CU Medical Systems i-PAD CU-SP1 AUTO Instructions For Use Manual

Page 1
1
Instructions for Use
i-PAD CU-SP1 AUTO
The information in these Instructions for Use applies to the i-PAD CU-SP1 AUTO. This
information is subject to change. Please contact CU Medical Systems, Inc. or its authorized
representatives for information on revisions.
Revision History
Edition1
Publication Date: March 2014
Document No.: SPA-OPM-E-01
Published by: CU Medical Systems, Inc.
Printed in the Korea
Copyright
© 2014 CU Medical Systems, Inc.
No part of these Instructions for Use may be reproduced without the permission of CU
Medical Systems, Inc.
Medical Device Directive
The i-PAD CU-SP1 AUTO complies with the requirements of the Medical Device Directive
2007/47/EC and its revisions.
Important:
Quick defibrillation is needed if sudden cardiac arrest occurs. Since the chance of success is
reduced by 7% to 10% for every minute that defibrillation is delayed, defibrillation must be
performed promptly.
Page 2
2
The i-PAD CU-SP1 AUTO is manufactured by:
CU Medical Systems, Inc.
130-1, Dongwhagondan-ro, Munmak-eup,
Wonju-si, Gangwon-do, Korea
220-801
Authorized EU Representative
Medical Device Safety Service
Schiffgraben 41, 30175 Hannover, Germany
Contact Us
Product and Order Inquiries
Overseas Sales Team
CU Medical Systems, Inc. 5F, Cheonggye Bldg, 221, Anyangpangyo-ro, Uiwang-si, Gyeonggi-do, Korea Tel : +82 31 421 9700 / Fax: +82 31 421 9911 E-mail addresses : sales@cu911.com
German Branch Office
CU Medical Germany GmbH, Ernst-Augustin-Str, 5-11, 12489 Berlin, Germany Tel: +49 30 6781 7804 Fax: +49 30 6782 0901
Service and Technical Support
Customer Service Team
CU Medical Systems, Inc. 4F, Cheonggye Bldg, 221, Anyangpangyo-ro, Uiwang-si, Gyeonggi-do, Korea Tel : +82 31 421 9700 / Fax: +82 31 421 9911 E-mail addresses : service@cu911.com
Page 3
3
Page 4
4
Table of contents
INTRODUCTION ........................................................................................................................... 6
OVERVIEW ................................................................................................................................... 7
1. INTRODUCTION .................................................................................................................. 8
1.1 DEVICE DESCRIPTION ........................................................................................................... 8
1.2 INDICATED USE .................................................................................................................... 8
1.3 LOCAL PROTOCOL ................................................................................................................ 9
1.4 ADDITIONAL INFORMATION .................................................................................................... 9
2. DEVICE FEATURES .......................................................................................................... 10
3. PREPARATION FOR USE ................................................................................................. 13
3.1 STANDARD PACKAGE CONTENTS ......................................................................................... 13
3.2 SETTING UP THE I-PAD CU-SP1 AUTO .............................................................................. 14
4. HOW TO USE THE I-PAD CU-SP1 AUTO......................................................................... 16
4.1 CHAIN OF SURVIVAL ........................................................................................................... 16
4.2 PREPARATION FOR DEFIBRILLATION ..................................................................................... 17
Step 1: Place pads on the patient.......................................................................................... 21
Step 2: Automatic shock delivery if needed. ......................................................................... 22
Step 3: Perform CPR. ............................................................................................................ 24
4.3 DEFIBRILLATION PROCEDURES IN PEDIATRIC MODE ............................................................. 26
5. AFTER USING THE I-PAD CU-SP1 AUTO ....................................................................... 27
5.1 MAINTENANCE AFTER EACH USE ........................................................................................ 27
5.2 SAVING AND TRANSFERRING TREATMENT DATA .................................................................... 28
5.2.1 Device Usage .......................................................................................................... 28
5.2.2 Transferring Treatment Data ................................................................................... 28
5.3 DEVICE SETTING ................................................................................................................ 31
5.3.1 CPR Guide Setting .................................................................................................. 31
5.3.2 Setting the CPR Guide ............................................................................................ 31
6. MAINTENANCE ................................................................................................................. 34
6.1 DEVICE STORAGE .............................................................................................................. 34
6.2 MAINTENANCE ................................................................................................................... 35
6.2.1 Device Inspection .................................................................................................... 35
Page 5
5
6.2.2 Replacing Supplies ................................................................................................. 35
Disposable Battery Pack ................................................................................................................. 35
Replacing the Pads ......................................................................................................................... 38
6.2.3 Cleaning the i-PAD CU-SP1 AUTO ......................................................................... 39
7. DISPOSAL ......................................................................................................................... 39
8. TROUBLESHOOTING ....................................................................................................... 40
8.1 SELF-TESTS ...................................................................................................................... 40
8.2 DEVICE STATUS ................................................................................................................. 42
8.3 TROUBLESHOOTING ........................................................................................................... 43
8.3.1 Troubleshooting While the Device is Operating ...................................................... 43
8.3.2 Troubleshooting While the Device is not Operating ................................................ 44
9. DEVICE SERVICE .............................................................................................................. 45
APPENDIX .................................................................................................................................. 47
A . RESCUE PROTOCOL ........................................................................................................... 47
B . PARTS AND ACCESSORIES ............................................................................................... 50
B.1 STANDARD ACCESSORIES ..................................................................................................... 50
B.2 OPTIONAL ACCESSORIES ...................................................................................................... 50
C . DESCRIPTION OF SYMBOLS ............................................................................................. 51
C.1 I-PAD CU-SP1 AUTO DEFIBRILLATOR .................................................................................. 51
C.1 I-PAD CU-SP1 AUTO DEFIBRILLATOR (CONTINUE) . 오류! 책갈피가 정의되어 있지 않습니다.
C.2 I-PAD CU-SP1 AUTO PACKAGING ....................................................................................... 53
C.3 ACCESSORIES ...................................................................................................................... 54
C.3.1 Disposable Battery Pack (CUSA1103BB, CUSA1103BS) ........................................... 54
C.3.2 Pads (CUA1007S, CUA1102S) .................................................................................... 55
D . GLOSSARY ........................................................................................................................... 56
E . DEVICE SPECIFICATIONS ................................................................................................... 61
F . ELECTROMAGNETIC COMPATIBILITY .............................................................................. 69
Page 6
6
Introduction
These Instructions for Use contain information necessary for the correct use of this device.
Please contact us regarding any questions or issues on the use of the device arising from
information found in these Instructions for Use [Chapter 9: Device Service].
The company or its authorized distributor is not responsible for any injury incurred by the user or
patient due to any apparent negligence or improper use by the user.
Hereinafter,
"device" refers to [CU-SP1 AUTO]
"We" or “Us" refers to CU Medical Systems, Inc.
"Pads" refers to defibrillation electrode pads,
"Battery Pack" refers to a disposable battery pack.
These Instructions for Use emphasizes the safety procedures and precautions for the device
use by using the terms below. Please acquaint yourself with the warnings, cautions and
references stated in these Instructions for Use in order to safely use the device.
WARNING
Conditions, hazards, or unsafe practices that can result in serious personal injury or loss of
life.
CAUTION
Conditions, hazards, or unsafe practices that can result in minor or moderate personal injury,
damage to the device, or loss of treatment data stored in the device, particularly if
precautionary steps are not taken.
Used to denote items that are important during installation, operation, or maintenance of the
device.
Page 7
7
Overview
Thank you for purchasing the i-PAD CU-SP1 AUTO. This device can be effectively and safely
used for a long period if you familiarize yourself with the instructions, warnings, precautions, and
notices contained in these Instructions for Use prior to its use.
WARNING
A defibrillator discharges electric shock with high voltage and current. You must be well-
acquainted with the instructions, warnings, and precautions contained in these Instructions for
Use.
You must follow the instructions, warnings, cautions, and notices in these Instructions for Use
when using this device.
The manufacturer will not be responsible for any problems involving the device that are
caused by the user's negligence.
This device shall be serviced only by the manufacturer or its authorized service centers.
If the Device is intended to be connected to equipment other than those stated in these
Instructions for Use, contact the manufacturer.
If this Device does not operate properly, contact the manufacturer or its authorized service
center.
Page 8
8
1. Introduction
1.1 Device Description
CU-SP1 AUTO is an easy-to-use Fully Automated External Defibrillator (AED) that is small, light,
and portable, and uses a battery.
The AED automatically reads the patient's electrocardiogram (ECG) and determines if a cardiac
arrest that requires defibrillation has occurred, so that both medical professionals and the
general public can easily operate it. Cardiac arrest can occur anytime to anyone at any place
and may threaten the patient's life if the appropriate CPR and/or electric shock with a
defibrillator are not applied within a few minutes.
The i-PAD CU-SP1 AUTO is a Fully automated external defibrillator (AED). If connected to a
patient, the i-PAD CU-SP1 AUTO automatically acquires and analyzes the electrocardiogram
(ECG) of the patient for the presence of Ventricular Fibrillation or Ventricular Tachycardia (also
known as shockable rhythms). If a shockable rhythm is detected, the device automatically
charges itself. After charging is completed, the equipment delivers an electric shock
automatically.
