
................................
................................
................................
.........................
A note on instrument limitations
................................
................................
................................
................................
.........................
................................
................................
Checking the items and their condition
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
Using the Parameter Fields to Search
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
............................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
..............................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
Indirect Property Calculations
................................
................................
Altering the Test Method during the Test
................................
................................
Changing the Units of Measurement
................................
................................
................................
................................
................................
................................
................................
................................
Testing with Wakefield 120 Thermal Grease
................................
Chemical Testing with RTV Tabs
................................
................................
................................
................................
................................
................................
................................
Using Excel for added Analysis Power
................................
................................
................................
................................
Overlaying Measurements/Tests
................................
................................
................................
................................
................................

................................
................................
................................
................................
................................
............................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
............................
g a User Calibration in the Wizard
................................
User Calibration Overview
................................
................................
Create Calibration Materials
................................
................................
Create Calibration Material Group
................................
................................
................................
................................
................................
................................
...........................
................................
................................
................................
Software Setup and Maintenance
................................
................................
................................
................................
................................
................................
................................
................................
Manage Users and User Groups
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
Reactivating a Locked Account
................................
................................
................................
................................
................................
................................
................................
System Calibration and Maintenance
................................
................................
................................
................................
................................
Unlock Calibrations/User Calibration
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
General Information required
................................
................................
................................
................................
.............................
................................
................................
................................
................................
................................
................................
Sensor temperature calibration
................................
................................
point effusivity calibration
................................
................................
Material groups and power levels
................................
................................
................................
................................
................................
Thermal conductivity (k) calibration
................................
................................
................................
................................
Temperature considerations in effusivity calibration
................................
Temperature considerations in conductivity calibration
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
................................
.............................
Appendix A: System Components and Accesso
................................
.............................
Therm TCi Controller Specifications
................................
.............................
Therm TCi Sensor Specifications
................................
................................
d Computer Specifications
................................
.............................

Appendix C: Materials of Construction and Chemical Resistance Data
................................
Appendix D: Reference Material Standard Values
................................
...............................
Appendix E: Permission Default Values
................................
................................
................................
................................
................................
Appendix G: Units and Conversions
................................
................................
Appendix H: Contact Agent MSDS
................................
................................
................................
................................
................................

technology expands the capabilities of
destructive thermal testing instrument to a whole new level. Designed to provide
simple, highly accurate thermal characterization
of thermal conductivity and thermal effusivity
and production environments, the
requires no calibration or sample preparation. The system has
broad testing capabilities
a wide range of temperatures
capabilities and benefits of the system include the ability to test:
Materials ranging from ~ 0 to 100 W/mK in thermal conductivity or 0
Solids, liquids, powders and pastes
.A wide temperature range from
cy and precision under 5% and 1% respectively. (Stated from 0
system is engineered to operate in lab environments (including QA/QC labs),
line testing. The base technology is migratable to the production
be integrated into many manufacturing processes as an in
line monitoring and control system
provides many additional thermo
physical properties by indirect calculation,
Thermal Conductivity (calculated from Thermal Effusivity)
Heat Capacity, volumetric specific heat
A full description of the indirect calculations is provided in Appendix
e base platform technology developed by Mathis Instruments
Ltd. in 2004/2005 for the pharmaceutical industry, leveraging the solid
state sensor technology as
well as associated electronics and improved software.
destructive testing technology, the C
maintains the ability to test
with little to no sample preparation
(approximately the size of a nickel coin). There are no limitations as to
large a sample can be due to the transient method of measurement the system employs.
Testing times are typically within a few seconds and automated testing remains a key
convenience. Dual sensor testing is also available for higher testing throughput, as
accessories for specific applications.
platform offers modular expansion of the system’s capabilities (e.g. calibration
ranges, sensor upgrades, accessories, etc.), and a flexible software architecture that offers a full,
elational database platform for data management
software introduces the option for
user calibration, available with the “Max k” module. The factory calibrations provide immediate
and broad testing capability while the user calibration feature of
fers the ability to focus on a
narrower region of interest at fixed temperatures.
This operating manual provides in
depth guidance to users in leveraging the
intended use. It is very important that users invest the time in reviewing t
within this document for the safe and effective application of this advanced technology. Upon
reviewing the document, should users have any further questions we encourage them to contact

