T-xxx Transducer Assembly Upgrade Kit
Instructions for Rosemount™ Gas Ultrasonic
Meters
Safety and approval information
This Rosemount product complies with all applicable European directives when properly installed in accordance with the
instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The EU declaration
of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions are available
on the internet at www.emerson.com or through your local Emerson support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at http://
www.emerson.com.
For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the user manual.
Product data sheets and manuals are available from the Emerson website at http://www.emerson.com.
Return policy
Follow Emerson procedures when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Emerson employees. Emerson will not accept your
returned equipment if you fail to follow Emerson procedures. Return procedures and forms are available on our web support site
at Emerson.com, or by phoning the Emerson Customer Service department.
Emerson Flow customer service
Email:
• Worldwide: http://flow.support@emerson.com
• Asia-Pacific: http://APflow.support@emerson.com
Telephone:
North and South America
United States800 522 6277U.K.0870 240 1978Australia800 158 727
Welcome to the Rosemount™ 3410 and Mark III Series Ultrasonic meter upgrade manual
for T-xxx transducers.
This manual has been designed to provide you with a step-by-step set of instructions for
upgrading a Rosemount Gas Ultrasonic meter with T-xxx series transducers.
NOTICE
Please read the Before you begin section to ensure you have all the components necessary
to perform the upgrade before taking the meter out of service.
1.2Before you begin
1.2.1
This section shows all the parts and tools required to perform the upgrade. Ensure that all
these items are available before taking the meter out of service.
Required parts
Ensure that you have received the T-xxx Series upgrade kit. The kit will include:
1. Assembled T-xxx transducer assemblies.
2. Hard copy of a Zero Flow Calibration report including the lengths and serial
numbers of all the components of the T-xxx transducer assemblies.
3. USB flash drive containing:
a. Electronic copy of the Excel Zero Flow Calibration report, which can be
imported into the Transducer Swap-Out wizard in MeterLink™ and written
into the meter’s configuration to eliminate manual entry of these values.
b. Latest production firmware for Mark III electronics and 3410 electronics.
Instructions manual7
IntroductionInstructions manual
February 202200809-0100-3106
c. MeterLink PC software installation file that supports T-xxx transducers.
Note
The T-xxx upgrade kits only work with ANSI 1500 and lower pressure class meters
designed for J, K, Y, or Z mounts. This nomenclature is marked on the side of the mount
itself (for example, J00, J25, J50, J75). If the mounts are marked with another letter (for
example, M, P, U), then the upgrade kit will not work for the meter. Contact Emerson
Customer Support for other options.
1.2.2Required tools and supplies
1. Torque wrench (adjustable to 35 ft.-lbs.)
2. 5/16” hex bit socket for torque wrench
3. Ethernet or Serial cable for connecting a laptop with MeterLink to the meter
4. Molykote 111 or similar O-ring lubricant
1.2.3
Before removing the meter from service
Procedure
1. Install the latest version of MeterLink on your computer.
The latest version at the time the upgrade kit was assembled is on the included USB
flash drive.
Note
The MeterLink installation program is available for download from the
Measurement and Control, Inc. website: emerson.com/en-us/catalog/meterlink.
Note
MeterLink versions starting with 1.60 support T-XXX transducers.
2. Make sure you can connect MeterLink to the meter.
3. Check the firmware version in the meter using the Meter Information dialog under
the Meter pull-down menu. The minimum firmware versions supported are noted
below. Emerson recommends if the USB flash drive contains a more recent firmware
version, upgrade the meter to the latest version before proceeding.
a) For Mark III electronics, version 1.82 or later firmware supports T-xxx
transducers.
b) For 3410 electronics, version 1.42 or later firmware supports T-xxx
transducers.
4. Using Edit/Compare Configuration from the Tools pull-down menu, read and save
the configuration to file.
5. Using Maintenance Log, take a two-minute maintenance log to have a record how
the meter was performing prior to the upgrade.
6. Open the Transducer Swap-Out wizard from the Tools pull-down menu.
• If you receive the message: This meter supports the Transducer
Swap-Out using zero flow calibration, but it has not
previously been entered, you need to make sure you have the meter
housing lengths for each chord of the meter body.
The meter housing lengths for the body are stamped on a metal tag (values are
labeled BODY FACE TO FACE) attached to the meter body and can also be
obtained from an original Zero Flow Report that was generated when the meter
was first assembled.
Figure 1-1: Metrology Tag
If you cannot get the values from the tag and you do not have the original Zero
Flow Report or CMM report, contact Customer Service and they can provide this
information for you.
Important
Make sure to have the meter body serial number available before contacting
support.
• If you did not receive the message above when opening the Transducer Swap-
Out wizard, make sure the Meter Housing chord lengths are not zero in the
dialog or you will need to locate the meter housing chord lengths.
7. Once you have this information, you can proceed with the upgrade.
1. Follow all company procedures and safe practices to remove the meter from
service. The meter must be fully vented to the atmosphere prior to proceeding with
the next section.
2. While the transducer cabling is all intrinsically safe, it is still recommended to
remove power from the meter before beginning the upgrade as an additional level
of safety.
