The Dakota Ultrasonics model PZX-7 is a precision ultrasonic micrometer. Based on
the same operating principles as SONAR, the PZX-7 is capable of measuring the
thickness of various materials with accuracy as high as 0.0001 inches, or 0.001
millimeters. The principle advantage of ultrasonic measurement over traditional
methods is that ultrasonic measurements can be performed with access to only one
side of the material being measured.
Dakota Ultrasonics maintains a customer support resource in order to assist users
with questions or difficulties not covered in this manual. Customer support may be
reached at any of the following:
Dakota Ultrasonics Corporation
1500 Green Hills Road, #107
Scotts Valley, CA 95066
Tel: (831) 431-9722
Fax: (831) 431-9723
www.dakotaultrasonics.com
1.1 Disclaimer
While the PZX-7 is a basic digital instrument, it is equipped with a number of
measurement modes and transducer options for additional versatility. It is strongly
recommended that the contents of this manual be read in its entirety to avoid
erroneous measurements based on operator error. The user is solely responsible for
proper use, setup and interpretation of the measurements acquired.
1
Page 6
CHAPTER TWO
KEYPAD, MENU, DISPLAY & CONNECTORS
The Keypad
2.1 ON/OFF/ENTER Key
The ON/OFF/ENTER key powers the unit ON or OFF. Since the same key is also
used as an ENTER key, the gauge is powered off by pressing and holding down the
key until the unit powers off.
Once the gauge is initially powered on, this key will function as the ENTER key,
similar to a computer keyboard. This key will be used to select or set a menu option.
Note: Unit will automatically power off when idle for 5 minutes. All current settings
are automatically saved prior to powering off.
2.2 PRB 0 Key
The PRB 0 key is used to “zero” the PZX-7 in much the same way that a mechanical
micrometer is zeroed. If the gauge is not zeroed correctly, all of the measurements
that the gauge makes may be in error by some fixed value. This feature is only used
2
Page 7
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
when using single element contact transducers. Refer to page 15 for further
explanation of this important feature.
2.3 CAL Key
The CAL key is used to enter and exit the PZX-7's calibration mode. This mode is
used to adjust the sound velocity value that the PZX-7 will use when calculating
thickness. The tool will either calculate the sound-velocity from a sample of the
material being measured, or allow a known velocity value to be entered directly. This
provides increased linearity between transducers. Refer to page 16 for an
explanation on the various calibration options.
2.4 GAIN Key
The PZX-7 has an adjustable gain feature for additional versatility to control the
overall output of the gauge, to compensate for various materials and applications.
There are five gain settings (VLOW, LOW, MED, HIGH, VHI) when pulse-echo or
thin PLAS mode is selected, and three gain settings (LOW, MED, HIGH) for all other
modes where automatic gain control is used. Refer to page 29 for details regarding
the gain feature.
2.5 SCAN Key
The PZX-7 offers a high speed scan feature. This feature allows for scanning larger
areas on a given test material, while still offering reasonable representation of
thickness over the area scanned. Refer to page 30 for an explanation on the scan
feature.
2.6 +/- Increment/Decrement Key’s
The +/- Keys are used to increment/decrement values, navigate menus, and select
menu options.
2.7 MODE Key
The MODE key toggles the measurement modes. If a delay line transducer (dLY) is
selected from the PROB menu option, echo-echo (E-E) and interface-echo (I-E)
modes will be toggled. If a contact transducer (Con) is selected, pulse-echo (P-E)
and echo-echo (E-E) modes will be toggled. Finally, if the plastic transducer (PLAS)
3
Page 8
Dakota Ultrasonics
option is selected, the thickness options thin (thin) and thick (thK) will be toggled. The
measurement modes and transducer styles offer the flexibility to address specific
application scenarios. Refer to page 11 for an explanation on the various calibration
options.
2.8 MENU Key
The MENU key is used to access and set all of the additional features of the PZX-7
that are not at the top level of the keypad with a dedicated key. The features and
setting are outlined in the table below:
Menu Feature Items:
Matl Scan Alarm Diff Unit Lite Beep Probe VX
Aluminum On On On English On On Delay On
Steel Off Off Off Metric Off Off Plastic Off
Stainless Options Options OptionsOptions Contact
Iron Set Lo Set
Nominal
Plexiglass Set Hi High Med .12 5
PVC High .12 10
Plastic .25 5
Poly Urea .25 10
User 1 .37 5
User 2 .50 5
.50 10
* Diameter options are only available for contact transducers.
** Diameter/Frequency options are only available for contact transducers.
Low Lo Options
Here’s a quick overview of navigating through the various features in MENU:
*
Navigating the Features in Menu
1) Press the key once to enter the sub menu items.
2) Press the keys to toggle through the features.
3) To enable or edit the status of any feature, press the key.
4
Page 9
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
4) The edit icon will start blinking to indicate that the PZX-7 is currently in EDIT
mode.
5) Press the keys to toggle through the setting options.
6) Press the key to accept changes and return to the top level of
features, or the key at any time to abort changes and return the
measurement screen.
2.9 The Display
The PZX-7 uses a custom glass LCD backlit low temperature display for use in a
variety of climate conditions. It contains graphic icons, as well as both 7 and 14
segment display areas. Let’s take a closer look and what all these things are telling
us:
A. Edit: This icon will be displayed, and blinking, to let a user know when they
are in an edit mode to change a value or setting.
B. Large 7 segment: The thickness measurement, velocity or alpha message
will be displayed in this area.
5
Page 10
Dakota Ultrasonics
C. Measurement Modes: This group of icons indicates which measurement
mode the PZX-7 is currently using. The modes are pulse-echo, echo-echo,
interface-echo, and plastics (PLAS).
D. Stability/Repeatability Indicator: This is used in conjunction with the
thickness measurement as a reference for the validity of the measurement.
The PZX-7 takes multiple measurements per second, and when all the vertical
bars are illuminated, it’s a reference that the same thickness value is reliably
being measured multiple times per second.
E. Battery: Indicates the amount of battery life the PZX-7 has remaining.
F. Backlight : When this icon is illuminated, it indicates the backlight is on.
G. Small 7 Segment: The material velocity, speed the sound wave travels
through a given medium/material, is displayed in this area, informing the user
what material the PZX-7 is currently calibrated too. This area is also used for
alpha messages in the menu and edit modes.
H. Units: This combination of icons are illuminated in different sequences to
inform the user what measurement units are currently being displayed in the
small 7 segment area.