The i-PAD CU-SP1 AUTO is easy to use. It guides the you throughout a rescue operation using
voice prompts and indicators (LED and graphical indicators).
The i-PAD CU-SP1 AUTO is small, light, highly portable, and battery powered. It is highly
suitable for use in public, out-of-hospital settings.
1.2 Indicated Use
The i-PAD CU-SP1 AUTO is indicated for use on patients that are exhibiting the symptoms of
sudden cardiac arrest (SCA) with all of the following signs:
a) No movement and no response when shaken
b) No normal breathing
Do not use the i-PAD CU-SP1 AUTO on patients who show either of the following
signs:
a) Movement or response when shaken
b) Presence of normal breathing
Page 9
9
Intended Users
The i-PAD CU-SP1 AUTO is intended for use in or out of the hospital by emergency care
personnel or healthcare professionals or laypersons. The manufacturer recommends that users
train on the use of the device.
1.3 Local Protocol
Please contact your local health authority for information on the requirements of ownership and
usage of defibrillators.
1.4 Additional Information
Please contact CU Medical Systems, Inc. or its local distributors for any additional information
on the i-PAD CU-SP1 AUTO.
Page 10
10
2. Device Features
Page 11
11
Power Button
Turns the device on or off. (When the device is on, a green
LED is lit)
i-Button
Reports device usage (the total hours of the last usage
and number of shocks)
checks the S/W version
downloads events and ECG data via an IrDA and SD
Card
sets the CPR mode (the number of compressions,
breaths and cycles; compression rate per minute; pausing
time; detailed guide on/off)
and checks for errors
Status LCD
Displays the current status of the device, battery and pads.
Adult/Pediatric
Selection Switch
Selects Adult/Pediatric modes.
Adult/Pediatric
Selection Switch
Cover
Covers the Adult/Pediatric Selection Switch to prevent
accidental switching.
Defibrillator Pads
Connector
Connects with the connectors of the pads.
Pads Connector
Status Indicator
Indicates the connection status of the defibrillator pads
connector.
Pads Position
Indicators
Indicates the pads position on the patient.
Do-Not-Touch-Patient
Indicator
Warns when not to touch the patient.
Page 12
12
CPR Detection
Indicator
Indicates performance of CPR on the patient.
(The indicator is lit if CPR is performed, and flashes if CPR
is not performed)
Battery Pack
The disposable power source of the device.
IrDA Port
Transmits and receives treatment data between the device
and a personal computer.
SD Card (External
Memory) Port
Port for copying device records to a SD card.
Pads Storage
Compartment
Stores pads.
PROHIBITION
Do not open SHIELD RUBBER that covers IrDA port and SD card port during
defibrillation therapy
Page 13
13
3. Preparation for Use
3.1 Standard Package Contents
The following are the standard package contents of this device
CU-SP1 AUTO Fully automated External Defibrillator Instructions for Use
1 Pack of Adult Pads (Disposable)
Please contact the manufacturer for replacement supplies (refer to [Appendix B: Parts and
Accessories] of these Instructions for Use).
1 Battery Pack (Disposable)
Page 14
14
WARNING
Only parts and accessories recommended and approved by CU Medical Systems, Inc. must be
used with the i-PAD CU-SP1 AUTO. Using unapproved parts and accessories may compromise
the safety and effectiveness of the i-PAD CU-SP1 AUTO.
Extra battery packs and pads are recommended.
3.2 Setting up the i-PAD CU-SP1 AUTO
Do the following to set up the i-PAD CU-SP1 AUTO
Open the package and verify that it contains all the items listed in the packing list.
Familiarize yourself with the device features by referring to [Chapter 2: Device Features] of
these Instructions for Use.
Insert the battery pack into the battery compartment on the device as shown in the figure
below.
As the battery pack is inserted, the device starts a self-test. If the device status is normal,
is shown on the Status LCD. If , or is displayed on the Status LCD after the self-
test, please refer to [Chapter 8: Troubleshooting] of these Instructions for Use.
If you have a carrying case, please safely store the Device in the carrying case. If you want
to purchase the carrying case, please contact us by referring to [Appendix A: Accessories] of
these Instructions for Use.
Storage and maintenance considerations:
Page 15
15
Refer to [Section 6.1: Device Storage] for proper device storage instructions.
When the device is in storage, check the Status LCD periodically to ensure that the device
is in good condition.
Store the CU-SP1 AUTO in accordance with your local emergency first aid protocol.
Store the device in an easy-to-access location where its Status LCD can be checked
periodically and its technical alarms can be easily heard (e.g. alarm on low battery or other
device problems).
It is also recommended to place an emergency use telephone near the device’s storage
area so that emergency medical services can be easily called during emergencies.
Store the accessories along with the device in the device’s carrying case for easy and quick
access.
WARNING
Electromagnetic interference may affect the performance of the device. While the device
is in use, it should be kept away from devices that cause electromagnetic interference.
Devices that may cause such interference include motors, X-ray equipment, radio transmitters,
and cell phones. Refer to [Appendix E: Electromagnetic Compatibility] of these Instructions for
Use for more information.
The use of accessories or cables other than those referred to in these Instructions for Use
may increase electromagnetic radiation from the device or reduce the device’s electromagnetic
immunity. Only accessories and cables that are authorized by the manufacturer should be
used with the i-PAD CU-SP1 AUTO.
There is a possibility of explosion or fire if the product is used in the presence of flammable
agent or in an OXYGEN enriched atmosphere due to the arc discharge caused by electrical
shock.
Page 16
16
4. How to Use the i-PAD CU-SP1 AUTO
4.1 Chain of Survival
If you think that you are witnessing someone go down in sudden cardiac arrest, perform the
chain of actions recommended by the American Heart Association (AHA) in its Chain of Survival
emergency response to sudden cardiac arrest.
1 2 3 4 5
1. Immediate recognition and activation of the emergency response system.
Check for a response by tapping the victim on the shoulder and shouting at the victim.
Activate the community emergency response system (e.g. call 911 or the equivalent service
in your locality)
2. Early CPR
Perform CPR.
3. Early defibrillation
Use this device (i-PAD CU-SP1 AUTO).
Using this device can be summarized in 3 steps:
After pressing the Power Button,
Step 1: Place pads on the patient.
Step 2: Automatic shock delivery if needed.
Step 3: Perform CPR.
4. Effective advanced life support – Perform advanced care in order to restore spontaneous
circulation.
5. Integrated post-cardiac arrest care – Transfer the patient to a medical institution or a
specialized facility
If finding and/or operating the defibrillator takes time, monitor the patient's status until the
defibrillator is available, perform CPR if necessary.
Page 17
17
4.2 Preparation for Defibrillation
Set the Adult/Pediatric Selection Switch to match the victim.
Adult victim
• Open the switch cover
• Set the switch to adult defibrillation mode as shown in the following picture
Child victim (victim is under 25kg or 8 years old)
If the pediatric pads are attached, the i-PAD CU-SP1 AUTO automatically adjusts its
defibrillation energy output for pediatric defibrillation regardless of the position of the
Adult/Pediatric Selection Switch (i.e. the output will be pediatric even if the selection switch is
set to adult)
If there are no pediatric pads for the pediatric patient, adult pads may be used. Ensure that the
Adult/Pediatric Selection Switch is set to Pediatric Mode. If the switch has not been set yet,
move it to Pediatric Mode as shown in the figures below
• Open the switch cover
Page 18
18
• Set the switch to pediatric defibrillation mode as shown in the following picture
If a young victim is over 25kg or 8 years old, or if you are not sure of the exact weight or
age:
DO NOT DELAY TREATMENT
• Set the Adult/Pediatric Selection Switch to Adult mode.
• Use the adult pads.
WARNING
Never perform defibrillation in pediatric mode to a patient who is either heavier than 25 kg or
older than 8 years old. Ensure the slide key for Adult/Pediatric Mode is as shown on the
bottom.
You can switch the adult/pediatric selection switch before or after turning on the i-PAD CU-
SP1 AUTO. However, the defibrillation mode should be changed before placing the pads on
the patient. Once the pads are in place, you cannot change the defibrillation mode anymore.
When the mode is correctly selected, the defibrillation energy is set to an adult value (150 J)
or pediatric value (50 J).
Turn the device on by pressing the Power Button.
When the power turns ON the following occurs in sequence:
the beeper will beep for 1 seconds
Voice instruction: “Call emergency Medical services, now”
Page 19
19
Remove clothes from patient's chest.
CAUTION
Time is essential for the cardiac arrest patient. Tear or cut clothes if removing them will take
time.
Dry the patient's skin such that pads can adhere well on the chest. Shave hair on the chest if
necessary.
Remove the pads package from the Pads Storage Compartment at the bottom of the
device.
Open the pads package.
Page 20
20
Take pads out of the pads package.
Refer to the pictures on both pads.
Adult Pads
Pediatric Pads
CAUTION
The adhesive material on the pads starts to dry out as soon as the package is opened. Use
immediately after opening. Refer to [Section 6.2: Maintenance] of these Instructions for Use
for procedures on how to check the expiration date of the pads and pads maintenance.
