CS Instruments LD 500, LD 510 Instruction Manual

EN- English
LD500 V1.00 Seite 1 von 68
Instruction manual
Leak meter with camera
Table of Content
LD500 V1.00 Seite 2 von 68
1 Table of Content
2 Foreword ............................................................................................................................................... 4
3 Safety instructions ................................................................................................................................ 5
3.1 About this document ........................................................................................................................................................................... 5
3.2 Ensuring safety .................................................................................................................................................................................... 5
3.3 Environmental protection .................................................................................................................................................................... 5
4 General function description ................................................................................................................ 6
5 Technical data LD500 ............................................................................................................................ 7
6 Device components and controls ......................................................................................................... 8
6.1 LD 500 ................................................................................................................................................................................................. 8
6.2 Pre Amplifier module ........................................................................................................................................................................... 9
6.3 Acoustic trumpet with camera ............................................................................................................................................................ 9
6.4 Focus tube with focus tip ................................................................................................................................................................... 10
6.5 Gooseneck (Optional) ........................................................................................................................................................................ 10
6.6 Assembly with acoustic trumpet ....................................................................................................................................................... 11
6.7 Assembly with focus tube with focus tip ........................................................................................................................................... 11
7 Commissioning / Applikation LD 500 or LD510 .................................................................................. 12
7.1 Switch on ........................................................................................................................................................................................... 12
7.2 Headphone Volume Loud / Volume Down ........................................................................................................................................ 12
7.3 Sensitivity level .................................................................................................................................................................................. 12
7.4 Procedure leak detection / measurement ......................................................................................................................................... 13
8 Operation ............................................................................................................................................ 14
8.1 Initialization ....................................................................................................................................................................................... 14
8.2 Screen Leckage .................................................................................................................................................................................. 15
8.3 The following picture shows and describes the display elements. .................................................................................................... 15
9 Settings ................................................................................................................................................ 16
9.1 Sensitivity settings ............................................................................................................................................................................. 16
9.2 Laser On/Off ...................................................................................................................................................................................... 16
9.3 Parameters for measurement ........................................................................................................................................................... 17
9.4 Storing oft he measurement .............................................................................................................................................................. 18
9.4.1 Measuring point designation / selection .............................................................................................................................. 19
9.4.2 Parameter of measurement (Re-Check) ............................................................................................................................... 19
9.4.3 Comment .............................................................................................................................................................................. 20
9.4.4 Storing measurement data to internal SD-card .................................................................................................................... 20
9.5 ZeroPoint adjustment ........................................................................................................................................................................ 21
10 Basic settings menu LD 500 ................................................................................................................ 22
10.1 Configuration ..................................................................................................................................................................................... 22
10.2 Export/Import .................................................................................................................................................................................... 23
10.2.1 Export „Journal Data“ ........................................................................................................................................................... 24
10.2.2 Export of System settings ..................................................................................................................................................... 25
10.2.3 Import of system settings ..................................................................................................................................................... 26
10.2.4 Export / import customer database ..................................................................................................................................... 27
Table of Content
LD500 V1.00 Seite 3 von 68
10.3 View bitmaps ..................................................................................................................................................................................... 28
10.4 Device Settings .................................................................................................................................................................................. 29
10.4.1 Password settings ................................................................................................................................................................. 29
10.4.2 Device Settings ..................................................................................................................................................................... 30
10.4.3 Set backlight brightness ........................................................................................................................................................ 35
10.4.4 Cleaning ................................................................................................................................................................................ 36
10.4.5 System-Status ....................................................................................................................................................................... 36
10.4.6 About LD 500/LD510 ............................................................................................................................................................ 36
11 Charging the batteries ........................................................................................................................ 37
11.1 Protection of exhaustive discharge ................................................................................................................................................... 37
12 LD 510 ................................................................................................................................................. 38
12.1 Selection External sensor ................................................................................................................................................................... 38
12.2 Input signals ext. sensor LD510 ......................................................................................................................................................... 39
12.3 Cable cross section ............................................................................................................................................................................ 39
12.3.1 Sensor circuit points/Output signal: ..................................................................................................................................... 39
12.4 Connection diagrams of the different sensor types .......................................................................................................................... 40
12.4.1 Connector pin assignment for all sensors at PI 500 ............................................................................................................. 40
12.4.2 Connection for CS dew point- and consumption sensors, series FA/VA 5xx ........................................................................ 41
12.4.3 Connection with RS485 ........................................................................................................................................................ 41
12.4.4 Three- and four-wire power supply 0 - 1/10/30 VDC ........................................................................................................... 41
12.4.5 Analogue two-, three-, and four-wire current signal ............................................................................................................ 42
12.4.6 Two-, three- and four-wire connector pin assignments for PT100/PT1000/KTY81 ............................................................. 43
12.5 Dew Point Sensor FA 500 / FA 510 of type FA 5xx (RS 485 Modbus) ................................................................................................. 44
12.5.1 Settings Dew point sensor FA 500 FA 510 ............................................................................................................................ 45
12.6 Flow sensor of type VA 5xx (RS 485 Modbus) .................................................................................................................................... 48
12.6.1 Settings for Flow sensor VA 5xx ............................................................................................................................................ 49
12.7 Type Modbus ..................................................................................................................................................................................... 57
12.7.1 Selection and activation of Sensor-Type Modbus ................................................................................................................ 57
12.8 Data logger Settings ........................................................................................................................................................................... 61
13 Scope of delivery ................................................................................................................................. 65
Foreword
LD500 V1.00 Seite 4 von 68
2 Foreword
Dear Customer,
thank you for purchasing our leak meter with camera LD 500. The new leak meter LD 500 with integrated camera and leakage calculation are ideal measuring instruments which help to find and document even smallest leakages (0.1 l/min corresponds to approx. 1 € p. a.) easily even in far distances.
The LD 510 is the worldwide first leak meter with an additional freely assignable sensor input for all CS sensors. In addition to the leakage measurement and detection also all necessary measurements with regards to dew point, flow, pressure, and temperature ... can be carried out
Main functions:
Tracking and location of leaks
compressed air, gas, steam and vacuum systems condensate drain seals refrigeration systems
Documentation / storage of leaks with
Image of the leak position Date / Time Description of the leakage position with indication of company / department or hall /
machine
Size of the leak in liters / min (units adjustable) Leakage costs per year in € (currency freely definable)
Remark: By means of the additional available CS leak Reporter (Order No .: 0554 0105)
detailed reports with summary totals, subtotals (departments / warehouses etc)) as well as history reports (for temporal / continuous improvements) could be created.
