different ground potentials relative to the computer
providing it is connected properly.
Circulating currents in signal shield induce
WRONG APPROACH!
signal
lead
instrument
differential
Differential Inputs
DI-195B features two dedicated differential input
channels. Differential input channels respond only to
the difference in voltage between each (+) and (-)
analog input, effectively suppressing common mode
voltages (i.e., identical voltages appearing
simultaneously and in phase on both inputs). Since
most forms of interference or noise are applied with
equal intensity to both inputs of a differential
amplifier (that is, in the common mode), the
differential amplifier is very effective at rejecting
noise riding on a signal of interest.
Two types of signal sources are likely to be
encountered when configuring your DI-195B; the
Ground loop
caused by
The most common error made in connecting
differential amplifiers is the tendency to ground both
ends of a signal shield. This causes current to flow
through the shield and induces noise on the signal to
be measured. This problem is eliminated by
ensuring that only one ground exists on the signal
circuit.
6
gnd
amplifier
input terminals
(no path for current to circulate)
shield
signal source
√
One ground on the signal circuit eliminates
noise-inducing ground loops
input terminals
amplifier
gnd
shield
Floating signal source
Insulated outer jacket
Foil shield
Twisted signal pair
Stranded copper conductors
Drain wire
RIGHT APPROACH
instrument
signal
lead
Potential difference
differential
Isolated Signal Sources — An isolated signal source
is one that is not connected to ground at any point.
However the signal shield should still be grounded
as before to reduce noise.
instrument
signal
lead
differential
Any application requiring a differential amplifier
also defines a need for quality signal cable. Four
elements combine to ensure adequate quality signal
cable; a twisted signal pair with low resistance
stranded copper conductors, surrounded by a
multiple-folded foil shield, with a copper stranded
drain wire, all contained within an insulated outer
jacket.
In applications where such signal cable is used, a
dramatic decrease in noise pickup will be
experienced. The drain wire should be considered as
the shield and should be connected as described
previously. Signal cable meeting all four criteria for
quality is Belden No. 8641.
7
Fast Start to
Step 4
Acquisition
Recording
Waveforms
with
Point to the Edit menu and drag down to
Channels…
This causes the Ch: block cursor to blink in the
Channels: field on the status bar. The currently
active list of enabled channels are displayed in this
field, immediately to the left of the cursor (initially,
only channel 1 is enabled for recording).
W
After you have completed all hardware and software
installations and connected your signals, you are
ready to record.
Steps 4 through 9 below outline the “fast start”
procedure for recording waveforms to disk. These
are the basic steps you will take every time you
record with W
software.
The rest of this chapter contains additional recording
functions that will not necessarily be used in every
session or are not requisite for recording, but will
make future recording sessions easier. After your
initial, “get-acquainted” recording session, take a
few moments and look over these functions.
It is assumed at this point that you have
W
If not, click the Start button on the taskbar, point to
Programs, point to WINDAQ (or whatever you
named the group window during step 11 of the
installation), and click WinDaq Acq.
This step specifies which analog channels to record.
For example, suppose we have analog signals
connected to channels 1 and 2, and we want to
record both of them. Use this procedure to enable
channels 1 and 2 for recording:
INDAQ/Lite
INDAQ/Lite waveform recording
Enabling Channels for
INDAQ/Lite waveform recording software running.
2.Enter the desired channel number and press the
ENTER key. Repeat for each desired channel.
A single-ended channel is added to the list by
entering the channel’s number and pressing the
ENTER key.
A differential channel is added to the list by
entering the channel’s number followed by the
letter “D” (e.g., “3D”) and pressing the ENTER
key.
This causes the channel number to be displayed in
the enabled channels row.
3.Disable channel configuration by pressing any
key other than the letters “D” and “I”.
Note that a channel may be removed by typing the
channel number preceded by a minus (“-”) sign.
Make sure that you use the minus sign on the main
keyboard, not the minus sign that resides on the
keypad (the minus sign on the keypad performs an
entirely different function). For example, typing “1” removes channel 1 from the row. Note also that it
is not possible to remove all the channels from the
row (in other words, have no channels enabled). For
example, say you are using the DI-195B and you
have only channel 1 configured and it is a
differential channel (Channels: field reads “1D”).
Typing “-1D” will not remove it from the row.
Typing “1” will however, change it to a singleended channel, and vice-versa.
Channels may not be added after the RECORD
mode has been entered.
1.To enable channel configuration:
8
Step 5
Viewing Enabled
Channels
Step 6
This step allows you to display onscreen all enabled
channels.