I. Small 14 Segment: Displays the current gain setting of the PZX-7 DL. In P-E
mode the options are VLOW, LOW, MED, HIGH, VHI, and LOW, MED, HIGH
with modes E-E, I-E, and PLAS. MED is the default setting.
J. Units: This combination of icons are illuminated in different sequences to
inform the user what measurement units are currently being displayed in the
large 7 segment area. The plus/minus icon is illuminated when the DIFF
(differential) feature is activated.
K. Small 14 Segment: The material type is displayed in this area. If it is set to a
value of one of the materials in our material list, it will be displayed in alpha
characters indicating the material type. Otherwise it will be set to CUST,
indicating custom material type.
L. Features:
display which features are currently enabled. For a complete list of the menu
features in the PZX-7, Refer to page 4 for a list. The PZX-7 can be locked
once calibrated, to avoid accidently changing the calibration. When this icon is
illuminated, the PZX-7 is in lock mode. Refer to page 38 for an explanation on
locking the PZX-7.
The icons illuminated in this row across the bottom of the LCD
6
Page 11
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
Delay Line Contact
The Transducer is the “business end” of the PZX-7. It transmits and receives
ultrasonic sound waves that the PZX-7 uses to calculate the thickness of the material
being measured. The transducer connects to the PZX-7 using a single shielded
coaxial cable with microdot connectors.
Delay Line Transducer
The single element delay line transducer is commonly used with the PZX-7 for
materials with an approximate range of 0.006” to 1.00” in steel. The measurement
modes used with this style of transducer are echo-echo (E-E) and interface-echo (IE). Echo-echo mode will be used for the lower end of the measurement range, while
interface-echo for the upper end of the measurement range. Interface-echo mode
measures from the end of the delay line to the first return echo from the back wall of
the material, while echo-echo measurements are made between the first and second
return echoes from the back wall of the material. The frequency range for the
versions Dakota commonly supply are from 10MHz to 20MHz.
The Delay line is fastened to the transducer with a retaining ring. A drop of couplant
is applied between the delay line and transducer body. Refer to the diagram above.
The couplant should be checked on a regular basis to keep it from drying out. To
replenish, unscrew the retaining ring counterclockwise, separate the delay line from
the body, ‘clean both surfaces’, apply a drop of couplant, and reassemble.
Delay Line Graphite Tip (For use with Plastics Mode)
The PZX-7 has a special “PLAS” mode that uses the standard delay line transducer,
as described above, but with a replaceable graphite delay tip. This mode and tip
should be used when the velocity of the plastic material measured is similar to the
velocity of the standard acrylic delay tip, and not achieving any signal/measurement.
The graphite tip creates a ‘mismatch’ between the tip material and the plastic material
measured.
The Delay line is fastened to the transducer with a retaining ring. A drop of couplant
is applied between the delay line and transducer body. Refer to the diagram above.
The couplant should be checked on a regular basis to keep it from drying out. To
7
Page 12
Dakota Ultrasonics
replenish, unscrew the retaining ring counterclockwise, separate the delay line from
the body, ‘clean both surfaces’, apply a drop of couplant, and reassemble.
Contact Transducer
The single element contact style transducers are commonly used for materials with
an approximate range of 0.040” to the maximum range of the PZX-7 rated in steel
and depending on the transducers diameter and frequency. The frequency range is
5.0 to 10 MHz, with diameters available at 0.125, 0.250, 0.375 and 0.500 inches.
The measurement modes used with this style of transducer are pulse-echo (P-E) and
echo-echo (E-E).
Measuring
In order for the transducer to do its job, there must be no air gaps between the wearface and the surface of the material being measured. This is accomplished with the
use of a "coupling" fluid, commonly called "couplant". This fluid serves to "couple", or
transfer, the ultrasonic sound waves from the transducer, into the material, and back
again. Before attempting to make a measurement, a small amount of couplant
should be applied to the surface of the material being measured. Typically, a single
droplet of couplant is sufficient.
After applying couplant, press the transducer (wear face down) firmly against the
area to be measured. The Stability Indicator should have six or seven bars
darkened, and a number should appear in the display. If the PZX-7 has been
properly "zeroed" (see page 15), only if P-E mode is currently selected, and set to the
correct sound velocity (see page 16), the number in the display will indicate the
actual thickness of the material directly beneath the transducer.
If the Stability Indicator has fewer than five bars darkened, or the numbers on the
display seem erratic, first check to make sure that there is an adequate film of
couplant beneath the transducer, and that the transducer is seated flat against the
material. If the condition persists, it may be necessary to select a different transducer
(size or frequency) for the material being measured. See page 13 for information on
transducer selection.
While the transducer is in contact with the material that is being measured, the PZX-7
will perform four measurements every second, updating its display as it does so.
When the transducer is removed from the surface, the display will hold the last
measurement made.
2.10 Top & Bottom End Caps
8
Page 13
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
The top end cap is where all connections are made to the PZX-7. The diagram
above shows the layout and description of the connectors:
Transducer Connectors
Refer to Diagram: The transducer connectors and battery cover/probe zero disk are
located on the PZX-7’s top end cap. The transducer connector is a Lemo “00”.
Probe Zero Disk & Battery Cover
Refer to Diagram: The Battery cover is the large round disk shown in the diagram.
Note: This same disk is also used as a probe zero reference disk. Simply remove
the cover when replacing the batteries (2 AA cells). When performing a manual
probe zero function, simply place the transducer on disk making firm contact.
Important: Be sure the battery polarity is correct, which can be found on the back
label of the PZX-7.
Note: Rechargeable batteries can be used, however they must be recharged outside
of the unit in a standalone battery charger.
USB-C Connector
Refer to Diagram: The USB-C connector, located on the bottom end cap, is a mini
type C female connector. It is designed to connect directly from the PZX-7 to a
standard USB type A port on a PC. The cable supplied with the PZX-7 is a USB type
C to a USB type A (pt# N-003-0330). See page 41 for information on connectivity.
Note: This connector is also used to upgrade the PZX-7 with the latest version of
firmware.
9
Page 14
CHAPTER THREE
PRINCIPALS OF ULTRASONIC MEASUREMENT
3.1 Time versus thickness relationship
Ultrasonic thickness measurements depend on measuring the length of time it takes
for sound to travel through the material being tested. The ratio of the thickness
versus the time is known as the sound velocity. In order to make accurate
measurements, a sound velocity must be determined and entered into the
instrument.
The accuracy of a thickness measurement therefore depends on having a consistent
sound velocity. Some materials are not as consistent as others and accuracy will be
marginal. For example, some cast materials are very granular and porous and as a
result have inconsistent sound velocities.
