Danfoss UV Data sheet

Data sheet
760 nm 1 mm10 nm 380 nm
Ultraviolet Flame Sensor UV
Application
Function
Light rays are sent at different wavelengths. The light that the human eye can see lies in the range of 380 to 760 nm and is called visible light. In the blue flame oil burner the flame’s visible light is very dim and therefore cannot be detected by the LD/LDS sensor series, which is built on the photoresistance principle. In the case of UV sensors, the flame’s ultraviolet light is used instead, which is detected by a UV diode. In the sensors, the sensor signal is converted by an amplifier circuit into a signal that the control can use.
Ultraviolet light Visible light Infrared light
Danfoss UV sensors are used to detect the flame in blue flame oil burners. They are intended for use together with Danfoss controls in the OBC 80 and BHO 70 series. The UV sensors meet the requirements of the EN230:2005 standard and also meets the requirements of the RoHS and WEEE Directives.
To ensure unambiguous detection of the ultraviolet light, it is first transmitted through a lens that focusses the light rays onto the surface of the UV diode, and then through a filter that reduces the sensitivity to a desired wavelength range. Danfoss UV sensors have max. sensitivity at 375 nm, which is apparent from the logarithmic diagram, where sensitivity at 375 nm is given to 100%.
Wavelength
© Danfoss | 2019.07 VD.AU.A3.02 | 1
Data sheet Ultraviolet Flame Sensor UV
12
Electrical connection
Mounting
The UV sensors are connected to Danfoss controls in the OBC 80 and BHO 70 series as shown in the diagram.
The UV sensors are designed to be very direc­tional to limit the influence of other light sources, primarily the ignition spark that emits UV light. The sensors must be installed so that they are not affected by light from the ignition spark and, at the same time, so that the best possible input from the flame is achieved. It is very important that the mounting flange or piping used has the UV sensor fixed in the correct position.
Note that the blue wire must be connected to terminal 11 and not to the common 0 on terminal 2 or the appertaining auxiliary terminals in the base section.
3
The UV sensors’ sensitivity is shown as a function of the angle deviation in Fig. 1 and 2.
If a sufficient signal is not achieved, we recom­mend selecting a UV sensor of higher sensitivity. To determine the quality of the signal, use the measurement setup below. If there is no flame/darkness, the value should be ≤ 5 µA and when there is a flame/light it should be ≥65 µA.
1. Brown
2. Blue
3. Black
Fig. 1 Fig. 2
Fig. 3
2 | VD.AU.A3.02 © Danfoss | 2019.07
Loading...
+ 2 hidden pages