DataApex Alltech 3300 ELSD User Manual

Clarity Controls
Alltech 3300 ELSD
DET ENG
Code/Rev.: M151/70A
Date: 1/10/2017
Phone: +420 251 013 400 DataApex Ltd.
Fax: +420 251 013 401 Petrzilkova 2583/13
www.dataapex.com The Czech Republic
Clarity®, DataApex®and
®
are trademarks of DataApex Ltd. Microsoft®and Windows
TM
are
trademarks of Microsoft Corporation.
DataApex reserves the right to make changes to manuals without prior notice. Updated manuals can be downloaded from www.dataapex.com.
Author: MP
Alltech 3300 ELSD Table of Contents
Contents
1 Alltech 3300 ELSD Control Module 1 2 Requirements 2 3 Installation Procedure 3
3.1 Hardware - Wiring 3
3.2 Clarity Configuration 4
4 Using the control module 6
4.1 Alltech 3300 ELSD Setup 7
4.2 Digital Outputs of Alltech 3300 ELSD 8
4.3 Method Setup - Acquisition 9
4.3.1 Method Setup - Acquisition - Settings 10
4.3.2 Method Setup - Acquisition - Time Table 12
4.4 Method Setup - Advanced 13
4.5 Device Monitor 14
5 Report Setup 15 6 Troubleshooting 16
- i -
Table of Contents Clarity Controls
To facilitate the orientation in the Alltech 3300 ELSD manual and Clarity chromatography station, different fonts are used throughout the manual. Meanings of these fonts are:
Instrument (blue text) marks the name of the window to which the text refers.
Open File(italics) describes the commands and names of fields in Clarity, parameters that can be entered into them or a window or dialog name (when you already are in the topic describing the window).
WORK1 (capitals) indicates the name of the file and/or directory. ACTIVE (capital italics) marks the state of the station or its part.
The bold text is sometimes also used for important parts of the text and the name of the Clarity station. Moreover, some sections are written in format other than normal text. These sections are formatted as follows:
Note: Notifies the reader of relevant information.
Caution: Warns the user of possibly dangerous or very important
information.
▌ Marks the problem statement or trouble question.
Description: Presents more detailed information on the problem, describes its causes,
etc.
Solution: Marks the response to the question, presents a procedure how to remove it.
- ii -
Alltech 3300 ELSD 1 Alltech 3300 ELSD Control Module
1 Alltech 3300 ELSD Control Module
This manual describes the setting of the Alltech 3300 ELSD detector. The control module enables direct control of the instrument over Ethernet (LAN) connection.
Fig 1: Alltech 3300 ELSD detector
Direct control means that the detector can be completely controlled from the Clarity environment, including the digital data acquisition. That way, no A/D converter is needed. Instrument method controlling the analysis conditions will be saved in the measured chromatograms.
- 1 -
Alltech 3300 ELSD 2 Requirements
2 Requirements
l Clarity Installation CD ROM with LC Control module (p/n A24).
Caution: Minimal firmware version required for 3300 ELSD is 1.0.32.
l Crossed LAN cable (p/n SK08). l Both PC and detector connected to the same Ethernet network.
Note: Cables are not part of Clarity Control Module. It is strongly
recommended to order required cables together with the Control Module.
- 2 -
Alltech 3300 ELSD 3 Installation Procedure
3 Installation Procedure
3.1 Hardware - Wiring
The Alltech 3300 ELSD detector is controlled by LAN (Ethernet) communication. It uses standard cross LAN cable for direct connection with PC and uncrossed LAN cable when connected to a network hub/switch (however the most of modern switches are able to automatically adapt to both types of cable). For connecting the detector with other devices using synchronization inputs and outputs there is 14-Pin connector available on the backside of the detector. Pins 1-2 are Start input, used for starting the detector by external device, pins 9- 10 accommodate External Output, suitable for controlling some other device. For detailed description of this connector, please check the original manual supplied by manufacturer of the detector.
