Shimadzu LC-10/20 Pump 3 Installation Procedure
3 Installation Procedure
3.1 Shimadzu LC-10/20 Pump communication
The Shimadzu LC- 10/20 Pump is controlled by serial (RS232)
communication. The communication is provided via the Optical to USB
convertor connected to the computer via USB A-A cable (p/n SK10) and to
the pump via optical cable (p/n SK15). Both of these cables are part of the
DataApex Opt-USB converter set (p/n OPT01), but can also be ordered
separately.
Optical to USB converter
For communication an OPT-RS232 converter is needed. It can either be
the DataApex Opt-USB converter (p/n OPT01) or Shimadzu adaptor
(Shimadzu p/n 22-35480-91).
Pump Settings
Communication parameters should be set on the pump in the SERVICE
MODE (see pump documentation) as follows:
Baud rate 2400, parity N, bits 8, stop bit 1.
SYS (Set System Parameter) has to be set to 5 for isocratic mode or to 6
for gradient mode from the pump keyboard.
Caution: These values are not documented in the pump manual.
The other Pump system parameters should be set to:
Local=0 (Remote)
ADRS= any value (not used)
CBM LINK=0 (not used)
Shimadzu LC-20 settings
Shimadzu LC-20 pumps have to be set to the LC-10 emulation mode in
order to be able to work in Clarity. This is done from the pump's keyboard
according to this procedure:
l Press the VP button on the pump's front keyboard until the CALIBRATION
inscription is shown on the pump's display.
l Press the func button to enter the pump's configuration. The password
has to be filled in.
Note: The default password is 00000.
l Press the func button repeatedly until the OP MODE inscription is shown
on the pump's display.
l Enter the desired number (1 ) to switch the pump to the SCL-10Avp
emulation mode and press the enter button.
- 3 -