Page 21
21
Defibrillation in Adult Mode
Step 1: Place pads on the patient.
Remove pad 1 from the single liner and stick the pad to the patient’s upper chest as shown
below.
Remove pad 2 from the single liner, and stick the pad to the patient’s side torso as shown
below.
If the device detects the connection with the patient after placing the pads, follow the voice
instruction of the device.
Defibrillation can be done even if the pads are reversed. If the locations of pads are switched,
follow the next voice instruction without changing the directions of pads. It is more important to
begin defibrillation as soon as possible.
In the event the pad is not adhering well, check if the adhesive side of the pads is dry. Each
pad has an adhesive gel. If the gel does not adhere well, replace it with a new pads.
WARNING
Ensure the patient is not on a wet surface when performing defibrillation. If the patient's skin is
wet, dry the skin first prior to using the device.
Keep the pads well clear of other electrodes or metal parts in contact with the patient.
Page 22
22
Step 2: Automatic shock delivery if needed.
The device acquires and analyzes the patient's ECG immediately after being connected. The
device will instruct you not to touch the patient by flashing the Do-Not-Touch-Patient Indicator
and by issuing the voice prompt: “Do not touch the patient, analyzing heart rhythm”. After
analyzing the ECG, the device will determine whether or not the patient needs defibrillation.
WARNING
Do not move or touch the patient during ECG analysis.
If the patient needs defibrillation, the device will do the following:
The device announces that a defibrillation shock is needed, and instructs you to keep away
from the patient.
CAUTION
While the device is charging after a shockable rhythm is detected, the ECG of the patient is
continuously acquired and analyzed. The device disarms itself if the ECG rhythm changes to
a non-shockable rhythm before shock delivery.
When it is charged, the device activates the following indicators in sequence:
The beeper will beep continuously.
The voice guide saying "Stand by, for automatic shock delivery. Warning: No one should be
touching the patient." will be given.
After the voice guide saying "Shock will be delivered in", the device delivers a defibrillating
shock to the patient. If defibrillation is properly done, the device reports that an electric shock
has been delivered.
After shock delivery, the device indicates that you may touch the patient, and the CPR Mode
Indicator is lit. Then, the voice instruction for CPR starts.
Page 23
23
If the patient does not need defibrillation, the device will do the following in sequence:
• the device announces that the patient does not need a defibrillating shock and that you may
touch the patient.
• the CPR Mode Indicator is lit.
• voice instruction for CPR starts.
WARNING
Do not touch (you or anybody else) the patient during shock delivery.
Do not open SHIELD RUBBER that covers IrDA port and SD card port during defibrillation therapy.
CAUTION
Before defibrillation, make sure that there is no contact which may provide unwanted
pathways for the defibrillating current; patient’s body (such as exposed skin or head or limbs),
conductive fluids (such as gel, blood, or saline), metal objects (such as bed frame or
stretcher)
While analyzing ECG, keep the patient still and minimize movements around the patient. Do
not touch the patient and pads while the Do-Not-Touch-Patient Indicator is on. Electrical noise
(interference) may delay the ECG analysis.
During defibrillation, disconnect other medical electrical equipment which has no defibrillation-
proof applied parts from the patient.
If the device malfunctions during a rescue operation, it will instruct you to get a replacement
defibrillator and will start the voice instruction for CPR. Have CPR performed until the
replacement defibrillator is ready to use.
Page 24
24
Step 3: Perform CPR.
Perform CPR when the i-PAD CU-SP1 AUTO instructs you to do so.
By default, the CU-SP1 AUTO gives voice instruction for CPR during pause for CPR after a
shock delivery. When voice instruction for CPR is needed outside of the default setting, press
the flashing blue i-Button for at least 15 seconds.
[CPR Method]
1. Compression Point
Place the heel of your hand in the middle of the patient's chest between nipples (which is the
lower half of the sternum), and put the heel of your other hand on top of the first so that your
hands are overlapped and parallel.
2. Compression Speed and Depth
Compress the chest at least 5 cm deep, and at a rate of at least 100 compressions per minute.
3. Opening the Airway
While lifting the patient's chin up, tilt the head backward to open the airway.
4. Ventilation Method
Pinch the patient's nose as shown in the figure below, and give the patient enough breath to
make the chest rise significantly.
Page 25
25
If you have not been trained in CPR, you should perform only chest compression or follow the
instructions of the emergency medical services’ agent on the phone.
If you are trained for CPR and able to perform ventilation, perform the chest compression
along with ventilation.
The CPR guide can be set on an administrator's mode. Refer to [Section 5.3: Device Setting]
for more information.
CAUTION
While playing the CPR guide, the device does not analyze the patient's ECG. After the CPR
guide, the device automatically starts the reanalysis of the patient's ECG.
In order to turn the device off after use, press the Power Button for at least 1 second.
Page 26
26
4.3 Defibrillation Procedures in Pediatric Mode
When the patient is older than 1 year and younger than 8 years, defibrillation can be done using
the pediatric pads. When the device is in pediatric mode (pediatric pads are connected to the
device or the Adult/Pediatric Selection Switch is set to Pediatric), it automatically sets the
defibrillation energy to 50 J and provides pediatric CPR guide.
Place pads on the middle of the chest and back as illustrated above. Pads are not specific to
either chest or back.
If there are no pediatric pads for the pediatric patient, use adult pads but set the Adult/Pediatric
Selection Switch to Pediatric Mode, and then perform defibrillation according to the voice
instructions.
Follow the instructions below when giving first aid during pediatric cardiac arrest.
• When giving first aid during a pediatric cardiac arrest, ask others to call the emergency
medical center and to bring the i-PAD CU-SP1 AUTO while you are performing pediatric
CPR.
• When there is no one else around, perform CPR for 1 to 2 minutes, call emergency medical
services, and then get the i-PAD CU-SP1 AUTO.
• If you witnessed the child’s collapse, call emergency medical services immediately, and
then get the i-PAD CU-SP1 AUTO.
Page 27
27
5. After Using the i-PAD CU-SP1 AUTO
5.1 Maintenance After Each Use
• Check if the device for signs of damage and contamination.
• If there is dirt contamination, see Section 6.2.3 on how to clean the device.
• Run a battery insertion test. Refer to Section [8.1: Self-Tests] for the procedure.
If is displayed on the Status LCD after running the test, the device status is normal.
• Dispose of the used pads properly. Place a new pouch of defibrillator pads into the pads
storage compartment. See to it that the pads are not beyond their expiration date.
The i-PAD CU-SP1 AUTO uses disposable pads. Do not reuse them. Refer to Section
[6.2.2: Replacing Supplies] on how to replace the pads.
WARNING
You should use only the defibrillator pads provided and recommended by the manufacturer.
Do not open the pads package until immediately before use. Since the adhesive material on
the pads starts to dry out as soon as the package is opened, the pads may not be usable
regardless of the expiration date.
Page 28
28
5.2 Saving and Transferring Treatment Data
5.2.1 Device Usage
This device automatically saves the following treatment data:
• ECG data
• Usage information
The treatment data is automatically recorded in the internal memory. This data is not erased
even if the device is turned off. The recorded treatment data may be transferred to a personal
computer (PC).
CAUTION
This i-PAD CU-SP1 AUTO keeps the data of the 5 most recent treatment operations and can
save up to 3 hours of ECG data for each rescue operation. ECG data beyond 3 hours will not
be recorded.
When the device is used more than 5 times, it deletes the oldest treatment data to make room
for data from a new treatment operation. It is recommended to transfer treatment data to a PC
after each use of the device.
If the battery pack is removed while the device is operating, treatment data cannot be properly
recorded. If you wish to remove the battery pack, turn the power off by pressing the Power
Button for at least 1 second before removing the battery pack.
5.2.2 Transferring Treatment Data
The treatment data may be transferred via a SD card or IrDA. The entire treatment data of all
patients that is recorded on the device is transferred using only the SD card method, while the
treatment data of one patient can be transferred using only the IrDA method.
1. Copying Treatment Data by Using the SD Card
Format the SD card on the PC to FAT (FAT16) format.
Open the SD Card Cover on the device and insert the SD card into the port.
Page 29
29
When the i-Button is pressed for more than 1 second in standby mode, the mode changes
into administrator mode with voice guide.
The device then gives you a summary (the total hours of the last device use and the number
of defibrillation shocks delivered).
The voice guide gives the S/W version of the device.
When instructed by the voice guide to transfer the treatment history, press the i-Button to
copy the data onto the SD card.
If there is treatment data in the device’s internal memory:
• The device informs you that copying of the treatment data onto the SD card has started, and
starts to copy the data.
• When copying is completed, the device mode changes to CPR guide setting mode. Refer to
[Section 5.3: Device Setting] for details regarding CPR guide setting.
If there is no treatment data in the device’s internal memory:
• The device mode changes to CPR guide setting mode after informing you that no treatment
data exists.
If the file has already been transferred, the device will inform you that the same file exists in
the PC. Press the i-Button for more than 2 seconds to overwrite the existing file or press down
the i-Button shortly to cancel copying the file.