- 3 -
3 Installation Procedure Clarity Controls
3.2 Clarity Configuration
Fig 2: How to add Alltech 3300
l
Start the Clarity station by clicking on the icon on the desktop.
l Invoke the System Configuration dialog accessible from the Clarity
window using the System - Configuration... command.
l Press the Add button (see 3.2 on pg 4.) to invoke the Available Control
Modules dialog.
l You can specify the searching filter to simplify the finding of the driver.
l Select the Alltech 3300 ELSD and press the Add button.
- 4 -
Alltech 3300 ELSD 3 Installation Procedure
The Alltech 3300 ELSD Setup dialog will appear.
Fig 3: Alltech 3300 Setup
l Select the correct IP Address and press the AutoDetect button. If the
detector is connected correctly, detector name and serial number will be displayed.
l Fill in the signal name.
l Press the OK button.
The Alltech 3300 ELSD will appear in the Setup Control Modules list of the System Configuration dialog.
l Drag and drop the Alltech 3300 ELSD icon from the Setup Control
Modules list on the left side of the System Configuration dialog to the
desired Instrument tab on the right side (or use the button to do so).
Note: The configuration dialog of the Alltech 3300 ELSD detector can be
displayed any time by double-clicking on its icon or using the Setup button.
- 5 -
Alltech 3300 ELSD 4 Using the control module
4 Using the control module
New Acquisition tab is created in the Method Setup dialog. If there are any other detectors configured on the instrument, then the Alltech 3300 ELSD detector can be accessed by switching to the desired signal in the Select Detector section on the top of the dialog.
- 6 -
4 Using the control module Clarity Controls
4.1 Alltech 3300 ELSD Setup
The Alltech 3300 Setup dialog serves for the correct setting of the communication between Clarity and the detector.
Fig 4: Alltech 3300 Setup
IP Address
IP address of the detector in your local network.
AutoDetect
This button is used to test the connection to the Alltech 3300 ELSD detector using the previously set IP Address . Obtained data about the detector are then displayed in the dialog.
Signal 1 - Name
Enables to set the relevant detector name.
Digital Output Names
Displays the Digital Output Names dialog which enables to set the custom labels of Alltech 3300 ELSD digital outputs. The custom names are displayed in the Device Monitor window described in the chapter "Device Monitor" on pg 14.
Fig 5: DigitalOutput Names
Keyboard Lock
When checked, locks the keyboard of the detector and prevents other users on the same network to connect the machine.
- 7 -
Alltech 3300 ELSD 4 Using the control module
4.2 Digital Outputs of Alltech 3300 ELSD
Fig 6: DigitalOutputs of Alltech 3300 ELSD
Serves for setting the initial and actual state of the Alltech 3300 ELSD detector's auxiliary output.
- 8 -
4 Using the control module Clarity Controls
4.3 Method Setup - Acquisition
This tab serves for the setting of the template method that is used by the Alltech 3300 ELSD detector for the acquisition. All parameters set on this tab are automatically sent to the Alltech 3300 ELSD detector prior to each injection in the ACTIVE sequence and may be sent there manually by using the Send Method button in the lower part of this tab or in the Single
Analysis dialog.
Some analysis parameters cannot be modified neither from Clarity nor directly on the Alltech 3300 ELSD detector. These parameters are not changed when sending the method to the detector. Setting of the acquisition parameters is spited into the Settings and Time
Table tabs.
- 9 -
Alltech 3300 ELSD 4 Using the control module
4.3.1 Method Setup - Acquisition - Settings
Settings tab for adjusting the parameters of the detector..
Fig 7: Method Setup - Acquisition
Heater Lower Drift Tube
Turn on the heater of the lower part of drift tube.
Heater Upper Drift Tube
Turn on the heater of the upper part of drift tube.
Heater Optics Block
Turn on the heater of the optics block.
Gas Flow
Turn on the flow of carrier gas.
Laser
Turns on the laser.
