C-THERM TCi Operator's Manual

C-Therm TCi
Operator
TH
130041 Rev.
H
TCi Operator Manual
iii
Table of Contents
1
Introduction
................................
................................
................................
.........................
1
1.1
A note on instrument limitations
................................
................................
..................
2
1.2C-
Therm TCi Manual Intent
................................
................................
.........................
3
1.3C-
Therm TCi System
Setup
................................
................................
........................
4
1.3.1
Checking the items and their condition
................................
................................
....41.3.2
Checking grounded cable
................................
................................
.......................
4
1.3.3
Making connectio
ns
................................
................................
................................
4
1.3.4
Starting up the C
-
Therm TCi system
................................
................................
.......
4
2C-
Therm TCi Software Use
................................
................................
................................
.
5
2.1
Menu Items
................................
................................
................................
................
5
2.2
Table Use
................................
................................
................................
...................
8
2.2.1
Menu
s
................................
................................
................................
....................
9
2.2.2
Buttons
................................
................................
................................
.................
10
2.2.3
Table Options
................................
................................
................................
.......112.2.4
Using the Parameter Fields to Search
................................
................................
...112.3
Reco
rd Management
................................
................................
................................
11
2.3.1
Create a Record
................................
................................
................................
...122.3.2
View a Record
................................
................................
................................
......122.3.3
Edit a Record
................................
................................
................................
.......
13
2.3.4
Delete a Record
................................
................................
................................
...
13
2.3.5
Recover a Record
................................
................................
................................
14
2.4
Printing
................................
................................
................................
.....................
143Records
................................
................................
................................
............................
16
3.1
Project
................................
................................
................................
......................
16
3.2
Material Group
................................
................................
................................
..........
16
3.3
Material
................................
................................
................................
....................
17
3.4
Material Lot
................................
................................
................................
..............
18
3.5
Test Method
................................
................................
................................
.............
18
3.6
Contact Agents
................................
................................
................................
.........
20
3.7
Notes
................................
................................
................................
........................
20
3.8
Calibration Materials
................................
................................
................................
.213.9
Calibration Material G
roups
................................
................................
......................
22
3.10
Calibration Methods
................................
................................
................................
..234
Testing
................................
................................
................................
..............................
26
4.1
General Procedure
................................
................................
................................
...264.2
Foam Testing
................................
................................
................................
...........
31
4.3
Powder Testing
................................
................................
................................
........
31
4.4
Solid Testing
................................
................................
................................
.............
32
4.5
Liquid Testing
................................
................................
................................
...........
32
4.6
Blotter
or Thin Sample Testing
................................
................................
..................
33
4.7
Indirect Property Calculations
................................
................................
...................
35
4.8
Altering the Test Method during the Test
................................
................................
...364.9
Changing the Units of Measurement
................................
................................
.........
36
4.10
Auto Test
ing
................................
................................
................................
.............
36
4.11
Contact Agents
................................
................................
................................
.........
37
4.11.1
Testing with Wakefield 120 Thermal Grease
................................
.....................
38
4.12
Chemical Testing with RTV Tabs
................................
................................
..............
405Data Analysis
techniques and practices
................................
................................
............
42
5.1
Opening a Test
................................
................................
................................
.........
42
5.2
Using Excel for added Analysis Power
................................
................................
......42
5.2.1
Copying to the Clipboard
................................
................................
......................
42
5.3
Overlaying Measurements/Tests
................................
................................
...............
42
5.4
Editing a Test
................................
................................
................................
...........
42
TCi Operator Manual
iv
5.5
Reports
................................
................................
................................
....................
43
5.6
Exporting and Importing
................................
................................
............................
44
5.6.1
Export
................................
................................
................................
..................
44
5.6.2
Import
................................
................................
................................
...................
466User Calibration
................................
................................
................................
................
48
6.1
User Calibration Wizard
................................
................................
............................
48
6.2
Opening or Continuin
g a User Calibration in the Wizard
................................
............
55
6.3
User Calibration Overview
................................
................................
........................
58
6.4
Create Calibration Materials
................................
................................
......................
59
6.5
Create Calibration Material Group
................................
................................
.............
59
6.6
Create Cal
ibration Method
................................
................................
........................
60
6.7
Acquire Calibration Data
................................
................................
...........................
60
6.8
Compute Results
................................
................................
................................
......627
Software Setup and Maintenance
................................
................................
......................
65
7.1
Setup
................................
................................
................................
........................
65
7.2
System Policies
................................
................................
................................
........
72
7.3
Manage Users and User Groups
................................
................................
...............
74
7.3.1
Permissions
................................
................................
................................
..........
74
7.3.2
Users
................................
................................
................................
...................
75
7.3.3
Groups
................................
................................
................................
.................
77
7.3.4
Reactivating a Locked Account
................................
................................
.............
79
7.3.5
Clear Password History
................................
................................
........................
79
7.4
Change Password
................................
................................
................................
....798
System Calibration and Maintenance
................................
................................
................
81
8.1
Calibration
................................
................................
................................
................
81
8.1.1
Unlock Calibrations/User Calibration
................................
................................
.....818.2
Diag
nostics
................................
................................
................................
...............