2. Transferring Treatment Data via IrDA
The data may be transferred to a PC using the data management software (CU Expert Ver.3.50
or higher) from the manufacturer. CU Expert includes ECG review and printing functions.
Position the IrDA adapter to face the IrDA port on the device as shown in the figure below.
When the i-Button is pressed for at least 1 second in standby mode, the mode changes to
administrator mode with a voice guide.
Page 30
30
The device gives the you a summary (the total hours of the last device use and the number
of defibrillation shocks delivered).
The voice guide gives the S/W version of the device.
When instructed to transfer the treatment history, press the i-Button to transfer the data.
If there is treatment data in the device’s internal memory:
The voice guide reports the total number of individual treatment data recorded in the device.
By default, of a maximum of 5 individual treatment data, the first on the list is the most recent.
Press the i-Button shortly to change the transfer order of the treatment data as follows:
No.1 treatment data No.2 treatment data No.3 treatment data No.4 treatment data
No.5 treatment data
If you wish to transfer the selected treatment data, press down the i-Button for more than 2
seconds.
Run CU Expert on the PC. Refer to the CU Expert manual for detailed information regarding
how to receive data.
The device will be connected with the CU Expert within a few seconds, and data will be
automatically transferred.
When copying is completed, the mode changes to CPR guide setting mode. Refer to [Section
5.3: Device Setting] for details regarding changing CPR guide setting.
If there is no treatment data in the device’s internal memory:
The device mode changes to CPR guide setting mode after informing you that no treatment
data exists.
CAUTION
The distance between the IrDA port on the device and the IrDA adapter should be within 30
cm, while their angle should be within ±15°. Also, since external light source affects the IrDA,
try to use it in indoors and away from fluorescent and/or incandescent lamps.
Page 31
31
5.3 Device Setting
5.3.1 CPR Guide Setting
The default CPR setting on CU-SP1 AUTO is 5 cycles with 30 chest compressions and 2
breaths in accordance with the American Heart Association (AHA) 2010 CPR Guidelines.
However, you may customize these.
You can set the following:
• Number of chest compressions
• Number of ventilation
• Number of cycles
• Number of chest compressions per minute
• Pausing time
• Detailed guide selection
5.3.2 Setting the CPR Guide
When the i-Button is pressed for at least 1 second in standby mode, the mode changes into
administrator mode with a voice guide.
The device gives you a summary (the total hours of the last device use and the number of
electric shocks).
When instructed to transfer the treatment data, do not press the i-Button, but instead wait for
5 seconds.
When instructed to set the CPR guide, press the i-Button to enter the CPR guide setting
mode.
When instructed to enter a password, enter the set password.
Set the Adult/Child Select Switch to Child, press the i-Button for 5 times, set the Adult/Child
Select Switch to Adult, and press the i-Button once for the password.
Set to pediatric mode.
Set to adult mode
Page 32
32
The voice guide will give information regarding the current CPR guide setting.
Press the i-Button to change the setting, or wait for 3 seconds to go to the next step.
(When the setting is completed, the i-Button flashes twice in blue.)
Settings can then be changed in the following order: Number of Chest Compressions,
Number of Ventilation, Chest Compression rate, Pausing Time, and Detailed Guide
Selection. Refer to [Table 1] CPR Guide Setting Options below.
When the setting is completed, the voice guide will give information regarding the set CPR
guide, which may be saved or canceled.
Press the i-Button to save or the Power Button to cancel by following the voice instructions.
When the CPR guide setting is either saved or canceled, the device automatically shuts
down.
[Table 1] CPR Guide Setting Options
Number
Setting
Option
Range
Unit
Default
Description
1
Number of
Chest
Compression
15, 30
15
30
Perform 30 compressions.
2
Number of
Ventilation
0 to 2 1 2
Give 2 breaths.
3
Number of
Cycles
2 to 10 1 5
Perform 5 cycles of chest
compression and ventilation.
4
Chest
Compression
Rate
100 to
120
5
100
Compress the chest at a rate of
100 compressions per minute.
5
CPR Pause
time
30 to
180 sec.
30 sec.
120 sec.
Pause for 120 seconds
(2 minutes).
6
Detailed
Guide
Selection
On/Off
Off
Turns ON or OFF detailed voice
instructions for the chest
compression and ventilation when
performing CPR.
Page 33
33
By default, Detailed Guide Selection is OFF during CPR so that you can concentrate on the
compression rate and ventilation guidance. If you want the Detailed Guide Selection to be ON
during CPR, set it ON as outlined in the previous pages.
If the Detailed Guide Selection is OFF and the Number of ventilation is set to 0, the CU-SP1
AUTO provides only chest compression guidance for 2 minutes. After 2 minutes, the CU-SP1
AUTO automatically reanalyzes the patient’s ECG.
The CPR chest compression rate can only be set in pediatric mode. In adult mode, the chest
compression rate is fixed at 30 regardless of the set chest compression rate.
Page 34
34
6. Maintenance
6.1 Device Storage
Please refer to the precautions below when storing the Device in order to avoid device damage.
PROHIBITION
Do not operate or store the device in conditions that are beyond the following. specified
limits.
Storage Conditions
The device is stored together with the defibrillator pads and the battery pack is inserted
- ready to be used in an emergency
Temperature: 0 ~ 43 (32 ~ 109)
Humidity: 5% ~ 95% (non condensing)
Transport Environment
device only, no defibrillator pads and battery pack included
Temperature: -20 ~ 60 (-4 ~ 140)
Humidity: 5% ~ 95% (a location with no condensation)
Do not store the device in areas that are directly exposed to sunlight
Do not store the device in areas with highly fluctuating temperatures
Do not store the device near heating equipment
Do not store the device in areas where there is high vibration (in excess of
Road Transportation and Helicopter Minimum Integrity of MIL-STD-810G Method 514.5C)
Do not operate or store the device in environments with high concentration of flammable
gas or anesthetics.
Do not operate or store the device in areas with high concentration of dust
Only personnel authorized by the manufacturer may open the device for servicing. There
are no user serviceable components inside the device.
WARNING
No modification of this equipment is allowed
Page 35
35
6.2 Maintenance
6.2.1 Device Inspection
The i-PAD CU-SP1 AUTO has self-testing capability. The device performs a self-test as soon as
the battery is inserted, turns itself off when the test is done, and periodically wakes up to
perform the daily, weekly, and monthly self-tests. To initiate a battery insertion self test, remove
the battery pack and reinsert. Refer to [Section 8.1: Self-Tests] for more information.
CAUTION
Inspect the i-PAD CU-SP1 AUTO daily to ensure that it is always ready for an emergency.
Check the current status of the device, battery, and pads as displayed on the Status LCD.
Refer to [Section 8.2: Device Status] for information regarding the Status LCD.
6.2.2 Replacing Supplies
When the device is in storage, check the battery level indicator and the pads status on the
Status LCD daily to ensure that the device is always ready for an emergency. Replace the
battery pack or the defibrillator pads when it is depleted or when they go beyond their expiration
date, respectively.
Disposable Battery Pack
Replacement of the Disposable Battery Pack
• Replace the battery pack when it becomes depleted. Refer to [Chapter 8: Troubleshooting]
on how to check the battery status.
• Dispose of depleted battery packs in accordance with local environmental regulations.
• The battery pack is disposable. Do not recharge.
Replacing the Disposable Battery Pack
1. Remove the discharged battery pack by pulling it out while pressing the latch on the bottom
of the device. Refer to the figure below.
WARNING
Use only battery packs recommended and provided by the manufacturer. Using battery packs
not recommended and provided by the manufacturer will result in abnormal operation.
Page 36
36
2. Insert a new battery pack in the direction of the arrow with the label facing upward as shown
in the figure below.
3. Push the battery pack until you hear it click into place.
Page 37
37
CAUTION
Battery Pack Precautions
• Do not subject the battery pack to serious physical impact.
• Do not attempt to open or break apart the battery pack
• Do not let the battery pack come into contact with open flames or hot objects.
• Do not short-circuit the terminals of the battery pack.
• Keep out of the reach of children.
• If any leakage gets in the eye, immediately clean the eye with water and consult with a
doctor.
• Do not store the battery pack under direct sunlight.
• Do not store the battery pack in a wet or very humid place.
• Comply with local regulations when disposing of the battery pack.
• Do not destroy or incinerate the battery pack.
• Never attempt to recharge the disposable battery pack.
Page 38
38
Replacing the Pads
Check the pads status on the Status LCD daily. Do not use pads that are beyond the
expiration date.
• Check the pads package for damage.
• Check the cable outside the packaging pouch for possible defects.
WARNING
Only pads provided by the manufacturer should be used with the i-PAD CU-SP1 AUTO. Using
pads other than the one specified by the manufacturer may affect defibrillation effectiveness.
Replacing Pads
1. Check the expiration date of the pads. Refer to the figure below for checking the expiration
date.
The expiration date is marked to the left
of the “Multifunction Defibrillation
ADULT PADS” label on the pads
package.