While there are many different ultrasonic techniques to measure thickness, which will
be discussed below, all of them rely on using the sound velocity to convert from time
to thickness.
3.2 Suitability of materials
Ultrasonic thickness measurements rely on passing a sound wave through the
material being measured. Not all materials are good at transmitting sound.
Ultrasonic thickness measurement is practical in a wide variety of materials including
metals, plastics, and glass. Materials that are difficult include some cast materials,
concrete, wood, fiberglass, and some rubber.
3.3 Range of measurement and accuracy
The overall measurement capabilities, based on the wide variety of materials, are
determined by the consistency of the material being measured (homogeneity). The
range of thickness that can be measured ultrasonically depends on the material type
and surface, as well as the technique being used and the type of transducer and
frequency. The range will vary depending on the type of material being measured.
Accuracy, is determined by how consistent the sound velocity is through the sound
path being measured, and is a function of the overall thickness of the material. For
example, the velocity in steel is typically within 0.5% while the velocity in cast iron
can vary by 4%.
3.4 Couplant
All ultrasonic applications require some medium to couple the sound from the
transducer to the test piece. Typically a high viscosity liquid is used as the medium.
The sound frequencies used in ultrasonic thickness measurement do not travel
through air efficiently. By using a liquid couplant between the transducer and test
piece the amount of ultrasound entering the test piece is much greater.
10
Page 15
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
3.5 Temperature
Temperature has an effect on sound velocity. The higher the temperature, the slower
sound travels in a material. High temperatures can also damage transducers and
present a problem for various liquid couplants.
Since the sound velocity varies with temperature it is important to calibrate at the
same temperature as the material being measured.
Normal temperature range
Most standard transducers will operate from 0F to 250F.
High temperature measurements
Special transducers and couplants are available for temperatures above 250F up to
1000F with intermittent contact. It is necessary to cool the transducer by
submerging it in water between measurements.
Modes and temperature errors
In addition to errors caused by velocity changing with temperature, some modes
(measurement techniques) are affected more than others. Multi-echo techniques
offer temperature compensation to help minimize these errors.
3.6 Measurement Modes
This section will cover the different measurements modes of the PZX-7, the
transducers required, and the reasons for using specific modes:
Pulse-Echo (P-E) Mode:
Pulse-echo mode measures from the initial pulse (sometimes referred to as an
artificial zero) to the first echo (reflection), and is only used if the transducer type is
set to contact mode. In this mode, a manual zero using the disk (battery cover) will
need to be performed prior to calibrating the PZX-7. The transducer is placed on the
reference disk located on top of the PZX-7, and the PRB 0 key pressed to establish a
zero point for the transducer connected.
Errors from surface coatings and temperature variations can occur in pulse-echo
mode. Therefore, coatings should be removed, and frequent probe zero’s should be
performed if experiencing temperature gradients.
Echo-Echo (E-E) Mode – Through paint
The echo-echo mode measures between the first and second return
echoes/reflections. This mode is most commonly used with the high frequency delay
line style transducers for thin materials, and will also eliminate compensate for
temperature. This technique will also eliminate error due to epoxy based coated
surfaces by eliminating the error from the coating and measuring only the second
11
Page 16
Dakota Ultrasonics
layer or base material without having to remove the coating. The disadvantage is
that two return echoes are required to effectively measure the test material. As a
result, the type and thickness of the coating will affect the ability to achieve a
successful measurement. Both delay line and contact style transducers can be used
for through paint/coating measurements.
Interface-Echo (I-E) Mode
The interface-echo mode measures from the end of the “‘interface/delay line” to the
first return echo/reflection. This mode can be used for thicker materials up to 1”, and
for attenuative materials that don’t produce a sufficient second echo. The standard
measuring range is approximately 0.060” up to 1.00” rated in steel, and offers
temperature compensation for the heating and cooling of the transducer itself.
Plastics (PLAS) Mode
The plastics mode is a special setup for measuring thin plastics using a modified
interface-echo mode. When the standard acrylic delay tip is similar in velocity to the
plastic material measured, a return echo is not possible. Simply replacing the acrylic
tip with our graphite tip and using the ‘plastics’ mode will provide the necessary
mismatch necessary to measure the plastic successfully. The approximate range is
0.005” up to 0.300” depending on the type of plastic measured. This mode
additionally offers temperature compensation for the heating and cooling of the
transducer.
12
Page 17
CHAPTER FOUR
SELECTING THE MEASUREMENT MODE
4.1 Which mode & transducer do I use for my application?
High penetration plastics and castings
The most common mode for these types of applications is pulse-echo. Thicker cast
iron and plastics applications will generally require lower frequencies depending on
the material type and thickness. Frequencies from 5MHz to 10MHz are common.
Larger diameters offer greater penetration power based on the size of the crystal.
Thin Materials & Tough Access Areas
The high frequency delay line transducers are suitable when measuring thinner
consistent materials that pass sound easily, such as steel, aluminum, titanium, etc.
They offer temperature compensation with a suitable range up to 1” depending on
the delay line used. Replaceable cone tip options are also useful for difficult access
areas and diameters.
Thru Paint & Coatings
It’s common that users will be faced with applications where the material will be
coated with some type of protective paint or other epoxy based material. Since the
velocity of the coating is approximately 3 times slower than that of steel, pulse-echo
mode will produce an error if not removed. However, echo-echo mode can be used
with either delay line or contact style transducers to eliminate the error from the
coating, and measure only the base material. High damped or high frequency
transducers are most suitable for echo-echo measurement mode.
Plastics
Thin plastics are best suited using the delay line transducer with either the acrylic
delay tip in echo-echo or interface-echo modes. If the plastic measured is similar in
velocity to the acrylic tip and not measuring adequately, the replaceable graphite tip
and plastics mode will serve as the best option. Thicker plastics will generally require
a lower frequency contact style transducer for additional penetration. Increasing the
diameter of the transducer will also offer greater output.
High temperature
High temperature applications will require a special delay line material that isn’t as
absorptive of temperature.
Noisy Material
Materials such as titanium, stainless steel, and aluminum may have inherent surface
noise issues (mirroring effect). This can be avoided using a high frequency delay line
13
Page 18
Dakota Ultrasonics
transducer in either echo-echo or interface-echo modes. If using a contact style
transducer, a high frequency option should be considered.
Restricted access
Measuring materials with extreme curvatures or restricted access are best suited for
higher frequencies and smaller diameter transducers or tip options.
14
Page 19
CHAPTER FIVE
MAKING MEASUREMENTS
The steps involved in making measurements are detailed in this section. The
following sections outline how to setup and prepare your PZX-7 for field use.