Safety instructions
LD500 V1.00 Seite 5 von 68
3 Safety instructions
3.1 About this document
Read through carefully this documentation and familiarize yourself with the
product before putting it to use. Pay particular attention to the safety warnings to prevent injury and product damage.
Keep this documentation to hand for easy reference when needed. Pass on this documentation to any subsequent users of the product.
3.2 Ensuring safety
Only use the product as intended and within the parameters specified in the
technical data. Do not use force for operating..
Never measure with the device at or near live/energized parts!
During leak detection on electrical systems, please maintain a sufficient safety distance to avoid dangerous electric shocks!
Avoid any direct contact with hot and/or rotating parts. Always switch on the device before putting on the headphones! At high signal
levels (bar graph headphones in the red area), the volume can be correspondingly large. The sensitivity setting can be used to reduce the volume.
Never point the laser directly into the eyes! Absolutely devoid a direct irradiation
of the eyes of humans and animals! Laser module: corresponds to DIN EN 60825-1: 2015-07 Class 2 (<1mW)
Observe the prescribed storage and operating temperatures. Improper handling or violence will void the warranty. Any kind of interventions on the device, as far as they do not correspond to the
intended and described procedures, lead to the warranty expiration and to the disclaimer.
The device is intended solely for the described purpose.
3.3 Environmental protection
Disposal of faulty rechargeable batteries / empty batteries in accordance with
applicable legal regulations
Lead back the product after the end of the period of use to the separate
collection for electric and electronic devices (observe local regulations) or return the product to CS Instruments GmbH & Co.KG for disposal.
CS Instruments GmbH & Co.KG makes no warranty as to its suitability for any particular purpose and assumes no liability for any errors contained in this manual. Nor for consequential damages in connection with the delivery, performance or use of this device.
General function description
LD500 V1.00 Seite 6 von 68
4 General function description
When gases escape from leaks in piping systems (leaking screw connections, corrosion, etc.), noises are generated in the ultrasonic range. With the LD 500 even the smallest leaks, which are inaudible to the human ear and not visible due to their size, can be located several meters away.
The inaudible ultrasound is converted to audible frequencies in addition to the display emission level shown in the display. With the convenient, sound-proof headphones, these sounds can be heard even in noisy environments.
In addition, the new LD500 device calculates the costs associated with leaks, providing additional transparency about the state of the system under test or the potential cost savings. The loss is displayed in l / min as well as in a freely selectable currency. The cost per liter or per cubic meter of compressed air can be stored in the device.
The professional measuring instrument LD500 finds typical application in leak detection in compressed air systems and leak testing of pressure less systems.
With the help of an integrated laser pointer, which serves as a targeting, the leak can be pinpointed.
Depending on the leakage, the appropriate accessories may be used to increase the sensitivity of the LD500 to use, available accessories are:
Acoustic trumpet Focus tube with focus tip Gooseneck Parabolic mirrorarabol
Technical data LD500
LD500 V1.00 Seite 7 von 68
5 Technical data LD500
Dimensions
263 x 96 x 280 mm (incl. Pre-Amp module and acoustic trumpet)
Weight
0,55 kg incl. Pre-Amp module and acoustic trumpet , complete set with transportation case ca.3,0 kg
Frequency range
40kHz (+/- 2kHz)
Power supply
Internal 7.4 V lithium-ion battery
Operating time
> 9 h (continuous operation)
Operating temperature
-5 °C to +40 °C
Charging
Ext. battery charger (included in the scope of delivery)
Charging time
approx. 1.5 h
Storage temperature
-20 °C to +50 °C
Laser
Wavelength 645-660nm, output < 1mW (Laser class 2)
Connections
3.5 mm jack for headphones, power jack for connecting an external charger USB Connection
Display
3.5-Touchpanel TFT transmissiv
Interface
USB for data export / -import, SW update etc.
Datalogger
4 GB-Memory card (Micro SD Class 4)
Sensitivity
min: 0,1l/min at 6bar / 5m Distance
Device components and controls
LD500 V1.00 Seite 8 von 68
6 Device components and controls
6.1 LD 500
Picture 1
Camera
Power on/off button
Loudness buttons (+/-)
USB connection
3,5“ TFT-Color Display
with Touchpanel
Foil keypad (4 buttons)
Battery charge socket
Acoustic trumpet
Pre Amplifier module
Release button for preamplifier module
Store button
Device components and controls
LD500 V1.00 Seite 9 von 68
Picture 2
6.2 Pre Amplifier module
Picture 3
6.3 Acoustic trumpet with camera
Picture 4
3.5mm jack for the headphone
Opening of the laser pointer
Ultrasonic microphone
Camera connection
Camera connection
Ultrasonic microphone opening
Laser module
Device components and controls
LD500 V1.00 Seite 10 von 68
6.4 Focus tube with focus tip
Picture 5
6.5 Gooseneck (Optional)
Picture 6
Device components and controls
LD500 V1.00 Seite 11 von 68
6.6 Assembly with acoustic trumpet
The acoustic trumpet allows acoustic amplification by bundling the sound waves and specifies the location of the leak. Due to the special construction of the integrated laser pointer is still usable. The camera is integrated on the bottom of the acoustic trumpet and is electrically connected to the preamplifier module via the jack plug. Assembling is done by plugging the individual components until easy locking audible (plug in to the stop). The components are removed in the reverse order; for unlocking the preamplifier module, the release button must also be pressed.
Picture 7
6.7 Assembly with focus tube with focus tip
The focus tube with focus tip is used to detect very small leaks, to accurately locate them. Just like the acoustic trumpet, the tube can be plugged into the preamplifier with ultrasonic receiver. The use of the camera is no longer possible.
The components are removed in the reverse order; for unlocking the preamplifier module, the release button must also be pressed.
Picture 8
Commissioning
LD500 V1.00 Seite 12 von 68
7 Commissioning / Applikation LD 500 or LD510
Please first observe the safety instructions in section 3
7.1 Switch on
Hold down the power button for about 1 second, the power will turn on, and a start-up sequence will appear on the display. Pressing the button again switches the device off again.
On-Off button, see device components and controls
7.2 Headphone Volume Loud / Volume Down
The volume keys increase or decrease the volume in the headphone in 16 levels. Continuously pressing the button automatically increases / decreases the value.
Volume up / down buttons for headphone volume, see device components and controls
Please make sure the headphone level is <50% before putting on the headphones.
7.3 Sensitivity level
When starting a leak detection or after switching on the sensitivity level "Auto" should be selected. In the case of strong noise levels from the environment it can be switched to a manually adjustable gain level, see chapter 9.1 „Setting of Sensitivity level“
Manual sensitivity level at measurement start: 30 – 90dB
Procedure leak detection / measurement
LD500 V1.00 Seite 13 von 68
7.4 Procedure leak detection / measurement
LD 500
Switch on
Headphone level
< 50 %
Reduce
Headphone level
Sensivity slevel
„Auto“
Yes
No
strong noise levels
existing?