It should be noted that in applying an instrument such as the C
Therm TCi in testing a wide
variety of materials, material types and temperature ranges, careful consideration should be given
to the limitations in performance and/or use o
f this system under certain conditions. In many
Therm TCi can replace as many as four (4)
instruments that covered the same claimed range of testing. However, in simplifying thermal
conductivity testing and pr
oviding an instrument to test such a wide material range, there are
some limitations that can be encountered in specific circumstances where a general research tool
Therm TCi will not be appropriate (the proverbial square bolt in a round hole).
As such, every attempt is made to provide up
front information on any known limitations of the
system and to provide general recommendations that help to ensure the optimal use of the
Examples of performance constraints that may be encounter
ed and some recommended best
Contact with some materials is poor due to the material morphology or a process
condition and the fact that the sensor chip surface is not flexible or conforming. C
has provided contact agents for non
-porous solid materials testing and can establish
special calibrations for certain materials that are not compatible with liquid contact
Like all lab instruments, the C
Therm TCi should be used in a consistent manner to
ensure consistent results. The
user’s test methodology is extremely important and it is
recommended users review any new application of the instrument with a C
technical support person.
Therm TCi is designed to withstand the full temperature ranges specified, but has
en calibrated for the entire range given the constraints of available calibration
materials and typical requirements for specific Material Groups. Customized calibration of
the instrument is possible, either by a C
Therm representative or by a user certifi
advanced operation of the instrument.
Therm TCi is a very sensitive analytical instrument. It is the user’s responsibility to
provide stable environmental conditions in optimizing the performance of the instrument.
As such, the users’ test me
thodology is important to guarantee accurate and consistent
.g.thermal chambers used in sub
zero temperatures may generate frost
between the sample and sensor, affecting the results. The user needs to compensate
accordingly for these conditions)
.The sensor construction was chosen to meet the optimal combination of test temperature
requirements as well as respecting internal design constraints. Some materials of
construction may be aggressively affected by contact with certain chemicals
it isnecessary to be aware of those chemicals that are known to impact the integrity of the
sensor as well as those that have not been tested or determined through literature
searches to be safe. Appendix C provides a listing to inform users of these issues and
has provided means to test the
component, the RTV sealant. Test tabs
Therm for chemical aggression testing, avoiding undue
Therm will not take responsibility for sensors damaged due to
e to untested or known to be damaging chemicals.
Testing under compression greater
can result in a damaged sensor. Special
testing conditions requiring higher
pressure ratings can be investigated on an individual
The thermal conductivity of the sample material is primarily determined by the direct
measurement of the thermal effusivity (thermal inertia) of the material. In rare circumstances, an
unacceptable error is evident in testing materials wh
ich have a density or heat capacity which are
substantially different than the materials applied in the calibration of the system.

explains these issues and C
Therm technical support is available to provide
equired. Accordingly, testing materials which appropriately fit within the proper
Material Group is necessary for accurate results. Material groupings can be further optimized for
the specific needs of a user
Therm on a contract basis, or th
rough additional training
and software module extension for user calibration capabilities.
Open discussions are recommended on new projects or applications with a C
support representative in determining whether or not the
research tool. In general, the standard configuration of the
requirements of most users. However, special conditions required by a user can be provided for
with customized solutions specific to their needs. Ple
applications that are not represented in this manual.
Material Testing Summary (taken from
-50 to 5°C: 3 drops glycol
This manual covers the detailed use of the
including the test methods and
procedures to use for various types of testing. The manual is primarily directed to explain the
menus and the detailed steps for day
day test setup, display and data
Special accessories and test methods as well as software modules not part of the standard
operational package are included separately to those who purchase those items,
documentation able to be added to this manual binder as needed. A list of additional modules and
accessories available or planned are located in Appendix A.