If you enabled more channels in the previous step,
you will want to display them all onscreen.
1.To display enabled channels:
1. Point to the View menu and drag down to
Format Screen.
2. In the format box, click the desired
display format.
Any channel may be displayed as an overlapping or
non-overlapping display.
Specifying a Sample Rate
You must specify a sample throughput rate (or total
scanning rate) for waveform recording.
For example, if you wanted to acquire four channels
of data all at 60 Hz, you would require a sample rate
of 240 Hz.
To specify a sample rate:
1.Select the Sample Rate command:
Point to the Edit menu and drag down to
Sample Rate…
This displays the sample rate dialog box as follows:
An overlapped format allows you to closely examine
the time and amplitude relationship of two
waveforms. A non-overlapped format may be used
to isolate each waveform’s transition to within a
defined area. Whatever the timing relationship of
your input signals, any display format may be
enabled at any time without affecting the waveform
information being stored to disk. Display format
changes only affect the way you view waveforms on
your computer’s monitor.
2.Enter the desired throughput rate (or total
scanning rate) in samples per second in the
Sample Rate dialog box.
3.Click OK or press the ENTER key.
INDAQ/Lite automatically allocates the correct
W
amount of buffer space when you input a sample
rate. The Input Buffer Size and Disk Buffer Size
values are displayed for informational purposes
only.
Keep in mind that the sample rate is actually a
throughput rate. The sample rate per channel
(obtained by dividing the selected sample
throughput rate by the number of channels enabled)
is displayed in the lower left corner of the window
as S/s/CHAN:.
9
Step 7
Measurement Range
skip this step. Otherwise, continue with this step to
select a higher gain factor (thus narrower
measurement range) for the best possible
measurement resolution.
Since gain and measurement range can be assigned
per-channel, the desired channel must first be
selected.
1.Select an analog channel for gain adjustment:
Although thousands of different sample rates may be
configured, some discrete rates are not achievable.
This limitation is due to the inability to divide the
master clock in a manner that results in an evennumbered quotient. For example, say you enter a
sample rate of 200. What you might actually get,
because of the previously stated limitation, is 200.3.
Specifying Gain and
NOTE
DI-151RS, DI-195B, and DI-401 Users
DI-151RS, DI-195B and DI-401 users can
ignore this step and proceed with step 8.
DI-151RS instruments have a fixed gain of
1 (unity) and a fixed measurement range of
±10 volts. DI-195B and DI-401 instruments
have a fixed gain of 1 (unity) and a fixed
measurement range of ±5 volts.
Click the left mouse button in the unselected
channel’s annotation margin.
The focal point for selecting a channel for any type
of adjustment is the variable waveform assignment
indicator.
Choosing the proper gain factor for your input
signals allows you to get the best possible
measurement resolution out of your instrument.
The default W
gain of 1 provides the widest possible measurement
range. For example, say you’re using a DI-400-PGH.
This instrument features a maximum analog
measurement range of ±10 volts and programmable
gain factors of 1, 2, 4, and 8. At a gain of 1 (gain
factor = 1), the full scale measurement range is ±10
volts. However, if we set the gain to 2, the full scale
measurement range becomes ±5 volts. The full scale
measurement range gets even smaller for gain
factors of 4 (range = ±2.5 volts) and 8 (range =
±1.25 volts).
If the signals you plan to record need the wide
measurement range that a gain of 1 provides, then
INDAQ/Lite gain setting is 1 (unity). A
This two-element equality (X=Y) is displayed for
every enabled channel. Element “X” (1 in the
illustration above) is the window number. Element
“Y” (0 in the illustration above) is the analog
channel assigned to that window.
When selected, a box surrounds the variable
waveform assignment indicator, indicating that the
window is enabled for adjustment.
10
When the desired channel is selected, gain
adjustments to that channel are now possible.
2.Select the Channel Settings… command:
Point to the Edit menu and drag down to
Channel Settings…
This displays the “Channel X Settings” dialog box
(where X represents the enabled channel):
The gain factors you see displayed at the top of this
dialog box are hardware dependent. Displayed next
to each gain are the equivalent plus and minus full
scale voltages that result from selecting that gain
factor. These voltages vary by hardware device.
leave the Unipolar box unchecked, thus specifying
bipolar.
Reset Cal. Button
Whenever the gain or unipolar/bipolar setting is
changed, any calibration is preserved and adjusted to
reflect the new gain or unipolar/bipolar setting.
Selecting Reset Cal allows you to reset the selected
channel from a significant engineering units
calibration back to the default voltage (or
temperature, if using thermocouples) calibration.