Sensivity level to
manual
30 – 90dB
leak detection
start
Leakage found ?
Continue leak search ?
LD 500
Switch off
Yes
No
Yes
Yes
Yes
Data export to USB Stick
Save measurement
Leaktag Issue/ attach
Picture 9
Operation
LD500 V1.00 Seite 14 von 68
8 Operation
The operation is largely self-explanatory and menu-driven via the touch panel. The selection of the respective menu items occur via short "tapping" with the finger or a soft round pen.
Attention: Please use no pens or other objects with sharp edges! The foil can be damaged!
Inputs or changes can be made with all white deposit fields
8.1 Initialization
After switching on the LD 500, the initialization takes place and then switch to leakage display
Operation
LD500 V1.00 Seite 15 von 68
8.2 Screen Leckage
8.3 The following picture shows and describes the display elements.
Date / Time:
Battery condtion indicator
Battery condition:
22%
Power supply connected and battery is charging:
Displayed values for:
Signal-Levell in dB Leakage size Leakage costs per year
Laser „On“ Icon
Date/ Time and battery condition indicator
Bar graph of Signal level
LeakTag number
Actual picture from the camera
Laser „On/Off“ button
Setting button
Store button for
Measurement
Zero point adjustment
-button Button Sensivity level
Home Menu button
Operation
LD500 V1.00 Seite 16 von 68
9 Settings
The operation is largely self-explanatory and menu-driven via the touch panel. The selection of the respective menu items occur via short "tapping" with the finger or a soft round pen.
Attention: Please use no pens or other objects with sharp edges! The foil can be damaged!
9.1 Sensitivity settings
In order to cover a measuring range from the smallest leaks (0.1 l / min) up to large leaks, the LD 500 has different measuring sensitivity levels:
0 – 60dB for smallest leakages starting from 0.1 l/min 10 – 70dB 20 – 80dB 30 – 90dB
for bigger leakages up to max. 50 l/min
Remark: In case of high noise level to switch to next lower sensitivity level.
These steps can be selected manually by pressing the Sensitivity button.Levels are changed to the next level each time they are pressed.
The„Auto“ setting allows you to automatically switch to the preferred sensitivity level for the LD 500. However, this requires a min. measuring time of 2 seconds. Auto(1) correspond to 0–60dB , Auto (2) to 10-70dB, Auto(3) to 20-80dB and Auto(4) to 30-90dB.
9.2 Laser On/Off
The laser pointer can only be switched on or off via the laser on / off button in the display (not via the membrane keypad). When switched on, the display shows a laser warning symbol.
Please note the warnings for laser operation! V Avoid direct / indirect (via reflexion) irradiation of the eyes in humans and animals!
Laser „On“ Icon
Operation
LD500 V1.00 Seite 17 von 68
9.3 Parameters for measurement
To get correct readings, the parameters
• Existing pressure
• Distance to measuring point
are needed. By pressing the key „Parameter“ in the display, the parameter input window is opened.
Note: Distance refers to distance from measuring point to acoustic trumpet or
focus tube with focus tip. Appropriate input range is 1m to 6m for acoustic trumpet , for focus tube with focus tip fixed to 20cm.
Pressure input range is 0.3bar to max. 10bar, for vacuum leakages -0.1bar to -1bar.
The pressure entry can be made by selecting one of the 3 predefined values or via the text field. Max. Permissible pressure value is 10bar. In case of higher pressures, please insert 10bar)
For the distance input, there are 4 predefined values or a text field for the measurement­specific distance entry. Selection „20cm“ is reserved for the measurement with the focus tube with focus tip. Currently the max. Measuring distance is 6.00 m.
Operation
LD500 V1.00 Seite 18 von 68
9.4 Storing oft he measurement
To store the measurements please press either the button „Store on the foil keypad, see chapter
Device components and controls , or by button „Store in the display.
All data are stored on to the inetrnal SD card.
After pressing one of the two „Store keys, the corresponding information for the measuring point must be completed. The measuring point information of the last stored storage (company, building and location) is displayed, the numbering of the leaking day is increased by 1. e.g .:
If necessary, fill out the LeakTag-form and attach it to the measuring location.
Please use correct LeakTag­number.
Operation
LD500 V1.00 Seite 19 von 68
9.4.1 Measuring point designation / selection
Store  Meas. Point
9.4.2 Parameter of measurement (Re-Check)
Store  Parameter
At this point, it is again possible to check and correct the parameters Pressure" and Distance".
Changing the parameters gives new values for leakage and cost. Execution of the corrections see description chapter 9.3
All information about the measuring point can be changed by selecting the corresponding text field or the stored measuring points can be loaded from the internal database.
Then a menu opens with the available / saved entries. When selecting a saved value, select it (highlighted in green) and then take over with „OK“.
If a new entry is necessary, the input menu opens after pressing the „new“ button.
Input is accepted via „OK“.
This procedure is analogous to enter the information for company, building and location.
Using the „delete“ button, individual entries can be deleted too.
Operation
LD500 V1.00 Seite 20 von 68
9.4.3 Comment
Store  Textfield Comment
9.4.4 Storing measurement data to internal SD-card
Speichern  store
In addition to the details of the measuring point with company, building and location, it is possible to enter a comment (up to 32 characters).
To do this, select the text field „Comment“ and enter the comment.
Before final storage of the measurement on the internal SD card, a summary is created and the correctness is queried once more for safety.
Storage is done with the Yes key.
The No key returns to the previous menu.
Operation
LD500 V1.00 Seite 21 von 68
9.5 ZeroPoint adjustment
In high noise environments, an adjustment can be made to reduce it. This automatic process is carried out by pressing the „Adj. key in the display.
Adj. 
For this adjustment, the LD 500 should point to a quiet point and the adjustment should be started by pressing the „Automatic " button.
If a ZeroPoint adjustments was performed the following icon is displayed.
To withdraw the ZeroPoint Adjustment e.g. for a new measurement in a quieter location, please press button Default Value“.
Operation
LD500 V1.00 Seite 22 von 68
10 Basic settings menu LD 500
Home
10.1 Configuration
Home  Configuration
With the button „Home“ you access the basic menu of the LD 500.
Return to measurement by pressing
„Leckage“ button.
By selecting the national standard of „ISO“ or
„US“ you can store your production cost for „1000 Nm³or. „1000 SCF“. These inputs are
used as the basis for the cost calculation.