Checking the items and their condition
damage and report this immediately to your shipping department to address with the courier
company. The standard shipping items are listed in the shipping list, typically consis
Operating manual (software CD inside)
Pelican Case for transport
cable for your power outlet type is provided. If this is not the
distribution agent immediately for resolution.
Make the following connections:
Connect power cable to rear of
Connect power cable to laptop and into the wall outlet
serial cable to laptop and to rear of
If you only have one sensor,
always use the “Sensor 1” connector location
software by selecting the
Proceed to Diagnostics (button located on the main window)
and material reference check
System Calibration and Maintenance chapter
Review test procedures, appendices and the SOPs.

software main window has eight menus,
and a status bar in addition to the frame, where most windows within the software are displayed.
software main menus are located across the top of the main window.
g this option begins a new test. It has the same effect as clicking the New
Test button on the toolbar.
This option opens the Manage Test Results window. Any set of test results can be
This option is only availabl
e when a test or record is open. It closes the active window.
This option opens the printer selection window. It is only available when there is a report
This option shows how the current screen would look if prin
available when there is a report or table open.
This option allows printer options to be selected. It is only available when there is a
This option logs the current user out and opens the login
This option opens the Change Password window.
This option allows the records in a database to be stored and a new database

This option closes the software.
ables containing records of the selected type.
The project describes the common purpose behind a set of tests.
A material group contains all materials with a common characteristic.
al is any substance tested.
Calibration methods detail the timing parameters used in the
All calibration results stored in the database and on sensors
are listed in this table.
Calibration Material Groups
Calibration materials groups are sets of materials that
are used to calibrate the sensors for a type of material or value range. Materials tested
using that calibration must be of the same type and
fit within the range of thermal
effusivity or conductivity values given by the calibration materials.
Calibration materials are materials that have stable known
thermal values that can be used to calibrate the sensors.
Contact agents ensure good contact between the sensor surface and
. The instrument record lists the
controller serial number and the instrument family.
ds specify all parameters (other than timing) of the test. A
test can be run without materials or material groups being defined, but must have a test
method. Timing parameters are set in the calibration method.
Test results are displayed i
n the same format when opened as they are
when the test is being performed. Test results can also be found through the test results
lists options for the main window. Any of the options can be turned off by clicking
The toolbar holds the buttons in the
The status bar, located at the bottom of the main window, shows the user logged in
and the database being used.
The test results explorer is located on
the left of the main window and
allows the user to browse the test results by product, material, unit operation, instrument, or user.
The test results explorer also has a test results filter button. Keywords can be entered into the

lists the different reports that can be generated in the
The audit trail report lists all activities fitting entered parameters that occur
within the software, such as creating reco
rds, performing tests, importing and exporting.
The test report lists all test results fitting entered parameters.
Reference Material Test Report
The reference material test report gives the results of all
reference material tests perfo
rmed fitting entered parameters.
is not visible to users who do not have the permissions required to perform
any task related to user or user group management, permissions, or system policies.
System policies allow adm
inistrators to control user password changes and
All users must have unique login IDs so that their actions within the software
can be documented in the audit trail. Permissions can be granted to users that override g
Membership in a user group grants or limits the permissions a user has
to perform actions within the software.
Permission default values can be changed in the Manage Permissions
All calibration results fitting entered parameters are exported to a
chosen file type with all attached calibration material and calibration method records.
All calibration methods fitting the
exported to a chosen file type with all attached calibration material records.
Export Calibration Materials
All calibration materials fitting the entered parameters
are exported to a chosen file type.
thods fitting entered parameters are exported to a chosen file
type with all attached material, project, and calibration method records. Calibration
results must be exported separately.
All test results fitting entered parameters are expo
type with all attached test method, material, project, and calibration method records.
Records from a selected export file are imported.
Records from a selected export file are im
Records from a selected export file are imported.
Records from a selected export file are imported.
Records from a selected export file are imported.