WARNING
FLUID CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the
transducer holder.
Attempting to do so may release pressurized gas or fluid, resulting in serious injury
or equipment damage.
CAUTION
ESCAPING GASES OR FLUIDS HAZARD
The purchaser of the meter is responsible for the selection of Rosemount
components/seals and materials compatible with the chemical properties of gas
flow measurement.
Failure to select the suitable meter component/seals may cause escaping gases or
liquids, resulting in injury or equipment damage.
2.2Remove the T-xx transducers
Procedure
1. Remove the transducer cables from the back of the transducers.
a) For meters with standard chordsets, unthread the cable nut by hand and
remove the transducer chordset.
™
Instructions manual13
Removing the T-xx transducersInstructions manual
February 202200809-0100-3106
Figure 2-1: Transducer Assembly with Standard Chordset
A. Transducer mount
B. Mount bolt
C. T-slot transducer holder
D. Cable nut
E. Chordset
b) For meters with red mini chordsets, remove the transducer chordset by
removing the two hitch pins securing it to the back of the transducer. Once
the hitch pins are removed, pull the chordset away from the transducer.
Figure 2-2: Transducer Assembly with Red Mini Chordset
A. Transducer mount
B. Mount bolt
C. Hitch pin
D. Chordset
2. Use the 5/16” Allen wrench to remove the four to six mount bolts holding each
mount in place. Keep the bolts clean and together because they are used to attach
the new T-xxx transducers to the meter body.
3. Pull all the transducer assembles off the meter body and set them aside.
4. Use a cloth to wipe out the transducer ports, especially the surface to which the
transducer mount O-ring mated. Visually check the ports to make sure they are free
of debris as well. Clean as needed.
Instructions manualInstalling the T-xxx transducers and checkout
00809-0100-3106February 2022
3Installing the T-xxx transducers and
checkout
3.1Installing the T-xxx transducers
Procedure
1. Take the T-xxx transducer assemblies and apply a thin layer of Molykote 111 or
other similar O-ring lubricant. One at a time, install the transducer assemblies in the
correct transducer port.
The Zero Flow Report provided with the kit will show the port location (for example,
A1, B2, ….) for each transducer assembly. The serial number on the side of the stalk
can be compared to the serial number listed on the Zero Flow Report to know which
assembly you are installing and into which port location.
Figure 3-1: Installed Transducer Assembly with Standard Chordset
A. Transducer Assembly
B. Mount bolt
C. Cable nut
D. Chordset
Instructions manual15
Installing the T-xxx transducers and checkoutInstructions manual
February 202200809-0100-3106
Figure 3-2: Installed Transducer Assembly with Red Mini Chordset
A. Transducer Assembly
B. Mount bolt
C. Hitch pin
D. Chordset
2. Use the mount bolts to attach the T-xxx transducer assemblies to the meter body.
Tighten bolts to 35 ft-lbs.
3. Attach the transducer cables to the back of each transducer with the reverse steps
taken to remove them from the old transducer assemblies.
3.2Configuring the meter
Procedure
1. If power was removed from the meter, ensure the environment is safe and reapply
power to the meter.
2. Connect to the meter using MeterLink.
3. Open the Transducer Swap-Out wizard from the Tools pull-down menu.
Note
If you receive a message that This meter supports the Transducer Swap-
Out using zero flow calibration…, click Yes to use this newer version of
the wizard.
4. While in the Transducer Swap-Out wizard, change the Transducer type selection to
the appropriate transducer type (for example. T-200). This will clear out the fields
for the components being exchanged.
5. Click Open and select the Excel zero flow report on the provided USB flash drive to
import all the new component lengths and serial numbers for the new T-xxx
Instructions manualInstalling the T-xxx transducers and checkout
00809-0100-3106February 2022
transducer assemblies. Alternatively, you can manually enter all the values from the
hard copy of the zero flow report.
6. If the Meter Housing fields are zero, you need to enter these values from the
metrology tag attached to the meter body, or from an original zero flow report for
the meter.
7. Click Write to write these new parameters to the meter electronics.
8. Click Export to create and save a new Excel report with all the new parameters and
existing parameters from the meter.
9. Close the Transducer Swap-Out wizard.
3.3Return the meter to service and checkout
Procedure
1. Follow all company procedures and safe practices to reapply pressure to the meter.
This should include at a minimum a soapy water leak test between the meter body
and the T-xxx transducer assemblies to ensure the new O-rings are sealing properly
and not damaged during installation.
2. Use the Meter Monitor dialog to ensure all chords are acquired and meter
performance is good. If there are any yellow or red lights, click Check Status to view
the errors and troubleshoot any problems.
Note
Chordal SOS spread may still be high until the meter is under flowing conditions and
thermal equilibrium has been achieved.
3. If flow is not applied, apply flow to the meter if possible and verify meter
performance remains good.
4. Using Maintenance Logs, take a two-minute maintenance log to have a record of
how the meter was performing after the upgrade.
Emerson Terms and Conditions of Sale are available upon
request. The Emerson logo is a trademark and service mark of
Emerson Electric Co. Rosemount is a mark of one of the
Emerson family of companies. All other marks are the property
of their respective owners.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.