82
8.2.1
Reference Material Test
................................
................................
.......................
82
8.2.2
System Checks
................................
................................
................................
....
86
8.3
Cleaning Procedure
................................
................................
................................
..878.4
Troubleshooting
................................
................................
................................
........
88
8.4.1
General Information required
................................
................................
................
88
8.4.2
Troubleshooting List
................................
................................
.............................
899Theory of Operation
................................
................................
................................
..........
92
9.1
TCi Calibrations
................................
................................
................................
........
92
9.1.1
Sensor temperature calibration
................................
................................
.............
92
9.1.2
Basic multi
-
point effusivity calibration
................................
................................
....939.2
Material groups and power levels
................................
................................
..............
98
9.3
Contact ag
ent
................................
................................
................................
...........
99
9.4
Thermal conductivity (k) calibration
................................
................................
.........
100
9.4.1
Limitations for
k
measurements
................................
................................
..........
101
9.5
Temperature considerations in effusivity calibration
................................
................
102
9.6
Temperature considerations in conductivity calibration
................................
............
103
9.7
Error analysis
................................
................................
................................
.........
104
9.7.1
General
................................
................................
................................
..............
104
9.7.2
Material
................................
................................
................................
..............
105
9.7.3
Contact
................................
................................
................................
..............
105
9.7.4
Equipment
................................
................................
................................
..........
105
9.7.5
Calibration
................................
................................
................................
..........
106
9.7.6
Additional Notes
Heat Equation
................................
................................
.......
107
Appendices
................................
................................
................................
.............................
110
Appendix A: System Components and Accesso
ries
................................
.............................
110
Appendix B: C
-
Therm TCi Controller Specifications
................................
.............................
111C-
Therm TCi Sensor Specifications
................................
................................
.................
112
C-Therm TCi Standar
d Computer Specifications
................................
.............................
113
TCi Operator Manual
v
Appendix C: Materials of Construction and Chemical Resistance Data
................................
114
Appendix D: Reference Material Standard Values
................................
...............................
118
Appendix E: Permission Default Values
................................
................................
...............
119
Appendix F: CE
Certification
................................
................................
................................
121
Appendix G: Units and Conversions
................................
................................
....................
122
Appendix H: Contact Agent MSDS
Glycol
................................
................................
.........
124
Appendix I: Disclaimer
................................
................................
................................
.........
132
TCi Operator Manual
vi
TCi Operator Manual
1
1
Introduction
The third generation of
theC-
Therm
thermal conductivity
technology expands the capabilities of
this rapid, non
-
destructive thermal testing instrument to a whole new level. Designed to provide
simple, highly accurate thermal characterization
of thermal conductivity and thermal effusivity
forinlab
use
, quality
control
,
and production environments, the
C-Therm
TCiTMThermal
Conductivity
Analyzer
requires no calibration or sample preparation. The system has
broad testing capabilities
(0.0 to 100 W/mK)
across
a wide range of temperatures
(–50
˚ to 200˚C).
The key
capabilities and benefits of the system include the ability to test:
Materials ranging from ~ 0 to 100 W/mK in thermal conductivity or 0
19000 W
s/m2k in
effusivity.
Solids, liquids, powders and pastes
.A wide temperature range from
50 to 200°C.
Accura
cy and precision under 5% and 1% respectively. (Stated from 0
°
C to 50
°
C).
TheC-
Therm TCi
system is engineered to operate in lab environments (including QA/QC labs),
for off
-
line and at
-
line testing. The base technology is migratable to the production
floor and can
be integrated into many manufacturing processes as an in
-
line monitoring and control system
measure
(see C
-
Therm ESP product line).
TheC-
Therm TCi
provides many additional thermo
-
physical properties by indirect calculation,
including:
Diff
usivity
Thermal Conductivity (calculated from Thermal Effusivity)
Heat Capacity, volumetric specific heat
R-value (insulations)
Depth of penetration
A full description of the indirect calculations is provided in Appendix
G.TheC-
Therm TCi
system uses th
e base platform technology developed by Mathis Instruments
Ltd. in 2004/2005 for the pharmaceutical industry, leveraging the solid
-
state sensor technology as
well as associated electronics and improved software.
Keeping true to
the original concept
ofanon-
destructive testing technology, the C
-
Therm TCi
maintains the ability to test
materials
with little to no sample preparation
, and the versatility to
test
small
sample
volumes
(approximately the size of a nickel coin). There are no limitations as to
how
large a sample can be due to the transient method of measurement the system employs.
Testing times are typically within a few seconds and automated testing remains a key
convenience. Dual sensor testing is also available for higher testing throughput, as
are additional
accessories for specific applications.
TheC-
Therm TCi
platform offers modular expansion of the system’s capabilities (e.g. calibration
ranges, sensor upgrades, accessories, etc.), and a flexible software architecture that offers a full,
r
elational database platform for data management
. The 2
.3
software introduces the option for
user calibration, available with the “Max k” module. The factory calibrations provide immediate
and broad testing capability while the user calibration feature of
fers the ability to focus on a
narrower region of interest at fixed temperatures.
