Data Sheet
Pressure transmitter
Type MBS 1700 and MBS 1750
For general purpose
The compact pressure transmitters MBS 1700
and MBS 1750 are designed for use as a general
purpose transmitter, and oers a reliable
pressure measurement, even under harsh
environmental conditions.
The version MBS 1750 with integrated pulsesnubber is designed for use in applications with
severe medium inuences like cavitation, liquid
hammer or pressure peaks and oers a reliable
pressure measurement, even under harsh
environmental conditions.
Excellent vibration stability, robust
construction, and a high degree of EMC / EMI
protection equip the pressure transmitter to
meet the most stringent industrial
requirements.
Features
• Enslosure and wetted parts of acid-resistant
stainless steel (AISI 316L)
• Pressure ranges in relative (gauge) from 0 –
25 bar
• Output signal: 4 – 20 mA
• Pressure connections:
◦ G 1/4A & G 1/2A EN837 (MBS 1700)
◦ G 1/4 DIN 3852-E, Gasket DIN 3869-15 (MBS
1750)
• Fully digitally compensated
AI309034611216en-000201
Pressure transmitter, Type MBS 1700 and MBS 1750
Application
Application and media conditions (MBS 1750)
Figure 1: MBS 1750
Application
Cavitation, liquid hammer and pressure peaks may occur in liquid lled systems with changes in ow velocity, e.g.
fast closing of a valve or pump starts and stops. The problem may occur on the inlet and outlet side, even at rather
low operating pressures.
Media condition
Clogging of the nozzle may occur in liquids containing particles. Mounting the transmitter in an upright position
minimizes the risk of clogging, because the ow in the nozzle is limited to the start-up period until the dead volume
behind the nozzle orice is lled. The media viscosity has only little eect on the response time. Even at a viscosities
up to 100 cSt, the response time will not exceed 4 ms.
© Danfoss | Climate Solutions | 2021.02 AI309034611216en-000201 | 2
Accuracy (incl. non-linearity, hysteresis and repeatability)
≤ ± 0.5% FS (typ.)
Non-linearity BFSL (conformity)
Hysteresis and repeatability
Thermal zero point shift
≤ ± 0.1% FS/10K (typ.)
Thermal sensitivity (span) shift
≤ ± 0.1% FS/10K (typ.)
Response time
Air and gases (MBS 1700)
Overload pressure (static)
Durability, P: 10 – 90% FS
Nom. output signal (short-circuit protected)
Supply voltage [UB], polarity protected
Supply – current consumption
Supply voltage dependency
Load [RL] (load connected to 0 V)
RL ≤ (UB - 9 V)/0.02 A [Ω]
Ambient temperature range
Compensated temperature range
Transport / storage temperature range
Vibration stability
Sinusoidal
IEC 60068-2-6
20 g, 25 Hz – 2 kHz
EN 10088-1; 1.4404 (AISI 316 L)
EN 10088-1; 1.4404 (AISI 316 L)
Glass lled polyamid PA 6.6
Pressure transmitter, Type MBS 1700 and MBS 1750
Product specication
Technical data
Table 1: Performance (EN 60770)
Table 2: Electrical specications
Table 3: Environmental conditions
Table 4: Mechanical characteristics
© Danfoss | Climate Solutions | 2021.02 AI309034611216en-000201 | 3