Rosemount™ Gas Ultrasonic 3410 Series Electronics Retrofit Instructions
Retrofit Instructions June 2022
Safety and approval information
www.emerson.com or through your local Emerson support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at http://www.emerson.com.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the user manual. Product data sheets and manuals are available from the Emerson website at
http://www.emerson.com.
Return policy
Follow Emerson procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Emerson employees. Emerson will not accept your returned equipment if you fail to follow Emerson procedures. Return procedures and forms are available on our web support site at Emerson.com, or by phoning the Emerson Customer Service department.
Emerson Flow customer service
Email:
Worldwide: http://flow.support@emerson.com
Asia-Pacific: http://APflow.support@emerson.com
Telephone:
North and South America
United States 800 522 6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303 527 5200 The
Mexico +41 (0) 41 7686
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
111
Europe and Middle East Asia Pacific
Netherlands
France 0800 917 901 India 800 440 1468
Central & Eastern
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
+31 (0) 704 136 666
+41 (0) 41 7686 111
New Zealand 099 128 804
Japan +81 3 5769 6803
2 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
North and South America Europe and Middle East Asia Pacific
UAE 800 0444 0684
Contents
Plan..............................................................................................................................................5
Install......................................................................................................................................... 16
Operate......................................................................................................................................31
Mark III RS-232 and RS-485 Serial communications setup...........................................................44
Summary table of communications/output settings.................................................................. 52
3410 Communication and Output settings................................................................................ 56
3410 Series engineering drawings..............................................................................................72
Retrofit Instructions 3
Retrofit Instructions June 2022
4 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions

1 Plan

1.1 Introduction

1.1.1 General

Welcome to the Rosemount 3410 Series Ultrasonic Retrofit Instruction Manual.
This manual has been designed to provide you with a step-by-step set of instructions for retrofitting a meter with Mark III electronics to the new 3410 Series electronics.
NOTICE
Please read the “Before You Begin” section to make sure you have all the components necessary to perform the before taking the meter out of service.
Important
If retrofitting from Mark II electronics to 3410 Series electronics complete Sections 3 and 4 in the Rosemount Ultrasonic Mark III Upgrade Kit Instructions to determine the communications and output settings of the Mark II electronics. Use those settings and the instructions in this manual to configure the Series electronics.
Retrofit Instructions 5
Retrofit Instructions June 2022

1.2 Before you begin

This section shows all of the parts and tools required to perform the upgrade. Ensure that all of these items are available before taking the meter out of service.

1.2.1 Tools required

Figure 1-1: Required Tools
A. -in. flat-blade screw driver B. ¼-in. flat-blade screw driver C. 3/8-in. drive ratchet wrench
D. 3/8-in. drive - extension - at least 3.5-in. long
E. 3/8-in. drive - 7/16-in. socket F. Allen wrench - size 6 mm (included in Retrofit kit)
G. Crescent wrench for cable glands and electrical conduit
MeterLink™ is required to communicate with the meter electronics, access information and collect the configuration in the meter and then convert and download the configuration into the new 3410 Series electronics. Always use the latest version of MeterLink™. Upgrades to the latest version are available at: Emerson.com/Rosemount
6 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions

