C-Squared Hydrogen Analyzer User Manual

C-Squared Inc. 1440 S State College Blvd, Unit 4L, Anaheim, CA 92806
Tel: (714) 999-2791, Fax: (714) 999-2928, sales@c-squaredinc.com, www.c-squaredinc.com
Unpacking and Inspection
The user should always inspect the shipping package for damage. If there is damage evident, please save the package and notify the shipping company immediately.
The Hydrogen Analyzer is a hand-held device for measuring Hydrogen content in a gas mixture, and is specifically designed to confirm the Hydrogen content in a gas mixture similar to normal air or nitrogen. The Analyzer is powered by rechargeable batteries. (Figure 1 shows the top panel.)
Fig. 1. Unit top panel view.
The LCD displays the Hydrogen content from 0 to 100% with 0.1% resolution. When the low battery indicator LED is on, the unit’s battery needs to be recharged prior to use. The patented sample unit allows the user to flow a gas mixture through the sensor inside the unit, as well as to seal the gas mixture in the sensor for measurement.
Incorporated
Hydrogen Analyzer
Discrete Sampler
Instruction Manual
Hydrogen Discrete Analyzer Instruction Manual Page 2
C-Squared Inc. 1440 S State College Blvd, Unit 4L, Anaheim, CA 92806
Tel: (714) 999-2791, Fax: (714) 999-2928, sales@c-squaredinc.com, www.c-squaredinc.com
Operation
The operation of the unit is very simple and does not require a flow control regulator. The operation is as simple as 1, 2, 3.
Step 1. Pull up the sensor cap and flow gas mixture through the gas inlet with a steady flow for about 10 seconds. (Enough time to clear the line of any existing gas).
Step 2. Push the sensor cap down firmly to stop the gas flow and seal the gas mixture inside the device.
Step 3. Turn on the power switch and wait for the display to give a stable reading (less than one minute), which is the percentage content of helium in the gas mixture tested.
Note: Always turn off the power after you take the reading to extend the battery life between charging.
Gas
Pull
Loading...
+ 1 hidden pages