In pulse-echo mode the probe zero must be performed on the reference disk (battery
disk) attached to the top of the instrument. This compensates for variation in the
transducer. In all modes the sound velocity must be determined. The sound velocity
is used to convert the transit time to a physical length. The sound velocity can be
selected from the material chart at the end of this manual, or for greater precision
determined by measuring and calibrating from a known thickness of the actual
material. To enter the velocity from a table, look up the material on the chart in the
appendix of this manual and refer to the section below on Calibration to a Known
Velocity. To determine the velocity of a single sample, refer to the Material
Calibration section on page 16.
When measuring curved materials, it’s more accurate to calibrate from two test
points, one at the minimum limit of the target thickness and one at the maximum limit.
In this case the reference disk is not used, and considered a ‘two-point’ calibration
described on page 19.
5.1 Probe zero
A probe zero is only necessary if a contact style transducer is used in pulse-echo
mode, and the contact option is enabled in the menu. A probe zero should be done
prior to calibrating the instrument, and done on a periodic basis when a temperature
gradient is an issue. If the PZX-7 is not zeroed correctly, all measurements will be in
error by some fixed value. In order to perform a probe zero, you must be in pulseecho mode.
Setting the zero point of the PZX-7 is important for the same reason that setting the
zero on a mechanical micrometer is important. It must be done prior to calibration,
and should be done throughout the day to account for any temperature changes in
the probe. If the PZX-7 is not zeroed correctly, all the measurements taken may be
in error by some fixed value. The zero can only be performed with the measurement
mode set to pulse-echo (P-E). Therefore, if the PZX-7 is to use the echo-echo (E-E)
measurement mode and a manual zero is being performed, the PZX-7 will argue by
briefly displaying the message “nO prB0” (no probe zero), and automatically switch
modes to pulse-echo (P-E) where a zero can be performed.
The procedure to zero the transducer is outlined as below:
Performing a Manual Probe Zero (On Block)
15
Page 20
Dakota Ultrasonics
1) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the disk (battery cover) located at the top of the unit to
obtain a measurement.
2) Be sure all six repeatability/stability bars in the top left corner of the display
are fully illuminated and stable, and last digit of the measurement is toggling
only +/- .001” (.01mm).
3) Press the key to perform the manual zero. “PRB0” will briefly be
displayed on the screen, indicating the zero calculation is being performed.
5.2 Material Calibration
In order for the PZX-7 to make accurate measurements, it must be set to the correct
sound velocity of the material being measured. Different types of materials have
different inherent sound velocities. For example, the velocity of sound through steel
is about 0.233 inches per microsecond, versus that of aluminum, which is about
0.248 inches per microsecond. If the gauge is not set to the correct sound velocity,
all of the measurements the gauge makes will be erroneous by some amount.
The One Point calibration is the simplest and most commonly used calibration
method - optimizing linearity over large
greater accuracy over small
material velocity. The PZX-7 provides three simple methods for setting the sound-
velocity outlined below:
ranges by calculating both the probe zero, as well as the
ranges. The Two Point calibration allows for
Known Velocity
If the material velocity is known, it can be manually entered into the PZX-7, rather
than have the PZX-7 calculate the velocity value using a known thickness of the
same material type. The steps for entering the velocity are outlined below:
Using a Known Material Velocity
16
Page 21
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
1) With the transducer free from contact with the material, press the key
to display the current velocity.
2) Use the keys to scroll the velocity to the known target
value.
Note: The longer the keys are pressed and held, the faster the value will
increment/decrement.
Note: Pressing the key prior to pressing the key will abort the cal
routine without saving any changes.
3) Press the key to set the velocity value and return to the measurement
screen. The new velocity value will be shown at the top of the display.
Known Thickness
Often times the exact sound velocity of a material is unknown. However, a sample
with one or two known thicknesses can be used to determine the sound velocity. As
previously discussed, the PZX-7 has a one or two point calibration option. The one
point calibration option is most suited for linearity over large ranges. When using the
one point option, the calibration should be perform on the thickest side of the
measurement range for the best linearity for that range. For example, if the
measurement range is .100” (2.54mm) to 1.0” (25.4mm), the user should calibrate on
a known thickness sample close to 1.0” (25.4mm). Note:It’s always handy to carry
a set of mechanical calipers to use in conjunction with the PZX-7 for calibration of
various materials in the field:
One Point Calibration
17
Page 22
Dakota Ultrasonics
Note: Be sure that a probe zero has been performed prior to performing this
calibration procedure.
1) Physically measure an exact sample of the material, or a location directly on
the material to be measured, using a set of calipers or a digital micrometer.
Note: A sample or location on the test piece should be used as close to the
maximum thickness of the test range to minimize error.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the sample or actual test material. Be sure that the
reading is stable and the repeatability indicator in the top left corner of the
display is fully lit and stable.
3) Press the key to enter the calibration edit screen displaying the current
measurement value.
4) Use the keys to scroll to the known thickness value.
Note: The longer the keys are pressed and held, the faster the value will
increment/decrement.
Note: Pressing the key prior to pressing the key will abort the cal
routine without saving any changes.
5) Once the known thickness value is being displayed, press the key to
display the calculated material velocity edit screen.
Note: The calculated velocity can be edited, if needed, by pressing the
keys to scroll and edit the velocity value.
6) Press the key to set the calculated material velocity and return to the
measurement screen.
18
Page 23
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
Note: CHECK YOUR CALIBRATION! Place the transducer back on the
calibration point and verify the thickness. If the thickness is not correct, repeat
the steps above.
Two Known Thicknesses
The two point calibration should be considered when an application requires
improved accuracy over a small measurement range based on tolerance
requirements. This calibration option calculates both the ‘probe zero’ and ‘velocity
value. If the two point option is used, a probe zero is not required. For example, if
the measurement range was .080” (2.03mm) to .250” (6.35mm), two known samples
or locations on the test material would be needed for the minimum and maximum
boundaries of the test range. Using the range above, a one point calibration would
be performed at .250” (6.35mm) and a two point calibration at .080” (2.03mm), or
something close to the min/max values of the measurement range.
Note: The PZX-7 also offers the capability of setting the ‘probe zero’ to use any
reference standard as the ‘probe zero’ standard. For clarification, if it’s desired to use
a one inch reference of a specific material type as the ‘zero’ reference, performing
the first point of a two-point calibration sets the internal zero of the PZX-7. This
should be used only in manual probe zero mode “on block”. This can be done in ‘all’
measurement modes to correct the zero or offset as needed/required.