The basic costs are entered via the text box
„Cost / 1000 Nm³“ for „ISO“ or Cost / 1000 SCF“ for „US“.
The currency of the production costs can be stored as text in the text field. „Currancy“.
The inputs „Parameter“ and „Meas. Point“ Follow the same procedure as described in
chapter 9.3.
Acceptance of the values and return to the basic settings menu is done by pressing the „Home“. Button.
20.00
0.54
Operation
LD500 V1.00 Seite 23 von 68
10.2 Export/Import
With Export / Import, recorded "journal data" can be transferred to a USB stick, system settings and measuring points can be exported as well as imported.
Operation
LD500 V1.00 Seite 24 von 68
10.2.1 Export „Journal Data“
Export / Import  Export  Journal Data
Attention: With ERASE Journal Data“ all journal data are deleted.
With the help of the „Change -button you can set a period between Start“ and „End“.
Stored measurement data that lies within this period will be exported.
The selected date is always highlighted in green and the dates of the Sundays are - as in the calendar - red.
For days on which measurement data was recorded, the date numbers are visually exalted
If several measurements have been recorded on a date, they will appear after the date selection. Now you can easily select the desired recording. With „OK“. the start or end time is taken over.
Press the Export“- button to transfer the selected data to the USB stick In the example given, 3 measurements are exported.
With „ERASE Journal Data“ the Journal Database is deleted. For verification is still a security question.
With „Back“ you return to the main menu.
Operation
LD500 V1.00 Seite 25 von 68
10.2.2 Export of System settings
This feature is especially relevant to the version LD 510, here for storing the external sensor settings as well as e.g. display option for charts, sensor value etc.
Export / Import  Export  System settings
Here the definition of the storage location takes place .Selection for internal SD card with activation of key „SdCard“ or on USB stick with key „USB“. The selection of the desired folder is made by selecting and activating with „goto“ button.
If a new directory is required, this is done by pressing new File, this can be created by selecting new Directory
Saving a system file with a new name takes place analogously, then the key new File must be pressed
Entries are to be confirmed with „OK“.
With „Cancel“ you return to the previous menu.
Operation
LD500 V1.00 Seite 26 von 68
10.2.3 Import of system settings
Export / Import  Import System settings
Sequence of directory and file selection is analogous to file export. Selection of internal SD card with activation of key „SdCard“ or on USB stick with key „USB“.
The selection of the desired folder is made by selecting and activating with the „goto“ button, then select corresponding system file.
Selection to be confirmed with „OK“.
Since system-relevant changes are made here, a confirmation prompt is issued, which must be confirmed with „OK“.
Operation
LD500 V1.00 Seite 27 von 68
10.2.4 Export / Import Customer database
These functions allow the stored measuring point descriptions (companies, buildings and location) to be exported as an XML file or to be imported from another LD 500 exported database. That means it is also possible to create and import the database externally, but the prerequisite is the correct format of the XML file.
Export / Import  Export Customers Export / Import  Import Customers
As data changes are made during importing, a confirmation question needs to be confirmed with "„Yes“.
Operation
LD500 V1.00 Seite 28 von 68
10.3 View bitmaps
View Bitmaps  Select Screenshoot
This allows the stored pictures (measurement pictures) on the SD-Card or USB Stick to load and shown in the display again.
Please press button „Select Screenshot“ and select the required picture (bitmap).
The pictures are stored and organized in different directories
The directory structure is year / calendar week
Designation: BMyyCWxx yy = Year xx = calender week
The selection of the desired folder is made by selecting and activating with the „goto“button.
Select the desired image and then display with „OK“.
Settings
LD500 V1.00 Seite 29 von 68
10.4 Device Settings
The settings are all protected by a password!
Settings or changes are generally confirmed with OK!
Remark:
If you go back to main menu and then again one of the setting menus is called, you must enter the password again.
10.4.1 Password Settings
Settings Passwort Settings
Overview of the Settings
If an incorrect password is entered there appears
Enter password or New password repeat in red
font.
If you can’t remember the password, please use
Master password in order to enter a new password.
Remark:
The master password is supplied together with the instrument’s documentation.
Factory settings for password at the time of delivery: 0000 (4 times zero).
If required, the password can be changed in the
Password settings.
The new password must be entered two times in a row and in each case confirmed with OK
Settings
LD500 V1.00 Seite 30 von 68
10.4.2 Device Settings
Settings Device settings
10.4.2.1 Language
Settings Device settings Set language
Overview of Device settings
Here you can select one of 11 languages for the LD 500.
Settings
LD500 V1.00 Seite 31 von 68
10.4.2.2 Date & Time
Settings Device settings Date & Time
10.4.2.3 SD-Card
Settings Device settings SD-Card Reset Logger Database Settings Device settings SD-Card Erase SdCard
Settings Device settings SD-Card Test SdCard
By pushing the Time Zone description field and enter the correct UTC, you can set the correct time all over the world.
The summer and wintertime switchover is realized by pushing the Daylight Saving button.
By pressing Reset Logger Database all actual stored data on SD-Card will be blocked for use in LD500 / LD510. Nevertheless all data are still stored and available for external use only.
By pressing Erase SdCard all Data on the SD-Card will be deleted.
With activation of Test SdCard data are written and read to and from the SD-card.
The number of test cycles, as well as possible errors and error codes are display in the status line.
Press the Back button to returns to the device settings menu.
Settings
LD500 V1.00 Seite 32 von 68
10.4.2.4 System update
Settings Device settings System-Update
P1: V033 C1 I1
Update Kanäle
10.4.2.4.1 Check for Updates
Settings Device settings System-Update check USB-Stick for new Updates
P1: V033 C1 I1
Update Kanäle
If it is required to install an older software version, you have press the button „Force all“
Overview of System-Update-Functions.
f after pressing the button “Check USB Stick
for new Software updates” the following
messages appear in the window, is the LD 500 is not properly connected to the USB flash drive or there are no files available.
I If the LD 500 is correctly connected to the USB stick and there are new versions of the individual SW Parts, the new versions are marked in red.
The update is started by pressing the Update Selection“. button.
Settings
LD500 V1.00 Seite 33 von 68
10.4.2.4.2 Update Channels
Settings Device settings System-Update Update-Channels
If there is an update either for the internal and external channel (LD 510 only), it must be started separately
Important:
If the Reboot system button appears after the update, it must be pushed to restart the LD
500!
10.4.2.5 Factory Reset
10.4.2.5.1 Standard Einstellungen wiederherstellen
Settings Device settings System Reset to Defaults
10.4.2.5.2 Unique USB ID
For connections with the PC, a status and therefore a unique USB ID can be defined here. Relevant for simultaneous connection of several USB devices to the PC.