This operating manual provides in
-
depth guidance to users in leveraging the
C-Therm TCi
for its
intended use. It is very important that users invest the time in reviewing t
he information provided
within this document for the safe and effective application of this advanced technology. Upon
reviewing the document, should users have any further questions we encourage them to contact
their C
-
Therm representative.
TCi Operator Manual
2
1.1
A note on instr
ument limitations
It should be noted that in applying an instrument such as the C
-
Therm TCi in testing a wide
variety of materials, material types and temperature ranges, careful consideration should be given
to the limitations in performance and/or use o
f this system under certain conditions. In many
cases, the C
-
Therm TCi can replace as many as four (4)
traditional
thermal conductivity
instruments that covered the same claimed range of testing. However, in simplifying thermal
conductivity testing and pr
oviding an instrument to test such a wide material range, there are
some limitations that can be encountered in specific circumstances where a general research tool
like the C
-
Therm TCi will not be appropriate (the proverbial square bolt in a round hole).
As such, every attempt is made to provide up
-
front information on any known limitations of the
system and to provide general recommendations that help to ensure the optimal use of the
instrument.
Examples of performance constraints that may be encounter
ed and some recommended best
practices include:
Contact with some materials is poor due to the material morphology or a process
condition and the fact that the sensor chip surface is not flexible or conforming. C
-
Therm
has provided contact agents for non
-porous solid materials testing and can establish
special calibrations for certain materials that are not compatible with liquid contact
agents.
Like all lab instruments, the C
-
Therm TCi should be used in a consistent manner to
ensure consistent results. The
user’s test methodology is extremely important and it is
recommended users review any new application of the instrument with a C
-
Therm
technical support person.
The C
-
Therm TCi is designed to withstand the full temperature ranges specified, but has
not be
en calibrated for the entire range given the constraints of available calibration
materials and typical requirements for specific Material Groups. Customized calibration of
the instrument is possible, either by a C
-
Therm representative or by a user certifi
ed for
advanced operation of the instrument.
The C
-
Therm TCi is a very sensitive analytical instrument. It is the user’s responsibility to
provide stable environmental conditions in optimizing the performance of the instrument.
As such, the users’ test me
thodology is important to guarantee accurate and consistent
results (
e
.g.thermal chambers used in sub
-
zero temperatures may generate frost
between the sample and sensor, affecting the results. The user needs to compensate
accordingly for these conditions)
.The sensor construction was chosen to meet the optimal combination of test temperature
requirements as well as respecting internal design constraints. Some materials of
construction may be aggressively affected by contact with certain chemicals
it isnecessary to be aware of those chemicals that are known to impact the integrity of the
sensor as well as those that have not been tested or determined through literature
searches to be safe. Appendix C provides a listing to inform users of these issues and
has provided means to test the
least resistant
component, the RTV sealant. Test tabs
can be purchased from C
-
Therm for chemical aggression testing, avoiding undue
damage to the sensor. C
-
Therm will not take responsibility for sensors damaged due to
exposur
e to untested or known to be damaging chemicals.
Testing under compression greater
than90psig
can result in a damaged sensor. Special
testing conditions requiring higher
-
pressure ratings can be investigated on an individual
basis in contacting a C
-
Therm
representative.
The thermal conductivity of the sample material is primarily determined by the direct
measurement of the thermal effusivity (thermal inertia) of the material. In rare circumstances, an
unacceptable error is evident in testing materials wh
ich have a density or heat capacity which are
substantially different than the materials applied in the calibration of the system.
The
Error
TCi Operator Manual
3
Analysis
section
explains these issues and C
-
Therm technical support is available to provide
further insight if r
equired. Accordingly, testing materials which appropriately fit within the proper
Material Group is necessary for accurate results. Material groupings can be further optimized for
the specific needs of a user
-
either by C
-
Therm on a contract basis, or th
rough additional training
and software module extension for user calibration capabilities.
Open discussions are recommended on new projects or applications with a C
-
Therm technical
support representative in determining whether or not the
C-Therm TCi
TM
is
an appropriate
research tool. In general, the standard configuration of the
C-Therm TCi
TM
covers the test
requirements of most users. However, special conditions required by a user can be provided for
with customized solutions specific to their needs. Ple
ase contact C
-
Therm when considering
applications that are not represented in this manual.
Material Testing Summary (taken from
the Testing chapter
):
Material
Type
Minimum
Thickness
Power
Level
Temperature
Range (°C)
Sample Setup &
Preparation
Contact A
gent
Liquids
1 mm
90-50 to 192
Fill 50 mL beaker to
35mL mark. Place
sensor in beaker.
None
Powders
1 mm
90-50 to 192
Fill 50 mL beaker to
30 mL mark. Place
sensor in beaker.
Foams
2 mm
90-50 to 192
Place sample on
sensor. Place weight
on sample.
Polymers
5 mm
90-50 to 192
Place contact agent
on sensor. Place
sample on sensor.
Place weight on
sample.
-50 to 5°C: 3 drops glycol
5 to 70°C: 3 drops water
70 to
200
°C:
Wakefield
120 Thermal Grease
Ceramics
5 mm
90-50 to 192
Metals
5-12 mm
125-50 to 150
1.2
C-Therm TCi
Manual Intent
This manual covers the detailed use of the
C-Therm TCi
including the test methods and
procedures to use for various types of testing. The manual is primarily directed to explain the
operating windows and
menus and the detailed steps for day
-to-
day test setup, display and data
collection.