The expiration date is marked as
follows:
MM / YYYY
YYYY – Year
MM – Month
2. Used or expired pads should be replaced. Hold the top and bottom of the pads connector
with your fingers, pull it out, and take the pads out from the Pads Storage Compartment as
illustrated below.
Page 39
39
3. Insert the pads connector of the new pads into the Defibrillator Pads Connector , and then put
the pads package in the Pads Storage Compartment as illustrated below.
6.2.3 Cleaning the i-PAD CU-SP1 AUTO
Clean the device with a soft cloth. The following detergents may be used to clean the exterior of
the device:
• Dilute soap and water
• Dilute chlorine bleach (dilute 30 ml of chlorine bleach in one liter of water)
• Dilute ammonia-based cleaners
• Dilute hydrogen peroxide
CAUTION
Do not immerse the device or its accessories in liquids.
Be careful not to allow any liquids to get into the device.
If the device is immersed in liquids, immediately contact the manufacturer or its authorized
service center.
Giving excessive force or shock while cleaning the device may cause damage.
Do not use an acetone-based strong detergent or abrasive when cleaning the device. In
particular, the filter on the IrDA port may be damaged.
Do not use a detergent containing abrasive ingredients.
Do not sterilize the i-PAD CU-SP1 AUTO.
7. Disposal
Dispose of CU-SP1 AUTO and its accessories in accordance with local regulations.
Page 40
40
8. Troubleshooting
8.1 Self-Tests
The following table lists the self-tests done by the device.
Self-Test Type
Description
Battery Insertion
Test
Runs when the battery pack is inserted into the device.
Perform this test:
• Before the device is deployed
• After each use
• When replacing the battery
• When the device is suspected to be damaged
CAUTION
Do not run this test when you are about to use the device to treat a
sudden cardiac arrest victim because this test takes time (around 20
seconds).
If a new battery pack is inserted just before a treatment, do the following
to cancel this test:
• Press the Power Button
• Wait for the device to turn OFF.
• Press the Power Button again to turn the device ON.
Aside from testing its internal systems, the device also tests the following
during this self-test:
• i-Button – press the buttons one by one when instructed
• Defibrillator pads status – the device tests the connection status
(whether connected or not) and the expiration date of the defibrillator
pads.
If no error is detected, will be displayed on the Status LCD.
If an error is detected, will be displayed on the Status LCD and the i-
Button will flash in red. When the i-Button is pressed as directed by the
voice instructions, the device will report the error and turn itself off. Refer
to [Section 8.3: Troubleshooting] for more information.
Self-Test Type
Description
Page 41
41
Power ON Test
The device performs a self-diagnostic test when the Power Button is
pressed
Run-time Test
The device monitors itself in real-time during its operation.
Periodic
Self-Diagnostic
Test
This device performs self-diagnostic tests daily, weekly and monthly. The
periodic self-test checks important features of the device such as the
battery status, pads status and internal circuits.
If the device fails to perform any self-test during use and is unable to defibrillate, it will instruct
the you to replace the device and start the voice instruction for CPR. In order to check the error,
turn the device off by pressing the Power Button. If you press and hold the i-button, the voice
will direct the you to press the blinking red i-Button. You can verify the cause of the error via the
voice instruction by pressing the i-Button. Refer to [Section 8.3: Troubleshooting] for more
information.
CAUTION
It is recommended to run the battery insertion test only during the times enumerated in the
table above. The battery insertion test consumes battery power and will shorten battery life if
done more frequently than necessary.
Page 42
42
8.2 Device Status
The status of the device is indicated by the following symbols:
Indicator
Description
Note
Status LCD
Device Operation
The device is functioning normally.
Status LCD
Device Operation
The device has an error.
Status LCD
Battery Level Indicator
The battery is fully charged.
Status LCD
Battery Level Indicator
Less than half battery power remains.
Status LCD
Battery Level Indicator
Less than a quarter battery power remains.
Status LCD
Battery Level Indicator
Battery is low.
Status LCD
Pads Status
The expiration date of the pads is more than
3 months.
Status LCD
Pads Status
The pads will expire within 3 months.
Status LCD
Pads Status
The pads is used or expired.
Do-Not-Touch-Patient
Indicator: Off
You may touch the patient.
Do-Not-Touch-Patient
Indicator: Light
You may not touch the patient.
CPR Detection Indicator: Light
Indicates that CPR is being performed.
CPR Detection Indicator:
Flashing
Indicates that CPR is not performed or not
properly performed.
i-Button: Flashing in Red
The device detected an error.
Press the i-Button for more information.
Page 43
43
8.3 Troubleshooting
The device informs you of its current status or of problems via status indicators, beeps, and/or
voice instruction. Refer to the following for details:
8.3.1 Troubleshooting While the Device is Operating
Symptom/Voice Instruction
Cause
Resolution
Status LCD
Device Operation
An error has occurred in
the device.
Immediately replace the
defibrillator and perform
CPR if appropriate.
Status LCD
Battery Level
Indicator
The battery is low.
Replace the battery with a
new one.
Status LCD
Pads Status
The pads is expired.
Replace the pads with a
new one.
The pads has been used.
Voice Prompt :
“Low battery”,
“Replace the battery with a new
one.”
The battery is low.
Replace the battery with a
new one.
Voice Prompt :
“Plug the pads connector into the
device.”
The Pads Connector is
disconnected
Ensure the Pads
Connector is properly
connected.
Voice Prompt :
” Used pads”,
“Replace the pads with a new one”
The pads has been
previously used.
Replace the pads with a
new one.
Voice Prompt :
” The pads are beyond their
expiration date”,
“Replace the pads with a new one”
The pads has expired.
Replace the pads with a
new one.
Voice Prompt :
” Press the pads firmly to the bare
skin of the patient”
The pads is not properly
attached to the patient's
skin.
Check if the pads is
securely attached to the
patient's skin.
Voice Prompt :
” No shock delivered”
The pads is not properly
adhering to the patient's
skin.
Press the pads firmly to the
patient's skin.
Shave chest hair or wipe
off moisture if necessary
before attaching the pads.
Page 44
44
If the problem cannot be solved during an emergency, you should follow the following steps:
Quickly replace the defibrillator if possible.
If no replacement device is available, check the patient's condition and perform CPR as
necessary. Continuously check the patient's condition and perform CPR until the
emergency medical services arrives.
8.3.2 Troubleshooting While the Device is not Operating
Symptom
Cause
Resolution
Status LCD
Device Operation
System error
Press the i-Button and hold for at
least 1 second. The device then
goes into Administration Mode.
After going into Administration
Mode, the device will issue the voice
instruction “Press the flashing red i-
Button”
Press the flashing red i-Button and
the device will then announce
system error and the associated
error code.
Contact us by referring to [Chapter
9: Device Service].
Status LCD
Battery Level Indicator
The battery is low.
Replace the battery with a new one.
Status LCD
Pads Status
The pads is expired.
Replace the pads with a new one.
The pads has been
used.
If the problem is not resolved or if no replacement battery is available, contact the
manufacturer (refer to [Chapter 9: Device Service])
Page 45
45
9. Device Service
Device Warranty
Device Name
Model Name
Purchase Name
Serial No.
Distributor
Person in Charge
This device is warranted by CU Medical Systems, Inc. against defects in materials and
workmanship for five full years from the date of original purchase. During the warranty period,
we will repair or, at our option, replace at no charge a device that proves to be defective,
provided you return the device, shipping prepaid, to us or to our authorized representative.
This warranty does not apply if the device has been damaged by accident or misuse or as the
result of service or modification by entities other than CU Medical Systems, Inc. or its
authorized representatives. IN NO EVENT SHALL CU MEDICAL SYSTEMS BE LIABLE FOR
CONSEQUENTIAL DAMAGES.
Only devices with serial numbers and their accessories are covered under this warranty.
PHYSICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED
UNDER THE WARRANTY. Items such as cables and modules without serial numbers are not
covered under this warranty.
Warranty Disclaimer
The following renders this warranty null and void:
• Servicing by unauthorized personnel.
• If the factory seal is broken without proper authorization from CU Medical Systems, Inc.
• Failure or damage caused by a fall or external shock after purchase
• Damage by natural disasters such as fire, earthquake, flood and/or lightning
• Failure or damage by environmental pollution or abnormal voltage
• Damage caused by storage in conditions beyond the specified limits.
• Failure due to depletion of consumables
• Failure caused by sand and/or soil getting inside the device
• The purchase date, customer name, distributor name, batch number and other listed
information being arbitrarily changed
• No proof of purchase provided along with the device warranty
• Usage of accessories and parts not recommended by the manufacturer.
• Other failure or damage caused by inappropriate operation.
Page 46
46
Service
The i-PAD CU-SP1 AUTO must be serviced only by authorized personnel.
The i-PAD CU-SP1 AUTO will be serviced free of charge during the warranty period. After the
warranty period, the cost of material and service shall be shouldered by the user.
When the i-PAD CU-SP1 AUTO is not operating properly, immediately bring it for servicing to
an authorized service center.