Acquisition Method
Allows you to select the method used for recording
and displaying data (intelligent oversampling).
All Dataq Instruments hardware products
continuously sample data using a burst sampling
technique. The hardware samples data at one rate
(referred to as the burst rate) while your computer
reports (i.e., displays and stores) this data at another
rate (called the sample rate or throughput rate). For
example, let’s say we want to record one channel at
100 Hz and our burst rate is set at 50kHz. This
means that for every 500 data points sampled, only 1
will be reported. The dilemma becomes; which data
point out of the 500 gets reported? Fortunately, you
have a choice for reporting this single data point:
Average - Averages all of the data points in the burst
sample and reports this average as the single value
for storage and display.
3.Click on the desired gain factor.
4.Click OK or press the ENTER key.
As the gain factor for a channel is sequenced upward
and downward through gain ranges, ±full scale
annotation adjusts to reflect the application of the
various gain factors.
A channel's gain factor cannot be adjusted following
the initiation of data storage to disk. Only the SETUP operating mode allows gain factor adjustments.
Unipolar/Bipolar Measurement Range Option
You can specify whether the signal you will be
acquiring is unipolar or bipolar. A unipolar signal is
one that ranges from 0 to + full scale (never goes
below zero), and a bipolar signal is one that ranges
from some negative value to some positive value (is
both positive and negative going). If your signal is
only a positive going waveform or a negative going
waveform, always select the Unipolar check box. If
your signal is both positive and negative going,
Last Point - Reports the last input data point in the
burst sample, ignoring the rest.
Maximum - Reports the highest value data point in
the burst sample, ignoring the rest.
Minimum - Reports the lowest value data point in
the burst sample, ignoring the rest.
Input Types
Linear - The default setting, available on all
instruments, should be used with all linear inputs.
Thermocouple - Selectable only on instruments with
the hardware to support thermocouples.
Nonlinear - Also only on instruments that support
thermocouples. Used only with a specialized, nonlinear, non thermocouple type of sensor.
Next and Previous Buttons
Allow you to step through the next or previous
enabled channel (in order), setting up each channel
11
for gain, unipolar/bipolar, sample rate, etc., without
Step 8
Signals
closing this dialog box.
To calibrate a waveform using this high/low
calibration method:
Calibrating Your Input
Calibration allows you to view and record your
waveform data in significant engineering units.
Without calibrating, waveform data is scaled,
annotated, and displayed in volts, which in many
cases is not very meaningful. Calibration is used to
give meaning to the displayed data. Instead of volts,
your data can be scaled, annotated, and displayed in
significant engineering units (i.e., PSI, RPM, lbs,
ft/s, etc.).
This function allows your waveform data to be
calibrated in significant engineering units using any
known or verified transducer output values.
For example, suppose you were acquiring pressure
data with a transducer that had a known output of -1
volt at 0 psi and 4.5 volts at 100 psi. Whether these
output levels were read from the manufacturers
specifications or verified through use, they can be
used as calibration values.
With this method, you can even calibrate the data
without knowing the transducer output relationship.
Simply let W
output voltage values for you. For example, suppose
you were acquiring pressure data and you had no
idea what kind of output voltages to expect from the
transducer you’re using. You could vent the pressure
apparatus to quiescence or atmospheric pressure and
observe what voltage the transducer is outputting at
quiescence from the Input Level text box in the LowCalibration dialog box (select Low Calibration…
from the Edit menu). In this same dialog box, enter
what you would like this quiescent state to be, such
as “0”, in the Low Cal Value text box, and enter
“PSI” in the Engr. Units text box. Now activate the
OK command button to close the dialog box, and
increase the pressure until the manometer reads a
convenient value, say 100 psi. In a similar manner,
observe what voltage the transducer is outputting at
this 100 psi level from the Input Level text box in
the High Calibration dialog box (select High
Calibration… from the Edit menu). Enter “100” in
the High Cal Value text box, enter “psi” in the
Engr. Units text box, choose the OK command
button, and you’re done.
INDAQ/Lite software establish these
1.Select the window that contains the desired
waveform:
Click the left mouse button in the
unselected window’s annotation margin.
2.If not already enabled, enable the Current Data
and Data Display options (on the Options
menu).