–The Acquire Data module allows data collection for a selected calibration
The Compute Results module allows manipulation of calibration
data and calculation of final user calibration results.
The calibration report contains results for selected user
The diagnostics window displays the instrument’s status, calibration data, and
A reference material test checks the continu
calibration by comparing the result of the test against the known value of the reference material.
Metric, Imperial, and CGS units are available for use.
Calibrations are unlocked by importing
authorization codes from provided xml
The User Calibration Wizard allows
a more streamlined approach to user calibration
data collection and result computation.
lists all open windows within the software. Clicking one
brings that window to the front.
lists various ways to open the help file and software information. Topics can be
opened by clicking any keyword line or topic title.
The contents page shows the
The index lists individual topics and keywords in alphabetical order.
The user enters keywords and
a list of topics fitting the entered keywords is
The about window displays version information a
Tables can be opened by clicking a button from the Setup window or by selecting an option from

The table has File and Audit Trail menus at the top,
buttons beneath, and fields to enter
ords for searching. The area for table entries takes up the remaining space.
The table is opened with no entries. To display all entries, click the search button. Keywords can
be entered to refine the search.
The menus are positioned along the up
per left of the main window. Menu options are grayed out
if the user logged in does not have permission to perform the action attached to the menu option.
Clicking this option searches the database for all records matching
keywords. If no keywords have been entered, all records will be returned.
Clicking this option clears the keyword fields.
Clicking this option resets the column widths in the table (grid).
the print window to print the table.
This option shows how the table would look printed. There is a print button on
the preview so that the table can be printed from there.
This opens the printer options window.
Clicking this option copies the current table to the clipboard. The
table can then be pasted into a spreadsheet program or other document.
Clicking this option opens a Save window to save the rows as a CSV file, which
hen be imported into MS Excel.
Once a record has been selected, click this option to open the record. The record also

Select a record and click the Edit option. The information contained in the record can be
Clicking the Add option displays an empty record that can be filled and saved.
Select a record and click this button to delete it. The record is not permanently deleted
but is marked as inactive.
Once this option is clicked, the table displays only deleted
This option can be used to recover a record once it has
been selected. The Recover button can also be clicked.
Click this button to attach a note to a record. Notes may be attached to test methods,
materials, and test results.
Select a record and click this option. The record will automatically be
selected when starting a test.
king this option closes the table.
Audit trail for all Records
Clicking this option displays the audit trail table. If the Search
button is clicked, all entries relating to the current table records are displayed.
Clicking this option displays the audit trail table. If the Search
button is clicked, all entries pertaining to the selected record are displayed.
This button has the same function as the menu option. It searches the
records fitting the entered keywords, or returns all records if no keywords are entered.
Once a record has been selected, click this button to open the record. The record also
tton displays an empty record that can be filled and saved.
Creates a copy of the current record that can be edited and saved.
Select a record and click the Edit button. The information contained in the record can be
Select a record and click this button to delete it. The record is not permanently deleted
but is marked as inactive.
Once a record has been deleted, it can be recovered by setting the table to show
deleted records, selecting the record, a
nd clicking the recover button.
Click this button to attach a note to a record. Notes may be attached to test methods,
materials, and test results.

Click this button to close the current table.
When any row of the table is
Clicking this option copies the current table to the clipboard. The table can
then be pasted into a spreadsheet program.
Clicking this option resets the widths of the colu
changed and displays any columns that are hidden by default.
Clicking this option resets the widths of the columns if they have been changed.
Clicking this option displays all audit tr
Using the Parameter Fields to Search
Keywords are used to narrow the search for a particular record or set of records. If only part of a
for the remainder of the word. For exa
compounds, use benz* or benz% to get all results with “benz” as part of the name.
Material groups can be searched by material group name.
Materials can be searched by material group name
Projects can be searched by pr
Calibration methods can be searched by calibration method name
Calibration results can be searched by sensor serial number, calibration method, date range, and
Calibration material groups, calibration materia
ls, contact agents, and instruments are all
displayed and cannot be searched.
Test methods can be searched by test method name, project name,
Test results can be searched by test ID, date range, test method, project,
Users can be searched by user name
User groups can be searched by user group name.
Records can be created, edited, viewed,
deleted, and recovered through the
tables can be accessed through the setup window.