Update for Channels LD 500.
If needed with „Reboot System“ the LD 500 could be started(reboot) here
Bevor the settings are changed to the production default settings a safety prompt is displayed and must be confirmed by pressing the button „Yes“.
Settings
LD500 V1.00 Seite 34 von 68
10.4.2.6 Calibration of touchpanel
Settings Device settings calibrate touchscreen
If necessary, the touch-screen calibration can be changed here.
Push Calibrate and it appears, 1. left above,2. bottom right, 3. bottom left, 4.right above and 5. in the middle, a calibration cross that must be pushed consecutively.
If the calibration finished positive a message
“Calibration successful” appears and have to
be confirmed with OK.
Is this not the case, so you can repeat the calibration with the help of the Cancel and
Calibrate buttons.
Settings
LD500 V1.00 Seite 35 von 68
10.4.3 Set backlight brightness
Settings Set backlight
Remark:
At the first touch, the Backlight in our example is reset to 39%, after that a "normal" function operation is possible.
Important: If the Backlight dimming after button is not activated, then the Backlight stays permanently on, in the currently set brightness.
Here you adjust the desired Backlight (15-100%) of the display directly.
E.g. Backlight to 39 %
With the help of the Backlight dimming after button, after a definable time interval (here after 15 minutes), the Backlight can be reduced to the minimum.
As soon as the dimmed screen is operated again, the Backlight is committed automatically on the last set value before dimming.
To reduce the energy consumption (device runtime), you can switch off the display backlight by setting "Backlight off after".
15
Einstellungen
LD500 V1.00 Seite 36 von 68
10.4.4 Cleaning
Settings Cleaning
10.4.5 System-Status
E Settings System-Status
53.5 °C
11,74 V
5.01 V
53.5 °C 11,74 V
5.01 V
10.4.6 About LD 500/LD510
Settings about LD 500
05186002
Buy
Buy
Brief description of the Hardware and
Software Version, as well as the Serial Number of the LD 500.
Under options, you can buy four additional, different functions, if you have not done this by ordering.
This function can be used for cleaning the touch panel during running measurements.
If one minute is not enough time to clean, the process can be repeated at any time.
Is the cleaning faster finished, then you can push the to abort press long button (for one or two seconds) to cancel.
The menu item “System status” provides information about the power supply voltages and an operating hour counter.
Charging the batteries
LD500 V1.00 Seite 37 von 68
11 Charging the batteries
The battery is charged within the device. For this, the supplied plug-in power supply is connected to the built-in charging socket of the LD 500 and the 230V socket.
The LD 500 checks the charging status of the battery and starts the charging process automatically if necessary.
11.1 Protection of exhaustive discharge
Z To protect the Li-ION accumulator of exhaustive discharge the device is switching off automatically if a cell voltage of 6,4V will be reached.
Connect external power
Symbol „CHG“ is displayed during charging.
Scope of delivery
LD500 V1.00 Seite 38 von 68
12 LD 510
12.1 Selection External sensor
The use of an „external Sensors“ requires to switch to its mode.
Home Mode Externer Sensor
Home menu for external sensor connection
Settings
LD500 V1.00 Seite 39 von 68
12.2 Input signals ext. sensor LD510
12.3 Cable cross section
12.3.1 Sensor circuit points/Output signal:
AWG26, cable cross-sections: 0,14 mm2
Input signals
Current signal (0 – 20 mA / 4 – 20 mA) internal or external power supply
Measuring range
0 – 20 mA / 4 – 20 mA
Resolution
0,0001 mA
Accuracy
0,03 mA 0,05 %
Input resistance
50
Voltage signal (0 - 1V)
Measuring range
0 - 1 V
Resolution
0,05 mV
Accuracy
0,2 mV 0,05 %
Input resistance
100 k
Voltage signal (0 - 10 V / 30 V)
Measuring range
0 - 10 V/30 V
Resolution
0,5 mV
Accuracy
2 mV 0,05 %
Input resistance
1 M
RTD Pt100
Measuring range
-200 - 850 °C
Resolution
0,1 °C
Accuracy
0,2 °C at -100 - 400 °C 0,3 °C (further range)
RTD Pt1000
Measuring range
-200 - 850 °C
Resolution
0,1 °C
Accuracy
0,2 °C at -100 - 400 °C 0,3 °C ( further range )
Pulse Measuring range
minimal pulse length 100 μs frequency 0 - 1 kHz max. 30 VDC
Settings
LD500 V1.00 Seite 40 von 68
12.4 Connection diagrams of the different sensor types
12.4.1 Connector pin assignment for all sensors at PI 500
The interface connector to be used is a ODU Medi Snap 8 pin – Reference: K11M07-P08LFD0-6550
.
Available connection cables at CS-Instruments are: ODU with Open ends: Order no 0553 0501, cable length: 5 m.
Order no 0553 0502, cable length: 10 m.
ODU with M12 Connector: Order no 0553 0503, cable length: 5 m.
Extension cable (ODU/ODU): Order no 0553 0504, cable length: 10 m.
Connection scheme:
Connector housing
Plug insert
Collet
Clamping nut
View on welding pins of Medi Snap Connector
C a b l e l e n g h t 5 m / 1 0 m
1
wei ss whi te
2
bra un b r ow n
3
grü n gre en
4
gel b y el l ow
5
gra u g re y
6
ros a p ink
7
bla u b l ue
8
rot red
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
+ RS485
- RS485
SDI (CS-internal data transmission for all Dew point and Flow sensor FA/ VA 400)
ANALOG IN +
ANALOG IN –)
STROMQUELLE 500 µA
+VB, 24V DC Power supply for sensor
-VB, GND Sensor
White
Brown
Green
Yellow
Grey
Pink
Blue
Settings
LD500 V1.00 Seite 41 von 68
12.4.2 Connection for CS dew point- and consumption sensors, series FA/VA 5xx
2
1 3
+
-
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
4
Sensor
Weiss / White
Braun / Brown
Blau / Blue
Rot / Red
LD 510 FA 510 FA 500 VA 500 VA 520
The digital data transmission between LD 510 and the sensors FA 500/ FA 510 and VA 500/520 occurs via RS 485 (Modbus).