1.3 Before removing the Electronics

1.3.1 Procedure before removing the Mark III Electronics
Procedure
1. Before removing power from the meter being retrofitted software to connect to the meter.
2. Use the Edit/Compare Configuration screen in MeterLink to read the configuration from the meter and save it to a file on your
computer. An example for a file name might be “Meter Name, MkIII Final Config, 3-4-2013 10-15-02 AM.cfg”. MeterLink will be used later in this procedure (see Configure the 3410 Series Electronics) to convert the configuration and download it to the 3410 Series electronics.
CAUTION
Loss of set-up data
The current meter configuration file must be saved before beginning the 3410 Series electronics retrofit.
Failure to save the current configuration file can result in the loss of important set-up data.
3 through 8 provide more detailed instructions on obtaining the required information with respect to the communications set-up of the electronics, however, if these settings are already known, these steps can be skipped. Step 8 is for Frequency and Diagnostic outputs. For reference, the known values should be entered in the Summary
table of communications/output settings in Appendix B. This
information will be required in the setup of the Series electronics. For any setting which is not known, the applicable step(s) for obtaining that information can be used.
3. Use the File|Meter Directory menu path, then select the <meter name> you want to retrofit.
a) Click the Direct button on the lower left corner of the Meter
Directory dialog box.
b) The Direct Connection Properties For dialog appears.
c) Obtain the Modbus ID (same as Comms Address).
Retrofit Instructions 7
Retrofit Instructions June 2022
Figure 1-2: Accessing Modbus ID for Ports A and B
4. Based on the information contained in the connection properties window, enter the values for the following parameters in the
Summary table of communications/output settings in Appendix B:
Port A and B: Modbus ID = _____(enter Comms Address).
5. Exit MeterLink and disconnect your PC from the meter electronics. The remaining information required before doing the retrofit upgrade, is based on the wiring terminations and the settings of various switches on the boards that comprise the electronics.
6. Power down the electronics.
CAUTION
Equipment damage
Remove power from the meter electronics.
Failure to remove power from the meter electronics may result in damage to the equipment.
7. Remove the end caps from the upper electronics housing of the meter to access the electronic circuit boards. See Figure 1-4 for board layout of the 3400 stacks. See Figure 3-4 for the 3410 Series electronics for board layout.
8 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Figure 1-3: Remove End Caps from the Upper Electronics Enclosure
Figure 1-4: 3400 Electronics Assembly board layout
Retrofit Instructions 9
Retrofit Instructions June 2022
Figure 1-5: 3410 Retrofit Electronics Assembly
A. End cap and security screw B. O-ring C. Desiccant packet D. CPU module
E. Optional RS-232 I/O module
F. Optional RS-485 I/O module G. Blanking cover H. Guide plate
I. M20 reducer J. Ground lug - external
K. Transmitter electronics enclosure
L. Power supply
M. I.S. barrier board
N. Standoffs
O. Backplane board
P. O-ring
Q. End cap and security screw
R. O-ring
S. Acquisition cable and roll pin
T. Acquisition module assembly
U. Base enclosure
V. Cable glands
W. Gasket, base enclosure
X. Acquisition cable clamp
10 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Y. Acquisition module
8. Remove the Mark III electronics CPU board, the HART Option Board, and the Field Termination Board from the enclosure. Fill out the
Summary table of communications/output settings in Appendix B
based on the Field Connection Board wire terminations and/or switch settings of the CPU board and Option Board, if installed. This information will be used to configure the new 3410 Series electronics later in this procedure.
Figure 1-6: Mark III CPU Board Showing Switch Settings
Retrofit Instructions 11
Retrofit Instructions June 2022
Figure 1-7: Mark III CPU Board with I.S. Barrier Board Piggy­Backed on CPU Board

1.4 Removing Mark III Electronics

1.4.1 Procedure to remove the Mark III Electronics
Procedure
1. Having obtained all the information related to the set-up of the 3400 Series electronics configuration and communications, all the components related to the electronics can now be physically removed from the meter. Unscrew all wiring from the Field Termination Board. It may be helpful to label the wires to make it easier to connect the 3410 Series electronics.
2. Remove any conduit and wiring attached to all six conduit ports.
3. Remove any ground wire attached to the ground lug on the outside of the upper electronics housing. At this point there should be no wires connected to the upper electronics housing.
12 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Figure 1-8: Ground Wire Removed
4. Use the 6mm Allen wrench (supplied with retrofit kit) to remove the two bolts holding the base enclosure.
Figure 1-9: Remove Bolts from Base Enclosure Cover
5. Remove the upper electronics housing from the base enclosure.
a) Remove the Acquisition cable, which comes down from the
upper electronics to the Acquisition module in the base enclosure.
Retrofit Instructions 13
Retrofit Instructions June 2022
NOTICE
Figure 1-10 shows the Acquisition module for the 3400 Series
electronics.
b) Loosen the two screws on the ends of the Acquisition cable
connector, and disconnect it from the Acquisition module.
c) Set the upper electronics housing aside once the housing is
removed.
Figure 1-10: Remove the Acquisition Cable
6. Remove all the screws that hold the transducer wires in the terminal blocks that attach the transducers to the Acquisition module.
a) Remove the screws from the ends of the two connectors.
b) Disconnect the connectors and remove each of the
transducer wires from the connector.
7. Loosen the cable glands on the base enclosure and pull all of the transducer cables out of the base enclosure.
8. Unscrew the three screws holding the Acquisition module in the base enclosure and remove the Acquisition module.
The four mounting bolts holding the enclosure cover to the meter body are sealed with an RTV silicone adhesive/sealant.
14 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
9. Remove the RTV with a flat blade screwdriver or utility knife around the perimeter of the recessed opening and the RTV eases the removal of the sealant.
a) Use the 7/16-in. socket wrench to remove the four mounting
bolts.
Figure 1-11: Remove Old RTV
10. Remove the base enclosure from the meter body.
11. If RTV silicone adhesive/sealant was used to seal the base enclosure to the meter body, remove the adhesive/sealant so the 3410 base enclosure will have a clean mounting surface. The 3400 Series electronics should be completely removed from the meter body at this time.
Retrofit Instructions 15
Retrofit Instructions June 2022