Select the desired record.
–Open the record’s table.
Select the record type from the Manage menu or click the record type in the Setup window. For
example, to open a material click Material in the Setup window.
Enter keywords into the parameter fields.
ywords is given in the Using the Parameter Fields to Search section.
Click the Search button or press the Enter key.
A list of records fitting the entered parameters is displayed.
Select the desired record.
Edit the fields as needed.
Any field with a white or blue background can be changed.
Select the record type from the Manage menu or click the record type in
example, to open a material click Material in the Setup window.
Enter keywords into the parameter fields.
Information about keywords is given in the Using the Parameter Fields to Search section.
utton or press the Enter key.
A list of records fitting the entered parameters is displayed.

Select the desired record.
Select the record type from the Manage
menu or click the record type in the Setup window. For
example, to open a material click Material in the Setup window.
Click the File menu in the table.
Click the Show Deleted Records option from the Recover submenu.
ct the record to be recovered.
Click the Recover button.
Click the Show Deleted Records option from the Recover submenu to display the
Test reports, tables, records, and results screens can be printed. A prin
before anything can be printed.
If the print option is grayed out, the screen cannot be printed.
View a preview of the screen, table, report, or record by clicking the Print Previ
file menu. The screen, table, report, or record can be printed form the preview screen.

The following section describes how to use and generate the different types of records which
support the data generated from testing.
A project describes the common purpose beh
ind a set of tests. Using this field allows you to
separate data for easy reference and future quick access. Projects are not necessary to run tests.
A default project is included in the
software and is all that is required to perform a
As more data is collected, the project can be used to filter data.
The project name indicates the purpose or common theme to the tests attached to it.
Enter a description (optional).
A material group contains all materials with a common characteristic. Material groups are not
necessary to run tests. A default material group is included in the
all that is required to perform a test.
Open the material group table.

Enter the material group name.
Use the material group name to indicate the common characteristic of th
attached to the group. For example: lubricants, powders, liquids, or the material name (if different
grades are to be included in the group).
A material is any substance tested. All tests on a p
articular material can be easily found by
searching for the material name. Entering a material name is not necessary to run a test
however, the value of that test is put at risk if this information is not entered either before or after
must be specified in order to allow
. A default material name is included in the
software and is all that is
required to perform a test. As more data is collected, the material name can

If multiple grades of the material are in use, indicate the grade in the material name. Do not
specify the material lot as t
his is covered separately.
If no group is selected, the material group is set to the default material group by the software.
Click the Add Material Values button (optional).
Enter material properties (
Enter a temperature, then the effusivity,
density, diffusivity, heat capacity,
and the source of the values. Not all values must be entered. When values are entered, indirect
values (diffusivity, heat capacity, R
nd depth of penetration) can be calculated. Heat
capacity and diffusivity can be entered, or calculated if
, density and heat capacity must be entered.
, density must be entered.
capacity must be entered.
can be calculated without additional values.
A material lot differentiates between different batches of the same material.
quently created when a test is started but can also be created through a material record.
Open the material lot table.
Manage Material table, and open the record for the material. Click the Show Lots
Each lot number for a material must be unique.
A test method describes the parameters of a test: what type of measurements to take (thermal
conductivity or effusivity), wh
en to take measurements, how many measurements to take. Unlike