3.Select Low Calibration… as follows:
Point to the Edit menu and drag down to
Low Calibration…
This displays the Low Calibration dialog box:
This is where you enter the low calibration data
value that is to be assigned to the input voltage
level. For instance, referring back to our
previous examples:
a)For the case where the transducer output is
known,
12
-1 volt
at
4.5 volts
at
Input level
Cal value
Engr. units
This is where you enter the high calibration data
value that is to be assigned to the input voltage
level. For instance, referring back to our previous
examples:
0PSI
100PSI
we would enter “0” in the Low Cal Value text
box, “-1” in the Input Level text box, and “psi”
in the Engr. Units text box.
b) For the case where the transducer output is
unknown, we would also enter “0” in the LowCal Value text box (with the test apparatus
vented to atmospheric pressure), ignore the
Input Level text box (W
INDAQ/Lite software
has established the input voltage level for you),
and enter “psi” in the Engr. Units text box.
Press the TAB key to position the insertion
point in the various text boxes.
Type up to a four character optional engineering
unit tag (i.e., PSI, f/s, rpm, etc.).
4.Complete the low calibration entry as follows:
Click OK.
The Low Calibration dialog box disappears and if no
high calibration has been done yet, an offset is
applied to the waveform to achieve the entered
value. But you are only half done, the high
calibration value must also be entered.
5.Select High Calibration… as follows:
a) For the case where the transducer output is
known, we would enter “100” in the High CalValue text box, “4.5” in the Input Level text
box, and “psi” in the Engr. Units text box.
b) For the case where the transducer output is
unknown, we would also enter “100” in the
High Cal Value text box (after 100 psi has been
applied in our example), ignore the Input Level
text box (W
INDAQ/Lite software has established
the input voltage level for you), and enter “psi”
in the Engr. Units text box.
6.Complete the high calibration entry as follows:
Click OK.
The High Calibration dialog box disappears and the
waveform channel displays all waveform data values
calibrated in the chosen engineering units.
If desired, the upper and lower grid limit values may
be specified (and the waveform automatically scaled
to fit). This feature allows the channel's full scale
display span to be shown in whole numbers rather
than obscure fractional values, which sometimes
result from calibrating. For example, a waveform
scaled in mmHg can be automatically expanded or
contracted to fit within grid limits of 0 to 100
mmHg, or -200 mmHg to +200 mmHg, or whatever
is desired. Note that a channel spanning 0 - 200
mmHg is much easier to read and interpret on a per
division basis than one that spans 10.372 - 189.628
mmHg.
Point to the Edit menu and drag down to
High Calibration….
This displays the High Calibration dialog box:
To specify convenient upper and lower chart edge
values:
7.Select Scaling Limits… as follows:
Point to the Scaling menu and drag
down to Limits….
This displays the Display Limits dialog box:
13
8.Type convenient top and bottom chart-edge
Step 9
Disk
values in their respective text boxes.
9.Complete the chart-edge value entry as follows:
1.Select the Record command:
Point to the File menu and drag down to
Record.
Click OK.
Your specified values will now be displayed on the
upper and lower chart-edges and the waveform will
be automatically scaled to fit the new values.
The Next and Previous buttons allow you to step
through all of the enabled channels in order, entering
the desired display limits for each channel, without
closing this dialog box.
When using the DI-151RS, the following procedure
can be used to calibrate the thermistor transducer:
1. With the thermistor connected to the DI-151RS
as described in step 3 of this fast start
procedure, submerge the tip of the thermistor in
an ice water bath. Submerge only the tip of the
thermistor, do not submerge the DI-151RS. Let
it remain in the ice water long enough for the
thermistor output to stabilize (you can monitor
the thermistor output with W
time display).
2. Follow steps 1 through 4 of the calibration
procedure described above, but enter “32” in the
Low Cal Value text box, and “F” in the Engr.
Units text box. Note that W
established the Input Level value automatically.
INDAQ/Lite’s real
INDAQ/Lite has
This displays the Open dialog box. Here is where
you specify a file name (to create a new file) or open
an existing file (to append to it).
If creating a new file, you will be prompted to
specify a maximum file size. If appending to an
existing file, you will be prompted to specify how
much additional data to add to the existing file.
In either case, you can specify file size in kilobytes
or recording time. In the case of recording time, file
size is based on the currently selected sample rate.
Note that you can experiment with different file
sizes, sample rates, and recording time to get the
desired result. Move between the two text boxes
with the TAB key.
2.Enter the desired file size or accept the default.
3.Close the dialog box and begin recording by
clicking OK or pressing the ENTER key.
When you enter the desired file size or accept the
default, you are placed in the active record-to-disk
operating mode where enabled analog channel data
is simultaneously digitized, displayed, and streamed
to disk. The status field in the bottom annotation
line indicates RECORD.
3. Now remove the thermistor from the ice water
bath and allow it to stabilize at room
temperature.