Please fill out the following table with the necessary information when requesting for service
Device classification
Fully Automated External Defibrillator
Device Name
i-PAD
Model Number
CU-SP1 AUTO
Serial Number
Date of Purchase
Sales Representative
User
Information Name
Address
Contact no.
Brief description of the
problem
CAUTION
If there is a device problem (except battery pack and defibrillation electrode problems) please contact
service center.
Page 47
47
Appendix
A . Rescue Protocol
Page 48
48
Voice Prompt: Do not touch the patient. Analyzing heart rhythm.
A
i-PAD CU-SP1
detect/decide:
ECG Shockable?
YES
NO
Voice Prompt: Stand by for automatic shock delivery. Warning: No one should be touching the patient.
Charging
Complete?
YES
NO
Voice Prompt: Shock will be delivered in(3, 2, 1)
Voice Prompt: Be sure Emergency Medical Services have been called.
Voice Prompt: You may touch the patient.
Voice Prompt: Begin CPR, now.
Voice Prompt: No shock advised.
B
Voice Prompt: Shock advised.
Page 49
49
Page 50
50
B . Parts and Accessories
To order replacement parts and accessories, cite the part and ordering numbers given in the
following table.
B.1 Standard Accessories
Name
Part Number
Ordering Number
Adult Pads (disposable)
CUA1007S
SP1-OA04
Disposable Battery Pack(Long-life)
CUSA1103BB
SP1-OA03
Instructions for Use
SPA-OPM-E-01
-
B.2 Optional Accessories
Carrying Case
SP1-A-BAG-3010
SP1-OA01
Disposable Battery Pack(Standard)
CUSA1103BS
SP1-OA02
Pediatric Pads (disposable)
CUA1102S
SP1-OA05
IrDA Adapter
IR-220LPLUS
SP1-OA06
PC S/W
CU Expert ver. 3.50 or higher
SP1-OA07
SD Card
HD1-CARD-SD
SP1-OA10
SD Card Reader
HD1-CARD-READER
-
Page 51
51
C . Description of Symbols
C.1 i-PAD CU-SP1 AUTO Defibrillator
Symbol
Description
Power ON/OFF button
i-Button
Adult / Pediatric Selection Switch
Do-Not-Touch-Patient Indicator
CPR Detection Indicator
Defibrillation-proof type BF applied part
General warning sign
Refer to instruction manual/booklet
CE Mark; meets the requirements of the European Medical
Device Directive 2007/47/EC and its revisions.
Serial Number
MANUFACTURER
Date of manufacture
AUTHORISED REPRESENTATIVE IN THE EUROPEAN
COMMUNITY
Page 52
52
C.1 i-PAD CU-SP1 AUTO Defibrillator (Continue)
Symbol
Description
Separate collection for electrical and electronic equipment
Page 53
53
C.2 i-PAD CU-SP1 AUTO Packaging
Symbol
Description
Stack up to 6 cartons high only
This side up
Keep dry
Fragile; breakable
Use no hooks
Storage Temperature limits: 0 to 43(32 to 109)
recyclable
CE Mark; meets the requirements of the European Medical
Device Directive 2007/47/EC and its revisions.
Serial Number
AUTHORISED REPRESENTATIVE IN THE EUROPEAN
COMMUNITY
Date of manufacture
Page 54
54
C.3 Accessories
C.3.1 Disposable Battery Pack (CUSA1103BB, CUSA1103BS)
Symbol
Description
Lithium Manganese Dioxide battery
Direct current
Lot Number
Do not expose the battery to high heat or open flames.
Do not incinerate the battery.
Do not mutilate the battery or open the battery case
Do not crush the battery
Refer to instruction manual/booklet
Separate collection for electrical and electronic equipment Date of manufacture
MANUFACTURER
CE Mark; meets the requirements of the European Medical
Device Directive 2007/47/EC and its revisions.
Direction of battery pack insertion
Plus; Positive polarity
Minus; negative polarity
Data terminal
Page 55
55
C.3.2 Pads (CUA1007S, CUA1102S)
Symbol
Description
Temperature limits: 0 to 43(32 to 109)
Lot number
Expiration date
Single use only; do not reuse
Do not fold or bend.
Refer to instruction manual/booklet
General warning sign
Contains no latex
Expiration Date and Lot number sticker
Pads pull –out (direction indicator)
CE Mark; meets the requirements of the European Medical
Device Directive 2007/47/EC and its revisions.
Page 56
56
D . Glossary
1 CPR
1 CPR consists of 5 cycles. (When the device is set to 5
cycles as default)
1 Cycle
Refers to 30 chest compressions followed by 2 breaths during
CPR. (When the device is set to the default setting [30:2])
If you specify the number of compression and number of
breath, the cycle is performed in accordance with the
specified protocol. Refer to [Section 5.3: Device Setting] for
detailed setting method.
Abrasive
A material used to sharpen and clean the surface of metal,
glass, stone and wood, which includes emery, quartz powder
and glass dust. Do not use these abrasives to clean the
device.
Adhesive Material
on the Pads
(Gel)
The adhesive material on the pads is very important for
maintaining the optimum adhesion between the skin and
pads. Therefore, never open the pads package when the
pads is not needed, and periodically check the expiration date
of the pads.
Adult
The adult in these Instructions for Use is defined as a person
who is older than 8 years or heavier than 25 kg.
American Heart
Association (AHA)
2010 CPR
Guidelines
The default settings of this device direct the you to perform
CPR immediately after one electric shock in accordance with
the 2010 CPR Guidelines. Also, the CPR guide is composed
of 5 cycles with the chest compression to ventilation ratio of
30:2 (if the device is set to a default setting of 5 cycles, 30:2).
If you are not trained in ventilation, perform only the chest
compression. Refer to [Section 5.3: Device Setting] for the
CPR setting. Please contact the manufacturer for additional
information.
Page 57
57
Arrhythmia
An abnormal heart rhythm.
Battery Pack
A disposable battery that supplies power to the i-PAD CU-
SP1 AUTO.
Cardiac Arrest
Patient
A patient with cardiac arrest symptoms. This device should be
used for the patient with the following symptoms: No
response, no movement and no normal breathing.
Communication
Port
A port that sends and receives data between the device and
PC.
Condensation
Moisture has an adverse effect on the device when
condensation is formed on the device surface. The device
should be stored in a dry environment without excessive
humidity.
CPR Mode
The device provides guidance for CPR while pausing analysis
of the patient's ECG such that you can easily perform CPR.
The CPR mode on this device complies with AHA's 2010 CPR
Guidelines. Refer to [Section 4.3., Step 3: Perform CPR] for
more information.
Defibrillation
Is a process in which an electronic device gives an electric
shock to the heart. This helps reestablish normal contraction
rhythms in a heart having dangerous arrhythmia or in cardiac
arrest.
Defibrillator Pads
Connector
A connector on the device that is used to connect the device
with defibrillator pads.
Disposable Battery
Pack
A disposable battery pack that provides power to the device.
Never charge this battery pack.
ECG
An abbreviation for electrocardiogram. A record of the heart's
electrical rhythm as detected by the defibrillation pads.
Page 58
58
Electric Shock
This device charges large energy in a short time and performs
defibrillation via an electric shock.
Error
A status in which the device does not properly operate. Refer
to [Section 8.3: Troubleshooting] for more information.
Fibrillation
Refers to an irregularity of the heart causing ineffective
circulation. Ventricular fibrillation is accompanied with an
acute cardiac arrest.
Flashing
A status in which the indicator is flashing.
i-Button
A button to check the most recent device usage, to report
error messages, to transfer the ECG and event data, and to
change the CPR guide settings.
IrDA Port
A communication port that sends and receives data between
the device and computer. Since this IrDA port utilizes light
(infrared), care needs to be taken to reduce interference.
Refer to the [CU Expert] manual for more information.
Light
A status in which the indicator is lit.
Operation Mode
An on the Status LCD while the device is on indicating
that the device is properly operating.
Pads
The pads stated in these Instructions for Use refers to a pads
(disposable) for defibrillation.
Pad 1
Refers to a pad that is placed under the right clavicle. Please
refer to the picture on the pad. (The position may be switched
with pad 2.)
Pad 2
Refers to a pad that is placed on the ribs on the patient's
lower left chest directly under the armpit. Please refer to the
picture on the pads (the position may be switched with pad 1).
Page 59
59
Pads Connector
The connector on the pads that is used to connect the pads
with the i-PAD CU-SP1 AUTO.
Applied part
Defibrillation electrode pads
Patient isolation type BF, defibrillation protected
PC S/W CU Expert
(CU-EX1)
PC software used to modify the settings of the i-PAD CU-SP1
AUTO and to manage treatment data. Refer to the appendix
on accessories if you want to purchase this software.
Pediatric
The child in these Instructions for Use is defined as a person
who is older than 1 year and younger than 8 years as well as
lighter than 25 kg.
Power Button
A green button on the front of the device. The device turns on
when the Power Button is pressed during Standby Mode, and
it turns off when the Power Button is pressed for one second
while the device is on. If the Power Button is pressed during
the battery insertion test, the battery insertion test is canceled.