12.4.3 Connection with RS485
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
+
-
Sensor
+
-
Sensor with RS485 Interface
12.4.4 Three- and four-wire power supply 0 - 1/10/30 VDC
+
-
ß 0...1 / 10 / 30 Vdc
+
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
Sensor
Sensor with voltage output in 3-wire technology
+
-
ß 0...1 / 10 / 30 Vdc
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
Sensor
+
-
Sensor with voltage output in 4-wire technology
red
blue
white
brown
Yellow
Yellow
grey
blue
blue
red
red
Settings
LD500 V1.00 Seite 42 von 68
12.4.5 Analogue two-, three-, and four-wire current signal
Sensors with 4 - 20 mA-output in 2-wire technology
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
+
Sensor
ß 0 / 4 ...20mA
LD 510
Sensors with 4 - 20 mA output in 3-wire technology
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
+
-
Sensor
ß 0 / 4 ...20mA
LD 510
Sensors with 4 - 20 mA output in 4-wire technology
0 / 4 ...20mA
ß 0 / 4 ...20mA
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
Sensor
+
-
LD 510
Yellow
Yellow
Yellow
Grau
blue
blue
blue
red
red
Settings
LD500 V1.00 Seite 43 von 68
12.4.6 Two-, three- and four-wire connector pin assignments for PT100/PT1000/KTY81
Yellow
rd
pink
Sensor
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
2-wire PT100/PT1000/KTY81
Sensor
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
3-wire PT100/PT1000/KTY81
Sensor
I (500µA)
12345678
+ RS485
- RS485 SDI
Analog IN +
+VB 24Vdc
-VB GND
Analog IN -
4-wire PT100/1000/KTY81
Yellow
Yellow
grey
grey
Pink
Pink
red
red
Scope of delivery
LD500 V1.00 Seite 44 von 68
12.5 Dew Point Sensor FA 500 / FA 510 of type FA 5xx (RS 485 Modbus)
First step: choose an unused sensor digital channel
External sensor Settings Sensor settings C1
Second step: choose type FA 5xx
External sensor Settings Sensor settings C1 Type description field FA 5xx
External sensor Settings Sensor settings C1 text field “Name”
Now the Type FA 5xx is selected for the FA 5xx series and confirmed by pressing the OK button.
Input of a name, please enter the text field „Name“.
It is possible to enter a name with max. 24 characters.
Confirmation by pressing the OK-button.
The connection with the FA 5xx sensor is done after confirmation by pressing “OK”.
Settings
LD500 V1.00 Seite 45 von 68
12.5.1 Settings Dew point sensor FA 500 FA 510
12.5.1.1 Unit selection for temperature and humidity
External sensor Settings Sensor settings C1 arrow right (2.page)
12.5.1.2 Definition of the System pressure (relative pressure value)
External sensor Settings Sensor settings C1 arrow right (2.page)Pressure Setting
The system pressure is inderted by entering the values in the corresponding text field. The unit can be freely selected, selection menu is opened by pressing the corresponding button units
Confirm the settings by pressing the OK button.
Unit selection for temperature and humidity by pressing the button °C, °F,
g/m³ or mg/m³.
Confirm the settings by pressing the OK button.
Settings
LD500 V1.00 Seite 46 von 68
12.5.1.3 Definition of Reference pressure (absolute pressure value)
External sensor Settings Sensor settings C1 arrow right (2.page)Pressure Setting Text field Ref.Pressure
12.5.1.4 Calibration
External sensor Settings Sensor settings C1 arrow right (2.page) Calibration
26.45
Here, a one-point calibration can be performed.
For that purpose, please enter in the text box "Reference Value" the new correct dew point value.
Then by pressing the "Calibration" button taking over the inserted reference value.
Calibration can be put back to factory
Reference pressure is the pressure for that the dew point in relaxation will be back­calculated.
Default- Value is 1013 mbar (Atm. Pressure).
Confirm the settings by pressing the OK button.
Settings
LD500 V1.00 Seite 47 von 68
12.5.1.5 More Settings Analogue output 4-20mA
External sensor Settings Sensor settings C1 arrow right (2.page) More-Settings 4-20mA
22
<3.6
This menu allows the adjustment / assignment of the measurement value and the scaling of the analogue output.
Selection of the measurement value by selecting the appropriate measured value key in this example, DP °
C" for dew point ° Ctd.
In text fields "4mA" and "20mA" the appropriate scaling values are entered, here from -80 ° Ctd (4mA) to
-20 ° Ctd (20mA).
With "Error Val" is determined what is output in case of error at the analog output.
<3.6 Sensor error / System error 22 Sensor error / System error
4..20 Output according Namur (3.8mA – 20.5 mA)
< 4mA to 3.8 mA Measuring range under range >20mA to 20.5 mA Measuring range exceeding
Scope of delivery
LD500 V1.00 Seite 48 von 68
12.6 Flow sensor of type VA 5xx (RS 485 Modbus)
First step: choose an unused sensor digital channel
External sensor Settings Sensor settings C1
Second step: choose type VA 5xx
External sensor Settings Sensor settings C1 Type description field VA 5xx
External sensor Settings Sensor settings C1 Name description field
The connection with the VA 5xx sensor is done after confirmation by pressing “OK”.
Now the Type VA 5xx is selected for the VA 5xx series and confirmed by pressing the OK button.
Input of a name, please enter the text field „Name“.
It is possible to enter a name with max. 24 characters.
Settings
LD500 V1.00 Seite 49 von 68
12.6.1 Settings for Flow sensor VA 5xx
External senor Settings Sensor settings C1 arrow right (2.page)
12.6.1.1 Diameter settings (only for VA 500 or VA 550)
External senor Settings Sensor settings C1 arrow right (2.page) diameter description field External senor Settings Sensor settings C1 arrow right (2.page) diameter unit description field
Important: The inner diameter should be entered as precisely as possible, because otherwise the measurement results are not correct! There is no uniform standard for the tube inner diameter! (Please, inquire at the manufacturer or measure by your own!)
For each text field could be the either a value or a unit be set.
Settings by entering the text field and then input a value or select the unit for the appropriate field.
In case of VA 520 and VA 570 with integrated measuring section the diameter and diameter unit field are not access able.
Important: Only for VA 500 or VA 550 possible to change the inner diameter
Here the “inner diameter is set to
27.5mm.
Please confirm by pressing the OK button and go back with arrow left (1.page).
After pressing the Unit Text fields following units bare selectable.
Settings
LD500 V1.00 Seite 50 von 68
12.6.1.2 Gas Constant settings
External senor  Settings  Sensor settings  C1 arrow right (2.page)  Gas Constant description field
Attention:
Reference temperature and reference pressure (factory setting 20 °C, 1000 hPa): All volume flow values (m³/h) and consumption values indicated in the display are related to 20 °C, 1000 hPa (according to ISO 1217 intake condition) 0 °C and 1013 hPa (= standard cubic meter) can also be entered as a reference. Do not enter the operation pressure or the operation temperature under reference conditions!
All gases marked in blue and with (real) have been a real gas calibration curve stored in the sensor.