Gain
Sets the gain of the detector.
Shutdown
Sets the behavior of the detector after the analysis is completed.
Keep State
Detector keeps the ON state.
Set Off after analysis
Set the detector OFF after an analysis is performed.
Temperature
Sets the temperature of the drift tube.
- 10 -
4 Using the control module Clarity Controls
Gas Flow Rate
Sets the flow rate of the carrier gas.
Equilibry Time
Set the time required for equilibration of gas flow and temperatures. The detector will become ready if the actual temperature stays within ± 1.5 °C of the temperature set in the Temperature field and the flow stays within
0.3 ml/min of the value set in the Gas Flow Rate field. When any of the two parameters exceeds the limits, the timer set in the Equilibry Time field resets.
Lower section of the Method Setup - Acquisition dialog
In the right section on all Method Setup - Acquisition sub-tabs for Alltech 3300 ELSD detector new button emerges:
Det Status…
After pressing the Det Status… button the Hardware Configuration dialog opens. It displays the detector type, LAN address through which it is connected and the port used.
Fig 8: Hardware Configuration
- 11 -
Alltech 3300 ELSD 4 Using the control module
4.3.2 Method Setup - Acquisition - Time Table
This sub- tab serves for time program to change operating parameters during run. Available Event Types are : Autozero, setting the zero for detector output, and Gain, setting the gain of the detector.
Fig 9: Method Setup - Time Table
- 12 -
4 Using the control module Clarity Controls
4.4 Method Setup - Advanced
Sets other parameters of the method, most notably the Subtraction and Auxiliary Signals. When invoked by the menu command, the Method
Setup - Advanced dialog will open. The tab is common to all signals while
performing multidetector measurement.
Fig 10: Method Setup - Acquisition - Advanced
Subtraction
Set the chromatogram for subtraction from the measured signal.
Column Calculations
Set the parameters of the column.
Auxiliary Signal table
Set which auxiliary signals will be stored in the chromatogram file.
- 13 -
Alltech 3300 ELSD 4 Using the control module
4.5 Device Monitor
The window with the detector status can be invoked by the Monitor ­Device Monitor command from the Instrument window or using the LC Monitor icon. It displays the actual detector temperatures and their settings
in the Method Setup - Acquisition window, as well as other parameters set there.
Fig 11: Device Monitor
The detector can be monitored and partially controlled from the Device
Monitor window. There is a control command allowing the AutoZero
function, all other fields are informative and described in the Method
Setup - Acquisition dialog.
Auto Zero
This button performs the autozero operation.
Note: Values which exceeds expected limits are displayed in red color.
- 14 -
Alltech 3300 ELSD 5 Report Setup
5 Report Setup
All of the detector settings accessible on the Method Setup - Acquisition tab for the given signal are reported. To do so, the Instrument Control parameter on the Method tab of the Report Setup dialog must be checked.
Fig 12: Print Preview of the Report Setup
- 15 -
Alltech 3300 ELSD 6 Troubleshooting
6 Troubleshooting
When the remedy for some problem cannot be discovered easily, the recording of communication between Clarity and the detector can significantly help the DataApex support to discover the cause of the problem. The recording can be enabled by adding or amending the COMMDRV.INI file in the Clarity installation directory (C:\CLARITY\CFG by default). The file can be edited in any text editor (e.g. Notepad). Following section should be edited or added:
[TCP_IP 192.168.0.100:79] echo=on textmode=on filename=Alltech3300ELSD_%D.txt reset=off
Note: Instead of 192.168.0.100:79 type correct IP address and port used for
communication with the Alltech 3300 ELSD detector. This port number is displayed when the Det Status button in the Method Setup - Acquisition dialog is invoked.
Note: %D (or %d) in the filename parameter means that the log will be created
separately for each day. The reset=off parameter disables deleting the content of the log each time the station is started during the same day.
The created *.TXT files will greatly help in diagnosis of unrecognized errors and problems in communication.
- 16 -
Loading...