the records discussed previously, the test method
and must be selected before a test
with any parameters other than the default can be performed.
Enter a name for the test method.
The test method name describes the purpose of the test. Use very broad or specific names as
a test method is to be used to monitor thermal conductivit
material in a process, urethane foam after a curing process for example, an appropriate name
k after Curing.” If the purpose of the test is more general, powder
effusivity for example, an appropriate name would be
“Powder Effusivity” or ‘General
The method determines the
timing, power and material type
parameters used in the test.
Click the Get Ambient Temperature for
Each Measurement checkbox (Optional).
Check this option only if the temperature is expected to be changing during the test.
Enter the delay before the first measurement (optional).
If no delay is entered, measurements begin immediately upon the
Enter the minimum measurement period (optional).
A minimum measurement period is used to limit the measurements to a set cycle. A period of 10
minutes means that each sensor will take its measurement(s) every ten minutes.
not specified, the measurements are taken upon completion of the cooling time.
Enter the number of measurements to be taken.
If zero is entered, measurements will be taken until the test is stopped by the user.
e number of sensors to be used.
If zero is entered, any number of sensors can be selected when beginning a test.

Enter the number of samples per sensor per measurement.
This is the number of times the sensor fires during a single measurement int
Click the checkbox to enable staggered sampling, if desired.
When staggered sampling is enabled, a set number of sensors fire each rotation until all sensors
have been used. Without staggered sampling all of the sensors fire on a single
Select the prompts to be displayed.
standard contact agents are currently offered with the
and Wakefield 120 thermal
e correction factors to prevent them from
negatively affecting measurements.
More information on contact agents is given in the testing
Open the Contact Agent table.
Enter the contact agent na
Enter the correction factor equation
on is developed by testing materials of known thermal conductivity
with the new contact agent. See the Testing chapter for more information on developing a
Enter the correction factor equation B value.
Notes can be attached to any other record in the
software. They are accessible
through the Notes button on most records.
Notes can be attached to material, project, and test method records.

Enter the calibration material name.
Enter a description for the calibration material (optional).
button twice in the Cal Material Values frame.
Select Effusivity for the first row.
Enter the calculated effusivity value.
thermal effusivity value is calculated from the following equation:
ere k is the thermal conductivity
, and Cp is the Heat Capacity
Select Conductivity for the second row.
Enter the literature thermal conductivity value.
Calibration Material Groups
Calibration material groups contain at least three different materials of similar properties.
Preloaded groups include liquids and powders, foams, polymers, ceramics, and metals. Th
records cannot be edited.
Any user created calibration material group can be edited in the
Note: Permission to create and edit calibration material groups must be granted by the
Open the Calibration Material Group table.
ibration option from the Manage menu, and Calibration Material Group option from

Enter the Cal Material Group name.
Click the + button to add a calibration material to the grou
Select a calibration material from the drop
Repeat steps 4 and 5 as needed to enter all calibration materials.
Select a diagnostic material from the drop
The diagnostic material is the material teste
d when the calibration is selected for a reference
Calibration methods contain the parameters used to calibrate sensors, including
Note: Only user created calibration
created differently in the User Cal Wizard. Please see User Calibrations for more
Open the Calibration Method table.
Calibration option from the Manage menu, and Calibration Method option from the

Enter a Calibration Method Name.
Select an Instrument Family.
This may usually be left as ESP Family
Select the Calibration Material Group.
Five calibration material groups come standard with the software: liquids and powders, foams,
polymers, ceramics, and metals. Others may be created and changed by users.
sults from Sensor ID Chip option.
Enter the Test Time in seconds.
The test time is the time during which voltage is applied to the sensor.
The standard test time is
Note: If the test time is 5 seconds or greater, the frequency must be de
Enter the Start Time in seconds.
The start time is the time into a test after which the data is used for calculations.
time, the heat wave is moving through the chip and contact agent.
The standard start time is
Enter the Cooling Time in seconds.
The cooling time is the time the sensor requires to cool down after a measurement is taken. No
further measurements can be taken by a sensor during the cooling time.
Step10–Select the Frequency.
The frequency must be decreased to 50 Hz if the test time is 5 seconds or more but may remain
at 100 Hz for any shorter test time.
The power level is the amount of voltage applied to the sens
foams, liquids, polymers, and ceramics and