The following steps outline this procedure:
Two Point Calibration
1) Physically measure a minimum and maximum calibration point of the exact
sample material, or locations directly on the material to be measured, using
a set of calipers or a digital micrometer.
Note:A sample or location on the test piece should be used as close to the
minimum and maximum thickness of the test range to minimize error and
improve linearity.
19
Page 24
Dakota Ultrasonics
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with either the minimum or maximum sample or actual test
material. Be sure that the reading is stable and the repeatability indicator in
the top left corner of the display is fully lit and stable.
3) Press the key to enter the calibration edit screen displaying the current
measurement value.
4) Use the keys to scroll to the known thickness value.
Note: The longer the keys are pressed and held, the faster the value will
increment/decrement.
Note: Pressing the key prior to pressing the key will abort the cal
routine without saving any changes.
5) Once the known thickness value is being displayed, press the key to
display “1 of 2”, which sets the zero value and returns to the measurement
screen.
Note: The internal zero of the PZX-7 is now set. The procedure above can be
used to set the internal zero of the PZX-7 to use any reference standard as
the ‘probe zero’ standard if desired.
6) Repeat steps 2-4 on the second test point/location.
7) Press the key to display the calculated velocity edit screen.
Note: The calculated velocity can be edited, if needed, by pressing the
keys to scroll and edit the velocity value.
8) Press the key to set the calculated material velocity and return to the
measurement screen.
20
Page 25
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
Note: CHECK YOUR CALIBRATION! Place the transducer back on the
calibration points. The thickness readings should now match the known
thickness values with minimal error. If the thicknesses are not correct, repeat
the steps above.
Basic Material Type
If the material velocity is unknown, a sample thickness cannot be taken directly from
the material, but the general type of material is known, selecting a basic material type
from the common material (MATL) list in the PZX-7 would offer a reasonable
approximation of the thickness. There are 9 common materials and 2 user
programmable settings available. It’s important to note that these velocities will not
always be an exact representation of the material being tested. Use these values
only if a close approximation is acceptable. Follow the steps below to select a basic
material type:
Selecting a Basic Material Type
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
MATL feature is being displayed.
3) Press the key to edit the material setting. The edit icon will be
illuminated and flashing.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving any changes.
4) Use the keys to scroll through the material options.
21
Page 26
Dakota Ultrasonics
ALUMINUM
(2024)
STEEL (4340)
in/µs m/s
0.250 6350
0.233 5918
STAINLESS (303)
CAST IRON
PLEXIGLASS
PVC
POLYSTYRENE
POLYURETHANE
USER PROGRAMMABLE
0.223 5664
0.180 4572
0.106 2692
0.094 2388
0.092 2337
0.070 1778
5) When the desired MATL setting is displayed, press the key to set the
material velocity and return to the measurement screen.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving any changes.
6) If USR1 or USR2 were selected, the velocity edit screen will be displayed
and edit icon illuminated and flashing.
7) Use the keys to scroll to the desired material velocity.
Note: The longer these keys are held, the faster the velocity value is
incremented.
8) When the desired velocity setting is displayed, press the key to set the
material velocity and return to the measurement screen.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving any changes.
22
Page 27
CHAPTER SIX
THROUGH PAINT MEASUREMENT - MULTIMODE
6.1 Introduction
Through paint measurement is accomplished by measuring the time between two
repeat echoes from the back surface of the material. Since both of these back wall
echoes travel the same path through the paint or coating, the thickness of the coating
is subtracted out of the measurement so that only the actual material thickness is
measured. This avoids having to scrape or remove the coating from materials prior
to inspection. The primary purpose of thru paint measurement is to determine the
actual/nominal material thickness without error from the coating. Finally, this mode
will only work for typical epoxy based coatings.
6.2 Multi Mode Transducers
The multi echo measurement technique does have restrictions on the type of
transducers it can use successfully. The key requirement is that the transducers are
“high damped”, which refers to the duration of how long the transducer rings. In
order to improve the low end measurement range, being able to measure thin
materials, the cycles of ring must be limited so they don’t interfere with the internal
gating.
Since the PZX-7 is a basic easy to operate gauge without the adjustability you’d get
using an advanced A-Scan scope, echo-echo mode is subject to limitations
depending on type and thickness of the coating, as well as the type of transducer
used.
The procedure for activating the through paint (E-E) measurement mode is outlined
as follows:
Echo-Echo Multi Mode
Note: Be sure that a probe zero (if using a contact transducer) and “one point
calibration”, or a “two point calibration” has been performed prior to this
procedure.
23
Page 28
Dakota Ultrasonics
1) Press the key to toggle between the measurement modes, for a given
transducer type (contact or delay line), to echo-echo (E-E) mode.
Note: An icon will be illuminated in the top left portion of the display to indicate
the measurement mode the PZX-7 is currently using.
24
Page 29
CHAPTER SEVEN
VELOCITY GAUGE
7.1 Velocity Gauge (VX)
The PZX-7 includes a function to convert the unit into a dedicated velocity gauge.
With this feature enabled, the PZX-7 will display all measurements in terms of
velocity, inches per microsecond (IN /s) or meters per second (M /s), rather than
dimensional inches or millimeters. This is primarily useful for rudimentary “nodularity”
testing, as the velocity can be associated with density and used to determine the
hardness/strength of a given material. A casting manufacturer would typically use
this feature to control their processes and make sure the density/hardness is
sufficient for each part and batch within a specified tolerance.
Using this feature will require calibration on a “known” thickness that will remain
consistent at a specific location on a group of parts. The test will always be
performed at the same location for all parts in the group. The velocity will be
determined, and either accepted or rejected depending on the specified tolerances.
The procedure for enabling this feature is outlined below:
Velocity Gauge Option
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
VX feature is being displayed.
3) Press the key to edit the velocity gauge setting. The edit icon will be
illuminated and flashing.
4) Use the keys to toggle velocity on/off.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
25
Page 30
Dakota Ultrasonics
5) When the desired VX setting is displayed, press the key to set the
status and return to the measurement screen.
7.2 Calibration to a known thickness
In order to calibrate the PZX-7 a ‘known thickness’ on the material or part will be
used. The same location will be used for all the other parts in the group/batch to
determine the velocity.
The procedure is outlined as follows:
Calibration – Known Thickness
Note: Be sure that a probe zero has been performed prior to performing this
calibration procedure.
1) Physically measure an exact sample of the material, or a location directly on
the material to be measured, using a set of calipers or a digital micrometer.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the sample or actual test material. Be sure that the
reading is stable and the repeatability indicator, in the top left corner of the
display, is fully lit and stable.