4. Follow steps 5 through 9 above, but enter “72”
(or whatever your room temperature is) in the
High Cal Value text box, and “F” in the Engr.
Units text box. For convenient display limit
values, enter “100” in the Top Limit text box
and “0” in the Bottom Limit text box.
Recording Waveforms to
The Record function on the File menu initiates
recording to disk. To begin waveform recording:
To stop acquisition to disk:
Point to the File menu and drag down to
Stop.
14
This suspends data storage to disk, but the real time
display remains active. The status field in the
bottom annotation line indicates STBY when
acquisition has been stopped. The Stop function on
the File menu only becomes selectable when in the
record-to-disk mode.
As data is being recorded to disk, the amount of file
space consumed is displayed in the “Storage:” field
of the status bar as a percentage of total file space
consumed.
Data acquisition to disk may be started and stopped
as many times as desired. When the target data file
has been filled, data acquisition to disk will cease,
the “Storage:” field of the status bar will display
“100% used”, and the “Status” field of the status bar
will display FILE FULL.
cut and paste portions of waveforms, and do all
forms of interpretive analysis, launch the W
INDAQ
Waveform Browser by clicking or double-clicking
on the W
INDAQ Waveform Browser icon (created
during installation). Then refer to the section of this
manual titled “Reviewing Recorded Waveforms
with W
INDAQ Waveform Browser” for complete
details.
Each time recording is activated, an event marker is
automatically inserted in the data stream. This
allows acquisition break points to be reviewed using
INDAQ Waveform Browser playback and analysis
W
software.
This concludes the “fast start” procedure for
recording waveforms to disk. The next logical step
is to review the data you just recorded. This is done
with W
INDAQ Waveform Browser playback and
analysis software (WWB). To make timing and
amplitude measurements, review event markers,
overlap waveforms to examine interdependencies,
15
Additional
/Lite
Browse
/Lite
Pause Graphics
Waveform
Recording
Functions
These functions are not requisite for recording or
will not necessarily be used every time
INDAQ/Lite is used, but are included to make
W
recording easier and more productive, or to increase
display performance and visual appeal.
WINDAQ
periodically pointing to the View menu and
selecting Refresh.
AQ
WIND
Scroll Mode,
Oscilloscope Mode, &
Waveform information can be displayed on your
monitor in either a continuous smooth scroll mode
or in a triggered sweep oscilloscope mode.
This function starts WINDAQ Waveform Browser
playback and analysis software directly from
INDAQ/Lite recording software. This enables you
W
to have both software packages running
simultaneously to playback, review, analyze,
measure, quantify, etc. the data that was just
recorded.
If you are in the record mode when Browse is
selected, W
automatically open the very same file you are
actively recording to.
If you are not in the record mode when Browse is
selected, W
you for the file you wish to review.
To enable this multitasking feature:
INDAQ Waveform Browser will
INDAQ Waveform Browser will prompt
Point to the File menu and drag down to
Browse.
Either display mode may be enabled at any time
during data acquisition without affecting the
information being stored to disk. Display mode
changes only affect the way waveforms are
displayed on your monitor.
1.Select a display mode:
Point to the Options menu and drag down to
Scroll Mode or Oscilloscope Mode…
Note that in either display mode, the waveform
display can be paused. This allows you to get a
stationary look at some point on the waveform.
While paused, the Mode: field in the status bar
changes from the chosen display mode to Paused.
This starts the W
waveform data is constantly being written to this
INDAQ Waveform Browser file while
W
INDAQ/Lite is actively recording, you must
W
request updates to see all of the latest data. This is
done in W
INDAQ Waveform Browser. Since
INDAQ Waveform Browser by
16
To enable/disable the waveform display pause
feature:
Point to the Options menu and drag down to
Pause Graphics.
When pause is enabled, a subsequent pause attempt
disables the paused waveform. The display mode
that was active before pausing is restored.
If the display is paused while in the Record
operating mode, data acquisition to disk continues
unimpeded, only the display is paused.
Slope option buttons - The slope option buttons
allow you to select the slope (positive or negative) at
which the waveform display will trigger. Selecting
the + option button triggers the display to start at the
trigger level, but on the waveform’s rising slope.
Selecting the - option button triggers the display to
start at the trigger level, but on the waveform’s
falling slope.
Channel text box - Allows you to specify or select a
channel for the trigger source.
Level field - Indicates the current trigger level in
volts.