Special accessories and test methods as well as software modules not part of the standard
operational package are included separately to those who purchase those items,
with appropriate
documentation able to be added to this manual binder as needed. A list of additional modules and
accessories available or planned are located in Appendix A.
TCi Operator Manual
4
1.3
C-Therm TCi
System Setup
1.3.1
Checking the items and their condition
The
standard
C-Therm TCi
system
will
arrive in a
Pelican Case
. Inspect the
case
for shipping
damage and report this immediately to your shipping department to address with the courier
company. The standard shipping items are listed in the shipping list, typically consis
ting of:
Laptop computer
C-Therm TCi
Controller
C-Therm TCi
sensor(s)
C-Therm TCi
Operating manual (software CD inside)
Contact agents
Reference material(s)
C-Therm TCi
sensor base
Power cable and
USB to Serial cable
Pelican Case for transport
1.3.2
Checking
grounded cable
Check that a properly
grounded
cable for your power outlet type is provided. If this is not the
case, contact
C-Therm
or your
C-Therm
distribution agent immediately for resolution.
Do Not
Use the
C-Therm TCi
in an UNGROUNDED State.
1.3.3
Making c
onnections
Make the following connections:
Connect power cable to rear of
C-Therm TCi
and into the wall outlet
Connect power cable to laptop and into the wall outlet
Connect
USB to
serial cable to laptop and to rear of
C-Therm TCi
Connect
C-Therm TCi
sen
sor(s) at rear of
C-Therm TCi
If you only have one sensor,
always use the “Sensor 1” connector location
1.3.4
Starting up the
C-Therm TCi
system
Turn on the laptop
Turn on the
C-Therm TCi
controller
Initiate the
C-Therm TCi
software by selecting the
C-Therm
TCi
software icon.
Proceed to Diagnostics (button located on the main window)
-
Conduct a system check
and material reference check
(refer to
the
System Calibration and Maintenance chapter
).Review the manual for
guidelines on
selecting
calibration
s
, and g
eneral operation.
Review test procedures, appendices and the SOPs.
Proceed to
sample testing
by following
procedures outlined in
the
Testing
chapter
.
TCi Operator Manual
5
2C-
Therm TCi
Software Use
TheC-
Therm TCi
software main window has eight menus,
seven
buttons, test re
sults explorer,
and a status bar in addition to the frame, where most windows within the software are displayed.
2.1
Menu Items
TheC-
Therm TCi
software main menus are located across the top of the main window.
The
File menu
contains:
New Test
Clickin
g this option begins a new test. It has the same effect as clicking the New
Test button on the toolbar.
Open Test
This option opens the Manage Test Results window. Any set of test results can be
opened from there.
Close
This option is only availabl
e when a test or record is open. It closes the active window.
Print
This option opens the printer selection window. It is only available when there is a report
or table open.
Print Preview
This option shows how the current screen would look if prin
ted. It is only
available when there is a report or table open.
Print Setup
This option allows printer options to be selected. It is only available when there is a
report or table open.
Log Off
This option logs the current user out and opens the login
window.
Change Password
This option opens the Change Password window.
Create Archive
This option allows the records in a database to be stored and a new database
created.
TCi Operator Manual
6
Exit
This option closes the software.
The items in the
Manage menu
open t
ables containing records of the selected type.
Projects and Materials
Projects
-
The project describes the common purpose behind a set of tests.
Material Groups
-
A material group contains all materials with a common characteristic.
Materials
-
A materi
al is any substance tested.
Calibration
Calibration Methods
Calibration methods detail the timing parameters used in the
sensor calibrations.
Calibration Results
All calibration results stored in the database and on sensors
connected to the system
are listed in this table.
Calibration Material Groups
Calibration materials groups are sets of materials that
are used to calibrate the sensors for a type of material or value range. Materials tested
using that calibration must be of the same type and
fit within the range of thermal
effusivity or conductivity values given by the calibration materials.
Calibration Materials
Calibration materials are materials that have stable known
thermal values that can be used to calibrate the sensors.
Contact Age
nts
-
Contact agents ensure good contact between the sensor surface and
solid samples.
Instruments
An instrument is a
TCi controller
. The instrument record lists the
controller serial number and the instrument family.
Testing
Test Methods
Test metho
ds specify all parameters (other than timing) of the test. A
test can be run without materials or material groups being defined, but must have a test
method. Timing parameters are set in the calibration method.
Test Results
Test results are displayed i
n the same format when opened as they are
when the test is being performed. Test results can also be found through the test results
explorer.
The
View menu
lists options for the main window. Any of the options can be turned off by clicking
them.
Toolba
r
The toolbar holds the buttons in the
C-Therm TCi
main window.
Status bar
The status bar, located at the bottom of the main window, shows the user logged in
and the database being used.
Test Results Explorer
The test results explorer is located on
the left of the main window and
allows the user to browse the test results by product, material, unit operation, instrument, or user.
The test results explorer also has a test results filter button. Keywords can be entered into the
filter to search for
test results.