3) Press the key to enter the calibration edit screen displaying the current
velocity IN /s (M /s)value. The edit icon will be illuminated and flashing.
4) Press the key again to edit the known thickness value. The edit icon
will be illuminated and flashing and the units will be IN or MM, indicating
thickness.
26
Page 31
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
5) Use the keys to scroll to the known thickness value.
Note: The longer the keys are pressed and held, the faster the value will
increment/decrement.
Note: Pressing the key prior to pressing the key will abort the cal
routine without saving any changes.
6) Once the known thickness value is being displayed, press the key to
return to the measurement screen and display the calculated material
velocity.
Note: The known thickness value that was used to calibrate will be displayed in
the top right corner of the display for confirmation.
7.3 Calibration to a known velocity
The velocity can also be directly edited and set to a target velocity value that was
previously determined from a reference standard at an earlier time.
The procedure for directly entering the velocity is outlined below:
Calibration – Known Velocity
Note: Be sure that a probe zero has been performed prior to performing this
calibration procedure.
Note: This procedure requires that the operator know the sound-velocity of the
material to be measured. A table of common materials and their soundvelocities can be found in Appendix A.
27
Page 32
Dakota Ultrasonics
1) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the sample or actual test material. Be sure that the
velocity measurement is stable and the repeatability indicator, in the top left
corner of the display, is fully lit and stable.
2) Press the key to enter the calibration edit screen displaying the current
velocity IN /s (M /s)value. The edit icon will be illuminated and flashing.
3) Use the keys to scroll to the known velocity value.
Note: Pressing the key prior to pressing the key will abort the cal
routine without saving any changes.
4) Once the known velocity value is being displayed, press the key to
display the calculated thickness based on known velocity.
5) Press the key to return to the measurement screen and begin making
measurements.
Note: The known velocity value that was entered will be displayed, and the
thickness value calculated will appear in the top right corner of the display
for confirmation.
28
Page 33
CHAPTER EIGHT
ADDITIONAL FEATURES
8.1 Gain
The gain, or amplification of the return echoes, can be adjusted in the PZX-7 to
accommodate a variety of materials and applications. The setting of the gain is
crucial in order to obtain valid readings during the measurement process. Too much
gain may result in erroneous measurements, detecting on noise rather than the
actual material back wall surface. Not enough gain may result in intermittent
detection. If you turn it up too much, you can’t hear the music clearly. If it’s turned
down too much, you can’t hear it at all.
The PZX-7 has the option of using and selecting three different transducer types;
contact, delay line, and plastics. The contact transducers, in pulse-echo (P-E) and
echo-echo (E-E), as well as delay line style transducers in interface-echo (I-E) and
echo-echo (E-E) use automatic gain control (AGC) and fixed time dependent gain
(TDG). They have three gain settings (LOW, MED, HIGH) with a 3dB cut/boost to
the dynamic range of the AGC. Plastics (PLAS) mode, ‘ thin setting only’, uses a
manual gain with five settings (VLOW, LOW, MED, HIGH, VHI) in 3dB increments,
while the delay line style transducers use automatic gain control (AGC), and have 3
gain settings (LOW, MED, HIGH), similar to the contact and delay line style probes.
The PZX-7 has been optimized for the MED gain setting for all common applications.
It should be operated in this mode as standard. However, some applications may
require the lower or higher gain settings. When? The low settings may be necessary
for noisy or granular cast materials. How do I know when to lower the gain? If the
reading becomes sporadic and won’t settle down or resolve on a thickness value
because the material is either very noisy or granular. Setting the gain to a lower less
sensitive level, would potentially offer improved stability. How do I know when to
increase the gain? When a material is difficult to penetrate or pass sound through.
This could be due to the material type, overall thickness, the transducer diameter and
frequency, or a combination of all the above. Turning the gain up for additional
output could improve the ability to obtain a successful measurement. In any case,
the selectable gain settings offer improved versatility to resolve and overcome
potential application issues.
Note: When the pulse-echo (P-E), echo-echo (E-E),or plastics (PLAS) “thick plastic
setting option” measurement modes are selected for either contact, delay line, or
plastics modes, The automatic gain control (AGC) is enabled. The dynamic range of
the AGC can be adjusted with the following options (LOW, MED, HIGH), with MED
still being the optimized standard setting as above.
The procedure for editing the gain is outlined as follows:
29
Page 34
Dakota Ultrasonics
GAIN
1) Press the key to edit the gain setting. The edit icon will be illuminated
and flashing.
2) Use the keys to scroll through the gain settings in PLAS
setting” (LOW, MED, HIGH) until the desired setting is being displayed.
3) Press the key to set the gain and return to the measurement screen.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
8.2 High Speed Scan
The High Speed Scan feature of the PZX-7 increases the overall repetition rate to a
maximum of 140Hz with a high speed screen refresh rate of 25 times a second. This
allows for making scanned passes over an arbitrary length of the test material, while
still maintaining a reasonable representation of thickness over the area or region
scanned.
The procedure to use the scan feature is outlined below:
High Speed Scan
30
Page 35
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
1) Press the key to toggle SCAN on/off. The display will briefly display the
status and return to the measurement screen.
8.3 Alarm
The Alarm feature of the PZX-7 provides a method of setting tolerances, or limits, for
a particular application requirement. This feature may be used for a variety of
applications to verify the material thickness is within the manufacturer specifications.
The settings available are ON/OFF/BEEP, where beep enables the audible beeper.
Both the on and beep settings will illuminate the led alarm lights above the keys on
the keypad. There are two limit values HI/LO, that can be set according to specified
tolerances.
The procedure to use the alarm feature is outlined below:
ALARM
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
ALRM feature is being displayed.
3) Press the key to edit the alarm status. The edit icon will be illuminated
and flashing.
4) Use the keys to toggle alarm on/off/beep.
5) When the desired ALRM status is displayed, press the key to set the
status and edit the LO limit option.
6) Use the keys to scroll the LO limit value to the target value.
31
Page 36
Dakota Ultrasonics
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes to the LO limit value.
7) When the target LO limit is displayed, press the key to set the value
and advance to setting HI limit option.
8) Use the keys to scroll the HI limit value to the target value.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes to the HI limit value.
9) When the target HI limit is displayed, press the key to set the value
and return to the measurement screen.
8.4 Differential
The Differential Mode of the PZX-7 provides the user with the ability to set a
nominal value, according to what the expected thickness should be, and measure the
+/- difference from the nominal value entered. This feature is typically used in QA
incoming inspections on pipes, plate stock, coils, etc.
