Surge
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
1kV differential mode
2kV common mode
3 Vrms
150 kHz to 80MHz
1kV differential mode
2kV common mode
3 Vrms
150 kHz to 80MHz
Mains power quality should be that of a
typical commercial or hospital environment.
Portable and mobile RF communications
Portable and mobile RF communications
equipment should be used no closer to any
part of the FS-L4202D, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W)
3. Guidance and manufacturer's declaration - electromagnetic immunity
Immunity test
Radiated RF
IEC 61000-4-3
IEC 60601 Test level
3 V/m
80.0 MHz to 2.5 GHz
Complince level
3 V/m
80.0 MHz to 2.5 GHz
Electromagnetic environment -guidance
Recommended separation distance
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
deter-mined by an electromagnetic site
survey, a should be less than the compliance
level in each frequency range.
80MHz to 800MHz
80MHz to 2.5GHz
FS-L4202D is intended for use in the electromagnetic environment specified below.
The customer or the user of FS-L4202D should assure that it is used in such an environment.
Immunity test
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
IEC 60601 Test level
3.0A/m
<5 % U
(>95 % dip in U )
for 0.5 cycle
40 % U
(60 % dip in U )
for 5 cycle
70 % U
(30 % dip in U )
for 25 cycle
<5 % U¨‰
(<95 % dip in U¨‰ )
for 5 s
Complince level
3.0A/m
<5 % U
(>95 % dip in U )
for 0.5 cycle
40 % U
(60 % dip in U )
for 5 cycle
70 % U
(30 % dip in U )
for 25 cycle
<5 % U
(<95 % dip in U )
for 5 s
Electromagnetic environment -guidance
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Mains power quaility should
be that of a typical commercial or hospital
environment. If the user of FS-L4202D
requires continued operation during
power mains interruptions, it is
recommended that FS-L4202 be powered
from an uninterruptible power supply or a
battery
NOTE : U is the A.C. mains voltage prior to application of the test level.
4. Recommended separation distances between portable and mobile
RF communications equipment and the FS-L4202D
- The is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the FS-L4202D can help
- Prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile
- RF communications equipment (transmitters) and the FS-L4202D as recommended
below,according to the maximum output power of the communications equipment.
150kHz to 80MHz
V1=3Vrms
Rated
maximum output
power of tr
ansmitter [W]
Separation distance according to frequency of transmitter[m]
80MHz to 800MHz
E1=3V/m
800MHz to 2.5GHz
E1=3V/m