When Oscilloscope Mode… is chosen from the
Options menu, the Oscilloscope Triggering dialog
box is displayed as follows:
This dialog box allows you to select one trigger
channel and display the data in a triggered sweep
oscilloscope format or in a free-running oscilloscope
format. Checking the Free Run check box selects
the free running oscilloscope format. Leaving the
Free Run check box unchecked selects the triggered
sweep oscilloscope format.
If free running is selected, the waveform
continuously updates on your monitor in a manner
identical to an oscilloscope display and the rest of
this dialog box can be ignored (no triggering
conditions need be set).
Free Run check box - Allows you to select between
a triggered sweep oscilloscope display or a freerunning oscilloscope display (see discussion above).
Note that this box must be unchecked to set the
trigger level.
Erase Bar check box - Allows you to enable/disable
the display of the vertical erase bar (cursor) on the
waveform. Whether the erase bar is displayed or
disabled, the waveform information on screen is still
“erased” and replaced with a fresh sweep of data.
OK command button - Closes the Oscilloscope
Triggering dialog box and applies the just
configured display trigger conditions.
Set Level command button - Allows you to adjust
or set the trigger level. Note that the Free Run check
box must be unchecked to set a trigger level.
When this button is selected, the Oscilloscope
Triggering dialog box closes and causes the cursor
to point to a short horizontal line at the right side of
the annotation margin. This short horizontal line is
the trigger level indicator.
Trigger level indicator
If triggered sweep is selected, you can also set
display trigger conditions (such as channel, level,
and slope) that define the “trigger” or starting point
for displaying the waveform on your monitor. In this
mode, the waveform is not displayed until a trigger
condition is detected. When a trigger is detected, a
single sweep of the waveform will be displayed on
your monitor and will remain on your monitor until
the next trigger condition is detected.
At this point, mouse movement is limited to
dragging the line up or down to the desired trigger
level. Note that when the mouse is moved, the data
17
display in the left annotation margin changes to
/Lite
Open Reference File
/Lite
User Annotation
Current data enabled (shown in oscilloscope mode)
indicate trigger level. When set where desired,
clicking the left mouse button sets a new trigger
level while clicking the right mouse button retains
the existing level. Note that if the trigger level is set
somewhere above or below the current signal, the
signal will disappear from the display.
Cancel command button - Closes the Oscilloscope
Triggering dialog box and cancels any changes
made to trigger conditions (provided the changes
were not saved with the OK command button).
AQ
WIND
Allows you to use an existing data file as a “set up”
for a new data file. All recording parameters (such
as number of channels, scaling constants,
engineering units, display format, compression,
display mode, gain, etc.) are ported from the
existing file to become the default conditions for the
new data file.
Current Data - Enables/disables the display of the
current waveform data value in the left or right
annotation margin. The Data Display option must
first be enabled for this option to work.
When enabled in the oscilloscope mode, the current
waveform data value at the cursor position is
displayed in the annotation margin (the cursor does
not have to be enabled).
When enabled in the scroll mode, the current
waveform data value at the real time point (far right
screen edge) is displayed in the annotation margin.
When disabled (in either mode), the base line value
is displayed in the left/right annotation margin (if
Data Display is enabled).
To enable current waveform data (if disabled, or
disable it if enabled):
Point to the Options menu and drag down to
Current Data.
This function allows instantaneous setup, thus
making it easy to run similar but separate waveform
recording sessions.
To use an existing file’s recording parameters for a
new file:
Point to the File menu and drag down to
Open Reference File…
You will be prompted for the name of the existing
data file. This file must be one created by W
INDAQ
software and must already exist. When a filename is
entered and the dialog box is closed, the new file
immediately inherits the existing file’s recording
parameters.
This function is enabled only when in the SET-UP
operating mode (Status: field in bottom annotation
line indicates SET-UP).
AQ
WIND
Current Data, Data
Display, Limit Display &
Allows you to enable or disable on-screen
annotation options. Each annotation option is
described as follows:
When enabled, the annotation margin will have the
following appearance and a check mark (
√) will be
displayed on the Options menu immediately
preceding the Current Data command:
Data cursor
(vertical line)
Current waveform
value at cursor position
Data Display - Displays or hides the current
waveform data value (or the baseline value,
depending on the state of the Current Data option)
in the annotation margin. If Limit Display is not
enabled, the engineering units will toggle on and off
along with the value. To display data (if hidden, or
hide it if displayed):
Point to the Options menu and drag down to
Data Display.