Device
The Device referred to in these Instructions for Use is the i-
PAD CU-SP1 AUTO Fully Automated External Defibrillator
(AED).
Pads liner
The liner that protects the conductive gel of the pads during
storage inside the pads pouch.
SD Card
An external memory card that could be used to store
treatment data (ECG and event) from the internal memory of
the device.
Self- Test
Self diagnostic tests that verify the proper operation of the
subsystems of the device.
Page 60
60
Internal discharge
(disarm)
If the equipment determines that the patient does not need an
electric shock due to the change in the patient's ECG, the
charged energy is discharged within the equipment.
Fully Automated
External
Defibrillator
(AED)
A device that automatically de_brillating shock after analyzing
and recognizing a shockable rhythm.
Standby Mode
The mode of the i-PAD CU-SP1 AUTO when the Power
Button is OFF but the battery pack is inserted. If is
shown on the Status LCD while the device is in standby
mode, the device is ready to be used as needed in an
emergency).
We
Refers to CU Medical Systems Inc.
Page 61
61
E . Device Specifications
Model Name: CU-SP1 AUTO
Physical
Category
Nominal Specifications
Dimensions
260㎜ x 256㎜ x 69.5㎜ (Width x Length x Height)
Weight
2.4 (Including the battery pack and pads)
Environmental
Category
Nominal Specifications
Operational Status (The device is in emergency use)
Temperature: 0 ~ 43 (32 ~ 109℉)
Humidity: 5% ~ 95% (non condensing)
Standby Status (The device is stored together with the defibrillator pads and the battery pack
is inserted - ready to be used in an emergency)
Temperature: 0 ~ 43 (32 ~ 109℉)
Humidity: 5% ~ 95% (non condensing)
Transport Status (device only, no defibrillator pads and battery pack included)
Temperature: -20 ~ 60 (-4 ~ 140)
Humidity: 5% ~ 95% (non condensing)
Altitude
0 to 15,000 feet (operational and storage)
Drop
Withstands 1.2-meter drop to any edge, corner, or surface
Vibration
Operating: Meets MIL-STD-810G Fig.514.6E-1, random
Standby: Meets MIL-STD-810G Fig.514.6E-2, swept sine(helicopter)
Sealing
IEC 60529: IP55
ESD
Meets IEC 61000-4-2:2001
EMI (Radiated)
Meets IEC 60601-1-2 limits, method EN 55011:2007 +A2:2007,
Group 1, Class B
EMI (Immunity)
Meets IEC 60601-1-2 limits, method EN 61000-4-3:2006 +A1:2008 Level 3
(10V/m 80MHz to 2500MHz)
Page 62
62
Defibrillator
Category
Nominal Specifications
Operating Mode
Fully-automatic
Waveform
e-cube biphasic (Truncated exponential type)
Output Energy
150 J at 50 Ω load for adults
50 J at 50 Ω load for children
Charge Control
Controlled by an automated patient analysis system
Charging Time
Within 10 seconds from when the voice instruction, "An electric shock is
needed." is issued.
Energy Storage
Charging Time
10 seconds, typical
Using new, disposal LiMnO2 battery pack (at 20)
11 seconds, typical
Using new, disposable LiMnO2 battery pack, depleted by
15 discharges at 200 Joules per discharge. (at 20)
Time from initiation of rhythm analysis (voice instruction:
“DO NOT TOUCH PATIENT, ANALYZING HEART RHYTHM”)
to readiness for discharge (voice instruction: STAND BY,
FOR AUTOMATIC SHOCK DELIVERY. WARNING: NO ONE
SHOULD BE TOUCHING THE PATIENT.”)
New battery pack
10 Seconds, typical (at 20)
New battery pack: 16th shock
discharge 11 Seconds, typical (at
20)
Time from Power ON to readiness for discharge
(voice instruction: “SHOCK WILL BE DELIVERED IN 3, 2, 1”)
New battery pack: 16th shock
discharge 25 Seconds, typical
(at 20)
Charging
Indicator
Beeper
Voice Instruction (“Shock will be delivered in”)
Time from CPR to
Shock
At least 6 seconds from the completion of CPR to shock delivery
Discharge
The device performs a self-discharge in the following events:
When the patient's ECG changes to a rhythm that does not require defibrillation.
When the device is turned off by pressing the Power Button for at least second.
When the pads is detached from the patient's body or the pads connector is
detached from the device.
When the impedance of the patient is out of the range of defibrillation
(25 Ω ~ 175 Ω)
Automatic Shock
Delivery
Shock is automatically delivered if a shockable rhythm is detected.
Page 63
63
Defibrillator
Category
Nominal Specifications
Shock Delivery
Vector
Adult pads in the anterior-anterior position
Pediatric pads in the anterior-posterior position
Patient Isolation
Type BF, defibrillation protected
Biphasic Truncated Exponential Type.
The shock waveform profile is automatically compensated for the patient’s transthoracic impedance.
A = first phase duration
B = second phase duration0
C = interphase duration
D = peak current
Output Waveform for Adult (150 Joules)
Patient
Impedance
(Ohms, Ω)
First Phase
duration
(milliseconds, ms)
Second Phase
duration
(milliseconds, ms)
Peak
Current
(A)
Energy
(Joules, J)
Energy
Accuracy
(Joules, J)
25
2.4
2.4
64.5
147.8
150(±15%)
50
4.4
4.4
32.7
149.7
150(±15%)
75
6.3
6.3
22.5
151.5
150(±15%)
100
8.8
8.8
15.9
148.1
150(±15%)
125
10.7
10.7
13.0
149
150(±15%)
150
12.7
12.7
11.0
148.2
150(±15%)
175
15.0
15.0
9.5
148.8
150(±15%)
A B
time(ms)
current(A)
C
D
Page 64
64
Output Waveform for Child (50 Joules)
Patient
Impedance
(Ohms, Ω)
First Phase
duration
(milliseconds, ms)
Second Phase
duration
(milliseconds, ms)
Peak
Current
(A)
Energy
(Joules, J)
Energy
Accuracy
(Joules, J)
25
2.3
2.3
35.4
50.2
50(±15%)
50
4.3
4.3
18.4
50.7
50(±15%)
75
6.3
6.3
12.3
49.7
50(±15%)
100
8.5
8.5
9.1
49.5
50(±15%)
125
10.6
10.6
7.3
50.3
50(±15%)
150
12.7
12.7
5.8
49
50(±15%)
175
15.0
15.0
4.9
49.6
50(±15%)
ECG Acquisition
Category
Nominal Specifications
Acquired ECG Lead
Lead II
Frequency Response
1 Hz to 30 Hz
ECG Analysis System
Category
Nominal Specifications
Function
Determines the impedance of the patient and evaluates the ECG of the patient to determine whether it is shockable or non shockable
Impedance Range
25Ω to 175Ω (shock will not be delivered if the patient’s impedance is beyond this range).
Shockable Rhythms
Ventricular Fibrillation or Fast Ventricular Tachycardia
Non Shockable
Rhythms
ECG rhythms excluding ventricular fibrillation and ventricular tachycardia When a rhythm that does not require defibrillation is detected, the device directs you to perform CPR.
Analysis Protocol
Prepare for shock delivery of pause for CPR, depending on the results of analysis.
Sensitivity and
Specificity
Meets ANSI/AAMI DF80 guidelines
Page 65
65
ECG Analysis System - ECG Database Test
ECG
Rhythm
Class
Rhythms
Minimum
test sample
size
Performan
ce goal
Test
sample
size
Shock
Decision
No Shock
Decision
Observed
Performance
90% One
Sided Lower
Confidence
Limit
SHOCKABLE
Coarse VF
200
>90%
sensitivity
219
213
6
97.26%
(213/219)
sensitivity
95%
Fast VT*
50
>75%
sensitivity
137
111
26
81.02%
(111/137)
sensitivity
76%
NON SHOCKABLE
Normal Sinus
Rhythm
100
minimum
(arbitrary)
> 99%
specificity
100 0 100
100%
(100/100)
specificity
97%
AF,SB,
SVT, heart
block,
idioven-
tricular PVC’s
30
(arbitrary)
> 95%
specificity
219 1 218
99.54%
(218/219)
specificity
98%
Asystole
100
> 95%
specificity
132 5 127
96.21%
(127/132)
specificity
93%
*Ventricular Tachycardia: less than 150 bpm is non-shockable, greater than or equal to 150 bpm is shockable
For overall ECG waveform test result;
VF and VT
All other ECG rhythms
Shock
(A)
324 cases
(B)
6 cases
No shock
(C)
32 cases
(D)
445 cases
The sensitivity of the device for shockable rhythm is A/(A+C): 91.01% (324/356)
The true predictive value of the device is A/(A+B): 98.18% (324/330)
The specificity of the device for non shockable rhythm is D/(B+D): 98.67% (445/451)
The false positive rate of the device for shockable rhythm is B/(B+D): 1.33% (6/451)
Page 66
66
Control Devices, Indicators, Voice Instructions
Category
Nominal Specifications
Control Devices
Power Button, i-Button, Adult/Pediatric Selection Switch
Status LCD
Displays device status, battery level and pads status
Indicator
Do-Not-Touch-Patient Indicator: Lights when the defibrillator is analyzing or
delivering an electric shock.