Select the gas you require and confirm selection by pressing OK button.
Settings
LD500 V1.00 Seite 51 von 68
12.6.1.3 Definition of the reference conditions
Here, the desired measured media reference conditions for pressure and temperature can be defined
External senor Settings Sensor settings C1 arrow right (2.page) Ref. Pressure description field External senor Settings Sensor settings C1 arrow right (2.page) Ref. Pressure Unit description field
External senor  Settings  Sensor settings  C1 arrow right (2.page)  Ref. Temp. description Field External senor Settings Sensor settings C1 arrow right (2.page) Ref. Temp. Unit description Field
12.6.1.4 Definition Unit of flow and velocity
External senor Settings Sensor settings C1 arrow right (2.page) Flow description Field External senor Settings Sensor settings C1 arrow right (2.page) Velocity description Field
Settings
LD500 V1.00 Seite 52 von 68
12.6.1.5 Definition consumption counter value and consumption unit
External senor Settings Sensor settings C1 arrow right (2.page) Count Val. description Field External senor Settings Sensor settings C1 arrow right (2.page) Count Val. Unit description Field
Remark:
After confirmation with OK, the font is black again and the values and settings are accepted
The sensor allows taking over a starting counter value. Inserting the value by entering the ”Count. Val.” text field.
In the Count. Val. Unit field different units could be used. Selection by activation of the “Count. Val. Unit” text field
In case the counter value unit will be changed only the consumption counter value will be recalculated to the appropriate unit.
Selection to confirm selection by pressing
OK button.
Settings
LD500 V1.00 Seite 53 von 68
12.6.1.6 Settings analogue output 4-20mA of VA 5xx
External senor Settings Sensor settings C1 More-Settings 4-20mA Ch1
This menu allows the adjustment / assignment of the measurement value and the scaling of the analogue output by pressing the”4-20mA Ch1” button.
Selection of the analogue output measurement value by activating the appropriate measured value key in this example, “Flow". Possible outputs are flow, velocity and temperature. In case of no use, please select “Off”.
The analogue output scaling have to possibilities, automatic scaling (default) and a manual scaling by the user. Auto scaling is based on the calibration settings, means 4mA is set to zero and the 20mA value is based on the max. settings here 900m³/h
A “manual scaling” needs an activation of the “scale
manual” button.
In text fields "4mA" and "20mA" the appropriate scaling values are entered, here from zero m³h (4mA) to 300 m³/h (20mA).
With "Error Val" it is determined what is the output in case of an error at the analogue output.
2 mA Sensor error / System error 22 mA Sensor error / System error
4..20 Output according Namur (3.8mA – 20.5 mA)
< 4mA to 3.8 mA Measuring range under range >20mA to 20.5 mA Measuring range exceeding
Inputs / changes to be confirmed with “OK” button. Return to main menu with “Back”.
Settings
LD500 V1.00 Seite 54 von 68
12.6.1.7 Settings Pulse / Alarm output of VA 5xx
External senor  Settings  Sensor settings  C1 More-Settings  Pulse / Alarm
External senor Settings Sensor settings C1 More-Settings Pulse
The pulse output of the VA 5xx could be set functionally as pulse output or alarm output.
Function to activate by pressing either the
“Pulse” or “Alarm” button.
In case of no use, please select “none”.
To set up the pulse first the unit and the measurement value have to be defined.
Unit selection by pressing “unit” button and choice one of the possible units “kg”,
“cf”, “ltr” or “m³”).
Pulse weight setting by entering the text field “Value”. Here with defined 1 pulse per m³ and with positive polarity.
With „Polarity“ the switching state could be defined. Pos. = 0 1 neg. 1 0
pos
neg
open
closed
Settings
LD500 V1.00 Seite 55 von 68
External senor Settings Sensor settings C1 More-Settings Alarm
In case of use the pulse output as alarm following definitions needs to be set:
Unit selection by pressing “unit” button and choice one of the possible units
cfm”, “ltr/s”, “m³/h”, “m/s” ”, “°F ”, “°C” ”, “kg/s or ”, “kg/min”).
Alarm value setting by entering the text fields “Value”.
The limits „Highor „Low“ defines when the alarm is activated, selecting by pressing the appropriate button High: Value over limit
Settings
LD500 V1.00 Seite 56 von 68
12.6.1.8 Settings ZeroPoint or Low Flow Cut off for VA 5xx
External senor Settings Sensor settings C1 More-Settings Zeropoint
2.045
2.045
With these function following adjustments for the sensor VA 5xx could be done:
Zeropoint:
When, without flow, the installed sensor shows already a flow value of > 0 m³/h herewith the zero point of the characteristic could be reset
Cutoff:
With the low-flow cut off activated, the flow below the defined "LowFlow Cut off" value will be displayed as 0 m³/h and not added to the consumption counter.
For Zero Point the text field “ ZeroPoint”
to enter and insert the displayed actual flow, here 2.045
Sensor Type Modbus
LD500 V1.00 Seite 57 von 68
12.7 Type Modbus
12.7.1 Selection and activation of Sensor-Type Modbus
First Step: First step: choose an unused sensor channel
External sensor Settings Sensor settings C1
Second step: choose type Modbus
External sensor Settings Sensor settings C1 Type description field Modbus
Third step: confirm with OK.
External sensor Settings Sensor settings C1 arrow right (2.page) Va use
12.7.1.1 Modbus Settings
External sensor Settings Sensor settings C1 arrow right (2.page) Modbus Settings ID -text field
Via Modbus, it is possible to read out up to 8 Register-Values (from Input or Holding Register) of the sensor.
Selection by the Register Tabs Va –Vh and activation by pressing of the corresponding Use button.
Please insert here the specified Modbus ID of the sensor, allowed values are 1 -247, (e.g. here Modbus
ID = 12)
For setting the Modbus ID on the sensor, please see sensor-datasheet. In addition in the menu are the serial transmission settings Baudrate, Stopbit, Paritybit and Timeout time to define. . In case that the LD 510 is the end of the RS485 bus system with activating Term- & Bias- button the required termination and biasing could be activated.
Confirmation by pressing OK button.
For resetting to the default values please press Set to
Default.
Sensor Type Modbus
LD500 V1.00 Seite 58 von 68
External sensor Settings Sensor settings C1 Reg. Address description field
Important: Required is the correct register-address. It should be noted that the register-number could be different to the register-address (Offset). For this, please consult the sensor data sheet.