When the data display option is enabled, the
annotation margin will have the following
appearance and a check mark (
√) will be displayed
on the Options menu immediately preceding the
Data Display command:
18
Data display enabled
/Lite
compression
Sample rate
Channel setup
Enable channels
Start/stop recording
Open/close file
Open reference file
mode
mode
Scroll mode
Shift waveform down
Shift waveform up
Shrink waveform 2x
Expand waveform 2x
Change palette
Show/hide grid
Limit Display - Displays or hides the upper and
lower chart-edge limits in the left/right annotation
margin. If Data Display is disabled, engineering
units toggle on and off along with the limits. To
display upper and lower chart-edge limits (if hidden,
or hide them if displayed):
AQ
WIND
ToolBox
When enabled, a floating “toolbox” of buttons is
displayed on your screen. The toolbox buttons
provide quick access to commonly used commands.
To display the toolbox:
Point to the View menu and drag down to
ToolBox.
Point to the Options menu and drag down to
Limit Display.
When the limit display option is enabled, the
annotation margin will have the following
appearance and a check mark (
√) will be displayed
on the Options menu immediately preceding the
Limit Display command:
Limit display enabled
User Annotation - Toggles the display of user
annotation (entered from the Edit menu, User
Annotation… command) in the waveform strip. To
display user annotation (if hidden, or hide it if
displayed):
Point to the Options menu and drag down to
User Annotation.
When user annotation is enabled, the waveform strip
will display the specified channel’s annotation
(entered with Edit User Annotation…) and a check
√) will be displayed on the Options menu
mark (
immediately preceding the User Annotation
command:
Each button on the ToolBox provides quick access
(a shortcut) to commonly used functions as follows:
Calibration
Waveform
Commented
event
marker
Trigger
Oscilloscope
User annotation enabled
Screen annotation options can be enabled in
virtually any combination to provide the desired
result. However, in some instances one option will
override another. For example, if Data Display is
disabled, it is not possible to enable Current Data.
Similarly, if Limit Display is disabled, it is not
possible to enable Current Data.
Open Reference File — The same as choosing
Open Reference File… from the File menu.
Open/Close File — The same as choosing Open…
from the File menu or, when a file is already open,
closes the file.
19
Start/Stop Recording — When activated with a
/Lite
User Annotation…
/Lite
Compression /2
data file already open, recording starts immediately.
When activated without a data file open, the Open
dialog box appears prompting you to specify a new
file name or select an existing file. When activated
during recording, recording is stopped.
Enable Channels — The same as choosing
Channels… from the Edit menu.
Channel Setup — The same as choosing Channel
Settings… from the Edit menu.
Calibration — The same as choosing Fixed
Calibration… from the Edit menu.
Sample Rate — The same as choosing Sample
Rate… from the Edit menu.
Waveform Compression — The same as choosing
Compression… from the Edit menu.
Commented Event Marker — The same as
choosing Insert Commented Mark from the Edit
menu.
AQ
WIND
Allows you to display a custom message (up to 30
characters) on the baseline of each display channel
(used in tandem with the User Annotation function
on the Options menu).
The desired channel annotation must first be entered
before it can be toggled on or off. To enter channel
annotation:
Point to the Edit menu and drag down to
User Annotation…
This displays the Channel Annotation dialog box as
follows:
Show/Hide Grid — The same as choosing Grids
from the Options menu.
Change Palette — The same as choosing User
Palette… from the File menu.
Expand Waveform 2X — The same as choosing
Grow 2X from the Scaling menu.
Shrink Waveform 2X — The same as choosing
Shrink 2X from the Scaling menu.
Shift Waveform Up — The same as choosing
Waveform Up 10 Pixels from the Scaling menu.
Shift Waveform Down — The same as choosing
Waveform Down 10 Pixels from the Scaling menu.
Scroll Mode — The same as choosing Scroll Mode
from the Options menu.
Oscilloscope Mode — The same as choosing
Oscilloscope Mode… from the Options menu.
Triggered Mode — The same as choosing
Triggered Mode… from the Options menu.
This dialog box prompts you to enter the desired
channel annotation for the currently selected
channel, or any other channel. If not the currently
selected channel, enter the desired channel number
for which you want to provide annotation in the
Channel # = text box. In the Annotation = text
box, enter the desired channel annotation (up to 30
characters).
Once the channel annotation is entered, the display
of user annotation can be toggled on or off with the
User Annotation command on the Options menu.
The Next and Previous buttons allow you to step
through all of the enabled channels in order, entering
the desired channel annotation for each channel,
without closing this dialog box.
AQ
WIND
Compression…,
Compression x2,
Allows you to vary the number of data points used
to display the real time waveform on your computer
screen.