TCi Operator Manual
7
The
Reports Menu
lists the different reports that can be generated in the
C-Therm TCi
software
Audit trail Report
The audit trail report lists all activities fitting entered parameters that occur
within the software, such as creating reco
rds, performing tests, importing and exporting.
Test Report
The test report lists all test results fitting entered parameters.
Reference Material Test Report
The reference material test report gives the results of all
reference material tests perfo
rmed fitting entered parameters.
The
Security menu
is not visible to users who do not have the permissions required to perform
any task related to user or user group management, permissions, or system policies.
System Policies
System policies allow adm
inistrators to control user password changes and
database connections.
Manage Users
All users must have unique login IDs so that their actions within the software
can be documented in the audit trail. Permissions can be granted to users that override g
roup
permissions.
Manage User Groups
Membership in a user group grants or limits the permissions a user has
to perform actions within the software.
Manage Permissions
Permission default values can be changed in the Manage Permissions
window.
The
Tools menu
Export
Calibration Results
All calibration results fitting entered parameters are exported to a
chosen file type with all attached calibration material and calibration method records.
Calibration Methods
-
All calibration methods fitting the
entered parameters are
exported to a chosen file type with all attached calibration material records.
Export Calibration Materials
All calibration materials fitting the entered parameters
are exported to a chosen file type.
Test Methods
All test me
thods fitting entered parameters are exported to a chosen file
type with all attached material, project, and calibration method records. Calibration
results must be exported separately.
Test Results
All test results fitting entered parameters are expo
rted to a chosen file
type with all attached test method, material, project, and calibration method records.
Import
Calibration Results
Records from a selected export file are imported.
Calibration Methods
-
Records from a selected export file are im
ported.
Calibration Materials
Records from a selected export file are imported.
Test Methods
Records from a selected export file are imported.
Test Results
Records from a selected export file are imported.
TCi Operator Manual
8
User Calibration
Acquire Data
–The Acquire Data module allows data collection for a selected calibration
method and temperature.
Compute Results
The Compute Results module allows manipulation of calibration
data and calculation of final user calibration results.
Calibration Report
The calibration report contains results for selected user
calibrations.
Diagnostics
The diagnostics window displays the instrument’s status, calibration data, and
board inventory.
Reference Material Test
A reference material test checks the continu
ed validity of the
calibration by comparing the result of the test against the known value of the reference material.
Change Units of Measure
Metric, Imperial, and CGS units are available for use.
Authorization
Calibrations are unlocked by importing
authorization codes from provided xml
files.
User Cal
The User Calibration Wizard allows
a more streamlined approach to user calibration
data collection and result computation.
The
Windows menu
lists all open windows within the software. Clicking one
of the listed windows
brings that window to the front.
The
Help menu
lists various ways to open the help file and software information. Topics can be
opened by clicking any keyword line or topic title.
Contents
The contents page shows the
organizatio
n
of the help file.
Index
The index lists individual topics and keywords in alphabetical order.
Search
The user enters keywords and
a list of topics fitting the entered keywords is
displayed.
About
The about window displays version information a
bout the
C-Therm TCi
software.
2.2
Table Use
Tables can be opened by clicking a button from the Setup window or by selecting an option from
the Manage menu.
TCi Operator Manual
9
The table has File and Audit Trail menus at the top,
nine
buttons beneath, and fields to enter
keyw
ords for searching. The area for table entries takes up the remaining space.
The table is opened with no entries. To display all entries, click the search button. Keywords can
be entered to refine the search.
2.2.1
Menus
The menus are positioned along the up
per left of the main window. Menu options are grayed out
if the user logged in does not have permission to perform the action attached to the menu option.
The Table File menu:
Search
Clicking this option searches the database for all records matching
the search
keywords. If no keywords have been entered, all records will be returned.
Reset Query
Clicking this option clears the keyword fields.
Reset Grid
Clicking this option resets the column widths in the table (grid).
Print
This option opens
the print window to print the table.
Print Preview
This option shows how the table would look printed. There is a print button on
the preview so that the table can be printed from there.
Print Setup
This opens the printer options window.
Copy Grid
to Clipboard
Clicking this option copies the current table to the clipboard. The
table can then be pasted into a spreadsheet program or other document.
Save Rows As
Clicking this option opens a Save window to save the rows as a CSV file, which
can t
hen be imported into MS Excel.
View
Once a record has been selected, click this option to open the record. The record also
opens if it is double
-
clicked.
TCi Operator Manual
10
Edit
Select a record and click the Edit option. The information contained in the record can be
edited and saved.
Add
Clicking the Add option displays an empty record that can be filled and saved.
Delete
Select a record and click this button to delete it. The record is not permanently deleted
but is marked as inactive.
Recover
Recover has
two options.
Show Deleted Records
Once this option is clicked, the table displays only deleted
records.
Recover Selected Records
This option can be used to recover a record once it has
been selected. The Recover button can also be clicked.
Notes
Click this button to attach a note to a record. Notes may be attached to test methods,
materials, and test results.
Set Default Record
Select a record and click this option. The record will automatically be
selected when starting a test.
Close
Clic
king this option closes the table.
The Audit trail menu:
Audit trail for all Records
Clicking this option displays the audit trail table. If the Search
button is clicked, all entries relating to the current table records are displayed.