2 Install

2.1 Installing the 3410 Series Electronics

2.1.1 Procedure to install the 3410 Series Electronics
Procedure
1. Remove the 3410 components from the packaging. a) Use the 6 mm Allen wrench to remove the two screws from
the base enclosure cover.
Figure 2-1: Remove Bolts from Base Enclosure Cover
2. Lift the upper electronics housing from the base enclosure. The
acquisition cable connector from the upper electronics is attached to the acquisition module with two screws. The cable is also secured by a cable clamp to a mounting bolt.
a) Remove the screw securing the cable clamp and then loosen
the two screws on the acquisition cable connector and disconnect the acquisition cable connector from the acquisition module.
16 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Figure 2-2: Remove Acquisition Cable and Module
3. Unscrew the remaining two mounting screws from the acquisition module in the base enclosure and remove the acquisition board.
4. Place an insulating gasket on the meter body where the electronics will be mounted (see gasket in Figure 1-5).
5. Place the base enclosure on top of the meter body on top of the insulating gasket. Note that the transducer port openings on the base enclosure are labeled with transducer locations (i.e., A1, A2, B1, B2, C1, C2, D1, D2).
a) Orient the base so that the A1 transducer port is closest to the
A1 transducer location on the meter body. The meter body has the transducer location labels embossed in the body next to each transducer boss. The ports on the base enclosure must be oriented outward (towards the sides of the meter body, perpendicular to meter axis).
Figure 2-3: Base Enclosure Orientation
6. Use the four ¼-in. bolts with stainless steel washers and nylon shoulder washers to secure the base enclosure to the meter body.
a) Use a nut driver to tighten the bolts to a torque of 55 to 65
inch-lbs.
Retrofit Instructions 17
Retrofit Instructions June 2022
Figure 2-4: Insert ¼-in. Bolts
7. Fill up the base enclosure wells with RTV adhesive/sealant around and over the four ¼-in. bolts.
This will ensure a water-tight seal. Any excess sealant can be removed by using a scraper to leave the sealant flush with the top of the well. Excess sealant will adhere to the acquisition board when it is installed, and make future removal of the board difficult.
Figure 2-5: Apply RTV Sealant Around Bolts
8. Insert the Acquisition board in the base enclosure and secure it with the two screws as shown for now.
18 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Figure 2-6: Acquisition Module Attachment Screws
Figure 2-7: Acquisition Module Wiring
9. Route the transducer cables through the cable glands with the correct label number (i.e. A1) which matches the label on the transducer cable.
Figure 2-8: Base Enclosure Cable Glands Installation
10. Judge the proper length and cut off the excess.
Retrofit Instructions 19
Retrofit Instructions June 2022
11. Strip the outer insulation, outer shield, and inner insulation using a utility knife. Verify that insulation of individual wires was not cut while removing outer layers. Strip each wire 0.28-in.
Figure 2-9: Transducer Cable Cut Instructions
12. Insert tubing over wires, inner insulation and under shield approximately 1.5-in. of shield should overlap.
13. Insert and secure transducer cables through cable glands verify Line Mark is as referenced in Figure 2-8.
Figure 2-10: Line Mark
14. Using a torque screwdriver set to 34 ±2 oz.-in., secure the wires to the terminal plugs of the Acquisition Module.
15. Verify the cable lengths are still an appropriate length and if necessary, cut the transducer cables to the correct length with allowance for the wiring terminations and connector placement.
16. Ensure that the contacts clamp on the bare wires and not on the wire insulation when the wires are being terminated on the connector.
a) Leave the connector plugged into the Acquisition board while
terminating the individual wires.
b) Tighten the connector screws on the either end of the
connector after terminating all the wires on the applicable connector.
c) Repeat this procedure for the transducer cables on the other
side of the meter.
20 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Figure 2-11: Install Cable and Secure Wires
17. Use the cable ties to dress the transducer cables. They should be dressed in groups of two: A1 and C1, D1 and B1, A2 and C2, D2 and B2. Install one cable tie at three inches from the base enclosure and another about the point where the cables start to bend and separate out to their respective port.
Retrofit Instructions 21
Retrofit Instructions June 2022
Figure 2-12: Re-oriented and Secured Transducer Cables
18. Reconnect the acquisition cable connector to the acquisition module and secure the cable connector by tightening the two screws.
19. Install the cable clamp on the Acquisition cable and secure it with the final mounting screw for the Acquisition module.
22 Rosemount Gas Ultrasonic 3410 Series Electronics
June 2022 Retrofit Instructions
Figure 2-13: Installing the Acquisition Module Cable Clamp
A. Acquisition cable connector B. Acquisition cable C. Cable clamp and screw
20. Wrap the excess cable around the Acquisition Module and place the electronics housing into position on the base enclosure.
21. Ensure the O-ring on the base of the upper enclosure is properly seated. Position the upper enclosure housing so that it is in the desired orientation for reattaching the electrical conduit.
22. Install the 3410 meter upper enclosure on the base enclosure once the transducers are wired correctly.
23. Install the two screws with the 6 mm Allen wrench to secure the upper enclosure to the base enclosure.
Retrofit Instructions 23
Retrofit Instructions June 2022
Figure 2-14: Bolt the Upper Enclosure to Base Enclosure
24. Remove the end cap next to the conduit entries from the upper enclosure. #unique_21/
unique_21_Connect_42_fig_B809C95392804A6EA94B366E5DF175 C4 shows the layout of the electronics.
24 Rosemount Gas Ultrasonic 3410 Series Electronics