20
Point to the Edit menu and drag down to
/Lite
Time Base
Time per major division
/Lite
Beep on File Full
Compression x2.
To divide the current compression factor by 2:
This function is simply a “graphics adjuster” that
only affects how the display looks on your screen.
Data acquisition is independent of this function and
is always recorded (gap-free) at a compression factor
of 1 (no compression applied).
When compression is applied to the waveform,
multiple A-D readings (equal to the compression
factor) are evaluated for plotting. In this instance,
less CPU time is required for the display. For
example, if compression equals 100, then 100
consecutive A-D readings are evaluated to define a
waveform point.
When compression is selected, a true min/max is
performed on the waveform to ensure that transient
information is always displayed.
The degree of waveform compression is controlled
as follows:
Point to the Edit menu and drag down to
Compression…
Point to the Edit menu and drag down to
Compression /2.
Perhaps these two functions are most useful when
the accelerator keys (PGUP for Compression x2 and
PGDN for Compression /2) are used. This allows
you to press one key several times in succession to
find just the right amount of compression for a
smooth scrolling, visually pleasing display.
As the waveform compression factor is changed, the
time base field (ms /DIV: or sec /DIV:) displayed in
the bottom annotation line is automatically adjusted
to reflect the proper time scaling of displayed
waveforms.
AQ
WIND
Displays the time per major division at the bottom of
the real time screen as sec / DIV: or ms / DIV:. The
displayed value is a function of the selected sample
rate, the number of channels enabled, and the
selected waveform compression factor.
This displays the Compression dialog box as
follows:
To specify a degree of waveform compression:
Enter the desired compression factor (from 1
to 65,536) and click OK.
A fast and convenient way to experiment with many
different compression factors is with the
Compression x2 and Compression /2 commands.
These commands respectively multiply or divide the
current compression factor by 2.
To multiply the current compression factor by 2:
Time per major division is displayed at all times in
all operating modes.
AQ
WIND
When enabled, this option sounds an audible beep
when the target file is full.
21
To enable the beep on file full function:
/Lite
/Lite
Save Default Setup
/Lite
Commented Mark
Point to the Options menu and drag down to
Beep on File Full.
When enough data has been recorded to fill the
allocated file space, an audible beep sounds (when
this function is enabled) to warn you that acquisition
has stopped and W
INDAQ/Lite software is in the
STBY operating mode.
AQ
WIND
Close
Allows you to close the data file you are currently
working with. Closing a data file removes the file
from your screen, but does not quit W
software.
A new data file must first be opened and data must
be recorded to it before the Close command is
selectable (undimmed).
However, the Close command is immediately
selectable on an existing data file that you are
appending to.
Once Close becomes selectable, you can close a
currently open data file as follows:
Point to the File menu and drag down to
Close.
AQ
WIND
Allows all current WINDAQ/Lite recording
parameters to become default settings for all
subsequent recording sessions.
INDAQ/Lite
AQ
WIND
Insert Mark, Insert
This function allows you to insert event markers
(either commented or uncommented) at any point in
the data stream going to disk. Event markers (along
with appropriate comments) can be used to mark an
event, or flag an uncommon occurrence in the data
stream going to disk, thus providing a complete
record of what happened and why.
After the waveform recording session, W
Waveform Browser playback software allows you to
automatically search and move to the point in the
data file where the event markers were placed
during data acquisition.
A transition from the SET-UP or STBY operating
mode to the RECORD operating mode automatically
inserts a positive going time and date stamped event
marker in the data stream going to disk, thus
marking the start of acquisition. You must be in the
RECORD mode to insert additional event markers
into the data going to disk.
While in the RECORD operating mode,
point to the Edit menu and drag down to
Insert Mark for an uncommented event
marker or Insert Commented Mark… for a
commented event marker.
A commented event marker allows you to enter up
to a 39-character comment that will be associated
with the event marker. When Insert Commented
Mark… is chosen from the Edit menu, the Event
Marker Comment dialog box is displayed as follows:
INDAQ
When you have set up W
INDAQ/Lite exactly the way
you like, you can make this configuration the
default, so it always appears this way when started.
To make the current screen design the default
condition:
Point to the File menu and drag down to
Save Default Setup.
Virtually every W
INDAQ/Lite recording parameter is
saved to determine the default condition: display
format, sample rate, compression, channels enabled,
etc.
Enter the desired comment (up to 39 characters) in
the text box. The comment will be displayed with its
associated event marker when Event Markers… is
selected from the View menu in the W
INDAQ
Waveform Browser playback and analysis software
package. A commented event marker is anchored at
Loading...
+ 60 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.