Audit trail fo
r selected Records
Clicking this option displays the audit trail table. If the Search
button is clicked, all entries pertaining to the selected record are displayed.
2.2.2
Buttons
Search
This button has the same function as the menu option. It searches the
database for
records fitting the entered keywords, or returns all records if no keywords are entered.
View
Once a record has been selected, click this button to open the record. The record also
opens if it is double
-
clicked.
Add
Clicking the Add bu
tton displays an empty record that can be filled and saved.
Copy
Creates a copy of the current record that can be edited and saved.
Edit
Select a record and click the Edit button. The information contained in the record can be
edited and saved.
Del
ete
Select a record and click this button to delete it. The record is not permanently deleted
but is marked as inactive.
Recover
Once a record has been deleted, it can be recovered by setting the table to show
deleted records, selecting the record, a
nd clicking the recover button.
Notes
Click this button to attach a note to a record. Notes may be attached to test methods,
materials, and test results.
TCi Operator Manual
11
Close
Click this button to close the current table.
2.2.3
Table Options
When any row of the table is
right-clicked,
four
options are displayed.
Copy to Clipboard
Clicking this option copies the current table to the clipboard. The table can
then be pasted into a spreadsheet program.
Show all Columns
Clicking this option resets the widths of the colu
mns if they have been
changed and displays any columns that are hidden by default.
Reset Grid
Clicking this option resets the widths of the columns if they have been changed.
Show Audit Trail
for
This Record
Clicking this option displays all audit tr
ail entries for the
selected record.
2.2.4
Using the Parameter Fields to Search
Keywords are used to narrow the search for a particular record or set of records. If only part of a
keyword is known, use
wildcards
(
% or *
)
for the remainder of the word. For exa
mple: for benzyl
compounds, use benz* or benz% to get all results with “benz” as part of the name.
Material groups can be searched by material group name.
Materials can be searched by material group name
and
material name.
Projects can be searched by pr
oject name.
Calibration methods can be searched by calibration method name
and
instrument family.
Calibration results can be searched by sensor serial number, calibration method, date range, and
validity.
Calibration material groups, calibration materia
ls, contact agents, and instruments are all
displayed and cannot be searched.
Test methods can be searched by test method name, project name,
and
calibration method.
Test results can be searched by test ID, date range, test method, project,
product bat
ch, material,
material lot
,
instrument
, and workstation
.
Users can be searched by user name
and
user ID.
User groups can be searched by user group name.
2.3
Record Management
Records can be created, edited, viewed,
copied,
deleted, and recovered through the
tables. The
tables can be accessed through the setup window.
TCi Operator Manual
12
2.3.1
Create a Record
Each individual record type has a different creation process. Record creation is detailed in
the
Record chapter
.
2.3.2
View a Record
Step 1
Open the record’s table.
Select the
record type from the Manage menu or click the record type in the Setup window. For
example, to open a material click Material in the Setup window.
Step 2
Enter keywords into the parameter fields.
Information about ke
ywords is given in the Using the
Parameter Fields to Search section
.
Step 3
Click the Search button or press the Enter key.
A list of records fitting the entered parameters is displayed.
TCi Operator Manual
13
Step 4
Select the desired record.
Step 5
Click the View button.
2.3.3
Edit a Record
Step 1
–Open the record’s table.
Select the record type from the Manage menu or click the record type in the Setup window. For
example, to open a material click Material in the Setup window.
Step 2
Enter keywords into the parameter fields.
Information about ke
ywords is given in the Using the Parameter Fields to Search section.
Step 3
Click the Search button or press the Enter key.
A list of records fitting the entered parameters is displayed.
Step 4
Select the desired record.
Step 5
Click the Edit butt
on.
Step 6
Edit the fields as needed.
Any field with a white or blue background can be changed.
Step 7
Click the Save button.
2.3.4
Delete a Record
Step 1
Open the record’s table.
Select the record type from the Manage menu or click the record type in
the Setup window. For
example, to open a material click Material in the Setup window.
Step 2
Enter keywords into the parameter fields.
Information about keywords is given in the Using the Parameter Fields to Search section.
Step 3
Click the Search b
utton or press the Enter key.
A list of records fitting the entered parameters is displayed.
TCi Operator Manual
14
Step 4
Select the desired record.
Step 5
Click the Delete button.
2.3.5
Recover a Record
Step 1
Open the record’s table.
Select the record type from the Manage
menu or click the record type in the Setup window. For
example, to open a material click Material in the Setup window.
Step 2
Click the File menu in the table.
Step 3
Click the Show Deleted Records option from the Recover submenu.
Step 4
Sele
ct the record to be recovered.
Step 5
Click the Recover button.
Step 6
-
Click the Show Deleted Records option from the Recover submenu to display the
undeleted records.
2.4
Printing
Test reports, tables, records, and results screens can be printed. A prin
ter must be installed
before anything can be printed.
Step 1
-
Click the File menu.
Step 2
-
Click Print.
If the print option is grayed out, the screen cannot be printed.
View a preview of the screen, table, report, or record by clicking the Print Previ
ew option in the
file menu. The screen, table, report, or record can be printed form the preview screen.