Pads Patch Position Indicators: Flashes when the defibrillator is turned on;
turns off when the pads is attached on the patient.
Pads Connector Status Indicator: Flashes when the defibrillator is turned on
and the pads connector is not connected; lights when the pads connector is
connected.
CPR Detection Indicator: Lights if CPR is detected; flashes if CPR is not
detected.
Blue i-Button: Flashes when guiding CPR, transferring the treatment history
and setting the CPR mode.
Red i-Button: Flashes when an error occurs.
Speaker
Plays back voice instructions. The CU-SP1 AUTO analyzes the ambient noise
level during a treatment operation. If ambient noise level is high, it
automatically increases the voice instructions volume so that you can hear
them clearly.
Beeper
Various beeping output
Battery Level
The battery level is automatically performed during periodic self tests, power
ON self-test, and run-time self-test.
Low Battery
Indicator
Shown on the Status LCD, announced via voice instruction, and indicated via
the flashing red i-Button
Voice Instruction
Guides the user via voice instructions.
Page 67
67
Self-Diagnostic Test
Auto
Power On Self-Test, Run-time Self-Test
Daily, Weekly, and Monthly Self-Test
Manual
Battery Pack Insertion Test (done when the user inserts the battery pack into
the battery pack compartment of the device)
Disposable Battery Pack
Category
Nominal Specifications
Battery Type
12V DC, 2.8Ah LiMnO2, Disposable: Standard
12V DC, 4.2Ah LiMnO2, Disposable: Long-life
Capacity
Standard - At least 50 shocks for a new battery
or 4 hours of operating time at room temperature (at 20)
Long-life - At least 200 shocks for a new battery
or 8 hours of operating time at room temperature (at 20)
Standby Life (After
Inserting the Battery)
Standard - At least 3 years from the date of manufacture if stored and
maintained in accordance with the instructions in this document.
Long-life - At least 5 years from the date of manufacture if stored and
maintained in accordance with the instructions in this document.
Temperature Ranges
Operating
Temperature: 0 ~ 43 (32 ~ 109)
Storage
Temperature: -20 ~ 60 (-4 ~ 140)
Page 68
68
Adult Defibrillation Pads (CUA1007S)
Category
Nominal Specifications
Type
Adult
Electrode Area
120 cm
2
Cable Length
Total 120 cm (Inside the pouch: 95 cm, Outside the pouch: 25 cm)
Shelf life
At least 36 months from the date of manufacture
Pediatric Defibrillation Pads (CUA1102S)
Category
Nominal Specifications
Type
Pediatric
Electrode Area
46.43 cm2
Cable Length
Total 120 cm (Inside the pouch: 80 cm, Outside the pouch: 40 cm)
Shelf life
At least 30 months from the date of manufacture
Data Storage and Transfer
Category
Nominal Specifications
IrDA
For PC communications
Internal Memory Data
Capacity
5 individual treatments, up to 3 hours per treatment
SD Card
External memory. Data may be copied from the internal memory to the SD
Card.
Page 69
69
F . Electromagnetic Compatibility
Guidance and manufacturer’s declaration – electromagnetic emissions
The i-PAD CU-SP1 AUTO is intended for use in the electromagnetic environment specified
below. The customer or the user of the i-PAD CU-SP1 AUTO should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF Emissions
CISPR 11
Group 1
The i-PAD CU-SP1 AUTO uses RF energy only for
its internal function. Therefore, its emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The i-PAD CU-SP1 AUTO is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes. Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
WARNING
The i-PAD CU-SP1 AUTO should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the i-PAD CU-SP1 AUTO should be observed to verify
normal operation in the configuration in which it will be used.
Page 70
70
Guidance and manufacturer’s declaration – electromagnetic immunity
The i-PAD CU-SP1 AUTO is intended for use in the electromagnetic environment
specified below. The customer or the user of the i-PAD CU-SP1 AUTO should assure that
it is used in such an environment.
Immunity Test
IEC 60601-1 test
level
Complianc
e level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±6 kV
Contact
Floors shou ld be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
±8 kV air
±8 kV air
Electrical fast
transi ent/burst
IEC 61000-4-4
±2 kV for power
supply lines
Not
Applic able
Mains power quality should be that
of a typical commercial or hospital
enviro nment
±1 kV for input/output
lines
Surge
IEC 61000-4-5
±1 kV diffe rential
mode
Not
Applic able
Mains power quality should be that
of a typical commercial or hospital
enviro nment.
±2 kV common mode
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95% dip in
UT) for 0.5 cycles
Not
Applic able
Mains power quality should be that
of a typical commercial or hospital
enviro nment. If the user of the i-
PAD CU-SP1 AUTO image
intensifier requires continued
operation during power mains
interruptions, it is recommended
that the i-PAD CU-SP1 AUTO
image intensifie r be powered from
an uni nterruptible power supply.
40 % UT (60% dip in
UT) for 5 c ycles
70 % UT (30% dip in
UT) for 25 cycles
<5 % UT (>95% dip in
UT) for 5 s
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be a t levels characteristic of
a typical location in a typical
commercial or hospital
enviro nment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Page 71
71
Guidance and manufacturer’s declaration – electromagnetic immunity
The i-PAD CU-SP1 AUTO is intended for use in the electromagnetic environment specified below.
The customer or the user of the i-PAD CU-SP1 AUTO should assure that it is used in such an
environment.
Immunity
Test
IEC 60601 Test
level
Complia
nce level
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the i-PAD CU-SP1 AUTO, includi ng cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bandsa
3 Vrms
  󰇟
 
󰇠
10 Vrms
150 kHz to 80 MHz
in ISM bandsa
10 Vrms
  󰇟
 
󰇠
Radiated
RF IEC
61000-4-3 10 V/m
80 MHz to 2,5 GHz
10 V/m
  󰇟
 
󰇠
20 V/m
80 MHz to 2,5 GHz
20 V/m
  󰇟
 
󰇠
where P is the maximum output powe r rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m)b
Field strengths from fi xed RF transmitters, as
deter-mined by an electromagnetic site surveyc,
should be less than the compliance level in each
frequency ranged.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Page 72
72
NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and refection from structures, objec ts an d people.
a
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765
MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz
to 40,70 MHz
b
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the
frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas. For this reason, an additional factor of 10/3 is used in c alcul ating
the recommended separation distance for transmitters in these frequency ranges
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
teleph ones and land mobile radios, amateur radio, AM and FM radio broadc ast a nd TV
broadc ast cannot be predicted theoretically with accuracy. To assess the electromagn e tic
enviro nment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field stre ngth in the location in which the i-PAD CU-SP1 AUTO is
used exceeds the applicable RF compliance level above, the CU-SP1 AUTO should be
observed to verify normal operation. If abnormal performance is observed, add itional
measures may be necessary, such as reorienting or relocating the i-PAD CU-SP1 AUTO
d
Over the frequency range 150 kHz to 80 MHz, fiel d strengths should be l ess t han [V1] V / m.
Page 73
73
Recommended separation distances between portable and mobile RF
communications equipment and the CU-SP1 AUTO
The i-PAD CU-SP1 AUTO is inten ded for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the i-PAD CU-SP1 AUTO can help
preven t electromag netic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the i-PAD CU-SP1 AUTO as
recommended belo w, according to the maximum output power of the communications equipment.
Rated
maximum
output
power of
transmitter
[W]
Separation distance according to freque ncy of transmitter [m]
150 kHz to 80 MHz
outside ISM bands
  󰇟


󰇠
150 kHz to 80
MHz in ISM bands
  󰇟
 
󰇠
80 MHz to 800 MHz
  󰇟
 
󰇠
800 MHz to 2,5
GHz
  󰇟
 
󰇠
V1 = 3 Vrms
V2 = 10 Vrms
E1 =
10 V/m
E1 =
20 V/m
E1 =
10 V/m
E1 =
20 V/m
0.01
0.06
0.12
0.12
0.06
0.23
0.16
0.1
0.11
0.38
0.38
0.19
0.73
0.36 1 0.35
1.20
1.20
0.60
2.3.0
1.15
10
1.11
3.79
3.79
1.90
7.27
3.64
100
3.50
12.00
12.00
6.00
23.00
11.50
For transmitters rated at a maximum output power not listed above, the recommended
separa tion distance d in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output po wer rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1) At 80 MHz and 800 MHz, the sepa ration distance for the higher frequency range
applies.
NOTE 2) The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are
6,765 MHz to 6,795 MHz; 13,553 MHz to 1 3,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66
MHz to 40,70 MHz .
NOTE 3)An additional facto r of 10/3 is used in calculating the recommended separation distance
for transmitters in the ISM frequency bands between 150 kHz and 80 MHz a nd in the frequency
range 80 MHz to 2.5 GHz to decreas e the l ik elihood that mobile/portable communications
equipment could cause interference if it is inadvertently brought into patient areas.
NOTE 4) These gu ideli nes may not apply in all situations. Electromagnetic propagation is
affect ed by absorption and refection from structures, objects and people
Loading...