The steps below outline how to enable and enter the nominal value to use this
feature:
Differential
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
DIFF feature is being displayed.
32
Page 37
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
3) Press the key to edit the differential status. The edit icon will be
illuminated and flashing.
4) Use the keys to toggle differential on/off.
5) When the desired DIFF setting is displayed, press the key to set the
status and edit the NOMINAL value.
6) Use the keys to scroll the NOMINAL value to the target
value.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes to the NOMINAL value.
7) When the target value is displayed, press the key to set the value and
return to the measurement screen.
8.5 Units
The PZX-7 will operate in both English (inches) or Metric (millimeters) units. In either
option the resolution (RES) can be set to LOW (.001in/.01mm), or HIGH
(.0001in/.001mm).
The procedure to select the units is outlined as below:
Units
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
UNIT feature is being displayed.
33
Page 38
Dakota Ultrasonics
3) Press the key to edit the units setting. The edit icon will be illuminated
and flashing.
4) Use the keys to toggle English or Metric units.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
5) When the desired UNIT setting is displayed, press the key to set the
units and edit the RES (resolution) option.
6) Use the keys to scroll the RES (LOW, HIGH) options.
7) When the desired RES setting is displayed, press the key to set the
resolution and return to the measurement screen.
8) Note:Pressing the key prior to pressing the key will abort to
the measurement screen without saving changes.
8.6 Lite
The PZX-7 uses a custom glass segmented display that is equipped with a backlight
for use in low light conditions. The options are on/off/auto, where the auto setting
only lights the display when the gauge is coupled to the material and receiving a
measurement.
The steps below outline how to toggle the options:
Backlight
1) Press the key to access the menu items/features.
34
Page 39
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
2) Use the keys to scroll through the items/features until the
LITE feature is being displayed.
3) Press the key to edit the light setting. The edit icon will be illuminated
and flashing.
4) Use the keys to toggle status on/off/auto.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
5) When the desired LITE setting is displayed, press the key to set the
status and edit the BRT (brightness) option.
6) Use the keys to scroll through the BRT (LO, MED, HI)
options.
7) When the desired BRT setting is displayed, press the key to set the
brightness and return to the measurement screen.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
8.7 Beep
The PZX-7 also has a feature to use the internal beeper, most commonly used with
the alarm feature, for the key strokes on the keypad. When enabled, pressing any of
the keys on the keypad will sound the beeper.
The procedure to enable the keyboard beeper feature is outlined below:
Beeper
35
Page 40
Dakota Ultrasonics
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
BEEPER feature is being displayed.
3) Press the key to edit the beeper setting. The edit icon will be
illuminated and flashing.
4) Use the keys to toggle the beeper on/off.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
5) When the desired BEEP setting is displayed, press the key to set the
status and return to the measurement screen.
8.8 Probe Type
The PZX-7 DL can use three different probe/transducer styles for different application
requirements. The styles are; CONTACT, DELAY LINE, and plastics PLAS. The
corresponding option ‘must’ be selected to match the type of transducer connected to
the PZX-7 DL in order for the gauge to operate correctly.
The procedure to select the probe type is outlined below:
Probe
1) Press the key to access the menu items/features.
36
Page 41
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
2) Use the keys to scroll through the items/features until the
PROB feature is being displayed.
3) Press the key to edit the probe type. The edit icon will be illuminated
and flashing.
4) Use the keys to select the contact/delay line/plastics
option.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
5) When the desired PROB type is displayed, press the key to set the
type and return to the measurement screen.
Note: If the ‘contact’ option was selected, the diameter and frequency of the
contact style transducer must be selected in order to operate correctly.
6) Use the keys to select the diameter & frequency (.12 5,
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
7) When the desired PROB diameter is displayed, press the key to set
the diameter and return to the measurement screen.
8.9 Velocity (VX)
When the velocity setting (VX) is enabled, the PZX-7 will display will the material
velocity as the primary measurement quantity instead of dimensional thickness. The
feature is generally used for basic “nodularity” testing, as velocity is a key part of
density for determining hardness. An example might be casting manufacturers
where the density/hardness will determine the strength of the material.
37
Page 42
Dakota Ultrasonics
When this feature is enabled, the PZX-7 is operating in reverse to the standard option
of the gauge. Only the ‘one point’ calibration can be used with this feature active,
and a manual or auto zero is still required. The PZX-7 can be calibrated by entering
the known velocity or entering the know thickness of the material at a given position
on the test material. Refer to the ‘making measurements’ section on page 25 for a
complete explanation of the zero and one point calibration procedure.
The procedure to enable the velocity feature is outlined below:
Velocity Gauge
1) Press the key to access the menu items/features.
2) Use the keys to scroll through the items/features until the
VX feature is being displayed.
3) Press the key to edit the setting. The edit icon will be illuminated and
flashing.
4) Use the keys to select the on/off option.
Note: Pressing the key prior to pressing the key will abort to the
measurement screen without saving changes.
5) When the desired VX setting is displayed, press the key to set the
status and return to the measurement screen.
8.10 Lock
The lock feature was built into the PZX-7 for the purpose of locking the operators out
of editing any of the gauge settings, for purposes of consistency between operators.
When the lock feature is enabled, the gauge calibration functionality cannot be
38
Page 43
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
altered, as well as any of the individual features in the gauge. The only keys that are
always unlocked are the power and probe zero keys, as these must remain unlocked
for measurement functionality.
The procedure to enable/disable the lock feature is outlined below:
Lock
1) With the PZX-7 powered off, press and hold down the key while
powering the PZX-7 on . The lock icon will be illuminated on the
display.
2) To unlock the PZX-7 repeat step one, but hold down the key while
powering the PZX-7 on .
8.11 Factory Defaults
The PZX-7 can be reset to factory defaults at any time to restore the original gauge
settings. This should only be used if the gauge is not functioning properly, or
perhaps multiple features have been enabled and a clean start is needed.
The procedure to reset the gauge is outlined below:
Factory Reset
1) With the PZX-7 powered off, press and hold down the and keys
while powering the PZX-7 on .
39
Page 44
Dakota Ultrasonics
Note: Once the measurement screen is displayed the and can be
released.
40
Page 45
CHAPTER NINE
CONNECTIVITY & POWER
9.1 Connectivity
The PZX-7 is equipped with a USB-C connector and offers serial over USB using a
CDC (Communications Data Class) and software wedge to transfer data into
common text editors and spreadsheet software programs as needed. The PZX-7
does not have any internal memory to store data, but is used to connect to external
storage devices such as SPC data collectors and PC based software programs.