TCi Operator Manual
15
Change the print options by clicking Print Setup from the File menu.
Click the Printer button to select a printer (if multiple are available) or
to see the printer's status.
Step 3
Click the OK button.
TCi Operator Manual
16
3
Records
The following section describes how to use and generate the different types of records which
support the data generated from testing.
3.1
Project
A project describes the common purpose beh
ind a set of tests. Using this field allows you to
separate data for easy reference and future quick access. Projects are not necessary to run tests.
A default project is included in the
C-Therm TCi
software and is all that is required to perform a
test.
As more data is collected, the project can be used to filter data.
Step 1
Open the Project table.
Step 2
Click the Add button.
Step 3
Enter the project name.
The project name indicates the purpose or common theme to the tests attached to it.
Step 4
Enter a description (optional).
Step 5
Click the Save button.
3.2
Material Group
A material group contains all materials with a common characteristic. Material groups are not
necessary to run tests. A default material group is included in the
C-Therm TCi
software and is
all that is required to perform a test.
Step 1
Open the material group table.
Step 2
Click the Add button.
TCi Operator Manual
17
Step 3
Enter the material group name.
Use the material group name to indicate the common characteristic of th
e materials to be
attached to the group. For example: lubricants, powders, liquids, or the material name (if different
grades are to be included in the group).
Step 4
Click the Save button.
3.3
Material
A material is any substance tested. All tests on a p
articular material can be easily found by
searching for the material name. Entering a material name is not necessary to run a test
;
however, the value of that test is put at risk if this information is not entered either before or after
the test is conduc
record
must be specified in order to allow
calculation of
indirect values
. A default material name is included in the
C-Therm TCi
software and is all that is
required to perform a test. As more data is collected, the material name can
be used to filter
data.
Step 1
Open the material table.
Step 2
Click the Add button.
Step 3
Enter the material name.
TCi Operator Manual
18
If multiple grades of the material are in use, indicate the grade in the material name. Do not
specify the material lot as t
his is covered separately.
Step 4
Select a material group.
If no group is selected, the material group is set to the default material group by the software.
Step 5
Click the Add Material Values button (optional).
Step 6
Enter material properties (
optional).
Enter a temperature, then the effusivity,
thermal
conductivity,
density, diffusivity, heat capacity,
and the source of the values. Not all values must be entered. When values are entered, indirect
values (diffusivity, heat capacity, R
-
value, a
nd depth of penetration) can be calculated. Heat
capacity and diffusivity can be entered, or calculated if
the density
is entered.
For
diffusivity
, density and heat capacity must be entered.
For
heat capacity
, density must be entered.
For
density
, heat
capacity must be entered.
TheR-
value
and
depth of penetration
can be calculated without additional values.
Step 7
Click the Save button.
3.4
Material Lot
A material lot differentiates between different batches of the same material.
Material lots are
fre
quently created when a test is started but can also be created through a material record.
Step 1
Open the material lot table.
Open the
Manage Material table, and open the record for the material. Click the Show Lots
button.
Step 2
Click the Add butt
on.
Step3–
Enter the lot number.
Each lot number for a material must be unique.
Step4–
Click the Save button.
3.5
Test Method
A test method describes the parameters of a test: what type of measurements to take (thermal
conductivity or effusivity), wh
en to take measurements, how many measurements to take. Unlike
TCi Operator Manual
19
the records discussed previously, the test method
is required
and must be selected before a test
with any parameters other than the default can be performed.
Step 1
Open the test method tab
le.
Step 2
Click the Add button.
Step 3
Enter a name for the test method.
The test method name describes the purpose of the test. Use very broad or specific names as
the tests require.
If
a test method is to be used to monitor thermal conductivit
y of a specific
material in a process, urethane foam after a curing process for example, an appropriate name
would be “Urethane Foam
k after Curing.” If the purpose of the test is more general, powder
effusivity for example, an appropriate name would be
“Powder Effusivity” or ‘General
Powder
Effusivity.”
Step 4
Select a project.
Step 5
Select
a
calibration method.
The method determines the
timing, power and material type
parameters used in the test.
Step 6
Click the Get Ambient Temperature for
Each Measurement checkbox (Optional).
Check this option only if the temperature is expected to be changing during the test.
Step 7
Enter the delay before the first measurement (optional).
If no delay is entered, measurements begin immediately upon the
beginning of the test.
Step 8
Enter the minimum measurement period (optional).
A minimum measurement period is used to limit the measurements to a set cycle. A period of 10
minutes means that each sensor will take its measurement(s) every ten minutes.
If a period is
not specified, the measurements are taken upon completion of the cooling time.
Step 9
Enter the number of measurements to be taken.
If zero is entered, measurements will be taken until the test is stopped by the user.
Step 10
Enter th
e number of sensors to be used.
If zero is entered, any number of sensors can be selected when beginning a test.
TCi Operator Manual
20
Step 11
Enter the number of samples per sensor per measurement.
This is the number of times the sensor fires during a single measurement int
erval.
Step 12
Click the checkbox to enable staggered sampling, if desired.
When staggered sampling is enabled, a set number of sensors fire each rotation until all sensors
have been used. Without staggered sampling all of the sensors fire on a single
rotation.