External sensor Settings Sensor settings C1 Reg. Format description field
Supported Data types: Data Type: UI1(8b) = unsigned Integer => 0 - 255
I1 (8b) = signed integer => -128 - 127 UI2 (16b) = unsigned Integer => 0 - 65535 I2 (16b) = signed integer => -32768 - 32767 UI4 (32b) = unsigned Integer => 0 - 4294967295 I4 (32b) = signed integer => -2147483648 - 2147483647 R4 (32b) = floating point number Byte Order: The size of each Modbus-register is 2 Byte. For a 32 bit value two Modbusregister will be read out by
the DS500. Accordingly for a 16bit Value only one register is read. In the Modbus Specification, the sequence of the transmitted bytes is not defined clearly. To cover all
possible cases, the byte sequence in the DS500 is adjustable and must adapted to the respective sensor. Please consult here for the sensor datasheet.
e.g.: High byte before Low Byte, High Word before Low Word etc. Therefore the settings have to be made in accordance to the sensor data sheet.
The measurement values are kept in the registers of the sensor and can be addressed via Modbus and read by the PI 500 This requires setting the desired register addresses in the PI 500
Entering the register / data address is
With the buttons Input Register and
Holding Register the corresponding
Modbus-register type will be selected.
The number format and transmission order of each value needs to be defined by Data Type and Byte Order. Both have to be applied in correct combination.
Sensor Type Modbus
LD500 V1.00 Seite 59 von 68
Example:
Holding Register - UI1(8b) - Value: 18
Holding Register – UI4(32) - Value: 29235175522 AE41 5652
External sensor Settings Sensor settings C1 Unit- description field
Selection Register Type Holding Register, Data Type U1(8b) und Byte Order A / B
HByte LByte 18 => 00 12
Data Order 1. Byte 2. Byte A 00 12
Selection Register Type Holding Register, Data Type U1(32b) und Byte Order A-B-C-
D
HWord LWord HByte LByte HByte LByte 29235175522 => AE 41 56 52
By pressing the description field Unit the list with the available units appear
Please select the unit by pressing the respective button e.g. m³/h. For validation of the unit, please push the button OK To move through the list please press the button Page. In case the unit is not available it is possible to create a user defined unit. Therefore, please select one of the
User_X buttons.
Sensor Type Modbus
LD500 V1.00 Seite 60 von 68
External sensor Settings Sensor settings C1 Scale- description field
External sensor Settings Sensor settings C1 OK
The use of this factor allows adapting the output value by the same.
By default or value = 0 no scaling is applied and displayed in the field is don’t scale
By pressing the OK button, the inputs are confirmed and stored.
Data logger settings
LD500 V1.00 Seite 61 von 68
12.8 Data logger Settings
External sensor Settings Logger settings
Remark:
If more than 12 measurement data are recorded at the same time, the smallest possible time interval of the data logger is 2 seconds.
In addition, if more than 25 measurement data are recorded at the same time, the smallest possible time interval of the data logger is 5 seconds.
In the top row you can select the predefined Time intervals 1, 2, 5, 10, 15, 30, 60 and 120 seconds for recording.
A different, individual Time interval can be entered in the highlighted white description field right at the head, where the currently set Time interval is always displayed.
Remark:
The largest possible Time interval is 300 seconds.
Sensor Type Modbus
LD500 V1.00 Seite 62 von 68
External sensor Settings Logger settings force new Record File button
or
External sensor Settings Logger settings force new Record File button Comment description field
Main menu Settings Logger settings timed Start button
A new recording file will be created by pushing the force new record file button and a name or comment can be entered by the choice of the Comment description field.
Important:
If a new recording file should be created, the force new record file button
By pushing the timed Start button and then the date/time description field below, the date and the start time can be set for a data logger recording.
Remark:
If the start time is activated, it will automatically be set at the current time plus a minute.
Sensor Type Modbus
LD500 V1.00 Seite 63 von 68
External senor Settings Logger settings timed Stop button
External senor Settings Logger settings timed Start button/timed Stop button Date/Time description field
By pushing the timed Stop button and then the date/time description field below, the date and the stop time can be set for a data logger recording.
Remark:
If the stop time activated, it will automatically be set to the current time plus an hour.
After pushing the date/time description
field a window will appear where the
yellow marked area of the time or date can always be set and changed.
Sensor Type Modbus
LD500 V1.00 Seite 64 von 68
External senor Settings Logger settings timed Start button/timed Stop button Date/Time description field Cal button
External senor Settings Logger settings Start button
External senor Settings Logger settings Start button/Stop button
Important:
If a new recording file should be created, the force new record file button must be activated. Otherwise, the last applied recording file is used.
After the start and stop time activation and the created settings, the Start button will be pushed and the data logger is armed.
The data logger starts the recording at the set time!
The data logger can be started without activated time settings, use the Start and Stop buttons for activate and disable. Left below there will be shown how many values are recorded and how long there still can be recorded.
Remark:
With the Cal button the desired date can be easily select from the calendar.
Data logger settings
LD500 V1.00 Seite 65 von 68
13 Scope of delivery
The LD 500 is available either as a single unit or in a set. The set contains all the components and accessories that are protected in a a rugged and shock-resistant transport case.
.
In der nachfolgenden Tabelle sind die Komponenten mit ihren Bestellnummern aufgelistet.
Description
Order no.r
Set LD 500 consisting of:
0601 0105
LD 500 leak detector with acoustic trumpet, and integrated camera, 100 leak tags for marking the leakages on site
0560 0105 Sound-proof headset
0554 0104
Focus tube with focus tip
0530 0104
Battery charger(AC adapter plug)
0554 0009
Transportation case
0554 0106
Helix cable for connecting the ultrasonic sound sensor
0200 01402
Focus tube with focus tip (0530 0104)
Soundproof headset (0554 0104)
Ext. Power supply (0554 0009)
LD500 Leak meter (0560 0105)
Transport case (0554 0106)
Data logger settings
LD500 V1.00 Seite 66 von 68
Diese Erklärung beinhaltet keine Zusicherung von Eigenschaften. Die Sicherheitshinweise der mitgelieferten Produktdokumentation sind zu beachten.
Appendix
LD500 V1.00 Seite 67 von 68
Contact
LD500 V1.00 Seite 68 von 68
SALES OFFICE NORD
Am Oxer 28c D-24955 Harrislee DEUTSCHLAND Tel.: +49 (0) 461 80 71 50 - 0 Fax: +49 (0) 461 80 71 50 - 15
info@cs-instruments.com
www.cs-instruments.de
SALES OFFICE SOUTH
Zindelsteiner Straße 15 D-78052 VS-Tannheim DEUTSCHLAND Tel.: +49 (0) 7705 978 99-0 Fax: +49 (0) 7705 978 99-20
info@cs-instruments.com
www.cs-instruments.de
Loading...