We also offer a standard serial hardware option, as well as a wireless Bluetooth
module that can be installed on the PC board of the PZX-7. Refer to our price list for
additional information on communication module options and cables.
9.2 Installing Serial Driver (CDC)
When the PZX-7 is powered on and connected to the PC, and no CDC driver was
previously installed, the following message will be displayed:
The driver file that needs to be installed can be downloaded directly from our website
at www.dakotaultrasonics.comDUsoftware/CDC/CDC_Ultrasonics.inf
found in the examples folder created on your PC by installing our DakView PC
software, and located in the USB directory. The file name is CDC_Ultrasonics.inf,
and can be downloaded and saved anywhere on your local PC.
To install the driver, open Control Panel, followed by selecting Device Manager:
41
, or can be
Page 46
Dakota Ultrasonics
Right click on the CDC device as illustrated above, and select “update driver”. The
following window will appear:
Select “browse my computer” and navigate to the directory where the driver was
saved on your PC and install the driver. When the installation is completed, the
following window will be displayed:
42
Page 47
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
Open device manager again to reference the communications port assigned to the
CDC device, it will be displayed to the right of the CDC device. Now that the driver
has been successfully installed, a ‘keyboard wedge’ with serial communications
capability can be used to connect to the PZX-7 and will capture the measurements to
your PC when the ENTER key is pressed:
9.3 Line Power
The PZX-7 can be powered using the standard USB-C to USB-A data cable (N-003-
0330), by connecting directly to a USB port on your computer, or using a standard
cell phone power adapter directly to an outlet. This is a convenient way to power the
gauge for specific bench top applications in a factory line environment.
When measuring a piece of pipe to determine the thickness of the pipe wall,
orientation of the transducers is important. The transducer should be oriented so that
the gap (sound barrier) in the wear face is perpendicular (at a right angle) to the
length (long axis) of the tubing, allowing both sides of the transducer to make the
same amount of contact. The transducer orientation can either be parallel or
perpendicular for large diameter piping, as it’s much easier to ensure both sides are
making similar contact.
Measuring hot surfaces
The velocity of sound through a substance is dependent on its temperature. As
materials heat up, the velocity of sound through them decreases. In most
applications with surface temperatures less than about 200F (100C), no special
procedures must be observed. At temperatures above this point, the change in
sound velocity of the material being measured starts to have a noticeable effect upon
ultrasonic measurement.
At such elevated temperatures, it is recommended that the user perform calibration
on a sample piece of known thickness, which is at or near the temperature of the
material to be measured. This will allow the PZX-7 to correctly calculate the velocity
of sound through the hot material.
Expansion and contraction of the transducer based on temperature, and a varying
temperature gradient, will also affect the measurement in a pulse-echo (P-E)
measurement mode. It is recommended that a “transducer zero” be performed often
to account for the delay line changing length and adversely affecting the accuracy of
the measurements.
When performing measurements on hot surfaces, it may also be necessary to use a
specially constructed high-temperature transducer. These transducers are built using
materials which can withstand high temperatures. Even so, it is recommended that
46
Page 51
PZX-7 Ultrasonic Multi-Mode Thickness Gauge
the probe be left in contact with the surface for as short a time as needed
(intermittent contact) to acquire a stable measurement.
Measuring laminated materials
Laminated materials are unique in that their density (and therefore sound-velocity)
may vary considerably from one piece to another. Some laminated materials may
even exhibit noticeable changes in sound-velocity across a single surface. The only
way to reliably measure such materials is by performing a calibration procedure on a
sample piece of known thickness. Ideally, this sample material should be a part of
the same piece being measured, or at least from the same lamination batch. By
calibrating to each test piece individually, the effects of variation of sound-velocity will
be minimized. If the variation is relatively close, averaging the sound velocities to
minimize error is another option.
An additional important consideration when measuring laminates is that many
included air gaps or pockets which will cause an early reflection of the ultrasound
beam. This effect will be noticed as a sudden decrease in thickness in an otherwise
regular surface. While this may impede accurate measurement of total material
thickness, it does provide the user with positive indication of air gaps in the laminate.
Measuring through paint & coatings
Measuring through paints and coatings are also unique, in that the velocity of the
paint/coating will be significantly different from the actual material being measured. A
perfect example of this would be a mild steel pipe with .025” of coating on the
surface. Where the velocity of the steel pipe is .2330 in/sec, and the velocity of the
paint is .0850 in/sec. If the user is calibrated for mild steel pipe and measures
through both materials, the actual coating thickness will appear to be approximately 3
times thicker than it actually is, as a result of the differences in velocity.
47
Page 52
WARRANTY INFORMATION
Warranty Statement
Dakota Ultrasonics warrants the PZX-7 against defects in materials and
workmanship for a period of two years from receipt by the end user. Additionally,
Dakota Ultrasonics warrants transducers and accessories against such defects for a
period of 90 days from receipt by the end user. If Dakota Ultrasonics receives notice
of such defects during the warranty period, Dakota Ultrasonics will either, at its
option, repair or replace products that prove to be defective.
Should Dakota Ultrasonics be unable to repair or replace the product within a
reasonable amount of time, the customer's alternative exclusive remedy shall be
refund of the purchase price upon return of the product.
Exclusions
The above warranty shall not apply to defects resulting from: improper or
inadequate maintenance by the customer; unauthorized modification or misuse; or
operation outside the environmental specifications for the product.
Dakota Ultrasonics makes no other warranty, either express or implied, with
respect to this product. Dakota Ultrasonics specifically disclaims any implied
warranties of merchantability or fitness for a particular purpose. Some states or
provinces do not allow limitations on the duration of an implied warranty, so the
above limitation or exclusion may not apply to you. However, any implied warranty of
merchantability or fitness is limited to the five-year duration of this written warranty.
This warranty gives you specific legal rights, and you may also have other rights
which may vary from state to state or province to province.
Obtaining Service During Warranty Period
If your hardware should fail during the warranty period, contact Dakota
Ultrasonics and arrange for servicing of the product. Retain proof of purchase in
order to obtain warranty service.
For products that require servicing, Dakota Ultrasonics may use one of the
following methods:
- Repair the product
- Replace the product with a re-manufactured unit
- Replace the product with a product of equal or greater performance
- Refund the purchase price.
After the Warranty Period
If your hardware should fail after the warranty period, contact Dakota Ultrasonics
for details of the services available, and to arrange for non-warranty service.
48
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.