Step 13
Select the prompts to be displayed.
Step 14
Click the Save button.
3.6
Contact Agents
Three
standard contact agents are currently offered with the
C-Therm TCi
system: Water, Glycol
,
and Wakefield 120 thermal
grease
. Contact agents hav
e correction factors to prevent them from
negatively affecting measurements.
More information on contact agents is given in the testing
chapter.
Step 1
Open the Contact Agent table.
Step 2
Click the Add button.
Step 3
Enter the contact agent na
me.
Step 4
Enter the correction factor equation
A value
.
The equati
on is developed by testing materials of known thermal conductivity
with water and then
with the new contact agent. See the Testing chapter for more information on developing a
contact a
gent.
Step 5
Enter the correction factor equation B value.
Step 6
Click the Save button.
3.7
Notes
Notes can be attached to any other record in the
C-Therm TCi
software. They are accessible
through the Notes button on most records.
Step 1
-
Open a rec
ord.
Notes can be attached to material, project, and test method records.
TCi Operator Manual
21
Step 2
-
Click the Notes button.
The notes table is displayed.
Step 3
-
Click the Add button.
Step 4
-
Enter the note.
The note can be up to 200 characters long.
Step 5
-
Clic
k the Save button.
3.8
Calibration Materials
Calibration materials are materials with known stable thermal properties that are used to build a
calibration curve.
Note:
Calibration Materials are created differently in the User Cal Wizard. Please see User
Ca
librations for more information.
Step 1
Open the Calibration Material table.
Step 2
Click the Add button.
TCi Operator Manual
22
Step 3
Enter the calibration material name.
Step 4
Enter a description for the calibration material (optional).
Step 5
Click the Add
button twice in the Cal Material Values frame.
Step 6
Select Effusivity for the first row.
Step 7
Enter the calculated effusivity value.
The
thermal effusivity value is calculated from the following equation:
CpkEff

(W
s/m
2
K)
Wh
ere k is the thermal conductivity
(W/m
∙K),
ρ is the density
(
kg/m3)
, and Cp is the Heat Capacity
(
J/Kg/K
).Note: 1 J
= 1 W
s
Step 8
Select Conductivity for the second row.
Step 9
Enter the literature thermal conductivity value.
Step 10
Click th
e Save button.
3.9
Calibration Material Groups
Calibration material groups contain at least three different materials of similar properties.
Preloaded groups include liquids and powders, foams, polymers, ceramics, and metals. Th
ese
records cannot be edited.
Any user created calibration material group can be edited in the
normal way.
Note: Permission to create and edit calibration material groups must be granted by the
system administrator.
Step 1
Open the Calibration Material Group table.
Click the Cal
ibration option from the Manage menu, and Calibration Material Group option from
the Calibration submenu.
Step 2
Click the Add button.
TCi Operator Manual
23
Step 3
Enter the Cal Material Group name.
Step 4
Click the + button to add a calibration material to the grou
p.
Step 5
Select a calibration material from the drop
-
down list
.
Step 6
Repeat steps 4 and 5 as needed to enter all calibration materials.
Step 7
Select a diagnostic material from the drop
-
down
list
.
The diagnostic material is the material teste
d when the calibration is selected for a reference
material test.
Step 8
Click the Save button.
3.10
Calibration Methods
Calibration methods contain the parameters used to calibrate sensors, including
timing options
,
power level,
frequency,
and calibration
material group.
Note: Only user created calibration
methods can be edited
.
Calibration methods are
created differently in the User Cal Wizard. Please see User Calibrations for more
information.
Step 1
Open the Calibration Method table.
Click the
Calibration option from the Manage menu, and Calibration Method option from the
Calibration submenu.
Step 2
Click the Add button.
TCi Operator Manual
24
Step3–
Enter a Calibration Method Name.
Step4–
Select an Instrument Family.
This may usually be left as ESP Family
.
Step5–
Select the Calibration Material Group.
Five calibration material groups come standard with the software: liquids and powders, foams,
polymers, ceramics, and metals. Others may be created and changed by users.
Step6–
Click the Ignore Cal
Re
sults from Sensor ID Chip option.
Step7–
Enter the Test Time in seconds.
The test time is the time during which voltage is applied to the sensor.
The standard test time is
0.8s.
Note: If the test time is 5 seconds or greater, the frequency must be de
creased to 50 Hz.
Step8–
Enter the Start Time in seconds.
The start time is the time into a test after which the data is used for calculations.
Prior to that
time, the heat wave is moving through the chip and contact agent.
The standard start time is
0.3s.
Step9–
Enter the Cooling Time in seconds.
The cooling time is the time the sensor requires to cool down after a measurement is taken. No
further measurements can be taken by a sensor during the cooling time.
The standard cooling
time is 60s.
Step10–Select the Frequency.
The frequency must be decreased to 50 Hz if the test time is 5 seconds or more but may remain
at 100 Hz for any shorter test time.
Step11–
Select the Power Level.
The power level is the amount of voltage applied to the sens
or. Use
90
when testing powders,
foams, liquids, polymers, and ceramics and
125
when testing metals.
Step 1
2–Set the Valid Ranges.
Loading...
+ 108 hidden pages